TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
•In North America, call 1-800-833-9200.
•Worldwide, visit www.tek.com to find contacts in your area.
Table of Contents
Table of Contents
List of Tables................................................................................................................................................................................. 4
Important safety information..........................................................................................................................................................5
General safety summary........................................................................................................................................................5
Service safety summary.........................................................................................................................................................7
Terms in the manual...............................................................................................................................................................7
Terms on the product............................................................................................................................................................. 7
Symbols on the product......................................................................................................................................................... 7
Test record........................................................................................................................................................................... 20
Signal calibration.......................................................................................................................................................... 22
Check DC gain accuracy.............................................................................................................................................. 22
Check Aux In edge trigger sensitivity............................................................................................................................26
Table 2: Test record.....................................................................................................................................................................20
TBS1000C Series Oscilloscope Specification and Performance Verification 4
Important safety information
Important safety information
This manual contains information and warnings that must be followed by the user for safe operation and to keep the product in a safe
condition.
To safely perform service on this product, see the Service safety summary that follows the General safety summary.
General safety summary
Use the product only as specified. Review the following safety precautions to avoid injury and prevent damage to this product or any
products connected to it. Carefully read all instructions. Retain these instructions for future reference.
This product shall be used in accordance with local and national codes.
For correct and safe operation of the product, it is essential that you follow generally accepted safety procedures in addition to the safety
precautions specified in this manual.
The product is designed to be used by trained personnel only.
Only qualified personnel who are aware of the hazards involved should remove the cover for repair, maintenance, or adjustment.
Before use, always check the product with a known source to be sure it is operating correctly.
This product is not intended for detection of hazardous voltages.
Use personal protective equipment to prevent shock and arc blast injury where hazardous live conductors are exposed.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals
for warnings and cautions related to operating the system.
When incorporating this equipment into a system, the safety of that system is the responsibility of the assembler of the system.
To avoid fire or personal injury
Use proper power cordUse only the power cord specified for this product and certified for the country of use.
Ground the productThis product is grounded through the grounding conductor of the power cord. To avoid
electric shock, the grounding conductor must be connected to earth ground. Before making
connections to the input or output terminals of the product, ensure that the product is properly
grounded. Do not disable the power cord grounding connection.
Ground-referenced oscilloscope useDo not float the reference lead of this probe when using with ground-referenced oscilloscopes.
The reference lead must be connected to earth potential (0 V).
Power disconnectThe power cord disconnects the product from the power source. See instructions for the
location. Do not position the equipment so that it is difficult to operate the power cord; it must
remain accessible to the user at all times to allow for quick disconnection if needed.
Connect and disconnect properlyDo not connect or disconnect probes or test leads while they are connected to a voltage
source.
Use only insulated voltage probes, test leads, and adapters supplied with the product, or
indicated by Tektronix to be suitable for the product.
Observe all terminal ratingsTo avoid fire or shock hazard, observe all rating and markings on the product. Consult the
product manual for further ratings information before making connections to the product. Do
not exceed the Measurement Category (CAT) rating and voltage or current rating of the lowest
rated individual component of a product, probe, or accessory. Use caution when using 1:1 test
leads because the probe tip voltage is directly transmitted to the product.
Do not apply a potential to any terminal, including the common terminal, that exceeds the
maximum rating of that terminal.
TBS1000C Series Oscilloscope Specification and Performance Verification 5
Important safety information
Do not float the common terminal above the rated voltage for that terminal.
Do not operate without coversDo not operate this product with covers or panels removed, or with the case open. Hazardous
voltage exposure is possible.
Avoid exposed circuitryDo not touch exposed connections and components when power is present.
Do not operate with suspected
failures
Do not operate in wet/damp
conditions
Do not operate in an explosive
atmosphere
Keep product surfaces clean and dry Remove the input signals before you clean the product.
Provide proper ventilationRefer to the installation instructions in the manual for details on installing the product so it has
Provide a safe working environmentAlways place the product in a location convenient for viewing the display and indicators.
If you suspect that there is damage to this product, have it inspected by qualified service
personnel.
Disable the product if it is damaged. Do not use the product if it is damaged or operates
incorrectly. If in doubt about safety of the product, turn it off and disconnect the power cord.
Clearly mark the product to prevent its further operation.
Before use, inspect voltage probes, test leads, and accessories for mechanical damage and
replace when damaged. Do not use probes or test leads if they are damaged, if there is
exposed metal, or if a wear indicator shows.
Examine the exterior of the product before you use it. Look for cracks or missing pieces.
Use only specified replacement parts.
Be aware that condensation may occur if a unit is moved from a cold to a warm environment.
proper ventilation.
Slots and openings are provided for ventilation and should never be covered or otherwise
obstructed. Do not push objects into any of the openings.
Avoid improper or prolonged use of keyboards, pointers, and button pads. Improper or
prolonged keyboard or pointer use may result in serious injury.
Be sure your work area meets applicable ergonomic standards. Consult with an ergonomics
professional to avoid stress injuries.
Probes and test leads
Before connecting probes or test leads, connect the power cord from the power connector to a properly grounded power outlet.
Keep fingers behind the protective barrier, protective finger guard, or tactile indicator on the probes.
Remove all probes, test leads and accessories that are not in use.
