Tektronix TBS2104, TBS2102, TBS2074, TBS2072 Performance Verification

TBS2000 Series Oscilloscopes
Specification and Performance Verification
*P077114800*
077-1148-00
TBS2000 Series Oscilloscopes
Specification and Performance Verification
Warning
The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
Supports Product Firmware V1.0 and above
www.tek.com
077-1148-00
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tek.com to find contacts in your area.

Table of Contents

Important safety information ................................................................................................................................ iii
General safety summary ................................................................................................................................ iii
Service safety summary ................................................................................................................................ vi
Terms in the manual ...................................................................................................................................... vi
Terms on the product .................................................................................................................................... vi
Symbols on the product ................................................................................................................................ vii
Preface ................................................................................................................................................................ ix
Specifications
Model overview .............................................................................................................................................. 1
Vertical system analog channels .................................................................................................................... 1
Horizontal system analog channels ............................................................................................................... 3
Trigger system ............................................................................................................................................... 6
Input output ports ........................................................................................................................................... 7
Data storage ................................................................................................................................................... 8
Display system ............................................................................................................................................... 8
Power source ................................................................................................................................................. 8
Physical characteristics .................................................................................................................................. 9
EMC, environment, and safety ....................................................................................................................... 9
Performance verification
Required equipment ..................................................................................................................................... 11
Test record ................................................................................................................................................... 12
Performance verification procedures ........................................................................................................... 13
Self test ........................................................................................................................................................ 14
Signal path compensation (SPC) ................................................................................................................. 14
Check DC balance ....................................................................................................................................... 14
Check DC gain accuracy .............................................................................................................................. 15
Check bandwidth .......................................................................................................................................... 17
Check vertical offset accuracy ..................................................................................................................... 18
Check sample rate and horizontal position time accuracy ........................................................................... 19
TBS2000 Series Specification and Performance Verification i
Table of Contents
ii TBS2000 Series Specification and Performance Verification

Important safety information

This manual contains information and warnings that must be followed by the user for safe operation and to keep the product in a safe condition.
To safely perform service on this product, see the Service safety summary that follows the General safety summary.

General safety summary

Use the product only as specified. Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. Carefully read all instructions. Retain these instructions for future reference.
Comply with local and national safety codes.
For correct and safe operation of the product, it is essential that you follow generally accepted safety procedures in addition to the safety precautions specified in this manual.
The product is designed to be used by trained personnel only.
Only qualified personnel who are aware of the hazards involved should remove the cover for repair, maintenance, or adjustment.
Before use, always check the product with a known source to be sure it is operating correctly.
This product is not intended for detection of hazardous voltages.
Use personal protective equipment to prevent shock and arc blast injury where hazardous live conductors are exposed.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals for warnings and cautions related to operating the system.
When incorporating this equipment into a system, the safety of that system is the responsibility of the assembler of the system.
To avoid fire or personal injury
Use proper power cord. Use only the power cord specified for this product and certified for the country of use.
Use proper power cord. Use only the power cord specified for this product and certified for the country of use. Do not use the
provided power cord for other products.
Use proper voltage setting. Before applying power, ensure that the line selector is in the proper position for the source being used.
Ground the product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded. Do not disable the power cord grounding connection.
Ground the product. This product is indirectly grounded through the grounding conductor of the mainframe power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded. Do not disable the power cord grounding connection.
Power disconnect. The power switch disconnects the product from the power source. See instructions for the location. Do not position the equipment so that it is difficult to disconnect the power switch; it must remain accessible to the user at all times to allow for quick disconnection if needed.
Power disconnect. The power cord disconnects the product from the power source. See instructions for the location. Do not position the equipment so that it is difficult to operate the power cord; it must remain accessible to the user at all times to allow for quick disconnection if needed.
Use proper AC adapter. Use only the AC adapter specified for this product.
TBS2000 Series Specification and Performance Verification iii
Important safety information
Connect and disconnect properly. Do not connect or disconnect probes or test leads while they are connected to a voltage source. Use only insulated voltage probes, test leads, and adapters supplied with the product, or indicated by Tektronix to be suitable for the product.
Observe all terminal ratings. To avoid fire or shock hazard, observe all rating and markings on the product. Consult the product manual for further ratings information before making connections to the product. Do not exceed the Measurement Category (CAT) rating and voltage or current rating of the lowest rated individual component of a product, probe, or accessory. Use caution when using 1:1 test leads because the probe tip voltage is directly transmitted to the product.
Observe all terminal ratings. To avoid fire or shock hazard, observe all rating and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Do not float the common terminal above the rated voltage for that terminal.
The measurement terminals on this product are not rated for connection to mains or Category II, III, or IV circuits.
Do not operate without covers. Do not operate this product with covers or panels removed, or with the case open. Hazardous voltage exposure is possible.
Avoid exposed circuitry. Do not touch exposed connections and components when power is present.
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by qualified
service personnel.
Disable the product if it is damaged. Do not use the product if it is damaged or operates incorrectly. If in doubt about safety of the product, turn it off and disconnect the power cord. Clearly mark the product to prevent its further operation.
Before use, inspect voltage probes, test leads, and accessories for mechanical damage and replace when damaged. Do not use probes or test leads if they are damaged, if there is exposed metal, or if a wear indicator shows.
Examine the exterior of the product before you use it. Look for cracks or missing pieces.
Use only specified replacement parts.
Replace batteries properly. Replace batteries only with the specified type and rating.
Recharge batteries properly. Recharge batteries for the recommended charge cycle only.
Use proper fuse. Use only the fuse type and rating specified for this product.
Wear eye protection. Wear eye protection if exposure to high-intensity rays or laser radiation exists.
Do not operate in wet/damp conditions. Be aware that condensation may occur if a unit is moved from a cold to a warm
environment.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry. Remove the input signals before you clean the product.
Provide proper ventilation. Refer to the manual's installation instructions for details on installing the product so it has proper
ventilation.
Slots and openings are provided for ventilation and should never be covered or otherwise obstructed. Do not push objects into any of the openings.
Provide a safe working environment. Always place the product in a location convenient for viewing the display and indicators.
Avoid improper or prolonged use of keyboards, pointers, and button pads. Improper or prolonged keyboard or pointer use may result in serious injury.
iv TBS2000 Series Specification and Performance Verification
Important safety information
Be sure your work area meets applicable ergonomic standards. Consult with an ergonomics professional to avoid stress injuries.
Use care when lifting and carrying the product. This product is provided with handles for lifting and carrying.
WARNING. The product is heavy. To reduce the risk of personal injury or damage to the device get help when lifting or carrying the product.
WARNING. The product is heavy. Use a two-person lift or a mechanical aid.
Use only the Tektronix rackmount hardware specified for this product.
Probes and test leads
Before connecting probes or test leads, connect the power cord from the power connector to a properly grounded power outlet.
Keep fingers behind the finger guards on the probes.
Remove all probes, test leads and accessories that are not in use.
Use only correct Measurement Category (CAT), voltage, temperature, altitude, and amperage rated probes, test leads, and adapters for any measurement.
Beware of high voltages. Understand the voltage ratings for the probe you are using and do not exceed those ratings. Two ratings are important to know and understand:
The maximum measurement voltage from the probe tip to the probe reference lead.
The maximum floating voltage from the probe reference lead to earth ground.
These two voltage ratings depend on the probe and your application. Refer to the Specifications section of the manual for more information.
WARNING. To prevent electrical shock, do not exceed the maximum measurement or maximum floating voltage for the oscilloscope input BNC connector, probe tip, or probe reference lead.
Connect and disconnect properly. Connect the probe output to the measurement product before connecting the probe to the circuit under test. Connect the probe reference lead to the circuit under test before connecting the probe input. Disconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe from the measurement product.
Connect and disconnect properly. De-energize the circuit under test before connecting or disconnecting the current probe.
Connect the probe reference lead to earth ground only.
Do not connect a current probe to any wire that carries voltages above the current probe voltage rating.
Inspect the probe and accessories. Before each use, inspect probe and accessories for damage (cuts, tears, or defects in the probe body, accessories, or cable jacket). Do not use if damaged.
Ground-referenced oscilloscope use. Do not float the reference lead of this probe when using with ground-referenced oscilloscopes. The reference lead must be connected to earth potential (0 V).
Floating measurement use. Do not float the reference lead of this probe above the rated float voltage.
TBS2000 Series Specification and Performance Verification v
Important safety information
Risk assessment warnings and information

