Tektronix RMD5000 Instructions Manual

xx
RMD5000
ZZZ
Rackmount Kit
Instructions
www.tektronix.com
P075102000*
*
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O. Box 5 0 0 Beaverto USA
For product information, sales, service, and technical support:
n, OR 97077
In North America, call 1-800-833-9200. Worl dwid e, visi t www.tektronix.com to nd contacts in your area.
Service Safety Summary
Only qualied personnel should perform service procedures. Read this Service Safety Summary and the General Safety Summary in the product service manual or the instru
Do Not Service Alone. Do not perform internal service or adjustments of this product unless another person capable of rendering rst aid and resuscitation is present.
To prevent the instrument and rack from falling onto the operator, two or more installers should install the instrument into the rack cabinet. After completing the installation procedure, the installers should verify that the instrument and rack cabinet
Disconnect Power. To avoid electric shock, switch off the instrument power, then disconnect the power cord from the mains power.
UseCareWhenServicingWithPowerOn. Dangerousvoltagesorcurrentsmay
in this product. Disconnect power and disconnect test leads before removing
exist protective panels, soldering, or replacing components.
To av
ction manual.
will not tip forward while the instrument is in the extended position.
oid electric shock, do not touch exposed connections.
RMD5000 Rackmount Kit Instructions 1
Kit Description
Kit Description
Products
These instructions describe the installation of the rackmount kit to your standard bench-top instrument. The rackmount kit is a collection of parts that, once installed, c 19-inch equipment rack. (See Table 1.) If you require the instrument to slide in and out of the rack, order the optional slide kit. (See Table 2.)
DPO4000B Series All Serial Numbers
MSO4000B Series AllSerialNumbers
ongure the instrument for mounting in a xed position into a standard
DPO5000 S
MSO5000 Series AllSerialNumbers
eries AllSerialNumbers
Kit Parts List
The parts included in this rackmount kit are listed below. (See Table 1.) (See
e 1 on page 3.)
Figur
Table 1: RMD5000 Rackmount kit
Item Quantity Part number Description
1 1 each 407-5531-XX
2 1 each 407-5532-XX
3 1 each 407-5534-XX
4 1 each 407-5533-XX
5
6 1 each 016-2037-XX HARDWARE KIT
7
1 each 426-2636-XX
1 each 016-2038-XX
1 each 075-1020-XX
PLATE, MOUNTING; RACKMOUNT LEFT
PLATE, MOUNTING; RACKMOUNT RIGHT
PLATE, SUPPORT; RACKMOUNT TOP
PLATE, SUPPORT; RACKMOUNT BOTTOM
PLATE, SUPPO RT; RACKMOUNT FRONT
COMMON HARDWARE KIT
TECH MANUAL: INSTRUCTIONS, RACKMOUNT KIT
Optional Accessories
Table 2: Optional Accessories
Part number Description
351-1095-XX
2 RMD5000 Rackmount Kit Instructions
SLIDE ASSY; PAIR, W/STD HARDWARE KIT AND REAR BRACKET
Kit Description
Figure 1: Rackmount kit parts (part appearance may vary)
Figure 2: Optional slide assembly kit
RMD5000 Rackmount Kit Instructions 3
Kit Description
Warranted Cha
racteristics
When the instrument is installed according to the instructions in this document, the rack-mounted instrument meets all warranted characteristics listed in the instrument s Requirements.) Instruments mounted using methods other than those described in these instructions may not meet their warranted characteristics.
See the specications for your instrument model for tables of the warranted characteristics.
Cooling air enters on the bottom and right sides as shown. (See Figure 3.) You must provide adequate cool air to meet the ambient temperature requirements listed. (See Table 3 on page 5.)
pecication except for those listed. (See page 5, Environmental
Figure 3: Instrument cooling
4 RMD5000 Rackmount Kit Instructions
Environmental Requirements
The following environmental characteristics supersede those listed in the specications and performance verication sections of the manual for your instrument.
