Tektronix SPG8000A, SPG800AUP Instructions Manual

xx
SPG8000A Master Sync / Clock Reference Generator
ZZZ
SPG800AUP Field Upgrade
Instructions
www.tek.com
*
075-1092-00
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O. Box 500 Beaverto USA
For product information, sales, service, and technical support:
n, OR 97077
In North America, call 1-800-833-9200. Worldwide, visit www.tek.com to nd contacts in your area.
Table of Contents
Important safety information ............... ................................ ................................ ....... 1
General safety summary ...................................................................................... 1
Service safety summary........................ ................................ ............................... 3
Terms in thi
Symbols and terms on the product ........................................................................... 4
Introduction ......................................................................................................... 5
Standard accessories............................................ ................................ ............... 6
Product documentation................. ................................ ................................ ....... 7
How to install and congure Option DPW ...................................................................... 8
Require
How to prevent component damage .. ................................ ................................ ....... 8
Install the backup Power Supply module ...... ................................ ............................. 9
Congure the preferred (active) supply ........................ .................................. .......... 11
How to install an SFP module (Options LX and SX only).......................... .......................... 13
Install the SFP module ....................................................................................... 13
SFP m
SFP module transportation............................. .................................. .................... 14
How to install Option PTP............................................ .................................. .......... 15
Enter the new option key..................................................................................... 15
How to upgrade the instrument rmware....................................................................... 17
Determine if a rmware upgrade is needed................................................................ 17
ckup user les to a USB drive...... ................................ ................................ ...... 19
Ba
Firmware upgrade overview................................................................................. 20
USB rmware upgrade ......................... .................................. ............................ 21
Network rmware upgrade .................................................................................. 23
Verify the rmware upgrade................................................................................. 26
Restore user les from a USB drive .................... ................................ .................... 27
Troubleshoot the rmware upgrade............... ................................ .......................... 28
s manual . ................................ ................................ ......................... 4
d equipment ...................... ................................ ................................ ..... 8
odule removal ..... .................................. ................................ .................. 14
SPG800AUP Field Upgrade Instructions i
Table of Contents
ii SPG800AUP Field Upgrade Instructions
Important safety information
This manual contains information and warnings that must be followed by the user for safe operation and to keep the product in a safe condition.
To safely perform service on this product, additional information is p rovided at the end of this section. (See page 3, Service safety summary.)
General safety summary
Use the product only as specied. Review the following safety precautions to avoid inj Carefully read all instructions. Retain these instructions for future reference.
ury and prevent damage to this product or any products connected to it.
Comply wi
For correct and safe operation of the product, it is essential that you follow general in this manual.
The pro
Only qualied personnel who are aware of the hazards involved should remove the co
Before use, always check the product with a known source to be sure it is oper
This product is not intended for detection of hazardous voltages.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals for warnings and
utions related to operating the system.
ca
When incorporating this equipment into a system, the safety of that system is the
esponsibility of the a ssembler of the system.
r
th local and national safety codes.
ly accepted safety procedures in addition to the safety precautions specied
duct is designed to be used by trained personnel only.
ver for repair, maintenance, or adjustment.
ating correctly.
SPG800AUP Field Upgrade Instructions 1
Important safety information
To avoid re or
personal
injury
Use proper powe
certied for the country of use.
Ground the product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor m ust be connected to earth ground. Before making connections to the input or output terminals of the product, make sure that the product is properly g rounded.
Power disconnect. The power cord disconnects the product from the power source. See instructions for the location. Do not position the equipment so that it is difcult to operate the power cord; it must remain accessible to the u ser at all times to allow for quick disconnection if needed.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating o f that terminal.
Do not operate without covers. Do not operate this product with covers or panels removed, or with the case open. Hazardous voltage exposure is possible.
Avoid exposed circuitry. Do not touch exposed connections and components when powe r is p resent.
r cord. Use only the power cord specied for this product and
Do not operate with suspected failures. If you suspect that there is damage to this
product, have it inspected by qualied service personnel.
Disable the product if it is damaged. Do not use the product if it is damaged or operates incorrectly. If in doubt about safety of the product, turn it off and disconnect the power cord. Clearly mark the product to prevent its further operation.
Before use, inspect voltage probes, test leads, and accessories for mechanical damage and replace when damaged. Do not use probes or test leads if they are damaged, if there is exposed metal, or if a wear indicator shows.
Examine the exterior of the product before you use it. Look for cracks o r missing pieces.
Use only specied replacement parts.
Do not operate in wet/damp conditions. Be aware that condensation may occur if a unit is moved from a cold to a warm environment.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry. Remove the input signals before you clean
the product.
2 SPG800AUP Field U pgrade Instructions
Important safety information
Provide proper
details on installing the product so it has proper ventilation.
