TEL AFA1000 / 2 – Mk2, AFA1000 / 1 – Mk2 Operating And Instruction Manual

1
AFA 1000
FUME CUPBOARD MONITORS
Operating and Instruction Manual
Model AFA1000 / 1 – Mk2
Digital display
3 Relay inputs
3 Relay outputs
Com port
Used for alarm indication and monitoring on Fume Cupboards
Model AFA1000 / 2 – Mk2
Digital display
3 Relay inputs
3 Relay outputs
3 Pushbuttons
Com port
Used for alarm indication , monitoring and control on Fume Cupboards
Issue: Jan 08
2
Note :- Access to the Calibration and Configuration menus is password protected and is factory set. To access and or change the password contact the supplier for the Engineers Password and enter the Passwords in the Main Menu or alternatively use a Laptop connected to the Com port and the Upload/Download software provided
Velocity Bar Graph or Alarm Time Line
Velocity display m/sec or fpm
Control Pushbutton functions
Control Pushbuttons
LED indicators
Function and up/down buttons for Menu Configuration and Calibration. ENTER – also used as Mute button for audible alarm
OPERATOR DISPLAY PANEL
3
Connection details :-
Relay output 1
Rated 240V 2A AC1
Com port RS485
Relay input 1
PSU
Flying lead with plug connection for PCB 15 V DC
Flying lead with plug connection for PCB and SM6 airflow sensor 12V DC
Flying lead from Proxy switch with plug connection for PCB 12V DC
Sash High
Proximity switch
( Optional )
SM6
Airflow Sensor
R1
R2
R3
SENSOR SOCKET
0-10V
ECON
OUTPUT
15V DC POWER SUPPLY
INPUT
0-10V AUX.
OUTPUT
s/w
Sash Proximity
switch
OUTPUT RELAYS
COMMS
PORT
RELAY INPUTS
INPUT 1
INPUT 2
INPUT 3
Sash Micro
switch
REFER TO INSTALLATION
INSTRUCTIONS BEFORE CONNECTING
MADE IN ENGLAND
www.tel-uk.com
T
EL
WARNING
Relay output 2
Rated 240V 2A AC1
Relay output 3
Rated 240V 2A AC1
Relay input 2
Relay input 3
Plug – in power adapter for
mains socket.
(To suit local requirements )
Standard options :-
230V 50Hz -- UK 3-pin 220V 50Hz -- Euro 2-pin 120V 60Hz -- USA 2-pin
4
1.1 General Description
All systems comprise of the following components :-
1 – SM6 Airflow Sensor, 1 – AFA1000 Alarm unit, 1 – DC power supply
If the Sash Alarm System option is included there will also be a sash micro switch or proximity switch.
Operator Features
--- the alarm has the following operator features :-
Digital Display
The digital display is a back-lit, full graphic unit with a visual display of approx 56 x 27 mm. The display operates through the software allowing the generation of figures, wording and Icons.
The display shows the fume cupboard face velocity in m/sec or fpm when enabled or the alternative with no velocity reading but showing AIR FAIL / AIR SAFE as continuous display. All of the above are configurable via the alarm key pad.
An ‘ event time line ’ segmented into 20 x 3 minute segments will scroll across the display ( when enabled) .This takes the form of a graphical ‘ blip’ that will progress from the right hand side to the left hand side – representing events that have occurred during the past hour. The alternative to the event time line is a dynamic ‘ bar graph ’ representing the face velocity
The display shows a Horn icon ( with line through it ) when the audible alarm is in the Muted condition
Sash High – will be displayed when the Sash alarm is enabled and the sash is raised above the max safe working opening. This display will alternate on/off with the velocity reading.
Ext Alarm – will be displayed when the external alarm input is activated ( when enabled ) This display will alternate on/off with the velocity reading
Air Fail - will be displayed if the airflow is less than the Low air alarm point. This display will alternate on/off with the velocity reading
High Air - will be displayed if the airflow is more than the High air alarm point. This display will alternate on/off with the velocity reading
Set-back - will be displayed if the night set-back function is activated ( when enabled ) This display will alternate on/off with the velocity reading
Disabled - will be displayed if the alarm disable function is activated ( when enabled ) This display will alternate on/off with the velocity reading
5
LED Indicators
---- the alarm unit has three LED indicators :-
Red -- Alarm Amber -- Caution Green -- Safe
Audible Alarm sounder
-- the alarm has an audible alarm sounder with local or remote
Mute facility
Enter
--- the alarm has an Enter button -- this is multi-functional as follows :­ Press Enter momentarily when alarm is sounding will mute the alarm Press Enter for 5 secs will gain access to Calibration and Configuration
menus ( both menus password protected )
+ / - -- the alarm has + / - buttons that can be used to scroll through the calibration and
configuration menu or to select options or values
External Connections
-- the alarm unit will have the following connection points :-
Input 1 --- volt free relay input configurable for normally closed or normally open relays
This input can be configured as :­ Alarm disable
Night set-back External alarm Sash High High / Low
Input 2 --- volt free relay input configurable for normally closed or normally open relays This input can be configured as :­ Alarm disable
Night set-back External alarm Sash High High / Low
Input 3 --- volt free relay input configurable for normally closed or normally open relays This input can be configured as :­ Alarm disable
Night set-back External alarm Sash High High / Low
6
Output 1 --- volt free relay output configurable as normally closed or normally open relays.
Output 2 --- volt free relay output configurable as normally closed or normally open relays.
Output 3 --- volt free relay output configurable as normally closed or normally open relays.
Com Port --- RS 485 to enable connection to Laptop or PC for full diagnostics , logging or
setting up and for communications to building computer system ( BMS)
Power supply --- low voltage DC power supply
Airflow Sensor --- connection socket for the face velocity airflow sensor.
The following Pushbutton controls apply to the AFA1000 / 2 only :-
Pushbutton 1
-- This button can be configured to act as a Fan stop / start button for the
fume cupboard extract fan or as Night Set-back Override
Pushbutton 2
--This button can be configured to act as a Lights ON / OFF or UV Lights On/OFF for the fume cupboard or Pump stop / start button for a fume cupboard scrubber pump.
( Can be interlocked with Pushbutton 1 )
Pushbutton 3
-- This button can be configured to act as a Panic button or a Purge button.
Fan ON / OFF -- will be displayed if the Fan button is operated ( when enabled ).
Lights ON / OFF -- will be displayed if the Lights button is operated ( when enabled ). Pump ON / OFF -- will be displayed if the Pump button is operated ( when enabled ). UV Lights ON / OFF -- will be displayed if the UV Lights button is operated
( when enabled ). Purge ON / OFF -- will be displayed if the Purge button is operated ( when enabled ).
See Menu Block Diagram document for other specific operations and indications.