Tektronix TBS1052B, TBS1052B EDU, TBS1072B, TBS1072B EDU, TBS1102B Service

...
xx
TBS1000B and TBS1000B-EDU Series
ZZZ
Digital Storage Oscilloscopes
Service Manual
*P077089702*
077-0897-02
xx
TBS1000B and TBS1000B-EDU Series
ZZZ
Service Manual
Revision A
This document applies to TBS1000B and TBS1000B-EDU models with no display adapter module.
Warning
The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
www.tek.com
077-0897-02
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
OpenChoice is a registered trademark of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O . Bo x 50 0 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200. Worldwide, visit www.tek.com to find contacts in your area.
Table of Contents
Importantsafety information..................................................................................... iii
General safety summary ..................................................................................... iii
Service safety summary...................................................................................... vi
Terms in t h i
Symbols and terms on the product......................................................................... vii
Preface.............................................................................................................. ix
Manual Conventions.......................................................................................... ix
Related Documentation ...................................................................................... ix
Inspection andCleaning ........................................................................................... 1
General
Interior Cleaning .......................................................................................... 1
Exterior Cleaning ......................................................................................... 2
Exterior Inspection........................................................................................ 2
Interior Inspection ........................................................................................ 3
Cleaning Procedure, Interior............................................................................. 4
stment procedures............................................................................................. 5
Adju
Required equipment ........................................................................................... 5
Adjustmentprocedure......................................................................................... 6
Enable the Service menu ................................................................................. 6
Adjustment procedure.................................................................................... 8
Diagrams ........................................................................................................... 13
placeable Parts.................................................................................................. 15
Re
Parts Ordering Information .................................................................................. 15
Using theReplaceable Parts List............................................................................ 16
Parts Lists andExploded Views............................................................................. 17
s manual ........................................................................................ vii
Care............................................................................................... 1
TBS1000B/-EDU Series Oscilloscope Service Manual i
Table of Contents
List of Tables
Table 1: External inspection checklist............................................................................ 2
Table 2: Int
Table 3: Required equipment...................................................................................... 5
Table 4: Adjustment steps ......................................................................................... 9
Table 5: Replaceable parts list ................................................................................... 16
Table 6: Replaceable parts — (LCD, Front panel, andFront case).......................................... 17
Table 7: Replaceable parts — (Power supply, Main board, Rear case). .. ... . ............................... 18
ernal inspection checklist ............................................................................ 3
ii TBS1000B/-EDU Series Oscilloscope Service Manual
Important safety information
This manual contains information and warnings that must be followed by the user for safe operation and to keep the product in a safe condition.
To safely perform service on this product, additional information is provided at the end of this section. (See page vi, Service safety summary.)
General safety summary
Use the product only as specified. Revie avoid injury and prevent damage to this product or any products connected to it. Carefully read all instructions. Retain these instructions for future reference.
Comply with local and national safety codes.
For correct and safe operation of the product, it is essential that you follow generally accepted safety procedures in addition to the safety precautions specified in this manual.
The product is designed to be used by trained personnel only.
Only qualified personnel who are aware of the hazards involved should remove the cover for repair, maintenance, or adjustment.
Before use, always check the product with a known source to be sure it is operating correctly.
This product is not intended for detection of hazardous voltages.
Use personal protective equipment to prevent shock and arc blast injury where hazardous live conductors are exposed.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals for warnings and cautions related to operating the system.
w the following safety precautions to
When incorporating this equipment into a system, the safety of that system is the responsibility of the assembler of the system.
To avoid re or personal
injury
TBS1000B/-EDU Series Oscilloscope Service Manual iii
Use proper power cord. Use only the power cord specified for this product and certified for the country of use.
Do not use the provided power cord for other products.
Ground the product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, make sure that the product is properly grounded.
Do not disable the power cord grounding connection.
Important safety information
Power disconne
source. See instructions for the location. Do not position the equipment sothatit is difficult to disconnect the power switch; it must remain accessible to theuserat all times to allow for quick disconnection if needed.
Connect and disconnect properly. Do not connect or disconnect probes or test leads while they are connected to a voltage source.
Use only insulated voltage probes, test leads, and adapters supplied with the product, or indicated by Tektronix to be suitable for the product.