Service safety summary

The Service safety summary section contains additional information required to safely perform service on the product. Only qualified personnel should perform service procedures. Read this Service safety summary and the General safety summary before performing any service procedures.
To avoid electric shock. Do not touch exposed connections.
Do not service alone. Do not perform internal service or adjustments of this product unless another person capable of rendering
first aid and resuscitation is present.
Disconnect power. To avoid electric shock, switch off the product power and disconnect the power cord from the mains power before removing any covers or panels, or opening the case for servicing.
Use care when servicing with power on. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
Verify safety after repair. Always recheck ground continuity and mains dielectric strength after performing a repair.

Terms in the manual

These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.

Terms on the product

These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
vi TBS2000 Series Specification and Performance Verification
Important safety information

Symbols on the product

When this symbol is marked on the product, be sure to consult the manual to find out the nature of the potential hazards and any actions which have to be taken to avoid them. (This symbol may also be used to refer the user to ratings in the manual.)
The following symbols may appear on the product:
TBS2000 Series Specification and Performance Verification vii
Important safety information
viii TBS2000 Series Specification and Performance Verification

Preface

This manual contains specification and performance verification information for the TBS2000 Series Digital Storage Oscilloscopes.
TBS2000 Series Specification and Performance Verification ix
Preface
x TBS2000 Series Specification and Performance Verification

Specifications

All specifications apply to all models unless noted otherwise.

Model overview

TBS2072 TBS2102 TBS2074 TBS2104
Analog channels 2 2 4 4
Bandwidth 70 MHz 100 MHz 70 MHz 100 MHz
Sample rate 1 GS/s 1 GS/s 1 GS/s 1 GS/s
Record length 20 M points 20 M points 20 M points 20 M points

