Tektronix TLASWUP Field Install Instructions

Instructions
TLASWUP Software Installation Instructions
075-0860-03
This document applies to TLA System Software version 5.1.
The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
www.tektronix.com
*P075086003*
075086003
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patent s, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
H In North America, call 1-800-833-9200.
H Worldwide, visit www.tektronix.com to find contacts in your area.
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it.
To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals for warnings and cautions related to operating the system.
ToAvoidFireor
Personal Injury
Use Proper Power Cord. Use only the power cord specified for this product and certified for the country of use.
Connect and Disconnect Properly. Do not connect or disconnect probes or test leads while they are connected to a voltage source.
Ground the Product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
The inputs are not rated for connection to mains or Category II, III, or IV circuits.
Connect the probe reference lead to earth ground only.
Power Disconnect. The power cord disconnects the product from the power source. Do not block the power cord; it must remain accessible to the user at all times.
Do Not Operate Without Covers. Do not operate this product with covers or panels removed.
Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualified service personnel.
Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
TLASWUP Software Installation Instructions
1
General Safety Summary
Provide Proper Ventilation. Refer to the manual’s installation instructions for details on installing the product so it has proper ventilation.
Terms in this Manual
Symbols and Terms
on the Product
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
These terms may appear on the product:
H DANGER indicates an injury hazard immediately accessible as you read the
marking.
H WARNING indicates an injury hazard not immediately accessible as you
read the marking.
H CAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
CAUTION
Refer to Manual
Mains Disconnected
OFF (Power)
WARNING
High Voltage
Mains Connected
ON (Power)
2
Protective Ground
(Earth) Terminal
Standby
TLASWUP Software Installation Instructions
Earth Terminal
Chassis Ground
Service Safety Summary
Only qualified personnel should perform service procedures. Read this Service Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this product unless another person capable of rendering first aid and resuscitation is present.
Disconnect Power. To avoid electric shock, switch off the instrument power, then disconnect the power cord from the mains power.
Use Care When Servicing With Power On. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
To avoid electric shock, do not touch exposed connections.
TLASWUP Software Installation Instructions
3
Service Safety Summary
4
TLASWUP Software Installation Instructions
TLASWUP Description
The TLASWUP software upgrade kit provides instructions and software to upgrade your Tektronix Logic Analyzer to TLA application software, ver­sion 5.1. It also supports the upgrade to Version 1.8 Pattern Generator applica­tion software for those instruments with TLA7PG2 Pattern Generator modules. The TLA Application Software V5.1 and higher will only run on TLA instru­ments with Windows XP Professional or Windows 2000 Professional operating systems purchased from Tektronix. The software will also run on a PC with Microsoft Windows XP Professional or Windows 2000 Professional operating systems.
To prevent personal injury or damage to the instrument, consider the following requirements before attempting service:
H Read the General Safety Summary and Service Safety Summary found at the
beginning of this manual.
H The procedures in this manual should be performed by a qualified service
person.
Be sure to follow all warnings, cautions and notes.
This document supports Tektronix modification: 30825
Instruments
TLA7012 Portable Mainframe All serial numbers TLA7016 Benchtop Mainframe All serial numbers TLA7PC1 Benchtop Controller All serial numbers TLA5000 Series Logic Analyzers All serial numbers TLA5000B Series Logic Analyzers All serial numbers TLA715 Portable Mainframe All serial numbers TLA721 Benchtop Mainframe All serial numbers
NOTE. Unless otherwise noted all references to TLA5000 Series Logic Analyzers apply to both TLA5000B and TLA5000 instruments.
TLA System and PC Requirements
H Minimum 10 GB hard drive, 500 MB free space
H 256 MB RAM minimum, 512 MB recommended
TLASWUP Software Installation Instructions
5
TLASWUP Description
Minimum Tool and Equipment List
No special tools are required to install this software.
