Tektronix Z Instruction Manual

IV l/X IN I U A L
Tektronix, Inc.
S.W. Millikan Way P. O. Box 500 Beaverton, Oregon Phone Ml 4-0161 Cables: Tektronix
Tektronix International A. G.
Terrassenweg 1A Zug, Switzerland PH.042-49192 Cable: Tekintag, Zug Switzerland Telex 53.574
070-251
All Tektronix instruments are warranted against defective materials and workman ship for one year. Tektronix transformers, manufactured in our own plant, are w ar
ranted for the life of the instrument.
Any questions w ith respect to the w a r ranty mentioned above should be taken up with your Tektronix Field Engineer.
Tektronix repair and replacement-part service is geared directly to the field, there fore all requests for repairs and replace ment parts should be directed to the Tek tronix Field Office or Representative in your
area. This procedure w ill assure you the fastest possible service. Please include the
instrument Type and Serial number with all requests for parts or service.
Specifications and price change priv ileges reserved.
Copyright © 1960 by Tektronix, Inc., Beaverton, Oregon. Printed in the United
States of America. All rights reserved. Con tents of this publication may not be repro duced in any form w ithout permission of the copyright owner.
CONTENTS
W arranty
Section 1
Specifications
Section 2
Operating Instructions
Section 3
Circuit Description
Section 4
Maintenance
Section 5
Calibration
Section 6
Accessories
Section 7
Parts List and Diagrams
® Type z
m
TYPE Z PLUG-IN UNIT
CALIBRATED DIFFERENTIAL CO MPACTO R
.1 0 0 0 CM. DYNAMIC SCALE LENGTH
/UUU V.f'l- " _
COMPARISON VOLTAGE -x
VOLTS/CM
(ATTE NUATION !
VAR. ATTEN
I PUSH TO j
0 I 5 C O N N E C T
SIGNAL j
ptSCO N N E d
signal
u n i a eh
ATTEN.
GAIN
ADJUST
b a ia n c e
'position
VOLTS/CM
{ A T T I N U A T I O N )
onto on.
PORTIA I
rlATEONIX,
Type Z
http://manoman.sqhill.com
O INI 1
SPECI F I CATION
General Information
The Type Z Plug-In Unit is a calibrated differential com
parator preamplifier designed for use in all Tektronix
Type 530-, 540-, 550-, or 580-Series Oscilloscopes. The
unit may be used for three different modes of operation:
(1) as a conventional plug-in preamplifier, (2) as a differential
input preamplifier, or (3) as a calibrated differential com- pa rator.
As a conventional preamplifier, the Z Unit alone has a
risetime of 24 nanoseconds and a maximum sensitivity of
0.05 volt per centimeter of deflection. Table 1-1 summar izes the risetimes and bandwidths available when the Z Unit is used in combination with various types of Tektronix
oscilloscopes.
TABLE 1-1
Z Unit and
Type.-
Risetime in
Nanoseconds*
Bandwidth, 3 db,
Megacycles/Seconcl*
532
70 5
531 or 535
39 9
536
40
9
531 A, 533, or 535A
35
10
541, 543, 545, 541 A, 545A, 551, or 555
27 13
581 or 585**
27 13
*For signals which do not overscan the graticule.
**Type 81 Adapter must be used.
In differential input mode of operation, the dynamic range of r'zlOO volts allows the application of common mode signals up to 100 volts to be applied to the unit with out attenuation. The common-mode rejection ratio of 40,000 to 1 at dc or low frequencies allows measurement
of differential signals less than 50 mv in amplitude on
7+c 100-volt common-mode signals.
As a calibrated differential comparator, the Z Unit has
an effective screen height of ih2000 cm at maximum sensi
tivity. Within the dynamic range of :J:100 volts, calibra'ed
±DC comparison voltages can be added differentially to the input waveform to permit a maximum of 0.005% or 5 mv per mm to be accurately resolved.
Vertical Deflection Factors
0.05 to 25 volts per centimeter in nine calibrated steps; also continuously variable (uncalibrated) between steps and up to 60 volts per centimeter.
