Tektronix RSA3303B, RSA3308B, RSA3408B User manual

User Manual
RSA3303B, RSA3308B, & RSA3408B Option 26 1xEV-DO Analysis Software
071-2402-00
This document supports firmware version 4.0 and above.
www.tektronix.com
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200. Worldwide, visit www.tektronix.com to find contacts in your area.
Warranty 2
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and replacement products used by Tektronix for warranty work may be new or reconditioned to like new performance. All replaced parts, modules and products become the property of Tektronix.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

Table of Contents

Getting Started
Operating Basics
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
Related Manuals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Functional Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
1xEV-DO Forward Link Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Modulation Accuracy Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Code Domain Power Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Channel Power Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
OBW Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
ACPR Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
Spectrum Emission Mask Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
Gated Output Power Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-30
Intermodulation Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-33
CCDF Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-35
Pilot to Code Channel Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-38
1xEV-DO Reverse Link Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . 2-41
Modulation Accuracy Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-42
Code Domain Power Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-51
Spectrum Emission Mask Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-55
Pilot to Code Channel Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-57
Editing the Measurement Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-61
Using the Measurement Limits Editor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-61
Saving and Loading Measurement Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-70
Measurement Limit Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-71
Appendices
Setting Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Index
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual i

List of Figures

List of Figures
Figure 2-1: Menu diagram showing measurement functions available
in Option 26 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Figure 2-2: Modulation accuracy measurement . . . . . . . . . . . . . . . . . . . . 2-4
Figure 2-3: Modulation accuracy display . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Figure 2-4: EVM display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Figure 2-5: MagErr display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Figure 2-6: PhaseErr display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Figure 2-7: Symbol table display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Figure 2-8: Code domain power measurement . . . . . . . . . . . . . . . . . . . . 2-13
Figure 2-9: Code domain power display . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Figure 2-10: Power codogram display . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
Figure 2-11: IQ power graph display . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-21
Figure 2-12: Channel power measurement . . . . . . . . . . . . . . . . . . . . . . . 2-23
Figure 2-13: OBW measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
Figure 2-14: ACPR measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
Figure 2-15: Spectrum emission mask measurement . . . . . . . . . . . . . . . 2-26
Figure 2-16: Gated output power measurement . . . . . . . . . . . . . . . . . . . 2-30
Figure 2-17: Intermodulation measurement . . . . . . . . . . . . . . . . . . . . . . 2-33
Figure 2-18: CCDF measurement (single view) . . . . . . . . . . . . . . . . . . . . 2-35
Figure 2-19: Scale settings in the CCDF display . . . . . . . . . . . . . . . . . . . 2-37
Figure 2-20: Pilot to code channel measurement . . . . . . . . . . . . . . . . . . . 2-38
Figure 2-21: Modulation accuracy measurement-reverse link . . . . . . . . 2-42
Figure 2-22: Modulation accuracy display-reverse link . . . . . . . . . . . . . 2-46
Figure 2-23: EVM display-reverse link . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-47
Figure 2-24: MagErr display-reverse link . . . . . . . . . . . . . . . . . . . . . . . . 2-48
Figure 2-25: PhaseErr display-reverse link . . . . . . . . . . . . . . . . . . . . . . . 2-49
Figure 2-26: Symbol table display-reverse link . . . . . . . . . . . . . . . . . . . . 2-50
Figure 2-27: Code domain power measurement-reverse link . . . . . . . . . 2-51
Figure 2-28: Spectrum emission mask measurement-reverse link . . . . . 2-55
Figure 2-29: Pilot to code channel measurement-reverse link . . . . . . . . 2-57
Figure 2-30: Measurement limits editor . . . . . . . . . . . . . . . . . . . . . . . . . . 2-61
Figure 2-31: ACPR measurement limits editor . . . . . . . . . . . . . . . . . . . . 2-64
Figure 2-32: Spectrum emission mask measurement limits editor . . . . 2-66
Figure 2-33: Gated output power measurement limits editor . . . . . . . . 2-68
ii RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual

List of Tables

List of Tables
Table 2-1: Measurement limits items . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-62
Table 2-2: Measurement limit items for the ACPR measurement . . . . 2-65
Table 2-3: Measurement limit items for the spectrum emission mask
measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-66
Table 2-4: Measurement limit items for the gated output power
measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-68
Table 2-5: Forward link common limits . . . . . . . . . . . . . . . . . . . . . . . . . . 2-71
Table 2-6: Reverse link common limits . . . . . . . . . . . . . . . . . . . . . . . . . . 2-71
Table 2-7: Forward and reverse link: all band classes . . . . . . . . . . . . . . 2-72
Table 2-8: Forward link: BC (band class) 0, 2, 3, 5, and 9 . . . . . . . . . . . 2-72
Table 2-9: Forward link: BC (band class) 1, 4, and 8 . . . . . . . . . . . . . . . 2-73
Table 2-10: Forward link: BC (band class) 3b . . . . . . . . . . . . . . . . . . . . 2-73
Table 2-11: Forward link: BC (band class) 6 . . . . . . . . . . . . . . . . . . . . . . 2-74
Table 2-12: Forward link: BC (band class) 7 . . . . . . . . . . . . . . . . . . . . . . 2-74
Table 2-13: Reverse link: BC (band class) 0, 2, 3, 5, and 9 . . . . . . . . . . . 2-75
Table 2-14: Reverse link: BC (band class) 1, 4, and 8 . . . . . . . . . . . . . . . 2-75
Table 2-15: Reverse link: BC (band class) 6 . . . . . . . . . . . . . . . . . . . . . . 2-76
Table 2-16: Reverse link: BC (band class) 7 . . . . . . . . . . . . . . . . . . . . . . 2-76
Table 2-17: Forward and reverse link: BC (band class) 0, 1, 2, 3, 4, 5,
7, 8, and 9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-77
Table 2-18: Forward link: BC (band class) 6 . . . . . . . . . . . . . . . . . . . . . . 2-77
Table 2-19: Reverse link: BC (band class) 6 . . . . . . . . . . . . . . . . . . . . . . 2-78
Table 2-20: Forward link, all band classes, idle slot . . . . . . . . . . . . . . . . 2-78
Table 2-21: Forward link, all band classes, active slot . . . . . . . . . . . . . . 2-79
Table A-1: Display format and scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual iii
List of Tables
iv RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual

Preface

About This Manual

This manual provides operating instructions for the RSA3303B, RSA3308B, and RSA3408B Real-Time Spectrum Analyzer Option 26 1xEV-DO analysis software.
This manual supports the following instruments:
RSA3303B Option 26
RSA3308B Option 26
RSA3408B Option 26
This manual is composed of the following sections:
Getting Started provides a product description.
Operating Basics describes the measurement functions added by the option
and explains how to set up the analyzer for each measurement mode.
Appendices provides additional information about scale setting ranges.
The analyzer uses Microsoft Windows XP as the operating system. This manual does not describe common usage of Windows XP. Refer to your Windows manuals as necessary.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual v
Preface

Related Manuals

The following documents are also available for the analyzer:
RSA3303B & RSA3308B User Manual
(Standard accessory; Tektronix part number 071-2363-XX)
RSA3408B User Manual
(Standard accessory; Tektronix part number 071-2364-XX) Describes how to install the analyzer and how to work with the menus, and details the standard functions. Also provides the specifications.
RSA3000B Series Programmer Manual
(Standard accessory; PDF, Tektronix part number 071-2382-XX) Contains an alphabetical listing of the programming commands and other information related to controlling the analyzer over the GPIB interface.
PDF Manual
The RSA3000B Series Programmer Manual (PDF only) is included in the Documents CD (Tektronix part number 063-4089-XX).
vi RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
Getting Started

Getting Started

Product Description

This section provides an overview of the product capabilities.
The RSA3300B Series and RSA3408B Real-Time Spectrum Analyzer Option 26 is an analysis software option that adds transmitter measurements capability for 1xEV-DO forward link (3GPP2 C.S0032) and reverse link (3GPP2 C.S0033) to the analyzers.
You can perform the following 1xEV-DO measurements:
Modulation accuracy
Code domain power
Channel power
OBW (occupied bandwidth)
ACPR (Adjacent channel power ratio)
Spectrum emission mask
Gated output power (forward link only)
Intermodulation
CCDF
Pilot to code channel
In addition, the option provides predefined measurement limits for the ACPR, spectrum emission mask, and gated output power measurements. Using these measurement limits, you can easily check that the input signal meets the specification.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 1-1
Getting Started
1-2 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
Operating Basics

Functional Overview

This section provides an overview of the Option 26 measurement functions.
Figure 2-1 shows the measurement functions and how to access each measurement function from the front-panel menu key.
Other option name (if installed)
Mode
S/A
Demod
Time
Demod menu
Analog Demod
Digital Demod
Standard...
Standard menu
1xEV-DO-Fwd
1xEV-DO-Rev
Measurement functions added by Option 26
Modulation Accuracy
Code Domain Power
Channel Power
OBW
Measure menu
Measure menu
ACPR
Spectrum Emission Mask
Gated Output Power
Intermodulation
CCDF
Pilot to Code Channel
Modulation Accuracy
Code Domain Power
Channel Power
OBW
ACPR
Spectrum Emission Mask
Intermodulation
CCDF
Pilot to Code Channel
Figure 2-1: Menu diagram showing measurement functions available in Option 26
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-1
Functional Overview
Accessing a Measurement
Function
All of the measurement functions available in Option 26 can be selected from the Demod (demodulation) mode.
Perform the following procedure to access any of the measurement functions:
1. Press the Demod key to open the Demod menu.
2. Press the Standard... side key to open the Standard menu.
3. Press the 1xEV-DO Fwd or 1xEV-DO Rev side key to open the Measure
menu for the standard.
4. Press one of the side keys to select the measurement that you want to perform. If the desired measurement is not displayed on the current Measure menu, press the Go to page 2 (of 2) side key to open the next page.
5. If needed, set frequency, span, and amplitude of the input signal.
Refer to your instrument user manual for information on how to set frequency, span, and amplitude.
2-2 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual

1xEV-DO Forward Link Measurements

This section describes the functions and features of the 1xEV-DO forward link measurements. Each measurement description contains general information about the measurement, descriptions of the measurement displays and functions available through menu selections.
The information is divided into the following subsections:
Modulation accuracy measurement
Code domain power measurement
Channel power measurement
OBW measurement
ACPR measurement
Spectrum emission mask measurement
Gated output power measurement
Intermodulation measurement
CCDF measurement
Pilot to code channel measurement
NOTE. If you are not familiar with the operation of the analyzer, refer to your
instrument user manual before reading this section.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-3
1xEV-DO Forward Link Measurements

