The servicing instructions are for use by qualified
personnel only. To avoid personal injury, do not
perform any servicing unless you are qualified to
do so. Refer to all safety summaries prior to
performing service.
Copyright T ektronix, Inc. All rights reserved.
T ektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and price change privileges reserved.
Printed in the U.S.A.
T ektronix, Inc., P.O. Box 1000, Wilsonville, OR 97070–1000
TEKTRONIX and TEK are registered trademarks of T ektronix, Inc.
Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it. To avoid potential hazards, use this
product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the instrument.
Read the General Safety Summary in other related instrument manuals for
warnings and cautions related to operating the instrument.
To Avoid Fire or
Personal Injury
Use Proper Power Cord. Use only the power cord specified for this product and
certified for the country of use.
Connect and Disconnect Properly . Do not connect or disconnect probes or test
leads while they are connected to a voltage source.
Ground the Product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, ensure that the product is properly grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and marking on the product. Consult the product manual for further ratings
information before making connections to the product.
The common terminal is at ground potential. Do not connect the common
terminal to elevated voltages.
Do not apply a potential to any terminal, including the common terminal, that
exceeds the maximum rating of that terminal.
Do Not Operate Without Covers. Do not operate this product with covers or panels
removed.
Use Proper Fuse. Use only the fuse type and rating specified for this product.
Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present.
Do Not Operate With Suspected Failures. If you suspect there is damage to this
product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry .
TLA 711 Benchtop Mainframe Rackmount Installation
iii
General Safety Summary
Provide Proper Ventilation. Refer to the manual’s installation instructions for
details on installing the product so it has proper ventilation.
Symbols and Terms
T erms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
T erms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the
marking.
WARNING indicates an injury hazard not immediately accessible as you read the
marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
iv
WARNING
High Voltage
Protective Ground
(Earth) T erminal
CAUTION
Refer to Manual
Double
Insulated
TLA 711 Benchtop Mainframe Rackmount Installation
Service Safety Summary
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnect Power. To avoid electric shock, disconnect the main power by means
of the power cord or, if provided, the power switch.
Use Care When Servicing With Power On. Dangerous voltages or currents may
exist in this product. Disconnect power, remove battery (if applicable), and
disconnect test leads before removing protective panels, soldering, or replacing
components.
To avoid electric shock, do not touch exposed connections.
TLA 711 Benchtop Mainframe Rackmount Installation
v
Installation Instructions
These instructions provide information for installing the 020–2197–XX TLA 711
Benchtop Mainframe Rackmount kit.
You can install the rackmount to either recess the mainframe into the front of the
mainframe rack or to extend it out of the front of the mainframe rack. You can
recess the mainframe from 0.735 inches to 5.735 inches in 0.5-inch increments.
You can extend the mainframe at specified depths from 0.57 inches to 5.57
inches in 0.5-inch increments.
The steps for installing the rackmount kits consist of installing the rackmount
hardware to the mainframe and then installing the mainframe in the rack.
A list of the kit contents and replaceable parts appears at the end of this
document.
You will need the following tools to install the rackmount option:
WARNING. Because of the size and weight of the mainframe, use care when
lifting or moving the benchtop mainframe to avoid personal injury while
performing the installation procedures. Use two people to lift or move the
mainframe.
Install the Rackmount Kit Hardware
Refer to Figure 1 on page 2 and perform the following steps to install the
rackmount kit hardware:
1. Remove the rubber feet from the mainframe.
2. Determine whether you want to install the benchtop mainframe so that it
extends out of the rack or is recessed within the rack.
If you want to extend the mainframe from the front of the rack, position the
rackmount bracket as shown on the right side of Figure 1. If you want to
recess the mainframe in the rack, position the rackmount bracket as shown
on the left side of Figure 1.