Input Impedance
1 megohm :Ul % paralleled by 24 pf.
Maximum Allow able Combined DC and Peak AC Input
600 volts, ac-coupled.
Maximum Common-Mode Signal
r'rlOO volts at 0.05 volt per centimeter. Higher voltages
are permissible with larger vertical deflection factors.
Common-Mode Rejection Ratio
40,000 to 1 at 0.05 volt per centimeter with a 1-kc sine
wave, lower at other sensitivities and higher frequencies.
At 0.05 volt per centimeter, a 200-volt (p-p) 1-kc common
mode signal produces less than 1 mm of vertical deflection.
Comparison Voltages Available
Three voltage ranges are provided: 0 to r-l volt, 0 to
~10 volts, and 0 to : 100 volts. An accurate 10-turn potentiometer is used to select the comparison voltages over these ranges.
Comparison Voltage Regulator
A regulator circuit maintains comparison voltages es sentially constant and independent of normal power supply voltages supplied by the oscilloscope.
1-1
Specifications Type Z
Comparison Voltage Accuracy
Within 5 millivolts (0.5%) on the ±l-v o lt range. Within 20 millivolts (0.2%) on the ± 10-volt range. Within 150 millivolts (0.15%) on the ±100-volt range.
Maximum Trace Shift
2 mm due to input grid or gas current.
Comparison Voltage Drift
Less than 0.1 % per 100 hours operation.
Temperature compensated over normally expected tem perature range. Air filter in oscilloscope should be main tained clean, particularly when using the Z Unit.
10-Turn Potentiometer Linearity
0.05%
Measurement Resolution
Resolution accuracy, at 100-volts comparison voltage
0.005% Maximum resolution—5 millivolts per millimeter.
Transient Response and Permissible Signal Volt age Rate of Change
Rate of rise1 volt per 6 nsec, maximum. If this rate
is exceeded, grid current will flow in the input stages.
Rate of fall1 volt in 5 nsec, maximum. If this rate is
exceeded, the amplifier will momentarily cut off. If over
driven by a sufficiently fast pulse, the amplifier will run down" linearly at the above rate.
Because of the wide dynamic amplitude capabilities of the
Z Unit, transient response is a function of signal amplitude.
Mechanical Specifications
ConstructionAluminum-alloy chassis. Front panel—Photo-etched. Net weight—6 pounds.
Accessories
2Instruction Manuals
1-2
®I
A front-panel view of the Type Z Unit is shown in Fig. 2-1.
In addition, a brief functional description is given of the front-panel controls, input connectors, Securing Rod, and Mechanical Lock.
Connecting the Z Unit to the Oscilloscope
Connect the Z Unit to the associated oscilloscope by
inserting the unit into the plug-in compartment and tighten
ing the Securing Rod. With the Intensity control of the oscilloscope turned fully counterclockwise, switch on the oscilloscope power. Normally it is necessary to wait a few
minutes for the oscilloscope and plug-in to warm up and stabilize. Then turn up the intensity and set the oscilloscope triggering controls to produce a free-running sweep. Posi tion the trace on the screen using the POSITION control and the oscilloscope beam-position indicators.
Preliminary Operational Adjustments
After the Z Unit has warmed up and stabilized, check
its operation to see if adjustment of one or more of the following controls is necessary. Be sure that the oscilloscope
used in conjunction with the Z Unit is correctly calibrated in the vertical-deflection circuit, and that the calibrator output voltage is correct.
1. Amplifier DC Balance
If the trace cannot be centered on the screen when -he
POSITION control is set near midrange, the AMP DC BAL adjustment (see Fig. 2-2) must be checked. Need for this adjustment arises mostly when the Z Unit is transferred from one oscilloscope to another. To make the adjustment, set the POSITION control to midrange, remove the left side panel from the oscilloscope, and adjust the AMP DC
BAL control to position the trace behind the horizontal centerline of the graticule.
