Tektronix R453, 453, AN-USM-273 Service manual

TM 11-6625-1722-15
TECHNICAL MANUAL
OPE R A TOR ’ S , OR GA N I Z A TI ON A L , D I R E C T S U PPOR T
GE N E R A L S U PPOR T , A N D D E POT M A I N TE N A N CE
MANUAL
OSCI L LOSCOP E AN / U SM - 2 7 3
(NSN 6625-00-930-6637)
This copy is a reprint which includes current
HEADQUARTERS,
DEPARTMENT OF THE ARMY
JANUARY 1972
THIS MANUAL IS AN AUTHENTICATION OF THE MANU­FACTURER’S COMMERCIAL LITERATURE WHICH, THROUGH USAGE, HAS BEEN FOUND TO COVER THE DATA REQUIRED TO OPERATE AND MAINTAIN THIS EQUIPMENT. SINCE THE MANIJAL WAS NOT PREPARED IN ACCORDANCE WITH MILITARY SPECIFICATION, THE FORMAT HAS NOT BEEN STRUCTURED TO CONSIDER LEVEL OF MAINTENANCE NOR TO INCLUDE A FOR­MAL SECTION ON DEPOT MAINTENANCE STANDARDS.

WARNING

DANGEROUS VOLTAGES
EXIST IN THIS EQUIPMENT
DON’T TAKE CHANCES!
CAUTION
Special 3% silver solder is required on the ceramic terminal
strips in this equipment. A 40- to 75-watt soldering iron should be used and it should be tinned with the same special
solder. Additional quantities of the solder may be procured under FSN 3439-912-8698. Ordinary solder may be used only in dire emergency.
This Manual Contains Copyrighted Material Reproduced
Tektronix, Inc. All Rights Reserved
Permission Of
By
TM 11-6625-1722-15
T
ECHNICAL M ANUAL
HEADQUARTERS
DEPARTMENT OF THE ARMY
NO. 11–6625–1722–15
ASHINGTON
, D.C., 10 January 1972
W
Operator’s Organizational, Direct Support, General Support, and Depot
Maintenance Manual Including Repair Parts and Special Tools Lists
OSCILLOSCOPE AN/USM–273
S
ECTION A.
INRODUCTION
1.
CHARACTERISTICS
2.
OPERATING INSTRUCTIONS . .. . . . . . . . . . . . . . . . . . . . .
CIRCUIT DESCRIPTION . . . . . . . . . . . . . . . . . .
3.
4.
MAINTENANCE
PERFOR0MANCE CHECK . . . . . . . . . . . . . . . . . . . . .
5.
Page
A-1
1-1
2–1
3–1
4–1
5-1
6.
7.
8.
9.
10,
APPENDIX A .
B.
C.
D.
CALIBRATION
PREVENTIVE MAINTENANCE INSTRUCTIONS . . . . . . . . . . . . . . . . .
MECHNICAL PARTS IDENTIFICATION . . . . . . . . . . . . . . . . . . . . . . .
DIAGRAMS
RACKMOUNTING
REFERENCES
ITEMS COMPRISING AN OPERABLE EQUIPMENT . . . . . . . . . . . . . . . .
MAINTENANCE ALLOCATION . . . . . . . . . . . . . . .
REPAIR PARTS AND SPECIAL TOOLS LIST . . . . . . . . . . . . . .
6–1
7–1
8-1
9-1
10-1
A–1
B–1
C-1
D-1
TM 11-6625-1722-15
Fig. 1-1. Top; the Type 453 Oscilloscope. Bottom; the Type R453 Oscilloscope.
A-O
SECTION 0
INSTRUCTIONS
TM 11-6625-1722-15
0-1.
maintenance instructions.
Scope
This manual describes Oscilloscope
Throughout this manual, the AN/USM-273 is re-
AN/USM-273 (fig. 1-1) and provides
ferred to as the Tektronix Type 453 Oscilloscope. The maintenance allo-
cation chart appears in appendix C.
Repair parts and special tools lists
are contained in TM-6625-1722-24P.
0-2.
Indexes of Publications
a. DA Pam 310-4.