TLA 711 Benchtop Mainframe Rackmount Installation
1
Installation Instructions
Screws (2)
3. Determine how far you want to recess or extend the mainframe and install
the rackmount adapter as shown using two 8-32 screws (included with the
kit). Similarly, install the other bracket on the other side of the mainframe.
Latch is
located at
the rear.
Screws (5)
Handle
Screws (2)
Bracket
Figure 1: Installing the rackmount kit hardware
4. Remove the five 8-32 screws from both sides of the mainframe.
5. Attach the left and right inside tracks to the sides of the mainframe using
five 8-32 flathead screws. Orient the button latch so that it faces away from
the mainframe and toward the rear.
Track
Bracket
2
TLA 711 Benchtop Mainframe Rackmount Installation
Install the Mainframe in the Rack
Use the following procedure to install the benchtop mainframe in the rack:
1. Identify the right and left slide-out track assemblies: find the date code label
on each assembly. The assembly to be mounted in the left side of the rack
(the side nearest the left side of the mainframe when it is mounted in the
rack) has a date code that ends with LH (for left hand). The right assembly
has a date code ending with RH.
2. Measure the distance between the front and rear rail of the rack.
3. Align the rear bracket to the right slide-out track as shown in Figure 2. Note
that the rear bracket has two pairs of mount-through holes. When aligning
the bracket and the track, select a pair of holes that mount the rear bracket so
the flange-to-flange distance (see Figure 2) matches the front rail to rear rail
spacing that you just measured in step 2.
NOTE. If necessary, refer to the rackmount kit instructions for detailed installation information.
Installation Instructions
4. Attach the rear bracket to the right slide-out track using two 10-32 screws
and a bar nut as shown in Figure 2. Leave the screws loose so that you can
adjust the overall length of the slide-out track assembly when installing the
mainframe in the rack.
5. Repeat steps 3 and 4 to assemble the left slide-out track assembly.
TLA 711 Benchtop Mainframe Rackmount Installation
3
Installation Instructions
Button latch is closest
to top of track.
Front flange (mounts to
front rail of rack)
Front of
instrument
Inner track
Left slide-out track assembly
Figure 2: Assembling the slide-out track assemblies
Outer track
6. Select the mounting position in the rack: select two 0.5-inch spaced holes in
the front rail and verify that the 1.5-inch and 14.00-inch clearances exist
relative to the mounting holes as shown in Figure 3.
NOTE. The 14-inch vertical clearance is only valid for the mainframe without the
rubber feet and cable tray.
4
TLA 711 Benchtop Mainframe Rackmount Installation
304.8 mm
355.6 mm
(14.0 in)
Installation Instructions
(12.0 in)
Left-front rail
38.1 mm
(1.5 in)
12.7 mm (0.5 in)
(for correct position of
securing holes)
Figure 3: Vertical clearance for rack installation (standard mainframe)
7. Select the mounting method according to the rack type.
To mount the slide-out tracks with the front and rear flanges outside of the
front and rear rails, use method A shown in Figure 4 when doing step 8. Add
a bar nut to the installation only if the rails have untapped holes.
To mount with front and rear flanges inside of the rails, use mounting
method B shown in Figure 4 when doing step 8. This mounting method
assumes untapped holes.
8. Using the mounting method determined from the previous step, secure the
right slide-out track assembly to the front and rear rails. Seat the screws
fully, but lightly, to adjust the mounting later.
9. Tighten the screws left loose in step 4 to fix the front to rear flange spacing
of the slide-out track assembly.
TLA 711 Benchtop Mainframe Rackmount Installation
5
Installation Instructions
10. Repeat steps 6 through 9 to mount the left slide-out track assembly.
Left slide-out trackLeft slide-out track
Use a bar nut if front
rails are not tapped.
Left-front rail
10-32 Pan head screw (4)
Mounting method AMounting method B
Left-front rail
10-32 Flat head
screw (4)
Use two flat head screws if the cabinet
rails have countersunk mounting holes;
otherwise use two pan head screws.