2. Differential Balance
Differential balance may be quickly checked by applying 100 volts of calibrator signal to both input connectors. Set both VOLTS/CM switches to .05, the AC-DC switches to DC, and the Mode switch to A-B DIFF. Ignoring the posit ve and negative spikes, adjust the DIFF. BAL. control to elimin-
SECTION 2
O PERAT I N G
I NSTRUCTI ONS
ate any square-wave response (that is, to obtain a straight-
line appearance of the trace).
3. Variable Attenuator Balance
Any vertical shift of the oscilloscope trace when the VAR. ATTEN. control is rotated, with no input signal applied, indicates need for adjustment of the variable attenuator
balance. This adjustment must be made in conjunction with the differential balance adjustment (step 2) due to interaction between the circuits. Repeat the two adjust ments until both are correct.
Set the Mode switch to A ONLY and remove any input signals. Adjust the VAR. ATTEN. BALANCE control to eliminate any vertical shift of the crt trace as the VAR. ATTEN. control is rotated. Check to see if the trace can be centered as described in step 1 (Amplifier DC Balance). If not, repeat step 1.
4. Gain
The gain adjustment should be checked periodically to assure correct vertical deflection factors. It should also be checked when the Z Unit is transferred from one oscilloscope
to another. The adjustment is made using the calibrator
signal from the oscilloscope.
Set the Mode switch to A ONLY and apply 200 milli volts of calibrator signal to the A input connector. Make sure the VAR. ATTEN. control is set fully clockwise. Set the A VOLTS/CM switch to .05 and rotate the GAIN AD JUST control for exactly 4 centimeters of vertical deflection.
Input Signal Connections
It is often possible to make signal connections to the Z Unit with short-length, unshielded test leads. This is par ticularly true for high-level, low-frequency signals. When such test leads are used, you must also use a ground con nection between the oscilloscope and the chassis of the signal source. (Note: Excessively long test leads may cause parasitic oscillations.)
In many applications, however, unshielded leads are unsatisfactory for making signal connections because of pickup resulting from magnetic fields. In such cases, shielded cables should be used. You must be sure that the ground conductors of the cables are connected to the chassis of both the oscilloscope and the signal source.
In high-frequency work it is usually necessary to terminate
signal sources and connecting cables in their characteristic
2-1
Operating Instructions Type Z
VAR, ATTEN. Varies the verti cal d e fle ction factors betwee n ranges o f the V OLTS/C M co n trols.
Po larity Selects the co mpa r ison voltag e p o la rity.
VOLTS/C M Selects the ca l d e fle ction factors.
verti-
A INPUT Connecto r fo r coupl ing input dc or ac vo lta g e s to pream plifier.
AC-DC Determines wheth e r Inpu t signals are dc couple d or ac coupled.
PUSH TO DISCONNECT NAL D isconnects signal
pream p lifier input.
DIF F. BAL. A d justs the am p lifier fo r maximum
fe re ntia l re jection ratio.
PUSH TO DISCONNECT SIG NAL Disconnects signal from pream p lifier input.
AC-DC Determ ines wheth er
inp u t signals are dc couple d or
ac coupled.
B INPUT Connector fo r coupl ing Inp u t dc or ac voltages to pream plifie r.
Range Selects the ap p rop ri ate co m pariso n volta g e ra nge.
Mechanic al lock Locks H eli-
dial when pressed dow n ward.
H e iid ial Adju s ts the com parison voltag e ov er the range selected by the Range switch.
VAR. ATTEN, BALANCE A d justed to preve n t v e rtical trace shift as the VAR. ATTEN con tro l Is ro tated.
POSITIO N Sets the vertical positio n of the trace on the oscilloscope screen..
VOLTS/CM Selects the ve rti ca l deflec tio n factors.
GA IN ADJUST Sets the de flection factors o f the p re am p lifie r to ag ree w ith the
fro nt panel m arkings.
Fig. 2 -1 . Functions of fro n t-p a ne l c o ntro ls, inp ut connectors. Securing Rod, and Mechanical Lock.