Use only correct Measurement Category (CAT), voltage, temperature, altitude, and amperage rated probes, test leads, and adapters for
any measurement.
Beware of high voltagesUnderstand the voltage ratings for the probe you are using and do not exceed those ratings.
Two ratings are important to know and understand:
•The maximum measurement voltage from the probe tip to the probe reference lead
•The maximum floating voltage from the probe reference lead to earth ground
These two voltage ratings depend on the probe and your application. Refer to the
Specifications section of the manual for more information.
Warning: To prevent electrical shock, do not exceed the maximum measurement
or maximum floating voltage for the oscilloscope input BNC connector, probe tip, or
probe reference lead.
TBS1000C Series Oscilloscope Specification and Performance Verification 6
Important safety information
Connect and disconnect properlyConnect the probe output to the measurement product before connecting the probe to the
circuit under test. Connect the probe reference lead to the circuit under test before connecting
the probe input. Disconnect the probe input and the probe reference lead from the circuit under
test before disconnecting the probe from the measurement product.
Connect and disconnect properlyDe-energize the circuit under test before connecting or disconnecting the current probe.
Connect the probe reference lead to earth ground only.
Do not connect a current probe to any wire that carries voltages or frequencies above the
current probe voltage rating.
Inspect the probe and accessoriesBefore each use, inspect probe and accessories for damage (cuts, tears, or defects in the
probe body, accessories, or cable jacket). Do not use if damaged.
Service safety summary
The Service safety summary section contains additional information required to safely perform service on the product. Only qualified
personnel should perform service procedures. Read this Service safety summary and the General safety summary before performing any
service procedures.
To avoid electric shockDo not touch exposed connections.
Do not service aloneDo not perform internal service or adjustments of this product unless another person capable of
rendering first aid and resuscitation is present.
Disconnect powerTo avoid electric shock, switch off the product power and disconnect the power cord from the
mains power before removing any covers or panels, or opening the case for servicing.
Use care when servicing with poweronDangerous voltages or currents may exist in this product. Disconnect power, remove battery
(if applicable), and disconnect test leads before removing protective panels, soldering, or
replacing components.
Verify safety after repairAlways recheck ground continuity and mains dielectric strength after performing a repair.
Terms in the manual
These terms may appear in this manual:
Warning:
CAUTION: Caution statements identify conditions or practices that could result in damage to this product or other property.
Warning statements identify conditions or practices that could result in injury or loss of life.
Terms on the product
These terms may appear on the product:
•DANGER indicates an injury hazard immediately accessible as you read the marking.
•WARNING indicates an injury hazard not immediately accessible as you read the marking.
•CAUTION indicates a hazard to property including the product.
Symbols on the product
When this symbol is marked on the product, be sure to consult the manual to find out the nature of the potential hazards and
any actions which have to be taken to avoid them. (This symbol may also be used to refer the user to ratings in the manual.)
TBS1000C Series Oscilloscope Specification and Performance Verification 7
The following symbols may appear on the product:
`````````
Important safety information
TBS1000C Series Oscilloscope Specification and Performance Verification 8
Input sensitivity range1 mV/Div to 10V/Div in 1-2-5 sequence with probe attenuation set to 1X
Maximum input voltage300 V RMS, Installation Category II; derate above 4 MHz at 20 dB per decade to 200 MHz
Vertical offset range1 mV/div to 50 mV/div: +/- 1V
100 mV/div to 500 mV/div: +/- 10 V
1 V/div to 5 V/div: +/- 100 V
Bandwidth limit20 MHz (Typ)
Input couplingDC or AC
Input resistance, DC coupled1 MΩ, ±2 %
Input capacitance, DC coupled14 pF ± 2 pF
Vertical zoomVertically expand or compress a live or a stopped waveform
Acquisition modes
SampleAcquire sampled values
Peak DetectCaptures glitches as narrow as 3.5 ns at all
sweep speeds.
AverageFrom 2 to 256 waveforms included in
average.
Hi-ResAverages multiple sample of one acquisition
interval into one waveform point.
RollScrolls waveforms right to left across the
screen at sweep speeds slower than or equal
to 40 ms/div
TBS1000C Series Oscilloscope Specification and Performance Verification 9
Horizontal systems - analog channel
Deskew range+/- 100 ns with 1ns resolution
Horizontal zoomHorizontally expand or compress a live or a stopped waveform
Timebase accuracy25 ppm
Seconds\Division range2 ns/div to 100 sec/div in a 1-2-4 sequence.
Signal acquisition system characteristics
No of input channelsTwo
Input couplingDC or AC
AC coupling connects a capacitor in series with the input circuitry. The DC input impedance
becomes very high, since capacitance is in series with all paths to ground.
Input impedance, DC coupled1 MΩ, ±2 %
Input capacitance, DC coupled14pF ± 2pF
Specifications
Maximum input voltageA300 V RMS, CAT II; derate above 4 MHz at 20 dB per decade to 200 MHz
Number of digitized bits8 bits
Sensitivity range1 mV/div to 10V/div in 1-2-5 sequence with probe attenuation set to 1X.
5, 1, 500 m, 200 m, 100 m, 20 m, 10 m, 1 m V/A current scale factor.
This adjusts the display scale factor of the instrument to accommodate various probe types.
