This manual provides tutorial examples of how to use the SPECMON3 and SPECMON6 Real-Time Spectrum Analyzers to
take measurements in different application areas. To work through these examples on your instrument, you can use either
the sample data files provided on your hard drive or a live signal of your choice. If you use your own signal, you need to
reset the instrument to match your signal's parameters.
NOTE. You can use the mouse, keyboard, and touch screen to perform all of the tasks in this manual. Additionally, you can
use the knob and buttons on the analyzer front panel as shortcuts to perform some of the tasks.
You can operate y our analyzer like a conventional spectrum analyzer. The following example leads you through basic
functions of frequency, span, and shows y ou how to make amplitude and frequency measurements with markers.
1. Push the front-panel Preset button to set
the instrument to the default settings.
The following steps set up the appropriate
measurement parameters for the sample
signal.
2. Click Freq in the application menu bar
and type in 2GHz.
2 GHz is the frequency of the saved
signal that you will recall in a later step.
Enter the value with the front-panel
keypad or an external keyboard attached
through the USB port.
: Making a Basic Spectrum Measurement
3. Click the Settings icon in the application
menu bar.
4. Set the span to 1MHzin the resulting
Spectrum Settings lower screen pane.
To locate the Span screen item, be sure
the Freq & Span tab is selected.
5. Select File > Recall.
The file that you will recall is a saved
data file. It mimics a live signal for the
purpose of this example application.
6. Go to: C:/SPECMON
Files/SampleDataRecords.
Select Acq data with setup (TIQ) in the
drop-down list field.
The instrument places a m arker on the
highest level peak of the spectrum. It
displays the marker measurement in the
upper left of the display.
The first marker is labeled MR to indicate
that it is the reference marker.
10. Click M arkers Define in the bottom left
of the display.
Application 1: M
aking a Basic Spectrum Measurement
This br ing
control panel.
11. Click Add.
A diamond shape labeled M1 appears o n
top of the
frequency. This is a delta marker.
The four delta markers, M1, M2, M3, and
M4, meas
referenced to MR.
You can also assign markers to specific
traces a
12. Use your finger or the mouse to slide the
marker over to the next signal.
Alternatively, you can do the same task
with the knob or arrow key on the front
panel. Do this by assigning the control
to the marker by touching the marker
toolbar at the bottom of the screen.
The marker readout at the upper right
shows the frequency and amplitude
differences between the reference
marker MR and the M1 delta marker.
The readout to the upper left shows the
absolute value of the M1 marker.
easuring Channel Strength and Adjacent Channel Power
Application 2
: Measuring Channel Strength and Adjacent
Channel Power
The analyzer can take channel power, field strength, adjacent channel power, and multi-carrier channel power measurements.
This application demonstrates the settings used for taking channel power and field strength.
1. Push the front-panel Preset button to set
the instrument to the default settings.
Set up the appropriate measurement
paramete
2. Ensure th
rs for the sample signal.
at Freq is set to 1.5 GHz.
3. Click the Displays icon in the application
bar. Doing this will let you open the C han
Power/STR and ACPR display.
Do not select Data and setup because
that would loa
saved along with the recalled file. That
would overwrite the settings you made
in the previo
example.
23. View the results.
The absolut
appears in the middle of the graph. The
upper adjacent power ratio appears to
the right,
ratio appears to the left.
The gray-shaded bands illustrate the
space bet
makes ACPR power measurements
within the defined channels, represented
by the uns
The yellow bar at the r ight side of the
display represents the signal strength or field
strengt
strength.
easuring Channel Strength and Adjacent Channel Power
The following example shows how to use your analyzer, with Option 21 installed, to demodulate a QPSK signal and to
analyze the signal in multiple domains. You will use the instrument to do the following:
Demodulate a QPSK signal to show its constellation diagram.
Measure the EVM (Error Vector Magnitude) and other key indicators using the Signal Quality display.
View the phase of the signal changing over time.
Use markers to see how the results correlate between the Symbol Table display, Constellation display, and the Phase vs
Time display.
NOTE. The following examples are based on the QPSK sample data file. If desired, you can load the QPSK sample data file
(QPSKDemo.tiq) to recreate the steps used in this application. The signal settings in the following examples are based on
the signal in the sample file. If you use a live signal, your settings may differ.
The General Purpose Digital Demodulation displays share the same modulation and advanced parameter controls. These
controls are av
13. Select the EVM vs Time display, and then
click the Settings icon.
14. Select the Modulation tab.
15. S e t the Modulation Type to QPSK.
16. S et the Symbol Rate to 3.84 MHz.
erforming Modulation Analysis
ailable in the Settings control panel for each display.