2-2
http://manoman.sqhill.com
Operating Instructions Type Z
Fig. 2 -2 . Left rear side view of the Z U n it showin g location of the AMP DC BAL adjustment.
impedances. Unterminated connections result in reflections in the cables and cause distortion of the displayed wave
forms.
When input signal connections are made, consider the effect of loading upon the signal source due to the input circuit of the Z Unit. The input resistance of the Z Unit is 1 megohm which is generally adequate to limit low-frequency loading to a negligible value. At high frequencies, however, the input capacitance of the Z Unit and the distributed capacitance of cables become important. Capacitive load ing at high frequencies may be sufficient to adversely
affect both the displayed waveform and the operation of
the signal source. Both capacitive and resistive loading can usually be limited to negligible values through use of attenuator probes.
Use of Probes
Attenuator probes reduce loading of the signal source. However, in addition to providing isolation of the oscillo scope from the signal source, an attenuator probe also de
creases the amplitude of the displayed waveform by the attenuation factor of the probe. When making amplitude
measurements with an attenuator probe, be sure to multi ply the observed amplitude by the attenuation of the probe. (Additional information concerning probe attenuation will be found under Differential Preamplifier Operation and At tenuator Test Point portions of this section of the manual.)
An adjustable capacitor in the probe body compensates
for variations in input capacitance from one instrument to
another. To assure the accuracy of pulse and transient meas urements, this adjustment should be checked frequently.
To make this adjustment, set the oscilloscope calibrator controls for a calibrator output signal of suitable amplitude.
Touch the probe tip to the calibrator-output connector and adjust the oscilloscope controls to display several cycles
of the waveform. If the probe cable is connected to the A input connector on the Z Unit, adjust the probe capacitor for flat tops on the calibrator square wave. If it is connected to the B input connector, adjust for a flat bottom on the square wave.
Conventional Preamplifier Operation
When the Z Unit is used for conventional preamplifier operation, the Mode switch should be placed in either the A ONLY or the B ONLY position. Input signals should then be connected to the corresponding input connector. Operation of the unit in the two positions is essentially the same except that signals applied to the B input connector are inverted on the display. Positive voltages produce an upward deflection when applied to the A input connector
and a downward deflection when applied to the B input connector (see Fig. 2-3).
Fig. 2 -3 . W ave form s a pplie d to the A Inpu t connector p roduce an up righ t d isp lay , wh ile w aveform s ap p lied to the B Inpu t are in verted.
The amount of vertical deflection produced by a signal
is determined by the settings of the VAR. ATTEN. control
and the VOLTS/CM switch. Calibrated deflection factors
indicated by the settings of the VOLTS/CM switch apply only when the VAR. ATTEN. control is set fully clockwise. Serious errors in display measurements may result if the setting of this control is inadvertently moved away from
the fully clockwise position.
The range of the VAR. ATTEN. control is approximately
2.5 to 1 to provide continuously variable (uncalibrated) vertical-deflection factors between calibrated settings of the VOLTS/CM switches.
Voltage measurements may be made directly from the
oscilloscope screen by noting the deflection factor on the
appropriate VOLTS/CM switch dial, and the amount of
deflection on the screen. Then multiply the deflection on the screen by the setting of the VOLTS/CM switch and the attenuation factor, if any, of the probe.
Placing the AC-DC switch in the AC position inserts a dc blocking capacitor in series with the input circuit. In the AC position, the input time constant is 0.1 second and the low-frequency response is 3 db at 2 cps. Thus some
attenuation exists even at 60 cps. Because the input dc signal may be suppressed by means of the calibrated com
parison voltage feature, there are few occasions where the
ac-coupling mode will be needed. Two principle occasions
are: (1) When it is desired to get a quick look at the ac component of a signal which has a large dc component
and (2) where there is a difference in dc levels of the two signals to be observed during differential mode of operation.
The tolerances of the input circuit time constants (at
3 db frequency) are nominally =h2%. When tighter
®T
2-3
http://manoman.sqhill.com
Operating Instructions Type Z
matching between A and B input circuits is necessary, one input dc blocking capacitor may be padded" with a small additional capacitance, generally less than 0.001 /ifd.
