Tektronix Series 2600B System SourceMeter Instrument User's Manual User manual

Series 2600B
System SourceMeter® Instrument
User's Manual
tek.com/keithley
2600B
S-900-01 Rev. A August 2021
*P2600BS-900-01A*
2600B
S-900-01A
System SourceMeter® instrument
Series 2600B
User's Manual
© 2021, Keithley Instruments, LLC
Cleveland, Ohio, U.S.A.
All rights reserved.
Any unauthorized reproduction, photocopy, or use of the information herein, in whole or in part,
without the prior written approval of Keithley Instruments, LLC, is strictly prohibited.
These are the original instructions in English.
TSPTM and TSP-LinkTM are trademarks of Keithley Instruments, LLC. All Keithley Instruments
product names are trademarks or registered trademarks of Keithley Instruments, LLC. Other brand
names are trademarks or registered trademarks of their respective holders.
The Lua 5.0 software and associated documentation files are copyright © 1994 - 2015, Lua.org,
PUC-Rio. You can access terms of license for the Lua software and associated documentation at
the Lua licensing site (https://www.lua.org/license.html).
Microsoft, Visual C++, Excel, and Windows are either registered trademarks or trademarks of
Microsoft Corporation in the United States and/or other countries.
Document number: 2600BS-900-01 Rev. A August 2021

Safety precautions

The following safety precautions should be observed before using this product and any associated instrumentation. Although some instruments and accessories would normally be used with nonhazardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired. The types of product users are: Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring that the
equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained. Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the
instrument. They must be protected from electric shock and contact with hazardous live circuits. Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line
voltage or replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly trained service personnel may perform installation and service procedures.
Keithley products are designed for use with electrical signals that are measurement, control, and data I/O connections, with low transient overvoltages, and must not be directly connected to mains voltage or to voltage sources with high transient overvoltages. Measurement Category II (as referenced in IEC 60664) connections require protection for high transient overvoltages often associated with local AC mains connections. Certain Keithley measuring instruments may be connected to mains. These instruments will be marked as category II or higher.
Unless explicitly allowed in the specifications, operating manual, and instrument labels, do not connect any instrument to mains. Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test
fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30 V RMS, 42.4 V peak, or 60 VDC are present. A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 V, no conductive part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance-limited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, ensure that the line cord is connected to a properly-grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power disconnect device must be provided in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
For safety, instruments and accessories must be used in accordance with the operating instructions. If the instruments or accessories are used in a manner not specified in the operating instructions, the protection provided by the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories. Maximum signal levels are defined in the specifications and operating information and shown on the instrument panels, test fixture panels, and switching cards.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard. Chassis connections must only be used as shield connections for measuring circuits, NOT as protective earth (safety ground)
connections. If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use
of a lid interlock.
If a screw is present, connect it to protective earth (safety ground) using the wire recommended in the user documentation.
The symbol on an instrument means caution, risk of hazard. The user must refer to the operating instructions located in the user documentation in all cases where the symbol is marked on the instrument.
The symbol on an instrument means warning, risk of electric shock. Use standard safety precautions to avoid personal contact with these voltages.
The symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
The symbol indicates a connection terminal to the equipment frame.
If this symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp must be properly disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains hazards that might result in personal injury or death. Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the instrument. Such damage may invalidate the warranty.
The CAUTION heading with the symbol in the user documentation explains hazards that could result in moderate or minor injury or damage the instrument. Always read the associated information very carefully before performing the indicated procedure. Damage to the instrument may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans. Before performing any maintenance, disconnect the line cord and all test cables. To maintain protection from electric shock and fire, replacement components in mains circuits including the power
transformer, test leads, and input jacks must be purchased from Keithley. Standard fuses with applicable national safety approvals may be used if the rating and type are the same. The detachable mains power cord provided with the instrument may only be replaced with a similarly rated power cord. Other components that are not safety-related may be purchased from other suppliers as long as they are equivalent to the original component (note that selected parts should be purchased only through Keithley to maintain accuracy and functionality of the product). If you are unsure about the applicability of a replacement component, call a Keithley office for information.
Unless otherwise noted in product-specific literature, Keithley instruments are designed to operate indoors only, in the following environment: Altitude at or below 2,000 m (6,562 ft); temperature 0 °C to 50 °C (32 °F to 122 °F); and pollution degree 1 or 2.
To clean an instrument, use a cloth dampened with deionized water or mild, water-based cleaner. Clean the exterior of the instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., a data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper cleaning/servicing.
Safety precaution revision as of June 2018.
Introduction .............................................................................................................. 1-1
Welcome .............................................................................................................................. 1-1
Extended warranty ............................................................................................................... 1-1
Contact information .............................................................................................................. 1-1
Customer documentation ..................................................................................................... 1-2
Product software and drivers ............................................................................................... 1-3
General ratings ..................................................................................................................... 1-4
Installation ................................................................................................................ 2-1
Introduction .......................................................................................................................... 2-1
Cooling vents ....................................................................................................................... 2-1
Turning the instrument on and off ........................................................................................ 2-2
Placing a 2600B in standby ................................................................................................. 2-3
Warmup period ..................................................................................................................... 2-3
Line frequency configuration ................................................................................................ 2-4
System information .............................................................................................................. 2-4
Instrument description ............................................................................................ 3-1
Controls, indicators, and connectors .................................................................................... 3-1
Front panel ................................................................................................................................ 3-1
Rear panel................................................................................................................................. 3-5
Menu overview ................................................................................................................... 3-12
Menu navigation ...................................................................................................................... 3-12
Menu trees .............................................................................................................................. 3-12
Setting values .......................................................................................................................... 3-19
Beeper ................................................................................................................................ 3-21
Displayed error and status messages ................................................................................ 3-21
Display operations .............................................................................................................. 3-22
Display mode .......................................................................................................................... 3-22
Display functions and attributes .............................................................................................. 3-23
Display features ...................................................................................................................... 3-24
Display messages ................................................................................................................... 3-25
Input prompting ....................................................................................................................... 3-29
Indicators................................................................................................................................. 3-31
Local lockout ........................................................................................................................... 3-32
Load test menu ....................................................................................................................... 3-32
Running a test from the front panel ......................................................................................... 3-34
Connecting the USB flash drive ......................................................................................... 3-34
Using the web interface...................................................................................................... 3-35
How to access the web interface ............................................................................................. 3-35
Web interface Welcome page ................................................................................................. 3-36
Use the ID button to identify the instrument ............................................................................ 3-37

