Range (In Range, Out
of Range)
Trigger when the duration of a
user-defined logical pattern is
either between the upper and
lower limits or outside the limits.
Time-out Trigger at the end of a
user-defined time period if the
user-defined logical pattern of a
specified duration does not
occur.
Sequence The sequence is defined by a set
of start and end events. The start
or end event can be either a
pattern or an edge (positive or
negative slope).
Width (<Time, >Time) Trigger when the duration
between the user-defined start
and end events is less than or
greater than the user-defined
input.
Range (In Range, Out
of Range)
Trigger when the duration
between the user-defined start
and end events is either between
the upper and lower limits or
outside the limits.
Time-out Trigger at the end of a
user-defined time period if the
user-defined end event does not
occur within the specified time
after the start event.
State Trigger when a user-specified
logical pattern is detected and the
input defined as the clock goes
true or goes false. Each input
may be specified as a low (logic
0), high (logic 1), or don't care
(X). Inputs can be ANDed or
ORed.
Power Requirements
Line Voltage Range - 90 to 250 V
RMS
.
Line Frequency - 47 to 63 Hz.