Tektronix RSA507A, RSA607A, RSA603A Declassification

xx
RSA500A and RSA600A Series Real-Time Spectrum Analyzers
ZZZ
Declassication and Security
Instructions
www.tek.com
*
077-1202-01
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O. Box 500 Beaverto USA
For product information, sales, service, and technical support:
n, OR 97077
In North America, call 1-800-833-9200. Worldwide, visit www.tek.com to nd contacts in your area.
Table of Contents
Preface .............................................................................................................. iii
Memory devices ............... ................................ ................................ ..................... 1
Clear or sanitize a non-functional instrument .. . ..... . ... . . ..... . ..... . .... . ..... . ..... . ... . . ..... . ..... . .... . .. 2
RSA500A and RSA600A Series Declassication and Security Instructions i
Table of Contents
ii RSA500A and RSA600A Series Declassication and Security Instructions
Preface
This document helps customers with data security concerns to sanitize or remove memory devices from the Tektronix RSA500A Series, and RSA600A Series Real-Time Sp
This product has data storage (memory) devices and data output devices (USB ports). These instructions tell how to clear or sanitize the memory devices and disable the data output devices. The instructions also tell how to declassify an instrument that is not functioning.
ectrum Analyzers.
Reference
Products
Required documents
Terms that may be used in
this document
The procedures in this document are written to meet the requirements specied in:
NISPOM, DoD 5220.22–M, Chapter 8
ISFO Process Manual for Certication & Accreditation of Classied Systems under NISPOM
This document covers the following Tektronix products:
RSA503A
RSA507A
RSA513A
RSA518A
RSA603A
RSA607A
No external documents are required to perform the procedures in this document.
The following terms may be used in this document:
Clear. This removes data on media/memory before reusing it in a secured area. All reusable memory is cleared to deny access to previously stored information by standard means of access.
Erase. This is equivalent to clear.
Instrument Declassication. A term that refers to procedures that must be
undertaken before an instrument can be removed from a secure environment. Declassication procedures include memory sanitization and memory removal, and sometimes both.
Media storage/data export device. Any of several devices that can be used to store or export data from the instrument, such as a USB port.
Nonvolatile memory. Data is retained when the instrument is powered off.
RSA500A and RSA600A Series Declassication and Security Instructions iii
Preface
Power off. Some still supplied to the instrument. For the purpose of clearing data, putting the instrument in Standby mode does not qualify as powering off. For these products, you will need to either press a rear-panel OFF switch or remove the power source from the instrument.
Remove. Thisisaphysicalmeanstoclearthedatabyremovingthememory device from the instrument. Instructions are available in the product Service Manual.
Sanitize. This eradicates the data from media/memory so that the data cannot be recovered by other means or technology. This is typically used when the device wi non-secured area.
Scrub. T
User Accessible. User is able to directly retrieve the memory device contents.
User-modiable. The user can write to the memory device during normal
instrument operation, using the instrument interface or remote control.
User-modiable. The user can write to the memory device during normal instrument operation, using the instrument interface or remote control.
Volatile memory. Data is lost when the instrument is powered off.
ll be moved (temporarily or permanently) from a secured area to a
his is equivalent to sanitize.
instruments have a “Standby” mode, in which power is
Device terms
User data. Describes the type of information stored in the device. Refers to waveforms or other measurement data representing signals connected to the instrument by users.
User settings. Describes the type of information stored in the device. Refers to instrument settings that can be changed by the user.
Both. Describes the type of information stored in the device. It means that both user data and user settings are stored in the device.
None. Describes the type of information stored in the device. It means that neither user data nor user settings are stored in the device.
Directly. Describes how data is modied. It means that the user can modify the data.
Indirectly. Describes how data is modied. It means that the instrument system resources modify the data and that the user cannot modify the data.
iv RSA500A and RSA600A Series Declassication and Security Instructions
Memory devices
Volatile memory devices
Type
of user Type a nd min. size Function
CYUSB3014 USB3
interface
info
stored
Spectrum
data in
FIFOs
Backed up by battery
No Run the
Method of modication
instrument
Data Input method Location
Firmware operations
U3001 No Disconnect
User accessible To clear To sanitize
power for at least 5 seconds.
Disconnect power for at least 5 seconds.
Non-volati
le memory
Type and min. size Function
16 MB Flash
16 MB Fla (RSA500P instruments only)
sh
Holds USB and FPGA images, calibration constant and board ID
Holds FPGA image and
tion
calibra constants
Data export devices
devices
s,
Typ e of user info stored
None
None
Data Method of modicati
USB 3.0
USB 3.0
Input
on
method Location
Firmware
operations
e
Firmwar
operations
User accessibl
U3003 No Not applicable.
U1201 No Not appl
e
To clear To sanitiz
Does not contain user data or settings. Clearing
able
would dis instrument functionality.
icable. Does not contain user data
ings.
or sett Clearing would disable
ument
instr functionality.
e
Not applicable. Does not contain us data or settings. Clearing would dis instrument functionality.
Not appl Does not contain user data or Clearing would disable instr functionality.
er
able
icable.
settings.
ument
There are no media or export devices that require declassication or security procedures with this product.
RSA500A and RSA600A Series Declassication and Security Instructions 1
Clear or sanitize a non-functional instrument
Clearorsanit
Charges
ize a non-functional instrument
If your instr No user data is stored on the instrument when the power is removed. Clearing or sanitizing the instrument is not applicable.
Replacemen the time of replacement.
ument is not functioning, return the instrument to Tektronix for repair.
t of any missing hardware will be charged according to the rate at
2 RSA500A and RSA600A Series Declassication and Security Instructions