Modulation Accuracy Measurement

The modulation accuracy measurement measures Rho (ratio of the correlated power to the total power), EVM (Error Vector Magnitude), magnitude error, phase error, frequency error, and origin offset. Figure 2-2 shows an example of the modulation accuracy measurement.
Figure 2-2: Modulation accuracy measurement
2-4 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Forward Link Measurements
Display Elements
Setting the Acquisition and
Analysis Parameters
The following information is shown in the modulation accuracy measurement display:
Overview: This view can contain power versus time or spectrogram.
Subview: This view can contain spectrum, EVM, magnitude error, phase
error, IQ power graph, and constellation.
Main view: This view can contain modulation accuracy, EVM, magnitude
error, phase error, and symbol table.
The display contents in each view can be changed using the View Define menu. Refer to Changing the View Contents on page 2-7 for more information about the View Define me n u.
You can set the following timing parameters using the Acquisition/Analysis menu. Press the Acquisition/Analysis key to open the Acquisition/Analysis menu.
Acquisition (chips): Sets the acquisition length in chips. The range depends
on the acquisition memory size and the span setting.
Acquisition Length (s): Displays the acquisition length in seconds.
Acquisition History: Specifies the number of the block to display and
analyze. The latest block is number zero. Older blocks have larger negative numbers.
Analysis Interval: Sets the analysis interval in half slots. The range depends
on the acquisition length setting.
Analysis Offset: Sets the start point of the analysis with respect to the left end
of the acquisition length in half slots.
Spectrum Length: Displays the time length for FFT processing of the
spectrum displayed in the subview. This value is equivalent to one frame acquisition length.
Spectrum Offset: Sets the beginning of Spectrum Length with respect to the
left end of the acquisition length.
NOTE. Refer to your instrument user manual for detailed information about the
timing parameters.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-5
1xEV-DO Forward Link Measurements
Setting the Measurement
Parameters
You can set the following measurement parameters using the Meas Setup menu. Press the Meas Setup key to open the Meas Setup menu.
Analyze: Performs analysis for the time slots in the analysis range.
Measurement Level: Selects the measurement level for the measurement.
You can select Chip (chip level) or Symbol (symbol level). When Overall is selected in the Channel Type menu item, you cannot select Symbol.
Channel Type...: Select the channel type to be measured. When Symbol is
selected in the Measurement Level menu item, you can select MAC, Pilot, Data, or Preamble. When Chip is selected in the Measurement Level menu item, you can select Overall, MAC, Pilot, Data, or Preamble.
PN Offset: Sets the PN offset in units of 64 chips. You can set the value from
0 to 511.
Select HalfSlot: Sets the half slot for the measurement.
Select Code: Sets the code in the half slot for the measurement. This setting is
only available when Measurement Level is set to Symbol.
Active Channel Threshold: Sets the active channel threshold level (in dB
from the pilot) used for deciding whether a code channel is active or inactive. You can set the value from 100 dB to 0 dB.
Measurement Filter...: Selects the measurement filter to apply when
calculating EVM and other modulation accuracy results. You can select None, cdma2000, or cdma2000+EQ (equalizer).
IQ Swap: Sets whether to swap the I and Q data streams before demodulation.
Limits...: Sets the pass/fail limits for the modulation accuracy measurement.
When pressing this side key, the measurement limits editor appears. Refer to Editing the Measurement Limits on page 2-61 for more information.
2-6 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Forward Link Measurements
Changing the View
Contents
You can change the view contents in the overview, subview, and main view using the View Define menu. Press the View: Define key to display the View Define menu.
Show Views: Selects the view style on the screen. You can select Single or
Multi.
Overview Content...: Selects a view to display in the overview. You can
select Spectrogram or Waveform (power versus time).
Subview Content...: Selects a view to display in the subview:
Spectrum: Displays spectrum of the measured signal.
EVM: Displays changes of EVM (Error Vector Magnitude) over time.
MagErr: Displays changes of magnitude error over time.
PhaseErr: Displays changes of phase error over time.
IQ Power Graph: Displays changes of I and Q powers over time. Refer
to IQ Power Graph Display on page 2-21.
Constellation: Displays the signal as an I-Q constellation.
Mainview Content...: Selects a view to display in the main view. You can
select Modulation Accuracy, EVM, MagErr, PhaseErr, or Symbol Table. Refer to Scale and Format in the Main View on page 2-8 for more information about the views.
Menu Off: Hides the side menu. To display the menu again, press the MENU
side key or View: Define key.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-7
1xEV-DO Forward Link Measurements
Scale and Format in the
Main View
1: second half slot
This subsection describes the scale settings and display format in the main view.
Modulation Accuracy Display. When you select Mainview Content from the View Define menu and then select Modulation Accuracy from the Mainview Content submenu, an IQ rectangular graph and the measured value for Rho, peak code domain error, rms and peak EVM, magnitude error, phase error, frequency error, IQ origin offset, and Tau (timing error) are displayed (see Figure 2-3).The peak code domain error is displayed only when the Measurement Level menu item is set to Symbol, and Tau is displayed only when an external trigger signal is applied.
Half slot number
0: first half slot
Number of preamble chips
Number of data chips
IQ rectangular graph
(Vector display)
Figure 2-3: Modulation accuracy display
You can set the scale of the display using the Scale menu. Press the View: Scale/Lines key, and then press the View Scale... side key to open the menu.
Measurement Content...: Selects vector or constellation display.
Ve ct or : Selects vector display. A signal represented with phase and amplitude is displayed in rectangular (I and Q) coordinates. The red point indicates the symbol position on the measured signal, and the yellow trace indicates the locus of the signal between symbols.
Constellation: Selects constellation display. It is the same as the vector display, except that only symbols of the measured signal are indicated in red, and the locus between symbols is not shown. The cross marks indicate symbol
positions of an ideal signal.
NOTE. I and Q signals are normalized to prevent the scale from changing when
signal attenuation changes.
2-8 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Forward Link Measurements
EVM Display. When you select Mainview Content from the View Define menu and then select EVM from the Mainview Content submenu, changes of EVM (Error Vector Magnitude) over time are displayed for each chip or symbol (see Figure 2-4).
Symbol number
Horizontal Start
Code number and channel number
Vertical Scale
Vertical Start
Horizontal Scale
Figure 2-4: EVM display
You can set the scale of the display using the Scale menu. Press the View: Scale/Lines key, and then press the View Scale... side key to open the menu.
Auto Scale: Sets the start value and the scale of the vertical axis to display the
entire waveform.
Horizontal Scale: Sets the scale of the horizontal axis (number of chips or
symbols).
Horizontal Start: Sets the chip number or symbol number of the first (left)
value of the horizontal axis.
Vertical Scale: Sets the scale of the vertical axis.
Vertical Start: Sets the minimum (bottom) value of the vertical axis.
Full Scale: Sets the scale of vertical axis to the default full-scale value.
Measurement Content...: Selects the display content in the main view. You
can select EVM, MagErr, or PhaseErr. This selection can also be made from Mainview Content in the View Define menu.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-9
1xEV-DO Forward Link Measurements
MagErr Display. When you select Mainview Content from the View Define menu and then select MagErr from the Mainview Content submenu, changes of magnitude error over time are displayed for each chip or symbol (see Figure 2-5).
Symbol number
Horizontal Start
Code length and channel number
Vertical Scale
Horizontal Scale
Figure 2-5: MagErr display
You can set the scale of the display using the Scale menu. Press the View: Scale/Lines key, and then press the View Scale... side key to open the menu.
Auto Scale: Sets the start value and the scale of the vertical axis automatically
to display the entire waveform.
Horizontal Scale: Sets the scale of the horizontal axis (number of chips or
symbols).
Horizontal Start: Sets the chip number or symbol number of the first (left)
value of the horizontal axis.
Vertical Scale: Sets the scale of the vertical axis.
Vertical Offset: Sets the offset value of the vertical axis. You can set the value
from 200% to 200%.
Full Scale: Sets the scale of vertical axis to the default full-scale value.
Measurement Content...: Selects the display content in the main view. You
can select EVM, MagErr, or PhaseErr. This selection can also be made from Mainview Content in the View Define menu.
2-10 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Forward Link Measurements
PhaseErr Display. When you select Mainview Content from the View Define menu and then select PhaseErr from the Mainview Content submenu, changes of phase error over time are displayed for each chip or symbol (see Figure 2-6).
Symbol number
Horizontal Start
Code length and channel number
Vertical Scale
Horizontal Scale
Figure 2-6: PhaseErr display
You can set the scale of the display using the Scale menu. Press the View: Scale/Lines key, and then press the View Scale... side key to open the menu.
Auto Scale: Sets the start value and the scale of the vertical axis automatically
to display the entire waveform.
Horizontal Scale: Sets the scale of the horizontal axis (number of chips or
symbols).
Horizontal Start: Sets the chip number or symbol number of the first (left)
value of the horizontal axis.
Vertical Scale: Sets the scale of the vertical axis.
Vertical Offset: Sets the offset value of the vertical axis. You can set the value
from 450 to 450 degrees.
Full Scale: Sets the scale of vertical axis to the default full-scale value.
Measurement Content...: Selects the display content in the main view. You
can select EVM, MagErr, or PhaseErr. This selection can also be made from Mainview Content in the View Define menu.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-11
1xEV-DO Forward Link Measurements
Symbol Table Display. When you select Mainview Content from the View Define menu and then select Symbol Table from the Mainview Content submenu, the symbol table is displayed (see Figure 2-7). This display is only available when the Measurement Level menu item is set to Symbol.
Code length and channel number
Symbol number
Channel type
Figure 2-7: Symbol table display
You can set the radix for the display using the Scale menu. Press the View: Scale/Lines key, and then press the View Scale... side key to open the menu.
Radix...: Sets the radix for displaying the table. You can select Hex
(hexadecimal), Oct (octal), or Bin (binary).
2-12 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual

Code Domain Power Measurement

The code domain power measurement measures the distribution of signal power across the set of code channels, normalized to the total signal power. This measurement allows you to verify that each code channel is operating at its proper level. Figure 2-8 shows an example of the code domain power measurement.
1xEV-DO Forward Link Measurements
Figure 2-8: Code domain power measurement
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-13
1xEV-DO Forward Link Measurements
Display Elements
Setting the Acquisition and
Analysis Parameters
The following information is shown in the code domain power measurement display:
Overview: This view can contain power versus time or spectrogram.
Subview: This view can contain spectrum, EVM, magnitude error, phase
error, IQ power graph, and constellation.
Main view: This view can contain code domain power, power codogram, and
IQ power graph.
The display contents in each view can be changed using the View Define menu. Refer to Changing the View Contents on page 2-16 for more information about the View De fi n e m e n u .
You can set the following timing parameters using the Acquisition/Analysis menu. Press the Acquisition/Analysis key to open the Acquisition/Analysis menu.
Acquisition Length (chips): Sets the acquisition length in chips. The range
depends on the acquisition memory size and the span setting.
Acquisition Length (s): Displays the acquisition length in seconds.
Acquisition History: Specifies the number of the block to display and
analyze. The latest block is number zero. Older blocks have larger negative numbers.
Analysis Interval: Sets the analysis interval in half slots. The range depends
on the acquisition length setting.
Analysis Offset: Sets the start point of the analysis with respect to the left end
of the acquisition length in half slots.
Spectrum Length: Display the time length for FFT processing of the spectrum
displayed in the subview. This value is equivalent to one frame acquisition length.
Spectrum Offset: Sets the beginning of Spectrum Length with respect to the
left end of the acquisition length.
NOTE. Refer to your instrument user manual for detailed information about the
timing parameters.
2-14 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Forward Link Measurements
Setting the Measurement
Parameters
You can set the following measurement parameters using the Meas Setup menu. Press the Meas Setup key to open the Meas Setup menu.
Analyze: Performs analysis for the time slots in the analysis range.
Measurement Level: Selects the measurement level for the measurement.
You can select Chip (chip level) or Symbol (symbol level). This setting is only available when Mainview Content in the View Define menu is set to IQ Power Graph.
Channel Type...: Selects the channel type to be measured. You can select
MAC, Pilot, Data, or Preamble.
PN Offset: Sets the PN offset in units of 64 chips. You can set the value from
0 to 511.
Select HalfSlot: Sets the half slot for the measurement.
Select Code: Sets the code in the half slot for the measurement.
Active Channel Threshold: Sets the active channel threshold level (in dB
from the pilot) used for deciding whether a code channel is active or inactive. You can set the value from 100 dB to 0 dB.
Measurement Filter...: Selects the measurement filter to apply when
calculating code domain power results. You can select None, cdma2000, or cdma2000+EQ (equalizer).
IQ Swap: Sets whether to swap the I and Q data streams before demodulation.
Limits...: Sets the pass/fail limits for the code domain power measurement.
When pressing this side key, the measurement limits editor appears. Refer to Editing the Measurement Limits on page 2-61 for more information.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-15
1xEV-DO Forward Link Measurements
Changing the View
Contents
You can change the contents in the overview, subview, and main view using the View Define menu. Press the View: Define key to display the View Define menu.
Show Views: Selects the view style on the screen. You can select Single or
Multi.
Overview Content...: Selects a view to display in the overview. You can
select Spectrogram or Waveform (power versus time).
Subview Content...: Selects a view to display in the subview:
Spectrum: Displays spectrum of the input signal.
EVM: Displays changes of EVM (Error Vector Magnitude) over time.
MagErr: Displays changes of magnitude error over time.
PhaseErr: Displays changes of phase error over time.
IQ Power Graph: Displays changes I and Q powers over time. Refer to
IQ Power Graph Display on page 2-21.
Constellation: Displays the signal as an I-Q constellation.
Mainview Content...: Selects a view to display in the main view. You can
select Code Domain Power, Power Codogram, or IQ Power Graph. Refer to Scale and Format in the Main View on page 2-17 for more information about
the views.
Code Order...: Sets the way that code channels are arranged on the display.
You can select Hadamard or BitReverse.
Menu Off: Hides the side menu. To display the menu again, press the MENU
side key or View: Define key.
2-16 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Forward Link Measurements
Scale and Format in the
Main View
Code length and channel number
Marker readout
Vertical Stop
This subsection describes the scale settings and display format in the main view.
Code Domain Power Display. When you select Mainview Content from the View Define menu and then select Code Domain Power from the Mainview Content submenu, the code domain power for code channels is displayed (see Figure 2-9).
active, inactive, or pilot display
Channel type
Code domain power
Vertical Scale
Horizontal Scale
Horizontal Start
Figure 2-9: Code domain power display
In the code domain power display, the I and Q power bars are distinguished by using different colors as follows:
Blue: pilot channel
Red: active channel
Yellow: inactive channel
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-17
1xEV-DO Forward Link Measurements
You can set the scale of the display using the Scale menu. Press the View: Scale/Lines key, and then press the View Scale... side key to open the menu.
Auto Scale: Sets the start value and the scale of the vertical axis automatically
to display the entire waveform.
Horizontal Scale: Sets the scale of the horizontal axis. The range is dependent
on the Channel Type setting.
Horizontal Start: Sets the channel number of the first (left) value of the
horizontal axis.
Vertical Scale: Sets the scale of the vertical axis. You can set the value from
100
Vert ic al St op : Sets the maximum (top) value of the vertical axis. You can set
the value from 100 dB to 100 dB.
Full Scale: Sets the scale of the vertical axis to the default full-scale value.
Y Axis...: Selects whether to represent the vertical (amplitude) axis with
relative values or absolute values. When you select Relative, the vertical axis represents the power relative to the total power of all channels. When you select Absolute, the vertical axis represents the absolute power of each channel.
µdB to 100 dB.
Graph Number: Selects the number of graphs to be displayed in the screen.
You can select 1 or 2.
View Data...: Sets which channels to display when 1 is selected in the Graph
Number menu item. You can select I, Q, or IandQ.
2-18 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Forward Link Measurements
Power Codogram Display. When you select Mainview Content from the View Define menu and then select Power Codogram from the Mainview Content submenu, the code domain power is displayed in spectrogram (see Figure 2-10).
Code length and channel number
Marker readout
Color Scale
Vertical Si ze
Vertical Start
active, inactive, or pilot display
Channel type
Code domain power
Half slot number
Relative time from completion of data acquisition
Horizontal Scale
Horizontal Start
Figure 2-10: Power codogram display
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-19
1xEV-DO Forward Link Measurements
You can set the scale of the display using the Scale menu. Press the View: Scale/Lines key, and then press the View Scale... side key to open the menu.
Auto Scale: Sets the start value and the scale of the vertical axis to display the
entire waveform.
Horizontal Scale: Sets the scale of the horizontal axis. The range is dependent
on the Channel Type setting.
Horizontal Start: Sets the channel number of the first (left) value of the
horizontal axis.
Vertical Size: Sets the scale of the vertical axis.
Vertical Start: Sets the frame value at the bottom of the vertical axis.
Color Scale: Sets the scale (the value separating the minimum power value
from the maximum power value) of the color axis. The choices are 5, 10, 20, and 50 dB.
Color Stop: Sets the stop value of the color axis.
Full Scale: Sets the maximum value of the color axis to the reference level.
Y Axis...: Selects whether to represent the Y (color) axis with relative values
or absolute values. When you select Relative, the Y axis represents the power relative to the total power of all channels. When you select Absolute, the Y axis represents the absolute power of each channel.
Graph Number: Selects the number of graphs to be displayed in the screen.
You can select 1 or 2.
View Data...: Sets which channels to display when 1 is selected in the Graph
Number menu item. You can select I, Q, or IandQ.
2-20 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Forward Link Measurements
IQ Power Graph Display. When you select Mainview Content from the View Define menu and then select IQ Power Graph from the Mainview Content submenu, I and Q powers over time are displayed for each chip or symbol (see Figure 2-11).
Marker readout
Vertical Stop
Vertical Sc ale
Symbol number
Code length and channel number
Horizontal Scale
Horizontal Start
Figure 2-11: IQ power graph display
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-21
1xEV-DO Forward Link Measurements
You can set the scale of the display using the Scale menu. Press the View: Scale/Lines key, and then press the View Scale... side key to open the menu.
Auto Scale: Sets the start value and the scale of the vertical axis automatically
to display the entire waveform.
Horizontal Scale: Sets the scale of the horizontal axis (number of chips or
symbols).
Horizontal Start: Sets the start chip or symbol number of the horizontal axis.
Vertical Scale: Sets the scale of the vertical axis. You can set the value from
100
Vertical Stop: Sets the stop value of the vertical axis. You can set the value
from 100 dB to 100 dB.
Full Scale: Sets the scale of the vertical axis to the default full scale value.
Y Axis...: Selects whether to represent the vertical (amplitude) axis with
relative values or absolute values. When you select Relative, the vertical axis represents the power relative to the total power of all channels. When you select Absolute, the vertical axis represents the absolute power of each channel.
µdB to 100 dB.
Graph Number: Selects the number of graphs to be displayed in the screen.
You can select 1 or 2.
View Data...: Sets which channels to display when 1 is selected in the Graph
Number menu item. You can select I, Q, or IandQ.
2-22 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
Channel Power Measurement
The channel power measurement measures the channel power within a specified bandwidth and the power spectral density in dBm/Hz. Figure 2-12 shows an example of the channel power measurement.
1xEV-DO Forward Link Measurements
Figure 2-12: Channel power measurement
Setting the Measurement
Parameters
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-23
You can set the following measurement parameters using the Meas Setup menu. Press the Meas Setup key to open the Meas Setup menu.
Channel Bandwidth: Sets the frequency range for the measurement.
Measurement Filter Shape: Sets the filter to be used for the measurement.
You can select Rect (Rectangular), Gaussian, Nyquist, or Root Nyquist.
Rolloff Ratio: Enters the rolloff ratio when the Nyquist or Root Nyquist filter
is selected. You can set the value from 0.0001 to 1. The default value is 0.5.
Limits...: Sets the pass/fail limit for the channel power measurement. When
pressing this side key, the measurement limits editor appears. Refer to Editing the Measurement Limits on page 2-61 for more information.
NOTE. Refer to your instrument user manual for information about the scale
settings for the display.
1xEV-DO Forward Link Measurements

OBW Measurement

The OBW (Occupied Bandwidth) measurement measures the frequency bandwidth, using the ratio you specify for carrier signal power/power within the span setting. Figure 2-13 shows an example of the OBW measurement.
Figure 2-13: OBW measurement
Setting the Measurement
Parameters
2-24 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
You can set the following measurement parameters using the Meas Setup menu. Press the Meas Setup key to open the Meas Setup menu.
Power Ratio: Specifies the power ratio of the carrier and span regions for
calculating OBW. You can set the value from 80 to 99.99%. The default value is 99%.
Limits...: Sets the pass/fail limits for the OBW measurement. When you press
this side key, the measurement limits editor appears. Refer to Editing the Measurement Limits on page 2-61 for more information.
NOTE. Refer to your instrument user manual for information about the scale
settings for the display.

ACPR Measurement

1xEV-DO Forward Link Measurements
The ACPR (Adjacent Channel Power Ratio) measurement measures the ratio of carrier signal power to the signal power in channels specified by limits. Figure 2-14 shows an example of the ACPR measurement.
Figure 2-14: ACPR measurement
Setting the Measurement
Parameters
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-25
You can set the following measurement parameters using the Meas Setup menu. Press the Meas Setup key to open the Meas Setup menu.
Main Channel Bandwidth: Sets the bandwidth of the main channel.
Measurement Filter Shape...: Sets the filter to be used for the measurement.
You can select Rect (Rectangular), Gaussian, Nyquist, or Root Nyquist.
Scroll Table: Scrolls the table displayed in the view.
Limits: Sets the pass/fail limit for the ACPR measurement. When you press
this side key, the measurement limits editor appears. Refer to Editing the Measurement Limits on page 2-61 for more information.
NOTE. Refer to your instrument user manual for information about the scale
settings for the display.
1xEV-DO Forward Link Measurements

Spectrum Emission Mask Measurement

The spectrum emission mask measurement verifies that the base station is not transmitting excessive power outside of its designated channel. Since the Option 26 provides the predefined measurement limits for each band class, you can perform pass/fail evaluations easily. Figure 2-15 shows an example of the spectrum emission mask measurement.
Figure 2-15: Spectrum emission mask measurement
NOTE. When performing this measurement, be sure to apply a continuous active
slot signal or a continuous idle slot signal.
2-26 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Forward Link Measurements
Setting the Measurement
Parameters
You can set the following measurement parameters using the Meas Setup menu. Press the Meas Setup key to open the Meas Setup menu.
Slot Type: Selects the slot type to be measured. You can select Active or Idle.
For Active, the total power of the Pilot, MAC, and Data channels is measured. For Idle, the burst power of the Pilot and MAC channels is measured.
The displayed menu items change depending on which slot type is selected.
When Active is selected:
Measurement...: Sets the limit table used for limit testing. You can select
Offset From Channel or Inband Spurious. For Offset From Channel, frequency zones are specified by the difference from the center frequency. For Inband Spurious, frequency zones are specified by the absolute frequency values.
Ref. Channel Select: Selects the reference mode used to measure the spectrum
emission level. You can select Auto or Man (manual). For Auto, the reference channel level is measured from the input signal for the specified channel bandwidth. For Man, you can set the reference channel level in the Ref. Channel Level menu item.
Ref. Channel Level: Sets the reference channel level used to measure the
spectrum emission level. You can set the value from 150 dBm to 30 dBm. This setting is only available when Ref. Channel Select is set to Man (manual).
Channel Bandwidth: Sets the channel bandwidth for the measurement.
Measurement Filter Shape...: Sets the filter to be used for the measurement.
You can select Rect (Rectangular), Gaussian, Nyquist, or Root Nyquist. This setting is only available when Ref. Channel Select is set to Auto.
Rolloff Ratio: Enters the rolloff ratio when the Nyquist or Root Nyquist filter
is selected. You can set the value from 0.0001 to 1. The default value is 0.5. This setting is only available when Ref. Channel Select is set to Auto.
Limits...: Sets the pass/fail limits for the spectrum emission mask
measurement. When you press this side key, the measurement limits editor appears. Refer to Editing the Measurement Limits on page 2-61 for more information.
Scroll Table: Scrolls the table displayed in the view.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-27
1xEV-DO Forward Link Measurements
When Idle is selected:
NOTE. In this measurement, you cannot edit a frequency mask for trigger.
Measurement...: Sets the limit table used for limit testing. You can select
Offset From Channel or Inband Spurious. For Offset From Channel, frequency zones are specified by the difference from the center frequency. For Inband Spurious, frequency zones are specified by the absolute frequency values.
Gate Length: Sets the gate length to be measured. You can set the value from
180 µs to 840 µs.
Burst Sync...: Sets the burst point to be measured. You can select Rising Edge,
Mid Point, or Trigger Position. This setting is only available when Slot Type is set to Idle.
Ref. Channel Select: Selects the reference mode used to measure the spectrum
emission level. You can select Auto or Man (manual). For Auto, the reference channel level is measured from the input signal for the specified channel bandwidth. For Man, you can set the reference level in the Ref. Channel Level menu item.
Ref. Channel Level: Sets the reference channel level used to measure the
spectrum emission level. You can set the value from 150 dBm to 30 dBm. This setting is only available when Ref. Channel Select is set to Man (manual).
Analyze: Performs analysis for the time slots in the analysis range.
Limits...: Sets the pass/fail limits for the spectrum emission mask
measurement. When you press this side key, the measurement limits editor appears. Refer to Editing the Measurement Limits on page 2-61 for more information.
Scroll Table: Scrolls the table displayed in the view.
NOTE. Refer to your instrument user manual for information about the scale
settings for the display.
2-28 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Forward Link Measurements
Setting the Acquisition and
Analysis Parameters
Display Elements
Changing the View
Contents
When Idle is selected as the slot type, you can set the following timing parameter using the Acquisition/Analysis menu. Press the Acquisition/Analysis key to open the Acquisition/Analysis menu.
Acquisition History: Specifies the block number to be analyzed. Block
number “0” corresponds to the latest block.
When Idle is selected as the slot type, the following information is shown in the spectrum emission measurement display:
Overview: This view can contain power versus time or spectrogram.
Main view: This view can contain spectrum.
The display content in the overview can be changed using the View Define menu. Refer to the next subsection for more information about the View Define menu.
When Idle is selected as the slot type, you can change the contents in the overview using the View Define menu. Press the View: Define key to display the View Define menu.
Show Views: Sets the view style on the screen. You can select Single or Multi.
Overview Content...: Selects a view to display in the overview. You can
select Spectrogram or Waveform (spectrum).
Menu Off: Hides the side menu. To display the menu again, press the MENU
side key or View: Define key.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-29
1xEV-DO Forward Link Measurements