Refer to the latest issue of DA Pam 310-4 to deter-
mine whether there are new editions, changes, or additional publications
pertaining to the equipment.
DA Pam 310-7.
b .
Refer to DA Pam 310-7 to determine whether there
are modification work orders (MWO’S) pertaining to the equipment.
0-3.
Maintenance Forms, Records, and Reports
Reports of Maintenance and Unsatisfactory
a.
Equipment.
Department
of the Army forms and procedures used for equipment maintenance will be
those prescribed by TM 38-750, The Army Maintenance Management System.
Report of Packaging and Handling Deficiencies.
b.
Fill out and for-
ward DD Form 6 (Packaging Improvement Report) as prescribed in AR 700-58/
NAVSUPINST 4030.29/AFR 71-13/MCO P4030.29A, and DLAR 4145.8.
Discrepancy
c.
in Shipment Report (DISREP) (SF 361). Fill out and
forward Discrepancy in Shipment Report (DISREP) (SF 361) as prescribed in
AR 55-38/NAVSUPINST 4610.33B/AFR 75-18/MCO P4610.19C and DLAR 4500.15.
0-1
TM 11-6625-1722-15
0-4.
Reporting Equipment Improvement Recommendations (EIR)
If your Oscilloscope AN/USM-273 needs improvement, let us know.
You,
us an EIR.
the user,
don’t like about your equipment.
design.
Tell us why a procedure is hard to perform.
(Quality Deficiency Report).
and Electronics Materiel Readiness Command, ATTN:
Monmouth, NJ 07703.
0-5.
Administrative Storage
We’ll send you a reply.
are the only one who can tell us what you
Let us know why you don’t like the
Put it on an SF 368
Mail it to Commander, US Army Communications
DRSEL-ME-MQ, Fort
Administrative storage of equipment issued to and used by Army activi-
ties shall be in accordance with TM 740-90–1.
0-6
Destruction of Army Electronics Materiel
Destruction of Army electronics materiel to prevent enemy use shall be
Send
in accordance with TM 750-244-2.
0-7.
Reporting Errors and Recommending Improvements
You can help improve this manual. If you find any mistakes or if you
know of a way to improve the procedures, please let us know. Mail your
letter or DA Form 2028 (Recommended Changes to Publications and Blank
Forms) to
Readiness Command, ATTN:
either case,
Commander, US Army Communications and Electronics Materiel
DRSEL-ME-MQ, Fort Monmouth, NJ 07703. In
a reply will be furnished direct to you.
0-2
Change 1

SECTION 1

CHARACTERISTICS
TM 11-6625-1722-15
Introduction
The Tektronix Type 453 Oscilloscope is a transistorized portable oscilloscope designed to operate in a wide range of environmental conditions. The light weight of the Type 453 allows it to be easily transported, while providing the performance necessary for accurate high-frequency meas­urements. provides calibrated deflection factors from 5 millivolts to 10 volts/division. Channels 1 and 2 can be cascaded using an external cable to provide a one millivolt minimum defect­ion factor (both VOLTS/DIV switches set to 5 mV).
The trigger circuits provide stable triggering over the full range of vertical frequency response. Separate trigger con­trols are provided to select the desired triggering for the A and B sweeps. One of three sweep modes can be selected for the A sweep; automatic, normal or single sweep. The horizontal sweep provides a maximum sweep rate of 0.1 microsecond/division (10 nanosecond/division using 10X magnifier) along with a delayed sweep feature for accurate relative-time measurements. can be made with Channel 2 providing the vertical deflection, and Channel 1 providing the horizontal [deflection. (TRIGGER switch set to CH 1 ONLY, HORIZ DISPLAY switch set to EXT HORIZ). The regulated DC power supplies maintain con-
The dual-channel DC-to-50 MHz vertical system
Accurate X-Y measurements
ELEC TRI C A L C H A RA C TERI STI C S
VERTICAL DEFLECTlON SYSTEM
stant output over a wide variation of line voltages and fre­quencies. approximately 90 watts.