17. Set the Mea
Raised Cosine.
18. Set the Reference Filter to Raised
Cosine.
19. Set the Filter Parameter to 0.220.
20. Close the control panel.
surement Filter to Root
Analyze the Signal
You can
The Constellation display should look similar
to the illustration. You might need to click the
Autosc
to properly scale the graph display. For a
QPSK signal, the points should be located in
four t
settings for Frequency, Modulation Type,
Symbol Rate, and Filters.
Look
display. The graph shows the EVM value in
percent at each trace point in time. The RMS
valu
period is shown at the bottom of the display
window, along with the peak EVM value and
the
analyze the signal using both qualitative and quantitative methods.
The Time Overview display shows the entire acquisition record, illustrating the length and offset for Spectrum Time and
Analysis Time. The spectrum length is the period of time within the acquisition record for which the spectrum is calculated.
The analysis length is the period of time within the acquisition record where other measurements are made. The analysis
length can be automatically determined by measurement parameters such as symbol rate, or you can manually adjust
the analysis length.
NOTE. The Spectrum Length and Spectrum Offset cannot be set independently unless the Spectrum Time Mode is set
to Independent. You can change the Spectrum Time Mode on the Analysis > S pectrum Time control panel tab. The red
line that represents the Spectrum Time settings in the Time Overview display is only shown when the Spectrum Time
Mode is set to Independent.
1. In the Time Overview display, select the
Analysis Length button.
The analysis length is indicated by the
blue bar above the graph.
2. Increase the analysis length to 500 us.
You can do this two ways: by changing
the value in the number entry box or by
dragging the right edge of the unshaded
area. Click Replay to rerun the analysis
using this new Analysis Length setting.
Changing the Analysis Length setting
changes the amount of data used for
computing the measurements in the
displays. The shading in the display
shows the extent of the analysis period.
The increased analysis length causes
the instrument to automatically increase
the acquisition length setting to collect
enough samples to satisfy the new
analysis settings. By default, the
automatically determined acquisition
length is equal to or slightly greater than
the analysis length.
4. Increase the Analysis Offset setting to
600 μs.
If the anal
that the analysis period extends past
the end of the acquisition record, the
instrume
length to provide the additional data.
For a recalled signal, if you increase the
Analysis
the end of the available data, the instrument
will analyze only the data that exists within
the set a
about the discrepancy, the instrument adds a
text readout to the right of the numeric value
readou
results (you n
you m ake a change in Analysis Offset or
Length when viewing recalled data).
7. Increase the analysis offset again.
Because th
cannot run a new acquisition to capture a
longer data r ecord. When the requested
analysis
of the data record, the actual analysis
length is reduced.
The following example shows how to use your analyzer to measure frequency hops. You will use the instrument to do
the following:
Measure the transition time.
Measure the hop to hop frequency difference.
Measure the frequency overshoot.
View the spectrogram to see more detail in the frequency transitions versus time.
NOTE. The following examples were based on the TimeFrequency.tiq demonstration data file. If desired, you can load this
file to recreate the steps used in this application. The signal settings in the following examples were based on the signal inthe demonstration file. If you use a live signal, your settings may differ.
1. Click the Displays icon.
This opens the Select Displays window.
2. Click Application Presets...
: Performing Time and Frequency Analysis
3. Click Time-Frequency Analysis and
OK from
By using an application preset, you direct
the instrument to automatically do much
of the
These displays allow you to see
both time- and frequency- domain
representat
include a reference marker (MR) and a
delta marker (M1) to help measure the
hops.
erforming Time and Frequency Analysis
and Spectrum.
ions of hopping signals. They
The Freque
the deviation from the center frequency
value on the vertical axis and time on the
horizont
The Spectrum display shows log power
on the v ertical axis and frequency on the
horizontal axis.
The Spectrogram display shows time on
the vertical axis and frequency on the
horizontal axis. The color at each point
represents the amplitude at a particular
frequency at a particular time.
The Time Overview display shows log
power on the vertical axis and time on
the horizonta
9. MovethemousetotheSpectrogram
display.
10. Right click the mouse and select Zoom
from the resulting menu. Pull the mouse
vertical
on one or two hops of the spectrogram
signal.
One way t
graph is to think of it as a stack of
spectrum traces turned on edge.
l axis.
ly and horizontally to zoom in
o understand the spectrogram
Application 4: P
erforming Time and Frequency Analysis
11. Use the
in the Spectrogram display to a point of
interest. As you move the marker up
and do
changes in the marker in the Spectrum
display. The Time-Frequency Analysis
appli
Spectrum display to show the selected
spectrogram line.