Table of contents

Table of contents Series 2600B System SourceMeter® instrument User's Manual
Change the IP configuration through the web interface .......................................................... 3-37
Set the instrument password ................................................................................................... 3-38
Using the virtual front panel..................................................................................................... 3-38
View buffer data using the web interface ................................................................................ 3-39
Download reading buffer data using the web interface ........................................................... 3-40
Using TSB Embedded ............................................................................................................. 3-41
Send individual commands using the web interface ................................................................ 3-41
Review events in the LXI Event Log ........................................................................................ 3-41
Help ......................................................................................................................................... 3-44
Operation .................................................................................................................. 4-1
Basic operation .................................................................................................................... 4-1
Source-measure capabilities ................................................................................................ 4-1
Voltage and current ................................................................................................................... 4-2
Limits .................................................................................................................................... 4-2
Setting the limit from the front panel .......................................................................................... 4-4
Setting the limit from a remote interface .................................................................................... 4-4
Sink operation ...................................................................................................................... 4-5
Setting the sink mode using the front panel .............................................................................. 4-6
Setting the sink mode from a remote interface .......................................................................... 4-6
DUT test connections ........................................................................................................... 4-6
Input/output connectors ............................................................................................................. 4-8
2-wire local sensing connections ............................................................................................. 4-12
4-wire remote sensing connections ......................................................................................... 4-12
Contact check connections ..................................................................................................... 4-14
Guarding and shielding ........................................................................................................... 4-16
Interlock................................................................................................................................... 4-23
Test fixture .............................................................................................................................. 4-24
Floating a SMU ....................................................................................................................... 4-26
Front-panel source-measure procedure ............................................................................ 4-28
Step 1: Select and set the source level ................................................................................... 4-28
Step 2: Set the compliance limit .............................................................................................. 4-28
Step 3: Select the measurement function and range .............................................................. 4-29
Step 4: Turn the output on ...................................................................................................... 4-29
Step 5: Observe readings on the display. ............................................................................... 4-29
Step 6: Turn the output off....................................................................................................... 4-29
Sense mode selection ........................................................................................................ 4-30
Front-panel sense mode selection .......................................................................................... 4-30
Remote interface sense mode selection ................................................................................. 4-30
Output-off modes ............................................................................................................... 4-31
Normal output-off mode .......................................................................................................... 4-31
High-impedance output-off mode ............................................................................................ 4-31
Zero output-off mode ............................................................................................................... 4-31
Output-off function ................................................................................................................... 4-32
Output-off limits (compliance) .................................................................................................. 4-33
Remote programming output-off states quick reference ......................................................... 4-34
Maintenance ................................ ................................................................ ............. 5-1
Introduction .......................................................................................................................... 5-1
Line fuse replacement .......................................................................................................... 5-1
Series 2600B System SourceMeter® instrument User's Manual Table of contents
Front-panel tests .................................................................................................................. 5-3
Keys test ................................................................................................................................... 5-3
Display patterns test .................................................................................................................. 5-3
Upgrading the firmware ........................................................................................................ 5-4
Using TSB to upgrade the firmware .......................................................................................... 5-5
Displaying the serial number ................................................................................................ 5-6
Next steps ................................................................................................................. 6-1
Additional 2600B information ............................................................................................... 6-1
In this section:
Welcome .................................................................................. 1-1
Extended warranty ................................................................... 1-1
Contact information ................................ .................................. 1-1
Customer documentation ......................................................... 1-2
Product software and drivers.................................................... 1-3
General ratings ......................................................................... 1-4