Vertical system analog channels

Hardware bandwidth limits 20 MHz
Input coupling DC, AC, or GND
Input impedance 1 MΩ ± 2 %, 11.5 pF ± 2.5 pF
Sensitivity range 2 mV/Div to 5 V/Div in 1-2-5 sequence with probe attenuation set to 1X.
Vertical resolution 8 bits
Maximum input voltage 300 V RMS; derate above 4 MHz to 6 V RMS at 200 MHz.
Based upon sinusoidal or DC input signal. Maximum viewable signal while DC coupled is ±100 V offset ± 5 V/div at 4 divisions, or 120 V. AC coupling allows measuring signals on a DC level up to 300V. For non-sinusoidal waveforms, peak value must be less than 450 V. Excursion above 300 V should be less than 100 ms duration. RMS signal level must be limited to 300 V. If these values are exceeded, damage to the instrument may result.
Number of digitized bits 8 bits
Acquisition modes Sample, Peak Detect, Average, and Roll
TBS2000 Series Specification and Performance Verification 1
Specifications Vertical system analog channels (cont.)
Math modes
All units: Ch 1 - Ch 2
Ch 2 - Ch 1
Ch 1 + Ch 2
Ch 1 X Ch 2
FFT
4 channel units: Ch 3 - Ch 4
Ch 3 + Ch 4
Ch 4 - Ch 3
Ch 3 X Ch 4
DC balance ± (1 mV +0.1 div)
DC gain accuracy ± 3% 10 mV/div through 5 V/div, derated at 0.05%/ °C above 30 °C.
± 4% typical 2 mV/div and 5 mV/div
DC voltage measurement accuracy average mode
Average of > 16 waveforms ±((DC Gain Accuracy) X |reading - (offset - position)| + Offset Accuracy + 0.11 div + 1 mV)
Delta Volts between any two averages of ≥16 waveforms acquired with the same oscilloscope setup and ambient conditions
Vertical position range ± 5 divisions
Vertical offset ranges
Vertical offset accuracy ± (0.01 X |offset - position | + DC Balance)
Analog bandwidth, DC coupled
100 MHz models: DC to ≥100 MHz for 2 mV/div through 5 V/div.
70 MHz models: DC to ≥70 MHz for 2 mV/div through 5 V/div.
Upper-Frequency limit, 20 MHz bandwidth limited
±(DC Gain Accuracy X |reading| + 0.08 div + 1.4 mV)
Volts/Div setting Offset range, 1 MΩ
2 mV/div to 200 mV/div ± 0.8 V
> 200 mV/div to 5 V/div ± 20 V
≥ 20 MHz
Because the digital triggering system uses data that has been BW limited, all Trigger functions on the BW limited analog channel are affected. Each channel is separately limited, allowing different bandwidths on different channels of the same instrument.
Lower-Frequency limit, AC coupled, typical
2 TBS2000 Series Specification and Performance Verification
<10 Hz
≤1 Hz when 10X, passive probes are used.
Vertical system analog channels (cont.) Specifications
Rise time, typical 3.5 ns for 100 MHz Models.
5 ns for 70 MHz Models.
Common mode rejection ratio (CMMR), typical
Crosstalk (channel isolation) 100 MHz Models:
100:1 at 60 Hz, reducing to 10:1 with 50 MHz sine wave with equal Volts/div and coupling settings on each channel.
>100:1 with 100 MHz sine wave and with equal V/div settings on each channel.
70 MHz Models:
>100:1 with 70 MHz sine wave and with equal V/div settings on each channel.

