Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 5 0 0
Beaverto
USA
For product information, sales, service, and technical support:
n, OR 97077
In North America, call 1-800-833-9200.
Worl dwid e, visi t www.tektronix.com to find contacts in your area.
14: Data placement................................... .................................. ....................10
iiTLA7SA08 & TLA7SA16 Series Product Specifications & Performance Verification
Preface
Related Documentation
This document provides the specifications of the TLA7SA16 and TLA7SA08
Logic Protocol Analyzer Modules for the TLA7012 and TLA7016 mainframes
and high-lev
el procedures to verify that the products are functioning correctly.
The follow
ing table lists related documentation that is available for your Tektronix
logic analyzer family product. The documentation is available on the TLA
Documentation CD included with your instrument and on the Tektronix Web
site (www.tektronix.com). Refer to the Tektronix Web site for the most current
documentation.
To obtain documentation that is not specified in the table, contact your local
Tektronix representative.
Related documentation
ItemPurpose
TLA Quick Start User Manuals
Online Help
Installation Reference SheetsHigh-level installation information
Installation Manuals
XYZs of Logic Analyzers
Declassification and Securities instructionsData security concerns specific to sanitizing
Optional Service ManualsSelf-service documentation for modules and
High-level operational overview
In-depth operation and UI help
Detailed first-time installation information
Logic analyzer basics
or removing memory devices from Tektronix
products
Collection of logic analyzer application
specific notes
TLA Product specifications and performance
verification procedures
Detailed information for controlling the logic
analyzer using .NET
Upgrade information for your logic analyzer
mainframes
TLA7SA08 & TLA7SA16 Series Product Specifications & Performance Verificationiii
Preface
ivTLA7SA08 & TLA7SA16 Series Product Specifications & Performance Verification
Specifications and Characteristics
All specifications in this document are guaranteed unless noted Ty pic al.Typical
characteristics describe typical or average performance and provide useful
reference in
formation.
Specifications that are marked with the
indirectly
The performance limits in this specification are valid with these conditions:
The instrument must be in an environment with temperature, altitude,
humidity, and vibration within the operating limits described in these
specifica
The instrument must have had a warm-up period of at least 30 minutes.
For modules, the performance limits in this specification are valid with these
conditions:
The logic protocol analyzer modules must be installed in a Logic Analyzer
Mainframe.
The module must have been calibrated/adjusted at an ambient temperature
between +20 °C and +30 °C.
Atmospheric Characteristics
The following table lists the Atmospheric characteristics of the Tektronix logic
tocol analyzers.
pro
symbol are checked directly (or
) at your nearest Tektronix location.
tions.
Table 1: Atmospheric characteristics
CharacteristicDescription
Temperature
Relative Hum idity
Operating (no media in CD or DV D drive of the mainframe)
0 °C to +40 °C, 15 °C/hr maximum gradient, noncondensing, derated 1°C per 300 meters (~1000 ft)
above 1500 meters (~5000 ft) altitude
Nonoperating (no media)
-20 °C to +60 °C, 15 °C/hr maximum gradient, without disk media installed in disk drives
Operating (no media)
5% to 95% relative humidity (%RH), up to +30 °C
5% to 75% relative humidity above +30 °C up to +40 °C, noncondensing and as limited by a Maximum
Wet-bulb Temperature of +29.4 °C
Nonoperating (no media)
5% to 95% relative humidity (%RH) up to +30 °C, 75% RH up to 60 °C, noncondensing, and as limited
by a Maximum Wet-Bulb Temperature of +40 °C (derates relative humidity to 20% RH at +60 °C)
TLA7SA08 & TLA7SA16 Series Product Specifications & Performance Verification1
Specifications and Characteristics
Table 1: Atmospheric characteristics (cont.)
CharacteristicDescription
Altitude
Operating
Up to 3,000 meters (~10,000 ft), derated 1 °C per 300 meters (~1000 ft) above 1500 meters (~5000 ft)
altitude
Nonoperating
Up to 12,000 meters (~40,000 ft)
2TLA7SA08 & TLA7SA16 Series P roduct Specifications & P erformance Verification