xx
TLP058
FlexChannel
Instruct
ions
®
Logic Probe
Connecting the probe to your c ircuit
Attach the probe to the circuit using the connectors and adapters
shown on the back of these instructions. Select the best method
for your needs, and then go to Setting up the Probe.
Setting up the probe
The oscilloscope automatically detects and configures channels
when you connect a logic probe.
To set and view the digital channel parameters, add the channel
with the connected logic probe to the display. Double-tap
the Waveform badge for the logic probe channel to open the
configuration menu. Digital channel s ettings include threshold
voltage (default is 1.4 V), signal height (for a ll channels), and
channel labels.
Characteristic Description
Threshold voltage ± 40 V
Minimum detectable
pulse width
Maximum input
toggle rate
Maximum sample
rate
Digital to analog
trigger skew
Probe length
1ns
500 MHz
6.25 GS/s
5ns
1.0 m (3.28 ft)
Table 2: Environmental specifications
Characteristic Description
Temperature
Operating
Nonoperating
Humidity Noncondensing, and as limited by a maximum
Operating 5% to 90% relative humidity to +40 °C (104 °F)
Nonoperating
Altitude
Operating
Nonoperating
0 °C to +50 °C (+32 °F to +122 °F)
–40 °C to +71 °C (–40 °F to +160 °F)
Wet-Bulb temperature of +39 °C (+102 °F)
5% to 55% relative humidity +40 °C to +50 °C
(+104 °F to 122 °F)
5% to 90% relative humidity to +40 °C
(+104 °F)
5% to 39% relative humidity +40 °C to +60 °C
(+104 °F to 140 °F)
3,000 m (9,842 ft) maximum
12,000 m (39,370 ft) maximum
1
*P0713
51502*
071-3515-02
Product description
The TLP058 FlexChannel®logic probe connects all instruments
with FlexChannel
device under test (DUT). The probe contains 8 data channels.
All leads include a ground connection at the t
connect the probe leads separately to the target system, or group
the leads together using the probe tip holde
®
inputs to digital buses and signals on your
ip. You can
r.
Using the probe
See the oscilloscope documentation and Help topics for how to
display and trigger on digital logic signals and buses.
Functional check
Follow th
e instructions for connecting the probe to the
oscilloscope and setting up the probe. Connect your probe
leads to a
ctive logic signals on your DUT. You should see logic
activity immediately on all connected, active channels. If you
do not see
an active s ignal, use an analog probe to verify signal
activity on the logic signal points.
application
Typical
1. Use the logic probe to view digital signals on a system bus.
2. Use an analog probe to view analog waveform information
of a digital signal.
Accessories
The probe comes standard with a Logic Probe Accessory Kit
(Tektronix part number 020-3170-XX). See the illustration on
the following page. See the label in the accessories box lid for
information on the individual accessories.
Equipment Recycling. This symbol indicates that
this product complies with the applicable European
Union requirements according to Directives
2012/19/EU and 2006/66/EC on waste electrical
and electronic equipment (WEEE) and batteries.
For information about recycling options, check the
Tektronix Web site (www.tek.com/productrecycling).
Safety summary
Connect and disconnect
properly. Connect the probe output
to the measurement instrument before connecting the probe
to the circuit under tes
t. Disconnect the probe input and the
probe ground from the circuit under test before disconnecting
the probe from the measu
Observe all terminal ra
rement instrument.
tings. To avoid fire or shock hazard,
observe all ratings and markings on the product. Consult the
product manual for furt
her ratings information before making
connections to the product. Connect the probe reference lead to
earth ground only.
Do not operate without
covers.
Avoid exposed circuitry. Do not touch exposed connections
and components when po
Do not operate with sus
wer is present.
pected failures. If you suspect there is
damage to this product, have it inspected by qualified service
personnel.
Do not operate in wet/d
amp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces
Safety terms and symbo
clean and dry.
ls in this manual.
These terms may appear in this manual:
WARNING. Warning sta
tements identify conditions or
practices that could result in injury or loss of life.
CAUTION. Caution sta
tements identify c onditions or
practices that could result in damage to this product or
other property.
Symbols on the product. This symbol may appear on the
product:
Connecting the probe to the oscilloscope
1. Move the locking lever to the unlocked position then let
go to reset locking lever to the center position.
