Tektronix Logic Analyzer Family
Performance Verification and
Adjustment Procedures
Version 4.0 Software
071-0862-00
Warning
The servicing instructions are for use by qualified
personnel only. To avoid personal injury, do not
perform any servicing unless you are qualified to
do so. Refer to all safety summaries prior to
performing service.
Use, duplication, or disclosure by the Government is subject to restrictions as set forth in subparagraph (c)(1)(ii) of the
Rights in T echnical Data and Computer Software clause at DFARS 252.227-7013, or subparagraphs (c)(1) and (2) of the
Commercial Computer Software – Restricted Rights clause at F AR 52.227-19, as applicable.
T ektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and price change privileges reserved.
T ektronix, Inc., 14200 SW Karl Braun Drive, Beaverton, OR 97077
TEKTRONIX and TEK are registered trademarks of T ektronix, Inc.
WARRANTY
T ektronix warrants that the products that it manufactures and sells will be free from defects in materials and
workmanship for a period of one (1) year from the date of shipment. If a product proves defective during this
warranty period, T ektronix, at its option, either will repair the defective product without charge for parts and labor,
or will provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration
of the warranty period and make suitable arrangements for the performance of service. Customer shall be
responsible for packaging and shipping the defective product to the service center designated by T ektronix, with
shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a
location within the country in which the T ektronix service center is located. Customer shall be responsible for
paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate
maintenance and care. T ektronix shall not be obligated to furnish service under this warranty a) to repair damage
resulting from attempts by personnel other than T ektronix representatives to install, repair or service the product;
b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any
damage or malfunction caused by the use of non-T ektronix supplies; or d) to service a product that has been
modified or integrated with other products when the effect of such modification or integration increases the time
or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX IN LIEU OF ANY OTHER WARRANTIES, EXPRESS
OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’
RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND
EXCLUSIVE REMEDY PROVIDED TO THE CUST OMER FOR BREACH OF THIS WARRANTY.
TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT , SPECIAL,
INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR
THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it. To avoid potential hazards, use this
product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the system. Read
the General Safety Summary in other system manuals for warnings and cautions
related to operating the system.
To Avoid Fire or
Personal Injury
Use Proper Power Cord. Use only the power cord specified for this product and
certified for the country of use.
Connect and Disconnect Properly . Do not connect or disconnect probes or test
leads while they are connected to a voltage source.
Ground the Product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, ensure that the product is properly grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
The common terminal is at ground potential. Do not connect the common
terminal to elevated voltages.
Do not apply a potential to any terminal, including the common terminal, that
exceeds the maximum rating of that terminal.
Do Not Operate Without Covers. Do not operate this product with covers or panels
removed.
Use Proper Fuse. Use only the fuse type and rating specified for this product.
Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present.
Do Not Operate With Suspected Failures. If you suspect there is damage to this
product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry .
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
ix
General Safety Summary
Provide Proper Ventilation. Refer to the manual’s installation instructions for
details on installing the product so it has proper ventilation.
Symbols and Terms
T erms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
T erms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the
marking.
WARNING indicates an injury hazard not immediately accessible as you read the
marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
WARNING
High Voltage
x
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Protective Ground
(Earth) T erminal
CAUTION
Refer to Manual
Double
Insulated
Service Safety Summary
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone
Disconnect Power
Use Care When Servicing
With Power On
Do not perform internal service or adjustments of this product unless another
person capable of rendering first aid and resuscitation is present.
To avoid electric shock, disconnect the main power by means of the power cord
or, if provided, the power switch.
Dangerous voltages or currents may exist in this product. Disconnect power,
remove battery (if applicable), and disconnect test leads before removing
protective panels, soldering, or replacing components.
To avoid electric shock, do not touch exposed connections.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
xi
Service Safety Summary
xii
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Preventing Electrostatic Discharge
CAUTION. Static discharge can damage internal semiconductor components.
Follow the guidelines listed below to avoid product damage.
When performing service that requires internal access to an instrument, adhere to
the following precautions to avoid damaging internal modules or their components:
HAvoid handling modules or components in areas that have floors or work
surfaces capable of generating a static charge.