Table 3: Warranted characteristics
Characteristic Description
Temperature inside rack cabinet
Operating +10°Cto+45°C(+50°Fto+113°F)
Nonoperating
Shock
Operating and nonoperating
Clearance Requirements
Kit Description
–22 °C to +60 °C (–7.6 °F to +140 °F)
20 g, 11 ms, half sine, each axis, 3 drops each
The rack in which the rack adapted instrument is mounted must provide the
wing clearance requirements:
follo
A minimum of 8.75 in (222.25 mm) of vertical space
A minimum width of 17.72 in (450.09 mm) between the left- and right-front rails in the rack
A minimum depth of 12.75 in (323.85 mm)
CAUTION. Adhering to these clearance requirements provides the rack-mounted
strument with sufcient clearance for air circulation and accommodation of the
in power cord and mounting hardware. Failure to provide these clearances can result in overheating and can cause instrument faults or failure.
RMD5000 Rackmount Kit Instructions 5
Kit Description
re 4: Instrument with rack adapter installed
Figu
6 RMD5000 Rackmount Kit Instructions
Installation Instructions
This section contains the procedures needed to rackmount a DPO4000B, MSO4000B, DPO5000, or MSO5000 series instrument.
Minimum Tool and Equipment List
The following tools are required to attach the rack-adapter kit hardware, install cabling har equipment cabinet. All tools are standard tools that are readily available.
Table 4: Tools required for rackmount installation
Name Description
Screwdriver handle (magnetic)
No. 2 Phillips or PoziDriv tip
T-10 TORX tip TORX-drive tip for T-10 size screw heads (at-head)
T-15 TORX tip TORX-drive tip for T-15 size screw heads
T-20 TORX tip TORX-drive tip for T-20 size screw heads
1/4-inch wrench
Torque driver
dware, and mount the rack-adapted instrument into a standard
Installation Instructions
Accepts 1/4-inch hexagonal head driver tips
Phillips or PoziDriv-driver tip for number 2 size screw heads
Wrench or nut driver can be used to install slides
6.5 in long shaft; accepts 1/4-inch hexagonal head driver tips
These instructions are for qualied service personnel who are familiar with
vicing the product. If you need further details for disassembling or reassembling
ser the product, refer to the appropriate product manual. Contact your nearest Tektronix Service Center or Tektronix Factory Service for installation assistance.
WAR NI NG . To prevent the rackmounted instrument from tipping forward onto the
operator, install the instrument so that the operator will be able to access all of its rear-panel connectors without pushing down on the instrument.
Verify that the r ack does not become unstable with the instrument fully extended. Do not leave the instrument extended when nished accessing the rear panel.
WAR NI NG . To avoid electric shock, disconnect the power cord from the mains
power before removing the cover. Failure to do so could result in personal injury or death.
RMD5000 Rackmount Kit Instructions 7
Installation Instructions
Remove
Equipment Required: Screwdriver handle with T-15 tip.
This procedure prepares the instrument for installation of the rackmount parts.
NOTE. Retain all parts removed from the instrument for later use. Some of those
parts will be reinstalled as you perform this rack conversion; store the remainder in case reconversion to a standard instrument conguration is desired at a later time.
1. Disconnect the power cord from the rear of the instrument.
2. Remove th
case assembly. (See Figure 5.)
3. Remove t
e four T-15 TORX screws that secure the rear cover to the front
he rear cover.
Figure 5: Removing the rear case cover
8 RMD5000 Rackmount Kit Instructions
Install
Installation Instructions
This section describes mounting the rackmount parts to the instrument.
Equipment Required: Torque driver with T-15 and T-20 tips.
1. Assemble the usingsix6-32x0.250inchpanheadT-15 TORX-drive screws tightened to 8 in-lb. (See Figure 6.)
NOTE. The bottom support plate has holes to facilitate mounting optional
accessories. You can use screws or nylon straps to mount accessories.
left and right mounting plates to the bottom support plate
Figure 6: Assembling the left and right mounting plates
RMD5000 Rackmount Kit Instructions 9
Installation Instructions
2. Fasten the fron 6-32x0.250inchpanheadT-15TORX-drivescrews,liningupthedimplesin the back support plate with the slots in the mounting plates. (See Figure 7.) Tighten screws to 8 in-lb.
t support plate to the left and right mounting plates with six
Figure 7: Assembling the rackmount front support plate
3. Install the two handles using four 10-32 x 0.25 inch pan head T-20 TORX-drive screws tightened to 16 in-lb. Handles should curve towards the sides of the instrument.