As yo
observe that the power remains constant
over time in the Time Overview display
even
change over time in the Frequency vsTime display.
The
between the Spectrum and Spectrogram
displays. The marker time is correlated
acr
Time, and Time Overview displays.
12. Show the full screen view of the
Frequency vs Time display. This will help
youmorecaref
erforming Time and Frequency Analysis
ully analyze the signal.
13. Click the rig
Zoom from the mouse menu. Click and
hold the left mouse button and move the
mouse to pul
out horizontally and vertically until you
have isolated one or two hops on the
screen.
Zooming in will help you see a more
detailed view of the signal and thus more
accuratel
that you can see the signal and the
overshoot better, you c an also see that
the signal
will impair your ability to measure the
overshoot. So the next step is to clear
up the noi
do that is to minimize the span setting
as far as you can.
14. Click th
15. Click Span in the resulting Frequency vs
Time S ettings pane.
The span is the measurement bandwidth
control for all the measurements in the
General Signal Viewing folder, including
Frequency vs. Time. Changing it in any
of these displays will also change it in
the other displays.
ht mouse button and select
l the displayed waveform
y measure the overshoot. Now
contains a lot of noise, which
se in the signal. One way to
e Settings icon in the menu bar.
16. Click the down arrow and see the setting
change to 20 MHz.
Reducing the span decreases the
measurement bandwidth. Reducing
the measurement bandwidth reduces
the amount of noise present on the
frequency vs. time waveform, allowing
for better resolution of the frequency
transitions.
Continue clicking Span, pushing the
down arrow and
up the signal more and more until the
waveform breaks down.
Change the span settings to 10, 5, and
2MHz.At2MH
breaks down and looks w rong, as shown
at the right. It no longer includes the hop
that you wan
When you set span too s mall, you
reduced the measurement bandwidth too
far. You in
result because you not only eliminated
unwanted noise but also eliminated
much of th
measure.
clicking Replay to clean
z, the waveform clearly
t to measure.
validated the measurement
e signal that you wanted to
Application 4: P
erforming Time and Frequency Analysis
18. Push the up arrow key twice to get the
span setting back to 10 MHz.
19. Click Replay to restore the good
m. You can see your desired
wavefor
signal once again, and it appears much
cleaner than it did at the original 40 MHz
g.
settin
Notice in the screen shot to the right that
now that you have cleaned up the signal,
you can
NOTE. To optimize the measurement
even further, you can go back to step 9
and use the right-button, mouse-controlled
Span Zoom and CF Pan features of the
Spectrogram display instead of the Zoom
and Pan features. Then use Replay and
Autoscale. Such an approach might yield a
further reduction of the span setting and thus
an even cleaner signal on which to make
your measurement.
The M1 marker is at the peak of the
overshoot, the MR marker is at the
middle of the h
overshoot is 370.240 kHz. The overshoot
occurs 151.600 μs before reference
marker MR.
23. Move the markers to measure transition
time. If yo
location of the markers, try using the
general purpose knob.
The trans
the signal is about to make a hop
and ends at about the settled time of
the new fr
measurements for your own application
might use other methods, such as
g when some other signal occurs
startin
or ending when the frequency has settled
to within some tolerance of a specified
With the DPX Spectrum display, your analyzer can identify infrequently occurring transient signals and low-power signals that
may be obscured by stronger signals. After you find that these signals exist, you can use some of the following tools to
capture and examine the signal details to determine their cause:
Use the Max Hold function to verify the presence of signals other than the CW signal.
Use the DPX Spectrum display to view transient signals.
Create a frequency m ask and the use the Frequency Mask tri gger to capture any signal that violates the mask.
Use the Spectrogram with Frequency Mask Trigger to view the mask violations in the Time and Frequency domains.
Detecting Transient Signals Using the DPX Spectrum Display
The D PX Spe
signals so that you can see low-level and higher power signals that occur at the same frequency, but at different times.
1. Push the Preset button on the front
panel to s
default settings.
ctrum display uses a bitmap image in addition to line traces to view signals. Bitmaps can represent multi-value
The more you increase the Persistence
setting, the more quickly you will see
infreque
the more frequent signals appear in red;
infrequent signals will appear in blue.
These set
signals below the maximum signal level. For
example, a low-level signal in the presence
of a puls
Persistence and Intensity setting.
If your instrument has Option 02/52 installed, you can use the Mask Editor to create a frequency mask for triggering on
transient signals. Complete the following steps to get a good visual reference that you can use to build the frequency mask.
1. Push the Preset button on the front
panel to set the instrument to the default
settings.