Information given in this instruction monual applies to the Type R453 also unless otherwise noted. The Type R453 is electrically identical to the Type 453 but is mechanically adapted for mounting in a standard 19-inch rack. Rack­mounting instructions, a mechanical parts list and a dimen­sional drawing for the Type R453 are provided in Section
10 of this manual.
The electrical characteristics which follow are divided into two categories. Characteristics listed in the Performance Requirement column are checked in the Performance Check and Calibration sections of this manual. Items listed in the Operational Information column are provided for reference use and do not directly reflect the measurement capabili­ties of this instrument. The Performance Check procedure given in Section 5 of this manual provides a convenient method of checking the items listed in the Performance Requirement column. The following electrical characteristics
apply over a calibration interval of 1000 hours at an am­bient temperature range of -15°C to +55°C, except as
otherwise indicated. Warm-up time for given accuracy is 20 minutes.
Total power consumption of the instrument is
. .
Characteristic
Deflection Factor
Deflection Accuracy
Variable Deflection Factor
Bandwidth at Upper -3 dB point (with or without P6010 Probe)
20 mV to 10 VOLTS/DIV
10 mV/DIV
5 mV/DIV
Channels land 2 cascaded
Risetime (calculated). With or
without P6010 Probe. 20 mV to 10 VOLTS/DIV
10 mV/DIV
5 mV/DIV
Channels 1 and 2 cascaded
Performance Requirement
5 millivolts/division to 10 volts/division in 11 calibrated steps for each channel. One millivolt/ division when Channel 1 and 2 are cascaded.
Within ±3Y% of indicated deflection with VARl­ABLE control set to CAL. Cascaded deflection factor uncalibrated.
Uncalibrated deflection factor at least 2.5 times the VOLTS/DIV switch indication. This provides a maximum uncalibrated deflection factor of 25 volts/division in the 10 volts position.
DC to 50 MHz or greater
DC to 45 MHz or greater
DC to 40 MHz or greater
DC to 25 MHz or greater
Less than 7 nanoseconds
Less than 7.8 nanoseconds
Less than 8.75 nanoseconds
Less than 14 nanoseconds
Operational Information
Steps in 1-2-5 sequence
With gain correct at 20 mV
Driven from 25-ohm source
Measured at one millivolt/division
Risetime calculated from bandwidth measurement using the formula:
Where:
= Risetime in nanoseconds.
t
r
BW = Bandwidth in megahertz.
1-1
TM 11-6625-1722-15
Characteristic
Input RC Characteristics
Maximum lnput Voltage
lnput Coupling Modes
AC Low-Frequency Response
(lower -3 dB point) Without probe
With P6010 Probe
Trace Shift Due to Input Gate
Current (at 25°C)
Vertical Display Modes
Chopped Repetition Rate
Attenuator Isolation
Common Mode Rejection Ratio
Linear Dynamic Range Useful
for Common-Mode Relection in ADD Mode
Polarity Inversion
Signal Delay Line
Low-Frequency
Vertical Linearity
Trace Drift (after 20 minute
warm up) 20 mV to 10 VOLTS/DIV
10 mV/DIV
5 mV/DIV
VERTICAL (cont)
Performance Requirement
AC or DC, selected by front-panel switch
Negligible
Channel 1 only Channel 2 only
Dual-troce, alternate between channels Dual-trace, chopped between channels Added algebraically
Approximately one-microsecond segments from each channel dispiayed at repetition rate of 500
kHz, ±20%.
Greater than 10,000:1, DC to 20 MHz
Greater than 20:1 at 20 MHz for common-mode signals less than eight times VOLTS/DIV switch setting.
Signal on Channel 2 can be inverted
Less than 0.15 division compression or expansion
of two division signal when positioned to vertical
extremes of display area
Operational Information
Typically 1 megohm (±2%), paralleled
by 20 pF (±3%)
600 volts DC + peak AC (one kilohertz
or less). Peak-to-peak AC not to exceed 600 volts.
Typicaily 1.6 Hz, Input Coupling switch
set to AC
Typically 0.16 Hz
With optimum GAIN frequency
Less than 10% incremental signal dis-
tortion for instantaneous input voltage
-10 or +10 times VOLTS/DIV
of switch setting
Approximately 140 nanoseconds
Includes CRT linearity. Measured with
one-kilohertz square wave.