Accuracy of the probe used must be added to the accuracy specifications of instrument.
No automatic probe interface is provided, so you must verify that the settings match the probe
characteristics. The probe check function allows setting of the proper attenuation for voltage
probes.
Acquisition modesSample, Peak detect, Average
Retained front panel settingsFront panel settings are retained when the instrument power is turned off and on with the power
switch. The settings are retained when the line power is turned off and on.
The instrument periodically saves front panel settings after settings are changed. There is a
delay of three seconds after the last change and before the storage of the settings in memory.
Math modesChannel 1 – Channel 2
Channel 2 – Channel 1
Channel 1 + Channel 2
Channel 1 * Channel 2
FFT
TBS1000C Series Oscilloscope Specification and Performance Verification 10
DC balance0.2 div at 1 mVdiv and up with the input 1 MW-DC coupled and 50 W terminated.
✓DC gain accuracy, Sample or
Average acquisition
DC voltage measurement accuracy,
average acquisition mode
Analog bandwidthDefined in Section 4.6 of IEEE std 1057. The difference between the upper and lower
√ Analog bandwidth, DC coupled,
Sample or Average
This is the difference between the measured DC gain and the nominal DC gain, divided by the
nominal DC gain and expressed as a percent.
+/-3.0% step gain, derated at 0.1%/°C above 30°C.
This is the accuracy of DC voltage DC voltage measurements acquired using average of > 16
waveforms.
Average of > 16 waveforms: ±((DC gain accuracy) X |reading - (offset - position)| + offset
accuracy + 0.1 div + 1mV)
frequencies, at which the amplitude response, as seen in the data record, is 0.707 (-3 dB)
of the response seen in the data record at the specified reference frequency. Specifies only the
-3 dB point. It does not include the in-band response.
200 MHz models:
DC to >200 MHz for 1 mV/div through 10 V/div.
100 MHz models:
DC to >100 MHz for 1 mV/div through 10 V/div.
70 MHz models:
DC to >70 MHz for 1 mV/div through 10 V/div.
50 MHz models:
DC to >50 MHz for 1 mV/div through 10 V/div.
Analog bandwidth selections20 MHz bandwidth limit ON/OFF
Lower- Frequency limit, AC coupledThis is the lower frequency for Analog Bandwidth when the instrument is AC-coupled.
Rise time, typical
This is the upper frequency for Analog Bandwidth when the instrument has 20 MHz bandwidth
limiting turned on.
20 MHz, ±25%
Bandwidth of all trigger paths are similarly limited, except the Aux-In, which is not affected by
BW Limit function. Each channel is separately limited, allowing different bandwidths on different
channels of the same instrument.
≤ 10 Hz.
ModelExpected full bandwidth rise time
TBS1202C2.5 ns
TBS1102C4 ns
TBS1072C5.5 ns
Table continued…
TBS1000C Series Oscilloscope Specification and Performance Verification 11
Specifications
ModelExpected full bandwidth rise time
TBS1052C8.4 ns
Rise time is generally calculated from the following formula: Rise time in ns = 420 / Bandwidth
in MHz
√ Vertical position accuracy, offset
accuracy
Vertical position ranges± 5 divisions
Common Mode Rejection Ratio
(CMRR), typical
Crosstalk (Channel Isolation)Section 4.11.1 of IEEE std. 1057. It is the ratio of the level of a signal input into one channel to
This is the accuracy of the nominal voltage level represented by the code at the vendor of the
A-D converter's dynamic range.
± (0.01 X |offset – position | + DC Balance)
With the same signal applied to each channel, CMRR is the ratio of the acquired signal
amplitude to the amplitude of the MATH difference waveform, either (Channel 1 - Channel 2),
(Channel 2 - Channel 1)
ModelCommon Mode Rejection Ratio
TBS1202C
TBS1102C
TBS1072C
TBS1052C
that of the same signal present in another channel due to stray coupling.
ModelCommon Mode Rejection Ratio
TBS1202C (200 MHz models)>100:1 with 200 MHz sine wave and with
TBS1102C (100 MHz models)>100:1 with 100 MHz sine wave and with
TBS1072C (70 MHz models)>100:1 with 70 MHz sine wave and with equal
TBS1052C (50 MHz models)>100:1 with 50 MHz sine wave and with equal
100:1 at 60 Hz, reducing to 10:1 with 50 MHz
sine wave, with equal Volts/Div and Coupling
settings on each channel.
equal V/div settings on each channel.
equal V/div settings on each channel.
V/div settings on each channel.
V/div settings on each channel.
Waveform measurements
Automated measurements32, of which up to six can be displayed on-screen at any one time. Measurements include:
TBS1000C Series Oscilloscope Specification and Performance Verification 12
Specifications
GatingIsolate the specific occurrence within an acquisition to take measurements on, using either the
screen, between waveform cursors or full record length.
Time Base System
Sample-Rate rangeThis is the range of real-time rates, expressed in samples/second, at which a digitizer samples
signals at its inputs and stores the samples in memory to produce a record of time-sequential
samples. (IEEE 1057, 2.2.1)
1 GS/s all channels (TBS1202C, TBS1102C, TBS1072C, TBS1052C)
Waveform interpolationOnly (sin x)/x interpolation is provided.