Observe all terminal ratings. To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product. Do not exceed the Measurement Category (CAT) rating and voltage or current rating of the lowest rated individual component of a product, probe, or accessory. Use caution when using 1:1 test leads because the probe tip voltage is directly transmitted to the product.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Do not float the common terminal above the rated voltage for that terminal.
Do not operate without covers. Do not operate this product with covers or panels removed, or with the case open. Hazardous voltage exposure is possible.
ct. The power switch disconnects the product from the power
Avoid exposed circuitry. Do not touch exposed connections and components
when power is present.
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by qualified service personnel.
Disable the product if it is damaged. Do not use the product if it is damaged or operates incorrectly. If in doubt about safety of the product, turn it offand disconnect the power cord. Clearly mark the product to prevent its further operation.
Before use, inspect voltage probes, test leads, and accessories for mechanical damage and replace when damaged. Do not use probes or test leads if they are damaged, if there is exposed metal, or if a wear indicator shows.
Examine the exterior of the product before you use it. Look for cracks or missing pieces.
Use only specified replacement parts.
Use proper fuse. Use only the fuse type and rating specified for this product.
Wear eye protection. Wear eye protection if exposure to high-intensity rays or
laser radiation exists.
iv TBS1000B/-EDU Series Oscilloscope Service Manual
Important safety information
Do not operate i
a unit is moved from a cold to a warm environment.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry. Remove the input signals before you clean
the product.
Provide proper ventilation. Refer to the installation instructions in the manual for details on installing the product so it has proper ventilation.
Slots and openings are provided for ventilation and should never be coveredor otherwise obstructed. Do not push objects into any of the openings.
Provide a safe working environment. Always place the product in a location convenient for viewing the display and indicators.
Avoid improper or prolonged use of keyboards, pointers, and button pads. Improper or prolonged keyboard or pointer use may result in serious injury.
Be sure your work area meets applicable ergonomic standards. Consult with an ergonomics professional to avoid stress injuries.
n wet/damp conditions. Be aware that condensation may occur if
Probes and test leads
re connecting probes or test leads, connect the power cord from the power
Befo connector to a properly grounded power outlet.
p fingers behind the finger guards on the probes.
Kee
Remove all probes, test leads and accessories that are not in use.
Use only correct Measurement Category (CAT), voltage, temperature, altitude, and amperage rated probes, test leads, and adapters for any measurement.
Beware of high voltages. Understand the voltage ratings for the probe you are using and do not exceed those ratings. Two ratings are important to know and understand:
The maximum measurement voltage from the probe tip to the probe reference lead.
The maximum floating voltage from the probe reference lead to earth ground
These two voltage ratings depend on the probe and your application. Refer tothe Specifications section of the manual for more information.
WAR N ING. To prevent electrical shock, do not exceed the maximum measurement
or maximum oating voltage for the oscilloscope input BNC connector, probe tip, or probe reference lead.
TBS1000B/-EDU Series Oscilloscope Service Manual v
Important safety information
Connect and dis
product before connecting the probe to the circuit under test. Connect the probe reference lead to the circuit under test before connecting the probe input. Disconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe from the measurement product.
Connect and
connecting or disconnecting the current probe.
Connect th
Do not connect a current probe to any wire that carries voltages above the current probe vol
Inspect the probe and accessories. Before each use, inspect probe and accessories for damage (cuts, tears, or defects in the probe body, accessories, or cablejacket). Do not use if damaged.
Ground
when using with ground-referenced oscilloscopes. The reference lead mustbe connected to earth potential (0 V).
-referenced oscilloscope use. Do not float the reference lead of this probe
connect properly. Connect the probe output to the measurement
disconnect properly. De-energize the circuit under test before
e probe reference lead to earth ground only.
tage rating.
Servicesafetysummary
The Service safety summary section contains additional information required to safely perform service on the product. Only qualified personnel should perform service procedures. Read this Service safety summary and the General safety summary before performing any service procedures.
To avoid electric shock. Do not touch exposed connections.
Do not service alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is present.
Disconnect power. To avoid electric shock, switch off the product power and disconnect the power cord from the mains power before removing any covers or panels, or opening the case for servicing.
Use care when servicing with power on. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
Verify safety after repair. Always recheck ground continuity and mains dielectric strength after performing a repair.
vi TBS1000B/-EDU Series Oscilloscope Service Manual