Getting help and support............................................................................................................................................................. 11
Minimum system requirements.....................................................................................................................................13
Downloading and installing the software.......................................................................................................................14
Activate the license.......................................................................................................................................................14
View software version and license key details..............................................................................................................14
Setting up the test environment.................................................................................................................................................. 16
Setting up tests.................................................................................................................................................................... 16
Connection setup for host tests.................................................................................................................................... 18
Connection setup for module tests............................................................................................................................... 21
Test setup overview............................................................................................................................................................. 22
Search instruments connected to the application................................................................................................................ 22
Before you click start............................................................................................................................................................23
Starting the application................................................................................................................................................................26
Options menu functions....................................................................................................................................................... 28
TekExpress instrument control settings............................................................................................................................... 31
Setup panel: Configure the test setup.........................................................................................................................................32
DUT: Set DUT settings.........................................................................................................................................................34
Test Selection: Select the tests............................................................................................................................................ 36
Acquisitions: Set waveform acquisition settings.................................................................................................................. 37
Configuration: Set measurement limits for tests.................................................................................................................. 39
Preferences: Set the test run preferences........................................................................................................................... 42
Status panel: View the test execution status...............................................................................................................................43
View test execution status....................................................................................................................................................43
View test execution logs.......................................................................................................................................................44
Results panel: View summary of test results...............................................................................................................................45
Filter the test results.............................................................................................................................................................45
View a generated report.......................................................................................................................................................50
Saving and recalling test setup................................................................................................................................................... 51
Save the configured test setup............................................................................................................................................ 51
Load a saved test setup.......................................................................................................................................................52
Perform a test using pre-run session files............................................................................................................................53
Save the test setup with a different name............................................................................................................................57
Run a saved test in prerecorded mode................................................................................................................................57
About SCPI command......................................................................................................................................................... 59
Socket configuration for SCPI commands........................................................................................................................... 59
Set or query the device name of application........................................................................................................................65
Set or query the suite name of the application.....................................................................................................................65
Set or query the test name of the application...................................................................................................................... 66
Set or query the version name of the application.................................................................................................................68
Set or query the general parameter values..........................................................................................................................68
Set or query the acquire parameter values..........................................................................................................................70
Set or query the analyze parameter values......................................................................................................................... 74
Query the available devices in the DUT panel of the application.........................................................................................74
Query the available suites for the selected device...............................................................................................................75
Query the list of available tests of the application................................................................................................................75
Query the available version names of the application......................................................................................................... 76
Query the list of available instruments based on the specified instrument type...................................................................77
Set or query the IP address of the instrument based on the specified instrument type.......................................................77
Query the information of the generated report file............................................................................................................... 77
Query the information of the generated waveform files....................................................................................................... 78
Query the information of the generated image files............................................................................................................. 78
Query the active TekExpress application name................................................................................................................... 79
Set or query the DUTID of application................................................................................................................................. 79
Sets or query the acquire mode status................................................................................................................................ 79
Set or query the execution mode status.............................................................................................................................. 80
Generate the report for the current session......................................................................................................................... 80
Query the value of specified report header field in the report.............................................................................................. 81
Query the value of specified result detail available in report summary/details table............................................................81
Restore the setup to default settings................................................................................................................................... 82
Save the setup..................................................................................................................................................................... 82
Save the settings to a specified session.............................................................................................................................. 82
Open the setup from a specified session.............................................................................................................................83
Query the current setup file name........................................................................................................................................83
Run/stop/pause/resume the selected measurements execution in the application............................................................. 83
Query the current measurement execution status............................................................................................................... 84
Query whether the current setup is saved or not saved...................................................................................................... 84
Exit or close the application................................................................................................................................................. 84
Query the status of the previous command execution.........................................................................................................84
Query the last error occurred............................................................................................................................................... 85
Set or query the popup details............................................................................................................................................. 85
Sets or query the limit values in the limits editor window.....................................................................................................85
Set or query the waveform file recalled for the specified test name and acquire type.........................................................86
Set or query the enable/disable status of Verbose function.................................................................................................86
Query the enable or disable status of Continuous run function........................................................................................... 87
Set or query the enable/disable status of Continuous Run function.................................................................................... 87
Set or query the continuous run duration time value........................................................................................................... 88
Set or query the session create option in the continuous run function................................................................................ 88
Set or query the View report after generating option status.................................................................................................89
Returns the report as XML string......................................................................................................................................... 90
Copies all the images from current run session to the given destination location................................................................90
Selects the specified test(s) and deselect all other tests..................................................................................................... 90
Returns the complete information about the selected test...................................................................................................91
Set the default session.........................................................................................................................................................91
Save the run/config sessions............................................................................................................................................... 91
Load the run/config session................................................................................................................................................. 91
Delete the run/config session...............................................................................................................................................92
Run the run/config saved session........................................................................................................................................92
Query the available list in the run/config session.................................................................................................................92
Query the current run/config session................................................................................................................................... 93
Override the run/config session........................................................................................................................................... 93
Testing procedures and examples....................................................................................................................................... 94
Signal types used by tests............................................................................................................................................ 94
Signal validation procedure.......................................................................................................................................... 95
Single-ended output voltage range...............................................................................................................................96
Output AC Common-Mode voltage (RMS) (Host).........................................................................................................96
Signal Rise/Fall time (20% - 80%)................................................................................................................................ 98
Total jitter (Host)............................................................................................................................................................99
Data dependent jitter (Host)........................................................................................................................................101
Data dependent pulse width shrinkage (Host)............................................................................................................102
Eye mask hit ratio (Host).............................................................................................................................................105
Voltage modulation amplitude (p-p)............................................................................................................................107
Output AC Common-Mode voltage (RMS) (Module).................................................................................................. 112
Total jitter (Module)......................................................................................................................................................113
Data dependent jitter (Module)....................................................................................................................................114
Data dependent pulse width shrinkage (Module)........................................................................................................116
Eye mask hit ratio (Module)........................................................................................................................................ 117
File name extensions......................................................................................................................................................... 121
De-embed using filter files................................................................................................................................................. 122
Map the My TekExpress folder...........................................................................................................................................123
TekExpress is the Tektronix Test Automation Framework, developed to support your current and future test automation needs. TekExpress
uses a highly modular architecture that lets you deploy automated test solutions for various standards in a relatively short time.
Note: The skew between Data+ and Data- in the signal path from the DUT to the oscilloscope will be computed and compensated
before running the measurements.
