Tektronix TLA7S08, TLA7S16 Declassification And Security Instructions

Page 1
TLA7S08 and TLA7S16 Serial Analyzer Modules Declassication and Security Instructions
www.tektronix.com
071-2189-00
Page 2
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14200 SW Karl Braun Drive P.O. B o x 5 0 0 Beaverton, O USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200. Worl dwid e, vis it www.tektronix.com to nd contacts in your area.
R 97077
Page 3
Table of Contents
Preface .............................................................................................................. iii
Clear and Sanitize Procedures. . ..... . ..... . .... . . .... . ..... . ..... . ..... . ..... ..... . ..... . ..... . ..... . ... . . . .... . .. 1
Memory Devices............................................................................................... 1
Data Export De
Troubleshooting..................................................................................................... 3
How to Clear a Non-Functional Instrument ........................................... ..................... 3
vices........................................................................................... 2
TLA7S08 and TLA7S16 Declassication and Security Instructions i
Page 4
Table of Contents
ii TLA7S08 and TLA7S16 Declassication and Security Instructions
Page 5
Preface
This document helps customers with data security concerns to sanitize or remove memory devices from TLA7S08 and TLA7S16 Serial Analyzer Modules.
These products have data storage (memory) devices and data output devices (USB ports). These instructions tell how to do the following:
Clear or sanitize the memory devices
Clear or sanitize an instrument that is not functioning
Related Do
Products
cuments
Terms
The following Tektronix products are covered by this document:
TLA7S08
TLA7S16
The TLA7S08 and TLA7S16 Serial Analyzer Modules Service Manual, part number 071-2888-XX, is available on the Tektronix Web site at www.tektronix.com/manuals.
The following terms may be used in this document:
Clear. This removes data on media/memory before reusing it in a secured area. All reusable memory is cleared to deny access to previously stored information by standard means of access.
Erase. This is equivalent to clear.
Media storage/data export. Various devices that are used to store or export data
from the instrument, such as a USB port.
Nonvolatile memory. Data is retained when the instrument is powered off.
Remove. Thisisaphysicalmeanstoclearthedatabyremovingthememory
device from the instrument. Instructions are available in the product service manual.
Sanitize. This eradicates the data from media/memory so that the data cannot be recovered by other means or technology. This is typically used when the device will be moved (temporarily or permanently) from a secured area to a non-secured area.
Scrub. This is equivalent to sanitize.
User-modiable. The m emory device can be written to by the user during normal
instrument operation, using the instrument’s user interface or remote control.
Volatile memory. Data is lost when the instrument is powered off.
TLA7S08 and TLA7S16 Declassication and Security Instructions iii
Page 6
Preface
iv TLA7S08 and TLA7S16 Declassication and Security Instructions
Page 7
Clear and Sanitize Procedures
Memory Device
s
The following tables list the volatile and nonvolatile memory devices in the standard instrument and listed options.
YouonlyneedtoperformtheNonvolatile Memory Security Procedure to clear the instrument.
Table 1: Vol
Type and minimum size Function
DDR,
1.53 G X 8
DRAM, 4MX8
atile Memory Devices
Acquisition memory for storing acquired data
Firmware execution code
User Modiable Input method Location Process to clear
No
No Written by the
Table 2: Nonvolatile Memory Devices
Type and minimum size Function
FLASH, 2 M X 16,
NVRAM, 128 K X 8
Flash, 16 M X 16
Stores instrument rmware and FPGA data
Stores instrument serial number and calibration constants
Stores FPGA bit les
User Modiable Input method Location Process to clear
No Programmed at the
No. Written by
No Programmed at the
Written by FPGA
processor system
factory,nouser data
processor
factory,nouser data
Process to sanitize
Acquisition board Remove power
LPU board Remove power
LPU board Remove LPU
LPU board Remove LPU
Acquisition board
source from the instrument for at least 20 seconds.
source from the instrument for at least 20 seconds.
Process to sanitize
board and either securely store it or destroy it.
board and either securely store it or destroy it.
Remove Acquisition board and either securely store it or destroy it.
Nonvo
latile Memory
Security Procedure
User data is not stored in nonvolatile memory. To secure nonvolatile memory, proceed as follows:
1. Remove the LPU board and Acquisition board from the module. Refer to the TLA7S16 Serial Analyzer Module PCI Express Instruction Manual, 071-2087-XX on the Tektronix Web site at www.tektronix.com/manuals.
2. Since there is no way for you to erase nonvolatile memory, either store these circuit boards in a secure area or destroy them.
TLA7S08 and TLA7S16 Declassication and Security Instructions 1
Page 8
Clear and Sanitize Procedures
Data Export Devices
Table 3: Data Export Devices
The following table lists the data export devices in the standard instrument and listed options.
Type Function
Reference clock
VXI backplane
Reference
Provides a copy of the internal reference clock
Interfaces with TLA mainframe
Clock and VXI
Backplane
User Modiable
No User directed
No Instrument
Input method Location Process to disable
function and application software
Front panel of module
Backplane connectors
Unplug the Reference Clock cable from the output port on the front panel (See below).
Power off the TLA mainframe power and remove the module (See below).
To loc ate and remove the Reference Clock cable and module from the TLA mainframe, refer to the TLA7S16 Serial Analyzer Module PCI Express Instruction Manual, 071-2087-XX on the Tektronix Web site at www.tektronix.com/manuals.
2 TLA7S08 and TLA7S16 Declassication and Security Instructions
Page 9
Troubleshooting
How to Clear a Non-Functional Instrument
If your instrument is not functioning and you need to clear it, remove the power source from the instrument for at least 20 seconds.
TLA7S08 and TLA7S16 Declassication and Security Instructions 3
Page 10
Troubleshooting
4 TLA7S08 and TLA7S16 Declassication and Security Instructions