Slots and open otherwise obstructed. Do not push objects into any of the openings.
Provide a safe working environment. Always place the product in a location convenient for viewing the display and indicators.
Be sure your work area meets applicable ergonomic standards. Consult with an ergonomics professional to avoid stress injuries.
Use only the Tektronix rackmount hardware specied for this product.
Servicesafetysummary
The Service safety summary section contains additional information required to safely perform service on the product. Only qualied personnel should perform service procedures. Read this Service safety summary and the General safety
summar
To avoid electric shock. Do not touch exposed connections.
ventilation. Refer to the installation instructions in the manual for
ings are provided for ventilation and should never be covered or
y before performing any service procedures.
Do not service alone. Do not perform internal service or adjustments of this product unless another person capable of rendering rst aid and resuscitation is
ent.
pres
Disconnect power. To avoid electric shock, switch off the product power and disconnect the power cord from the mains power before removing any covers or panels, or opening the case for servicing.
e care when servicing with power on. Dangerous voltages or currents may exist
Us
in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
Verify safety after repair. Always recheck ground continuity and mains dielectric strength after performing a repair.
SPG800AUP Field Upgrade Instructions 3
Important safety information
Termsinthismanual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
Symbols and terms on the product
These ter
The following symbol(s) may appear on the product:
ms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the mark
WARNING indicates an injury hazard not immediately accessible as you read th
CAUTION indicates a hazard to property including the product.
ing.
e marking.
When this symbol is marked on the product, be sure to consult the manual to nd out the nature of the potential hazards and any actions which have to betakentoavoidthem.(Thissymbolmayalsobeusedtorefertheuserto ratings in the manual.)
4 SPG800AUP Field U pgrade Instructions
Introduction
Introduction
This document provides instructions for installing the following eld upgrades for the SPG8000A Master Sync / Clock Reference Generator:
Option DPW – adds a second, hot-swappable redundant (backup) Power Supply module
Option LX – adds a Gigabit Ethernet optical SFP module (1310 nm, single-mode, LC connector)
Option SX – adds a Gigabit Ethernet optical SFP module (850 nm, multimode, LC connector)
Option PTP – adds PTP (IEEE 1588) support, including support for SMPTE ST2059 and AES67 proles
Documen
t organization
Firmware upgrade
These instructions are composed of the following sections:
How to install and congure Option DPW describes how to install the backup Power Supply module into the SPG8000A generator. This section also describes how to congure one of the Power Supply module as the preferred (active) supply. (See page 8.)
How to install an SFP module (Options LX and SX only) describes how to install, remove, and transport an SFP module. (See page 13, How to install an
odule (Options LX and SX only).)
SFP m
How to install Option PTP describes how to enter the option k ey that is
uiredtoenableOptionPTP.(Seepage15.)
req
How to upgrade the instrument rmware describes how to upgrade the
grade the instrument rmware to the latest version. (See page 17.)
up
Tektronix releases software and rmware updates for products to add new features and to x product problems. You can nd the latest rmware for your product at the Tektronix Web site (www.tek.com/downloads).
We recommended that you verify your instrument has the latest rmware installed when performing upgrades. (See page 17, How to upgrade the instrument rmware.)
SPG800AUP Field Upgrade Instructions 5
Introduction
Standard acce
ssories
International power cord
options
The following accessories are shipped with the eld upgrade kit:
SPG800AUP Field Upgrade Instructions (this d ocument) Tektronix part number: 075-1092-XX
Option DPW: Power Supply module with the power cord option you ordered (see list below)
Option LX: Gigabit Ethernet optical SFP module (1310 nm, single-mode, LC connector)
Option SX: Gigabit Ethernet optical SFP module (850 nm, multimode, LC connector)
Option PTP: Document listing the unique module ID and Option Key for the instrument you are upgrading
All of the available power cord options except as noted below include a lock mechanism to keep the power cord attached to the instrument.
Opt. A0 – N orth America power (standard)
Opt. A1 – Universal EURO power
A2–UnitedKingdompower
Opt.
Opt. A3 – Australia power
Opt. A5 – Switzerland power
Opt. A6 – Japan power
Opt. A10 – China power
Opt. A11 – India power (no locking cord)
pt. A12 – Brazil power (no locking cord)
O
Opt. A99 – No power cord
6 SPG800AUP Field U pgrade Instructions
Introduction
Product documentation
The user documentation for your product is listed in the following table. Check the Tektronix Web site for the latest version of the documents (www.tek.com/downloads).