Welcome

Thank you for choosing a Keithley Instruments product. The Series 2600B System SourceMeter® instrument provides manufacturers of electronic components and semiconductor devices with an instrument that combines source and measurement capabilities in a single instrument called a source-measure unit (also called a SMU). This combination simplifies test processes by eliminating synchronization and connection issues associated with multiple instrument solutions. A 2600B provides a scalable, high throughput, highly cost-effective solution for precision dc, pulse, and low frequency ac source-measure testing that also maintains code compatibility throughout the Series 2600B SourceMeter instruments.
This manual describes basic operation for the Models 2601B, 2602B, 2604B, 2611B, 2612B, 2614B, 2634B, 2635B, and 2636B.

Extended warranty

Additional years of warranty coverage are available on many products. These valuable contracts protect you from unbudgeted service expenses and provide additional years of protection at a fraction of the price of a repair. Extended warranties are available on new and existing products. Contact your local Keithley Instruments office, sales partner, or distributor for details.

Contact information

If you have any questions after you review the information in this documentation, please contact your local Keithley Instruments office, sales partner, or distributor. You can also call the Tektronix corporate headquarters (toll-free inside the U.S. and Canada only) at 1-800-833-9200. For worldwide contact numbers, visit tek.com/contact-us.
Section 1

Introduction

Section 1: Introduction Series 2600B System SourceMeter® instrument User's Manual
1-2 2600BS-900-01 Rev. A / August 2021

Customer documentation

The documentation for the 2600B includes a Quick Start Guide, User's Manual, and Reference Manual. A Quick Start Guide is provided as a hard copy with the instrument. You can also access it from tek.com/keithley as an Adobe Acrobat PDF file.
Quick Start Guide: Provides unpacking instructions, describes basic connections, and reviews
basic operation information. If you are new to Keithley Instruments equipment, refer to the Quick Start Guide to take the steps needed to unpack, set up, and verify operation.
User's Manual: Includes installation, instrument description, operation, and maintenance
information.
Reference Manual: Includes advanced operation topics and maintenance information.
Programmers looking for a command reference and users looking for an in-depth description of how the instrument works (including troubleshooting and optimization) should refer to the Reference Manual.
Series 2600B System SourceMeter® instrument User's Manual Section 1: Introduction
2600BS-900-01 Rev. A / August 2021 1-3

Product software and drivers

Go to the Product Support and Downloads web page (tek.com/product-support) to download drivers and software for your instrument.
Available drivers and software include:
KickStart Software: Enables quick test setup and data visualization when using one or
more instruments.
Test Script Builder (TSB): This software provides an environment to develop a test program and
the ability to load the test program onto the instrument. Running a program loaded on the instrument eliminates the need to send individual commands from the host computer to the instrument when running a test.
IVI-COM Driver: An IVI instrument driver you can use to create your own test applications in
C/C++, VB.NET, or C# programming languages. It can also be called from other languages that support calling a DLL or ActiveX (COM) object. Refer to IVI Foundation (ivifoundation.org) for additional information.
LabVIEW
Software drivers: Drivers to communicate with National Instruments
LabVIEW Software.
Keithley I/O layer: Manages the communications between Keithley instrument drivers and
software applications and the instrument itself. The I/O Layer handles differences in communications required to support GPIB, serial, ethernet, and other communications buses so that drivers and software applications do not need to handle the differences themselves.
To identify IP addresses of instruments that are connected to the local area network (LAN) and support the VXI-11 discovery protocol, you can also use LXI Discovery Tool, available from the
Resources (lxistandard.org/Resources/Resources.aspx) page of the LXI Consortium website
(lxistandard.org).
Section 1: Introduction Series 2600B System SourceMeter® instrument User's Manual
1-4 2600BS-900-01 Rev. A / August 2021