10-32 Pan head
screw (4)
Figure 4: Installing the slide-out track assemblies in the rack (top view)
WARNING. Installing the benchtop mainframe in the rack requires at least two
people. Because of the size and weight of the mainframe, installing the mainframe alone can result in personal injury or equipment damage. One person
should be on the left side of the mainframe with one hand holding the left front
handle and with the other hand underneath the mainframe. The other person
should do likewise on the right side of the mainframe.
11. Fully extend the inner tracks of each slide-out section out of the front of the
rack until they lock.
12. Lift the mainframe and insert the left and right tracks that extend from the
rear of the mainframe into the ends of the tracks that you just extended.
Make sure that the tracks mounted on the mainframe slip just inside the inner
tracks that you extended earlier.
13. Slide the mainframe back until it stops.
14. Push to release the button latches, located on the outside of each track, and
continue to slide the mainframe all the way back into the rack cabinet.
6
TLA 711 Benchtop Mainframe Rackmount Installation
Installation Instructions
15. Tighten the four screws that you left loose at the rear of the rack when you
performed step 8 and pull the mainframe part way out of the rack. (Tighten
the 10-32 screws using 28 inch-lbs of torque.)
16. Be sure that the screws that you left loose at the front of the rack are loose
enough to allow the slide-out track assemblies to seek the normal positions.
17. Tighten the screws and push the mainframe all the way into the rack. If the
tracks do not slide smoothly, readjust the level using the method described in
steps 15 and 16. (Tighten the 10-32 screws using 28 inch-lbs of torque.)
18. Install the proper power cord at the rear of the mainframe.
TLA 711 Benchtop Mainframe Rackmount Installation
7
Replaceable Parts
This section contains a list of the replaceable parts for the 020–2197–XX
TLA 711 Benchtop Mainframe Rackmount Kit. Use this list to identify and order
replacement parts.
Parts Ordering Information
Replacement parts are available through your local Tektronix field office or
representative.
Changes to Tektronix products are sometimes made to accommodate improved
components as they become available and to give you the benefit of the latest
improvements. Therefore, when ordering parts, it is important to include the
following information in your order:
Part number
Instrument type or model number
Instrument serial number
Instrument modification number, if applicable
If you order a part that has been replaced with a different or improved part, your
local Tektronix field office or representative will contact you concerning any
change in part number.
Change information, if any, is located at the rear of this manual.
TLA 711 Benchtop Mainframe Rackmount Installation
9
Replaceable Parts
Using the Replaceable Parts List
This section contains a list of the mechanical and/or electrical components that
are replaceable for the 020–2197–XX TLA 711 Benchtop Mainframe Rackmount
Kit. Use this list to identify and order replacement parts. The following table
describes each column in the parts list.
Parts list column descriptions
ColumnColumn nameDescription
1Figure & index numberItems in this section are referenced by figure and index numbers to the exploded view
illustrations that follow.
2Tektronix part numberUse this part number when ordering replacement parts from Tektronix.
3 and 4Serial numberColumn three indicates the serial number at which the part was first effective. Column four
indicates the serial number at which the part was discontinued. No entries indicates the part is
good for all serial numbers.
5QtyThis indicates the quantity of parts used.
6Name & descriptionAn item name is separated from the description by a colon (:). Because of space limitations, an
item name may sometimes appear as incomplete. Use the U.S. Federal Catalog handbook
H6-1 for further item name identification.
7Mfr. codeThis indicates the code of the actual manufacturer of the part.
8Mfr. part numberThis indicates the actual manufacturer’s or vendor’s part number.
Abbreviations
Mfr. Code to Manufacturer
Cross Index
Abbreviations conform to American National Standard ANSI Y1.1–1972.
The table titled Manufacturers Cross Index shows codes, names, and addresses
of manufacturers or vendors of components listed in the parts list.
10
TLA 711 Benchtop Mainframe Rackmount Installation
Replaceable Parts