Table 1: Product documentation
Availability
Document Tektronix Part Number Description Print Web
Installation and Safety Instructions
User Manual
Specications and Performance Verication
Service Manual
Declassication and Security Instructions
Release Notes 077-1222-xx
Video Sync Pulse Generator and Electronic Changeover Unit System Integration Technical Reference
071-3479-xx (English)
077-1214-xx (Japanese)
077-1215-xx (Russian)
077-1216-xx (English)
077-1217-xx (Japanese)
077-1218-xx (Russian)
077-1219-xx
077-1220-xx Describes how to service the
077-1221-xx Describes how to clear or sanitize
077-0563-xx (ECO422D)
077-0877-xx (ECO8000, ECO8020)
Describes how to install the instrument and provides basic safety and operating information
Provides detailed operating information
Lists the product specications and provides procedures for verifying the performance of the instrument
instrument to the module level (such as circuit boards and fuses)
the data storage (memory) devices in the product for customers with data security concerns.
Describes the new features, improvements, and limitations of the instrument rmware
Provides information for system integrators who are designing systems for high-denition (HD) and standard-denition (SD) digital video content where Tektronix electronic changeover units and video sync pulse generators are to be deployed.
SPG800AUP Field Upgrade Instructions 7
How to install and c ongure Option DPW
How to install
Required
How to pr
equipment
event component damage
and congure Option DPW
This section backup Power Supply module into the SPG8000A generator:
Guidelines
Instructions for installing or replacing the Power Supply module
Instructions for how to congure the preferred (active) supply
No equipment is required to install the backup Power Supply module.
CAUTION. Electrostatic discharge (ESD) can damage components on the modules
and mainframe. To prevent ESD or other component damage, follow the steps below when installing, removing, or handling modules:
provides the following information on how to install the Option DPW
for preventing c omponent damage
Wear a grounded antistatic wrist strap to discharge the static voltage from your body while installing or removing modules from the mainframe.
Transport and store module s in a static-protected bag or container.
Do not slide the module over any surface.
dle modules as little as possible.
Han
Do not touch module components or connector pins.
Do not use any devices capable of generating or holding a static charge in the work area where you remove, install, or handle modules.
Avoid handling modules in areas that have a oor or work-surface covering capable of generating a static charge.
8 SPG800AUP Field U pgrade Instructions
How to install and congure Option DPW
Install the ba
ckup Power Supply module
Perform the following steps to install the backup Power Supply module or to replace a failed supply:
1. If your instrument was shipped with on the blank power supply cover from the Power Supply 2 slot by pushing the retaining latch to the left, and then pulling the cover out of the instrument as shown in the following gure. Proceedtostep3.
ly one Power Supply module, remove
Figure 1: Removing the shield from the Power Supply 2 slot
2. If your instrument has two Power Supply modules installed, perform the following steps to install the new supply:
a. Remove the power cable from the Power Supply module that you are
replacing.
WARNING. Disconnect the p ower cord from a Power Supply module before
you remove the module from the instrument. Also, do not connect a power cord to a Power Supply module while the module is not installed in the instrument. There are dangerous high voltages on the module when the power cord is connected.
CAUTION. To prevent an unexpected shutdown of the instrument, be sure to
unplug the correct power cord if you are replacing a faulty Power Supply module. (See Figure 2.)
When viewed from the front of the instrument, Power Supply 1 is the left Power Supply module and Power Supply 2 is the right Power Supply module. This matches the orientation of the PWR1 and PWR2 indicators on the front panel. The STATUS LED on the rear panel of each Power Supply module will match the status shown by the front-panel PWR1 and PWR 2 indicators.
SPG800AUP Field Upgrade Instructions 9
How to install and c ongure Option DPW
b. Pull out the han
c. While pushing the retaining latch to the left, use the handle to pull the
Power Supply m
3. Insert the replacement supply into the instrument until it latches into place.
4. If necessary, push the handle of the supply into the storage position.
5. Connect the power cable to the Power Supply module you installed.
6. Check that the status of the Power Supply module you just installed is bright
green or dim green.
7. If desired, change the preferred supply conguration. (See page 11, Congure the preferred (active) supply.)
dle located at the bottom of the supply. (See Figure 2.)
odule out of the instrument.
Figure 2: Removing an installed Power Supply module
10 SPG800AUP Field Upgrade Instructions
How to install and congure Option DPW
Congure the p
referred (active) supply
When two Power Supply modules are installed in the instrument, one is congured as the primary supply and the other is congured as the backup supply. In the event that th power to maintain instrument operation.
NOTE. The preferred supply conguration only applies if two Power Supply
modules are installed and both of the supplies are connected to a power source. If there are two Power Supply modules installed but one of the supplies has a failure, the good supply will be used to power the instrument regardless of the preferre
The recommended u sage strategy is to congure one supply as preferred, and use the second supply only in case the rst supply has a failure. This usage strategy allows the backup supply to have the maximum remaining life when it is needed.