HSpray carpeted work areas with a solution of equal parts of water and fabric
softener.
HWear clothing made from materials that do not accumulate static charges.
Wool (and some artificial fibers) build up static charges readily; cotton
conducts electricity and resists static accumulation.
HMinimize the handling of static-sensitive devices.
HTransport and store static-sensitive devices in their protected containers or on
a metal rail. Label any package that contains static-sensitive parts.
HService instruments and modules at grounded static-free work stations.
HDo not allow any device capable of generating a static charge on work
station surfaces.
HWear a grounding wrist strap while working with static-sensitive devices.
HHandle circuit boards by their edges, if possible.
HDo not slide static-sensitive components over any surface.
HDo not use high-velocity compressed air to clean or dry components or
modules.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
xiii
Preventing Electrostatic Discharge
xiv
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Preface
This manual contains performance verification and adjustment information for
the Tektronix Logic Analyzer Family hardware components. Each section covers
an individual hardware component such as a mainframe, module, or adjustment
fixture and includes the following information:
HAn introduction that describes the procedures
HEquipment lists required to perform the procedures
HFunctional verification procedures to quickly determine that the component
operates properly (as for a brief incoming inspection)
HCertification test procedures and Calibration Data Reports
HPerformance verification procedures to ensure the component operates
properly and meets warranted specifications
HAdjustment procedures used to adjust the component to meet its warranted
specifications
This manual also includes an introduction with important overview information,
and an appendix with a parts list and an exploded view of the Logic Analyzer
Adjustment and Verification fixture.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
xv
Preface
Naming Conventions
This manual uses the following abbreviated names for the full product name
throughout the text:
TLA7Lx, TLA7Mx, TLA7Nx, TLA7Px, or TLA7Qx Logic Analyzer
Module
TLA7Dx and TLA7Ex Digital Storage ScopeDSO module
TLA7PG2 Pattern Generator Modulepat gen module
Device Under Test / System Under TestDUT / SUT
portable mainframe
benchtop chassis
LA module or LA
Throughout this manual, the term “module” refers to a LA module, a DSO
module, an expansion module, or a benchtop controller module.
Tektronix provides service to cover repair under warranty as well as other
services that are designed to meet your specific service needs.
Whether providing warranty repair service or any of the other services listed
below, Tektronix service technicians are well equipped to service the Tektronix
Logic Analyzer Family products. Tektronix technicians train on Tektronix
products; they have access to the latest information on improvements to the
products as well as the latest new product upgrades. Services are pr ovided at
Tektronix Service Centers and on-site at your facility, depending on your
location.
Warranty Repair Service
xvi
Tektronix technicians provide warranty service at most Tektronix service
locations worldwide. The Tektronix product catalog lists all service locations
worldwide or you can visit us on our Customer Services World Center web site
at Tektronix.com/Measurement/Service. See our latest service offerings and
contact us by email.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Preface
Calibration and Repair
Service
In addition to warranty repair, Tektronix Service offers calibration and other
services which provide cost-effective solutions to your service needs and qualitystandards compliance requirements. Our instruments are supported worldwide by
the leading-edge design, manufacturing, and service resources of Tektronix to
provide the best possible service.
The following services can be tailored to fit your requirements for calibration
and/or repair of the Tektronix Logic Analyzer Family products.
Service Options. Tektronix Service Options can be selected at the time you
purchase your instrument. You select these options to provide the services that
best meet your service needs. These service options are listed on the TektronixService Options page at the beginning of this manual.
Service Agreements. If service options are not added to the instrument purchase,
then service agreements are available on an annual basis to provide calibration
services or post-warranty repair coverage for the Tektronix Logic Analyzer
Family products. Service agreements may be customized to meet special
turn-around time and/or on-site requirements.
Service on Demand. Tektronix also offers calibration and repair services on a
“per-incident” basis that is available with standard prices for many products.
Self Service. Tektronix supports repair to the replaceable-part level by providing
for circuit board exchange.
Use this service to reduce down-time for repair by exchanging circuit boards for
remanufactured ones. Tektronix ships updated and tested exchange boards. Each
board comes with a 90-day service warranty.