Key Features
•Allows conformance testing to the latest Compliance Test Specification (CTS)
•Reliable Testing:
•Conformance to SFP+ and QSFP+ interconnect module electrical testing ensures reliable results
•SFP+ QSFP+ Tx is SFP+ and QSFP+ compliance software
•Enables execution of the physical-layer (PHY), electrical tests, and SFP+/QSFP+ based electrical measurements
•Automated testing:
•Minimizes user intervention when conducting time-consuming testing
•Reduces the time required to conduct testing
•Selective testing:
•Performs fully-automated testing for transmitter measurements
•Allows you to select individual tests or test groups in the tree-structure
•Avoids repeated testing through accurate and reliable results from a single run
•Quick testing:
•One-button selection of multiple tests ensures faster testing
•Test margins and statistical information aid analysis to find answers quickly
•Wide range of tests for SFP+/QSFP+ Host devices and SFP+/QSFP+ Module devices enables complete validation
•Complete compliance solution with an elaborate test fixture and signal sources support for SMA cable-based solution provides
cost-effective way to perform compliance testing
•User defined mode supports PRBS7, PRBS11, PRBS15, PRBS20 and PRBS23 in addition to patterns supported in Compliance
mode including PRBS9, PRBS31 and 8180
•Customize the setup:
•Modify the test setup according to the DUT configuration.
•Run test measurements with different record lengths.
•Run test measurements with more than one signal.
•Detailed test reporting:
•Provides a Pass/Fail summary table
•Provides margin details on each test
•Provides a consolidated report for all tests
•Provides additional information such as skew and signal type selected for each measurement
Use the product documents for more information on the application functions, understand the theory of operation, how to remotely program
or operate the application, and do other tasks.
Table 1: TekExpress Application documents
To learn aboutUse this document
How to use the application
How to remotely control the instrument
Conventions
This application help uses the following conventions:
TekExpress SFP+ QSFP+ Help
PDF version of this document can be downloaded from www.tek.com/downloads
Compiled HTML (CHM) version is integrated with the application. Press F1 key from the
keyboard to start the help.
Tektronix Part Number: 077-xxxx-xx
•The term "Application," and "Software" refers to the TekExpress SFP+ QSFP+ application.
•The term “DUT” is an abbreviation for Device Under Test.
•The term “select” is a generic term that applies to the two methods of choosing a screen item (button control, list item): using a mouse
or using the touch screen.
•A Note identifies important information.
Table 2: Icons used in the help
IconDescription
This icon identifies important information
This icon identifies conditions or practices that could result in loss of data.
This icon identifies additional information that will help you use the application more
efficiently.
Tektronix values your feedback on our products. To help us serve you better, please send us your suggestions, ideas, or comments on
your application or oscilloscope. Contact Tektronix through mail, telephone, or the Web site. See Contacting Tektronix at the front of this
document for contact information.
When you contact Tektronix Technical Support, please include the following information (be as specific as possible):
General information
•All instrument model numbers
•Hardware options, if any
•Modules used
•Your name, company, mailing address, phone number, FAX number
•Please indicate if you would like to be contacted by Tektronix about your suggestion or comments.
Application specific information
•Software version number
•Description of the problem such that technical support can duplicate the problem
•If possible, save the setup files for all the instruments used and the application
•If possible, save the TekExpress setup files, log.xml, *.TekX (session files and folders), and status messages text file
The following table describes the minimum system requirements for the TekExpress SFP+ QSFP+ application.
Table 3: Minimum system requirements
InstrumentsDescription
ProcessorSame as the oscilloscope
Operating SystemSame as the oscilloscope (Windows 7, 64-bit only)
MemorySame as the oscilloscope
Hard DiskSame as the oscilloscope
Display
Firmware
Same as the oscilloscope
TekScope 10.3.3 for non-SX series digital oscilloscopes
TekScope 10.3.0 for SX series digital oscilloscopes
1
Getting started
Software
•DPOJET, Jitter and Eye Diagram Analysis Tool, version 10.0.0.35 or later (64-bit)
•MathWorks MATLAB Runtime 8.0 2 (Windows 7)
•IronPython 2.7.3
•PyVisa 1.0.0.25
•Microsoft .NET 4.0 Framework
•Microsoft Internet Explorer 6.0 SP1 or later
•Microsoft Photo Editor 3.0 or equivalent software for viewing image files
•Adobe Reader 7.0 or equivalent software for viewing portable document format (PDF) files
Other Devices
•Microsoft compatible mouse or compatible pointing device
•Four USB ports (two USB ports minimum)
•PCI-GPIB or equivalent interface for instrument connectivity
Supported instruments
The TekExpress SFP+ QSFP+ Tx application runs on the following Tektronix oscilloscopes:
•DPO/DSA/MSO71604C and DPO/DSA/MSO72004C Series Digital Oscilloscopes
•DPO/DSA72504D and DPO/DSA73304D Series Digital Oscilloscopes
•DPO/MSO72304DX, DPO/MSO72504DX, and DPO/MSO73304DX Series Digital Oscilloscopes
•DPO70000SX Series Digital Oscilloscopes
3
1
If TekExpress is running on an instrument having a video resolution lower than 800x600 (for example, a sampling oscilloscope), it is recommended that you connect
a secondary monitor, which must be enabled before launching the application.
2
MatLab Runtime Compiler is required for performing the Host Output TWDPc test. This test is purchase option SFP-WDP.
3
If TekExpress is installed on a Tektronix oscilloscope, TekExpress will use the virtual GPIB port for communicating with oscilloscope applications. If external GPIB
communication devices such as USB-GPIB-HS or equivalent are used for instrument connectivity, make sure that the Talker Listener utility is enabled in the GPIB
menu of the DPO/DSA oscilloscope. For ease of use, connect to an external (secondary) monitor.
Complete the following steps to download and install the latest TekExpress SFP+ QSFP+ application.
1. Go to www.tek.com.
2. Click Downloads. In the Downloads menu, select DOWNLOAD TYPE as Software and enter the application name in the MODEL ORKEYWORD field and click SEARCH.
3. Select the latest version of software and follow the instructions to download the software. Copy the executable file into the
oscilloscope.
4. Double-click the executable and follow the on-screen instructions.
The software is installed at C:\Program Files\Tektronix\TekExpress\TekExpress SFP+ QSFP+.
5. Select Application > TekExpress SFP+ QSFP+ from the Oscilloscope menu, to open the application.
Activate the license
Activate the license using the Option Installation wizard in the TekScope application:
1. In the TekScope application menu bar, click Utilities > Option Installation. The TekScope Option Installation wizard opens.
2. Push the F1 key on the oscilloscope keyboard to open the Option Installation help topic.
3. Follow the directions in the help topic to activate the license.
View software version and license key details
To view version information of the application, click Options > About TekExpress.
Set up tests using the tabs in the Setup panel. Settings in the DUT tab use a top-down, left-to-right logic flow, so that any parameter that
affects or acts as a filter for other parameters appears either above or to the left of the affected parameters.
Test options: standards, views, and modes
Test options include two standards, QSFP+ Tx and SFP+ Tx (selected in the DUT tab of the Setup panel), two views, Compliance and
Advanced (selected in the DUT tab of the Setup panel), and two modes, Compliance and User Defined (selected in the configuration
settings of the Setup panel).
