Tektronix SG5030 Specifications Sheet

The SG 5030 is the only choice for anyone who needs leveled output amplitude to calibrate analog or digital scopes with bandwidth to 550 MHz — under either local or programmable control. That’ s because no other programmable leveled sine wave generator is built specifically to fill scope calibration requirements. The SG 5030’s leveled output is variable from 0.1 Hz to 550 MHz with a reference frequency of 50 kHz. Crystal-controlled frequency accuracy eliminates drift so there’ s no second-guessing results.
Accurately calibrated output voltage is
provided from 4.5 mV to 5.5 V peak-to­peak into 50 . Absolute amplitude accuracy is ±1.5% from 0.1 Hz to 50 kHz, with flatness from ±1.5 to ±4% over the remainder of the frequency range to 550 MHz.
Aremote leveling head is standard and plugs directly into the oscilloscope to ensure that the output signal is level at all times. All other signal generators are spec’d at the front-panel BNC connector, not at the end of the cable going to the instrument under test.
Easy test setup — either locally or over the GPIB
The SG 5030’s front panel is simple to operate, providing complete status and error information through a bright, 8-digit display. To simplify manual use and reduce bus traffic in ATE applications, you can store up to 20 front-panel settings in non­volatile memory.
Frequency and amplitude along with front-panel store/recall settings are all fully programmable. Automating test procedures makes scope calibration some four times faster than manual test methods allow — and virtually eliminates the risk of operator error .
Configure the SG 5030 with a CG 5011 Programmable Calibration Generator
As three-wide TM 5000 Series modules, they conveniently fill a six-slot TM 5006A mainframe to form a complete, cost-effective benchtop or rackmount calibration system.
Finally, count on TEGAM to offer more support of your purchase, including comprehensive manuals, applications assistance, updates and a worldwide service network.
SG 5030 Pro grammab le Leveled Sine Wa ve Generator for Scope Calibration
• 0.1 Hz to 550 MHz frequency range with a reference frequency of 50 kHz
• 4.5 mV to 5.5 V amplitude range
• Amplitude flatness from ±1.5 to ±4% of 50 kHz reference
• Synthesized frequency accurac y
• Excellent VSWR
• 20 stored front-panel settings capability with non-volatile memory
• Self-test on power-up
• Easy integration into GPIB systems
• Companion to CG 5011 Programmable Calibration Generator
MODULAR TEST INSTRUMENTS
TM 500 / 5000
THE GLOBAL SOURCE FOR PROVENTEST
AND MEASUREMENT TECHNOLOGY.
Prices and specifications subject to change without notice.
SG 5030 Specifications
Frequency
Range/Resolution
0.1 Hz - 4.9999 kHz 0.1 Hz step 5 kHz - 49.999 kHz 1 Hz step 50 kHz - 550 MHz 10 Hz steps
Accuracy With Internal Timebase (within 1 year of adjustment)
ppm of Setting + Hz
0.1 Hz - 4.9999 kHz ±(3 + 0.06) 5 kHz - 49.999 kHz ±(3 + 0.3) 50 kHz - 550 MHz ±(3 + 3)
Accuracy With External Timebase (10 MHz)
0.1 Hz - 4.9999 kHz ±(ext. t.b. error + 0.06 Hz)
5 kHz - 49.999 kHz ±(ext. t.b. error
+ 0.3 Hz)
50 kHz - 550 MHz ±(ext. t.b. error
+ 3 Hz)
Amplitude
Range/Resolution
4.5 mV - 55 mV 0.02 mV/step
55.2 mV - 550 mV 0.2 mV/step
552 mV - 5.5 V 2 mV/step
-42.95 dBm -
+18.75 dBm 0.05 dBm/step
Accuracy (within 1 year of adjustment)
0.1 Hz - 50 kHz ±1.5% of setting
Flatness (within 1 year of adjustment)
% of 50 kHz Ref. Freq. 50 kHz - 100 MHz ±1.5% 100 MHz - 250 MHz ±3% 250 MHz - 550 MHz ±4%
Output
Leveling Settling Time Less than 20 ms Source Resistance 50 , ±1% DC Offset ±20 mV VSWR Less than 1.2:1
up to 550 MHz
Harmonic Distortion
0.1 Hz - All Harmonics
49.999 kHz and Spurs less
than -50 dBc
50 kHz- 2nd Harmonic 550 MHz less than -30 dBc
All others less than -35 dBc Nonharmonics less than -40 dBc Phase noise less than -85 dBc/Hz at 10 kHz offset
Internal Timebase Output
Frequency 10 MHz Accuracy ±3 ppm Amplitude 400 mV p-p
into 50
Impedance 50
External Timebase Input
Frequency -10 MHz,
±1.5 ppm
Amplitude l0dBm to
+10dBm (70 to 700 mV rms)
Impedance 50 Ωac,
500 dc
Environmental
Temperature
Operating 0°C to 40°C Non-operating -40 ° C t o + 71 ° C
Humidity
Operating 95% ±5% RH,
+ 10°C to +30°C 75% ±5% RH, +30°C to +40°C RH not controlled below + 10°C
General
W eight 6.5 lb. (2.4 kg) Height 5" (12.7 cm) Width 8" (20.32 cm) Length 11" (27.94 cm)
MODULAR TEST INSTRUMENTS
TM 500 / 5000
Ordering Information
SG 5030 Prgm. Leveled Sine Wave
Generator Includes: Operators Manual (070-7705-01), Instrument Interfacing Guide (070-7704-00), Reference Guide (070-7706-00), Leveling Head (015-2350-01).
TM 5003 3 Wide Power Module
Mainframe, GPIB TM 5003/RI TM 5003 w/Rear Interface
TM 5006A 6 Wide Power Module
Mainframe, GPIB TM 5006A/R TM 5006A w/Rack Mount TM 5006A/RI TM 5006Aw/Rear Interface TM 5006A/R/RI TM 5006A w/Rack Mt &
Rear Interface TM 5006A/EMC TM 5006A w/EMC Shielding
Mainframe Power Plug Options
Standard 120V North American UE220 220V Universal Euro
& Switzerland UK240 240V United Kingdom A240 240VAustralian NA240 240V North American S220 220VSwitzerland
Warranty
One year on materials and workmanship.
Calibration Documentation
Contact TEGAM for OPTION Z540 NIST Traceable Compliance Certificate and Test Data.
Calibration & T echnical Services
For warranty and remedial repair, calibration services and spare parts, or for additional information on TEGAM sales and service offices around the world, contact us at 440-466-6100 (ph) or 440-466-6110 (fx).
SG5030.991PG
THE GLOBAL SOURCE FOR PROVENTEST
AND MEASUREMENT TECHNOLOGY.
TEN TEGAM WAY • GENEVA,OHIO 44041
440-466-6100 • FAX 440-466-6110
www.tegam.com • e-mail:sales@tegam.com