Time
Typically less than Typically less than
0.03 division/hour
Typically less than
0.05 division/hour
Typically less than
0.08 division/hour
adjustment at low
Temperature
0.0075 division/de­gree C
Typically less than
0.0125 division/de­gree C
Typically less than
0.02 division/de-
gree C
Source
Coupling
Polarity
1-2
TRIGGERING (A AND B SWEEP)
Internal from displayed channel or from Channel
1 only Internal from AC power source External External divide by 10
AC
AC low-frequency reject AC high-frequency reject
DC
Sweep can be triggered from positive-going or
negative-going portion of trigger signal
TRIGGERING (cont)
TM 11-6625-1722-15
Characteristic
Internal Trigger Sensitivity
AC
LF REJ
HF REJ
DC
External Trigger Sensitivity
AC
LF REJ
HF REJ
DC
Auto Triggering (A sweep only)
Single Sweep (A sweep only)
Display Jitter
Maximum Input Voltage
External Trigger Input RC
Characteristics (approximate)
LEVEL Control Range
Performance Requirement
0.2 division of deflection, minimum, 30 Hz to 10 MHz; increasing to 1 division at 50 MHz
0.2 division of deflection, minimum, 30 Hz to 10
0.2 division of deflection, minimum, 30 kHz to 10
0.2 division of deflection, minimum, 30 Hz to 50 kHz
0.2 division of deflection, minimum, DC to 10 MHz; increasing to 1 division at 50 MHz
50 millivolts, minimum, 30 Hz to 10 MHz; increas­ing to 200 millivolts at 50 MHz
50 millivolts, minimum, 30 kHz to 10 MHz; increas­ing to 200 millivolts at 50 MHz
50 millivolts, minimum, 30 Hz to 50 kHz
50 millivolts, minimum, DC to 10 MHz; increas-
ing to 200 millivolts at 50 MHz
Stable display presented with signal amplitudes
given under Internal and External Trigger Sensi­tivity above 20 Hz. Presents a free-running sweep for lower frequencies or in absence of trigger signal.
A Sweep Generator produces only one sweep
when triggered. Further sweeps are locked out
until RESET button is pressed. Trigger sensitivity same as given above.
Less than 1 nanosecond at 10 nanoseconds/divi­sion sweep rate (MAG switch set to X10)
At least ±2 volts, SOURCE switch in EXT posi­tion. At least ±20 volts, SOURCE switch in EXT
÷10 position
Operational Information
Typical -3 dB point, 16 Hz
Typical -3 dB point, 16 kHz
Typical -3 dB points, 16 Hz and 100 kHz
Typical -3 dB point, 16 Hz
Typical -3 dB point, 16 kHz
Typical -3 dB points, 16 Hz and 100
kHz
600 volts DC + peak AC (one kilohertz
or less). Peak-to-peak AC not to exceed
600 volts.
1 Megohm paralleled by 20 pF, except in LF REJ
Sweep Rates
A Sweep
B sweep
Sweep Accuracy-A and B
Sweep 5 s to 0.1 s/DIV
50 ms to 0.1 µs/ DIV
Variable Sweep Rate
HORIZONTAL DEFLECTION SYSTEM
A and B Sweep Generator
0.1 microsecond/division to 5 seconds/division in
24 calibrated stem
0.1 microsecond/division to 0.5 second/division
in 21 calibrated steps
0°C to +40°C
Within ±3% of indi- Within ±5% of indi-
cated sweep rate
Within ±3% of indi- Within ±4% of indi-
cated sweep rate
Uncalibrated sweep rate to at least 2.5 times the
TIME/DIV indication, or a maximum of at least
12.5 seconds/division in the 5 s position (B sweep,
maximum of 1.25 seconds/division in the .5 s
position.