Gated Output Power Measurement

The Gated Output power measurement measures the total RF power in a selected channel. Figure 2-16 shows an example of the gated output power measurement.
Display Elements
Figure 2-16: Gated output power measurement
The following information is shown in the gated output power measurement display:
Overview: This view can contain power versus time or spectrogram.
Subview: This view can contain spectrum.
Main view: This view can contain power versus time.
The display content in the overview can be changed using the View Define menu. Refer to Changing the View Contents on page 2-32 for more information about the View De fi n e m e n u .
2-30 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Forward Link Measurements
Setting the Acquisition and
Analysis Parameters
Setting the Measurement
Parameters
You can set the following timing parameter using the Acquisition/Analysis menu. Press the Acquisition/Analysis key to open the Acquisition/Analysis menu.
Acquisition History: Specifies the block number to be analyzed. Block
number “0” corresponds to the latest block.
You can set the following measurement parameters using the Meas Setup menu. Press the Meas Setup key to open the Meas Setup menu.
In the Meas Setup menu, you can set the following parameters:
Analyze: Performs analysis for the time slots in the analysis range.
Slot Type: Selects the slot type to be measured. You can select Active or Idle.
For Active, the total power of the Pilot, MAC, and Data channels is measured. For Idle, the burst power of the Pilot and MAC channels is measured.
NOTE. When you select Idle, the trigger souse need to be set to Level. For
information about setting the trigger souse, refer to your instrument user manual.
Burst Sync...: Sets the burst point to be measured. You can select Rising Edge,
Mid Point, or Trigger Position. This setting is only available when Idle is selected as the slot type.
Burst Offset: Sets the burst offset between the trigger position and the burst
position for the measurement. You can set the value from 1 ms to 1 ms. This setting is only available when Burst Sync is set to Trigger Position.
Ref. Channel Select: Selects the reference mode used to measure the power
level. You can select Auto or Man (manual). For Auto, the reference channel level is measured from the input signal. For Man, you can set the reference channel level in the Ref. Channel Level menu item.
Ref. Channel Level: Sets the reference channel level used to measure the
power level. You can set the value from 150 dBm to 30 dBm. This setting is only available when Ref.Channel Select is set to Man (manual).
Limits...: Sets the pass/fail limits for the gated output power measurement.
When you press this side key, the measurement limits editor appears. Refer to Editing the Measurement Limits on page 2-61 for more information.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-31
1xEV-DO Forward Link Measurements
Changing the View
Contents
Scale in the Main View
You can change the contents in the overview using the View Define menu. Press the View: Define key to display the View Define menu.
Show Views: Sets the view style on the screen. You can select Single or Multi.
Overview Content...: Selects a view to display in the overview. You can
select Spectrogram or Waveform (power versus time).
Menu Off: Hides the side menu. To display the menu again, press the MENU
side key or View: Define key.
You can set the scale of the main view using the Scale menu. Press the View: Scale/Lines key, and then press the View Scale... side key to open the menu.
Auto Scale: Sets the start value and the scale of the vertical axis to display the
entire waveform.
Horizontal Scale: Sets the scale of the horizontal axis.
Horizontal Start: Sets the minimum (left) value of the horizontal axis.
Vertical Scale: Sets the scale of the vertical axis.
Vertical Stop: Sets the maximum (top) value of the vertical axis.
Full Scale: Sets the scale of the vertical axis to the default full scale value.
2-32 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual

Intermodulation Measurement

The intermodulation measurement measures the third-order and fifth-order harmonic distortion components of a modulated signal. Figure 2-17 shows an example of the intermodulation measurement.
1xEV-DO Forward Link Measurements
Figure 2-17: Intermodulation measurement
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-33
1xEV-DO Forward Link Measurements
Setting the Measurement
Parameters
You can set the following measurement parameters using the Meas Setup menu. Press the Meas Setup key to open the Meas Setup menu.
2nd Channel Offset: Sets the center frequency of the second channel where
the interfering carrier exists.
Main Channel Integration BW: Sets the main channel integration
bandwidth. You can set the value from 500 kHz to 5 MHz.
Measurement Filter Shape...: Sets the filter to be used for the measurement.
You can select Rect (Rectangular), Gaussian, Nyquist, or Root Nyquist.
Rolloff Ratio: Enters the rolloff ratio when the Nyquist or Root Nyquist filter
is selected. You can set the value from 0.0001 to 1. The default value is 0.5.
Limits...: Sets the pass/fail limits for the intermodulation measurement. When
you press this side key, the measurement limits editor appears. Refer to Editing the Measurement Limits on page 2-61 for more information.
NOTE. Refer to your instrument user manual for information about the scale
settings for the display.
2-34 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual

CCDF Measurement

1xEV-DO Forward Link Measurements
The CCDF (Complementary Cumulative Distribution Function) measurement displays the peak-to-average power ratio along the horizontal axis, and the probability that the ratio is exceeded along the vertical axis. This display is useful in designing digital communications systems. Figure 2-18 shows an example of the CCDF measurement.
Figure 2-18: CCDF measurement (single view)
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-35
1xEV-DO Forward Link Measurements
Setting the Acquisition and
Analysis Parameters
Setting the Measurement
Parameters
You can set the following timing parameters using the Acquisition/Analysis menu. Press the Acquisition/Analysis key to open the Acquisition/Analysis menu.
Acquisition Length (s): Sets the acquisition length in seconds.
Acquisition History: Specifies the number of the block to display and
analyze. The latest block is number zero. Older blocks have larger negative numbers.
Spectrum Length: Display the time length for FFT processing of the spectrum
displayed in the subview. This value is equivalent to one frame acquisition length.
Spectrum Offset: Sets the beginning of Spectrum Length with respect to the
left end of the acquisition length.
Output Trigger Indicator: Selects whether to enable the output trigger
indicator.
NOTE. Refer to your instrument user manual for detailed information about the
time parameters.
You can set the following measurement parameters using the Meas Setup menu. Press the Meas Setup key to open the Meas Setup menu.
Display Gaussian Line: Selects whether to display the Gaussian line on the
graticule.
Display Reference Line: Selects whether to display the most recently stored
reference line on the graticule.
Store Reference Line: Stores a new reference line.
Reset Measurement: Restart the CCDF measurement. When you press this
key, the CCDF accumulation counter on the upper right side of the display is reset to 0.
CCDF Auto-Scaling: Selects whether to automatically set the scaling of the
horizontal axis. When you set it to On, the signal peak value is displayed as the maximum value on the right edge of the graticule. When you set it to Off, the scaling of the horizontal axis can be set manually with CCDF Scale. The default value is Off.
CCDF Scale: Sets the full-scale value of the horizontal axis of the CCDF
display when CCDF Auto-Scaling is set to Off. You can set the value from 1dB to 100 dB.
Threshold: Sets the threshold which defines the samples to be included in the
CCDF calculation. You can set the value from 250 dBm to 130 dBm.
Limits...: When you press this side key, the measurement limits editor appears.
Refer to Editing the Measurement Limits on page 2-61 for more information.
2-36 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Forward Link Measurements
Changing the View
Contents
Scale in the Main View
You can change the contents in the overview using the View Define menu. Press the View: Define key to display the View Define menu.
Show Views: Sets the view style on the screen. You can select Single or Multi.
When you set it to Multi, the CCDF, spectrum, and power versus time of the measured signal are displayed simultaneously.
Overview Content...: Selects a view to display in the overview. You can
select Spectrogram or Waveform (power versus time).
Menu Off: Hides the side menu. To display the menu again, press the MENU
side key or View: Define key.
You can set the scale in the main view using the Scale menu. Press the View: Scale/Lines key, and then press the View Scale... side key to open the menu.
Auto Scale: Sets the start value and the scale of the vertical axis automatically
to display the entire waveform.
Horizontal Scale: Sets the scale of the horizontal axis.
Horizontal Start: Sets the minimum (left) value of the horizontal axis.
Vertical Stop: Sets the maximum (top) the vertical axis. The setting ranges are
twice the Vertical Start value to 100% in a 1-2-5 sequence.
Vertical Stop
Vertical Start
Vertical Start: Sets the minimum (bottom) value of the vertical axis. The
setting ranges are 10
Full Scale: Sets the scale of the vertical axis to the default full scale value.
Sub Grid: Sets whether to display the sub-grid on the graticule.
Horizontal Start
-5
to half the Vertical Stop value in a 1-2-5 sequence.
Horizontal Scale
Figure 2-19: Scale settings in the CCDF display
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-37
1xEV-DO Forward Link Measurements

Pilot to Code Channel Measurement

The pilot to code channel measurement measures the time difference between the pilot channel and other code channels detected in the received signal. Figure 2-20 shows an example of the pilot to code channel measurement.
Figure 2-20: Pilot to code channel measurement
2-38 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Forward Link Measurements
Display Elements
Setting the Acquisition and
Analysis Parameters
The following information is shown in the pilot to code channel measurement display:
Overview: This view can contain power versus time or spectrogram.
Subview: This view can contain spectrum, EVM, magnitude error, phase
error, constellation.
Main view: This view lists the following measurement results for the active
channels:
Power level Timing referred to the pilot channel Phase referred to the pilot channel Code domain error
The display content in the subview can be changed using the View Define menu. Refer to Changing the View Contents on page 2-40 for more information about the View Define me n u.
You can set the following timing parameters using the Acquisition/Analysis menu. Press the Acquisition/Analysis key to open the Acquisition/Analysis menu.
Acquisition Length (chips): Sets the acquisition length in chips. The range
depends on the acquisition memory size and the span setting.
Acquisition Length (s): Displays the acquisition length in seconds.
Acquisition History: Specifies the number of the block to display and
analyze. The latest block is number zero. Older blocks have larger negative numbers.
Analysis Interval: Sets the analysis interval in half slots. The range depends
on the acquisition length setting.
Analysis Offset: Sets the start point of the analysis with respect to the left end
of the acquisition length in half slots.
Spectrum Length: Display the time length for FFT processing of the spectrum
displayed in the subview. This value is equivalent to one frame acquisition length.
Spectrum Offset: Sets the beginning of Spectrum Length with respect to the
left end of the acquisition length.
NOTE. Refer to your instrument user manual for detailed information about the
time parameters.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-39
1xEV-DO Forward Link Measurements
Setting the Measurement
Parameters
You can set the following measurement parameters using the Meas Setup menu. Press the Meas Setup key to open the Meas Setup menu.
Analyze: Performs analysis for the time slots in the analysis range.
Channel Type...: Select the channel type to be measured. You can select
MAC, Data, or Preamble.
PN Offset: Sets the PN offset in units of 64 chips. You can set the value from
0 to 511.
Select HalfSlot: Sets the half slot for the measurement.
Scroll Table: Scrolls the table displayed in the view.
Active Channel Threshold: Sets the active channel threshold level (in dB
from the pilot) used for deciding whether a code channel is active or inactive. You can set the value from 100 dB to 0 dB.
Measurement Filter...: Selects the measurement filter to apply when
calculating pilot to code channel results. You can select None, cdma2000, or cdma2000+EQ (equalizer).
IQ Swap: Sets whether to swap the I and Q data streams before demodulation.
Limits...: Sets the pass/fail limits for the pilot to code channel measurement.
When you press this side key, the measurement limits editor appears. Refer to Editing the Measurement Limits on page 2-61 for more information.
Changing the View
Contents
You can change the contents in the overview and subview using the View Define menu. Press the Define key to display the View Define menu.
Overview Content...: Selects a view to display in the overview. You can
select Spectrogram or Waveform (power versus time).
Subview Content...: Selects a view to display in the subview:
Spectrum: Displays spectrum of the input signal.
EVM: Displays changes of EVM (Error Vector Magnitude) in symbols.
MagErr: Displays changes of magnitude error in symbols.
PhaseErr: Displays changes of phase error in symbols.
Constellation: Displays the signal as an I-Q constellation.
2-40 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual

1xEV-DO Reverse Link Measurements

This section describes the functions and features of the 1xEV-DO reverse link measurements. Each measurement description contains general information about the measurement, descriptions of the measurement displays and functions available through menu selections.
The information is divided into the following subsections:
Modulation accuracy measurement
Code domain power measurement
Spectrum emission mask measurement
Pilot to code Channel measurement
The following measurement items available in the 1xEVDO reverse link measurements are the same as those in the 1xEV-DO forward link measurements. Refer to 1xEV-DO Forward Link Measurements beginning on page 2-3.
Channel power measurement
OBW measurement
ACPR measurement
Intermodulation measurement
CCDF measurement
NOTE. If you are not familiar with the operation of the analyzer, refer to your
instrument user manual before reading this section.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-41
1xEV-DO Reverse Link Measurements