10 RMD5000 Rackmount Kit Instructions
Installation Instructions
4. Install the top (See Figure 8.)
For DPO4000B i toward the front of the enclosure so that the “DPO4000B” stamping on the plate is visible.
For DPO5000 instruments, install the top support plate with the ange toward the back of the enclosure.
support plate; note the orientation for the different instruments:
nstruments, install the top support plate with the ange
Figure 8: Installing the top support plate
RMD5000 Rackmount Kit Instructions 11
Installation Instructions
5. Install the ins inch at-head T-15 TORX-drive screws tightened to 8 in-lb. (See Figure 9.)
trument into the rackmount enclosure using six 6-32 x 0.250
Figure 9: Installing the instrument in the enclosure
12 RMD5000 Rackmount Kit Instructions
Installation Instructions
This completes this kit. The instrument is now ready to mount in a xed location, to the front rails of a standard instrument rack.
To mount the instrument in a xed location, proceed to Finishing the Installation. (See page 21, Finishing the Installation.)
If you require the instrument to s lide in and out of the instrument rack, order the optional slide assembly kit and use the instructions that follow. (See page 14, Mount the Track Hardware (Optional).)
the installation of the standard rack-adapter hardware included in
RMD5000 Rackmount Kit Instructions 13
Installation Instructions
Mount the Track Hardware
(Optional)
Install the opt
1. Remove the front section of each of the two tracks. (See Figure 10.)
Figure 10: Preparing the tracks for mounting
2. Install the front left- and right-side track sections on the instrument using four 6-32 x 0.250 inch pan h ead T-15 TORX-drive screws tightened to 8 in-lb. (See Figure 11.)
ional slide assembly kit as follows:
Figure 11: Installing the left- and right-side track sections
14 RMD5000 Rackmount Kit Instructions
Installation Instructions
WAR NI NG . To ensure that the rackmount track locks, make sure the track button
latches are oriented correctly: the right-side latch is located toward the bottom of the rackmount panel and the left-side latch is located toward the top of the rackmount panel. (See Figure 12.) If the button latches are not oriented correctly the instrument could slide from the rack, which could cause personal injury or damage to th
e instrument.
Figure 12: Track orientation
This completes the installation of the slide hardware to the instrument. To
lete the installation, proceed to: Rackmount the Rack- Adapted Instrument.
comp (See page 16.)
RMD5000 Rackmount Kit Instructions 15
Installation Instructions
Rackmount the Rack-Adapted Instrument
This procedure assembles and installs the slide-out tracks in the equipment rack, and then installs the rack-adapted instrument in the rack.
The slide-out tracks permit the rack-adapted instrument to be extended out of the rack for rear-panel and connector maintenance without removing the instrument from the rack.
WARNING. To prevent the rackmounted instrument from tipping forward onto the
operator, rear devices without pushing down on the instrument.
install the instrument so that the operator will be able to access all of its
Install Track Assembly and
Instrument into the Rack
Verify th Do not leave the instrument extended when nished accessing the rear panel.
Equipm and one torque driver.
NOTE. The rack hardware kit contains hardware for mounting the instrument in
several congurations. Not all of the hardware in the kit will be needed.
A standard equipment rack has rails with universal hole spacing. If you use a rack with other than universal hole spacing, you may have to drill additional mounting holes in the rack.
1. Assemble the slide-out track:
at the rack does not become unstable with the instrument fully extended.
ent Required: One screwdriver handle, one number two PoziDriv tip,
a. Measure the distance between the front and rear rail of the equipment rack.