2. Insert the probe into a FlexChannel channel until fully
seated and the lock mechanism clicks.
3. Mov
e the locking lever to the locked position. The status
light should be a solid green.
E. It is normal for the logic probe Status light to flash
NOT
green when the oscilloscope is powering on, and will turn to a
ady green once the oscilloscope is fully running.
ste
NOTE. If the status LED continues flashing green, flashing
, or is a steady red, contact Tektronix Customer Support
red
for assistance.
4. To r
emove the probe, move and hold the locking lever at
the unlocked position and pull out the probe.
Specifications
Table 1: Electrical and mechanical specifications
Characteristic Description
Input channels 8 digital
tresistance
Inpu
Input capacitance 3.0 pF
Input signal swing
Minimum 400 mV p-p
Maximum
imum nondest–
Max
ructive input signal
100 kΩ ±1.0%
30 V p-p, ≤200 MHz (centered around
DC threshold voltage) at the probe tip
the
10 V p-p, ≥200 MHz (centered around
the DC threshold voltage) at the probe tip
p-p, ±42 V peak, ±50 V
30 V
Contacting Tektronix
Web site: www.tektronix
Phone: 1-800-833-9200
Address: Tektronix, Inc.
Email:
ranty information
DC
War
Department or name (if known)
14200 SW Karl Braun
Drive P.O. Box 500
Beaverton, OR 9
USA
techsupport@tektronix.com
.com
7077
For warranty information, go to ww w.tektronix.com/warranty.
x

xx
TLP058
FlexChannel
®
ロジック・プローブ
取扱説明書
2
15-02
071-35
製品の説明
TLP058型FlexChannel®ロジック・プローブは、FlexChannel®
入力を装備した計測器を被測定デバイス(DUT)のデジタル・
バス/信号に接続するのに使用します。TLP058ロジック・プ
ローブはどのFlexChannelオシロスコープ入力チャンネルに
も接続できます。
すべてのリードのチップにはグランド接続があります。プロー
ブ・リードは、ターゲット・システムに個別に接続したり、プロー
ブのチップ・ホルダを使用してグループ化したりすることがで
きます。
プローブと測定回路の接続
プローブを回路に接続するには、本書の末尾に記載されて
いるコネクタとアダプタを使用します。最適な方法を選んで、
次の「プローブのセットアップ」に進んでください。
プローブのセットアップ
ロジック・プローブを接続すると、オシロスコープが自動的に
チャンネルを検出して構成します。
デジタル・チャンネルのパラメータを設定したり表示するに
は、プローブが接
続されているチャンネルをディスプレイに追
加します。ロジック・プローブ・チャンネルの波形バッジをダブ
ルタップして、設
定メニューを開きます。デジタル・チャンネ
ルの設定には、スレッショルド電圧(デフォルトは1.4V)、信号
高さ(全チャンネ
ル)、チャンネル・ラベルがあります。
プローブの使用
デジタル・ロジック信号とバスの表示方法および始動方法に
ついては、オシロ
スコープのマニュアルとヘルプ・トピックを
参照してください。
機能チェック
プローブとオシロスコープの接続およびプローブの設定につ
いては、次の指示
に従ってください。プローブのリードをDUT
のアクティブなロジック信号に接続します。アクティブなチャ
ンネルにすべて接
続したらすぐにロジック状態を確認してく
ださい。アクティブな信号が表示されない場合は、アナログ・
プローブを使用し
てロジック信号ポイントの信号状態を確認
してください。
主な用途
1. ロジック・プローブは、システム・バスのデジタル信号の