Record lengthThis is the total number of samples contained in a single acquired waveform record (Memory
Length in IEEE 1057.2.2.1).
20k, 2k and 1k samples per record.
Seconds/Division range
ModelRange
TBS1202C, TBS1102C2 ns/div to 100 sec/div in a 1-2-4 sequence
TBS1072C, TBS1052C
√ Long-Term sample rate and
horizontal position time accuracy
ZoomThe zoom function enables a user to select a part of the display to be magnified. Both the
Delta time measurement accuracyThis is the accuracy of delta time measurements made on any single waveform. The
Time measurement functionsFrequency, Period , Rise, Fall, Pwidth, Nwidth, Pduty, Nduty, DelayRR, DelayRF, DelayFR,
This is the maximum, total, long-term error in sample-rate or horizontal position time accuracy,
expressed in parts per million.
±25 x 10-6 over any >1 ms interval.
original waveform and the zoomed waveform are displayed. The user chooses the waveform
with the Multipurpose knob.
specification is related to the long-term sampling rate.
The following limits are given for signals having an amplitude ≥ 5 divisions, a slew rate at the
measurement points of ≥ 2.0 divisions/ns, and acquired ≥ 10 mV/div.
ConditionTime measurement accuracy
Single shot, sample mode, full bandwidth
selected
> 16 averages, full bandwidth selected
The Sample Interval is the time between the samples in the waveform record.
DelayFF, Burst width, Phase.
±(1 Sample Internal + 25 X 10-6 X |reading| +
0.6 ns)
±(1 Sample Internal + 25 X 10-6 X |reading| +
0.4 ns)
TBS1000C Series Oscilloscope Specification and Performance Verification 13
Specifications
Waveform Math
ArithmeticAdd, Subtract, and Multiply waveforms
FFTSpectral magnitude. Set FFT Vertical Scale to Linear RMS or dBV RMS, and FFT Window to
Rectangular, Hamming, Hanning, or Blackman-Harris.
Triggering system
Trigger types
EdgePositive or negative slope on any channel. Coupling includes DC, HF reject, LF reject, and noise
reject.
Pulse widthTrigger on width of positive or negative pulses that are >, <, =, or ≠ a specified period of time.
RuntTrigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing
the first again.
Trigger types
Trigger source selectionAnalog channels, Aux In (External), AC Line.
Horizontal trigger positionThe trigger position is set by the Horizontal Position knob.
Aux In input impedance1 MΩ ±2 %, 13 pF ±2 pF
Aux In maximum input voltage300 V
EdgePositive or negative slope on any channel.
Coupling includes DC, HF reject, LF reject,
and noise reject.
Pulse WidthTrigger on width of positive or negative pulses
that are >, <, =, or ≠ a specified period of
time.
RuntTrigger on a pulse that crosses one threshold
but fails to cross a second threshold before
crossing the first again.
, Installation Category II; derate at 20 dB/decade above 100 kHz to 13 V peak AC at 3
RMS
MHz and above
Based on sinusoidal or DC input signal. The maximum viewable signal while DC coupled is ±50
V offset ±5 V/div at 4 divisions, or 70 V. AC coupling allows measuring signals on a DC level
up to 300 V. For non sinusoidal waveforms, peak value must be less than 450 V. Excursions
above 300 V should be less than 100 ms duration and the duty factor is limited to < 44%. RMS
signal level must be limited to 300 V. If these values are exceeded, damage to the instrument
may result.
Line trigger characteristicsLine Trigger mode provides a source to synchronize the trigger with the AC line input.
Input Amplitude requirements: 85 VAC - 265 VAC.
Input Frequency requirements: 45 Hz - 440 Hz.
TBS1000C Series Oscilloscope Specification and Performance Verification 14
Specifications
Edge trigger
√ Sensitivity, Edge-Type trigger, DC
coupled
Trigger modesAuto, Normal, Single sequence
Trigger coupling (analog
channels)
Trigger slopeRising edge, Falling edge
Measurement Style A: The minimum signal levels for achieving stable frequency indication on
the Trigger Frequency Counter within 1% of correct indication.
Measurement Style B: Section 4 10.2 in IEEE Std. #1057. The minimum signal levels required
for stable edge triggering of an acquisition when the trigger Source is DC coupled.
Trigger sourceSensitivity
Channel inputsAll products
AC, DC, Noise Reject, High Freq Reject, Low Freq Reject
The Aun In path does not have a DC blocking capacitor
ahead of the trigger input circuit. The roll off associated with
AC coupling happens after the input circuit. When attempting
to trigger on an AC signal that has a DC offset, use care to
avoid overloading the input of the Aux In circuit. For signals
that have a large DC offset, using Channel 1 or Channel 2
with AC coupling is preferred.