Horizontal system analog channels

Sample rate TBS207x, TBS210x: 500 MS/s all channels, 1 GS/s half channels (at most one channel from each
pair {channel1, channel2} and {channel3, channel4} is active);
Sample rate for time/div versus record length (1 GS/s)
Time/Div Real Time Sampling Rate = 1 GS/s
RL= 2 k RL= 20 k RL= 200 k RL= 2 M RL= 20 M RL= Auto
2 ns 1 GS/s 1 GS/s 1 GS/s 1 GS/s 1 GS/s 1 GS/s
4 ns 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
10 ns 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
20 ns 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
40 ns 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
100 ns 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
200 ns 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
400 ns 250 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
1 μs 125 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
2 μs 62.5 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
4 μs 31.25 MS/s 250 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
10 μs 12.5 MS/s 125 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
20 μs 6.25 MS/s 62.5 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
40 μs 3.125 MS/s 31.25 MS/s 250 MS/s 500 MS/s 500 MS/s 500 MS/s
100 μs 1.25 MS/s 12.5 MS/s 125 MS/s 500 MS/s 500 MS/s 500 MS/s
200 μs 625 kS/s 6.25 MS/s 62.5 MS/s 500 MS/s 500 MS/s 500 MS/s
400 μs 312.5 kS/s 3.125 MS/s 31.25 MS/s 250 MS/s 500 MS/s 500 MS/s
1 ms 125 kS/s 1.25 MS/s 12.5 MS/s 125 MS/s 500 MS/s 500 MS/s
2 ms 62.5 kS/s 625 kS/s 6.25 MS/s 62.5 MS/s 500 MS/s 500 MS/s
4 ms 31.25 kS/s 312.5 kS/s 3.125 MS/s 31.25 MS/s 250 MS/s 250 MS/s
10 ms 12.5 kS/s 125 kS/s 1.25 MS/s 12.5 MS/s 125 MS/s 125 MS/s
20 ms 6.25 kS/s 62.5 kS/s 625 kS/s 6.25 MS/s 62.5 MS/s 62.5 MS/s
TBS2000 Series Specification and Performance Verification 3
Specifications Horizontal system analog channels (cont.)
Time/Div Real Time Sampling Rate = 1 GS/s
RL= 2 k RL= 20 k RL= 200 k RL= 2 M RL= 20 M RL= Auto
40 ms 3.125 kS/s 31.25 kS/s 312.5 kS/s 3.125 MS/s 31.25 MS/s 31.25 MS/s
100 ms 1.25 kS/s 12.5 kS/s 125 kS/s 1.25 MS/s 12.5 MS/s 12.5 MS/s
200 ms 625 S/s 6.25 kS/s 62.5 kS/s 625 kS/s 6.25 MS/s 6.25 MS/s
400 ms 313 S/s 3.125 kS/s 31.25 kS/s 312.5 kS/s 3.125 MS/s 3.125 MS/s
1 s 125 S/s 1.25 kS/s 12.5 kS/s 125 kS/s 1.25 MS/s 1.25 MS/s
2 s 63 S/s 625 S/s 6.25 kS/s 62.5 kS/s 625 kS/s 625 kS/s
4 s 31 S/s 313 S/s 3.125 kS/s 31.25 kS/s 312.5 kS/s 312.5 kS/s
10 s 13 S/s 125 S/s 1.25 kS/s 12.5 kS/s 125 kS/s 125 kS/s
20 s 6 S/s 63 S/s 625 S/s 6.25 kS/s 62.5 kS/s 62.5 kS/s
40 s 3 S/s 31 S/s 313 S/s 3.125 kS/s 31.25 kS/s 31.25 kS/s
100 s 1 S/s 13 S/s 125 S/s 1.25 kS/s 12.5 kS/s 12.5 kS/s
Sample rate for time/div versus record length (500 MS/s)
Time/Div Real Time Sampling Rate = 500 MS/s
RL= 2 k RL= 20 k RL= 200 k RL= 2 M RL= 20 M RL= Auto
1 ns
2 ns
4 ns 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
10 ns 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
20 ns 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
40 ns 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
100 ns 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
200 ns 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
400 ns 250 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
1 μs 125 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
2 μs 62.5 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
4 μs 31.25 MS/s 250 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
10 μs 12.5 MS/s 125 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
20 μs 6.25 MS/s 62.5 MS/s 500 MS/s 500 MS/s 500 MS/s 500 MS/s
40 μs 3.125 MS/s 31.25 MS/s 250 MS/s 500 MS/s 500 MS/s 500 MS/s
100 μs 1.25 MS/s 12.5 MS/s 125 MS/s 500 MS/s 500 MS/s 500 MS/s
200 μs 625 kS/s 6.25 MS/s 62.5 MS/s 500 MS/s 500 MS/s 500 MS/s
400 μs 312.5 kS/s 3.125 MS/s 31.25 MS/s 250 MS/s 500 MS/s 500 MS/s
1 ms 125 kS/s 1.25 MS/s 12.5 MS/s 125 MS/s 500 MS/s 125 MS/s
2 ms 62.5 kS/s 625 kS/s 6.25 MS/s 62.5 MS/s 500 MS/s 62.5 MS/s
4 ms 31.25 kS/s 312.5 kS/s 3.125 MS/s 31.25 MS/s 250 MS/s 250 MS/s
10 ms 12.5 kS/s 125 kS/s 1.25 MS/s 12.5 MS/s 125 MS/s 125 MS/s
20 ms 6.25 kS/s 62.5 kS/s 625 kS/s 6.25 MS/s 62.5 MS/s 62.5 MS/s
4 TBS2000 Series Specification and Performance Verification
Horizontal system analog channels (cont.) Specifications
Time/Div Real Time Sampling Rate = 500 MS/s
RL= 2 k RL= 20 k RL= 200 k RL= 2 M RL= 20 M RL= Auto
40 ms 3.125 kS/s 31.25 kS/s 312.5 kS/s 3.125 MS/s 31.25 MS/s 31.25 MS/s
100 ms 1.25 kS/s 12.5 kS/s 125 kS/s 1.25 MS/s 12.5 MS/s 12.5 MS/s
200 ms 625 S/s 6.25 kS/s 62.5 kS/s 625 kS/s 6.25 MS/s 6.25 MS/s
400 ms 313 S/s 3.125 kS/s 31.25 kS/s 312.5 kS/s 3.125 MS/s 3.125 MS/s
1 s 125 S/s 1.25 kS/s 12.5 kS/s 125 kS/s 1.25 MS/s 1.25 MS/s
2 s 63 S/s 625 S/s 6.25 kS/s 62.5 kS/s 625 kS/s 625 kS/s
4 s 31 S/s 313 S/s 3.125 kS/s 31.25 kS/s 312.5 kS/s 312.5 kS/s
10 s 13 S/s 125 S/s 1.25 kS/s 12.5 kS/s 125 kS/s 125 kS/s
20 s 6 S/s 63 S/s 625 S/s 6.25 kS/s 62.5 kS/s 62.5 kS/s
40 s 3 S/s 31 S/s 313 S/s 3.125 kS/s 31.25 kS/s 31.25 kS/s
100 s 1 S/s 13 S/s 125 S/s 1.25 kS/s 12.5 kS/s 12.5 kS/s
Waveform interpolation Only (sin x)/x interpolation is provided.
Waveform interpolation is activated for sweep speeds of 100 ns/div and faster
Record length 20 M, 2 M, 200 k, 20 k, 2 k samples per record, user selectable or in the AUTO mode automatically
select the shortest record length which supports the highest sample rate available for the Time/Div settings.
Seconds division range TBS207x, TBS210x: 2 ns/div to 100 sec/div in a 1-2-4 sequence
Deskew range Analog channels only:
TBS207x, TBS210x: ±100 ns with 2 ns resolution
Long term sample rate and
horizontal position time accuracy
Delta time measurement accuracy The limits are given in the following table for signals having amplitude ≥ 7 divisions, slew rate at the
± 25 x 10-6 over any ≥ 1 ms interval.
measurement points of ≥ 2.0 divisions/ns, and acquired at ≥ 10 mV/Div:
Condition Time Measurement Accuracy
Single shot, full bandwidth selected ± (1 Sample Interval + 25 X 10-6 X |reading| +
0.6 ns)
> 16 averages, full bandwidth selected ± (1 Sample Interval + 25 X 10-6 X |reading| +
0.4 ns)
TBS2000 Series Specification and Performance Verification 5
Specifications