General ratings

Category
Specification
Supply voltage range
100 V ac to 240 V ac, 50 Hz or 60 Hz (autosensing), 240 VA maximum
Input and output connections
See Front panel (on page 3-1) and Rear panel (on page 3-5)
Environmental conditions
For indoor use only. Altitude: Maximum 2,000 meters (6,562 feet) above sea level Operating: 0 °C to 50 °C, 70% relative humidity up to 35 °C. Derate 3%
relative humidity per °C, 35 °C to 50 °C
Storage: −25 °C to 65 °C Pollution degree: 1 or 2
In this section:
Introduction .............................................................................. 2-1
Cooling vents ........................................................................... 2-1
Turning the instrument on and off ............................................ 2-2
Placing a 2600B in standby ...................................................... 2-3
Warmup period ......................................................................... 2-3
Line frequency configuration .................................................... 2-4
System information .................................................................. 2-4

Introduction

This section provides the information you need to install the 2600B, make communications connections, and power up the instrument.

Cooling vents

The 2600B has side and top intake and rear exhaust vents. One side must be unobstructed to dissipate heat.
Excessive heat could damage the 2600B and degrade its performance. Only operate the 2600B in an environment where the ambient temperature does not exceed 50 °C.
Do not place a container of liquid (water or coffee, for instance) on the top cover. If it spills, the liquid may enter the case through the vents and cause severe damage.
To prevent damaging heat build-up and ensure specified performance, use the following guidelines.
The rear exhaust vent and either the top or both side intake vents must be unobstructed to properly dissipate heat. Even partial blockage could impair proper cooling.
Do not position any devices adjacent to the 2600B that force air (heated or unheated) toward its cooling vents or surfaces. This additional airflow could compromise accuracy.
When rack mounting the 2600B, make sure there is adequate airflow around both sides to ensure proper cooling. Adequate airflow enables air temperatures within approximately one inch of the 2600B surfaces to remain within specified limits under all operating conditions.
Section 2

Installation

Section 2: Installation Series 2600B System SourceMeter® instrument User's Manual
2-2 2600BS-900-01 Rev. A / August 2021
If high power dissipation equipment is rack mounted next to the 2600B, it could cause excessive heating. To produce specified 2600B accuracies, maintain the specified ambient temperature around the surfaces of the 2600B. In rack configurations with convection cooling only, proper cooling practice places the hottest non-precision equipment (for example, the power supply) at the top of the rack away from and above precision equipment (such as the 2600B).
Mount precision equipment as low as possible in the rack, where temperatures are coolest. You can add space panels above and below the 2600B to help provide adequate airflow.

Turning the instrument on and off

The 2600B operates from a line voltage of 100 V to 240 V at a frequency of 50 Hz or 60 Hz. Line voltage is automatically sensed (there are no switches to set). Make sure the operating voltage in your area is compatible.
Follow the procedure below to connect the 2600B to line power and turn on the instrument.
Operating the instrument on an incorrect line voltage may cause damage to the instrument, possibly voiding the warranty.
To turn a 2600B on and off:
1. Before plugging in the power cord, make sure that the front-panel POWER switch is in the off (O) position.
2. Connect the female end of the supplied power cord to the ac receptacle on the rear panel.
3. Connect the other end of the power cord to a grounded ac outlet.
The power cord supplied with the 2600B contains a separate protective earth (safety ground) wire for use with grounded outlets. When proper connections are made, the instrument chassis is connected to power-line ground through the ground wire in the power cord. In addition, a chassis ground connection is provided through a screw on the rear panel. This terminal should be connected to a known protective earth. In the event of a failure, not using a properly grounded protective earth and grounded outlet may result in personal injury or death due to electric shock.
Do not replace detachable mains supply cords with inadequately rated cords. Failure to use properly rated cords may result in personal injury or death due to electric shock.
Series 2600B System SourceMeter® instrument User's Manual Section 2: Installation
2600BS-900-01 Rev. A / August 2021 2-3
Hazardous voltages may be present in the test system. To prevent injury or death, remove power from the instrument or test system and discharge any energy storage components (for example, capacitors or cables) before changing any connections that might allow contact with an uninsulated conductor.
On some sensitive or easily damaged devices under test (DUTs), the instrument power-up and power-down sequence can apply transient signals to the DUT that may affect or damage it. When testing this type of DUT, do not make final connections to it until the instrument has completed its power-up sequence and is in a known operating state. When testing this type of DUT, disconnect it from the instrument before turning the instrument off.
To prevent any human contact with a live conductor, connections to the DUT must be fully insulated and the final connections to the DUT must only use safety-rated safety-jack-socket connectors that do not allow bodily contact.
4. To turn your instrument on, press the front-panel POWER switch to place it in the on (I) position.
5. To turn your instrument off, press the front-panel POWER switch to place it in the off (O) position.