The selected standard determines which measurement standard will be used. The selected view determines where the test configuration
settings are displayed. The selected mode determines whether you can change test configuration settings to include those that are
noncompliant.
•Compliance View selected with Compliance Mode
View configuration options in the Test Selection tab of the Setup panel. Tests will run automatically with little or no user intervention.
You will not be able to change test parameters to anything that deviates from the compliance standards. The only test configuration
parameters that you can change in this mode are the Global Settings.
•Compliance View selected with User-Defined Mode
View configuration options in the Test Selection tab of the Setup panel. Tests will run automatically but you will be able to change the
parameters before starting the tests.
•Advanced View selected with Compliance Mode
View configuration options in the Configuration tab of the Setup panel. Tests will run automatically with little or no user intervention. You
will not be able to change test parameters to anything that deviates from the compliance standards. The only configuration parameters
that you can change in this mode are the Global Settings.
•Advanced View selected with User-Defined Mode
View configuration options in the Configuration tab of the Setup panel. Tests will run automatically but you will be able to change test
parameters before starting the tests.
Supported tests
The application supports the following tests, grouped by profile.
Host Profile Tests
•SFF-8431 Table 11 Output Electrical Specifications at B:
•Single Ended Output Voltage Range
•Output AC Common Mode voltage (RMS)
•SFF-8431 Table 12 Jitter and Eye Mask Specifications at B:
•SFF-8431 Table 33 Output Specifications at B for Cu:
Note: The group SFF-8431 Table 33 Output Specifications at B for Cu is associated with the 10GSFP+ Cu Direct Attach
Cable device profile (the check box for this profile is located on the DUT tab of the Setup panel). This group of tests is
available only when this check box is selected.
•Voltage Modulation Amplitude (p-p)
•Transmitter Qsq (for Cu)
•Output AC Common Mode voltage (RMS) (for Cu)
•Host Output TWDPc
Note: An evaluation version of the Host Output TWDPc test is included in the SFP+ QSFP+ Tx installation. You are
allowed 10 free trials of this test. Each time you run the test, regardless of how often you open the application, one of the
free trials is used. To use this test after the evaluation period, purchase the SFP-WDP option license key.
Module Profile Tests
•SFF-8431 Table 16 Transmitter Input Electrical Specifications at B:
•AC Common-Mode Voltage Tolerance
•Single-Ended Input Voltage Tolerance
•SFF-8431 Table 17 Transmitter Input Tolerance Signal Calibrated at B:
Setting up the test environment
•Crosstalk Source Rise/Fall Time (20%-80%)
•Crosstalk Source Amplitude (p-p Differential)
•Output AC Common-Mode Voltage (RMS)
•Total Jitter
•Data Dependent Jitter
•Data Dependent Pulse Width Shrinkage
•Uncorrelated Jitter
•Eye Mask Hit Ratio
•Additional Supported Measurements for QSFP+ only
•Total Jitter @ J2
•Total Jitter @ J9
Instrument connection setup
To run tests, you need the following instrument (for details, see Minimum System Requirements):
TF-QSFP-TPA-PRQSFP+ Host Compliance Board Plug and Module Compliance Board Receptacle
TF-DC-BLOCK-KITDC Block Kit
Equipment connections can vary by Device Profile (Host and Module) and by individual test. Refer to the following connection diagrams
when setting up equipment for testing.
A test setup includes configuration parameters and report options. Use the options in the Setup panel and Reports panel to select and
configure tests.
1. Select the DUT parameters
2. Select one or more tests
3. Select acquisitions
4. Configure test parameters
5. Select test notification preferences
6. Select report options
Search instruments connected to the application
Use the TekExpress Instrument Control Settings dialog box to search the instruments (resources) connected to the application. The
application uses TekVISA to discover the connected instruments.
The instruments required for the test setup must be connected and detected by the application, before running the test.
Note:
To refresh the list of connected instruments:
1. Select Options > Instrument Control Settings.
2. In the Search Criteria section of the Instrument Control Settings dialog box, select the connection types of the instruments to
search. Instrument search is based on the VISA layer, but different connections determine the resource type, such as LAN, GPIB, and
USB. For example, if you choose LAN, the search will include all the instruments supported by the TekExpress that are communicating
over the LAN.
3. Click Refresh. The TekExpress application searches for the connected instruments.
4. When the search is complete, a dialog box lists the instrument-related details based on the search criteria. For example, for the Search
Criteria as GPIB, the application displays all the GPIB instruments connected to the application.
TekExpress Instrument Control Settings window.
The details of the instruments are displayed in the Retrieved Instruments table. The time and date of instrument refresh is displayed in
the Last Updated field.
Running tests
After selecting and configuring tests, review the pre-run checklist and then click Start to perform the tests. The application acquires and
analyzes the data, then displays a report when the tests are finished.
While the tests are running, other applications may display windows in the background. The TekScope application takes precedence over
other applications, but you can switch to other applications by using the Alt+Tab key combination. To keep the TekExpress SFP+ QSFP+
Tx application on top, select Keep On Top from the SFP+ QSFP+ Tx Options menu.
Before you click start
Before you run tests for the first time, do the following:
1. Understand where your test files are stored on the instrument.
After you install and launch TekExpress SFP+ QSFP+ Tx, it creates the following folders on the oscilloscope:
Every time you launch TekExpress SFP+ QSFP+ Tx, an Untitled Session folder is created in the SFP-Tx folder. The
Untitled Session folder is automatically deleted when you exit the SFP+ QSFP+ Tx application. To preserve your test
session files, save the test setup before exiting the TekExpress application.
CAUTION: Do not modify any of the session files or folders because this may result in loss of data or corrupted session files.
Each session has multiple files associated with it. When you save a session, a .TekX file, and a folder named for the session
that contains associated files, is created on the oscilloscope X: drive.
2. Map the shared My TekExpress folder as X: (X drive) on the instruments used in test setups running Microsoft Windows Operating
The My TekExpress folder has the share name format <domain><user ID>My TekExpress. Or, if the instrument is not
connected to a domain, the share name format is <instrument name><user ID>My TekExpress. This shared folder
is used to save the waveform files and is used during other file transfer operations.
Note: If the X: drive is mapped to any other shared folder, the application will display a warning message asking you to
disconnect the X: drive manually.
3. Make sure that the My TekExpress folder has read and write access, and that the contents are not set to be encrypted:
a. Right-click the folder and select Properties.
b. Select the General tab and then click Advanced.
c. In the Advanced Attributes dialog box, make sure that the option Encrypt contents to secure data is NOT selected. Example.
4. See the pre-run checklist before you run a test.
Pre-run checklist
Do the following before you click Start to run a test. If this is the first time you are running a test on a setup, refer to the information in
Before you click start.
1. Make sure that all the required instruments are properly warmed up (approximately 20 minutes).
2. Perform Signal Path Compensation (SPC).
a. On the oscilloscope main menu, select the Utilities menu.
b. Select Instrument Calibration.