Sensitivity
(Measurement style
A), typical
0.4 division from DC
to 50 MHz
(Measurement style
B)
0.8 div from DC to 10
MHz > 2 mV/div)
0.6 divisions >50 MHz
to 100 MHz
0.8 divisions >100
MHz to 200 MHz
TBS1072C3 div between 10 MHz
and 70 MHz
TBS1102C3 div between 10 MHz
and 100 MHz
TBS1052C3 div between 10 MHz
and 50 MHz
TBS1202C3 div between 10 MHz
and 200 MHz
Aux In300 mV from DC to
100 MHz
500 mV from 100
MHz to 200 MHz
TBS1202CTBS1202C
2.5 div from DC to 10
MHz (2 mV/Div)
1.5 div between 10
MHz and 70 MHz
1.5 div between 10
MHz and 100 MHz
1.5 div between 10
MHz and 50 MHz
1.5 div from 10 MHz
and 100 MHz
2.0 div above 100
MHz to 200 MHz
200 mV from DC to
100 MHz
350 mV from 100
MHz to 200 MHz
Trigger frequency feadout typically stabilizes at 50% more signal than generates a stable visual
display.
TBS1000C Series Oscilloscope Specification and Performance Verification 15
Same as DC coupled limits for frequencies 50 Hz and above
Reduces DC coupled trigger sensitivity by 2X.
Same as DC coupled limits for frequencies above 300 kHz.
Trigger frequency counter resolution 3 digits
Trigger frequency counter signal
source
Trigger level accuracy, DC coupled,
typical
Pulse-Width trigger
Edge selected trigger source only
This is the amount of deviation allowed between the level on the waveform at which triggering
occurs and the level selected for DC-coupled triggering signals. A sine wave with 20 ns rise
time corresponds to about 18 MHz.
±0.2 div for signals within ±4 divisions from the center screen, having rise and fall times of ≥ 20
ns.
Aux In: ±(6% of setting + 40 mV) for signals less than ±800 mV
Pulse-Type trigger, minimum
pulse, rearm time, minimum
transition time
TBS1000C Series Oscilloscope Specification and Performance Verification 16
Specifications
Pulse width range33 ns ≤ width ≤ 10 seconds
Pulse width resolution16.5 ns or 1 part per thousand, whichever is larger
Equal guardband
Not equal guardband
Pulse-Width trigger pointEqual: The oscilloscope triggers when the trailing edge of the
t > 330 ns: ±5% < guardband < ±(5.1% + 16.5 ns)
t ≤ 330 ns: guardband = ±16.5 ns.
All pulses, even from the most stable sources, have some
amount of jitter. To avoid disqualifying pulses that are
intended to qualify but are not absolutely correct values,
Tektronix provides an arbitrary guardband. Any measured
pulse width within the guardband will qualify. If you are
looking for pulse width differences that are smaller than
the guardband width, offsetting the center should allow
discriminating differences down to the guardband accuracy.
330 ns < 1: ±5% ≤ guardband < ±(5.1% + 16.5 ns)
165 ns < 1 < 330 ns: guardband = -16.5 ns/+33 ns
t ≤ 165 ns: guardband = ±16.5 ns
All pulses, even from the most stable sources, have some
amount of jitter. To avoid disqualifying pulses that are
intended to qualify but are not absolutely correct values,
Tektronix provides an arbitrary guardband. Any measured
pulse width outside the guardband will qualify. If you are
looking for pulse width differences that are smaller than
the guardband width, offsetting the center should allow
discriminating differences down to the guardband accuracy.
Not equal has slightly better ability to deal with small pulse
widths than equal. The accuracy is not better.
pulse crosses the trigger level.
Not Equal: If the pulse is narrower than the specified
width, the trigger point is the trailing edge. Otherwise, the
oscilloscope triggers when a pulse continues longer than the
time specified as the Pulse Width.
Less than: The trigger point is the trailing edge.
Greater than (also called the time out trigger): The
oscilloscope triggers when a pulse continues longer than the
time specified as the Pulse Width.
Display specifications
Display type7 inch TFT Liquid Crystal Display (LCD).
Display ResolutionThis is the number of individually addressable pixels
800 horizontal by 480 vertical displayed pixels
FormatYT and XY
TBS1000C Series Oscilloscope Specification and Performance Verification 17
Waveform stylesVectors, variable persistence, and Infinite persistence.
Specifications
Luminance, typical
Contrast ratio and control, typicalAvailable black room contrast ratio, full black to full white. 400 minimum, 500 typical.
300 cd/m2, typical.
Interfaces and output ports specifications
USB Interface1 High Speed USB2.0 Host and 1 High Speed USB2.0 Device connector are standard in all
models.
Kensington style lockRear-Panel security slot connects to standard Kensington-style lock
Probe compensator, output voltage
and frequency, typical
The Probe compensator output voltage is in peak-to-peak volts and frequency is in Hertz.
0-5V amplitude +/- 10%, 1 kHz +/- 10%
Data handling characteristics
Retention of front panel settingsFront panel settings are stored periodically in memory. The settings are not lost when the
instrument is turned off or if there is a power failure.
Stored waveforms and multiple front
panel settings
Two Channel 1, Channel 2, or Math waveforms can be stored in nonvolatile waveform memory
A or B. One, both, or neither of A or B waveform memories can be displayed. Ten user
setups of the current instrument settings can be saved and restored from nonvolatile memory.
Additional storage is available when an appropriate mass storage device is connected via USB.
Power distribution system
Power consumption30 W (typ)
Source voltageFull range: 100 to 240 VACRMS ± 10%, Installation Category II (covers range of 90 to 264 VAC)
Power source100 to 240 VAC RMS ±10%
Power source frequency45 Hz to 65 Hz (100 to 240 V).