Trigger system

Waveform update rate, typical Minimum triggered acquisition rate is 10,000 wfm/sec.
Trigger types Edge, Pulse Width, and Runt.
Trigger source Analog channels and AC Line
Trigger coupling analog channels DC, Noise Reject, High Freq Reject, Low Freq Reject.
Line trigger characteristics Line Trigger mode provides a source to synchronize the trigger with the AC line input.
Matches the AC power Source Voltage and Source Frequency listed in the Power Supply System section.
Sensitivity, edge–type trigger, DC coupled
Edge–Type trigger sensitivity, not DC coupled, typical
Trigger level ranges Input channels: ± 4.90 divisions from center screen
Trigger level accuracy, DC coupled, typical
Lowest frequency for successful operation of Set Level to 50% function.
Pulse–Type runt trigger sensitivity, typical
Trigger Source Sensitivity
Analog inputs 0.4 division from DC to 50 MHz
Trigger Coupling Typical Sensitivity
HF reject Same as DC Coupled limits from DC to 85 kHz.
LF reject 1.2 times the DC Coupled limits for frequencies
Noise reject 2.5 times the DC Coupled limits.
±0.2 div for signals within ±4 divisions from center screen, having rise and fall times of ≥20 ns.
50 Hz. Using a 10 X probe will not affect the operation of this function.
0.75 divisions, from DC to max bandwidth.
0.6 divisions >50 MHz to 100 MHz
Attenuates signals above 85 kHz.
above 65 kHz. Attenuates signals below 65 kHz.
Pulse–Type trigger width sensitivity, typical
Pulse–Type trigger, minimum pulse rearm time
Time range for pulse width or runt triggering
6 TBS2000 Series Specification and Performance Verification
3.5 ns.
Pulse Class Minimum Pulse Width Minimum Rearm Time
Runt 2 ns 2 ns
Width 2 ns 2 ns
Rise/Fall Time 2 ns 2 ns
2 ns to 8 s
Trigger system (cont.) Specifications
Time accuracy for pulse width triggering
Trigger frequency counter Provides the user a higher accuracy means of identifying the frequency of trigger signals. Since
Resolution 6 digits
Accuracy, typical ±25 x 10-6 including all reference errors and ±1 count errors.
Frequency range, typical AC coupled, 10 Hz minimum to rated bandwidth
Signal source Edge selected trigger source only.
± 2 ns.
averaging takes place over a longer time span, the number of stable digits is improved over the Automatic Measurement of the same type.
Frequency counter measures the selected trigger source at all times in edge mode, including when the oscilloscope acquisition is halted due to changes in run status, or acquisition of a single shot event has completed. Counts all edges of sufficient amplitude.

Input output ports

TekVPI interface The probe interface allows installing, powering, compensating and controlling a wide range of
probes offering a variety of features.
Total probe power, typical TBS2xx4: 24 W, derated at 0.3 W/ °C above 30 °C
TBS2xx2: 12 W
Ethernet interface One 10/100M BaseT port
Wi-Fi interface Available as an optional USB dongle, supports 802.11 b/g/n.
GPIB interface Available as an optional accessory that connects to USB Device and USB Host Ports,
TEK-488 GPIB to USB Adapter. Control interface is incorporated in the instrument UI.
USB interface 2 High Speed 2.0 Host and 1 High Speed Device connector are standard in all models.
Probe compensator Front-panel pins
Output voltage and frequency, typical
Aux Out HIGH to LOW transition indicates the trigger occurred.
5 V amplitude ± 10% square wave, 1 kHz ± 10%.
TBS2000 Series Specification and Performance Verification 7
Specifications

Data storage

Nonvolatile memory retention time, typical
Real-Time clock A programmable clock providing time in years, months, days, hours, minutes, and seconds.
No time limit for Front Panel Settings, saved waveforms, setups, and calibration constants.

Display system

Display type 9 inch (228 mm) wide format liquid crystal TFT color display.
Display resolution 800 horizontal by 480 vertical displayed pixels (WVGA).
Waveform styles Vectors, Variable Persistence, and Infinite Persistence.
Graticules Grid, None.
Format YT and XY.

Power source

Power consumption 30 W typical, 80 W max at 85 to 275 VAC input.
Power source voltage Full range: 100 to 240 VAC RMS ±10%, Installation Category II (Covers range of 90 to 264 VAC)
Source frequency 45 Hz to 65 Hz over entire source voltage range.
360 Hz to 440 Hz, 100 to 132 VAC RMS source voltage
8 TBS2000 Series Specification and Performance Verification
Specifications

Physical characteristics

Weight
TBS2xx2: 2.62 kg (5.75 lbs.), standalone instrument.
5.1 kg (11.2 lbs.), when packaged for domestic shipment.
TBS2xx4: 4.17 kg (9.15 lbs.), stand-alone instrument.
7 kg (15.4 lbs.), when packaged for domestic shipment.
Dimensions
TBS2xx2: Height:
Handle folded down: 174.9 mm (6.89 in)
Handle Up: 224.9 mm (8.85 in)
Width: 372.4 mm (14.66 in)
Depth: 103.3 mm (4.07 in)
TBS2xx4: Height:
Handle folded down: 201.5mm (7.93 in)
Handle Up: 251.5 mm (9.9 in)
Width: 412.8 mm (16.25 in)
Depth: 128.1 mm (5.04 in)
Cooling method TBS2xx4: Forced air flow, with fan.
TBS2xx2: Convection air flow, no fan.
Cooling clearance 50 mm (2 in) required on left side and rear of instrument.

EMC, environment, and safety

Temperature
Operating: 0 °C to +50 °C, with 5 °C/minute maximum gradient, noncondensing, up to 3000 m altitude.
Instrument will be in specification after a 10 minute settling time and performance of SPC
Nonoperating: -40 °C to +71 °C, with 5 °C/minute maximum gradient.
Humidity
Operating: 5% to 95% relative humidity (% RH) at up to +30 °C, 5% to 60% RH above +30 °C up to +50 °C,
noncondensing.
Nonoperating: 5% to 95% RH (Relative Humidity) at up to +30 °C,
5% to 60% RH above +30 °C up to +60 °C, noncondensing.
TBS2000 Series Specification and Performance Verification 9
Specifications EMC, environment, and safety (cont.)
Altitude
Operating: Up to 3,000 meters (9,842 feet).
Non-Operating: Up to 12,000 meters (39,370 feet).
Altitude is limited by possible damage to LCD at higher altitudes. This damage is independent of operation.
10 TBS2000 Series Specification and Performance Verification