3. Deskew channels.
4. Verify that the application is able to find the DUT. If it cannot, perform a search for connected instruments.
a. In SFP+ QSFP+ Tx, select the Setup panel and then click the Test Selection tab.
b. Select any test and then click Configure.
c. In the Configuration section, click Global Settings.
To start the TekExpress SFP+ QSFP+, select from the oscilloscope menu bar Applications > TekExpress SFP+ QSFP+.
During start, a "My TekExpress" folder is created in the Documents folder of the current user and gets mapped to "X" drive. When the
application is closed properly, the "X" drive gets unmapped. Session files are then stored inside the X:\SFP+ QSFP+ folder. If this
file is not found, the application runs an instrument discovery program to detect connected instruments before starting TekExpress SFP+
QSFP+.
To keep the TekExpress SFP+ QSFP+ application on top of any application, select Keep On Top from the options menu. If the application
goes behind the oscilloscope application, select Applications >TekExpress SFP+ QSFP+ to bring the application to the front.
This section describes the application controls with functionality and its details.
Table 5: Application control description
ItemDescription
Options menuMenu to display global application controls.
Test panelControls that open tabs for configuring test settings and options.
Starting the application
Start / Stop buttonUse the Start button to start the test run of the measurements in the selected order. If prior
acquired measurements are not cleared, then new measurements are added to the existing
set. The button toggles to the Stop mode while tests are running. Use the Stop button to
abort the test.
Pause / Continue buttonUse the Pause button to pause the acquisition. When a test is paused, this button changes
as Continue.
Clear buttonUse the Clear button to clear all existing measurement results. Adding or deleting a
measurement, or changing a configuration parameter of an existing measurement, also
clears measurements. This is to prevent the accumulation of measurement statistics or
sets of statistics that are not coherent. This button is available only on Results panel.
Note: This button is visible only when there are results data on the panel.
Application window move iconPlace the cursor over the top of the application window to move the application window to
Mini view / Normal viewMini view displays the run messages with the time stamp, progress bar, Start / Stop button,
and Pause / Continue button. The application moves to mini view when you click the Start
button.
Options menu functions
To access the Options menu, click in the upper-right corner of the application. It has the following selections:
Table 6: Options menu settings
MenuFunction
Default Test SetupOpens a new test setup with default configurations.
Open Test SetupOpens a previously saved test setup. Displays the list of previously saved test setup file
names. Make the selection and click OK to open the test setup.
Save Test SetupSaves the current test configurations with the specified file name.
Save Test Setup AsSaves the current test setup with a different file name or file type.
Open RecentDisplays the recently opened test setup file names. Make the selection and click OK to
open the test setup.
Instrument Control Settings
Keep On Top
Email SettingsConfigures email options for test run and result notifications.
DeskewLoads oscilloscope channel deskew settings into the application.
HelpDisplays the TekExpress SFP+ QSFP+ help.
Table continued…
Detects, lists, and refreshes the connected instruments found on the specified connections
(LAN, GPIB, USB, Serial, Non-VISA Resources, TekLink, and VXI).
Always keeps the TekExpress SFP+ QSFP+ application on top of all the applications.
Use the Email Settings utility to get notified by email when a measurement completes or produces any error condition. Follow the steps to
configure email settings:
Figure 1: Email settings window
1.Select Options > Email Settings to open the Email Settings dialog box.
2.(Required) For Recipient email Address(es), enter one or more recipient email addresses. To include multiple addresses, separate
the addresses with commas.
3.(Required) For Sender’s Address, enter the email address used by the instrument. This address consists of the instrument name,
followed by an underscore, followed by the instrument serial number, then the @ symbol, and the email server ID. For example:
user@yourcompany.com.
4.(Required) In the Server Configuration section, type the SMTP Server address of the Mail server configured at the client location,
and the SMTP Port number, in the corresponding fields.
If this server requires password authentication, enter a valid login name, password, and host name in the corresponding fields.
If any of the above required fields are left blank, the settings will not be saved, and email notifications will not be sent.
Note:
5.In the Email Attachments section, select from the following options:
•Reports: Select to receive the test report with the notification email.
•Status Log: Select to receive the test status log with the notification email. If you select this option, then also select whether you
want to receive the full log or just the last 20 lines.
6.In the Email Configuration section:
•Enter a maximum file size for the email message. Messages with attachments larger than this limit will not be sent. The default is
5 MB.
•Enter the number in the Number of Attempts to Send field, to limit the number of attempts that the system makes to send a
notification. The default is 1. You can also specify a timeout period.
7.Select the Email Test Results When complete or on error check box. Use this check box to quickly enable or disable email
notifications.
Use the TekExpress Instrument Control Settings dialog box to search the instruments (resources) connected to the application. You
can use the Search Criteria options to search the connected instruments depending on the connection type. The details of the connected
instrument is displayed in the Retrieved Instruments window.
To access, click Options > Instrument Control Settings. Select GPIB as search criteria for TekExpress application and click Refresh.
The connected instruments displayed in the Retrieved Instruments window and can be selected for use under Global Settings in the test
configuration section.
Figure 2: TekExpress Instrument Control Settings window
Use the DUT tab to select parameters for the device under test. These settings are global and apply to all tests of current session. DUT
settings also affect the list of available tests in the Test Selection tab.
Figure 6: DUT tab
Click Setup > DUT to access the DUT parameters:
Table 7: DUT tab configuration
SettingDescription
DUT IDAdds an optional text label for the DUT to reports. The default value is DUT001. The maximum
number of characters supported is 32. You cannot use the characters (.,..,...,\,/:?”<>|*) in an ID
name.
Opens a comments dialog box which allows you to enter optional text to add to a report.
Comments icon (to the right of the
DUT ID field)
Acquire live waveformsAcquire active signals from the DUT for measurement and analysis.
Use prerecorded waveform filesRun tests on a saved waveform. Also refer Load a saved test setup.
ViewSelect the View type from the drop-down:
Table continued…
You can enter a maximum number of 256 characters. Refer Select report generation
options on page 47 to enable or disable comments which displays on the test report.
•Compliance
•Advanced
Note: If you select Advanced View, the configuration settings are included in the Setup
steps. If you select Compliance View, the configuration step is not included as a
Use Acquisitions tab to view the test acquisition parameters. The contents displayed on this tab depends on the DUT type and the tests
selected.
Figure 8: Acquisition tab
Table 9: Acquisitions tab configuration
SettingDescription
Data Source (+), Data Source (–)Sets the channels used for positive and negative data sources. When QSFP+
Tx is the selected standard, data sources for each lane to be tested can be
configured.
DeskewProvides the option of setting deskew values on the scope either in an automated
way or manual way.
Test NameDisplays the name of the selected test to which the acquisitions apply. One or
more tests can perform the same acquisitions.
AcquisitionThe acquisition of the waveform for the corresponding test
Signal TypeShows the acquisition parameters of the signal. This column is displayed only if
the Show Acquire Parameters check box is selected in the Acquisition and Save
Options section.