観測に使用します。
2. アナログ・プローブはアナログ波形情報の確認に使用し
ます。
アクセサリ
このプロー
ブには、標準付属品としてロジック・プローブ・ア
クセサリ・キット(当社部品番号:020-3170-XX)がついていま
す。次のペ
ージの図をご覧ください。アクセサリの個別の情
報については、アクセサリの箱のふたに貼付されているラベ
ルをご覧く
ださい。
特性 概要
スレッショルド電圧
検出可能最小パルス
幅
最大入力トグル・レー
ト
最大サンプル・レート
DAトリガ・スキュ
プローブ長
±40 V
1ns
500 MHz
6.25 GS/s
5ns
ー
1.0 m(3.28フィート)
表 4: 環境仕様
特性 概要
温度
動作時
非動作時
湿度 結露なし、最高湿球温度39℃(+102 ゚F)
動作時
非動作時
高度
動作時
非動作時
0 ℃ ~+50 ℃(+32 °F~+122 °F)
–40 °C~+71 °C(–40 °F~+160 °F)
相対湿度5%~90%(+40 ℃(104°F)まで)
相対湿度5%~ 55%(+40 ℃~ +50 ℃(+
104 °F~122 °F))
相対湿度5%~90%(+40 ℃(+104 °F)ま
で)
相対湿度5%~39%(+40 ℃~60 ℃(+104
°F~140 °F))
最高3,000m(9,842フィート)
最高12,000 m(39,370 フィート)
機器のリサイクル:このマークは、本製品がWEEE
(廃棄電気・
令2012/19/EUおよび2006/66/ECに基づき、EU
の諸要件に準拠していることを示しています。リサ
イクル方法については、当社のWebサイトのサー
ビス・セクション(www.tek.com/productrecycling)
を参照してください。
電子機器)およびバッテリに関する指
安全にご使用いただくために
接続と切断は正しく行うこと。: 測定対象の回路にプローブ
を接続する前に、プローブ出力を測定機器に接続してくださ
い。測定機器からプローブを外す前に、測定対象の回路か
らプローブの入力とグランドを外してください。
すべての端子の定格に従うこと。: 火災や感電の危険を避
けるために、本製品のすべての定格とマーキングに従ってく
ださい。本製品に電源を接続する前に、定格の詳細につい
て、製品マニュアルを参照してください。プローブの基準リー
ドは、グランドにのみ接続してください。
カバーを外した状態で動作させないこと。:
露出した回路への接触は避けること。: 電源が投入されて
いるときに、露出した接続部分やコンポーネントに触れない
でください。
故障の疑いがあるときは使用しないこと。: 本製品に故障
の疑いがある場合、資格のあるサー ビス担当者に検査して
もらってください。
湿気の多いところでは動作させないこと。:
爆発性のガスがある場所では使用しないこと。:
製品の表面を清潔で乾燥した状態に保つこと。:
安全に関する用語と記号
このマニュアルでは次の用語を使用します。
警告: 人体や生命に危害をおよぼすおそれのある状態
や行為を示します。
注意: 本製品やその他の接続機器に損害を与えるおそ
れのある状態や行為を示します。
本製品の記号:: 本製品は以下の記号に注意してご使用く
ださい。
オシロスコープへのプローブの接続
1. ロック・レバーをアンロック位置に移動させ、ロック・レバー
が中央に戻るまで待ちます。
2. FlexChannelチャンネルにプローブを挿入し、プローブが
完全に挿入されてロック・メカニズムのカチッという音が
するまで押し込みます。
3. ロック
・レバーをロック位置に移動させます。ステータス・
ライトが緑色に点灯します。
ジック・プローブのステータス・ライトは正常であれば、
注: ロ
オシロスコープの電源が入ったときには緑色に点滅し、オシ
ープが完全に作動すると緑色の点灯に変わります。
ロスコ
注: ステータスLEDが緑色に点滅し続ける、赤色に点滅し
、または赤色に点灯する場合には、当社カスタマ・サ
続ける
ポートにご連絡ください。
4. プロー
ブを取り外すには、ロック・レバーをアンロック位
置に移動させたままプローブを引き抜きます。
x
仕様
表 3: 電気仕様と機械仕様
特性 概要
入力チャンネル数
入力抵抗
入力キャパシタンス
入力信号スイング
最小値
最大値
最大非破壊入力信
号
8デジタル
100 kΩ ±1.0%
3.0 pF
400 mV p-p
30 V p-p、200 MHz以下( プ ローブ・
チップで、DCスレッショルド電圧を中
心に)
10 V p-p、200 MHz以上( プ ローブ・
チップで、DCスレッショルド電圧を中
心に)
30 V p-p、±42Vピーク、±50 V
当社へのお問合せ
Webサイト:
電話番号:
住所:
電子メール・アド
レス:
www.tektronix.com
1-800-833-9200
Tektronix, Inc.
部署名または個人名(わかる場合)
14200 SW Karl Braun
ve P.O. Box 500
Dri
Beaverton, OR 97077
USA
techsupport@tektronix.com
保証
DC
詳細については、http://www.tektronix.com/warranty
保証の
にアクセスしてください。