Placing a 2600B in standby

Hazardous voltages may be present on all output and guard terminals. To prevent electrical shock that could cause injury or death, never make or break connections to the 2600B while the instrument is powered on. Turn off the equipment from the front panel or disconnect the main power cord from the rear of the 2600B before handling cables. Putting the equipment into standby does not guarantee that the outputs are powered off if a hardware or software fault occurs.
When the instrument is on, the output may be placed in an active output state (output on) or a standby mode (output off). From the front panel, pressing the OUTPUT ON/OFF control toggles the output using the present instrument configuration. You can also place the output in standby over the remote interface by sending the following command:
smuX.source.output = smuX.OUTPUT_OFF
Even though the instrument is placed in standby, the output may not actually be off.

Warmup period

The 2600B must be turned on and allowed to warm up for at least two hours to achieve rated accuracies.
Section 2: Installation Series 2600B System SourceMeter® instrument User's Manual
2-4 2600BS-900-01 Rev. A / August 2021

Line frequency configuration

The factory configures the 2600B to detect the power line frequency automatically at each power-up. This detected line frequency (either 50 Hz or 60 Hz) is used for aperture (NPLC) calculations.
In noisy environments, you can manually configure the instrument to match the actual line frequency.
To configure the line frequency from the front panel:
1. Press the MENU key, then turn the navigation wheel to select LINE-FREQ, and then press the ENTER key.
2. Turn the navigation wheel to select the appropriate frequency and then press the ENTER key. To configure the instrument to automatically detect line frequency at each power-up, select AUTO.
3. Press the EXIT (LOCAL) key to return to the main display.
To configure the line frequency from a remote interface:
Set the localnode.linefreq or the localnode.autolinefreq attribute. To set the line frequency to 60 Hz, send:
localnode.linefreq = 60
To configure the instrument to automatically detect line frequency at each power-up:
localnode.autolinefreq = true

System information

You can retrieve serial number, firmware revision, calibration dates, and memory usage from the instrument.
To view the system information from the front panel:
1. Press the MENU key.
2. Select SYSTEM-INFO.
3. Select one of the following:
FIRMWARE SERIAL# CAL MEMORY-USAGE
To retrieve system information from a remote interface:
To retrieve the firmware revision and serial number, send the *IDN? query. For more information on system information, see the Series 2600B Reference Manual. In particular:
To determine memory usage, see the meminfo() function.
To determine when calibration was last run, see smuX.cal.date.
To determine when calibration is due, see smuX.cal.due.
In this section:
Controls, indicators, and connectors ........................................ 3-1
Menu overview ....................................................................... 3-12
Beeper ................................................................................... 3-21
Displayed error and status messages .................................... 3-21
Display operations .................................................................. 3-22
Connecting the USB flash drive ............................................. 3-34
Using the web interface .......................................................... 3-35

Controls, indicators, and connectors

2600B controls, indicators, and the USB port are on the front panel (on page 3-1). Make connections to the 2600B through connectors on the rear panel (on page 3-5).

