Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Table of Contents
Service Safety Summary.......... .................................. ................................ ...............1
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary located in the RSA5100A
Series Real-
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnect Power. To avoid electric shock, switch off the instrument power, then
disconnect the power cord from the mains power.
UseCareWhenServicingWithPowerOn. Dangerousvoltagesorcurrentsmay
exist in this product. Disconnect power, remove battery (if applicable), and
discon
components.
Time Signal Analyzers Service Manual before performing any service
nect test leads before removing protective panels, soldering, or replacing
To avo
id electric shock, do not touch exposed connections.
RSA5100A Series Option 55 Upgrade1
Kit Description
Kit Description
Products
Kit Parts List
This kit describes the installation of a Digital I and Q Output (Option 55) in an
RSA5100A Series Real-Time Signal Analyzer.
RSA5100A Series. All serial numbers
QuantityPart numberDescription
1050-XX
1
1
S
-XX
-XX
13-XX
MANUAL, TECH INSTALLATION, RSA5UP OPTION 55
CA ASS Y, SP; IDC, TW FLAT, I & Q OUTPUT TO REAR
PANEL, 50 POS; SAFETY CONTROLLED; RSA51XXA
OPTION 5
CIRCUIT BD ASSY; REALTIME I/Q OUTPUT,
FUN
389367200, WIRED
DATA SHEET; SOFTWARE OPTION KEY
AUTHORIZATION CERTIFICATE, UPGRADE
TS
KI
ABEL, MANUFACTURED; OPTION KEY UPGRADE
L
LABEL 2.100 X 2.700, SAFETY CONTROLLED
5
SY ELEC: 50P MDR CONN, 1.5M L, 18 TWIST
)
W, MACHINE; 2.5MM X 0.45 X 6MM, PNH, STEEL,
CTIONAL BOARD TESTED LEVEL, 679589800,
1 ea071-2880
2 ea174-5106
2 ea174-5195-XX
1 ea174-52
1 ea200-5195-XX
4 ea211-0450-XX
8 ea211-
1 ea664-5898-XX
a
1e
1ea
1
NS — Not saleable
NS
N
2RSA5100A Series Option 55 Upgrade
Installation Instructions
This section contains all procedures needed to install the Digital I and Q Output
option (Option 55) in RSA5100A Series instruments.
Minimum Tool and Equipment List
The following tools are required to remo ve and install the instrument covers and
panels. All
ItemNameDescription
1
2No. 2 Phil
3
4
5
tools are standard tools that should be readily available.
Screwdriver handle (magnetic)
T15 TORX tipTORX driver tip for T-15 size screw
T20 TOR
5/32” hex wrench
lips or Pozidriv tip
Xtip
Installation Instructions
Torque dri
hex-head driver tips
Phillips or Pozidriv-driver tip for number
2 size screw heads
heads
TORX dr
heads
Hex wre
screws at front of top cover
ver. Accepts 14 inch
iver tip for T-20 size screw
nch to remove Allen head
These instructions are for qualified service personnel who are familiar
with servicing the product. If you need further details for disassembling or
reassembling the product, refer to the the RSA5100A Series Real-Time Signal
Anal
Remove Cosmetic Covers
NOTE. Right-side or left-side references in these instructions assume you are
viewing the instrument from the front panel.
WAR NING. To avoid electric shock, switch off the instrument power, then
disconnect the power cord from the mains power. Failure to do so can cause
injury or death.
1. Remove the power cord.
2. If it is installed, pull the front cover off the instrument.
yzers Service Manual, Tektronix part number 077-0522-XX.
RSA5100A Series Option 55 Upgrade3
Installation Instructions
3. Remove the two T
of the instrument. (It is not necessary to remove the black metal handles.)
(See Figure 1.)
4. Remove four T15 black screws along each side that secure the top and bottom
covers to the instrument, and two T20 Torx-head screws near the front edge of
the top cover (next to the folding handles).
5. Remove the top and bottom covers. Remove the top cover by pulling straight
back about 1 inch. Then pull out on the sides of the top cover outward,
flexing them slightly to clear the instrument chassis, and pull it away from the
instrument.
15 screws that secure the plastic carrying handle to the side
4RSA5100A Series Option 55 Upgrade
Installation Instructions
Figure 1: Remove cosmetic covers
RSA5100A Series Option 55 Upgrade5
Installation Instructions
Install RTIQ B
oard
1. Remove the 18 T15 screws that secure the internal top cover to the chassis,
and then lift the internal top cover away.