Parts List
Table 1: Parts list
Quantity Part number Description
1ea NS Tektronix Logic Analyzer Family Application Software V5.1
1ea NS Tektronix Logic Analyzer Family Product Documentation
1ea NS Nero OEM Suite CD
1ea 075--0860--03 TLASWUP Software Upgrade Kit Instructions
1ea NS CheckIt Utilities sticker
CD
1 sheet NS Software revision labels
4ea NS Kit labels: TLA7KUP, TLA5KUP, TLA7UP
NS - Not Saleable
6
TLASWUP Software Installation Instructions
Installation Instructions
The following instructions direct you through the process of upgrading the TLA application software on your TLA logic analyzer. These instructions apply only to instruments with TLA Application Software V4.0 or higher. To install the TLA application software on a PC, see Installing the TLA Application Software onaPCbeginning on page 19.
Contact your nearest Tektronix, Inc., Service Center or Tektronix Factory Service for installation assistance.
Upgrade Instruction Overview
The basic upgrade procedure is to manually remove the older software, update the drives, and reinstall the new software. The upgrade steps are listed in this document.
While performing the upgrade procedures, keep the following items in mind:
H Throughout this procedure you will see the term “log on as Administrator.”
The instrument is set up to automatically log on as Administrator (with no password) so you may not see the log in prompt. If the network setups have been changed on your instrument, make sure that you log on as Administra­tor or as a user who has administrator privileges. Failure to do so can prevent the software upgrade from completing successfully.
H After logging on as administrator, quit any applications.
H If your instrument does not display the file extensions, you need to change
your folder options to display the file extensions. (From the Control Panel, double-click Folder Options, select the View tab, and then deselect the “Hide file extensions for known file types” selection.)
H Use the Windows Add or Remove Programs Control Panel utility to remove
existing software.
H You will be asked to restart Windows after you install each software
package. If you do not restart Windows when prompted your software may not install properly and can cause unpredictable behavior.
H If Windows does not restart normally, press and hold the On/Standby button
for five or six seconds to force a shutdown.
H You may be asked to remove some software packages that may not be loaded
on your instrument. If this is the case, skip those instructions.
TLASWUP Software Installation Instructions
7
Installation Instructions
Remove the Old Software
Complete the following steps to remove old software from your logic analyzer. Skip any steps that mention software that isn’t installed on your logic analyzer.
NOTE. Refer to the release notes for a list of software version numbers to determine which software you may need to upgrade. The release notes are included on the TLA application software CD.
1. If you have not already done so, power on the logic analyzer and log on as
2. Exit the TLA application.
3. Open the Control Panel.
4. Double-click the Add or Remove Programs icon.
5. For each of the following software packages, select the software (the exact
Administrator.
names may vary depending on the software version), click the Change/Re­move button, and follow the on-screen instructions to remove the software:
H Adobe Acrobat Reader
H CheckIt Utilities (V6.5.5)
H Logitech Mouseware (V9.41.2 or earlier)
H Click’N Burn Pro
H NI-488.2
H NI-Max
H QA+Win32
H SnagIt
H Macromedia Flash Player
H Stomp RecordNow MAX
H Tektronix Pattern Generator
H Tektronix TLA Application
6. Windows will uninstall the selected application. If any messages appear
asking you for permission to remove unused shared files or any read-only files, select Yes to Al l.
7. After the software has been successfully removed, restart Windows.
8
TLASWUP Software Installation Instructions
Installation Instructions
Update the TLA Drivers (Required for TLA Application Software V4.4 or Earlier)
Complete the following steps to remove the old PCI-VXI Bridge driver and install the new one required by the TLA. Skip this section if your TLA Applica­tion Software is V5.0 or higher.
1. After rebooting the instrument, insert the second CD of the TLA Application
software set.
2. Open the Windows Control Panel.
3. Open the Device Manager in the Hardware page of the System Properties
dialog box.
4. Select PCI-VXI Bridge from the PciVxi listing.
5. Right-click to uninstall the device and then confirm that you want to delete
the device.
6. Restart the instrument.
After restarting, the instrument detects a new device and displays a dialog box.
7. Click Cancel to exit the dialog box.
8. Open the Device Manager in the Hardware page of the System Properties
dialog box.
9. Under the PciVxi listing, Right-click Other PCI Bridge Device and select Properties.
10. Select the Driver page and click Update Driver.
The Upgrade Device Driver Wizard displays.