The PUSH TO DISCONNECT SIGNAL buttons allow the signal to be momentarily removed from the input without the bother of disconnecting the probe or coaxial input con nector. This provides an easy method for finding the base
line of zero voltage level on the crt. (When utmost ac curacy in measurement is required, trace deviation from exact zero due to gas or grid current must be considered.) The POSITION control may then be adjusted to make the zero level lie at any graticule mark desired.
Differential Preamplifier Operation
The primary purpose of differential operation is to elim inate undesirable common-mode signals. The term common- mode signal" is defined as that signal which is common to both inputs of a differential amplifier. It most commonly, but not necessarily, represents unwanted hum or noise.
This feature can be used, for example, to observe the signal across one circuit element while effectively eliminat ing the remainder of the circuit from the observations. Tnis is accomplished by connecting the signal at one end of the element to one input of the Z Unit and the signal at the other end of the element to the other input of the unit
Differential operation between the two inputs is obtained
when the Mode switch is in the A-B DIFF. position. Maximum
common-mode rejection is obtained when both input at tenuators are set to XI. Common-mode rejection is a func tion of frequency in practical amplifiers. It is 40,000 to 1 for dc common-mode signals in the Z Unit and remains near that value through audio frequencies, decreasing as the frequency increases.
The differential or common-mode rejection ratio of the Z Unit describes the ability of the unit to reject common mode signals. Common-mode rejection ratio is best defined
as the ratio of amplifier response to that part of the input
signal not common to both, with respect to the response
of the amplifier to any input signal which is common to
both inputs. It is defined numerically in the following ex
ample.
If an input signal consists of 100 volts (p-p) of 60-cps hum
and 0.1 volt of desired signal, the 100-volt hum would cause
-nr
----
;7
-------
times 100 volts or 2000 cm of deflection, and
.05 volts/cm the signal would cause an additional 2 cm of deflection.
If conventional preamplifier operation were used, the de
sired signal would be deflected off the screen and could not be observed. However, if differential operation is
used and common-mode rejection were 2000 to 2 or 1000 to 1, then the hum and desired signal would each produce 2 cm of deflection. The resulting combined wave-
i
r s r
i
/
/
la ) 100-volt, 6 0 ~ hum plus 0.1-volt, 1-kc square
wave. Vertical Sensitivity: 25 v/cm. Mode: A ONIY. Sweep Rate: 5 miliisec/cm.
7
r
7
, V
L
j
\
/
V
/
\
u.
V /
(c) 100-volt 60 hum only. Control settings similar
to (a) except Mode is B ONLY.
(b) Top 4-cm of waveform (a ) at vertical sensitivity
of 0.05 v/cm. Mode: A-Vc, decoupled.
V c S i + S O v .
Sweep Rate: 0.5 millisec/cm.
++++
4+B ++++ t+H
W+
H + +
+-H4 ++++
m+
++++
id ) 60 hum is differentially suppressed and only the
1-kc square wave is displayed. Vertical Sensitivity:
0.05 v/cm. Mode: A-B DIFF., dc-coupled. Sweep Rate:
0.5 millisec/cm.
2-4
Fig. 2 -4. Com m on-mode rejection b y the Z U nit.
Operating Instructions Type Z
form could be seen but not easily measured. With a common-mode rejection of 40,000 to 1, the hum would be reduced to 0.5 mm and the desired signal alone could be accurately observed and measured.
The preceding example is shown pictorially in the series of waveform photographs shown in Figs. 2-4(a), (c), and (d). A combined 100-volt hum and 0.1-volt square-wave signal is shown in Fig. 2-4(a). Although the square wave does not seem to be present, it can be made visible by increasing the sensitivity of the Z Unit to 0.05 volts/cm, which increases the effective height of the waveform to 2000 cm. A dc com parison voltage (using differential comparator operation explained later in this section) of approximately 50 volts is used to bring the top 4-cm portion of the combined signals into view, shown by the waveform photograph in Fig. 2-4(b). The 2-cm high square-wave signal, riding on top of the hum but not synchronized with it, causes the separated appearance of the hum signal. Fig. 2-4(c) shows the 100-volt (p-p) hum signal which, by itself, is applied to the B input connector. The resulting display using common mode rejection is shown in Fig. 2-4(d).