-15°C to +55°C
cated sweep rate
cated sweep rate
A sweep is main and delaying sweep
B sweep is delayed sweep
A VARIABLE and B TIME/DiV VARl­ABLE controls set to CAL
1-3
TM 11-6625-1722-15
A and B Sweep Generotor
Characteristic
Sweep Length
A sweep
B sweep
Sweep Hold-off-A sweep
5s to 10
Sweep Magnification
Magnified Sweep Accuracy 1% tolerance added to speclfled sweep accuracy
Magnified Sweep Linearity
Normal/Magnified Registration
Calibrated Delay Time Range
DELAY-TIME MULTIPLIER
Dial Range
Delay Time Accuracy
5s to 0.1 s/DIV
50 ms to 1
Incremental Multiplier Linearity
Delay Time Jitter
Variable from less than 4 divisions to 11.0, ±0.5 division
11.0 divisions, ±0.5 division
Less than one times the A TIME/DIV switch set-
ting
Less than 2.5 microseconds
Each sweep rate can be increased 10 times the Extends fastest sweep rate to 10 nano-
indicated sweep rate by horizontally expanding seconds/division the center division of display
±1.5% for any eight division portion of the total magnified sweep length (excluding first
and last 60 nanoseconds of magnified sweep)
±0.2 division, or less, trace shift at graticuie center when switching MAG switch from X10 to OFF
Continuous from 50 seconds to 1 microsecond A VARIABLE control set to CAL for indi-
0.20 to 10.20
0°C to +40°C -15° C to +55° C
Within ±2.5% of indi­cated delay cated delay Includes incremental multiplier linearity
Within ±1.5% of indi­cated delay cated delay
±0.2%
Less than 1 part in 20,000 of 10 times A TIME/ DIV switch setting
Performance Requirement
Sweep Magnifier
Sweep Delay
Within ±3.5% of indi-
‘Within ±2% of indi-
±0.3%.
Operational Information
A TIME/DIV switch set to 1 ms
B TIME/DIV switch set to 1 ms
cated delay
Equal to 0.5 division, or less, with the A TIME/DIV switch set to 1 ms and the B TIME/DIV switch set to 1
Input to Channel 1 (TRIGGER
switch in CH 1 ONLY) Deflection factor
Accuracy
X Bandwidth at Upper -3 dB
Point
Input RC characteristics
Phase difference between X
and Y amplifiers at 50 kHz
Input to EXT HORIZ Connector
Deflection factor
1-4
External Horizontal Amplifier
5 millivolts/division to 10 volts/division in 11 cali­brated steps
0°C to +40°C
Within ±5% of indi- Within ±8% of indi-
cated deflection
5 MHz or greater
Less than 3°
B SOURCE switch in EXT; 270 millivolts/division, ±15%. B SOURCE switch in EXT
±20%
cated deflection
Steps in 1-2-5 sequence.
Channel 1 VARIABLE control does not affect horizontal deflection
With external horizontal gain correct
at 20 mV
Typically 1 megohm (±2%), paralleled
by 20 pF (±3%)
Ex t e r na l H o r iz o n ta l A m p l i fi e r (c ont)
X Ba nd w id th a t Up p er -3 5 M Hz o r g re a te r
TM 11-6625-1722-15
Operational Information
Input RC c h ara c te ristic s
(approxima te]
Ph a se d if fe re n c e b etw e e n X
and Y amplifiers at 50kHz
CALIBRATOR
Waveshap e Po la rit y Output Voltage
Output Current
Sq u a r e w a v e Po sit ive g o in g w it h b a se line a t ze ro v o lt s
0.1 vo lt o r 1 volt, p ea k to p eak
5-milliamp ere s through PROBE LOO P on side
Re p e t i t i o n Ra t e
Volta ge Accuracy Current Accuracy Re p e t it i o n Ra t e A c c u r a c y Ri se t im e Duty C ycIe
.-
±1.5%
±0.5%
Le ss t ha n 1 m ic r o se c o n d
49% to 51%
Output Resistance
Z A XI S I NPUT
Se n s i t i v i t y
5 v olt p ea k-to-p ea k signal p rod uc e s no tic e able modulation
Usa b le Freq ue n c y Ra ng e D C t o g re ate r t ha n 50 M Hz Input Re sista nc e a t DC Input C o up ling DC coupled Po la rity o f O p e ra tio n
1 me gohm, para llele d by 20 pF
Le ss t h a n 3 °
Selected by CALIBRATOR switch on side
panel
Approximately 200 ohms in 1 V position. Approximately 20 ohms in .1 V position.