Modulation Accuracy Measurement

The modulation accuracy measurement measures Rho (ration of the correlated power to the total power), EVM (Error Vector Magnitude), magnitude error, phase error, frequency error, and origin offset. Figure 2-21 shows an example of the modulation accuracy measurement.
Figure 2-21: Modulation accuracy measurement-reverse link
2-42 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Reverse Link Measurements
Display Elements
Setting Acquisition and
Analysis Parameters
The following information is shown in the modulation accuracy measurement display:
Overview: This view can contain power versus time or spectrogram.
Subview: This view can contain spectrum, EVM, magnitude error, phase
error, IQ power graph, and constellation.
Main view: This view can contain modulation accuracy, EVM, magnitude
error, phase error, and symbol table.
The display contents in each view can be changed using the View Define menu. Refer to Changing the View Contents on page 2-45 for more information about the View Define me n u.
You can set the following acquisition and analysis parameters using the Acquisition/Analysis menu. Press the Acquisition/Analysis key to open the menu.
Acquisition Length (chips): Sets the acquisition length in chips. The setting
range depends on the acquisition memory size and the span setting.
Acquisition Length (s): Displays the acquisition length in seconds.
Acquisition History: Specifies the number of the block to display and
analyze. The latest block is number zero. Older blocks have larger negative numbers.
Analysis Interval: Sets the analysis interval in half slots. The setting range
depends on the acquisition length setting.
Analysis Offset: Sets the start point of the analysis with respect to the left end
of the acquisition length in half slots.
Spectrum Length: Display the time length for FFT processing of the spectrum
displayed in the subview. This value is equivalent to one frame acquisition length.
Spectrum Offset: Sets the beginning of Spectrum Length with respect to the
left end of the acquisition length.
NOTE. Refer to your instrument user manual for detailed information about the
acquisition and analysis parameters.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-43
1xEV-DO Reverse Link Measurements
Setting Measurement
Parameters
You can set the following measurement parameters using the Meas Setup menu. Press the Meas Setup key to open the Meas Setup menu.
Analyze: Performs analysis for the time slots in the analysis range.
Measurement Level: Selects the measurement level for the measurement.
You can select Chip (chip level) or Symbol (symbol level).
Select HalfSlot: Sets the half slot for the measurement.
Select Code: Sets the code in the half slot for the measurement. This setting is
only available when Measurement Level is set to Symbol.
Longcode Mask I: Sets the long code mask value for the I signal. You can set
the value from 0 to 3FFFFFFFFFF in the hexadecimal format.
Longcode Mask Q: Sets the long code mask value for the Q signal. You can
set the value from 0 to 3FFFFFFFFFF in the hexadecimal format.
Active Channel Threshold: Sets the active channel threshold level (in dB
from the pilot) used for deciding whether a code channel is active or inactive. You can set the value from 100 dB to 0 dB.
Measurement Filter...: Selects the measurement filter to apply when
calculating EVM and other modulation accuracy results. You can select None, cdma2000, or cdma2000+EQ (equalizer).
IQ Swap: Sets whether to swap the I and Q data streams before demodulation.
Limits...: Sets the pass/fail limits for the modulation accuracy measurement.
When you press this side key, the measurement limits editor appears. Refer to Editing the Measurement Limits on page 2-61 for more information.
2-44 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Reverse Link Measurements
Changing the View
Contents
You can change the view contents in the overview, subview, and main view using the View Define menu. Press the View: Define key to display the View Define menu.
Show Views: Selects the view style on the screen. You can select Single or
Multi.
Overview Content...: Selects a view to display in the overview. You can
select Spectrogram or Waveform (power versus time).
Subview Content...: Selects a view to display in the subview:
Spectrum: Displays spectrum of the measured signal.
EVM: Displays changes of EVM (Error Vector Magnitude) over time.
MagErr: Displays changes of magnitude error over time.
PhaseErr: Displays changes of phase error over time.
IQ Power Graph: Displays changes of I and Q powers over time. Refer
to IQ Power Graph Display on page 2-21.
Constellation: Displays the signal as an I-Q constellation.
Mainview Content...: Selects a view to display in the main view. You can
select Modulation Accuracy, EVM, MagErr, PhaseErr, or Symbol Table. Refer to Scale and Format in the Main View on page 2-46 for more information about the views.
Menu Off: Hides the side menu. To display the menu again, press the MENU
side key or View: Define key.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-45
1xEV-DO Reverse Link Measurements
Scale and Format in the
Main View
1: second half slot
ACK or NACK indicator
This subsection describes the scale settings and display format in the main view.
Modulation Accuracy Display. When you select Mainview Content from the View Define menu and then select Modulation Accuracy from the Mainview Content submenu, an IQ rectangular graph and the measured value for Rho, peak code domain error, rms and peak EVM, magnitude error, phase error, frequency error, and IQ origin offset (see Figure 2-22). The peak code domain error is displayed only when the Measurement Level menu item is set to Symbol.
Half slot number
0: first half slot
DRC code value
IQ rectangular graph
(vector display)
DRC value
Figure 2-22: Modulation accuracy display-reverse link
You can set the scale of the display using the Scale menu. Press the View: Scale/Lines key, and then press View Scale... side key to open the menu.
Measurement Content...: Selects vector or constellation display.
Ve ct or : Selects vector display. A signal represented with phase and amplitude is displayed in rectangular (I and Q) coordinates. The red point indicates the symbol position on the measured signal, and the yellow trace indicates the locus of signal between symbols.
Constellation: Selects constellation display. It is the same as the vector display, except that only symbols of the measured signal are indicated in red, and the locus between symbols is not shown. The cross marks indicate symbol
positions of an ideal signal.
NOTE. I and Q signals are normalized to prevent the scale from changing when
signal attenuation changes.
2-46 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Reverse Link Measurements
EVM Display. When you select Mainview Content from the View Define menu and then select EVM from the Mainview Content submenu, changes of EVM (Error Vector Magnitude) over time are displayed for each chip or symbol (see Figure 2-23).
Chip number
Vertical Scale
Vertical Start
Horizontal Scale
Horizontal Start
Figure 2-23: EVM display-reverse link
You can set the scale of the display using the Scale menu. Press the View: Scale/Lines key, and then press View Scale... side key to open the menu.
Auto Scale: Sets the start value and the scale of the vertical axis to display the
entire waveform.
Horizontal Scale: Sets the scale of the horizontal axis (number of chips or
symbols).
Horizontal Start: Sets the chip number or symbol number of the first (left)
value of the horizontal axis.
Vertical Scale: Sets the scale of the vertical axis.
Vertical Start: Sets the minimum (bottom) value of the vertical axis.
Full Scale: Sets the scale of vertical axis to the default full-scale value.
Measurement Content...: Selects the display content in the main view. You
can select EVM, MagError, or PhaseError. This selection can also be made from Mainview Content in the View Define menu.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-47
1xEV-DO Reverse Link Measurements
MagErr Display. When you select Mainview Content from the View Define menu and then select MagErr from the Mainview Content submenu, changes of magnitude error over time are displayed for each chip or symbol (see Figure 2-24).
Chip number
Vertical Scale
Horizontal Scale
Horizontal Start
Figure 2-24: MagErr display-reverse link
You can set the scale of the display using the Scale menu. Press the View: Scale/Lines key, and then press View Scale... side key to open the menu.
Auto Scale: Sets the start value and the scale of the vertical axis automatically
to display the entire waveform.
Horizontal Scale: Sets the scale of the horizontal axis (number of chips or
symbols).
Horizontal Start: Sets the chip number or symbol number of the first (left)
value of the horizontal axis.
Vertical Scale: Sets the scale of the vertical axis.
Vertical Offset: Sets the offset value of the vertical axis. You can set the value
from 200% to 200%.
Full Scale: Sets the scale of vertical axis to the default full-scale value.
Measurement Content...: Selects the display content in the main view. You
can select EVM, MagError, or PhaseError. This selection can also be made from Mainview Content in the View Define menu.
2-48 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Reverse Link Measurements
PhaseErr Display. When you select Mainview Content from the View Define menu and then select PhaseErr from the Mainview Content submenu, changes of phase error over time are displayed for each chip or symbol (see Figure 2-25).
Chip number
Vertical Scale
Horizontal Scale
Horizontal Start
Figure 2-25: PhaseErr display-reverse link
You can set the scale of the display using the Scale menu. Press the View: Scale/Lines key, and then press View Scale... side key to open the menu.
Auto Scale: Sets the start value and the scale of the vertical axis to display the
entire waveform.
Horizontal Scale: Sets the scale of the horizontal axis (number of chips or
symbols).
Horizontal Start: Sets the chip number or symbol number of the first (left)
value of the horizontal axis.
Vertical Scale: Sets the scale of the vertical axis.
Vertical Offset: Sets the offset value of the vertical axis. You can set the value
from 450 to 450 degrees.
Full Scale: Sets the scale of vertical axis to the default full-scale value.
Measurement Content...: Selects the display content in the main view. You
can select EVM, MagError, or PhaseError. This selection can also be made from Mainview Content in the View Define menu.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-49
1xEV-DO Reverse Link Measurements
Symbol Table Display. When you select Mainview Content from the View Define menu and then select Symbol Table from the Mainview Content submenu, the symbol table is displayed (see Figure 2-26). This display is only available when the Measurement Level menu item is set to Symbol.
Symbol number
Code length and Channel number
Figure 2-26: Symbol table display-reverse link
You can set the radix for the display using the Scale menu. Press the View: Scale/Lines key, and then press View Scale... side key to open the menu.
Radix...: Sets the radix for displaying the table. You can select Hex
(hexadecimal), Oct (octal), or Bin (binary).
2-50 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual

Code Domain Power Measurement

The code domain power measurement measures the distribution of signal power across the set of code channels, normalized to the total signal power. This measurement allows you to verify that each code channel is operating at its proper level. Figure 2-27 shows an example of the code domain power measurement.
1xEV-DO Reverse Link Measurements
Figure 2-27: Code domain power measurement-reverse link
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-51
1xEV-DO Reverse Link Measurements
Display Elements
Setting Acquisition and
Analysis Parameters
The following information is shown in the code domain power measurement display:
Overview: This view can contain power versus time or spectrogram.
Subview: This view can contain spectrum, EVM, magnitude error, phase
error, IQ power graph, and constellation.
Main view: This view can contain code domain power, power codogram, and
IQ power graph.
The display contents in each view can be changed using the View Define menu. Refer to Changing the View Contents on page 2-54 for more information about the View De fi n e m e n u .
You can set the following acquisition and analysis parameters using the Acquisition/Analysis menu. Press the Acquisition/Analysis key to open the menu.
Acquisition Length (chips): Sets the acquisition length in chips. The setting
range depends on the acquisition memory size and the span setting.
Acquisition Length (s): Displays the acquisition length in seconds.
Acquisition History: Specifies the number of the block to display and
analyze. The latest block is number zero. Older blocks have larger negative numbers.
Analysis Interval: Sets the analysis interval in half slots. The setting range
depends on the acquisition length setting.
Analysis Offset: Sets the start point of the analysis with respect to the left end
of the acquisition length in half slots.
Spectrum Length: Displays the time length for FFT processing of the
spectrum displayed in the subview. This value is equivalent to one frame acquisition length.
Spectrum Offset: Sets the beginning of Spectrum Length with respect to the
left end of the acquisition length.
NOTE. Refer to your instrument user manual for detailed information about the
acquisition and analysis parameters.
2-52 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Reverse Link Measurements
Setting Measurement
Parameters
You can set the following measurement parameters using the Meas Setup menu. Press the Meas Setup key to open the Meas Setup menu.
Analyze: Performs analysis for the time slots in the analysis range.
Measurement Level: Selects the measurement level for the measurement.
You can select Chip (chip level) or Symbol (symbol level). This setting is only available when Mainview Content in the View Define menu is set to IQ Power Graph.
Select HalfSlot: Sets the half slot for the measurement.
Select Code: Sets the code in the half slot for the measurement.
Longcode Mask I: Sets the long code mask value for the I signal. You can set
the value from 0 to 3FFFFFFFFFF in the hexadecimal format.
Longcode Mask Q: Sets the long code mask value for the Q signal. You can
set the value from 0 to 3FFFFFFFFFF in the hexadecimal format.
Active Channel Threshold: Sets the active channel threshold level (in dB
from the pilot) used for deciding whether a code channel is active or inactive. You can set the value from 100 dB to 0 dB
Measurement Filter...: Selects the measurement filter to apply when
calculating code domain power results. You can select None, cdma2000, or cdma2000+EQ (equalizer).
IQ Swap: Sets whether to swap the I and Q data streams before demodulation.
Limits...: Sets the pass/fail limits for the code domain power measurement.
When you press this side key, the measurement limits editor appears. Refer to Editing the Measurement Limits on page 2-61 for more information.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-53
1xEV-DO Reverse Link Measurements
Changing the View
Contents
You can change the contents in the overview, subview, and main view using the View Define menu. Press the View: Define key to display the View Define menu.
Show Views: Selects the view style on the screen. You can select Single or
Multi.
Overview Content...: Selects a view to display in the overview. You can
select Spectrogram or Waveform (power versus time).
Subview Content...: Selects a view to display in the subview:
Spectrum: Displays spectrum of the input signal.
EVM: Displays changes of EVM (Error Vector Magnitude) over time.
MagErr: Displays changes of magnitude error over time.
PhaseErr: Displays changes of phase error over time.
IQ Power Graph: Displays the code domain power for each symbol.
Refer to IQ Power Graph Display on page 2-21.
Constellation: Displays the signal as an I-Q constellation.
Mainview Content...: Selects a view to display in the main view. You can
select Code Domain Power, Power Codogram, or IQ Power Graph. Refer to Scale and Format in the Main View on page 2-17 for more information about
the views.
Code Order...: Sets the way that code channels are arranged on the display.
You can select Hadamard or BitReverse.
Menu Off: Hides the side menu. To display the menu again, press the MENU
side key or View: Define key.
2-54 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual

Spectrum Emission Mask Measurement

The spectrum emission mask measurement verifies that the base station is not transmitting excessive power outside of its designated channel. Because Option 26 provides predefined measurement limits for each band class, you can perform pass/fail evaluations easily. Figure 2-28 shows an example of the spectrum emission mask measurement.
1xEV-DO Reverse Link Measurements
Figure 2-28: Spectrum emission mask measurement-reverse link
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-55
1xEV-DO Reverse Link Measurements
Setting the Measurement
Parameters
You can set the following measurement parameters using the Meas Setup menu. Press the Meas Setup key to open the Meas Setup menu.
Measurement...: Sets the limit table used for limit testing. You can select
Offset From Channel or Inband Spurious. For Offset From Channel, frequency zones are specified by the difference from the center frequency. For Inband Spurious, frequency zones are specified by the absolute frequency values.
Ref. Channel Select: Selects the reference mode used to measure the spectrum
emission level. You can select Auto or Man (manual). For Auto, the reference channel level is measured from the input signal for the specified channel bandwidth. For Man, you can set the reference channel level in the Ref. Channel Level menu item.
Ref. Channel Level: Sets the reference channel level used to measure the
spectrum emission level. You can set the value from 150 dBm to 30 dBm. This setting is only available when Ref. Channel Select is set to Man (manual).
Channel Bandwidth: Sets the channel bandwidth for the measurement.
Measurement Filter Shape...: Sets the filter to be used for the measurement.
You can select Rect (Rectangular), Gaussian, Nyquist, or Root Nyquist. This setting is only available when Ref. Channel Select is set to Auto.
Rolloff Ratio: Enters the rolloff ratio when the Nyquist or Root Nyquist filter
is selected. You can set the value from 0.0001 to 1. The default value is 0.5. This setting is only available when Ref. Channel Select is set to Auto.
Limits...: Sets the pass/fail limits for the spectrum emission mask
measurement. When you press this side key, the measurement limits editor appears. Refer to Editing the Measurement Limits on page 2-61 for more information.
Scroll Table: Scrolls the table displayed in the view.
NOTE. Refer to your instrument user manual for information about the scale
settings for the display.
2-56 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual

Pilot to Code Channel Measurement

The pilot to code channel measurement measures the time difference between the pilot channel and other code channels detected in the received signal. Figure 2-29 shows an example of the pilot to code channel measurement.
1xEV-DO Reverse Link Measurements
Figure 2-29: Pilot to code channel measurement-reverse link
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-57
1xEV-DO Reverse Link Measurements
Display Elements
Setting Acquisition and
Analysis Parameters
The following information is shown in the modulation accuracy measurement display:
Overview: This view can contain power versus time.
Subview: This view can contain spectrum, EVM, magnitude error, phase
error, constellation.
Main view: This view lists the following measurement results for the active
channels:
Power level Timing referred to the pilot channel Phase referred to the pilot channel Code domain error
The display content in the subview can be changed using the View Define menu. Refer to Changing the View Contents on page 2-59 for more information about the View De fi n e m e n u .
You can set the following acquisition and analysis parameters using the Acquisition/Analysis menu. Press the Acquisition/Analysis key to open the menu.
Acquisition Length (chips): Sets the acquisition length in chips. The setting
range depends on the acquisition memory size and the span setting.
Acquisition Length (s): Displays the acquisition length in seconds.
Acquisition History: Specifies the number of the block to display and
analyze. The latest block is number zero. Older blocks have larger negative numbers.
Analysis Interval: Sets the analysis interval in half slots. The setting range
depends on the acquisition length setting.
Analysis Offset: Sets the start point of the analysis with respect to the left end
of the acquisition length in half slots.
Spectrum Length: Display the time length for FFT processing of the spectrum
displayed in the subview. This value is equivalent to one frame acquisition length.
Spectrum Offset: Sets the beginning of Spectrum Length with respect to the
left end of the acquisition length.
NOTE. Refer to your instrument user manual for detailed information about the
time parameters.
2-58 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
1xEV-DO Reverse Link Measurements
Setting Measurement
Parameters
You can set the following measurement parameters using the Meas Setup menu. Press the Meas Setup key to open the Meas Setup menu.
Analyze: Performs analysis for the time slots in the analysis range.
Select HalfSlot: Sets the half slot for the measurement.
Select Code: Sets the code in the half slot for the measurement.
Longcode Mask I: Sets the long code mask value for the I signal. You can set
the value from 0 to 3FFFFFFFFFF in the hexadecimal format.
Longcode Mask Q: Sets the long code mask value for the Q signal. You can
set the value from 0 to 3FFFFFFFFFF in the hexadecimal format.
Active Channel Threshold: Sets the active channel threshold level (in dB
from the pilot) used for deciding whether a code channel is active or inactive. You can set the value from 100 dB to 0 dB.
Measurement Filter...: Selects the measurement filter to apply when
calculating EVM and other modulation accuracy results. You can select None, cdma2000, or cdma2000+EQ (equalizer).
IQ Swap: Sets whether to swap the I and Q data streams before demodulation.
Limits...: Sets the pass/fail limits for the pilot to code channel measurement.
When you press this side key, the measurement limits editor appears. Refer to Editing the Measurement Limits on page 2-61 for more information.
Changing the View
Contents
You can change the contents in the overview and subview using the View Define menu. Press the View: Define key to display the View Define menu.
Overview Content...: Selects a view to display in the overview. You can
select Spectrogram or Waveform (power versus time).
Subview Content...: Selects a view to display in the subview:
Spectrum: Displays spectrum of the input signal.
EVM: Displays changes of EVM (Error Vector Magnitude) over time.
MagErr: Displays changes of magnitude error over time.
PhaseErr: Displays changes of phase error over time.
Constellation: Displays the signal as an I-Q constellation.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-59
1xEV-DO Reverse Link Measurements
2-60 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual

Editing the Measurement Limits

This section tells how to set measurement limits that can be used to perform pass/fail testing for the 1xEV-DO measurements.
Refer to Measurement Limit Defaults on page 2-71 for information about the default settings of each measurement limit.