For More Information. Contact your local Tektronix service center or sales
engineer for more information on any of the Calibration and Repair Services just
described.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
xvii
Preface
Contacting Tektronix
Phone1-800-833-9200*
AddressTektronix, Inc.
Department or name (if known)
14200 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
Web sitewww.tektronix.com
Sales support1-800-833-9200, select option 1*
Service support1-800-833-9200, select option 2*
Technical supportEmail: techsupport@tektronix.com
1-800-833-9200, select option 3*
1-503-627-2400
6:00 a.m. – 5:00 p.m. Pacific time
*This phone number is toll free in North America. After office hours, please leave a
voice mail message.
Outside North America, contact a Tektronix sales office or distributor; see the
Tektronix web site for a list of offices.
xviii
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Introduction
Introduction
This manual contains instructions for testing the performance and adjustment of
the following instruments:
HLogic analyzer probes (tested with the associated logic analyzer)
HLogic Analyzer Module adjustment/verification fixture
HTLA7Dx and TLA7Ex Digitizing Oscilloscope Module
HTLA7PG2 Pattern Generator Modules
HPattern generator probes (tested with the associated pattern generator
module)
Four levels of testing are provided:
HFunctional verification procedures
HPerformance verification procedures
HCertification test procedures and Calibration Data Reports
HAdjustment procedures
Not all of the components of the Tektronix Logic Analyzer Family require all
four levels of testing. Table 1–1 on page 1–2 lists the components of the logic
analyzer family and the levels of testing described in this manual.
Certain instruments, such as the Termination Board (used with pattern generator
modules), do not require any testing or certification.
Table 1–2 lists the typical order of procedures needed for verifying the performance and adjustment of the instruments. These procedures assume that you are
familiar with operating the instruments.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
2. If module fails performance verification procedure,
then run the adjustment procedure
1. Run adjustment procedure
2. Run performance verification procedure
3. Rerun performance verification if step 2 is done
Logic Analyzer
Run performance verification procedure
2
Run performance verification procedure
Adjustment and
Verification fixture
Pattern Generator
Modules
DSO modules1. Run performance verification procedure
1
Refer to the appropriate certification sections in this document for all certification procedures
2
The Logic Analyzer Adjustment and Verification fixture requires calibration every two years
1–2
Run the functional verification and the performance
verification procedures
2. If module fails performance verification procedure,
then run the adjustment procedure
3. Rerun performance verification if step 2 is done
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Run the functional verification and the performance
verification procedures
1. Run performance verification procedure
2. If module fails performance verification procedure,
then run the adjustment procedure
3. Rerun performance verification if step 2 is done
1
following board repair
Functional Verification Procedures
Functional verification procedures verify the basic functionality of the instrument inputs, outputs, and basic instrument actions. These procedures include
power-on and extended diagnostics, self calibration, as well as semi-automated
or manual check procedures. These procedures can be used for incoming
inspection purposes.
Certification Test Procedures
Certification procedures are used to certify the accuracy of an instrument and
provide a traceability path to national standards. Certification data is recorded on
Calibration Data Reports provided with this manual. Calibration Data Reports
are produced for the TLA600 Logic Analyzers, LA modules, and DSO modules
as an output from the PV/Adjust software.
Introduction
Performance V erification Procedures
Performance verification procedures confirm that a product meets or exceeds the
performance requirements for the published specifications documented in the
Tektronix Logic Analyzer Family User Manual. Refer to Figure 1–1 on
page 1–4 for a graphic overview of the procedures.
Adjustment Procedures
Adjustment procedures check for and, if necessary, correct any adjustment errors
discovered when performing functional or performance verification procedures.
The adjustment procedures documented in this manual are controlled by
software, except for the procedures for the adjustment/verification fixture.
Some procedures require user intervention to move probes or change test
equipment settings. The adjustment software is part of the self-calibration
routines in the logic analyzer application software and the PV/Adjust software
provided with the product CD-ROM.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Yes
Done
Test Equipment
The procedures in this manual use external, traceable signal sources to directly
test characteristics that are designated as checked (n) in the Specifications
section of the Tektronix Logic Analyzer Family User Manual. Table 1–3 shows
the required equipment list. Always warm up the equipment for 30 minutes
before beginning the procedures.
