Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specifications and price change privileges rese rved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O . Bo x 50 0
Beaverto
USA
For product information, sales, service, and technical support:
n, OR 97077
In North America, call 1-800-833-9200.
Worl d wide , visi t www.tektronix.com to find contacts in your area.
Service Safety Summary
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary located in your instrument’s
Service manu
before performing any service procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnect Power. To avoid electric shock, switch off the instrument power, then
disconnect the power cord from the mains power.
UseCareWhenServicingWithPowerOn. Dangerousvoltagesorcurrentsmay
exist in this product. Disconnect power, remove battery (if applicable), and
nect test leads before removing prote ctive panels, soldering, or replacing
discon
components.
al (the SPECMONB Service manual part number is 077-0909-XX)
To avo
id electric shock, do not touch exposed connections.
SPECMONB Series Memory Upgrade1
Kit description
Kit description
Products
Option 53 kit parts list
This kit describes the installation of the following upgrade options:
Option 53: Acquisition Memory
SPECMON3B Series. All instruments
SPECMON6B Series. All instruments
SPECMON26B Series. All instruments
y
Quantit
1075-1065-XX
1071-2191-XX
1167-1
1
1
Part num
N/ALABEL, MANUFACTURED; OPTION KE Y
N/ALABEL, MANUFACTURED; PRODUCT LABEL,
ber
473-XX
Descrip
SPECMONBUP OPTION 53 UPGRADE
INSTRU
MANUAL, TECH; SVCPT-UPG; TEKTRONIX
SUPPL
THE PEOPLES REPUBLIC OF CHINA; CHINA
ROHS
This section contains all procedures needed to install the required components.
Installation instructions
Minimum tool
Remo
ve cosmetic covers and shield
and equipment list
The following tools are required to for installation of this kit. All tools are
standard tools that are readily available.
Item
no.NameDescription
1
2
3
These i
with servicing the product. If you need further details for disassembling or
reassembling the product, refer to the product service manual. Service manuals
can be downloaded from the web at www.tektronix.com/manuals. Search by
model number to find the service manual for your instrument.
Screwdriver
handle
(magnetic)
T15 TORX
tip
T20 TORX
tip
nstructions are for qualified service personnel who are familiar
TORX-driver tip for T15 size screws on the instrument covers
TORX driver tip for T20 size screw heads
NOTE. Right-side or left-side references in these instructions assume you are
viewing the instrument from the front panel.
WAR N ING. To avoid electric shock which may result in injury or loss of life, switch
off the instrument power, then disconnect the power cord from the mains power.
1. Remove the power cord.
f it is installed, pull the front protective cover off the instrument.
2.I
3. Remove the two T15 Torx-head screws that secure the plastic carrying handle
to the side of the instrument. (It is not necessary to remove the black metal
handles.) (See Figure 1.)
SPECMONB Series Memory Upgrade3
Installation instructions
4. Remove four T15
bottom covers to the instrument, and two T20 Torx-head screws near the front
edge of the top cover (next to the folding handles).
5. Remove the top cover (leave the bottom cover in place). Remove the top
cover by pulling straight back about 1 inch. Then pull the sides of the top
cover outward, flexing them slightly to clear the instrument chassis, and pull
it away from the instrument.
6. Remove the eighteen T15 Torx-head screws that secure the top shield to the
chassis and remove the shield.
Torx-head screws along each side that secure the top and
Figure 1: Removing covers
4SPECMONB Series Memory Upgrade
Replace the Existing Acquisition Memory
The acquisition memory is located near the rear panel, on the left side.
Installation instructions
1. Use the
Place the memory module into a static-shielded bag.
2. In
the gold “fingers” are fully seated. Then rotate the module down toward the
board until the latches click into place.
ejector latches to release the existing acquisition memory module.
sert the new memory module at an approximately 45° angle, making sure
SPECMONB Series Memory Upgrade5
Installation instructions
Reinstall top shield and cosmetic covers
1. Place the instrument on its bottom feet.
2. Place the top shield on top of the instrument and reinstall the eighteen T15
Torx-head screws.
3. Place the instrument on the rear feet, so the front panel is facing up and the
top is toward you.
4. Place the top cover over the top of the instrument and slide it toward the
front panel. Make sure that the top cover wraps around the flanges on the
rear panel on all three sides.
5. ReinstallthefourT20Torxheadscrews(twooneachside)nearthefront
edge of the top cover (next to the folding handles) that secure the top cover to
the instrument. Torque these screws to 8.0 in/lb.
6. Rotate the instrument so the bottom faces you.
7. Place the bottom cover on the instrument, with the flip feet towards the front.
8. Align the four screw holes on each side in the top and bottom covers with the
holes in the chassis, and install eight T15 Torx-head screws, four on each
side. Torque these screws to 8.0 in/lb.
9. Position the plastic carrying handle and its bracket on the right side of the
instrument, and install the two T15 screws that secure it in place. Torque
these screws to 8.0 in/lb.
6SPECMONB Series Memory Upgrade
Install option key
Installation instructions
To activate your new option, you must enter a new option key.
1. Power on the instrument.
NOTE. When the application launches, it will display an error message indicating
that the current option key does not support the new option hardware. Click
OK to clear the error message.
2. Verify the latest version of instrument software is installed. Tektronix
recommends to update to the latest available software.
a. Select Help > About to check the software version.
b. Use your Web browser to go to: www.tektronix.com/software.
c. Enter your instrument model number in the Search by keyword text box
and click Go.
d. Follow the link for the analyzer software.
e. If the
f.Foll
3. Select Tools > Install Upgrades... to start the upgrade installation process.
4. Click Continue from the Install Upgrades introduction screen.
installed software is older than that available, download the
software.
ow the instructions on the Web page to install the software.
SPECMONB Series Memory Upgrade7
Installation instructions
5. Enter the optio
instructions to install the option.
n key provided by Tektronix, and follow the on-screen
Attach labels
ach the option key label
Att
Attach the product/option
label
6. Power off the instrument, then power back on.
7. In the H
8. Verify the new option is listed.
Please attach all labels provided in this kit on the instrument’s rear panel.
Place the new option key label over the existing label on the rear panel.
Place the new product label over the existing label on the rear panel.
elp menu, select About Tektronix Real-Time Analyzer.
8SPECMONB Series Memory Upgrade
Verify instrument software version
Verify instru
ment software version
Tektronix re
your instrument.
1. Select Help
2. Use your Web browser to go to: www.tektronix.com/software.
3. Search for your instrument’s model number and follow the link to the software.
4. If the installed software is older than that available, download the software.
5. Follow the instructions on the Web page to install the software.
■ End of D
commends to update to the latest available application software for
> About to check the software version.
ocument ■
SPECMONB Series Memory Upgrade9
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.