11. Click Next.
12. In the Install Hardware Device Drivers page, select Search for a suitable driver for my device (recommended) andthenclickNext.
13. In the Locate Drive Files page, select Specify a location and click Next.
14. Click Browse and go to the directory that contains the new pci2tla.inf driver file. (D:\Drivers\your instrument\PCI-VXI Bridge\pci2tla.inf).
15. Select the pci2tla.inf driver and then click Open.
16. Click OK to install the driver.
17. In the Driver Files Search Results page, make sure that Install one of the other drivers checkbox is selected and then click Next.
TLASWUP Software Installation Instructions
9
Installation Instructions
18. Select the PCI to TLA Bridge driver and then click Next.
19. Click Finish to complete the installation.
Install Adobe Acrobat Reader
1. Log on as Administrator.
2. Install TLA V5.1 Application Software CD 1 in the CD drive.
3. Browse to the Documentation folder on the CD and double click AdbeRdr70_enu_full.exe.
4. Follow the on-screen instructions.
5. After the software has been successfully installed, restart Windows.
Install SnagIt
Install NI-GPIB
1. Log on as Administrator.
2. Browse to the SnagIt folder on the CD and double-click Setup.exe.
3. Follow the on-screen instructions and clear the check box for starting SnagIt
automatically. Use the remaining default choices provided by the installer.
4. After the software has been successfully installed, restart Windows.
1. Log on as Administrator.
2. Browse to the NI-GPIB-USB folder on the CD and run the Setup.exe
program.
3. Follow the on-screen instructions and note the items below:
a. When prompted, select the Typical installation option.
b. When the Add GPIB wizard appears, select GPIB-USB-B.
c. After restarting the instrument, the NI-488.2 Getting Started wizard
displays. This is not needed; select Do not show at Windows startup.
10
d. When you first connect the iView cable, the instrument will detect the
new hardware. Select Install the software automatically (Recom- mended).
TLASWUP Software Installation Instructions
Install the TLA Software
Installation Instructions
The TLA software uses a wizard to install the following software applications (if they are not already installed) for your instrument:
H TLA Application
H Pattern Generator Application
H Microsoft’s .NET Framework
H OpenChoice TekVisa
H TPI Client
H PPI Client
1. Log on as Administrator.
2. Browse to the TLA Application folder on the CD and double-click Setup.exe.
Install CheckIt Utilities
3. Follow any on-screen instructions. If you are asked for permission to overwrite any read-only files, select Yes to All.
4. After the software has been successfully installed, restart Windows.
1. Log on as Administrator.
2. Exit the TLA application.
3. Install TLA V5.1 Application Software CD 2 in the CD drive.
4. Browse to the CheckIt folder on the CD and double-click Start.exe.
5. Click Next to start the installation.
6. Read and accept the license agreement.
7. When prompted for the CheckIt serial number, use U7-999999.
8. Follow any on-screen instructions and use all of the default choices provided
by the installer.
9. Before clicking Finish, clear any check boxes that ask you to start the application or to view the Readme file.
10. After the software has been successfully installed, restart Windows.
TLASWUP Software Installation Instructions
11
Installation Instructions
Install Flash Player
1. Log on as Administrator.
2. Exit the TLA application.
3. Browse to the Macromedia folder on the CD and double-click flashplay­er6_winax.exe.
4. Follow the on-screen instructions.
5. After the software has been successfully installed, restart Windows.
Install Logitech Mouseware
1. Log on as Administrator.
2. Exit the TLA application.
3. Browse to the Mouseware folder on the CD and double-click Setup.exe.
4. Follow the on-screen instructions.
5. After the software has been successfully installed, restart Windows.
6. Follow the on-screen instructions in the Device Setup Wizard to set the
mouse to the factory recommended settings.
Install the Nero Suite 3 Software
The Nero OEM Suite software is available on a separate CD.
1. Log on as Administrator.
2. Exit the TLA application.
3. Insert the Nero CD into the CD drive.
4. Follow the on-screen instructions to install the software.
5. Enter the serial number when prompted (provided on the CD envelope).
6. Click Exit when done and then remove the CD.
7. Restart Windows.
12
TLASWUP Software Installation Instructions
Optional Software
Installation Instructions
Tektronix offers the following service-related software for running the perfor­mance verification software.