The preferred supply conguration is not saved as part of the preset system. Therefore, the preferred supply conguration will not change when you activate any of the system presets, including the Factory Default preset.
Perform the following steps to congure which Power Supply module is the preferred (active ) supply:
e primary supply fails, the backup supply automatically provides
d supply conguration.
1. Check the status of the front-panel PWR1 and PWR2 indicators:
Bright green indicates the preferred (active) supply.
Dim green indicates the inactive (backup) supply.
NOTE. If the c olor of the PWR1 or PWR2 indicators is other than bright or
dim green, service the faulty Power Supply module before proceeding.
2. Check the temperature weighted hours of the installed supplies:
NOTE. It is recommended that you congure the supply with the most temperature
weighted hours to be used as the primary preferred supply. This allows for the newer supply to be preserved so it can provide the best possible backup should the primary supply fail. For more information about temperature weighted hours, see the SPG8000A User Manual.
a. Press the SYSTEM button to access the SYSTEM menu.
b. Press the up ()ordown() arrow button to select SYSTEM :
DIAGNOSTICS, and then press the ENTER button to access the
DIAGNOSTICS menu.
SPG800AUP Field Upgrade Instructions 11
How to install and c ongure Option DPW
c. Press the up (▲)
DIAGNOSTICS : POWER SUPPLY 1. The second line of the display
will list the number of hours the supply has been the active supply and the backup supply.
d. Press the right () arrow button to display the number of temperature
weighted hours of the rst supply as shown below. Note the number of hours of hours shown.
e. Press the up ()ordown() arrow button to select SYSTEM :
DIAGNOSTICS : POWER SUPPLY 2.
f. Press the right () arrow button to display the number temperature
weighted hours of the second supply. Note the number of hours of hours shown.
3. If the Power Supply module with the greatest number of temperature weighted hours i Otherwise, you do not need to change the preferred supply conguration.
4. If you button to access the SYSTEM menu.
s not the currently the preferred supply, then continue this procedure.
want to change the preferred supply conguration, press the SYSTEM
or down () arrow button to select SYSTEM :
5. Pres
6. Pre
NOTE. When viewed from the front of the instrument, Power Supply 1 is the left
Power Supply module and Power Supply 2 is the right Power Supply module. This m
7. If status of the supply is Inactive, press the ENTER button to change the
stheup(▲)ordown(▼)arrowbuttontoselectSYSTEM : POWER
SUPPLY PREFERRED SELECT.
ss the left ()orright(►) arrow button to display the supply you want to be the primary (active) supply: Power Supply 1 or Power Supply 2. The second line of the display shows the supply status: Active, Preferred or Inactive.
atches the orientation of the PWR1 and PWR2 indicators on the front panel.
status to Active, Preferred. The front-panel indicator for the selected supply should change to bright green.
12 SPG800AUP Field Upgrade Instructions
How to install an SFP module (Options LX and SX only)
How to install
an SFP module (Options LX and SX only)
Install the SFP module
There are two installed in the instrument:
Option LX: G
LC connector
Option SX:
connector
To install the SFP module, you will rst need to remove the p lug from the SFP connector. Insert the SFP module into the PTP connector on the rear-panel of the instrument as shown below. The module will latch into place when fully inserted.
optional, optical SFP modules available when Option PTP is
igabit Ethernet optical SFP module, 1310 nm, single-mode,
Gigabit Ethernet optical SFP module, 850 nm, multimode, LC
SPG800AUP Field Upgrade Instructions 13
How to install an SFP module (Options LX and SX only)
SFP module rem
oval
To remove the SFP module, lift up on the latch and then pull the module out of the PTP connector as shown below.
SFP module transportation
The SFP module should be removed from the generator while the instrument is being transported.
CAUTION. To prevent static damage to the SFP module while you are transporting
the instrument, always transport the SFP module in a antistatic bag or container.
14 SPG800AUP Field Upgrade Instructions
How to install Option PTP
How to install
Option PTP
Although you upgrade, we recommended that you verify your instrument has the latest rmware installed when performing upgrades. Proceed as follows:
Enter the new option key
To enable Option PTP, you must enter the option key supplied w ith this kit. Each option key applies to only one instrument.
Locate the option key document, and then perform the following steps to enter the option key:
NOTE. This procedure uses the instrument front panel to enter the option key.
You can also use the SPG8000A Web interface to remotely enter the option key. Refer to the SPG8000A User Manual for information about using the SPG8000A Web i
do not need to upgrade the rmware as part of an Option PTP
Perform the How to upgrade the instrument rmware procedure. (See
page 17.)
Perform the Enter the new option key procedure. (See page 15.)
nterface to enter the option key.