The following five operational notes provide helpful in formation to obtain optimum performance from the Z Unit using this mode of operation.
1. Large signal response of the Z Unit as a differential am plifier is shown in Table 2-1.
TABLE 2-1
Signal Handling Capabilities of the Z Unit
Attenua
tion
Maximum
Calibrated
Sensitivity
(volts/cm)
Minimum
Uncalibrated
Sensitivity
(volts/cm)
Maximum
Signal or
Common
Mode (volts)
Magni fication Factor
XI
.05 .12
100
2000X
X2
.1
.24
200
1000X
X5
.25 .6
500 400X
X10
.5
1.2 1000*
200X
X20
1
2.4 1000*
100X
X50
2.5
6
1000*
40X
XI00
5
12
1000*
20X
fox
X200
10
24
1000*
X500 25
60
1000*
5X
^Maximum continuous dc input level with AC-DC switch set to DC: 800
volts; momentary maximum dc level should not exceed 1000 volts.
2. Any difference in attenuation factors of the two attenua tors (VOLTS/CM switches) will decrease the differential capabilities of the Z Unit to the extent shown in Table 2-2.
3. Both AC-DC switches should be in the DC position if possible as explained under Conventional Preamplifier Operation.
4. Differential balance may decrease slightly as the VAR. ATTEN. control is adjusted away from the calibrated (clockwise) position.
5. Either input signal alone may be viewed without switch ing back to the A ONLY or B ONLY positions of the Mode switch by depressing the PUSH TO DISCONNECT SIGNAL button of the other input channel.
TABLE 2-2
Attenuator Performance
Attenuation
Attenuator
Accuracy, ± %
Input Resistance
Tolerance, ± %
XI 0
1
X2
1
1
X5
1.5 1
X10 and up 2
1
If conventional probe is used, the probe resistor tolerance of 1 % de creases the Z Unit attenuator accuracy. These inaccuracies can be ac curately measured as explained later under Attenuator Test Point.
Calibrated Differential Comparator Operation
When the Mode switch is in the A-Vc or Vc-B position, the Z Unit is a calibrated differential comparator or slide- back voltmeter. The calibrated comparison voltage, which has a range of 0 to 100 volts, may be added (differentially) to either input signal.
In this mode a calibrated dc voltage is internally applied to cancel out any unwanted dc component in the applied signal, thereby allowing accurate measurements of relative ly small ac signals riding on top of relatively large dc signals.
When the Mode switch is in the A-Vc position, the com parison voltage is applied internally to the amplifier input where the B signal is ordinarily applied during differential mode of operation. The switches and input connector in the B ONLY section of the front panel are not used. Signals applied to the B input connector will not be ob served.
In Vc-B position of the Mode switch, the comparison voltage is applied to the amplifier input where the A input signal is normally applied during differential mode of op eration. All switches and the input connector in the A ONLY section of the front panel are not used.
The dc comparison voltage is set by 3 controls: the COM PARISON VOLTAGE Polarity, Range, and Helidial. The Range control has ranges of 0 to 1, 0 to 10, and 0 to 100 volts. The Helidial varies the comparison voltages over this range and indicates the precise comparison voltage at any particular setting.
NOTE The regulator circuit in the Z Unit maintains con stant, accurate, comparison voltages as long as
the oscilloscope 150- and + 225-volt power supplies are in regulation and within their output voltage tolerance ratings. Be sure these and other regulated power supplies in the oscilloscope are
operating properly.
Differential comparator mode of operation may be used to make the following voltage measurements: (1) dc volt age measurements, (2) ac signal measurements superimposed on dc, and (3) high-amplitude ac signal measurements.