H TLA Logic Analyzer PV
H Pattern Generator PV
H TLACAL
You should only need to install this software if you are running performance verification procedures with the appropriate test fixtures.
1. Log on as Administrator.
2. Exit the TLA application.
3. Browse to the appropriate folder on the CD and double-click Setup.exe.
4. Follow the on-screen instructions.
Upgrade the Firmware
TLA7000 Series
Mainframes
5. After the software has been successfully installed, restart Windows.
The firmware upgrade procedures depend on your logic analyzer.
Complete the following steps to upgrade the instrument firmware (Refer to TLA700 Series Logic Analyzers and Modules on page 15 to upgrade module firmware):
CAUTION. You MUST perform these steps to ensure that your Mainframe firmware has been upgraded. Failure to do so will result in the instrument not functioning properly and the instrument may need to be repaired at the factory.
1. Log on as Administrator.
2. Exit the logic analyzer application.
3. Click Start All Programs Tektronix Logic Analyzer TLA Firmware
Loader.
4. Select your instrument from the Firmware Loader -- Connection dialog box
and click Connect.
5. Click Load in the upper portion of the Load Firmware dialog box to identify
or to load mainframe firmware.
TLASWUP Software Installation Instructions
13
Installation Instructions
6. Compare the version numbers in the two columns. If the Running version does not match the Required version, you MUST upgrade the firmware.
7. In the Frame column, click the Mainframe name that corresponds to the firmware that requires upgrading and then click Update Firmware.
8. The next dialog box will have the correct .upd file selected in the file name box. Click Open.
9. When the next window appears, click Start Flash Operation. The operation may take several minutes.
10. Follow the remaining on-screen instructions.
11. Repeat the above steps for each mainframe firmware that is out of date.
TLA5000 Series Logic
Analyzers
1. Log on to the instrument as Administrator.
2. Exit the logic analyzer application.
3. Click Start Programs Tektronix Logic Analyzer TLA Firmware
Loader.
4. Select your instrument from the Firmware Loader -- Connection dialog box and click Connect.
5. You may see a pop-up window reminding you to cycle the power on the instrument after completing the upgrade operation. Click Yes to continue.
6. Select the product from the Supported list box near the top of the window.
7. Select Load Firmware from the Execute menu.
8. If you have a TLA5000 Series Logic Analyzer click the TLA520x.lod file.
9. Click Open. You will be asked to confirm your action; click Yes when
prompted.
10. Exit the firmware loader program and power down the logic analyzer. You must power down the logic analyzer to allow the TLA application to start up properly.
14
TLASWUP Software Installation Instructions
Installation Instructions
TLA700 Series Logic
Analyzers and Modules
If you need to upgrade the firmware on your logic analyzer modules, a startup message appears telling you that you need to upgrade.
1. Note the modules that require new firmware.
a. For TLA7Axx/TLA7NAx series logic analyzer modules skip steps 2–8
andgodirectlytostep9onpage16.
b. For all other module types continue with step 2.
2. Turn off the logic analyzer and disconnect the power cord.
3. Remove the modules from the logic analyzer.
4. Refer to Figure 1 and locate the flash programming pins on the rear of each
module.
5. Install a jumper on the flash programming pins.
6. Reinstall the module(s) in the logic analyzer.
Figure 1: Flash programming pins
TLASWUP Software Installation Instructions
Flash programming pins
15
Installation Instructions
7. Reconnect the power cord and then power on the logic analyzer.
8. Log on as Administrator and then wait for the TLA application to start.
NOTE. Any modules with the flash programming jumper installed will not display in the System window.
9. Exit the TLA application.
10. Click Start Programs Tektronix Logic Analyzer TLA Firmware
Loader.
11. Select your instrument from the Firmware Loader -- Connection dialog box and click Connect.
12. Select the modules that you want to update from the list of modules displayed in the Supported list box near the top of the window. This selects the modules for the upgrade. If you are upgrading more than one module at a time, note the slot location of the module in the selection box.
13. Select Load Firmware from the Execute menu.
14. Click the proper .lod file for the selected module; refer to Table 2 if
necessary.
