Attaching the Calibration Unit
To properly attach the YBAC3 calibration unit to the YBA250:
1. Gently push the Open, Short, or Load connector straight onto the Test Port. DO NOT turn or twist the body of the calibration unit while pushing it onto the Test Port.
Hold the body of
the unit to keep it
from turning
Removing the Calibration Unit
To properly remove the YBAC3 calibration unit from the YBA250:
1. Hold the body of the calibration unit still while turning the nut to loosen it. Turn ONLY the nut to remove the connector.
2. Gently pull the calibration unit straight away from the RF input connector. DO NOT turn or twist the body of the connector while pulling it off the Test Port.
General Characteristics
Table 1: Physical characteristics
Connector
Impedance 50 (nominal)
Frequency Range DC to 5.0 GHz
Recommended torque 0.7Nmto1.1Nm
Gauge 5.33 mm to 5.84 mm
Calibration 1 year cycle
Table 2: Precision Open characteristics
Reflection Coefficient
Type N Male
--0.09 dB to 5.0 GHz
www.tektronix.com
*P071183800*
071-1838-00
CAUTION. The calibration unit is
exceptionally sensitive to mishandling. Use the procedures that follow when attaching and removing the calibration unit or you will damage its connectors. If its precision connectors become damaged, the calibration unit cannot provide dependable measurements.
2. To secure the connector, hold the body of the connector still while turning the nut until the nut is finger-tight. Turn ONLY the nut to secure the connector.
Hold the body of
the unit to keep it
from turning
Tighten connector by turning only the
connector nut
Table 3: Precision Short characteristics
Reflection Coefficient
Table 4: Precision Load characteristics
Frequency Range
Return Loss
--0.09 dB to 5.0 GHz
DC to 5.0 GHz
42 dB from DC to 2.5 GHz38 dB from 2.5 GHz to 5 GHz
Warranty Information
For warranty information, go to www.tektronix.com, click Support, and then use the provided links to search for your product’s warranty.
Contacting Tektronix
Tektronix, Inc. 14200 SW Karl Braun Drive PO Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
H In North America, call 1-- 800--833--9200.
H Worldwide, visit www.tektronix.com to find
contacts in your area.
Copyright E Tektronix, Inc. All rights reserved
Loading...