360 Hz to 440 Hz (100 to 132 V).
Fuse rating3.15 Amps, T rating, 250 V; IEC and UL approved.
Mechanical characteristics
Weight
Instrument only1.9794.36
Instrument with accessories3.1126.86
Cooling clearance50 mm (2 in) required on left side, right side and rear of instrument.
kgLb
TBS1000C Series Oscilloscope Specification and Performance Verification 18
Specifications
RM2000B rackmount
Shipping dimensions
Dimensions
mmIn
Height177.87.0
Width482.619.0
Depth108.04.25
mmIn
Height266.710.5
Width476.218.75
Depth228.69.0
mmIn
Height- Handle folded down154.956.1
Height-Handle up213.368.4
Width325.1212.8
Depth106.684.2
Cooling methodConvection cooled
Environmental performance
Temperature
Operating:0° C to +50° C (32 °F to 122 °F)
Non-operating:–30° C to +71° C (–40 °F to 159.8 °F), with 5° C/minute maximum gradient
Humidity
Operating:5% to 90% relative humidity (% RH) at up to +30° C
5% to 60% RH above +30 °C up to +50 °C, non-condensing.
Non-operating:5% to 90% RH (Relative Humidity) at up to +30 °C,
5% to 60% RH above +30° C up to +60° C, non-condensing
Altitude
Operating:Up to 3,000 meters (9,842 feet)
Non-operating:Up to 12,000 meters (39,370 feet).
Regulatory
ElectromagneticEC Council Directive 2014/30/EU
CompatibilityUL61010-1, UL61010-2-030, CAN/CSA-C22.2 No. 61010.1, CAN/CSA-C22.2 No. 61010-2:030
SafetyCompiles with the Low Voltage Directive 2014/35/EU for Product Safety
TBS1000C Series Oscilloscope Specification and Performance Verification 19
Performance verification
Performance verification
This chapter contains performance verification procedures for the specifications marked with the check mark. The following equipment, or
a suitable equivalent, is required to complete these procedures.
Required equipment
Table 1: Performance verification
DescriptionMinimum requirementsExamples
DC voltage source 17.5 mV to 7 V, ±0.5% accuracyWavetek 9100 Universal Calibration System
with Oscilloscope Calibration Module (Option
Leveled sine wave Generator50 kHz and 200 MHz, ±3% amplitude accuracy
250)
Time mark generator10 ms period, ±10 ppm accuracy
50Ω BNC cableBNC male to BNC male, ≈ 1 m (36 in) longTektronix part number 012-0482-XX
50Ω BNC cableBNC male to BNC male, ≈ 25 cm (10 in) longTektronix part number 012-0208-XX
50 Ω feed through terminationBNC male and female connectorsTektronix part number 011-0049-XX
Dual banana to BNC adapterBanana plugs to BNC femaleTektronix part number 103-0090-XX
BNC T adapterBNC male to dual BNC female connectorsTektronix part number 103-0030-XX
Splitter, powerFrequency range: DC to 4 GHz. Tracking: >2.0% Tektronix part number 015-0565-XX
Adapter (four required)Male N-to-female BNCTektronix part number 103-045-XX
AdapterFemale N-to-male BNCTektronix part number 103-0058-XX
Fluke 5500A Multi-product Calibrator with
Oscilloscope Calibration Option (Option
5500A-SC)
Pomona #B-18-0
Pomona #B-18-2
Test record
Table 2: Test record
Instrument Serial Number:
Temperature:
Date of Calibration:
TBS1000C Series Oscilloscope Specification and Performance Verification 20
Certificate Number:
RH %:
Technician:
Performance verification
Instrument performance test
Channel 1
DC Gain Accuracy
MinimumIncomingOutgoingMaximum
5 mV/div
33.6 mV36.4 mV
200 mV/div1.358 V1.442 V
2 V/div13.58 V14.42 V
Channel 2
DC Gain Accuracy
5 mV/div
33.6 mV36.4 mV
200 mV/div1.358 V1.442 V
2 V/div13.58 V14.42 V
Channel 1 Bandwidth2.12 V
Channel 2 Bandwidth2.12 V
Sample Rate and Delay Time
-2 divs+2 divs
Accuracy
Channel 1 Edge Trigger SensitivityStable trigger
Channel 2 Edge Trigger SensitivityStable trigger
Aux In Edge Trigger SensitivityStable trigger
Channel 1 Vertical Position
0—
Accuracy, Minimum margin
Channel 2 Vertical Position
0—
Accuracy, Minimum margin
—
—
—
—
—
1
1
2
2
2
Performance verification procedures
Before beginning these procedures, two conditions must be met:
•The oscilloscope must have been operating continuously for twenty minutes within the operating temperature range specified in the
Environmental Performance table.
•You must perform the Self Calibration operation described below. If the ambient temperature changes by more than 5 °C, you must
perform the Self Calibration operation again.
The time required to complete the entire procedure is approximately one hour.
Warning:
making or changing any interconnections.
Self test
This internal procedure is automatically performed every time the oscilloscope is powered on. No test equipment or hookups are required.
Verify that no error messages are displayed before continuing with this procedure.
1
The bandwidth test does not have a high limit.
2
The limits vary by model. Check the procedure for the correct limits.