NOTE. Be sure to select the proper .lod file for each module. Note the slot number in the title bar so that you select the correct module for the .lod file.
Table 2: TLA firmware files
Module type .lod file
TLA7NAx logic analyzer modules TLA7NAX.lod
TLA7Axx logic analyzer modules TLA7AXX.lod
TLA7Nx, TLA7Px, TLA7Qx logic analyzer modules TLA7NPQ.lod
TLA7Lx and TLA7Mx logic analyzer modules TLA7LM.lod
TLA7Dx and TLA7Ex DSO modules FirmwareDSO.lod
TLA7PG2 Pattern Generator modules
1
The .lod files for the pattern generator modules are located in the C:\Program Files\Tektronix Pattern Generator\Firmware folder. All other files are located in the C:\Program files\TLA 700\Firmware folder.
1
tla7pg2.lod
16
15. Click OK. When prompted to confirm your action; click Yes.
TLASWUP Software Installation Instructions
Installation Instructions
NOTE. The program will not allow you to load firmware to an incompatible module. For example, the program will not load DSO firmware to a logic analyzer module.
The program will load the firmware for each module one at a time. The process may take several minutes per module.
16. When the upgrade is complete, exit the program.
17. For TLA7Axx/TLA7NAx series logic analyzer modules skip the remaining
steps in this section and continue with Verify Proper Operation on page 17. For all other module types continue with step 18.
18. Turn off the logic analyzer and disconnect the power cord.
19. Remove the module from the instrument and remove the jumper from the
Flash programming pins on the rear of the module. Keep the jumper for future updates.
Verify Proper Operation
WARNING. Dangerous voltages may be present when the instrument is connected to the Target system. To avoid electrical shock, always tighten the hold-down screws that secure the module to the mainframe.
20. Reinstall the module in the logic analyzer, latch it in place, and tighten the
hold-down screws.
21. Reconnect the power cord and power on the instrument.
22. Verify that the power-on diagnostics pass.
After you have installed all of the latest software and upgraded the module firmware, you are ready to verify operation.
1. If you have not already done so, power on the logic analyzer and log on as
Administrator.
2. Verify that the Power-on diagnostics pass (if any Power-on diagnostics fail,
the Power-on Diagnostics dialog box automatically displays in the System window).
3. In the System window, verify that all installed modules appear in the
window.
TLASWUP Software Installation Instructions
17
Installation Instructions
4. If you have a TLA7000 series or TLA700 series instrument with any missing modules in the System window, you can verify the status of the modules by doing the following steps:
a. Select System Properties in the System menu.
b. Click the Modules tab.
c. Note the information under the Messages column for any installed
modules. If any modules indicate invalid or similar messages, you must update the firmware for those modules. (See Upgrade the Firmware on page 13).
d. If any of the modules indicate that the jumper is installed, power down
the logic analyzer, remove the module, and then remove the jumper from the rear of the module. Reinstall the module in the instrument and try again.
Calibrate the Logic Analyzer and DSO Modules
After verifying proper operation it is recommended that you calibrate the logic analyzer and DSO modules if they are installed in your instrument (pattern generator modules do not require calibration).
1. Allow the instrument to warm up for at least 30 minutes before continuing.
2. In the System window, select Calibration and Diagnostics from the System
menu.
3. Click the Self Calibration tab.
4. Select all of the modules in the dialog box and Click Run.
NOTE. The calibration process may take several minutes for each module.
5. Verify that the Status column for each module changes to Calibrated.
6. After all of the modules have been calibrated close the dialog box and power
down the instrument.
18
TLASWUP Software Installation Instructions
Installing the TLA Application Software on a PC
You can install the TLA application software on a PC for the following purposes:
H To run the TLA application software in the Offline mode (similar to TLAVu
in earlier software versions).
H To control a logic analyzer with the same TLA application software version
over a network.
Installation Instructions
Install Third-Party
Software
To use all of the features of the TLA application software on your PC, you must install additional third-party software. The following third-party software is available on the TLA Application Software CD V5.1, or higher:
H The NI-GPIB-USB software allows you to use the iView software with your
PC.
