Tektronix TBS1052C, TBS1072C, TBS1102C, TBS1202C Declassification and Security Instructions

TBS1000C Series Digital Storage Oscilloscope Declassification and Security Instructions
*P077158500*
077-1585-00
TBS1000C Series Digital Storage Oscilloscope Declassification and Security Instructions
Warning
The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
Supports Product Firmware V1.0 and above
www.tek.com
077-1585-00
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tek.com to find contacts in your area.

Table of Contents

Preface ................................................................................................................................................. 3
Products .......................................................................................................................................... 3
Required documents ....................................................................................................................... 3
Terms used in this document .......................................................................................................... 4
Device terms ................................................................................................................................... 5
Memory devices
Volatile memory ............................................................................................................................. 1
Non-volatile memory devices ......................................................................................................... 2
Data export devices ........................................................................................................................ 3
Clear and sanitize procedures
Clear the flash memory ................................................................................................................... 5
Clear or sanitize a non-functional instrument
Acquisition board ........................................................................................................................... 7
USB flash drive .............................................................................................................................. 7
Charges ........................................................................................................................................... 7
TBS1000C Declassification and Security i
Table of Contents
ii TBS1000C Declassification and Security

Preface

Products

This document helps customers with data security concerns to sanitize or remove memory devices from the listed products. These products have data storage (memory) devices and data output devices (USB ports). These instructions tell you how to clear or sanitize the memory devices, and also tell you how to declassify an instrument that is not functioning.
Instrument code and calibration settings reside in nonvolatile flash memory.
Instrument setups and reference waveforms may also be stored in flash memory or on USB drives connected to the instrument.
If you have any questions, contact the Tektronix Technical Support Center at
https://www.tek.com/support
The following Tektronix products are covered by this document:
TBS1052C
TBS1072C

Required documents

TBS1102C
TBS1202C
To perform the procedures in this document, you will need to have access to the manuals listed below. These manuals are available on the Tektronix Web site at
https://www.tek.com/product-support.
TBS1000C Series Digital Storage Oscilloscope Service Manual (Tektronix part number, 077-1584-00).

Terms used in this document

The following terms may be used in this document:
Clear. This removes data on media/memory before reusing it in a secured area. All reusable memory is cleared to deny access to previously stored information by standard means of access.
Erase. This is equivalent to clear.
Instrument Declassification. A term that refers to procedures that must be undertaken before an instrument can be removed from a secure environment. Declassification procedures include memory sanitization and memory removal, and sometimes both.
Media storage/data export device. Any of several devices that can be used to store or export data from the instrument, such as a USB port.
Nonvolatile memory. Data is retained when the instrument is powered off.
Power off. Some instruments have a “Standby” mode, in which power is still supplied to the instrument. For the purpose of clearing data, putting the instrument in Standby mode does not qualify as powering off. For these products, you will need to either press a rear-panel OFF switch or remove the power source from the instrument.
Remove. This is a physical means to clear the data by removing the memory device from the instrument. Instructions are available in the product Service Manual.
Sanitize. This eradicates the data from media/memory so that the data cannot be recovered by other means or technology. This is typically used when the device will be moved (temporarily or permanently) from a secured area to a non-secured area.
Scrub. This is equivalent to sanitize.
User Accessible. User is able to directly retrieve the memory device contents.
User-Modifiable. User can write to the memory device during normal instrument operation, using the instrument interface or remote control.
Volatile memory. Data is lost when the instrument is powered off.

Device terms

The following terms are used with the memory devices in this document:
User data. Describes the type of information stored in the device. Refers to waveforms or other measurement data representing signals connected to the instrument by users.
User settings. Describes the type of information stored in the device. Refers to instrument settings that can be changed by the user.
Both. Describes the type of information stored in the device. It means that both user data and user settings are stored in the device.
None. Describes the type of information stored in the device. It means that neither user data nor user settings are stored in the device.
Directly. Describes how data is modified. It means that the user can modify the data.
Indirectly. Describes how data is modified. It means that the instrument system resources modify the data and that the user cannot modify the data.