16 RMD5000 Rackmount Kit Instructions
Installation Instructions
Figure 13: Slide-out track assemblies
b. Align the rear bracket to the right slide-out track as shown. (See
re 13.) Note that the rear bracket has multiple pairs of mount-through
Figu holes. When aligning the bracket and track, be sure to select a pair of holes that mount the rear bracket so that the ange-to-ange distance matches the distance between the front rail and rear rail measured in step a.
c. Using a screwdriver with a number two PoziDriv tip, secure the rear
bracket to the right slide-out track using two s crews (10-32) and a bar nut as illustrated. Leave the screws loose so that the overall length of the slide-out track assembly can be adjusted when installing it in the rack.
d. Repeat steps b and c to assemble the left slide-out track assembly.
RMD5000 Rackmount Kit Instructions 17
Installation Instructions
2. Mount the slide in the kit parts list:
a. Select the mou
the front rail. Verify that the 4.36 inch and 10.500 inch clearances exist relative to those mounting holes. (See Figure 14.)
-out track assemblies, using the slide drawer hardware listed
nting position in rack: Select two 0.5 inch-spaced holes in
Figure 14: Vertical clearances for rack installation (left-front rail shown)
18 RMD5000 Rackmount Kit Instructions
Installation Instructions
b. Select the moun
To mount the slide-out tracks with their front and rear anges outside of the front and r substep c. (See Figure 15.) Add a bar nut to the installation only if the rails have untapped holes.
To mount with front and rear anges inside of rails, use the mounting method B. This mounting method assumes untapped holes.
ting method according to rack type:
ear rails, use the mounting method A shown when doing
Figure 15: Installing slide-out track assemblies in rack (top view)
c. Install in rack: Using the method and hardware determined from
substep b, se cure the right slide-out track assembly to its front and rear rails. The screws should be lightly seated so mounting can be adjusted later.
RMD5000 Rackmount Kit Instructions 19
Installation Instructions
d. Fixthelengtho
loose in step 1, substep c, applying 28 inch-lb of torque, to x the front to rear ange spacing of the slide-out track assembly.
e. Mount the left slide-out track assembly: Repeat substeps c through d to
mount the left slide-out track assembly.
3. Mount the instrument in the rack:
WARNING. To prevent the instrument from tipping or dropping onto the installers,
two or more people should install this instrument into the rack cabinet.
After completing the installation p rocedure, the installers should verify that the instrument and rack cabinet will not tip forward while the instrument is in the extended
position.
a. Install the instrument:
Working from the front of the rack, slide the inner track of each slide-out track assembly until it extends out the front of the rack. Continue to slide them out until they lock.
Insert the left and right tracks that extend from the rear of the instrument into the ends of the tracks just extended. Make sure the tracks mounted on
strument slip inside the inner tracks extended earlier.
the in
f the slide-out track assembly: Tighten the screws left
Slide the instrument backward until it stops.
Push to release the button latches, located on the outside of each track, and continue to slide the instrument all the way into the cabinet.
b. Level the rackmounted instrument:
Tighten the four screws that were left loose a t the rear of the rack when you did step 2, substep c, and then pull the instrument partially out of the rack.
Be sure that the four screws that were left loose at the front of the rack are loose enough to allow the slide-out track assemblies to seek their normal positions.
Retighten the four screws and push the instrument all the way into the rack. If the tracks do not slide smoothly, readjust the level using the method just detailed.
When leveling is completed, tighten the 10-32 screws using 28 inch-lb of torque.
20 RMD5000 Rackmount Kit Instructions
Installation Instructions
Finishing the Installation
If you intend to to complete your installation. If you mounted the slide track kit to the instrument, you should also perform this step to secure the instrument to the rack.
WAR NI NG . To prevent the rack-mounted instrument from sliding forward and
causing personal injury or instrument damage, always secure the instrument to the rack if the rack is moved (for example, if the rack is repositioned or relocated to another r
Secure the instrument to the rack using four 10-32 x 0.625 in screws and rectangular washer spacers as shown. (See Figure 16.)
mount the instrument to the rack without slides, use this procedure
oom).
re 16: Securing the instrument to the rack
Figu
dofdocument*
*En
RMD5000 Rackmount Kit Instructions 21