Performance verification

Required equipment

Table 3: Performance verification
Description Minimum requirements Examples
DC voltage source 17.5 mV to 7 V, ±0.5% accuracy Wavetek 9100 Universal Calibration System
Leveled sine wave Generator 50 kHz and 200 MHz, ±3% amplitude
accuracy
Time mark generator 10 ms period, ±10 ppm accuracy
50Ω BNC cable BNC male to BNC male, ≈ 1 m (36 in) long Tektronix part number 012-0482-XX
50Ω BNC cable BNC male to BNC male, ≈ 25 cm (10 in) long Tektronix part number 012-0208-XX
50 Ω feed through termination BNC male and female connectors Tektronix part number 011-0049-XX
Dual banana to BNC adapter Banana plugs to BNC female Tektronix part number 103-0090-XX
BNC T adapter BNC male to dual BNC female connectors Tektronix part number 103-0030-XX
Splitter, power Frequency range: DC to 4 GHz. Tracking:
>2.0%
Adapter (four required) Male N-to-female BNC Tektronix part number 103-045-XX
Adapter Female N-to-male BNC Tektronix part number 103-0058-XX
Leads, 3 black Stacking banana plug patch cord, ≈ 45 cm
(18 in) long
Leads, 2 red Stacking banana plug patch cord, ≈ 45 cm
(18 in) long
with Oscilloscope Calibration Module (Option
250) Fluke 5500A Multi-product Calibrator with Oscilloscope Calibration Option (Option 5500A-SC)
Tektronix part number 015-0565-XX
Pomona #B-18-0
Pomona #B-18-2
TBS2000 Series Specification and Performance Verification 11
Performance verification

Test record

Table 4: Test record
Instrument Serial Number: Temperature: Date of Calibration:
Certificate Number: RH %: Technician:
Instrument performance test Passed Failed
Self test
Signal path compensation (SPC)
Table 5: DC balance
Channel Coupling Low limit Test result High limit
Channel 1 DC -21 mV 21 mV
GND -21 mV 21 mV
Channel 2 DC -21 mV 21 mV
GND -21 mV 21 mV
Channel 3
Channel 4
1
DC -21 mV 21 mV
GND -21 mV 21 mV
1
DC -21 mV 21 mV
GND -21 mV 21 mV
Table 6: DC gain accuracy
Channel Vertical scale Low limit Test result High limit
Channel 1 200 mV/div 1.358 V 1.442 V
Channel 2 200 mV/div 1.358 V 1.442 V
Channel 3
Channel 4
1
1
200 mV/div 1.358 V 1.442 V
200 mV/div 1.358 V 1.442 V
Table 7: Bandwidth
Channel Low limit Test result High limit
Channel 1 2.12 mV -- --
Channel 2 2.12 mV -- --
Channel 3
Channel 3
1
1
1
Channels 3 and 4 are only on four channel instruments
2.12 mV -- --
2.12 mV -- --
12 TBS2000 Series Specification and Performance Verification
Performance verification
Table 8: Vertical position range
Channel V/div setting Trace position Offset DC Voltage source Pass/Fail
Channel 1 200 mV/div Top -0.8 V -1.8 V
Bottom +0.8 V +1.8 V
5 V/div Top –20 V -45 V
Bottom +20 V +45 V
Channel 2 200 mV/div Top -0.8 V -1.8 V
Bottom +0.8 V +1.8 V
5 V/div Top –20 V -45 V
Bottom +20 V +45 V
Channel 3 200 mV/div Top -0.8 V -1.8 V
Bottom +0.8 V +1.8 V
5 V/div Top –20 V -45 V
Bottom +20 V +45 V
Channel 4 200 mV/div Top -0.8 V -1.8 V
Bottom +0.8 V +1.8 V
5 V/div Top –20 V -45 V
Bottom +20 V +45 V
Table 9: Sample rate and delay time accuracy
Instrument performance test Low limit Test result High limit
Sample Rate and Delay Time Accuracy -2.5 divs +2.5 divs

Performance verification procedures

NOTE.
If your instrument firmware version is v1.02, it should be updated before performing the performance verification procedures. Download the latest firmware from www.tek.com/software.
The following three conditions must be met prior to performing these procedures:
The instrument must have been operating continuously for twenty (20) minutes in an environment that meets the operating range specifications for temperature and humidity.
You must perform a signal path compensation (SPC) before beginning these procedures. If the operating temperature changes by more than 10 °C (18 °F), you must perform the signal path compensation again.
You must connect the instrument and the test equipment to the same AC power circuit. Connect the instrument and test instruments into a common power strip if you are unsure of the AC power circuit distribution. Connecting the instrument and test instruments into separate AC power circuits can result in offset voltages between the equipment, which can invalidate the performance verification procedure.
The time required to complete the entire procedure is approximately one hour.
TBS2000 Series Specification and Performance Verification 13
Performance verification
WARNING. Some procedures use hazardous voltages. To prevent electrical shock, always set voltage source outputs to 0 V before making or changing any interconnections.

Self test

This procedure uses internal routines to verify that the instrument functions and passes its internal self tests. No test equipment or hookups are required. Start the self test with these steps:
1. Disconnect all probes and cables from the instrument inputs.
2. Push the front-panel Default Setup button to set the instrument to the factory default settings.
3. Push the Utility menu button.
4. Push the Utility Page bezel button, the Diagnostics bezel button, the Self Test bezel button, and turn Multipurpose knob a to
select Loop Times.
5. Push the Multipurpose knob a to select Loop Times, and turn the Multipurpose knob a to select Loop 1 Times.
6. Push the Multipurpose knob a to set the Loop Times to 1.
7. Turn Multipurpose knob a to select Run Self Test, and push the Multipurpose knob a to start the self tests.
8. Wait while the self test runs. When the self test completes, a dialog box displays the results of the self test.
9. Push the Menu Off button to clear the dialog box and Self Test menu.