2-5
Operating Instructions Type Z
1. DC Voltage Measurements.
When the Z Unit is used to make any dc voltage measure ments, it is first necessary to establish a reference line on the screen of the oscilloscope. This line will usually be the horizontal centerline of the graticule. To establish the reference, set the COMPARISON VOLTAGE Polarity switch to 0 and press the appropriate PUSH TO DISCONNECT SIGNAL button to disconnect the input signal. (As noted
previously under Conventional Preamplifier Operation, slight trace deviation from exact zero must be taken into account to obtain best accuracy.) Use the POSITION control to set the oscilloscope trace at the centerline of the graticule.
To measure a dc voltage component of ±100 volts or less, apply the input signal to one of the connectors. For ex ample, suppose the signal is applied to the A input connector, then proceed as follows:
(a) Place the A VOLTS/CM switch to .05 and the Mode
switch to the A-Vc position.
(b) If the dc voltage component to be measured is posi tive, place the COMPARISON VOLTAGE Polarity switch to the + position. If it is between 10 and 100 volts, set the
COMPARISON VOLTAGE Range switch to 100 V.
(c) Rotate the COMPARISON VOLTAGE Helidial to bring the desired portion of the trace onto the screen. Set the trace exactly on the reference line with the Helidial.
(d) Recheck the reference as described in the first para graph.
(e) When the zero reference line and signal trace appeiar at the same place on the screen, the input voltage and the comparison voltage are equal.
2. AC Signal Measurements Superimposed on DC
Small ac signals superimposed on a dc component can be measured accurately by first using the comparison volt age to effectively eliminate the dc component. The ac signal can then be measured in the same manner as in con ventional preamplifier operation. The VAR. ATTEN. control
must be kept clockwise to obtain correct results.
3. High-Amplitude AC Signal Measurements.
High-amplitude ac signals, subject to the rise rate and fall rate limitations listed in the Specifications section, can also be measured with the Z Unit at maximum sensitivity. This type of measurement is very similar to dc measurements except that it is not necessary to establish a zero voltage reference line unless measurements of both ac and dc volt age levels are to be made. To measure the voltage differ ence between two points on the waveform, proceed as fol lows:
(a) Set the Helidial to zero and position one point on the waveform to the horizontal centerline of the graticule.
(b) Use the Helidial and Range controls to bring the other
desired point to the centerline.
(c) The voltage difference between the two points is
read from the Helidial and the setting of the Range switch.
AC-DC Voltage Measurements Exceeding ±1 0 0 Volts.
If ac, dc, or both ac and dc voltage components are greater than ±100 volts, the .05 settings of the VOLTS/CM switches cannot be used. It will be necessary to use a lower sensitivity in order to prevent overdriving the preamplifier and to prevent exceeding the comparison voltage available. To obtain the correct voltage measurement, use the multi
plication factor which appears below the volts per centi meter setting on the VOLTS/CM knob. The product of the multiplication factor times the comparison voltage used
is the input signal voltage.
ATTENUATOR TEST POINT
Two applications included here show how the ATTEN TEST PT connector, located on the left side of the Z Unit, can be used to make attenuation factor or ratio checks on probes used with the Z Unit and on the internal attenuators of the unit. The test point, when connected to a voltage divider network under test, forms a bridge circuit as shown in Fig. 2-5. The crt is the null indicator and the Helidial reading is used to determine the attenuation factor or ratio of the divider resistors.
A third application describes how the unit may be used to measure external resistors.
± 1 0 0 V
Fig. 2 -5 . S im plified diagra m o f the brid g e circuit fo rm ed durin g use of the ATTEN TEST PT connector.
2-6
Operating Instructions Type Z
Checking Attenuation Accuracy of VOLTS/CM Switches
The attenuation accuracy of the X2, X5, XI0, and X20 positions of the VOLTS/CM switches can be determined by utilizing the 100-volt comparison voltage available at the
ATTEN TEST PT connector. The following procedure ex
plains how to check the X2 position of the A VOLTS/CM switch. The other positions are checked in the same manner,
1. Set the A VOLTS/CM switch to X2, the AC-DC switch to DC, and the Mode switch to A-Vc.
2. Place the COMPARISON VOLTAGE Polarity switch to 0
and position a free-running trace to the center of the screen for reference,
3. Remove the left side panel from the oscilloscope. Connect a test lead from the A input connector to the ATTEN TEST PT connector (see Fig. 2-6).