Some procedures use hazardous voltages. To prevent electrical shock, always set voltage source outputs to 0 V before
TBS1000C Series Oscilloscope Specification and Performance Verification 21
Performance verification
Signal calibration
The self calibration routine lets you quickly optimize the oscilloscope signal path for maximum measurement accuracy. You can run the
routine at any time, but you should always run the routine if the ambient temperature changes by 5 °C or more.
1. Disconnect all probes and cables from the channel input connectors (channels 1 and 2).
2. Push the Utility button and select the Do Self Cal option to start the routine. The routine takes approximately one minute to complete.
3. Verify that self calibration passed.
Check DC gain accuracy
This test checks the DC gain accuracy of all input channels.
1. Set the DC voltage source output level to 0 V.
2. Set up the oscilloscope using the following table:
Push menu buttonSelect menu optionSelect setting
Default Setup--
Channel 1Probe1X
AcquireAverage16
MeasureSourceChannel under test
MeasurementsMean
3. Connect the oscilloscope channel under test to the DC voltage source as shown in the following figure:
4. For each vertical scale (volts/division) setting in the following table, perform the following steps:
a. Set the DC voltage source output level to the positive voltage listed and then record the mean measurement as V
b. Reverse the polarity of the DC voltage source and record the mean measurement as V
c. Calculate V
diff
= V
pos
- V
and compare V
neg
to the accuracy limits in the following table:
diff
neg
.
pos
.
Vertical Scale (volts/div)settingDC voltage source output levelsAccuracy limits for V
diff
5 mV/div+17.5 mV, -17.5 mV33.6 mV to 36.4 mV
200 mV/div+700 mV, -700 mV1.358 V to 1.442 V
2 V/div+7.00 V, -7.00 V13.58 V to 14.42 V
5. Set DC voltage source output level to 0 V.
TBS1000C Series Oscilloscope Specification and Performance Verification 22
Performance verification
6. Disconnect the test setup.
7. Repeat steps 1 on page 22 through 6 on page 23 for all input channels.
Check bandwidth
This test checks the bandwidth of all input channels.
1.Set up the oscilloscope using the following table:
Push menu buttonSelect menu optionSelect setting
Default Setup--
Channel 1Probe1X
AcquireAverage16
Trig MenuCouplingNoise Reject
MeasureSourceChannel under test
MeasurementsPeak-Peak
2.Connect the oscilloscope channel under test to the leveled sine wave generator as shown in the following figure:
3.Set the oscilloscope Vertical Scale (volts/division) to 500 mV/div.
4.Set the oscilloscope Horizontal Scale (seconds/division) to 10 μs/div.
5.Set the leveled sine wave generator frequency to 50 kHz.
6.Set the leveled sine wave generator output level so the peak-to-peak measurement is between 2.98 V and 3.02 V.
7.Set the leveled sine wave generator frequency to:
•200 MHz if you are checking a TBS1202C
•100 MHz if you are checking a TBS1102C
•70 MHz if you are checking a TBS1072C
•50 MHz if you are checking a TBS1052C
8.Set the oscilloscope Horizontal Scale (seconds/division) to 10 ns/div.
9.Check that the peak-to-peak measurement is ≥ 2.12 V.
10. Disconnect the test setup.
11. Repeat steps 1 on page 23 through 10 on page 23 for all input channels.
TBS1000C Series Oscilloscope Specification and Performance Verification 23
Check sample rate accuracy and delay time accuracy
This test checks the sample rate and delay time accuracy (time base).
1.Set up the oscilloscope using the following table:
Push menu buttonSelect menu optionSelect setting
Default setup--
Channel 1Probe1X
2.Connect the oscilloscope to the time mark generator as shown in the following figure:
Performance verification
3.Set the time mark generator period to 10 ms.
4.Set the oscilloscope Vertical Scale (volts/division) to 500 mV/div.
5.Set the oscilloscope Main Horizontal Scale (seconds/division) to 1 ms/div.
6.Push the Trigger Level knob to activate the Set To 50% feature.
7.Use the Vertical Position control to center the test signal on screen.
8.Use the Horizontal Position control to set the position to 10.00 ms.
9.Set the oscilloscope Horizontal Scale (seconds/division) to 200 ns/div.
10. Check that the rising edge of the marker crosses the center horizontal graticule line within ±2 divisions of the vertical center graticule
line, as shown in the following figure:
One division of displacement from graticule center corresponds to a 25 ppm time base error.
Note:
TBS1000C Series Oscilloscope Specification and Performance Verification 24
Performance verification
11. Disconnect the test setup.
Check edge trigger sensitivity
This test checks the edge trigger sensitivity for all input channels.
1.Set up the oscilloscope using the following table:
Push menu buttonSelect menu optionSelect setting
Default setup--
Channel 1Probe1X
Trig MenuModeNormal
AcquireSample-
MeasureSourceChannel under test
MeasurementsPeak-Peak
2.Connect the oscilloscope channel under test to the leveled sine wave generator as shown in the following figure:
3.Set the oscilloscope Vertical Scale (volts/division) to 500 mV/div.
4.Set the oscilloscope Horizontal Scale (seconds/division) to 20 ns/div.
5.Set the leveled sine wave generator frequency to 10 MHz.