1. Press the front-panel SYSTEM button to access the SYSTEM menu.
2. Press the up ()ordown() arrow button until SYSTEM : OPTIONS is
displayed. The second line of the display lists the options installed in the
strument. Option PTP should not b e already listed.
in
3. Press the ENTER button to enter the OPTIONS submenu.
4. Press the up ()ordown() arrow button until SYSTEM : OPTIONS : UNIQUE ID is displayed.
5. Verify that the displayed unique ID matches the module ID listed on the
option key document.
SPG800AUP Field Upgrade Instructions 15
How to install Option PTP
6. Enter the optio
a. Press the up ()ordown() arrow button to select SYSTEM :
OPTIONS : KEY.
b. Press the ENTER button to enter the option key edit mode. The
underscore
c. Use the up () or down ()arrowbuttontoselecttherst character of
the option k
d. Use the left ()orright(►) arrow button to move the underscore
character
e. Enter all of the option key characters, and then press ENTER button to
conrm th
f. Press the BACK button to exit the OPTIONS submenu.
7. Cycle power on the instrument to ensure the newly enabled option(s) is fully
operational.
8. Verify that Option PTP is enabled:
a. Press the up ()ordown() arrow button until SYSTEM : OPTIONS
is displayed.
nkey:
The existing option key string is displayed.
character (_) appears under the rst character of the option key.
ey.
to the next character in the option key.
e selection.
b. Check that PTP is displayed in the second line of the display. This
veries that Option PTP has been enabled.
c. If Option PTP is not enabled, use the SYSTEM : OPTIONS : UNIQUE
ID display and the SYSTEM : OPTIONS : KEY display to verify that you are updating the correct instrument and that you entered the new option key correctly.
16 SPG800AUP Field Upgrade Instructions
How to upgrade the instrument rmware
How to upgrade
the instrument rmware
Tektronix re and to x product problems. You can nd the latest rmware for your product at the Tektronix Web site (www.tek.com/downloads).
leases software and rmware updates for products to add new features
Determine if a rmware upgrade is needed
Before you begin the upgrade, perform the following steps to determine if the rmware on your instrument should be upgraded and to backup your user les.
CAUTION. The rmware upgrade process replaces all installed user les with the
factory default les. To prevent the loss of your user les, including custom test signals, logo and ID text overlays, and saved presets, be sure to backup all of your user data before you perform the upgrade. The following procedure includes instructions for backing up your user data.
1. Power on the instrument.
2. Verify the current rmware version installed on the instrument:
a. Press the SYSTEM button to access the SYSTEM menu.
b. Press the up ()ordown() arrow button to select SYSTEM :
VERSION INFO (F/W).
c. In the second line of the display, note the rmware version number of
the instrument.
Installed rmware version number
3. Verify the latest version of software at the Tektronix Web site:
a. On a computer, use a Web browser to navigate to the following Tektronix
Web site:
http://www.tek.com/downloads
b. On the Downloads Finder Web page, search by model number (such
as SPG8000A) and lter by software and software type to locate the rmware-upgrade package for the instrument.
c. Note the latest version number of the rmware-upgrade package(s).
Web site rmware version number
SPG800AUP Field Upgrade Instructions 17
How to upgrade the instrument rmware
4. If the latest fir on your instrument, you do not need to upgrade the instrument rmware.
5. If the latest instrument, download the latest rmware package to your computer.
6. After the fir into the directory of your choice: transfer.exe and rmware.pkg.
NOTE. If a readme.txt file is included in the rmware package, be sure to read the
le before you upgrade the rmware.
7. If the rmware in your instrument needs to be upgraded, verify which options are currently installed before you install the software. After you install the upgrade, you will verify that any previously installed options are still enabled.
a. Connect the instrument and computer to your local Ethernet network.
b. Connect the power cable to power on the instrument.
c. Press the SYSTEM button to access the SYSTEM menu.
d. Press
OPTIONS. The second line of the display lists the hardware options
(GPS, BG, AG, SDI) and software option (PTP) that are installed in the instrument as shown below.
mware version at the Web site is the same as the version installed
rmware version at the Web site is newer than the version on your
mware package ZIP le is downloaded, extract the following les
the up ()ordown(▼) arrow button to select SYSTEM :
e. Ma
8. Proceed as follows:
18 SPG800AUP Field Upgrade Instructions
ke a note of the installed options.
If your instrument needs a rmware upgrade, proceed to Backup user
les to a USB drive.(Seepage19.)
If your instrument does not need a rmware upgrade, proceed to Enter
the new option key.(Seepage15.)