Signal path compensation (SPC)

This process corrects for DC inaccuracies caused by temperature variations and/or long term drift.
1. Remove all input signals (probes and cables) from channel inputs. Input signals with AC components adversely affect SPC.
2. Push the front-panel Utility button, and then push the Utility Page bezel button.
3. Push the Calibration bezel button.
4. Turn the Multipurpose button a to select Signal Path, and then push Multipurpose knob a to select Calibration Signal Path.
5. Push the Compensate Signal Paths bezel button.
6. Wait while the Signal Path Compensation runs. On completion a dialog box informs you whether the Compensation
completed successfully or not.
7. Push the Menu Off button to clear the dialog box and Self Test menu.

Check DC balance

This test checks the DC balance of each channel. You do not need to connect the instrument to any equipment to run this test.
14 TBS2000 Series Specification and Performance Verification
Performance verification
1. Push the front-panel Default Setup button to set the instrument to the factory default settings.
2. Turn the Horizontal Scale knob to 1 ms/div.
3. Push the Trigger Menu front-panel button.
4. Push the Measurement source bezel button for the channel you are testing.
5. Select the AC Line trigger source with Multipurpose knob a. You do not need to connect an external signal to the instrument
for this DC Balance test.
6. Push the front-panel Acquire button.
7. Push the Acquire Mode bezel button.
8. Turn Multipurpose knob a to select Average, and then push Multipurpose knob a to turn on Average mode.
NOTE. When using averaging, allow the instrument to acquire all the samples before taking the measurement.
9. If needed, adjust the number of averages to 16 with Multipurpose knob a.
10. Push the front-panel channel button for the instrument channel to test, as shown in the test record (for example, 1, 2, 3, or
4).
11. Set the channel being tested to 200 mV/div using the Vertical Scale knob.
12. Attach a 50 Ω terminator to the instrument input channel being tested.
13. Push the Coupling bezel button and use the Multipurpose knob to select DC or GND coupling, as given in the test record.
14. Push the front-panel Resources Measure button.
15. Push the bezel channel button for the instrument channel to test, as shown in the test record (for example, CH1, CH2, CH3,
or CH4).
16. Use Multipurpose knob a to select the Mean measurement.
17. Push the Multipurpose knob to add the Mean measurement, and then push the Menu Off button.
18. View the mean measurement value in the display and enter that mean value as the test result in the test record.
19. Push the front-panel channel button, and repeat steps 13 through 18
20. Repeat steps 5 through 19 for each remaining channel.

Check DC gain accuracy

This test checks the DC gain accuracy of all input channels.
1. Push the front-panel Default Setup button to set the instrument to the factory default settings.
2. Push the front-panel Horizontal Acquire button, then push the Mode bezel button, and then use the Multipurpose knob to
select and then turn averaging on.
NOTE. When using averaging, allow the instrument to acquire all the samples before taking the measurement.
3. If necessary, use Multipurpose knob a to set the number of averages to 16.
TBS2000 Series Specification and Performance Verification 15
Performance verification
4. Set the DC voltage source to 0 V, and then connect it to channel 1, as shown. If using a Fluke 9500 as the voltage source, connect the calibrator head to channel 1.
5. Push the front panel button to select the channel to be tested (1, 2, 3, or 4).
6. Push the Probe Setup bezel button, and then use the Multipurpose knob to select Set to 1X.
7. Push the Resources Measure button, and then push the bezel button to select the channel to be tested (1, 2, 3, or 4).
8. Use Multipurpose knob a to select and add the Mean measurement, and then push the Menu Off button.
9. For each Volts/div line in the following worksheet, perform these steps:
a. Set the DC voltage source output level to the positive voltage listed and record the Mean measurement as V
b. Set the DC voltage source to the negative level listed, and record the Mean measurement as V
c. Calculate V
690 mV and V
d. Enter V
Ch Volts/div)
= V
– V
diff
pos
is -685 mV, then V
neg
in the worksheet, and in the test record.
diff
, and then enter V
neg
diff
is 1.375 V.
DC voltage source setting V
setting
Positive Negative
in the test record. As an example, on the 200 mV/div setting, if V
diff
pos
V
neg
V
diff
neg
Accuracy limits for V
Ch 1 200 mV/div +700 mV, -700 mV 1.358 V to 1.442 V
Ch 2 200 mV/div +700 mV, -700 mV 1.358 V to 1.442 V
Ch 3 200 mV/div +700 mV, -700 mV 1.358 V to 1.442 V
Ch 4 200 mV/div +700 mV, -700 mV 13.58 V to 14.42 V
10. Set the DC voltage source to 0 V, and move the BNC cable to the next channel to be tested.
11. Repeat steps 5 through 10 for each remaining channel.
.
pos
.
is
pos
diff
16 TBS2000 Series Specification and Performance Verification
Performance verification

Check bandwidth

This test checks the bandwidth of all input channels.
1. Connect the output of the leveled sine wave generator (for example, Fluke 9500) to the channel 1 input as shown:
2. Push the front-panel Default Setup button to set the instrument to the factory default settings.
3. Push the front-panel Trigger Menu button.
4. Push the Coupling bezel button, and then use the Multipurpose knob to select and then set Noise Reject (DC Low
Sensitivity).
5. Push the front-panel Trigger Menu button.
6. Push the Source bezel button and use Multipurpose knob a to select the channel being tested as the trigger source.
7. Push the Menu Off button, so you can see the screen.
8. Push the channel button (1, 2, 3, or 4) for the channel that you want to check.
9. Push the Probe Setup bezel button, and then use the Multipurpose knob to select Set to 1 X.
10. Push the front-panel Measure button, and then push the bezel button for the channel you are testing.
11. Use Multipurpose knob a to select the Peak-to-peak measurement.
12. Turn the Vertical Scale knob to set the vertical scale to 500 mV/div.
13. Turn the Horizontal Scale knob to 400 μs/div.
14. Set the leveled sine wave generator frequency to 1 kHz.
15. Set the leveled sine wave generator output level so the peak-to-peak measurement is between 2.98 V and 3.02 V.
16. Set the leveled sine wave generator frequency to:
100 MHz if you are checking a TBS2104 or TBS2102
70 MHz if you are checking a TBS2074 or TBS2072
17. Use the Horizontal Scale knob to set the instrument to 10 ns/div.
18. Check that the peak-to-peak measurement is ≥2.12 V. Enter this measurement in the test record.
19. Move the input cable to the next channel to be tested.
20. Repeat steps 3 through 19 for all input channels.
TBS2000 Series Specification and Performance Verification 17
Performance verification

