Tektronix RSA3303B, RSA3308B, RSA3408B Instructions Manual

Instructions
RSA3303B, RSA3308B, & RSA3408B Real-Time Spectrum Analyzers Declassification and Security
071-2371-00
www.tektronix.com
071237100
071237100
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14200 SW Karl Braun Drive or P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tektronix.com to find contacts in your area.

Table of Contents

Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Clear and Sanitize Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Memory Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Media and Data Export Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
How to Clear or Sanitize a Non-Functional Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . 5
RSA3303B, RSA3308B, & RSA3408B Declassification and Security i
Table of Contents
ii RSA3303B, RSA3308B, & RSA3408B Declassification and Security

Preface

This document helps customers with data security concerns to sanitize or remove memory devices from the RSA3303B, RSA3308B, and RSA3408B Real-Time Spectrum Analyzers.
These products have data storage (memory) devices and data output devices (USB ports). These instructions tell how to clear or sanitize the memory devices and disable the data output devices. The instructions also tell how to declassify an instrument that is not functioning.
Products
Terms
The following Tektronix products are covered by this document:
RSA3303B
RSA3308B
RSA3408B
The following terms may be used in this document:
Clear. This removes data on media/memory before reusing it in a secured area. All reusable memory is cleared to deny access to previously stored information by standard means of access.
Erase. This is equivalent to clear.
Media storage/data export device. Any of several devices that can be used to
store or export data from the instrument, such as a USB port.
Nonvolatile memory. Data is retained when the instrument is powered off.
Power off. Some instruments have a “Standby” mode, in which power is still
supplied to the instrument. For the purpose of clearing data, putting the instrument in Standby mode does not qualify as powering off. For these products, you will need to either press a rear-panel OFF switch or remove the power source from the instrument.
Remove. A physical means to clear the data by removing the memory device from the instrument. Instructions are available in the product service manual.
Sanitize. This eradicates the data from media/memory so that the data cannot be recovered by other means or technology. This is typically used when the device will be moved (temporarily or permanently) from a secured area to a non-secured area.
RSA3303B, RSA3308B, & RSA3408B Declassification and Security iii
Preface
Scrub. This is equivalent to sanitize.
User-modifiable. The memory device can be written to by the user during normal
instrument operation, using the instrument’s user interface or remote control.
Volatile memory. Data is lost when the instrument is powered off.
iv RSA3303B, RSA3308B, & RSA3408B Declassification and Security

Clear and Sanitize Procedures

Memory Devices

The following tables list the volatile and nonvolatile memory devices in the standard instrument and listed options. Detailed procedures to clear or sanitize these devices, if any, are shown following each table.
Table 1: Volatile memory devices
Type and minimum size Function
SDRAM Standard: 64 MB Option 02: 256 MB
Holds active acquisition data
Table 2: Nonvolatile memory devices
Type and minimum size Function
40 GB Fixed Hard Drive
Holds instrument operating system and application software. Holds all user-storable data such as waveforms, measurement results, and instrument settings. (Operating system restore image and calibration data are stored on the D drive.)
User modifiable Input method Location Process to sanitize
No Firmware operations A42 board Remove the power source
from the instrument for at least 20 seconds.
User modifiable Input method Location Process to clear Process to sanitize
Yes Firmware
operations, user input
Internal Perform a recovery
procedure to reinstall the operating system and application software. After performing the recovery procedure erase the free space on the hard drive with commercial erasure software.
Remove the hard drive. Store the removed hard drive in a secure area or destroy it. (See page 2, Fixed Hard Drive Removal.) When it is removed, no user data remains in the instrument.
RSA3303B, RSA3308B, & RSA3408B Declassification and Security 1
Clear and Sanitize Procedures
Clearing Hard Disk Drives
Fixed Hard Drive Removal
If your organization’s security protocols allow the use of software to purge or sanitize hard drives, you can use commercial software to erase free space on the hard drive before sending the instrument out for upgrades or repair. Follow the instructions that come with the software to ensure that the erasure of sensitive data from the hard drive complies with your organization’s security protocols.
Sequence for clearing secure data from a hard drive:
1. Perform a System Recovery
2. Upgrade the instrument application software if necessary.
3. Erase the free space on the hard drive using commercial software.
For a detailed removal and replacement procedure for the fixed hard drive, see the RSA3303B & RSA3308B Service Manual (Tektronix part number 071-2367-XX) or RSA3408B Service Manual (Tektronix part number 071-2366-XX). Refer to the Removal and Installation Procedures section of the manual.
WAR NING. Only qualified service personnel should perform the hard drive
removal procedure. Read the Service Safety Summary and the General Safety Summary in the RSA3303B & RSA3308B Service Manual (Tektronix part number 071-2367-XX) or RSA3408B Service Manual (Tektronix part number 071-2366-XX) before proceeding.
WAR NING. Before working on the spectrum analyzer, disconnect the power cord
from the line voltage source. Failure to do so could cause serious injury or death.
CAUTION. To avoid damaging the hard drive, perform the procedure in a
static-safe environment with proper electro-static discharge controls in place.
2 RSA3303B, RSA3308B, & RSA3408B Declassification and Security

Media and Data Export Devices

The following table lists the media storage and data export devices in the instrument.
Table 3: Media and data export devices
Clear and Sanitize Procedures
Type and minimum size Function
720 KB/1.44 MB floppy disk drive
USB device ports Holds all user-storable
LAN Ethernet connection
GPIB connector Data transfer. N/A Remote control Rear panel N/A
Holds all user-storable data such as waveforms, measurement results, and instrument settings.
data such as waveforms, measurement results, and instrument settings.
Support remote control and data transfer.
User modifiable Input method Location Process to disable
Yes Firmware
operations, User input
Yes User writable Side panel
N/A Remote control
via TekVISA
Side panel Format any disks (do not use “Quick
Format”) and remove them from the drive. Store the disks or destroy them.
Side panel Disconnect from Network cable.
RSA3303B, RSA3308B, & RSA3408B Declassification and Security 3
Clear and Sanitize Procedures
4 RSA3303B, RSA3308B, & RSA3408B Declassification and Security

Troubleshooting

How to Clear or Sanitize a Non-Functional Instrument

If your instrument is not functioning, perform the following actions and return the instrument for Tektronix for repair. Describe the initial problem with the product. Tektronix will install replacement parts and then repair and return the instrument.
Hard Disk Drive
Floppy Disk Drive
External Memory Drive
Remove the internal hard disk using the procedures in the RSA3303B & RSA3308B Service Manual (Tektronix part number 071-2367-XX) or RSA3408B Service Manual (Tektronix part number 071-2366-XX).
Remove any disk in the drive. Store the disk or destroy it.
Remove any USB flash drives or external hard drives from the instrument.
RSA3303B, RSA3308B, & RSA3408B Declassification and Security 5
Troubleshooting
6 RSA3303B, RSA3308B, & RSA3408B Declassification and Security