Front panel

The front panel of the 2600B is shown below. The descriptions of the front-panel controls, USB port, and indicators follow the figure.
Figure 1: Front panel of single-channel SMU (2601B, 2611B, and 2635B)
Section 3

Instrument description

Section 3: Instrument description Series 2600B System SourceMeter® instrument User's Manual
3-2 2600BS-900-01 Rev. A / August 2021
Figure 2: Front panel of two-channel SMU (2602B, 2604B, 2612B, 2614B, 2634B, and 2636B)
1. Power switch, display and configuration keys
Toggles between the source-measure displays and the user message mode. Configures a function or operation.
Power switch. The in position turns the 2600B on (I); the out position turns it off (O).
2. SMU setup, performance control, special operation, and numbers
SMU (source-measure unit) setup SRC
Selects the source function (voltage or current) and places the cursor in the source field for editing.
MEAS
Cycles through measure functions (voltage, current, resistance, or power).
LIMIT
Places the cursor in the compliance limit field for editing. Also selects the limit value to edit (voltage, current, or power).
MODE
Selects a meter mode (I-METER, V-METER, OHM-METER, or WATT-METER).
Performance control DIGITS
Sets the display resolution (4½, 5½, or 6½ digits).
SPEED
Sets the measurement speed (FAST, MEDium, NORMAL, HI-ACCURACY, or OTHER). Speed and accuracy are set by controlling the measurement aperture.
REL
Controls relative measurements, which allows a baseline value to be subtracted from a reading.
FILTER
Enables or disables the digital filter. You can use this filter to reduce reading noise.
Series 2600B System SourceMeter® instrument User's Manual Section 3: Instrument description
2600BS-900-01 Rev. A / August 2021 3-3
Special operation
LOAD
Loads a test for execution (FACTORY, USER, or SCRIPTS).
RUN
Runs the last selected factory or user-defined test.
STORE
Accesses reading buffers and makes readings:
TAKE_READINGS: Use to make readings and store them in a reading buffer. SAVE: Use to save a reading buffer to nonvolatile memory or to a user-installed
flash drive (USB1) in CSV or XML format.
Readings can include measurements, source values, and timestamp values.
RECALL
Recalls information (DATA or STATISTICS) stored in a reading buffer:
DATA: Includes stored readings, and if configured, source values and timestamp
values.
STATISTICS: Includes MEAN, STD DEV, SAMPLE SIZE, MINIMUM,
MAXIMUM, PK-PK.
TRIG
Triggers readings.
MENU
Accesses the main menu (on page 3-12). You can use the main menu to configure many functions and features.
EXIT
Cancels the selection and returns to the previous menu or display. Also used as a LOCAL key to take the instrument out of remote operation.
ENTER
Accepts the selection and moves to the next choice or exits the menu.
Numbers
Number keys
When enabled and in EDIT mode, the number keys (0 to 9, +/-, 0000) allow direct numeric entry. Press the navigation wheel to enter EDIT mode. For more information, see Setting a value (on page 3-19).
3. Range keys
Selects the next higher source or measure range.
Enables or disables source or measure autorange.
Selects the next lower source or measure range.
In addition to selecting range functions, the up and down range keys change the format for numbers that are not ranges, such as the limit value.
Section 3: Instrument description Series 2600B System SourceMeter® instrument User's Manual
3-4 2600BS-900-01 Rev. A / August 2021
4. Cursor keys
Use the CURSOR keys to move the cursor left or right. When the cursor is on the source or compliance value digit, press the navigation wheel to enter edit mode, and turn the navigation wheel to edit the value. Press the navigation wheel again when you finish editing. Use the CURSOR keys or the navigation wheel to move through menu items. To view a menu value, use the CURSOR keys for cursor control, and then press the navigation wheel to view the value or submenu item.
5. Navigation wheel
Turn the navigation wheel to:
Move the cursor to the left and the right (the cursor indicates the selected value
or item)
While in edit mode, increase or decrease a selected source or compliance value
Push the navigation wheel to:
Enable or disable edit mode for the selected source or compliance value Open menus and submenu items Select a menu option or a value
6. Output control
Turns the source output on or off.
7. USB port
Use the USB port to connect a USB flash drive to the instrument. You can use the USB flash drive to store reading buffer data, scripts, and user setups. You can also use it to upgrade the firmware.
Series 2600B System SourceMeter® instrument User's Manual Section 3: Instrument description
2600BS-900-01 Rev. A / August 2021 3-5
8. Display indicators (not shown)
The items listed below represent the possible display indicators and their meanings.
Indicator
Meaning
4W
Remote (4-wire) sense is selected
AUTO
Source or measure autorange is selected
EDIT
Instrument is in editing mode
ERR
Questionable reading or invalid calibration step
FILT
Digital filter is enabled
LSTN
Instrument is addressed to listen
REL
Relative mode is enabled
REM
Instrument is in remote mode
SRQ
Service request is asserted
TALK
Instrument is addressed to talk
* (asterisk)
Readings are being stored in the buffer