Waveform FileLists the name of the waveform files used for the test. Select waveform files
by clicking the button in the row and selecting any waveform file using the
standard File Open window. This option is available only when Use pre-recordedwaveform files is selected in the Setup panel DUT tab.
Acquisition and Save OptionsAll waveforms are saved before analysis.
Show Acquire ParametersWhen selected, the signal acquisition parameter for each test displays in the
Signal Type column.
Signal ValidationDetermines how the application responds to the source signal
Note: Signal validation is valid only for Live acquisitions
Prompt me if signal fails
Use signal as is - Don’t Check Skips the signal validation process. Does not test the signal for its characteristics
Skip test if signal failsMoves to the next test (if applicable) if the signal fails
TekExpress SFP+ QSFP+ application saves all acquisition waveforms to files by default. Waveforms are saved in a unique folder for each
session (a session is started when you click the Start button). The folder path is X:\TekExpress SFP+ QSFP+\UntitledSession\<dutid>\<date>_<time>. Images created for each analysis, XML files with result values, reports, and other
information specific to that particular execution are also saved in this folder.
If the signal fails, pauses the measurement operation and displays a dialog box
with the following options:
•Reacquire: Try acquiring the signal again.
•Use Anyway: Use the signal even if it cannot be validated.
•Skip Test: Abort the test if the signal cannot be validated.
Saving a session moves the session file contents from the Untitled Session folder to the specified folder name and changes the session
name to the specified name.
Use Configuration tab to view and configure the Global Settings and the measurement configurations. The measurement specific
configurations available in this tab depends on the selections made in the DUT panel and Test Selection panel.
Note: The Use Filter File for De-embedding (only for TWDPC) check box is available only if the test Host Output TWDPC is
selected in the Test Selection tab.
Note: If you change the channel selection for Data+ (DP) and Data– (DN) signals, make sure that you change the corresponding
trigger source in the Analyze tab for each of the measurements. If you do not do this, the waveform might not trigger and the
measurements might not be completed.
Table 10: Configuration tab: Common parameters
SettingsDescription
Limit Editor
Displays the upper and lower limits for the applicable measurement using different types of comparisons.
Table 11: Configuration tab: Global Settings configuration
SettingDescription
Compliance ModeSelect to use Compliance Mode values. You cannot change
most test parameters in Compliance mode but you can view the
parameters.
User Defined ModeSelect to run tests with custom parameters.
Instruments DetectedDisplays the instruments connected to this application. Click on the
instrument name to open a list of available (detected) instruments.
Select Options > Instrument Control Settings to refresh the
connected instrument list refer TekExpress instrument control
settings.
Record LengthSpecifies the length of the record (5M is the default)
Data RateSelect the signal data rate as one of 9.95328 Gbps, 10.3125 Gbps
(default), 10.51875 Gbps, or 11.10 Gbps.
Use Filter File for De-embedding (All Test except TWDPC)When selected, provides a browse button and field for selecting a
filter to use to compensate for cable lengths.
Use Filter File for De-embedding for TWDPCWhen selected, provides a browse button and field for selecting a
Signal TypeSpecifies the signal type of the acquisition for the test selected in the tree view of the
Measurements tab. Running tests in User Defined Mode allows you to perform the tests on
different types of signal. The default signal type varies by test. Not all tests support all the
signal types. For each test, the application includes the signal type options that are best
suited to the measurements.
Window Size (%)Host test Transmitter Qsq only. Set at 10% for Compliance Mode. In User Defined Mode,
Use Preferences tab to set the application action on completion of a measurement. The Preferences tab has the feature to enable or
disable certain options related to the measurement execution.
Figure 11: Preferences tab
Refer the below table for the options available in the Preferences tab:
Table 13: Preferences tab settings
SettingDescription
On Test Failure, stop and notify me of the failure
Select to stop the test run on Test Failure, and to get notified via
email. By default, it is unselected. Click Email Settings to configure
the email settings to receive notifications.
The Status panel contains the Test Status and Log View tabs, which provides status on the test acquisition and analysis (Test Status) and
listing of test tasks performed (Log View tab). The application opens the Test Status tab when you start to execute the test. Select the
Test Status or the Log View tab to view these items while the test execution is in progress.
View test execution status
The tests are grouped and displayed based on the Clock and Data lane. It displays the tests along with the acquisition type, acquire, and
analysis status of the tests. In pre-recorded mode, Acquire Status is not valid.
The Test Status tab presents a collapsible table with information about each test as it is running. Use the symbols to expand (
collapse () the table rows.
) and
Figure 12: Test execution status view in Status panel
Table 14: Test execution status table headers
Table HeaderDescription
Test NameDisplays the measurement name.
AcquisitionDescribes the type of data being acquired.
Acquire StatusDisplays the progress state of the acquisition:
Analysis StatusDisplays the progress state of the analysis:
•To be started
•In progress
•Completed
•Aborted
View test execution logs
The Test Status tab displays the detailed execution status of the tests. Also, displays each and every execution step in detail with its
timestamp information. The log details can be used to troubleshoot and resolve any issue/bug which is blocking the test execution process.
Figure 13: Log view in Status panel
Table 15: Status panel settings
ControlDescription
Message HistoryLists all the executed test operations and timestamp information.
Auto Scroll
Clear LogClears all the messages from the log view.
Save
Show Detailed LogSelect the check box to record a detailed history of test execution.
Enables automatic scrolling of the log view as information is added to the log during the test execution.
Saves the log file into a text file format. Use the standard Save File window to navigate to and specify
the folder and file name to save the log text.
Results panel: View summary of test results
Results panel: View summary of test results
When a test execution is complete, the application automatically opens the Results panel to display a summary of test results.
In the Results table, each test result occupies a row. By default, results are displayed in summary format with the measurement details
collapsed and with the Pass/Fail column visible.
Figure 14: Results panel with measurement results
Note: Results of Crosstalk measurements, Total Jitter @ J2, and Total Jitter @ J9 are not included in the Overall Test Results
because these measurements are informative only. If any of the Crosstalk tests or Total Jitter @ J2 and Total Jitter @ J9 fails, it
will not cause the overall test results to fail.
Click icon on each measurement in the row to expand and to display the minimum and maximum parameter values of the
measurement.
Filter the test results
Each column in the result table can be customized and displayed by enabling or disabling any column as per your requirement. You can
change the view in the following ways:
•To remove or restore the Pass/Fail column, select Preferences > Show Pass/Fail.
Click Reports panel to configure the report generation settings and select the test result information to include in the report. You can use
the Reports panel to configure report generation settings, select test content to include in reports, generate the report, view the report,
browse for reports, name and save reports, and select report viewing options.
Select report generation options
This section describes the report generation settings you can configure in the Reports panel. Select report settings before running a test or
when creating and saving test setups. Report settings configured are included in saved test setups.
