Tektronix TLS216 Data Sheet

TLS216
Features
Specs
Ordering Information
Logic Scope
TLS216
P6240
Signal Acquisition System
Sample Rate 2 GS/s Bandwidth 500 MHz Channels 16 Samplers 16 Sensitivity
50 mV to 10 V/div*
1
Position Range ±5 Div
Offset ±1 V from 1 to 99.5 mV/div
*1 Magnification used above 2 V/div range.
Vertical System
DC Gain Accuracy - ±1.5% after probe calibration. Vertical Resolution - 8-Bits (256 levels over 10.24 vertical
divisions), 11-Bits with averaging.
Analog Bandwidth Selection - 20 MHz and full. Input Coupling - DC. Input Impedance - 50 Ohm with coax adapter, 1 megaohm
with probe. Maximum Input Voltage - Max nondestructive input ±25 V
(DC + Peak AC). Dynamic Range - +15 V to -12 V with probe.
Time Base System
Time Bases Main, Delayed Time/Division Range 500 ps to 5 s/div Time Base Accuracy 0.01% over any interval <1ms Record Length 500 to 2,000 pts Pre-trigger Position 20% to 80% of record
Acquisition Modes
Sample - Sample data only. Envelope - Max/min values acquired over one or more
acquisitions. Average - Waveform averages selectable from 2 to 10,000.
Display
Mixed-mode Data Display - Logic Scope offers Analog, Timing and BusForm
TM
display formats that simplify recognition of digital timing faults and signal anomalies. BusForm display compresses many input signals into the least amount of display area. The acquired data can be displayed in any of the three display modes without re-acquiring.
Waveform Style - Dots, vectors, variable persistence selectable from 250 ms to 10 s, infinite persistence and intensified samples.
Color - Standard palettes and user definable colors for waveforms, text, graticules and cursors. Measurement text and cursor colors matched to waveform. Waveform collision areas highlighted with different color. Statistical waveform distribution shown with color grading through variable persistence.
Color Grading - With variable persistence selected, historical timing information is represented by temperature or spectral color scheme providing "z-axis" information about rapidly changing waveforms.
Graticules - Full, grid, cross hair, frame. Format - YT. Fit to Screen - Entire acquisition memory displayed on screen.
Triggering System
Triggers - Main, Delayed. Main Trigger Modes - Auto, Normal, Single. Delayed Trigger - Delayed by time and/or events. Time Delay Range - 16 ns to 250 s (time/div <10 µs); 15.1 ns
to 250 s (time/div <25 µs).
Events Delay Range - 1 to 9,999,999 events. AUX Trigger Input - TTL compatible; Max input voltage is
-1 V to +6 V.
Measurement System
The Logic Scope provides a comprehensive suite of automatic measurements specifically designed to improve efficiency when troubleshooting digital hardware. Twenty-eight on-board measurements, including setup time, hold time and skew, speed identification of common digital problems. Whereas traditional DSOs confine waveform measurements to establishing the relationship of a single channel to a clock, the Logic Scope simultaneously samples on all channels, so you can see the relationships between multiple channels.
Automatic Waveform Measurements
Period Frequency High Low + Width - Width Maximum Minimum Rise Fall Peak to Peak Amplitude + Duty Cycle - Duty Cycle + Overshoot - Overshoot Propagation Delay Burst Width Mean Cycle Mean rms Cycle rms Area Cycle Area Phase Setup Time Hold Time Skew
Continuous update of up to four measurements on any combination of waveforms.
Gated - Any region of the record may be isolated for measurement using vertical bars.
Snapshot - Performs all measurements on any one waveform, showing results from one instant in time.
Cursor Measurements - Absolute, Delta; Volts, Time, Frequency, Binary Readout.
Cursor Types - Horizontal bars (volts); Vertical bars (time); operated independently or in tracking mode.
Computer Interface
GPIB (IEEE 488.2) Programmability - Full talk/listen modes. Control of all modes, settings, and measurements.