T able 1–3: Test equipment
Item number and descriptionMinimum requirementsExampleWhere used
Introduction
1. MainframeTLA700 Series MainframeBenchtop Chassis or
Portable Mainframe
2. ControllerTLA700 Series ControllerBenchtop Controller used
with the Benchtop Chassis or
the embedded controller in
the Portable Mainframe
3. TLA700 Series Logic Analyzer
Performance Verification &
Adjustment Software
4. Adjustment/verification fixture,
with one of the following
Power Supplies:
USA/CAN
Europe
Japan
United Kingdom
5. DSO probe calibration fixtureOne requiredTektronix part number
6. Oscilloscope1 GHz bandwidth Delay time accura-
7. DSO probesThree required, with < one-inch
8. Logic analyzer probesTwo requiredTektronix P6417 or P6418
9. High density logic analyzer
probe (optional)
10. Pattern Generator moduleOne requiredTLA7PG2Merge and Deskew pattern
11. Pattern Generator probe
cable
Included with the Tektronix Logic
Analyzer Family CD-ROM
Tektronix part number
671-3599-XX
Tektronix part numbers:
12 V , 1.5 A
12 V , 1.5 A
12 V , 1.5 A
12 V , 1.5 A
Tektronix P6243 or P6245
probe, with std. accessories
Logic Analyzer probes
Analyzer probe
012-A212-00
All procedures, except
TLA600 procedures
All procedures, except
TLA600 procedures
All procedures for the
TLA600 Logic Analyzers,
LA modules and
DSO modules
Most procedures for
the LA module
DSO functional verification
procedure
fixture procedures and
pattern generator procedures
Most procedures for
the DSO and Pattern
Generator modules
Most procedures for TLA600
and the LA module
Most procedures for TLA600
and the LA module
generator procedures
Pattern generator procedures
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
1–5
Introduction
T able 1–3: Test equipment (Cont.)
Item number and descriptionWhere usedExampleMinimum requirements
12. TTL/CMOS Pattern Generator probes
13. ECL Pattern Generator
probes
14. Termination boardTwo requiredTektronix part number
15. Frequency counterFrequency accuracy: <0.0025%
16. Function generatorAmplitude: 4V
17. Power supplyVoltage: –2 V
18. BNC CableImpedance: 50 W,
19. T-ConnectorOne requiredTektronix part number
20. SMB-to-BNC cableOne requiredTektronix P6041PV procedures for the
21. Adapter, male BNC-to-male
BNC
22. Miniature probe-to-square
pin adapters
23. 0.025-inch square pin3/4-inch length, (two required)Tektronix part number
24. Digital multimeter
with test leads
25. Connector, dual-bananaFemale BNC-to-dual bananaTektronix part number
26. Voltage referenceAccuracy: ≤0.01%Data Precision 8200PV procedures for the
27. Capacitor
28. Cable, dual-inputFemale BNC-to-dual male BNCTektronix part number
29. Adapter, N-to-BNCMale type N-to-female BNCTektronix part number
30. Adapter, SMA-to-BNCFemale type SMA-to-female BNCTektronix part number
1
Four requiredTektronix P6470 Pattern
Generator probes
Four requiredTektronix P6471 Pattern
Generator probes
067-A016-00
Hewlett Packard 5314APortable Mainframe and
Frequency range: 1 kHz to 100 MHz
T ektronix AFG2020Pattern generator procedures
Offset: 2 V (50 W termination)
Frequency: 1 MHz or higher
METRONIX 524BPattern generator procedures
Current: 1 A
Tektronix part number
Length: 24 in (two required)
One requiredTektronix part number
Two requiredTektronix part number
DCV accuracy: 0.1% from –10 V to
+100 V
0.1 mF, 200 VTektronix part number
012-1342-XX
103-0030-XX
103-0029-XX
103-0177-XX
131-1426-XX
Tektronix DMM 900 SeriesAll procedures except
103-0090-XX
283-0189-XX
067-0525-XX
103-0045-XX
015-1018-XX
Most pattern generator
procedures
Most pattern generator
procedures
Most pattern generator
procedures
Benchtop Controller
procedures
Pattern generator procedures
Pattern generator procedures
Benchtop Controller
DSO procedures
Portable Mainframe
(power supply check)
Portable Mainframe
(power supply check)
Benchtop Controller
PV procedures for the
LA and DSO module
LA and DSO module
PV procedures for the
LA and DSO module
PV procedures for
the DSO module
PV procedures for
the DSO module
DSO procedures
1–6
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
T able 1–3: Test equipment (Cont.)