2. Install the Option 5 5 (RTIQ) board into Slot 3 (counting from the rear of the
instrument). (See Figure 2.)
CAUTION. Be
installing the RTIQ board.
careful not to bend the pins on the Digital Interface Board when
Figure 2: Installing the RTIQ board
Install Rear Panel Connectors
1. Remove the blank, rear-panel cover from the chassis (attached with eight
screws). Discard the blank panel.
2. Install the two IQ Output cables (Tektronix part number 174-5106-XX) on
the rear panel cover (Tektronix part number 200-5195-XX) using four screws
6RSA5100A Series Option 55 Upgrade
Installation Instructions
(Tektronix par
using a torque driver set to 5.5 in-lb. (See Figure 3.)
t number 211-0450-XX) provided in the kit. Tighten the screws
Figure 3
3. Instal
: Securing the output cables to the rear panel cover
l the RTIQ Ground cable (Tektronix part number 174-5132-XX) onto
the spade connector on the rear panel cover.
Install Rear Panel Cover Onto Rear Panel
1. Install the rear panel cover with attached cabling on the rear panel using eight
T15 screws (Tektronix part number 211-1050-XX). Torque these screws to
10.0 in/lb. (See Figure 4.)
RSA5100A Series Option 55 Upgrade7
Installation Instructions
Figure 4: I
Connect Rear Panel Cables
1. Use the board ejectors to lift the RTIQ board up out of its slot. Connect the
2-wire RTIQ ground cable from the rear-panel spade connector to J20 on the
RTIQ board. Gently pull on the cable to ensure the locking tabs are engaged.
(See Figure 5.)
nstalling the rear panel cover on the rear panel of the chassis
Figure 5: Connecting the output cables
8RSA5100A Series Option 55 Upgrade
Installation Instructions
3. Connect the IQ o
4. Push the RTIQ board back into the Digital Interface board so that it is fully
seated. Be car
5. Reinstall the top internal cover.
a. Place the top internal cover onto the instrument, aligning the two
protrusions on the cover with the two slots in the chassis.
b. Replace the 18 T15 screws that secure the top internal cover to the chassis.
Torque these screws to 8.0 in/lb.
utput cables as shown. (See Figure 5.)
eful not to bend the pins on the Digital Interface board.
RSA5100A Series Option 55 Upgrade9
Installation Instructions
Reinstall Cos
metic Covers
1. Reinstall the top and bottom covers.
a. Place the instrument on its rear feet, so the front panel is facing up and the
top is towards you.
b. Place the top cover over the top of the instrument and slide it toward
the front panel. Make sure that the top cover wraps around the flanges
on the rear panel on all three sides.
c. Reinstall the four T20 Torx-head screws (two on each side) near the
front edge of the top cover (next to the folding handles) that secure the
top cover
d. Rotate the instrument so the bottom faces you.
e. Place the bottom cover on the instrument, with the flip feet towards the
front.
f.Align the four screw holes on each side in the top and bottom covers with
the holes in the chassis, and insta ll eight T15 screws, four on each side.
Torq u
g. Position the plastic carrying handle and its bracket on the right side of
the i
Torque these screws to 8.0 in/lb.
to the instrument.
e these screws to 8.0 in/lb.
nstrument, and install the two T15 screws that secure it in place.
Install Option Key
To activate the Option 55 Digital I and Q Output option, you must enter a new
option key.
1. Power on the analyzer.
NOTE. When the analyzer application launches, it will display an error message
indicating that the current option key does not support the new option hardware.
Click OK to clear the error message.
2. In the signal analyzer application, select Tools > Install Upgrades .The
Install Upgrade wizard starts. (See Figure 6.) Click Continue.
10RSA5100A Series Option 55 Upgrade
Installation Instructions
Figure 6: S
3. The Insta
Figure 7: Entering the option installation key
4. Type the option installation key into the empty text box and click Continue.
tarting the option upgrade
ll Upgrade Enter Key screen appears as shown.
An Installation Success dialog box appears after you click Continue.
5. Click Close. Exit and restart the RSA5100A application to enable the new
feature.
6. Select Help > About Tektronix Real Time Signal Analyzer.
7. Look in the Installed Options listing and verify that Signal Output Package
(Opt 55) appears in the listing.
8. Click OK.
RSA5100A Series Option 55 Upgrade11
Installation Instructions
Installing th
e Option Key Label
Place the new option key label over the existing label on the instrument rear panel.
This completes the installation of the Option 55 Digital I and Q Output upgrade kit.
■ End of Document ■