Hard Copy/Desktop Publishing
Printer - HP ThinkJet, Epson, PostScript, DeskJet, LaserJet, DPU 411/412.
Export File Formats - EPS (Encapsulated PostScript), Interleaf, TIFF, PCX, BMP, RLE.
Plotter Support - HPGL. I/O Ports - GPIB, Centronics, RS-232 (Talk only). Floppy Disk Drive - 3.5 in., 1.44 MB DOS format. VGA Output - 15-Pin analog output (Color).
Storage
Waveforms - 16 full 2,000 pt waveforms. 119,008 500 pt waveforms on 1.44 MB floppy disk.
Setups - 10 front-panel setups. 78,947 setups on 1.44 MB floppy disk.
Trigger Types (Main and Delayed)
Edge Trigger when either a rising or
falling edge (positive or negative
slope) is detected.
Pulse The pulse may be positive or
negative and defined by a
duration.
Width (<Time, >Time) Trigger when the duration of
either a positive or negative pulse
is less than or greater than the
user-defined input.
Range (In Range, Out of Range)
Trigger when the duration of
either a positive or negative pulse
is either between the upper and
lower limits or outside the limits.
Time-out Trigger at the end of a
user-defined time period if the
trailing edge of either a positive
or negative pulse is not detected.
Pattern Each input may be defined as a
low (logic 0), high (logic 1), or
don't care (X). Inputs can be
ANDed or ORed. The pattern
may be positive true or positive
false.
Width (<Time, >Time) Trigger when the duration of a
user-defined logical pattern is
less than or greater than the
user-defined input.
Range (In Range, Out of Range)
Trigger when the duration of a
user-defined logical pattern is
either between the upper and
lower limits or outside the limits.
Time-out Trigger at the end of a
user-defined time period if the
user-defined logical pattern of a
specified duration does not
occur.
Sequence The sequence is defined by a set
of start and end events. The start
or end event can be either a
pattern or an edge (positive or
negative slope).
Width (<Time, >Time) Trigger when the duration
between the user-defined start
and end events is less than or
greater than the user-defined
input.
Range (In Range, Out of Range)
Trigger when the duration
between the user-defined start
and end events is either between
the upper and lower limits or
outside the limits.
Time-out Trigger at the end of a
user-defined time period if the
user-defined end event does not
occur within the specified time
after the start event.
State Trigger when a user-specified
logical pattern is detected and the
input defined as the clock goes
true or goes false. Each input
may be specified as a low (logic
0), high (logic 1), or don't care (X). Inputs can be ANDed or
ORed.
Power Requirements
Line Voltage Range - 90 to 250 V
RMS
.
Line Frequency - 47 to 63 Hz.
Power Consumption - 300 W max.
Environmental and Safety
Temperature - Operating: +4°C to +50°C. Nonoperating:
-22°C to +60°C. Humidity - Operating and Nonoperating: Operating to 80% at
or below 29°C, to 20% from +30°C to +50°C. Nonoperating to 90% at or below 41°C, to 5% from +41°C to 50°C.
Altitude - Operating: 15,000 ft., Nonoperating: 40,000 ft. Electromagnetic Compatibility - Meets 89/336/EEC. Safety - UL 3111, CSA 1010.1, EN61010-1, IEC61010-1.
Probes
Active Probes - 2.5 pF, 1 megaohm FET.
Physical Characteristics
Dimensions mm in.
Height with feet 236 9.3 without feet 193 7.6 Width with handle 445 17.5 Depth with front cover installed 432 17
Weight kg lbs.
Net approximately 12.3 27 Shipping approximately 20 44
Top of Page
Features
Specs
Ordering Information
Product(s) complies with IEEE Standard
488.1-1987, RS-232-C, and with Tektronix Standard Codes and Formats.
Tektronix Measurement products are manufactured in ISO registered facilities.
49A-10733-4p154, 06/1997, 08/10/2000
© Copyright Tektronix, Inc | webmaster@tek.com | Privacy Policy