YBAC3
Calibration Kit
Instructions
Attaching the Calibration Unit
To properly attach the YBAC3 calibration unit to
the YBA250:
1. Gently push the Open, Short, or Load
connector straight onto the Test Port. DO
NOT turn or twist the body of the calibration
unit while pushing it onto the Test Port.
Hold the body of
the unit to keep it
from turning
Removing the Calibration Unit
To properly remove the YBAC3 calibration unit
from the YBA250:
1. Hold the body of the calibration unit still
while turning the nut to loosen it. Turn
ONLY the nut to remove the connector.
2. Gently pull the calibration unit straight away
from the RF input connector. DO NOT turn
or twist the body of the connector while
pulling it off the Test Port.
General Characteristics
Table 1: Physical characteristics
Connector
Impedance 50 Ω (nominal)
Frequency Range DC to 5.0 GHz
Recommended torque 0.7Nmto1.1Nm
Gauge 5.33 mm to 5.84 mm
Calibration 1 year cycle
Table 2: Precision Open characteristics
Reflection Coefficient
Type N Male
≥ --0.09 dB to 5.0 GHz
www.tektronix.com
*P071183800*
071-1838-00
CAUTION. The calibration unit is
exceptionally sensitive to mishandling. Use
the procedures that follow when attaching
and removing the calibration unit or you
will damage its connectors. If its precision
connectors become damaged, the
calibration unit cannot provide dependable
measurements.
2. To secure the connector, hold the body of the
connector still while turning the nut until the
nut is finger-tight. Turn ONLY the nut to
secure the connector.
Hold the body of
the unit to keep it
from turning
Tighten connector
by turning only the
connector nut
Table 3: Precision Short characteristics
Reflection Coefficient
Table 4: Precision Load characteristics
Frequency Range
Return Loss
≥ --0.09 dB to 5.0 GHz
DC to 5.0 GHz
≥ 42 dB from DC to 2.5 GHz
≥ 38 dB from 2.5 GHz to 5 GHz
Warranty Information
For warranty information, go to
www.tektronix.com, click Support, and then use
the provided links to search for your product’s
warranty.
Contacting Tektronix
Tektronix, Inc.
14200 SW Karl Braun Drive
PO Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and
technical support:
H In North America, call 1-- 800--833--9200.
H Worldwide, visit www.tektronix.com to find
contacts in your area.
Copyright E Tektronix, Inc. All rights reserved