Figure 15: Reports panel
Report Update Mode Settings
Table 16: Report Update Mode Settings
ControlDescription
Report Generation
Generate new reportEach time when you click Run and when the test execution is complete, it will create a
new report. The report can be in either .mht, .pdf, or .csv file formats.
Append with previous run sessionAppends the latest test results to the end of the current test results report. Each time
when you click this option and run the tests, it will run the previously failed tests and
replace the failed test result with the new pass test result in the same report.
Replace current test in previous run sessionReplaces the previous test results with the latest test results. Results from newly added
Displays the name and path of the TekExpress SFP+ QSFP+ report. The default location
is at \My Documents>\My TekExpress\SFP+ QSFP+\Reports. The
report file in this folder gets overwritten each time you run a test unless you specify a
unique name or select to auto increment the report name.
To change the report name or location, do one of the following:
•In the Report Path field, type the current folder path and name.
•Double-click in the Report Path field and then make selections from the popup
keyboard and click Enter.
Be sure to include the entire folder path, the file name, and the file extension.
For example: C:\Documents and Settings\your user name\MyDocuments\My TekExpress\SFP+ QSFP+\DUT001.mht.
Note: You cannot set the file location using the Browse button.
Open an existing report
Click Browse, locate and select the report file and then click View at the bottom of the
panel.
Save as typeSaves a report in the specified file type, selected from the drop-down list. The report is
saved in .csv, .pdf, or .mht.
Note: If you select a file type different from the default, be sure to change the
report file name extension in the Report Name field to match.
Auto increment report name if duplicateSets the application to automatically increment the name of the report file if the
application finds a file with the same name as the one being generated. For example:
DUT001, DUT002, DUT003. This option is enabled by default.
Contents To Save
Include pass/fail results summarySelect to include pass/fail information in the details table of the report.
Include detailed resultsSelect to include detailed results in the report.
Include plot imagesSelect to include the plot images in the report.
Include setup configurationSets the application to include hardware and software information in the summary box
at the top of the report. Information includes: the oscilloscope model and serial number,
the oscilloscope firmware version, and software versions for applications used in the
measurements.
Include user commentsSelect to include any comments about the test that you or another user have added in
the DUT tab of the Setup panel. Comments appear in the Comments section, below the
summary box at the beginning of each report.
Group Test Results by
Test NameSelect to group the test results based on the test name in the report.
LaneSelect to group the test results based on the Lanes in the report
Other settings in report panel
View report after generatingAutomatically opens the report in a Web browser when the test execution is complete.
Setup InformationThe summary box at the beginning of the report lists setup configuration information. This information
includes the oscilloscope model and serial number, optical module model and serial number, and software
version numbers of all associated applications.
Test Name Summary TableThe test summary table lists all the tests which are executed with its result status.
MeasurementThe measurement table displays the measurement related details with its parameter value.
User commentsIf you had selected to include comments in the test report, any comments you added in the DUT tab are
You can save the test setup and recall it later for further analysis. Saved setup includes the selected oscilloscope, general parameters,
acquisition parameters, measurement limits, waveforms (if applicable), and other configuration settings. The setup files are saved under
the setup name at X:\TekExpress SFP+ QSFP+
Figure 17: Example of Test Setup File
Use test setups to:
•Recall a saved configuration.
•Run a new session or acquire live waveforms.
•Create a new test setup using an existing one.
•View all the information associated with a saved test, including the log file, the history of the test status as it executed, and the results
summary.
•Run a saved test using saved waveforms.
Images that are shown in this Saving and recalling test setup chapter are for illustration purpose only and it may vary
Note:
depending on the TekExpress application.
Save the configured test setup
You can save a test setup before or after running a test. You can create a test setup from already created test setup or using a default
test setup. When you save a setup, all the parameters, measurement limits, waveform files (if applicable), test selections, and other
configuration settings are saved under the setup name. When you select the default test setup, the parameters are set to the application’s
default value.
•Select Options > Save Test Setup to save the opened setup.
To save a test setup with a different name, follow the steps:
1. Select Options> Save Test Setup As.
Saving and recalling test setup
2. Enter the new test setup name and click Save.
Run a saved test in prerecorded mode
Use this option to rerun a complete test using just the oscilloscope and the saved test setup files, provided that you selected to save the
captured waveforms when you originally ran the saved test.
Note: When you run a saved test in prerecorded mode and then save it under the same name, the test results are saved in a
new session folder named for the date and time of the session. Any test settings that you changed for the session will be saved
as a new test session file and be paired with a folder of the same name. Example. When you open a test setup that has multiple
You can use the Standard Commands for Programmable Instruments (SCPI) to communicate remotely with the TekExpress application.
Complete the TCPIP socket configuration and the TekVISA configuration in the oscilloscope or in the device where you are executing the
script.
Note: If you are using an external PC to execute the remote interface commands, then install TekVISA in the PC to make the
configurations.
Socket configuration for SCPI commands
This section describes the steps to configure the TCPIP socket configuration in your script execution device and the steps to configure the
TekVISA configuration in the oscilloscope to execute the SCPI commands.
TCPIP socket configuration
1. Click Start > Control Panel > System and Security > Windows Firewall > Advanced settings.
2. In Windows Firewall with Advanced Security menu, select Windows Firewall with Advanced Security on Local Computer >
Inbound Rules and click New Rule…
2. Click Search Criteria. In Search Criteria menu, click LAN to Turn-on. Select Socket from the drop-down list, enter the IP address of
the TekExpress device in Hostname and type Port as 5000. Click to configure the IP address with Port.
Enter the Hostname as 127.0.0.1 if the TekVISA and TekExpress application are in the same system, else enter the IP address of the
oscilloscope where the TekExpress application is running.
3. Click Search to setup the TCPIP connection with the host. Check whether the TCPIP host name is displayed in OpenChoice
Instrument Manager > Instruments.
4. Double-click OpenChoice Talker Listener and enter the Command *IDN? in command entry field and click Query. Check that the
Operation is successful and Talker Listener Readout displays the Command / Data.
A session should be run earlier and the report should be generated to get the information of the report.
Syntax
TEKEXP:INFO? REPORT (Query)
Returns
<FileSize>:: <String>
<FileName>:: <String>
Examples
TEKEXP:INFO? REPORT command returns the information of the generated report in the format ("1215”,"DUT001.mht”).
Query the information of the generated waveform files
This command queries the information of the generated waveform files in the format.
<File1Size,"File1Name”>.
If there are more than one waveform, the waveform file names are displayed with the comma separated values in the format
<File1Size,"File1Name">,<File2Size,"File2Name">.
Syntax
TEKEXP:INFO? WFM (Query)
Returns
<FileSize>:: <String>
<FileName>:: <String>
Examples
TEKEXP:INFO? WFM command returns the information of the generated waveform in the format (20000858,"X:\SFP+ QSFP+\Untitled
Session\DUT001\20200916_041609\Iter1_Short Record‐length for SCOPE Period_NoSSC_DIFF.wfm”).