6.Set the leveled sine wave generator output level to approximately 500 mV
so that the measured amplitude is approximately 500
p-p
mV. (The measured amplitude can fluctuate around 500 mV.)
7.Push the Trigger Level knob to activate the Set To 50%. Rotate the Trigger Level knob to adjust the trigger level as necessary and
then check that triggering is stable.
8.Set the leveled sine wave generator frequency to:
•200 MHz if you are checking a TBS1202C
•100 MHz if you are checking a TBS1102C
•70 MHz if you are checking a TBS1072C
•50 MHz if you are checking a TBS1052C
9.Set the oscilloscope Horizontal Scale (seconds/division) to 4 ns/div.
10. Set the leveled sine wave generator output level to approximately 750 mV
so that the measured amplitude is approximately 750
p-p
mV. (The measured amplitude can fluctuate around 750 mV.)
TBS1000C Series Oscilloscope Specification and Performance Verification 25
Performance verification
11. Push the Trigger Level knob to activate the Set To 50% feature. Rotate the Trigger Level knob to adjust the trigger level as
necessary and then check that triggering is stable.
12. For the TBS1202C model, set the frequency to 150 MHz, and increase the amplitude to 1 Vp-p. Verify stable triggering.
13. Set the oscilloscope Horizontal Scale (seconds/division) to 2 ns/div.
14. Change the oscilloscope setup using the following table:
Push menu buttonSelect menu optionSelect setting
Trig menuSlopeFalling
15. Push the Trigger Level knob to activate the Set To 50% feature. Rotate the Trigger Level knob to adjust the trigger level as
necessary and then check that triggering is stable.
16. Disconnect the test setup.
17. Repeat steps 1 on page 25 through 16 on page 26 for all input channels.
Check Aux In edge trigger sensitivity
This test checks the edge trigger sensitivity for the Aux In.
1.Set up the oscilloscope using the following table:
Push menu buttonSelect menu optionSelect setting
Default setup--
Channel 1Probe1X
Trig MenuSourceAux In
ModeNormal
AcquireSample-
MeasureSourceCH1
MeasurementsPeak-Peak
2.Connect the oscilloscope to the leveled sine wave generator as shown in the following figure, using channel 1 and AUX IN.
TBS1000C Series Oscilloscope Specification and Performance Verification 26
Performance verification
3.Set the oscilloscope Vertical Scale (volts/division) to 100 mV/div.
4.Set the oscilloscope Horizontal Scale (seconds/division) to 20 ns/div.
5.Set the leveled sine wave generator frequency to 10 MHz.
6.Set the sine wave generator output level to approximately 300 mV
into the power splitter. This is about 200 mV
p-p
on channel 1 of
p-p
the oscilloscope.
The AUX IN input will also be receiving approximately 200 mV
. Small deviations from the nominal 200 mV
p-p
oscilloscope display
p-p
are acceptable.
7.Set the leveled sine wave generator frequency to:
•200 MHz if you are checking a TBS1202C
•100 MHz if you are checking a TBS1102C
•70 MHz if you are checking a TBS1072C
•50 MHz if you are checking a TBS1052C
8.Set the oscilloscope Horizontal Scale (seconds/division) to 4 ns/div.
9.Push the Trigger Level knob to activate the Set To 50% feature. Rotate the Trigger Level knob to adjust the trigger level as necessary
and then check that triggering is stable.
10. Set the oscilloscope Horizontal Scale (seconds/division) to 2 ns/div.
11. Push the Trigger Level knob to activate the Set To 50% feature. Rotate the Trigger Level knob to adjust the trigger level as necessary
and then check that triggering is stable.
12. Change the oscilloscope setup using the following table:
Push menu buttonSelect menu optionSelect setting
Trig menuSlopeFalling
13. Push the Trigger Level knob to activate the Set To 50% feature. Rotate the Trigger Level knob to adjust the trigger level as
necessary and then check that triggering is stable.
14. Disconnect the test setup.
TBS1000C Series Oscilloscope Specification and Performance Verification 27
Performance verification
Check Vertical Offset Accuracy
This test checks the offset range for each channel.
1.Connect the oscilloscope to a DC voltage source to run this test. If using the Fluke calibrator as the DC voltage source, connect the
calibrator head to the oscilloscope channel to test.
2.Set the oscilloscope to the factory default settings. Push Default Setup button on the front-panel.
3.Push the Channel button (CH1 or CH2) for the channel that you want to check.
4.Push the Probe Setup button, and then use the Multipurpose knob to set to 1X.
5.Push the Position button, use the Multipurpose knob to select Set To 0 divs if it is not set to 0 already.
6.Push -more- Page 1/2 button for more selection. Set the Bandwidth to 20MHz.
7.Push the Acquire button to set the acquire mode to Average and select for an average of 16 waveforms.
8.Push the Trigger Menu button to set the trigger Source to AC Line.
9.Use the Horizontal Scale knob to set the oscilloscope to 1ms/div.
10. Push Measure button on the front-panel to configure the mean measurement. Push the button next to the channel you want to check
and highlight it as the measurement source.
11. Use the Multipurpose knob to select for Mean measurement.
12. Push the Multipurpose knob to add the Mean measurement.
13. Set the DC Source output value, oscilloscope Vertical Scale, and Offset value as given in the table below.