Fig. 2 -6. Left side vie w of the Z U nit sh o w ing the lo ca tion o f the ATTEN TEST PT.
CAUTION
Use care in making the test lead connections to prevent shorting the 100 volts at the ATTEN TEST PT to ground. An accidental short circuit may damage the CAL. 4 adjustment (R7684).
4. Set the COMPARISON VOLTAGE Polarity switch to +
and the Range switch to 100 V.
5. Rotate the Helidial to return the trace to the reference level obtained in step 2. Use the vertical trace indicator lights to aid in returning the trace. In this example the Helidial should end up in the vicinity of 5.00 (or 50 volts).
6. Divide the Helidial reading into 10 to obtain the at
tenuation factor. (Assume the Helidial reading to be 5.02.
5.02 divided into 10 is 1,99 or an attenuation factor of XI.99.
The attenuation ratio is 1.99 to 1).
Checking Attenuation Accuracy of Probes
Passive attenuator probes can be checked for attenuation accuracy by a method similar to that used for determining the VOLTS/CM attenuation accuracy. Referring to Fig. 2-5,
divider resistor R1 is the 1-meg input resistor in the oscil
loscope, R2 is the resistor in the probe. The procedure which follows describes a method for checking a 10X attenuator probe.
1. Set the A VOLTS/CM switch to XI, the AC-DC switch to DC, and the Mode switch to A-Vc.
2. Place the COMPARISON VOLTAGE Polarity switch to 0
and position a free-running trace to the center of the screen
for reference.
3. Remove the left side panel from the oscilloscope [if not already removed). Connect the probe cable connector to
the A Input connector on the Z Unit and connect the probe
tip to the ATTEN TEST PT.
4. Set the COMPARISON VOLTAGE Polarity switch to -j-
and the Range switch to 100 V,
5. Rotate the Helidial to return the trace to the reference level obtained in step 2. In this example the Helidial should end up near a reading of 10 volts,
6. Divide the Helidial reading into 100 to obtain the at
tenuation factor of the probe.
Checking External Resistors
External resistors can be checked for value against the
1-megohm input resistance of the unit by using the bridge circuit of Fig, 2-5. In this application, R2 is the resistance to be measured and R1 is the internal 1-megohm resistor.
The following procedure can be used:
1. Set the A VOLTS/CM switch to XI, the AC-DC switch to
DC, and the Mode switch to A-Vc.
2. Place the COMPARISON VOLTAGE switches to 0 and
100 V. Position a free-running trace to the center of the screen for reference.
3. Remove the left side panel of the oscilloscope. Connect one end of R2 to the A input connector and the other end to the ATTEN TEST PT (see Fig. 2-6) by using a jumper wire.
4. Set the COMPARISON VOLTAGE Polarity switch to -f.
5. Adjust the Helidial to return the trace to the reference level obtained in step 2. Use the vertical trace indicator
lights to aid you in returning the trace.
6. Let H be the Helidial reading. Then use the folowing equation to find the value of R2.
R2 = R1 1 ^ - = ^ Megohms.
If R2 is small compared to 1 megohm, the Helidial read ing will be very close to 10,00 making it difficult to obtain an accurate measurement. When this occurs, R1 can be shunted by an accurate external resistor which will permit Helidial readings around 5.00. The only restriction on the shunting is that the total resistance to ground from the ATTEN TEST PT should not be less than 20 k. This limits the current drawn from the test point to a maximum of 5 ma. If the 1-megohm resistor is shunted, R1 in the equation will have to be changed to the new value of the 1-megohm and shunt resistors in parallel.
@1
2-7
http://manoman.sqhill.com
A
INPUT
t
B
INPUT
r
»
Loading...
+ 30 hidden pages