Service safety summary
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary located in the Service Manual
before perfo
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
rming any service procedures.
Disconnec
disconnect the power cord from the mains power.
UseCareWhenServicingWithPowerOn. Dangerousvoltagesorcurrentsmay
exist in this product. Disconnect power, remove battery (if applicable), and
disconnect test leads before removing protective panels, soldering, or replacing
compon
To avoid electric shock, do not touch exposed connections.
tPower. To avoid electric shock, switch off the instrument power, then
ents.
RSA5100A/B Series Digital I/Q and Zero Span Upgrades1
Kit description
Kit description
This kit describes the installation of the following Options:
Products
RSA5100B Ser
Option 65: Digital I and Q Outputs
Option 66: Zero Span Output
Option 6566: Digital I and Q Outputs and Zero Span Output
RSA5100A Series.
Option 5566: Digital I and Q Outputs and Zero Span Output
RSA5103B. All serial numbers
RSA5106B. All serial numbers
15B. Allserialnumbers
RSA51
RSA5126B. All serial numbers
RSA5103A. All serial numbers
ies.
5106A. All serial numbers
RSA
RSA5115A. Allserialnumbers
RSA5126A. All serial numbers
2RSA5100A/B Series Digital I/Q and Zero Span Upgrades
Kit parts list
Kit description
The following table provides a list of parts contained in this kit. The kit contains
the same parts, regardless of which of the three options you ordered. The actual
option (or options) available is controlled with the option key provided in the kit.
NOTE. These cables are not needed to install the Option.
They are supplied for external connections to the I and Q
connectors.
Output
CABLE ASSY, RIBBON; STATIC GROUND, 1X2; SAFETY
OLLED
CONTR
CABLE, COAX BNC-ST TO SMB-RA
COVER, REAR PANEL, OPTION (I AND Q / ZERO SPAN)
W, MACHINE; 2.5MM X 0.45 X 6MM, PNH, STEEL,
SCRE
ZINC PLATED, T8 TORX DRIVE
EW, MACHINE; 6-32 X 0.312 L, PNH, STEEL, ZINC
SCR
FINISH, T15
T BD SUBASSY; REAL TIME IQ AND LOG VIDEO
CK
OUTPUT, FUNCTIONALLY TESTED
RSA5100A/B Series Digital I/Q and Zero Span Upgrades3
1
/A
N
ABEL, EXPORT CONTROL CLA SSIFICATION
L
NOTE. Export control label only applies to RSA5100B
Series instruments when installing Option 65 or Option 6566.
1
1
N/ALABEL, MANUFACTURED; OPTION KEY UPGRADE
LABEL
N/ALABEL, MANUFACTURED; PRODUCT LABEL, SAFETY
CONTROLLED
Installation instructions
Installation instructions
This section contains all procedures needed to install the required components.
Minimum tool
and equipment list
The following tools are required to for installation of this k it. All tools are
standard tools that are readily available.
ItemNameDescription
1.
2.
3.
4.
5.
6.
These instructions are for qualified service personnel who are familiar
with servicing the product. If you need further details for disassembling or
reassembling the product, refer to the product’s Service Manual.
Screwdriver handle (magnetic)
T8 TORX tipTORX driver tip for T-8 size screw
T15 TORX tipTORX driver tip for T-15 size screw
T20 TORX tipTORX driver tip for T-20 size screw
5/8" hex wrenchOpen end wrench to install BNC
5/32" hex wrench
NOTE. Only needed for early models
of the RSA5100A Series instruments.
Torque driver. Accepts 14 inch
hex-head driver tips
heads
heads
heads
connector
Hex wrench to remove Allen head
screws at front of top cover
ove cosmetic covers and shield
Rem
NOTE. Right-side or left-side references in these instructions assume you are
viewing the instrument from the front panel.
WARNING. To avoid electric shock, switch off the instrument power, then
disconnect the power cord from the mains power. Failure to do so can cause
injury or death.
1. Remove the power cord.
2. If it is installed, pull the front cover off the instrument.
4RSA5100A/B Series Digital I/Q and Zero Span Upgrades
Installation instructions
3. Remove the two T
15 Torx-head screws that secure the plastic carrying handle
to the side of the instrument. (It is not necessary to remove the black metal
handles.) (See Figure 1.)
4. Remove four T15 Torx-head screws along each side that secure the top and
bottom covers to the instrument, and two T20 Torx-head screws near the fr ont
edge of the top cover (next to the folding handles).
NOTE. Early models of the RSA5100A Serie s instruments used four 5/32″ Allen
socket cap screws in place of the T15 Torx-head screws.
5. Remove the top and bottom covers. Remove the top cover by pulling straight
back about 1 inch. Then pull out on the sides of the top cover outward,
flexing t
hem slightly to clear the instrument chassis, and pull it away from the
instrument.
6. Remove t
he eighteen T15 Torx-head screws that secure the top shield to the
chassis and remove the shield.