How to upgrade the instrument rmware
Backup user fil
es to a USB drive
If your instrument has custom user les installed, perform the following steps to backup the user les to a USB drive. You will restore the user les at the end of the upgrade p
1. Insert a USB memory device into the USB port on the front of the instrument.
2. Verify that the USB device is properly installed:
a. Press the SYSTEM button to access the SYSTEM menu.
b. Use the up ()ordown()arrowbuttontoselectSYSTEM : USB
(Mounted).
NOTE. The display will read “Mounted” if the USB device is properly
installed. If the display reads “Unmounted” or “Not Connected”, try to
ll the USB device.
reinsta
If the instrument still does not recognize the USB device, attempt this
ure again using a different USB device from a well known brand that
proced is 16 GB or less.
3. Use the up ()ordown() arrow button to select SYSTEM : BACKUP/RESTORE.
rocedure.
4. Use the left ()orright() arrow button to select Backup All User Data to USB, and then press the ENTER button.
TION. To prevent the loss of your user les, be sure to use the Backup All
CAU
User Data to USB option in the backup procedure.
5. In the overwrite conrmation display, press the ENTER button to perform
he selected backup function.
t
6. During the backup process, progress messages are displayed. When the message Backup completed successfully is displayed, press any button to return to the SYSTEM menu display.
7. Use the up ()ordown()arrowbuttontoselectSYSTEM : USB (Mounted).
8. Press the ENTER button to safely unmount the USB device. You can now
use the USB device to transfer the saved les to another SPG8000A generator or to re store user les after a rmware upgrade.
SPG800AUP Field Upgrade Instructions 19
How to upgrade the instrument rmware
Firmware upgrade overview
There are two methods for upgrading the instrument rmware:
Using a USB memory device (See page 21, USB rmware upgrade.)
UsinganEthe rmware upgrade.)
The time req
CAUTION. The instrument automatically reboots at the end of the upgrade process.
Do not perform the upgrade while the instrument is in use as a sync reference.
Once the instrument has started to erase internal ash, do not remove power from the instrument. If you do so, the instrument ash may be corrupted and you will have to run the recovery procedure. (See page 28, Troubleshoot the rmware
NOTE. If power to the instrument is lost before it begins erasing the internal
ash, y the instrument displays only a blank screen after power is reapplied, see the troubleshooting section for help. (See page 28, Troubleshoot the firmware upgrade.)
If you have loaded the SPG8000A memory with a large number of test signals or frame picture les, you may not be able to upgrade the instrument rmware because the memory is too full. If you receive a memory error while attempting to upgrade the rmware, you need to delete some of the test signal or frame picture le
upgrade.)
ou can restart the rmware upgrade after the instrument reboots. If
s and then perform the upgrade.
rnet network (requires a Windows PC) (See page 23, Network
uired to complete the rmware upgrade is approximately 10 minutes.
20 SPG800AUP Field Upgrade Instructions
How to upgrade the instrument rmware
USB rmware up
grade
Use the following steps to upgrade the instrument rmware using a USB memory device.
1. Insert a USB memory device into a USB port of the computer containing the latest instrument rmware.
2. On the computer, navigate to the USB device.
3. On the USB device, create a directory named SpgUpgrd.
CAUTION. To prevent upgrade failure, be sure to corre ctly name the upgrade
directory. The directory name is case sensitive, with the “S” and the “U” in SpgUpgrd
4. Open the SpgUpgrd directory and copy the rmware.pkg le from the
5. Eject
6. If necessary, power on the instrument and wait for the instrument to fully
being upper case.
downlo directory path of SpgUpgrd\rmware.pkg.
boot u instrument.
aded ZIP le to the USB device. The USB device should have a
and safely remove the USB device from the computer.
p before inserting the USB device into the front-panel USB port on the
7. Pres
8. Verify that the USB device is properly installed by pressing the up ()or
NOTE. The display will read “Mounted” if the USB device is properly installed. If
the display reads “Unmounted”, try to reinstall the USB device.
If the instrument still does not recognize the USB device, attempt this procedure again using a different USB device from a well known brand that is 16 GB or less.
9. Press the up ()ordown(▼)arrowbuttontoselectSYSTEM :
stheSYSTEM button to access the SYSTEM menu.
n(▼) arrow button to select SYSTEM : USB (Mounted).
dow
FIRMWARE UPGRADE.
SPG800AUP Field Upgrade Instructions 21
How to upgrade the instrument rmware
10. Press the left ( press the ENTER button. A conrmation message is displayed as shown below.
NOTE. If the instrument displays the message ERROR: SpgUpgrd/rmware.pkg
not found, make sure you have the correct directory path on the USB device and
then restart this procedure at step 6.