Item number and descriptionWhere usedExampleMinimum requirements
Introduction
31. Cable, precision
50 W coaxial
32. Generator, functionFrequency range: 1 Hz to 20 MHz
33. Generator, sine waveFrequency range: 100 kHz to
50 W, 36 in, male-to-male BNC
connectors
Frequency accuracy: 0.1%
Amplitude range: 10 mV to 20 V
1005 MHz
Tektronix part number
012-0482-XX
All procedures except
Benchtop Chassis
Hewlett Packard 3325BPV procedures for
the DSO module
p-p
Gigatronics 6061PV procedures for the
DSO and LA module
Frequency accuracy: <6 ppm
Amplitude: 2 mV to 1.5 V
RMS
Accuracy: <0.35 dB
34. Power meter with sensorBandwidth: >1.2 GHz Accuracy:
0.2 dB Sensitivity: 500 pW to 20 mW
35. Adapter, N-to-BNC (used
Female type N-to-male BNCTektronix part number
only with power meter)
36. 50 W power divider (used
only with power meter)
37. SMA-to-BNC adapters (used
only with power meter)
38. Printer with cable
(optional, used to print
Calibration Data Reports)
1
The capacitor is installed across the Data Precision 8200 output terminals to reduce noise. If your voltage reference
Provide load isolation between
equipment Maximum VSWR: 1.50
For use with the 50 W power divider
(three required)
PC-compatible, continuous feed,
prints ASCII text, connects to EPP
connector
Rohde & Schwarz NRVS
with model NRV-Z4 Sensor
DSO adjustment
procedure #4
DSO adjustment
103-0058-XX
Tektronix part number
015-0565-XX
Tektronix part number
015-1018-XX
procedure #4
DSO adjustment
procedure #4
DSO adjustment
procedure #4
Any general purpose printerPrinting test results for
the LA and DSO module
produces <4 mVp-p of noise, external noise reduction is not necessary.
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
1–7
Introduction
Software Installation and Removal Instructions
These procedures describe loading and unloading the PV/Adjust software. It is
recommended you have ≥10 MB of free space on the hard drive before installing
the software. The Performance Verification software is located on Disc 1 of the
Tektronix Logic Analyzer Family Application Software CD.
NOTE. This installation program uses parameters you supply to create a custom
start-up file in your hard disk directory.
The batch file enables the software to configure your instrument properly before
it runs the program.
1. Power on the instrument.
2. Exit the Application.
Verify PV/Adjust Software
Version
Verify Directories
If your logic analyzer already has PV/Adjust software loaded on it, you must
verify that the version is the same as the version printed on Disc 1 of the
Tektronix Logic Analyzer Family Application Software CD.
If the version of the PV/Adjust software loaded on your logic analyzer is an
earlier version, you must delete the earlier version before you can load the newer
version.
If your logic analyzer already has a directory named Tekcats or Temptek on the
hard drive, the software installation cannot be completed. Follow these instructions to verify the directory is not present:
1. Select Start → Programs → Windows Explorer.
2. From the Exploring window select Tools → Find → Files or Folders.
3. Select the Name & Location tab, and search for the following directories:
HTekcats
HTemptek
4. If either directory is found follow the instructions under Removing the
Software beginning on page 1–9 to remove the software and the directories.
1–8
Tektronix Logic Analyzer Family Performance Verification and Adjustment Procedures
Loading...
+ 223 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.