Using the Measurement Limits Editor

All of the measurement limits are set in the measurement limits editor, which you access from the Limits... side key on the Meas Setup menu.
Figure 2-30 shows the measurement limits editor for the 1xEV-DO forward link measurements.
Figure 2-30: Measurement limits editor
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-61
Editing the Measurement Limits
Table 2-1 describes all of the measurement limit items available in the measurement limits editor.
Table 2-1: Measurement limits items
Limit item Description Lower limit range Upper limit range
Channel power Sets the lower and upper limits for the Channel
Power measurement.
ACPR Select this item to access another measurement
limits editor to edit ACPR Limits.
Intermod 3rd Order Sets the upper limit of the third harmonic distortion
component for the Intermodulation measurement.
Intermod 5th Order Sets the upper limit of the fifth harmonic distortion
component for the Intermodulation measurement.
SEM Offset From Channel Select this item to access another measurement
limits editor to edit SEM offset from channel limits.
SEM Inband Spurious Select this item to access another measurement
limits editor to edit SEM inband spurious limits.
RMS EVM Sets the upper limit for the RMS EVM measurement. Not settable 0 to 100%
Sets the upper limit for the Peak EVM
Peak EVM
Peak Code Domain Error Sets the upper limit for the Peak code domain error
Rho Sets the lower limit of Rho for the modulation
Pilot To Code Timing Sets the upper limit for the pilot to code timing
Pilot To Code Phase Sets the lower and upper limits for the pilot to code
OBW Sets the upper and lower limits for the OBW
Tau Sets the lower limit of Tau for the modulation
Gated Output Power (Idle)
Gated Output Power (Active)
* These items are only available in the 1xEV-DO forward link measurements.
measurement. Not settable 0 to 100%
measurement.
accuracy measurement.
measurement.
phase measurement.
measurement.
accuracy measurement.
*
Select this item to access another measurement limits editor to edit gated output power limits.
*
Select this item to access another measurement limits editor to edit gated output power limits.
200 to 200 dBm 200 to 200 dBm
------------------ ------------------
Not settable -200 to 200 dBc
Not settable -200 to 200 dBc
------------------ ------------------
------------------ ------------------
Not settable 200 to 200 dB
0 to 1 Not settable
1 to 1 s 1 to 1 s
100 to 100 degree 100 to 100 degree
0 to 30 MHz 0 to 30 MHz
0 to 1 s Not settable
------------------ ------------------
------------------ ------------------
2-62 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
Editing the Measurement Limits
Setting a Measurement
Limit
Perform the following procedure to set measurement limits:
1. Press the Meas Setup menu key to open the Meas Setup menu.
2. Press the Limits... side key to open the measurement limits editor with
LIMITS menu.
3. Press the Select row to edit side key.
4. Turn the rotary knob or press the arrow keys to select the row that you want to
edit.
5. Press the Enable Limit side key to enable or disable the pass/fail testing for
the specified limit(s).
6. Press the Lower Limit side key to set the lower limit for the selected
measurement item, if applicable.
7. Press the Upper Limit side key to set the upper limit for the selected
measurement item.
When you want to set limits for the spectrum emission mask or gated output power measurement, perform the following steps to open another measurement limits editor:
1. Press the Select row to edit side key.
2. Turn the rotary knob or press the arrow keys to select ACPR, SEM Offset
From Channel, SEM Inband Spurious, Gated Output Power (Idle), or Gated Output Power (Active) row.
3. Press the Edit ACPR Limits..., Edit SEM Offset Limits..., Edit SEM
Inband Limits..., Edit Gated Output Power (Idle) Limits..., or Edit Gated Output Power (Active) Limits... side key. This opens the limits editor to set
more complex limits for the ACPR, spectrum emission mask, or gated output power measurement. Figure 2-31 shows the measurement limits editor for the ACPR measurement.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-63
Editing the Measurement Limits
Setting the ACPR Limits
When you select the ACPR row in the measurement limits editor, another measurement limits editor for the ACPR measurement is displayed. Figure 2-31 shows the measurement limits editor for the ACPR measurement.
Figure 2-31: ACPR measurement limits editor
2-64 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
Editing the Measurement Limits
Table 2-2 lists the measurement limit items that can be set in the measurement limits editor.
Table 2-2: Measurement limit items for the ACPR measurement
Item Description Limit ranges
Enable Enables or disables use of the limits in the specified zone (A to L). -------------------
Frequency Sets the frequency offset for the zone. 8 to 8 GHz Measurement BW Sets the measurement bandwidth for the zone. 8 to 8 GHz
Limit (Abs) Sets an absolute level limit for the zone. 200 to 200 dBm
Limit (Rel) Sets a relative level limit for the zone. 200 to 200 dBc Failure Mode Selects the fail mode to detect fail conditions between the measurement results and the
test limits. The choices are:
Absolute: Fail is detected when the absolute measurement result is larger than the
limit for Limit (Abs).
Relative: Fail is detected when the relative measurement result is larger than the
limit for Limit (Rel).
-------------------
Perform the following steps to set measurement limits in the limits editor:
1. Press the Select column to edit side key.
2. Turn the rotary knob or press the arrow keys to select the column that you want
to edit.
3. Press the Enable side key to enable or disable use of the limits for the zone.
4. Press the side key corresponding to the limit item, and then use the rotary knob
or arrow keys to set the value.
5. Press the Failure Mode... side key to select the failure mode.
6. Press the Default limits to... side key if you want to use a given measurement
limit for the measurement. Refer to Measurement Limit Defaults on page 2-71 for information about the default settings of each measurement limit.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-65
Editing the Measurement Limits
Setting the Spectrum
Emission Mask Limits
When you select the SEM Offset From Channel or SEM Inband Spurious row in the measurement limits editor, another measurement limits editor for the spectrum emission mask measurement is displayed. Figure 2-32 shows the measurement limits editor for the spectrum emission mask measurement.
Figure 2-32: Spectrum emission mask measurement limits editor
Table 2-3 lists the measurement limit items that can be set in the measurement limits editor.
Table 2-3: Measurement limit items for the spectrum emission mask measurement
Item Description Limit ranges
Enable Enables or disables use of the mask in the specified zone (A, B, C, D, or E). ------------------­Beginning Frequency Sets the beginning frequency for the zone. 8 to 8 GHz
Ending Frequency Sets the ending frequency for the zone. 8 to 8 GHz
Measurement Bandwidth Sets the measurement bandwidth. 8 to 8 GHz Offset from Carrier* Specifies which offset side is to be measured. You can select Upper (positive),
Lower (negative), or Both.
-------------------
2-66 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
Editing the Measurement Limits
Table 2-3: Measurement limit items for the spectrum emission mask measurement (cont.)
Item Description Limit ranges
Fail if signal exceeds Selects the fail mode to detect fail conditions between the measurement results and
the test limits. The choices are:
-------------------
Absolute: Fail is detected when one of the absolute measurement results is
larger than the limit for Beginning Absolute Limit and/or Ending Absolute Limit.
Relative: Fail is detected when one of the relative measurement results is
larger than the limit for Beginning Relative Limit and/or Ending Relative Limit.
Rel OR Abs: Fail is detected when one of the absolute measurement results is
larger than the limit for Beginning Absolute Limit and Ending Absolute Limit OR one of the relative measurement results is larger than the limit for Beginning Relative Limit and Ending Relative Limit.
Rel AND Abs: Fail is detected when one of the absolute measurement results
is larger than the limit for Beginning Absolute Limit and Ending Absolute Limit AND one of the relative measurement results is larger than the limit for
Beginning Relative Limit and Ending Relative Limit. Beginning Absolute Limit Sets an absolute level limit at Beginning Frequency. 200 to 200 dBm
Ending Absolute Limit Sets an absolute level limit at Ending Frequency. 200 to 200 dBm
Beginning Relative Limit Sets a relative level limit at Beginning Frequency. 200 to 200 dBc Ending Relative Limit Sets a relative level limit at Ending Frequency. 200 to 200 dBc
* This item is only available when SEM Offset From Channel is selected.
Perform the following steps to set measurement limits in the limits editor:
1. Press the Select column to edit side key.
2. Turn the rotary knob or press the arrow keys to select the column that you want
to edit.
3. Press the Enable side key to enable or disable use of the limits for the zone.
4. Press the side key corresponding to the limit item, and then use the rotary knob
or arrow keys to set the value.
5. Press the Default limits to... side key if you want to use a given measurement limit for the measurement. Refer to Measurement Limit Defaults on page 2-71 for information about the default settings of each measurement limit.
6. When you press the Rescale Graph side key, the graph is rescaled to match the set limits.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-67
Editing the Measurement Limits
Setting the Gated Output
Power Limits
When you select the Gated Output Power (Idle) row or Gated Output Power (Active) row in the measurement limits editor, another measurement limits editor for the gated output power measurement is displayed. Figure 2-33 shows the measurement limits editor for the gated output power measurement.
Figure 2-33: Gated output power measurement limits editor
Table 2-4 lists the measurement limit items that can be set in the measurement limits editor.
Table 2-4: Measurement limit items for the gated output power measurement
Item Description Limit ranges
Enable Enables or disables use of the mask in the specified zone (A, B, C, D, or E). ------------------­Beginning Time Offset Sets the beginning time from the center of the zone. 10 to 10 s
Ending Time Offset Sets the ending time from the center of the zone. 10 to 10 s
Limit Type Sets the limit type for the measurement. You can select Upper (upper limit) or Lower
(lower limits).
Offset Side Specifies which offset side is to be measured. You can select Upper (positive),
Lower (negative), or Both.
-------------------
-------------------
2-68 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
Editing the Measurement Limits
Table 2-4: Measurement limit items for the gated output power measurement (cont.)
Item Description Limit ranges
Fail if signal exceeds Selects the fail mode to detect fail conditions between the measurement results and
the test limits. The choices are:
-------------------
Absolute: Fail is detected when one of the absolute measurement results is
larger than the limit for Beginning Absolute Limit and/or Ending Absolute Limit.
Relative: Fail is detected when one of the relative measurement results is
larger than the limit for Beginning Relative Limit and/or Ending Relative Limit.
Rel OR Abs: Fail is detected when one of the absolute measurement results is
larger than the limit for Beginning Absolute Limit and Ending Absolute Limit OR one of the relative measurement results is larger than the limit for Beginning Relative Limit and Ending Relative Limit.
Rel AND Abs: Fail is detected when one of the absolute measurement results
is larger than the limit for Beginning Absolute Limit and Ending Absolute Limit AND one of the relative measurement results is larger than the limit for
Beginning Relative Limit and Ending Relative Limit. Beginning Absolute Limit Sets an absolute level limit at Beginning Time Offset. 200 to 200 dBm
Ending Absolute Limit Sets an absolute level limit at Ending Time Offset. 200 to 200 dBm
Beginning Relative Limit Sets a relative level limit at Beginning Time Offset. 200 to 200 dB Ending Relative Limit Sets a relative level limit at Ending Time Offset. 200 to 200 dB
Perform the following steps to set measurement limits in the limits editor:
1. Press the Select column to edit side key.
2. Turn the rotary knob or press the arrow keys to select the column that you want
to edit.
3. Press the Enable side key to enable or disable use of the limits for the zone.
4. Press the side key corresponding to the limit item, and then use the rotary knob
or arrow keys to set the value.
5. Press the Default limits to... side key if you want to use a given measurement limits for the measurement. Refer to Measurement Limit Defaults on page 2-71 for information about the default settings of each measurement limit.
6. When you press the Rescale Graph side key, the graph is rescaled to match the set limits.
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-69
Editing the Measurement Limits

Saving and Loading Measurement Limits

Measurement limits that set in the measurement limits editor can be saved as a file and loaded from a file. This section describes how to save and load measurement limits.
Refer to your instrument user manual, for further information about file operation including how to enter a file name and how to delete a file.
Saving Measurement
Limits
Loading Measurement
Limits
Perform the following steps to save the current measurement limits to a file:
1. When you open the spectrum emission mask measurement limits editor or gated output power measurement limits editor, press the Cancel-Back
(MENU) side key to return the measurement limit editor (see Figure 2-30).
2. Press the Save Limits side key to open the Save to File menu.
You can save the limits by using the preset file name or by entering a new file name.
3. To use the preset file name, press one of the following side keys: Save to LimitsA, Save to LimitsB, or Save to LimitsC.
4. To enter a new file name, type a file name in the text box at the top of the menu,
and then press the Save File Now side key.
The file extension *.lmt is automatically added to the limits file.
Perform the following steps to load measurement limits from a file:
1. When you open the spectrum emission mask measurement limits editor or gated output power measurement limits editor, press the Cancel-Back
(MENU) side key to return the measurement limit editor (see Figure 2-30).
2. Press the Load Limits side key to open the Load From File menu.
3. To load limits from the preset file, press one of the following side keys: Load from LimitsA, Load from LimitsB, or Load from LimitsC.
4. To load limits from an existing file, select the file from the file list, and then press the Load File Now side key.
2-70 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual

Measurement Limit Defaults

Tables 2-5 through 2-21 list the default limit settings used to perform the pass/file testing for the 1xEV-DO measurements. Refer to Using the Measurement Limits Editor on page 2-61 for procedures to edit the measurement limits.
Editing the Measurement Limits
Common Limits
Tables 2-5 and 2-6 list the common measurement limits that can be used for the measurements other than the ACPR, Spectrum Emission Mask and Gated Output Power measurements.
Table 2-5: Forward link common limits
Limit Enable Lower Upper Units
Channel Power No −150 30 dBm
Intermod 3rd Order Yes N.A −20 dBc
Intermod 5th Order Yes N.A −40 dBc RMS EVM No N.A 10 percent
Peak EVM No N.A 31 percent
Peak Code Domain Error No N.A 0.4 dB Rho Yes 0.912 N.A N.A
Pilot To Code Timing Yes 50 n 50 n s
Pilot To Code Phase Yes −0.15 0.15 radian OBW Yes N.A 1.48 M Hz
Tau Yes N.A 0.00001 s
Table 2-6: Reverse link common limits
Limit Enable Lower Upper Units
Channel Power No −150 30 dBm
Intermod 3rd Order Yes N.A −18 dBc Intermod 5th Order Yes N.A −36 dBc
RMS EVM No N.A 10 percent
Peak EVM No N.A 24 percent Peak Code Domain Error No N.A 0.25 dB
Rho Yes 0.994 N.A N.A
Pilot To Code Timing Yes 10 n 10 n s Pilot To Code Phase Yes −0.15 0.15 radian
OBW Yes N.A 1.48 M Hz
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-71
Editing the Measurement Limits
ACPR Limits
Table 2-7 list the measurement limits that can be used for the ACPR measurement.
Table 2-7: Forward and reverse link: all band classes
Zone Enable Frequency Measure BW Limit (Abs) Limit (Rel) Fail Mode
A Yes 750 kHz 30 kHz 27 dBm 45 dBc Relative BYes 750 kHz 30 kHz 27 dBm 45 dBc Relative C Yes 1.995 MHz 30 kHz 27 dBm 55 dBc Relative DYes 1.995 MHz 30 kHz 27 dBm 55 dBc Relative E Yes 3.125 MHz 30 kHz 13 dBm 55 dBc Relative FNo 3.125 MHz 30 kHz 13 dBm 55 dBc Relative G No 4 MHz 30 kHz 13 dBm 55 dBc Relative HNo 4 MHz 30 kHz 13 dBm 55 dBc Relative I No 7.5 MHz 30 kHz 13 dBm 55 dBc Relative JNo 7.5 MHz 30 kHz 13 dBm 55 dBc Relative K No 0 Hz 30 kHz 13 dBm 55 dBc Relative L No 0 Hz 30 kHz 13 dBm 55 dBc Relative
SEM Offset from Channel
Limits
Tables 2-8 through 2-16 list the measurement limits that can be used for the SEM Offset From Channel measurement.
Table 2-8: Forward link: BC (band class) 0, 2, 3, 5, and 9
Zone ABCDE
En ab le Yes Ye s Yes No N o
Start Frequency 750 kHz 1.98 MHz 4 MHz 4 MHz 20 MHz
Stop Frequency 1.98 MHz 4.00 MHz 20 MHz 3 GHz 8 GHz
Measurement BW 30 kHz 30 kHz 1 MHz 1 MHz 1 MHz Offset Side Both Both Both Both Both
Failure Mode Relative Rel OR Abs Absolute Absolute Absolute
Absolute Start 30 dBm 27 dBm 36 dBm 13 dBm 13 dBm Absolute Stop 30 dBm 27 dBm 36 dBm 13 dBm 13 dBm
Relative Start 45 dBc 55 dBc 0 dBc 0 dBc 0 dBc
Relative Stop 45 dBc 55 dBc 0 dBc 0 dBc 0 dBc
2-72 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
Editing the Measurement Limits
Table 2-9: Forward link: BC (band class) 1, 4, and 8
Zone ABCDE
Enable Yes Yes Yes Yes No
Start Frequency 885 kHz 1.25 MHz 2.25 MHz 4 MHz 4 MHz
Stop Frequency 1.25 MHz 2.25 MHz 4 MHz 20 MHz 3 GHz Measurement BW 30 kHz 30 kHz 1 MHz 1 MHz 1 MHz
Offset Side Both Both Both Both Both
Failure Mode Relative Rel OR Abs Absolute Absolute Absolute Absolute Start 30 dBm 9 dBm 13 dBm 36 dBm 13 dBm
Absolute Stop 30 dBm 9 dBm 13 dBm 36 dBm 13 dBm
Relative Start 45 dBc 45 dBc 0 dBc 0 dBc 0 dBc Relative Stop 45 dBc 45 dBc 0 dBc 0 dBc 0 dBc
Table 2-10: Forward link: BC (band class) 3b
Zone ABCDE
Enable Yes Yes Yes No No
Start Frequency 750 kHz 1.98 MHz 4 MHz 4 MHz 20 MHz Stop Frequency 1.98 MHz 4 MHz 20 MHz 3 MHz 8 GHz
Measurement BW 30 kHz 30 kHz 1 MHz 1 MHz 1 MHz
Offset Side Both Both Both Both Both Failure Mode Relative Relative Absolute Absolute Absolute
Absolute Start 30 dBm 27 dBm 36 dBm 13 dBm 13 dBm
Absolute Stop 30 dBm 27 dBm 36 dBm 13 dBm 13 dBm
Relative Start 45 dBc 60 dBc 0 dBc 0 dBc 0 dBc Relative Stop 45 dBc 60 dBc 0 dBc 0 dBc 0 dBc
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-73
Editing the Measurement Limits
Table 2-11: Forward link: BC (band class) 6
Zone ABCDE
En ab le Yes Ye s Yes Yes Ye s
Start Frequency 885 kHz 1.25 MHz 1.45 MHz 2.25 MHz 4 MHz
Stop Frequency 1.25 MHz 1.45 MHz 2.25 MHz 4 MHz 20 MHz Measurement BW 30 kHz 30 kHz 30 kHz 1 MHz 1 MHz
Offset Side Both Both Both Both Both
Failure Mode Relative Rel OR Abs Rel OR Abs Absolute Absolute Absolute Start 30 dBm 13 dBm 13 dBm 13 dBm 36 dBm
Absolute Stop 30 dBm 13 dBm 26.6 dBm
Relative Start 45 dBc 45 dBc 45 dBc 0 dBc 0 dBc Relative Stop 45 dBc 45 dBc 45 dBc 0 dBc 0 dBc
1 Based on -[13+17x( f-1.45 MHz)] dBm when f is 2.25 MHz.
1
13 dBm 36 dBm
Table 2-12: Forward link: BC (band class) 7
Zone ABCDE
En ab le Yes Ye s Yes Yes Ye s
Start Frequency 750 kHz 1.98 MHz 3.25 MHz 4 MHz 4 MHz Stop Frequency 1.98 MHz 3.25 MHz 4 MHz 20 MHz 20 MHz
Measurement BW 30 kHz 30 kHz 30 kHz 1 MHz 1 MHz
Offset Side Both Both Both Both Both Failure Mode Relative Rel OR Abs Rel OR Abs Absolute Absolute
Absolute Start 30 dBm 27 dBm 46 dBm 36 dBm 36 dBm
Absolute Stop 30 dBm 27 dBm 46 dBm 36 dBm 36 dBm
Relative Start 45 dBc 55 dBc 55 dBc 0 dBc 0 dBc Relative Stop 45 dBc 55 dBc 55 dBc 0 dBc 0 dBc
2-74 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
Editing the Measurement Limits
.
Table 2-13: Reverse link: BC (band class) 0, 2, 3, 5, and 9
Zone ABCDE
Enable Yes Yes Yes No No
Start Frequency 885 kHz 1.98 MHz 4 MHz 4 MHz 20 MHz
Stop Frequency 1.98 MHz 4 MHz 20 MHz 3 GHz 8 GHz Measurement BW 30 kHz 30 kHz 1 MHz 1 MHz 1 MHz
Offset Side Both Both Both Both Both
Failure Mode Rel AND Abs Rel AND Abs Absolute Absolute Absolute Absolute Start 54 dBm 54 dBm 36 dBm 13 dBm 13 dBm
Absolute Stop 54 dBm 54 dBm 36 dBm 13 dBm 13 dBm
Relative Start 42 dBc 55 dBc 0 dBc 0 dBc 0 dBc Relative Stop 42 dBc 55 dBc 0 dBc 0 dBc 0 dBc
.
Table 2-14: Reverse link: BC (band class) 1, 4, and 8
Zone ABCDE
Enable Yes Yes Yes No No
Start Frequency 1.25 MHz 1.98 MHz 4 MHz 4 MHz 20 MHz Stop Frequency 1.98 MHz 4 MHz 20 MHz 3 GHz 8 GHz
Measurement BW 30 kHz 30 kHz 1 MHz 1 MHz 1 MHz
Offset Side Both Both Both Both Both Failure Mode Relative Rel OR Abs Absolute Absolute Absolute
Absolute Start 54 dBm 54 dBm 36 dBm 13 dBm 13 dBm
Absolute Stop 54 dBm 54 dBm 36 dBm 13 dBm 13 dBm
Relative Start 42 dBc 50 dBc 0 dBc 0 dBc 0 dBc Relative Stop 42 dBc 50 dBc 0 dBc 0 dBc 0 dBc
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-75
Editing the Measurement Limits
.
Table 2-15: Reverse link: BC (band class) 6
Zone ABCDE
En ab le Yes Ye s Yes No N o
Start Frequency 1.25 MHz 1.98 MHz 2.25 MHz 2.25 MHz 4 MHz
Stop Frequency 1.98 MHz 2.25 MHz 4 MHz 4 MHz 20 MHz Measurement BW 30 kHz 30 kHz 1 MHz 1 MHz 1 MHz
Offset Side Both Both Both Both Both
Failure Mode Rel AND Abs Rel OR Abs Rel OR Abs Absolute Absolute Absolute Start 54 dBm 54 dBm 13 dBm 13 dBm 36 dBm
Absolute Stop 54 dBm 54 dBm 14.75 dBm
2
13 dBm 36 dBm
Relative Start 42 dBc 50 dBc 45 dBc 0 dBc 0 dBc Relative Stop 42 dBc 50 dBc 45 dBc 0 dBc 0 dBc
2 Based on -[13+1x( f-2.25 MHz)] dBm when f is 4 MHz.
.
Table 2-16: Reverse link: BC (band class) 7
Zone ABCDE
En ab le Yes Ye s Yes Yes Ye s
Start Frequency 885 kHz 1.98 MHz 2.25 MHz 4 MHz 4 MHz Stop Frequency 1.98 MHz 2.25 MHz 4 MHz 20 MHz 20 MHz
Measurement BW 30 kHz 30 kHz 30 kHz 1 MHz 1 MHz
Offset Side Both Both Both Both Both Failure Mode Rel AND Abs Rel AND Abs Rel AND Abs Absolute Absolute
Absolute Start 54 dBm 54 dBm 35 dBm 36 dBm 36 dBm
Absolute Stop 54 dBm 54 dBm 35 dBm 36 dBm 36 dBm
Relative Start 42 dBc 54 dBc 55 dBc 0 dBc 0 dBc Relative Stop 42 dBc 54 dBc 55 dBc 0 dBc 0 dBc
2-76 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
Editing the Measurement Limits
SEM Inband Spurious
Limits
Tables 2-17 through 2-19 list the measurement limits that can be used for the SEM Inband Spurious measurement.
Table 2-17: Forward and reverse link: BC (band class) 0, 1, 2, 3, 4, 5, 7, 8, and 9
Zone ABCDE
Enable Yes Yes No No No
Start Frequency 1 GHz 925 MHz 935 MHz 1.805 GHz 4 MHz Stop Frequency 3 GHz 935 MHz 960 MHz 1.880 GHz 8 GHz
Measurement BW 1 MHz 100 kHz 100 kHz 100 kHz 1 MHz
Failure Mode Absolute Absolute Absolute Absolute Absolute Absolute Start 36 dBm 67 dBm 79 dBm 71 dBm 36 dBm
Absolute Stop 36 dBm 67 dBm 79 dBm 71 dBm 36 dBm
Relative Start 0 dBc 0 dBc 0 dBc 0 dBc 0 dBc Relative Stop 0 dBc 0 dBc 0 dBc 0 dBc 0 dBc
Table 2-18: Forward link: BC (band class) 6
Zone ABCDE
Enable Yes Yes Yes Yes No
Start Frequency 1.8935 GHz 925 MHz 935 MHz 1.805 GHz 4 MHz Stop Frequency 1.9196 GHz 935 MHz 960 MHz 1.880 GHz 20 MHz
Measurement BW 300 kHz 100 kHz 100 kHz 100 kHz 1 MHz
Failure Mode Absolute Absolute Absolute Absolute Absolute Absolute Start 41 dBm 67 dBm 79 dBm 71 dBm 36 dBm
Absolute Stop 41 dBm 67 dBm 79 dBm 71 dBm 36 dBm
Relative Start 0 dBc 0 dBc 0 dBc 0 dBc 0 dBc
Relative Stop 0 dBc 0 dBc 0 dBc 0 dBc 0 dBc
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-77
Editing the Measurement Limits
Table 2-19: Reverse link: BC (band class) 6
Zone ABCDE
En ab le Yes Ye s Yes Yes Ye s
Start Frequency 1.8935 GHz 876 MHz 921 MHz 1.710 GHz 1.805 GHz
Stop Frequency 1.9196 GHz 915 MHz 960 MHz 1.785 GHz 1.880 GHz Measurement BW 300 kHz 100 kHz 100 kHz 100 kHz 100 kHz
Failure Mode Absolute Absolute Absolute Absolute Absolute
Absolute Start 41 dBm 98 dBm 57 dBm 98 dBm 47 dBm Absolute Stop 41 dBm 98 dBm 57 dBm 98 dBm 47 dBm
Relative Start 0 dBc 0 dBc 0 dBc 0 dBc 0 dBc
Relative Stop 0 dBc 0 dBc 0 dBc 0 dBc 0 dBc
Gated Output Power Limits
Tables 2-20 and 2-21 list the measurement limits that can be used for the gated output power measurement.
Table 2-20: Forward link, all band classes, idle slot
Zone ABCDE
En ab le Yes Ye s Yes No N o
Start Time 0 s 0 s 97.33 µs0 s 0 s
Stop Time 90.33 µs 97.33 µs 416.665 µs 416.665 µs 416.665 µs Upper/Lower Lower Upper Upper Lower Upper
Offset Side Both Both Both Both Both
Failure Mode Relative Relative Relative Relative Relative
Absolute Start 150 dBm 30 dBm 30 dBm 150 dBm 30 dBm Absolute Stop 150 dBm 30 dBm 30 dBm 150 dBm 30 dBm
Relative Start 2.5 dB 2.5 dB 7 dB 2.5 dB 2.5 dB
Relative Stop 2.5 dB 2.5 dB 7 dB 2.5 dB 2.5 dB
2-78 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
Editing the Measurement Limits
Table 2-21: Forward link, all band classes, active slot
Zone ABCDE
Enable No No No Yes Yes
Start Time 0 s 0 s 97.33 µs0 s 0 s
Stop Time 90.33 µs 97.33 µs 416.665 µs 416.665 µs 416.665 µs Upper/Lower Lower Upper Upper Lower Upper
Offset Side Both Both Both Both Both
Failure Mode Relative Relative Relative Relative Relative Absolute Start 150 dBm 30 dBm 30 dBm 150 dBm 30 dBm
Absolute Stop 150 dBm 30 dBm 30 dBm 150 dBm 30 dBm
Relative Start 2.5 dB 2.5 dB 7 dB 2.5 dB 2.5 dB Relative Stop 2.5 dB 2.5 dB 7 dB 2.5 dB 2.5 dB
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual 2-79
Editing the Measurement Limits
2-80 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual
Appendices

Appendix A: Setting Range

This section lists the setting range of the horizontal and vertical scale for the views.
Table A-1: Display format and scale
Display format Horizontal range Vertical range
Spectrum 0 Hz to 8 GHz 200 to 100 dBm
Spectrogram 0 Hz to 8 GHz 15999 to 0 frame
63999 to 0 frame (Option 02)
Code domain power Forward link:
MAC: 64 codes Pilot: 32 codes Data: 16 codes Preamble: 32 codes
Reverse link: 16 codes
Modulation accuracy Fixed Fixed
EVM Foward link:
Channel Chip Symbol Overall 1024 ------­MAC 128 2 Pilot 96 3 Data max 800 max 50 Preamble max 800 max 25
Reverse link:
Chip: 1024 Symbol: max 256
Amplitude (Mag) error Same as the horizontal range of EVM 300 to 300% Phase error Same as the horizontal range of EVM 675 to 675 degrees
Power codogram Same as the horizontal range of Code domain power 6144 to 0 frame
Symbol table Forward link:
MAC: 2 Pilot: 3 Data: max 50 Preamble: max 25
Reverse link: max 256
IQ power graph Same as the horizontal range of EVM 6144 to 0 frame
Constellation Fixed Fixed
Relative: 200 to 100 dB Absolute: 140 to -160 dBm
100 to 200%
24579 to 0 frame (Option 02)
NA
24579 to 0 frame (Option 02)
RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual A-1
Appendix A: Setting Range
A-2 RSA3303B, RSA3308B, & RSA3408B Option 26 User Manual