RSA5100A/B Series Digital I/Q and Zero Span Upgrades5
Installation instructions
Figure 1: Remove covers
6RSA5100A/B Series Digital I/Q and Zero Span Upgrades
Attach cables to rear-panel cover insert
1. Install the cables to the new rear-panel cover provided in this kit. (See
Figure 2.)
Install the two IQ Output ribbon cables onto the rear panel cover using
four T8 Torx-head screws provided in the kit. Tighten the screws using a
torque driver set to 5.5 in-lb.
Install the Ground cable onto the spade connector on the rear-panel cover.
Install the Zero Span Output BNC cable to the rear-panel cover. Use the
BNC washer and 5/8″ nut that is provided with the BNC connector.
Installation instructions
Fig
RSA5100A/B Series Digital I/Q and Zero Span Upgrades7
ure 2: Installing cables
Installation instructions
Install rear panel cover insert into instrument
1. Remove the blank, rear-panel cover from the chassis (attached with eight T15
Torx-head screws). Discard the blank panel.
2. Install the new rear-panel cover (with attached cabling) on the rear panel
chassis using eight T15 Torx-head screws. Torque these screws to 10.0 in/lb.
(See Figure 3.)
Figure
3: Installing the rear panel cover on the rear panel of the chassis
8RSA5100A/B Series Digital I/Q and Zero Span Upgrades
Install option board
Installation instructions
1. Slide the option board (Digital I and Q / Zero Span) into Slot 3 (counting from
the rear of the instrument). (See Figure 4.)
CAUTION. Be careful not to bend the pins on the Interface Board when installing
the option board.
e 4: Installing the option board
Figur
2. Conn
RSA5100A/B Series Digital I/Q and Zero Span Upgrades9
ect the cables from the rear-panel cover to the Option board. (See
Figure 5.)
nect the 2-wire ground cable.
Con
Connect the I a nd Q output cables.
Connect the Zero Span output cable.
Installation instructions
Figure 5: Connecting the output cables
Reinstall top shield and cosmetic covers
1. Place t
2. Place the top shield on top of the instrument and reinstall the eighteen T15
Torx-
3. Place the instrument on its rear feet, so the front panel is facing up and the
top i
4. Place the top cover over the top of the instrument and slide it toward the
fro
rear panel on all three sides.
5. Re
edge of the top cover (next to the folding handles) that secure the top cover to
the instrument. Torque these screws to 8.0 in/lb.
6. Rotate the instrument so the bottom faces you.
7.P
he instrument on its bottom feet.
head screws.
s toward you.
nt panel. Make sure that the top cover wraps around the flanges on the
install the four T20 Torx-head screws (two on each side) near the front
lace the bottom cover on the instrument, with the flip feet towards the front.
10RSA5100A/B Series Digital I/Q and Zero Span Upgrades
Installation instructions
Install option key
8. Align the four s
holes in the chassis, and install e ight T15 Torx-head screws, four on each
side. Torque these screws to 8.0 in/lb.
9. Position the plastic carrying handle and its bracket on the right side of the
instrument, and install the two T15 Torx-head screws that secure it in place.
Torque these screws to 8.0 in/lb.
NOTE. RSA5100A Series - If adding Option 5566 to an RSA5100A Series
instrument, the software most likely requires updating to recognize the new
feature. This should be done before installing the new option key.
See the procedure “Verify instrument software version”.
To activate your new option, you must enter a new option key.
1. Power on the instrument.
NOTE. When the application launches, it will display an error message indicating
that the current option key does not support the new option hardware. Click
OK to clear the error message.
crew holes on each side in the top and bottom covers with the
2. Select Tools > Install Upgrades... to start the upgrade installation process.
3. Click Continue from the Install Upgrades introduction screen.
RSA5100A/B Series Digital I/Q and Zero Span Upgrades11
Installation instructions
4. Enter the optio
instructions to install the option.
n key provided by Tektronix, and follow the on-screen
Attach labels
Attach the option key label
Attach the product/option
lab
5. Power off the instrument, then power back on.
6. In the Help menu, select About Tektronix Real-Time Analyzer.
7. Ver if
Plea
Place the new option key label over the existing label on the rear panel.
Place the new product label over the existing label on the rear panel.
y the new option is listed.
se attach all labels provided in this kit on the instrument’s rear panel.
el
12RSA5100A/B Series Digital I/Q and Zero Span Upgrades
Verify instrument software version
Attach the Export Control
classification label
Place the Expor
installed the Digital I and Q Outputs option.
NOTE. The Export Control label applies only to the RSA5100B Series instruments
when installing Option 65 or Option 6566.
t Control classification label on the instrument rear panel if you’ve
Verify instrument software version
Tektronix recommends to update to the latest available application software for
your instrument.
1. Select Help > About to check the software version.
2. Use your W
3. Search for your instrument’s model number and follow the link to the software.
4. If the installed software is older than that available, download the software.
5. Follow the instructions on the Web page to install the software.
eb browser to go to: www.tektronix.com/software.
■ End of Document ■
RSA5100A/B Series Digital I/Q and Zero Span Upgrades13
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.