Check vertical offset accuracy

This test checks the offset range for each channel.
1. Connect the instrument to a DC voltage source to run this test. If using the Fluke calibrator as the DC voltage source, connect the calibrator head to the instrument channel to test.
2. Push the front-panel Default Setup button to set the instrument to the factory default settings.
3. Push the channel button (1, 2, 3, or 4) for the channel that you want to check.
4. Push the Probe Setup button, and then use the Multipurpose knob to select Set to 1 X.
5. Use the Vertical Scale knob to set the instrument to 200 mV/div.
6. Use the Vertical Position knob to place the trace at the bottom of the display (-5 divisions).
7. Press the Offset bezel button and use the Multipurpose knob to set the Offset to +0.8 V.
8. Set the DC Voltage source to +1.8 V.
9. Check that the vertical trace is now within 0.2 divisions of the Zero volt line. Record Pass or Fail in the test record.
10. Set the DC Voltage source to 0 V.
11. Push the Offset bezel button and use the Multipurpose knob to select Set to 0V.
12. Use the Vertical Position knob to place the trace at the top of the display (+5 divisions).
13. Press the Offset bezel button and use the Multipurpose knob to set the Offset to –0.8 V.
14. Set the DC Voltage source to –1.8 V.
15. Check that the vertical trace is now within 0.2 divisions of the Zero volt line. Record Pass or Fail in the test record.
16. Set the DC Voltage source to 0 V.
17. Push the Offset bezel button and use the Multipurpose knob to select Set to 0V.
18. Use the Vertical Scale knob to set the instrument to 5 V/div.
19. Use the Vertical Position knob to place the trace at the bottom of the display (-5 divisions).
20. Press the Offset bezel button and use the Multipurpose knob to set the Offset to +20.00 V.
21. Set the DC Voltage source to +45 V.
22. Check that the vertical trace is now within 0.2 divisions of the Zero volt line. Record Pass or Fail in the test record.
23. Push the Offset bezel button and use the Multipurpose knob to select Set to 0V.
24. Use the Vertical Position knob to place the trace at the top of the display (+5 divisions).
25. Press the Offset bezel button and use the Multipurpose knob to set the Offset to –20.00 V.
18 TBS2000 Series Specification and Performance Verification
Performance verification
26. Set the DC Voltage source to –45 V.
27. Check that the vertical trace is now within 0.2 divisions of the Zero volt line. Record Pass or Fail in the test record.
28. Set the DC Voltage source to 0 V.
29. Push the Offset bezel button and use the Multipurpose knob to select Set to 0V.
30. Move the DC Voltage source cable to the next channel to be tested.
31. Push the channel button (1, 2, 3, or 4) for the next channel to check.
32. Repeat steps 4 through 31 for each of the remaining channels.

Check sample rate and horizontal position time accuracy

This test checks the sample rate and horizontal position time accuracy (time base).
1. Connect the output of the time mark generator to the channel 1 input using a 50 Ω cable and 50 Ω feed through terminator.
2. Set the time mark generator period to 1 ms. Use a time mark waveform with a fast rising edge.
3. Push the front-panel Default Setup button to set the instrument to the factory default settings.
4. Push the channel 1 button.
5. Push the Probe Setup bezel button, and then use the Multipurpose knob to select Set to 1 X.
6. Set the Vertical SCALE to 500 mV/div.
7. Set the Horizontal SCALE to 1 ms/div.
8. If adjustable, set the time mark generator amplitude to approximately 1 Vp-p
9. Push the Trigger Level knob, to set the trigger level to 50%.
10. Adjust the Vertical POSITION knob to center the time mark signal vertically on the screen.
11. If necessary, adjust the Horizontal POSITION knob to move the trigger location to the center of the screen (50%).
12. Turn the Horizontal POSITION knob counterclockwise to set the delay to close to 1 ms.
13. Set the Horizontal Scale to 10 ns/div.
14. If necessary, turn the Horizontal Position knob to set the delay to exactly 1.0000 ms.
TBS2000 Series Specification and Performance Verification 19
Performance verification
15. Compare the rising edge of the marker with the center horizontal graticule line. The rising edge should cross the 0 V center within ±2.5 divisions (±25 ns) of the center graticule line. Enter the deviation in the test record.
NOTE. One division of displacement from graticule center corresponds to a 10 ppm time base error.
20 TBS2000 Series Specification and Performance Verification

Index

B
bandwidth test, 17
C
R
Required equipment, 11
calibration
Do Self Cal, 14
D
DC balance test, 14 DC gain accuracy test, 15
H
Horizontal position time accuracy test, 19
P
Performance verification, 11 Performance verification procedures, 13
S
Sample rate and horizontal position time accuracy test, 19 self calibration, 14 Self test, 14 Signal path compensation, 14 Specifications, 1
T
Test record, 12
V
Vertical offset test, 18
TBS2000 Series Specification and Performance Verification 21
Index
22 TBS2000 Series Specification and Performance Verification
Loading...