Query the information of the generated image files
This command queries the information of the generated image files in the format.
<File1Size,"File1Name”>.
If there are more than one image, the image file names are displayed with the comma separated values in the format
TEKEXP:INFO? IMAGE command returns the information of the generated image in the format (109058, "X:\SFP+ QSFP+
\Untitled Session\DUT001\20200916_041609\Iter1_Short Record‐length for SCOPE Period_NoSSC_DIFF.png";22794,"X:\SFP+ QSFP+
\UntitledSession\DUT001\20 200916_041609\ScopePeriodPlot_Iteration1WithCursor.png").
Query the active TekExpress application name
This command queries the active TekExpress application name running on the oscilloscope.
Syntax
TEKEXP:*IDN? (Query)
Returns
<String>
Examples
TEKEXP:*IDN? command returns the active TekExpress application name running on the oscilloscope.
Set or query the DUTID of application
This command sets or queries the DUTID of the application.
Syntax
TEKEXP:VALUE DUTID,"<Value>" (Set)
TEKEXP:VALUE? DUTID (Query)
Command arguments
Argument NameArgument Type
<Value><String>
Returns
<String>
Examples
TEKEXP:VALUE DUTID,"DUT001" command sets the DUTID of the application to DUT001.
TEKEXP:VALUE? DUTID command returns the DUTID of the application.
Sets or query the acquire mode status
This command sets or queries the acquire mode status.
TEKEXP:VALUE LIMIT,<TestName>,<LimitHeader>,<Value1>,<CompareString>,<Value2> command
sets the limits value for the specified testname and limit header.
TEKEXP:VALUE? LIMIT,<TestName>,<LimitHeader> command returns the limits value for the specified testname and
limit header.
Set or query the waveform file recalled for the specified test name and acquire
type
This command set or queries the waveform file recalled for the specified test name and acquire type.
If there are more than one waveform, the waveform file names are displayed with the symbol "$" separated values in the format
TEKEXP:VALUE WFMFILE,<TestName>,<AquireType>,<WaveformFileName> command recalls the sepcified
waveform file for the specified testname and acquire type.
TEKEXP:VALUE? WFMFILE,<TestName>,<AquireType> command returns the waveform file name recalled for the
specified testname and acquire type.
Set or query the enable/disable status of Verbose function
This command sets or queries the enable/disable status of Verbose function.
SameSession_ClearResults - Clears the test results of the current
session and starts the test execution. The session results will be
added in the same session, by erasing the previous run results.
Returns
NewSession | SameSession_ClearResults
Examples
TEKEXP:VALUE? ContinuousRun_RunSessionOptions command returns the option for session creation in the
continuous run function.
TEKEXP:VALUE ContinuousRun_RunSessionOptions,"Value" command sets the option for session creation in the
continuous run function.
Set or query the View report after generating option status
This command sets or queries the enable/disable status of the View report after generating function.
Syntax
TEKEXP:VALUE? GENERAL,"View Report After Generating" (Query)
TEKEXP:VALUE GENERAL,"View Report After Generating",<value> (Set)
Arguments
Argument NameArgument value
<Value>{True | False} or {1 | 0}
It represents enabled or disabled.
Where,
•True or 1 - enabled
•False or 0 - disabled
Returns
{True | False} or {0 | 1}
Examples
TEKEXP:VALUE? GENERAL,"View Report After Generating" command returns the enable or disable status of
view report after generating option.
Eye Mask Hit RatioPRBS31PRBS7, PRBS9, PRBS11, PRBS15, PRBS20, PRBS23, PRBS31
User-Defined mode support
8180
PRBS7, PRBS9, PRBS11, PRBS15
Oscilloscope setup
The following tables show the generalized oscilloscope setup parameters for all tests, except where noted.
Table 20: Horizontal setup
ModeSample RateRecord LengthAcquisition Mode
Manual100 GS/s5M, 10M100 GS/s
Table 21: Vertical setup
Channel 1Channel 2
PositionScaleOffsetPositionScaleOffset
0Vertical setup
procedure
00Vertical setup
procedure
0
Signal validation procedure
Each measurement goes through signal validation, signal acquisition, and measurement steps. Before running any measurement, check
the signal for validity. To do this, set up the oscilloscope to acquire the signal and then test the signal for its characteristics such as
voltage level and/or frequency. Once the characteristics of the signal are correct, the signal is valid and ready for use in compliance
measurements. The signal validation procedure is the same for both Host and Module Profile measurements.
SFP+ solution supports three types of signals: PRBS9, PRBS31, and 8180.
PRBS20, PRBS23, PRBS31 pattern validation is not done. The signal is assumed to be PRBS20, PRBS23, or PRBS31
Note:
respectively.
Acquire the signal according to the SFF-8431 specification. The oscilloscope setup differs for each measurement. For all measurements,
use the following guidelines to set up the oscilloscope. For the oscilloscope setup for individual tests, see the individual test topics.
After the signal is acquired, the measurement is conducted using either the base oscilloscope, DPOJET, or the TekExpress algorithm
library.
The following describes the oscilloscope setup and signal validation measurement for PRBS7, PRBS9, PRBS11, PRBS15, and 8180
signals.
1. Connect the input signal to the appropriate channels. (The channels are those selected as data sources in the Acquisitions tab of the
Setup panel.)
The vertical setup routine runs automatically to find the best fitting vertical scale.
Horizontal setup uses these parameters:
•Mode: Manual
•Sample Rate: 100 GS/s
•Record Length: 5M/10M
•Acquisition Mode: Sample
2. On the oscilloscope, set Math1 = (Ch1 – Ch3).
3. Call the Signal Validation algorithm to verify the signal type.
The signal validation results are returned from the algorithm. If signal validation fails, display a signal validation message.
Host test: SFF-8431 table 11 tests
Single-ended output voltage range
Single-ended output voltage range is measured as the single-ended peak-to-peak output voltage for the positive and negative data
channels on a single acquisition. This test is one of the SFF-8431 Table 11 Output Electrical Specifications at B tests.
Connect the equipment as shown in the diagram at Connection Setup for Host Tests: All other Host tests.
Signal validation
Signal validation for PRBS31 pattern signals is not done by the SFP+ QSFP+ Tx application. The software assumes that the PRBS31
pattern is valid.
Measurement procedure
Single-ended output voltage range is measured using base DPOJET.
•Use DPOJET measurement name: Ampl > Max and Ampl > Min for Dp and Dn
Limits
For each of Dp and Dn: Minimum = –0.3 V; Maximum = 4 V
Snapshot and cursor placement details
There will be two snapshots:
•A snapshot of Dp with cursors placed at maximum and minimum voltages of single-ended Pk-Pk measurement value of Dp. Cursor
values can be derived using DPOJET: Ampl > Cycle Max and Ampl > Cycle Min for +ve and –ve peak, respectively for Dp.
