Tektronix SPECMONBUP (Option 53) Field Install Instructions

xx
SPECMONBUP (Option 53) SPECMONB Series Real-Time Spectrum Analyzers
ZZZ
Acquisition Memory Upgrade
Instructions
www.tektronix.com
*
075-1065-00
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specications and price change privileges rese rved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O . Bo x 50 0 Beaverto USA
For product information, sales, service, and technical support:
n, OR 97077
In North America, call 1-800-833-9200. Worl d wide , visi t www.tektronix.com to nd contacts in your area.
Service Safety Summary
Only qualied personnel should perform service procedures. Read this Service Safety Summary and the General Safety Summary located in your instrument’s
Service manu before performing any service procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this product unless another person capable of rendering rst aid and resuscitation is present.
Disconnect Power. To avoid electric shock, switch off the instrument power, then disconnect the power cord from the mains power.
UseCareWhenServicingWithPowerOn. Dangerousvoltagesorcurrentsmay exist in this product. Disconnect power, remove battery (if applicable), and
nect test leads before removing prote ctive panels, soldering, or replacing
discon components.
al (the SPECMONB Service manual part number is 077-0909-XX)
To avo
id electric shock, do not touch exposed connections.
SPECMONB Series Memory Upgrade 1
Kit description
Kit description
Products
Option 53 kit parts list
This kit describes the installation of the following upgrade options:
Option 53: Acquisition Memory
SPECMON3B Series. All instruments
SPECMON6B Series. All instruments
SPECMON26B Series. All instruments
y
Quantit
1 075-1065-XX
1 071-2191-XX
1 167-1
1
1
Part num
N/A LABEL, MANUFACTURED; OPTION KE Y
N/A LABEL, MANUFACTURED; PRODUCT LABEL,
ber
473-XX
Descrip
SPECMONBUP OPTION 53 UPGRADE INSTRU
MANUAL, TECH; SVCPT-UPG; TEKTRONIX SUPPL THE PEOPLES REPUBLIC OF CHINA; CHINA ROHS
IC, MEMORY, DRAM; 512MX64, DDR2, 667MHZ, MT16HTS51264HY-667A1, SODIMM 200
UPGRADE LABEL
SAFETY CONTROLLED
tion
CTIONS
EMENTAL INFORMATION SHEET FOR
2 SPECMONB Series Memory Upgrade
Installation instructions
This section contains all procedures needed to install the required components.
Installation instructions
Minimum tool
Remo
ve cosmetic covers and shield
and equipment list
The following tools are required to for installation of this kit. All tools are standard tools that are readily available.
Item no. Name Description
1
2
3
These i with servicing the product. If you need further details for disassembling or reassembling the product, refer to the product service manual. Service manuals can be downloaded from the web at www.tektronix.com/manuals. Search by model number to nd the service manual for your instrument.
Screwdriver handle (magnetic)
T15 TORX tip
T20 TORX tip
nstructions are for qualied service personnel who are familiar
Torque driver handle. Accepts 1/4-in. hex-head driver tips
TORX-driver tip for T15 size screws on the instrument covers
TORX driver tip for T20 size screw heads
NOTE. Right-side or left-side references in these instructions assume you are
viewing the instrument from the front panel.
WAR N ING. To avoid electric shock which may result in injury or loss of life, switch
off the instrument power, then disconnect the power cord from the mains power.
1. Remove the power cord.
f it is installed, pull the front protective cover off the instrument.
2.I
3. Remove the two T15 Torx-head screws that secure the plastic carrying handle
to the side of the instrument. (It is not necessary to remove the black metal handles.) (See Figure 1.)
SPECMONB Series Memory Upgrade 3
Installation instructions
4. Remove four T15 bottom covers to the instrument, and two T20 Torx-head screws near the front edge of the top cover (next to the folding handles).
5. Remove the top cover (leave the bottom cover in place). Remove the top cover by pulling straight back about 1 inch. Then pull the sides of the top cover outward, exing them slightly to clear the instrument chassis, and pull it away from the instrument.
6. Remove the eighteen T15 Torx-head screws that secure the top shield to the chassis and remove the shield.
Torx-head screws along each side that secure the top and
Figure 1: Removing covers
4 SPECMONB Series Memory Upgrade
Replace the Existing Acquisition Memory
The acquisition memory is located near the rear panel, on the left side.
Installation instructions
1. Use the Place the memory module into a static-shielded bag.
2. In the gold “fingers” are fully seated. Then rotate the module down toward the board until the latches click into place.
ejector latches to release the existing acquisition memory module.
sert the new memory module at an approximately 45° angle, making sure
SPECMONB Series Memory Upgrade 5
Installation instructions
Reinstall top shield and cosmetic covers
1. Place the instrument on its bottom feet.
2. Place the top shield on top of the instrument and reinstall the eighteen T15
Torx-head screws.
3. Place the instrument on the rear feet, so the front panel is facing up and the top is toward you.
4. Place the top cover over the top of the instrument and slide it toward the front panel. Make sure that the top cover wraps around the anges on the rear panel on all three sides.
5. ReinstallthefourT20Torxheadscrews(twooneachside)nearthefront edge of the top cover (next to the folding handles) that secure the top cover to the instrument. Torque these screws to 8.0 in/lb.
6. Rotate the instrument so the bottom faces you.
7. Place the bottom cover on the instrument, with the ip feet towards the front.
8. Align the four screw holes on each side in the top and bottom covers with the
holes in the chassis, and install eight T15 Torx-head screws, four on each side. Torque these screws to 8.0 in/lb.
9. Position the plastic carrying handle and its bracket on the right side of the instrument, and install the two T15 screws that secure it in place. Torque these screws to 8.0 in/lb.
6 SPECMONB Series Memory Upgrade
Install option key
Installation instructions
To activate your new option, you must enter a new option key.
1. Power on the instrument.
NOTE. When the application launches, it will display an error message indicating
that the current option key does not support the new option hardware. Click OK to clear the error message.
2. Verify the latest version of instrument software is installed. Tektronix recommends to update to the latest available software.
a. Select Help > About to check the software version.
b. Use your Web browser to go to: www.tektronix.com/software.
c. Enter your instrument model number in the Search by keyword text box
and click Go.
d. Follow the link for the analyzer software.
e. If the
f. Foll
3. Select Tools > Install Upgrades... to start the upgrade installation process.
4. Click Continue from the Install Upgrades introduction screen.
installed software is older than that available, download the
software.
ow the instructions on the Web page to install the software.
SPECMONB Series Memory Upgrade 7
Installation instructions
5. Enter the optio instructions to install the option.
n key provided by Tektronix, and follow the on-screen
Attach labels
ach the option key label
Att
Attach the product/option
label
6. Power off the instrument, then power back on.
7. In the H
8. Verify the new option is listed.
Please attach all labels provided in this kit on the instrument’s rear panel.
Place the new option key label over the existing label on the rear panel.
Place the new product label over the existing label on the rear panel.
elp menu, select About Tektronix Real-Time Analyzer.
8 SPECMONB Series Memory Upgrade
Verify instrument software version
Verify instru
ment software version
Tektronix re your instrument.
1. Select Help
2. Use your Web browser to go to: www.tektronix.com/software.
3. Search for your instrument’s model number and follow the link to the software.
4. If the installed software is older than that available, download the software.
5. Follow the instructions on the Web page to install the software.
End of D
commends to update to the latest available application software for
> About to check the software version.
ocument
SPECMONB Series Memory Upgrade 9
Loading...