11. A second conrmation is displayed as shown below to remind you that the upgrade w
CAUTION. If you did not backup your user les as part of the upgrade preparation
process, it is recommended that you press the BACK button to cancel the upgrade and then backup your user les before performing the upgrade. The instructions for backing up user les is described in the section To transfer user les from an instrument to a USB drive. (See page 19.) You will restore the user les at the end of
the upgrade procedure.
) or right () arrow button to select USB Upgrade,andthen
ill delete all user-created les on the instrument.
12. Press the ENTER button to proceed with the upgrade or press the BACK button to cancel the upgrade.
13. When you press ENTER, the upgrade process begins. During the upgrade process, the instrument displays messages indicating the upgrade status.
CAUTION. Once the instrument has started to erase internal ash, do not remove
power from the instrument. If you do so, the instrument ash may be corrupted and you will have to run the recovery procedure. (See page 28, Troubleshoot
he rmware upgrade.)
t
14. If the instrument does not reboot at the completion of the rmware upgrade, remove and then reconnect the power cable.
22 SPG800AUP Field Upgrade Instructions
How to upgrade the instrument rmware
15. Remove the USB d
CAUTION. To prevent a loss of data on the USB device, use the SYSTEM menu to
unmount the USB device before removing it from the instrument.
16. Proceed t
Network rmware upgrade
Use the f connection:
1. Connec
evice after the upgrade is complete:
a. Press the SYSTEM button.
b. Press the up
(Mounted), and then press the ENTER button.
c. Remove the
now safely remove the USB drive appears.
o Verify the rmware upgrade. (Seepage26.)
ollowing steps to upgrade the instrument rmware using a network
t the instrument and a Windows PC to your local Ethernet network or
directly connect the instrument to a Windows PC.
()ordown(▼)arrowbuttontoselectSYSTEM : USB
USB device from the instrument when the message You may
2. Power
3. Prepare the instrument for the rmware upgrade:
a. Press the SYSTEM front panel button.
b. Press the up ()ordown() arrow button to select SYSTEM :
c. Press the left ()orright()arrowbuttontoselectNetwork Upgrade,
on the instrument.
FIRMWARE UPGRADE.
andthenpresstheENTER button. A conrmation message is displayed as shown below.
SPG800AUP Field Upgrade Instructions 23
How to upgrade the instrument rmware
d. Asecondconfirm
the upgrade will delete all user-created les on the instrument.
CAUTION. If you did not backup your user les as part of the upgrade
preparation process, it is recommended that you press the BACK button to cancel the upgrade and then backup your user les before performing the upgrade. The instructions for backing up user les is described in the section To trans f e r will restore the user les at the end of the upgrade procedure.
e. Press the ENTER button to proceed with the upgrade or press the BACK
button to cancel the upgrade.
f. When you press ENTER, a Waiting for network connection message is
displayed along with the IP address of the instrument. This indicates the instrument is ready to be upgraded. An example display is shown below.
user les from an instrument to a USB drive. (See page 19.) You
ation is displayed as shown below to remind you that
g. If no network connection is established, cycle power on the instrument
and verify the network connection before attempting the rmware upgrade again. See the SPG8000A User Manual for details on conguring
network connections.
the
24 SPG800AUP Field Upgrade Instructions
How to upgrade the instrument rmware
4. Perform the rm
a. On the PC, double-click the transfer.exe le to launch the transfer
program.
b. Type the IP address or DNS name of the instrument you are upgrading,
and then pre indicating the progress of the upgrade appear in the transfer program window and on the instrument display.
CAUTION. Once the instrument has started to erase internal ash, do not
remove power from the instrument. If you do so, the instrument ash may be corrupted and you will have to run the recovery procedure. (See page 28, Troubles
NOTE. The upgrade will not proceed until all of the rmware data is
transfe will attempt to complete the upgrade even if it loses contact with transfer.exe. So that you can monitor the progress of the upgrade, it is suggested that you do not close the transfer.exe window until the transfer utility asks for another IP address.
hoot the rmware upgrade.)
rred to the instrument. Once the upgrade has started, the instrument
ware upgrade:
ss Enter.Thisstartsthermware upgrade process. Messages
Figure 3: Sample of transfer.exe window after the upgrade is complete
5. If the instrument does not reboot at the completion of the rmware upgrade, remove and then reconnect the power cable.
6. Proceed to Verify the rmware upgrade. (Seepage26.)
SPG800AUP Field Upgrade Instructions 25
How to upgrade the instrument rmware
Verify the rm
ware upgrade
1. If your instrument had a failure during the rmware upgrade, proceed to
Troubleshoot the rmware upgrade. (See page 28.)
2. After the instrument reboots at the end of the upgrade process, press the SYSTEM button to access the SYSTEM menu.
3. Press the up ()ordown()arrowbuttontoselectSYSTEM : OPTIONS.
The second line of the display lists the hardware options (GPS, BG, AG, SDI) and so as shown below.