H The SnagIt software is useful for copying screen shots of logic analyzer data
for use with other applications.
NOTE. Several third-party software applications are included to use with your logic analyzer. These applications may include software license agreements. Be sure to abide by those license agreements.
Install the NI-GPIB Software. Complete the following steps to install the NI-GPIB software on your PC:
1. Install the TLA Application Software CD in the CD drive.
2. Browse to the NI-GPIB-USB folder on the CD and run the Setup.exe
program.
3. Follow the on-screen instructions and note the items below:
a. When prompted, select the Typical installation option.
b. When the Add GPIB wizard appears, select GPIB-USB-HS.
c. After restarting the instrument, the NI-488.2 Getting Started wizard
displays. This is not needed; select Do not show at Windows startup.
d. When you first connect the iView cable, the instrument will detect the
new hardware. Select Install the software automatically (Recom­mended).
TLASWUP Software Installation Instructions
19
Installation Instructions
Install the SnagIt Software. Complete the following steps to install the SnagIt software on your PC:
1. Install the TLA Application Software CD in the CD drive.
2. Browse to the SnagIt folder on the CD and run the Setup.exe program.
3. Follow the on-screen instructions.
Install the TLA Application
Software
Start the TLA Application
Browse to the TLA Application software folder on the CD and run the Setup.exe program. Follow the on-screen instructions to install the TLA application software.
If you have an older version of the software on the hard disk, the installation program will detect it and ask if you want to remove it. Follow the on-screen instructions to remove the software, answering “Yes” to any prompts. Restart the instrument when prompted and run the Setup.exe program again.
If you have installed the TLA application software on a PC for remote operation, start the software by double-clicking on the TLA Application icon. The TLA Connection dialog box displays.
1. Select an instrument in the TLA Connection dialog box and then click Connect. If your instrument is connected to a network, you can connect to
any unused instrument on your network.
2. Click Offline to start an offline version of the TLA application software. The TLA Offline software allows you to run the TLA application without connecting to an instrument. You can analyze previously acquired data from a logic analyzer, create or modify reference memories, or perform system tests without being connected to an instrument.
Attach the Upgrade Kit Label
After completing all the previous steps, you need to install the upgrade kit label to individual modules and on the instrument to indicate that the kit is installed.
NOTE. If there already is an upgrade kit label installed, install the new label just above or below the old one. However, ensure that you install the software revision label over the old revision label to indicate the current version of software or firmware on your instrument.
20
TLASWUP Software Installation Instructions
Installation Instructions
TLA5000 Series
Instruments
TLA700 or TLA7000
Portable Mainframes
Locate the big label on the rear of the instrument, and then attach the software revision label and the kit label as shown in Figure 2.
TLA5202
Place revision label here.
Kit label
Figure 2: TLA5000 revision and kit label locations
Locate the big label on the bottom of the instrument, and then attach the software revision label and the kit label as shown in Figure 3.
Place revision
label here.
Kit label
Figure 3: TLA700 or TLA7000 revision and kit label locations
Benchtop Mainframes and
Instrument Modules
On the Benchtop Mainframes and instrument modules, the software label is located on the side of the controller or module. You must remove the modules from the mainframe before you can install the software revision label.
TLASWUP Software Installation Instructions
21
Installation Instructions
CAUTION. To avoid damaging the mainframe or the modules, always power down the mainframe before removing any modules.
1. Power down the mainframe.
2. Loosen the four hold-down screws that secure the front of the module to the
mainframe.
3. Remove the module from the mainframe and set it on a clean working surface.
4. Locate the big label on the side of the module and attach the software revision label and the kit label as shown in Figure 4. Repeat the process for each instrument module where you updated the firmware.
Place revision
label here.
Kit label
Figure 4: TLA700 or TLA7000 revision and kit label locations
WARNING. Dangerous voltages may be present when the instrument is connected to the Target system. To avoid electrical shock, always tighten the hold-down screws that secure the module to the mainframe.
5. Reinstall the modules in the benchtop mainframe, tighten the hold-down screws.
6. Power on the mainframe and verify that all diagnostics pass.
g End of document g
22
TLASWUP Software Installation Instructions
Loading...