Software Upgrade
This kit includes materials to upgrade the basic and optional software for the
following instruments with Windows XP operating system:
HWCA230A and WCA280A Wireless Communication Analyzers
HRSA2203A and RSA2208A Real-Time Spectrum Analyzers
HRSA3303A and RSA3308A Real-Time Spectrum Analyzers
HRSA3408A Real-Time Spectrum Analyzer
The software is contained in the CD-ROM supplied with this kit. You have to
install the software via LAN from a CD-ROM drive of an external PC to the
analyzer. Prepare a Windows PC with which the analyzer is able to share the
CD-ROM drive.
There are two ways to connect the analyzer and the PC (see Figure 1):
HConnect the analyzer to a single PC using the LAN cross cable supplied with
HConnect the analyzer to a PC via LAN.
this kit.
LAN cross cable supplied with this kitEthernet LAN
The software is installed via network from the CDĆROM drive of an external PC to the analyzer.
Figure 1: Connection between the analyzer and PC
NOTE. When you connect the analyzer to LAN (right case in Figure 1), contact
your system administrator for the network settings.
The details are provided on the following pages.
WCA & RSA Software Upgrade Instructions
or
1
Software Upgrade
Before Starting the Software Installation
Before starting installation of the latest version of software, you are requested to
perform setup for the analyzer and for the PC.
Setting Up the Analyzer
Checking the Version of Existing Application Software.
1. Turn the power to the analyzer on.
2. Press the SYSTEM key on the front panel.
3. Press the Versions and Installed Options... side key.
The screen shown in Figure 2 appears.
Main system version
Optional software versions
Figure 2: Checking the software version (an example of RSA3408A)
Note the versions of the main system and optional software (if installed).
Connecting a Mouse and a Keyboard. Connect a USB mouse and a USB keyboard
to the USB ports on the side panel. You can connect them to either or both ports.
You can also connect the mouse to the USB port on the keyboard.
WCA & RSA Software Upgrade Instructions
2
Software Upgrade
USB ports
Front
Figure 3: USB ports (side panel)
Displaying Windows XP Desktop. Display the Windows XP desktop on the
analyzer screen using these steps:
1. With the mouse, move the pointer to the bottom of screen.
The task bar appears.
2. Place the pointer on the icon indicating the analyzer model name located on
the task bar, and click on it with the right mouse button. A menu opens.
The figure below is an example of WCA280A.
3. Select Close from the menu.
The analyzer system program stops, and the Windows XP desktop displays.
WCA & RSA Software Upgrade Instructions
3
Software Upgrade
Network Settings for the Analyzer. No network setting is made for the analyzer
when it is shipped from the factory. Configure your analyzer using the following
procedure:
HWhen you connect the analyzer to a single PC with the LAN cross cable:
Use the following steps to set the networking parameters.
HWhen you connect the analyzer to a PC via LAN:
Contact your system administrator for the network settings. If the analyzer is
already connected to the LAN, set up the CD-ROM drive of a shared PC
(other than the analyzer) and insert the CD-ROM (supplied with this upgrade
kit) into the drive to upgrade the software in accordance with Installing theSoftware on page 10.
1. From the Windows desktop of the analyzer, select the following:
Start → Setting → Control Panel → Network Connections.
2. Double-click Local Area Connection.
WCA & RSA Software Upgrade Instructions
4
Software Upgrade
3. In the Local Area Connection Properties dialog box, select Internet Protocol (TCP/IP) in the General tab and then click the Properties button.
4. In the Internet Protocol (TCP/IP) Properties dialog box, select Use the following IP address and then enter the IP address and the subnet mask.
For example, enter 192.168.10.1 for the IP address and 255.255.255.0 for the
subnet mask.
When finished, click the OK button.
WCA & RSA Software Upgrade Instructions
5
Software Upgrade
In the Local Area Connection Properties dialog box, click the
5.
Advanced tab
.
6. Clear the Internet Connection Firewall checkbox and then click OK.
WCA & RSA Software Upgrade Instructions
6
Software Upgrade
In the Internet Connection Firewall dialog box, click Ye s .
7.
8. When you finish the network settings, close all the windows being open.
9. Press the ON/STANDBY switch on the front panel to turn the power off.
Setting Up the PC
Network Settings for the PC. Set networking parameters on the PC in the same
manner as for the analyzer. Refer to Network Settings for the Analyzer on
page 4 for the procedure.
But in step 4 on page 5, enter a different PC IP address than the analyzer
address. For example, enter 192.168.10.2 for the IP address and 255.255.255.0
for the subnet mask as shown in the figure below.
WCA & RSA Software Upgrade Instructions
7
Software Upgrade
Setting Up the CDĆROM Sharing.
1. On the Windows desktop, double-click the My Computer icon.
2. Select the CD-ROM drive icon and then click on it with the right
mouse button. A menu appears.
3. From the menu, select Sharing...
4. The CD-ROM Property dialog box opens. From the Sharing tab, select Share this folder.
5. Insert the CD-ROM supplied with this kit into the shared CD-ROM drive.
Connecting the Analyzer with the PC. Connect the analyzer with the PC via LAN.
Figure 4 shows the Ethernet connector location on the analyzer.
Figure 4: Ethernet connector (side panel)
Ethernet 10/100BASEĆT connector
Front
WCA & RSA Software Upgrade Instructions
8
Software Upgrade
1. Connect a LAN cable.
NOTE. Leave the mouse and the keyboard connected to the analyzer.
HWhen you connect the analyzer with a single PC:
Connect the LAN cross cable supplied with this kit to the analyzer and
to the PC Ethernet port.
LAN cross cable
included in the kit
HWhen you connect the analyzer to a PC via LAN:
NOTE. When you connect the analyzer to your LAN, contact your system
administrator for the network settings such as IP address. Do not use the LAN
cross cable (supplied with the kit) for the connection in this case.
Ethernet LANEthernet LAN cable
2. Power on the analyzer.
3. Exit the analyzer system program to display the Windows XP desktop.
(Refer to Displaying Windows XP Desktop on page 3 for the procedure.)
WCA & RSA Software Upgrade Instructions
9
Software Upgrade
Installing the Software
Search for the PC to be shared, and then install the new software from the
CD-ROM supplied with the upgrade kit.
Searching for the PC
Being Shared
Search for the PC having the shared CD-ROM drive:
1. From the Windows desktop of the analyzer, click the My Network Places
icon, and then click the right mouse button. From the menu, select
Search for Computers...
2. In the Search Results – Computers dialog box, enter the shared PC name inthe Computer name field (for example, “PC123”), and then click the
Search button to start searching for the PC. If the PC has been found, thename appears in the Name list.
10
WCA & RSA Software Upgrade Instructions