Rear panel

The rear panels of the 2600B is shown below. The descriptions of the rear-panel components follow the figures.
Figure 3: 2601B and 2611B rear panel
Section 3: Instrument description Series 2600B System SourceMeter® instrument User's Manual
3-6 2600BS-900-01 Rev. A / August 2021
Figure 4: 2602B and 2612B rear panel
Figure 5: 2604B rear panel
Series 2600B System SourceMeter® instrument User's Manual Section 3: Instrument description
2600BS-900-01 Rev. A / August 2021 3-7
Figure 6: 2614B rear panel
Figure 7: 2634B rear panel
Section 3: Instrument description Series 2600B System SourceMeter® instrument User's Manual
3-8 2600BS-900-01 Rev. A / August 2021
Figure 8: 2635B rear panel
Figure 9: 2636B rear panel
Series 2600B System SourceMeter® instrument User's Manual Section 3: Instrument description
2600BS-900-01 Rev. A / August 2021 3-9
1. SMU connector
Channel A
2601B, 2602B, 2604B, 2611B, 2612B, 2614B
Channel B 2602B, 2604B, 2612B, 2614B
This connector provides input/output connections for HI and LO, sense (S HI/S LO), and guard (G). Connections are as follows: LO = LO
S LO = Sense LO G = Guard S HI = Sense HI HI = HI
Channel A 2634B, 2635B, 2636B
Channel B 2634B, 2636B
These triaxial connectors provide input/output connections for HI and LO, sense HI and sense LO, guard, and ground. See connections at left. The connector conductors are as follows: * Center conductor
** Inner shield *** Outer shield
2. Cooling exhaust vents
Exhaust vent for the internal cooling fan. Keep the vent free of obstructions to prevent overheating. Also see
Cooling vents (on page 2-1).
Section 3: Instrument description Series 2600B System SourceMeter® instrument User's Manual
3-10 2600BS-900-01 Rev. A / August 2021
3. Digital I/O
2601B, 2602B, 2611B, 2612B, 2635B, 2636B
Female DB-25 connector. Use a cable equipped with a male DB-25 connector (L-com part number CSMN25MF-5).
Pins provided: Fourteen digital input or output pins, seven GND pins, and three +5 V pins. The 2601B, 2602B, and 2604B have an output enable pin. The 2611B, 2612B, 2635B, and 2636B have an interlock pin.
2604B
Pins provided: One output enable pin, seven GND pins, and three +5 V pins. The digital input and output pins are not available on the 2604B.
2614B, 2634B
Pins provided: One interlock pin, seven GND pins, and three +5 V pins. The digital input and output pins are not available on the 2614B and 2634B.
4. IEEE-488
Connector for IEEE-488 (GPIB) operation. Use a shielded cable such as the Keithley Instruments Model 7007-1.
5. LAN
RJ-45 connector for a local area network (LAN). The LAN interface supports Auto-MDIX, so you can use either a CAT-5e crossover cable or a normal CAT-5e straight-through cable.
6. USB port
Use the USB-2.0 receptacle (Type B) on the rear panel to connect the instrument to a computer. You can use this connection to send commands to the instrument.
Series 2600B System SourceMeter® instrument User's Manual Section 3: Instrument description
2600BS-900-01 Rev. A / August 2021 3-11
7. Ground
2601B, 2602B, 2604B, 2611B, 2612B, 2614B
Ground terminal for connection output HI or LO to chassis ground. Ground screw for connections to chassis ground.
2635B
Triaxial connector on ground module. Screw terminal connector on ground module.
2634B, 2636B
Triaxial connector on ground module. Screw terminal connector on ground module.
8. TSP-Link
Expansion interface that allows a 2600B and other TSP-enabled instruments to trigger and communicate with each other. Use a category 5e or higher LAN crossover cable. The TSP-Link is not available on the 2604B, 2614B, and 2634B.
9. Power module
Contains the ac line receptacle and power line fuse. The instrument can operate on line voltages of 100 V ac to 240 V ac at line frequencies of 50 Hz or 60 Hz.
10. RS-232
Female DB-9 connector. For RS-232 operation, use a straight-through (not null modem) DB-9 shielded cable to connect to the computer.
Section 3: Instrument description Series 2600B System SourceMeter® instrument User's Manual
3-12 2600BS-900-01 Rev. A / August 2021

Menu overview

The following topics describe how to work with the front-panel menus.