Memory devices

Volatile memory

DDR3 SDRAM
Type and size
Function
Type of user information stored
Backed up by battery?
Method of modification
Data input method
Location
User accessible
To clear
Process to sanitize
Real-time clock
Type and size
Function
1 GB, (256 M X 16 bit) X two banks
Holding and processing waveforms and processor system RAM
Both user data and user settings
No
Indirect
Firmware operations
Main Acquisition board.
No
Remove power from the instrument for at least 20 seconds.
Remove power from the instrument for at least 20 seconds.
Real-time clock, which contains battery backed-up memory
Holds date/time data
Type of user information stored
Backed up by battery?
Method of modification
None
Yes
Indirect
TBS1000C Declassification and Security 1
Memory devices

Non-volatile memory devices

MMC Nand flash
Type and size
Function
Type of user information stored
Backed up by battery?
Method of modification
Data input method
Location
User accessible
To clear
Process to sanitize
QSPI/NOR flash
Type and size
Function
4 G Byte (8G X 4) <- Check with Gene.
Kernel code, application code, waveform data, backup data, and user stored data.
Both user data and user settings.
No.
Indirect.
Firmware operations
Main acquisition board.
No.
Remove power from the instrument for at least 20 seconds.
Remove power from the instrument for at least 20 seconds.
16 M byte.
U-boot, FPGA configurable code and constant parameters.
Type of user information stored
Backed up by battery?
Method of modification
Data input method
Location
User accessible
To clear
Process to sanitize
Both user data and user settings.
No.
Indirect.
Firmware operations.
Main acquisition board.
No.
Remove power from the instrument for at least 20 seconds.
Remove power from the instrument for at least 20 seconds.
2 TBS1000C Declassification and Security
Memory devices
SPI flash
To clear
Process to sanitize
Nand flash
To clear
Process to sanitize

Data export devices

USB host port
Type and size
Function
Method of modification
Data input method
Location
Perform the procedure Clear the flash memory on page 5.
Perform the procedure Clear the flash memory on page 5.
Perform the procedure Clear the flash memory on page 5.
Perform the procedure Clear the flash memory on page 5.
USB host port
Supports removable USB flash drive. User storage of reference waveforms, screen images, and instrument setups.
Directly
Save
Front panel
To disable
USB device port
Type and size
Function
Method of modification
Data input method
Location
To disable
Files can be deleted or overwritten on the oscilloscope or a PC; USB flash drive can be removed and destroyed. The USB host port cannot be disabled.
USB device port
Supports remote control and data transfer to a PC
Directly
Remote control via USBTMC
Rear panel
The USB device port cannot be disabled.
TBS1000C Declassification and Security 3
Memory devices
4 TBS1000C Declassification and Security

Clear and sanitize procedures

Clear the flash memory

This procedure does not erase or change factory calibration constants.
1. Push the front panel DEFAULT SETUP button to recall the default setup.
2. Push the front panel SAVE/RECALL button.
3. Push the Action option (side bezel) button until Save Setup is selected.
4. Push the Save To option button until Setup is selected.
5. Push the Setup option button until 1 is selected.
6. Push the Save option button to overwrite setup 1 with the default setup.
7. Push the Setup and Save option buttons again to overwrite the next setup
with the default setup. Repeat this step until setup 1 is selected again.
8. Push the front panel CH 1 MENU button, and then push the Coupling option button until Ground is selected.
9. Push the front panel SAVE/RECALL button.
10. Push the Action option button until Save Waveform is selected. Push the
Save To option button until Ref is selected. Push the Source option button until CH1 is selected.
11. Push the To option button until RefA is selected. Push the Save option button to overwrite reference waveform RefA with the null waveform.
12. Push the To and the Save option buttons again to overwrite the next reference waveform with the null waveform. Repeat this step until reference RefA is selected again.
13. For TBS and TDS oscilloscopes, insert a USB flash drive into the USB Flash Drive port on the front of the oscilloscope. For TPS oscilloscopes, insert a compact flash card into the Compact Flash slot on the front of the oscilloscope.
14. Push the front panel UTILITY button.
15. Push the Options option button.
16. Push the Printer Setup option button.
17. Push the PRINT Button option until Saves Image To File is selected.
18. For TBS and TDS oscilloscopes, push the File Format option button until
JPG is selected. For TPS oscilloscopes, select BMP instead of JPG.
19. Remove any probes from the BNC front panel connectors.
TBS1000C Declassification and Security 5
Clear and sanitize procedures
20. Push the front panel PRINT button to overwrite the hard copy image file with an image that does not contain any useful information.
21. Power off the oscilloscope, and then power on the oscilloscope to complete the process.
6 TBS1000C Declassification and Security

Clear or sanitize a non-functional instrument

Acquisition board

Remove the Acquisition board and return the product to Tektronix. A new Acquisition board will be installed, and the instrument will be repaired and adjusted as necessary.
For removal instructions, refer to your product service manual (manual part numbers are listed under Required documents near the front of this manual.
After removal of the Acquisition board, refer to your company's internal policies regarding handling or disposal of the board.

USB flash drive

Remove the USB flash drive and return the instrument to Tektronix for repair.
After removal of the USB flash drive, refer to your company's internal policies regarding handling or disposal of the flash drive.

Charges

Replacement of any missing hardware will be charged according to the rate at the time of replacement.
TBS1000C Declassification and Security 7
Clear or sanitize a non-functional instrument