2. Tune the instrument to the frequency of
your signa
Viewing Transient Signals in Time and Frequency Domains
Spectrograms allow you to see how signals change over time. You can use the Spectrogram display to examine the transient
signals that violated the mask. Combining the Spectrogram display with the Frequency Mask Trigger allows you to see
how often the violations occur and to troubleshoot the cause of the problem.
1. Click Displays to open the Select
Displays dialog box.
2. Add the Spectrogram and Time Overview
displays.
Spectrum traces 1, 2, 3, and 4 show the spectrum for the Spectrum Time selected in the Time Overview display or in the
Spectrum Time tab of the Analysis control panel. The Spectrogram, by c omparison, covers the Analysis Time selected in
the Time Overview display or in the Analysis Time tab of the Analysis control panel.
Pulsed RF measurements have historically been difficult to perform. Some measurem ents required custom-built and
dedicated test tools, plus trained experts to properly use the tools to achieve accuracy and repeatability. Tektronix Real-Time
Spectrum Analyzers have revolutionized pulse measurements through automation. An analyzer, with Option 20 installed, can
replace specialized test equipment formerly required for pulsed RF measurements.
This application shows how to accomplish the following pulsed RF measurement tasks:
Capture a series of RF pulses in a single acquisition record.
Select measurements to display in the Pulse Table.
Examine the pulse shape and measure reference points w ith the Pulse Trace display.
View Trend and FFT analysis on the measurement results with the Pulse Statistics display.
NOTE. To complete the following example, you w ill need a pulsed signal or an appropriate saved data record. This example
uses the PulseDemo.tiq file, which is located in the folder C:\SPECMO N Files\Sample D ata Records.
Capture the Pulses
1. Push the Preset button on the front
panel to set the instrument to the default
settings.
2. Click Displays to open the Select
Displays dialog box.
Analysis Length to include several
pulses. Decre
to about 10 ms so you can see the first
pulse in detail. Adjust the Spectrum
Offset so the
on time of this pulse.
16. Click Replay to run the measurements
over these new analysis and spectrum
time period
17. Select the Pulse Table display and then
select the Settings icon.
ase the horizontal scale
Spectrum Time covers the
s.
Application 6: T
aking Pulse Measurements
18. Select th
19. Select th
interested in. (For this example, select
Average ON Power, Pulse Width, and
Rise Tim
20. Close th
21. When you see the data in the Pulse
Table display, click Replay to recalculate
the Pulse Table measurements.
e Measurements tab.
e measurements that you are
e).
e control panel.
Quick Tip
can take measurements while the instrument is running or while it is stopped. Stopping the instrument may make it
You
easier to read the measurements from c aptured data.
Measure the Parameters of the Captured Pulses
After you have captured the pulses, you can use the Pulse Trace display to view the details of specific measurements.
in the Pulse Table display. For example,
click the cell
of Pulse 1.
The Pulse Trace display shows an
amplitude v
selected result on the selected pulse.
Blue lines and arrows show how the
measuremen
The green arrow in the display shows the
power threshold used to detect pulses. If
this thres
no pulses will be detected. You can set
the power threshold on the Settings >
Params ta
The new pulse appears in the Pulse
Trace display and is selected in the
Pulse Table d
Pulse Trace display and the Pulse Table
display together to view and analyze
pulse measur
You can select a different result in the
Pulse Trace display and it will also be
selected in
3. Use the Scale and Offset controls to
zoom in on details of the selected pulse.
For example, you can adjust the controls
to get a close look at the details of the
Rise Time measurement as shown.
isplay. You can use the
ements.
the Pulse Table display.
Application 6: T
aking Pulse Measurements
Quick Tip
Click Autoscale to optimize the vertical and horizontal offset and scale settings.
When using scale or offset, adjust the offset control to move the area of interest to the far left side of the screen, and then
adjust the scale to expand the area of interest. Another way to change scaling is to right-click in the graph and select
Pan or Zoom, then use the mouse or the touchscreen to drag in the graph.
Review Measurement Statistics Across All Measured Pulses
You can use the Pulse Statistics display to show the trend or an FFT across all measured pulses. To get the best frequency
resolution and dynamic range in the display, you need to include many pulses in the analysis period.
When Trend is th
Pulse Statistics display plots the results
of the selected measurement for every
measured puls
7. Select the Φ Di
Pulse-to-pulse phase measurements are
good examples to show the trend and
FFT statisti
8. Change the St
FFT shows a spectrum-like trace of the
amplitude (
result in the set) versus frequency. This
can be useful for identifying interference
in the puls
spike appears around 60 Hz, it might
indicate coupling from the AC power
supply.