Menu navigation

To navigate through the menus and submenus, the 2600B must not be in edit mode (the EDIT indicator is not illuminated).
Selecting menu items
To navigate the Main and Configuration menus, use the front-panel keys as follows:
Press either CURSOR arrow key to highlight an option.
Rotate the navigation wheel (clockwise or counterclockwise) to highlight an option.
Press the ENTER key (or the navigation wheel) to select an option.
Use the EXIT (LOCAL) key to cancel changes or to return to the previous menu or display.
For quick menu navigation, turn the navigation wheel to highlight an option and then press the navigation wheel to select the highlighted option.

Menu trees

You can configure instrument operation through the menus that are accessed from the front panel.
Main menu
The main menu structure is summarized in the following figure and table. For other menu items, see
Configuration menus (on page 3-15).
Series 2600B System SourceMeter® instrument User's Manual Section 3: Instrument description
2600BS-900-01 Rev. A / August 2021 3-13
Figure 10: Main menu tree
1 Mutually exclusive
2 TSP-Link is not available on the 2604B, 2614B, and 2634B
3 DIGOUT is not available on the 2604B, 2614B, and 2634B
The following table contains descriptions of the main menu options and cross-references to related information.
To access a menu option, press the MENU key, turn the navigation wheel to move the cursor to select an item, and press the navigation wheel.
Section 3: Instrument description Series 2600B System SourceMeter® instrument User's Manual
3-14 2600BS-900-01 Rev. A / August 2021
Menu selection
Description
For more information, see:
SCRIPT
Saves and recalls user scripts
Series 2600B Reference Manual, “Manage scripts”
- LOAD
Loads scripts into nonvolatile memory
- SAVE
Saves scripts
SETUP
Saves and recalls user and factory setup options
Series 2600B Reference Manual, “Saved setups”
- SAVE
Saves user setup options
- RECALL
Recalls user setup options
- POWERON
Sets the configuration used during startup
GPIB
Configures the GPIB interface options
Series 2600B Reference Manual, “Remote communications interfaces”
- ADDRESS
Configures the address for the GPIB interface
- ENABLE
Enables and disables the GPIB interface
LAN
Configures the local area network (LAN)
Series 2600B Reference Manual, “LAN communications”
- STATUS
Displays LAN connection status
- CONFIG
Configures the LAN IP address and gateway
- APPLY_SETTINGS
Applies changes made using the CONFIG menu
- RESET
Restores the default settings
- ENABLE
Enables and disables the LAN interface
RS232
Controls the options for the RS-232 interface
Series 2600B Reference Manual, “Remote communications interfaces”
- BAUD
Sets the baud rate
- BITS
Configures the number of bits
- PARITY
Sets the parity
- FLOW-CTRL
Configures the flow control
- ENABLE
Enables and disables the RS-232 interface
TSPLINK
Configure the instrument in a TSP-Link® network
Series 2600B Reference Manual, TSP-Link system expansion interface”
- NODE
Selects the instrument node identifier
- RESET
Resets the TSP-Link network
UPGRADE
Upgrades the firmware from a USB flash drive
Upgrading the firmware (on page 5-4)
DISPLAY
Accesses display functions
Front-panel tests (on page 5-3)
- TEST
Runs the display test
See Numeric entry method in Setting a
value (on page 3-19)
- NUMPAD
Enables and disables the numeric keypad
DIGOUT
Controls digital outputs
Series 2600B Reference Manual, “Digital I/O”
- DIG-IO-OUTPUT
Selects the digital I/O values
- WRITE-PROTECT
Write-protects specific digital I/O lines
- LEGACY-MODE
Sets digital I/O to work like an older SourceMeter instrument
BEEPER
Controls the key beeps
Beeper (on page 3-21)
- ENABLE
Enables the key beeps
- DISABLE
Disables the key beeps
LINE-FREQ
Configures the line frequency
Line frequency configuration (on page 2-4)
- 50Hz
Set the line frequency to 50 Hz
- 60Hz
Set the line frequency to 60 Hz
- AUTO
Enables automatic line-frequency detection during start up
SYSTEM-INFO
Displays the system information
System information (on page 2-4)
- FIRMWARE
Displays the version of firmware installed
- SERIAL#
Displays the serial number of the unit
- CAL
Displays the last calibration date
- MEMORY-USAGE
Displays memory usage in percentage
RESET-PASSWORD
Resets the system password
Series 2600B Reference Manual, Password management”
Loading...
+ 72 hidden pages