•A snapshot of Dn with cursors placed at maximum and minimum voltages of single-ended Pk-PK measurement value of Dn. Cursor
values can be derived using DPOJET: Ampl > Cycle Max and Ampl > Cycle Min for +ve and –ve peak, respectively for Dn.
Output AC Common-Mode voltage (RMS) (Host)
Output AC common-mode voltage is measured as the AC RMS voltage of the common mode ((DP+DN) ÷ 2) signal. This test is one of the
SFF-8431 Table 11 Output Electrical Specifications at B tests.
Connect the equipment as shown in the diagram at Connection Setup for Host Tests: All other Host tests.
Signal validation
Signal validation for PRBS31 pattern signals is not done by the SFP+ QSFP+ Tx application. The software assumes that the PRBS31
pattern is valid.
Measurement procedure
Output AC common-mode voltage is measured using the base oscilloscope.
1. Math1 = (Dp + Dn) ÷ 2. Switch off channels connected to Dp and Dn.
2. Draw a histogram bounding box from the top left of the oscilloscope to the bottom right, and choose Histogram Vertical.
3. Select Math1 as the histogram source.
4. From the Measure menu, select Histogram Measurements > Standard Deviation.
5. Do a single step run and measure the mean value of the histogram standard deviation. This is the measurement result.
Limits
Maximum = 15 mV
Snapshot and cursor placement details
A snapshot where the histogram on math and the standard deviation value of the histogram is visible on the screen. The standard
deviation value visible on the screen should be the value reported in the report. No cursors are required.
Host test: SFF-8431 table 12 tests
Crosstalk source Rise/Fall time (20% - 80%) (Host)
Crosstalk source rise/fall time is measured as the 20% – 80% (of VMA) Rise time and 80% – 20% (of VMA) Fall time of the differential
Crosstalk Source introduced to the signal. This test is one of the SFF-8431 Table 12 Jitter and Eye Mask Specifications at B tests.
Connect the equipment as shown in the diagram at Connection Setup for Host Tests: Table 12 Crosstalk Source Tests.
Signal validation
For compliance, perform Signal Validation Procedure.
Measurement procedure
The crosstalk source rise/fall time measurement is performed using the algorithm library.
Inputs to algorithm
•Differential waveform obtained on a Math channel using Dp and Dn after deskewing and applying 12 GHz filter
•Signal Type ("SFPPlus.SignalType")
Algorithm
•Input the differential waveform.
•Obtain values of vHigh and vLow as done in the VMA measurement (averaged over all the patterns found in the waveform).
•Find the position of the rising and falling edges in the input waveform.
•On each rising edge, traverse forward until the first point crossing the upper limit (80%) is found (= End) and traverse backward until
the first point crossing the lower limit (20%) is found (= Start).
•On each falling edge, traverse backward until the first point crossing the upper limit (80%) is found (= Start) and traverse forward until
the first point crossing the lower limit (20%) is found (= End).
Crosstalk source amplitude is measured as the difference between the nominal one and zero levels, which are defined as voltages
measured in the mid 20% of the high (eight 1s) and low (eight 0s) regions respectively, of the differential crosstalk source introduced to the
signal. This test is one of the SFF-8431 Table 12 Jitter and Eye Mask Specifications at B tests.
Connect the equipment as shown in the diagram at Connection Setup for Host Tests: Table 12 Crosstalk Source Tests.
Signal validation
For compliance, perform Signal Validation Procedure.
Measurement procedure
The crosstalk source amplitude (p-p differential) measurement is performed using the algorithm library.
Inputs to algorithm
Differential waveform obtained on a Math channel using Dp and Dn after deskewing and applying 12 GHz filter
Algorithm
•Input the differential waveform.
•On the input waveform, identify the regions with the pattern 11111111 or 00000000.
•On each such pattern, measure the mean voltage in the middle 20% of the pattern.
•The mean voltages corresponding to the Signal High regions is taken as vHigh, and the mean voltages corresponding to the Signal
Low is taken as vLow. The vHigh and vLow values are averaged across all regions.
•The required result Amplitude is given by, Amplitude = vHigh – vLow.
•End
Limits
Target value = 1000 mV
Snapshot and cursor placement details
This measurement has one snapshot with cursor1 = +ve high of VMA, cursor2 = –ve high of VMA.
Signal Rise/Fall time (20% - 80%)
Signal rise/fall time is measured as the 20% – 80% (of VMA) Rise time and 80% – 20% (of VMA) Fall time of the input signal. This test is
one of the SFF-8431 Table 12 Jitter and Eye Mask Specifications at B tests.
Connect the equipment as shown in the diagram at Connection Setup for Host Tests: All other Host tests.
For compliance, perform Signal Validation Procedure.
Measurement procedure
The signal rise/fall time (20% – 80%) measurement is performed using the algorithm library.
Inputs to algorithm
•Differential waveform obtained on a Math channel using Dp and Dn after deskewing and applying 12 GHz filter
•Signal Type ("SFPPlus.SignalType")
Algorithm
•Obtain values of vHigh and vLow as done in the “VMA” measurement. (Averaged over all the patterns found in the waveform.)
•Find the position of rising and falling edges in the input waveform.
•On each rising edge, traverse forward until the first point crossing the upper limit (80%) is found (=End) and traverse backward until the
first point crossing the lower limit (20%) is found (= Start).
•On each falling edge, traverse backward until the first point crossing the upper limit (80%) is found (=Start) and traverse forward until
the first point crossing the lower limit (20%) is found (=End).
•Compute Transition Time, TT = (End – Start) This value is averaged across all edges.
•End
Limits
Minimum value = 34 ps
Snapshot and cursor placement details
This measurement has two snapshots: first with cursors placed at the rise time, second with cursors placed at the fall time.
•Rise time: Zoom to the cursor positions. Cursor1 = starting point of RT of first rising edge, cursor2 = ending point of RT of first rising
edge.
•Fall time: Fall time: Zoom to the cursor positions. Cursor1 = starting point of FT of first falling edge, cursor2 = ending point of FT of first
falling edge.
Total jitter (Host)
TJ, as used here, is the Level 1 definition for Total Jitter as described in the FC-MJSQ, where TJ is the crossing width, defined as the late
time at which the BER is 10–12 minus the early time at which the BER is 10–12. This test is one of the SFF-8431 Table 12 Jitter and Eye
Mask Specifications at B tests.
Connect the equipment as shown in the diagram at Connection Setup for Host Tests: All other Host tests.
Signal validation
Signal validation for PRBS31 pattern signals is not done by the SFP+ QSFP+ Tx application. The software assumes that the PRBS31
pattern is valid.
Measurement procedure
Total jitter is measured using base DPOJET.
•Deskew the Data+ and Data- inputs. Compute the differential signal on Math1 = Dp – Dn.
•Use DPOJET TJ@BER measurement on the Math signal and Skew measurement on the single-ended input signals.