4. Verify that the options listed match those that you noted before the upgrade.
5. Press the up () or down ()arrowbuttontoselectSYSTEM : VERSION INFO (F/W). The second line of the display lists the installed rmware
version.
6. Verify that the displayed rmware version is the same version as the rmware package you used to perform the upgrade.
ftware options (3G, DBT) that are installed in the instrument
7. Proceed as follows:
If you backed up your user les to a USB drive before you upgraded the
rmware, proceed to Restore user les from a USB drive.
If you did not backup your user les, proceed to Enter the new option
key.(Seepage15.)
26 SPG800AUP Field Upgrade Instructions
How to upgrade the instrument rmware
Restore user
les from a USB drive
If you backed u p your user les to a USB drive before you upgraded the rmware, perform the following steps to restore the user les:
1. Insert a USB memory device containing the user les you want to install on the SPG8000A generator into the USB port on the front of the instrument.
2. Verify that the USB device is properly installed:
a. Press the SYSTEM button to access the SYSTEM menu.
b. Use the up ()ordown()arrowbuttontoselectSYSTEM : USB
(Mounted).
NOTE. The display will read “Mounted” if the USB device is properly
installed. If the display reads “Unmounted” or “Not Connected”, try to
ll the USB device.
reinsta
If the instrument still does not recognize the USB device, attempt this
ure again using a different USB device from a well known brand that
proced is 16 GB or less.
3. Use the up ()ordown() arrow button to select SYSTEM : BACKUP/RESTORE.
4. Use the left ()orright()arrowbuttontoselectRestore All User Data From USB, and then press the ENTER button.
TION. To prevent the loss of your user les, be sure to use the Restore All
CAU
User Data from USB option.
5. In
6.D
7. Use the up ()ordown()arrowbuttontoselectSYSTEM : USB
8. Press the ENTER button to safely unmount the USB device.
9. Proceed to Enter the new option key. (Seepage15.)
the overwrite conrmation display, press the ENTER button to perform
the selected restore function.
uring the restore process, progress messages are displayed. When the message Restore completed successfully is displayed, press any button to return to the SYSTEM menu display.
(Mounted).
SPG800AUP Field Upgrade Instructions 27
How to upgrade the instrument rmware
Troubleshoot
the rmware upgrade
If a power failure occurs during the upgrade and the instrument has a blank screen when you reapply power, perform the following procedure.
1. Connect an Ethernet cable between the instrument and your computer or local Ethernet network.
2. Remove the power cable to turn off the instrument.
3. Press and hold the STATUS button and then reconnect the power cable. Continue to hold the STATUS button until the message Select upgrade network speed: appears on the top line of the LCD display.
4. Press the left ()orright() arrow button to select a network speed
(10BaseT, 100BaseT, or 1000BaseT), and then press the ENTER button. If you do recommended.
5. When th left ()orright(►) arrow button to select from the following modes and then press the ENTER button.
not know your network speed, the 100BaseT selection is
e message Select upgrade network mode: is displayed, press the
DHCP – Use this selection if the instrument is connected to a network
with DHCP support. An IP address will be assigned to the instrument from your network server.
Local IP (192.168.1.1) – Use this selection i f the instrument is connected
to a network without DHCP support or is connected to a standalone computer, and you want to use the factory default, static IP addres s of
192.168.1.1 for the upgrade.
TE. When you select either the Local IP or the Saved static IP upgrade
NO
mode, you cannot edit the displayed IP address. The IP address cannot be edited when the instrument is in recovery mode.
Saved static IP (IP address) – Use this selection if the instrument is
connected to a network without DHCP support or is connected to a standalone computer, and you want to use the previously saved static IP address assigned to the instrument for the upgrade. The saved IP address is displayed.
Quit and continue booting – Use this selection to quit the upgrade
process and leave the instrument in its current condition.
6. If you did not select Quit and continue booting in the previous step, the top line of the instrument display will show the message Waiting for network connection and the second line will list an IP address. Make a note of the displayed IP address.
28 SPG800AUP Field Upgrade Instructions
How to upgrade the instrument rmware
7. Enter the IP add application window on your PC to restart the upgrade process. See step 4 of the network rmware upgrade procedure. (See page 25.).
CAUTION. To prevent upgrade failure, leave the instrument power on during the
upgrade process. The upgrade process may take up to 15 minutes to complete.
8. When the upgrade is complete, the instrument should reboot. If it does not, cycle the power.
If you continue to experience an upgrade problem, contact Tektronix customer support or your local Tektronix representative for assistance.
ress you noted in the previous step into the upgrade transfer
SPG800AUP Field Upgrade Instructions 29