Approximately 47 kilohms
Po sitiv e-g o in g inp u t sig n a l d e c re a se s trac e intensity Neg ative-g oing signal inc rea ses trace
Maximum Input Voltage
A and B Ga te
Waveshap e Amplitude Po la rity Duration
Output resistance
Vertica l Signal Out (CH 1 only)
Output voltage
Ba n d w id t h
Output coupling
Output resistance
O UTPUT SIGN ALS
Re c t a n g u l a r p u l se
Po slt lve-g o in g w it h b a se line a t a b o u t - 0.7 v o lt s.
I Same d ura tio n a s the re sp ec t ive sw eep
I
25 m illivo lts, o r g rea ter/ division o f CRT d isplay int o 1 meg oh m load .
DC to 25 MHz or g rea te r when ca sca ded with
Channel 2 or into 50-ohm load.
DC coup led
200 vo lts c om bined DC and pea k AC
A GATE duration variable between about 4 and 11 times the A TIME/DIV sw i tc h se t t in g
with the A SWEEP
LENG TH c o n t ro l . Approximately 1.5 kilohms
Approximately 50 ohms
1-5
TM 11-6625-1722-15
Characteristic
Line Voltage
Voltage Ranges (AC, RMS)
115-volts nominal
230-volts nominal
Line Frequency
Maximum Power Consumption
at 115 Volts, 60 Hz
Tube Type
Phosphor
Accelerating Potential
Graticule
Type
Area
Illumination
Unblinking
Raster Distortion
Trace Finder
POWER SUPPLY
Performance Requirement
115 volts nominal or 230 volts nominal
90 to 110 volts 104 to 126 volts 112 to 136 volts
180 to 220 volts 208 to 252 volts 224 to 272 volts
——— 48 to 440 Hz
CATHODE-RAY TUBE (CRT)
I
Internal
Six divisions vertical by 10 divisions horizontal.
Each division equals 0.8 centimeter.
0.1 division or less total
Limits display within graticule area when pressed.
Operational Information
.—
Line voltage and range selected by Line Voltage Selector assembly on rear panel. Voltage ranges apply for wave­form distortion which does not reduce the peak line voltage more than 5% below the true sine-wave peak value.
92 watts (105 volt-amperes)
Tektronix T4530-31-1 rectangular
P31 standard. Others available on
special order.
Approximately 10 kV total (cathode potential -1.95 kV).
Variable edge lighting
Bias-type, DC coupled to CRT grid.
Adjustable with Geometry and Y Axis Align adjustments.
Characteristic
Temperature
Operating
Non-operating
Altitude
Operating
Non-operating
Humidity
Non-operating
Vibration
Operating and non-operating
Shock
Operating and non-operating
ENVIRONMENTAL CHARACTERISTICS
The following environmental test limits apply when tested in accordance with the recom­mended test procedure. This instrument will meet the electrical characteristics given in this section following environmental test. including failure criteria, etc., may be obtained from Tektronix, Inc. Contact your local Tektronix Field Office or representative.
Performance Requirement
-15°C to +55°c
-55° to +75°C
15,000 feet maximum
50,000 feet maximum
Five cycles (120 hours) of Mil-Std-202C, Method 106B
15 minutes along each of the three major axes at a total displacement of 0.025-inch peak to peak (4 g at 55 c/s) with frequency varied from 10-55­10 c/s in one-minute cycles. Hold at 55 c/s for three minutes on each axis.
Two shocks of 30 g, one-half sine, 11 millisecond
duration each direction along each major axis.
Complete details on environmental test procedures,
Supplemental Information
Fan at rear circulates air throughout ins­trument. cutout protects instrument from over­heating.
Derate maximum operating tempera-
ture by 1°C/1000 feet change in altitude above 5000 feet.
Exclude freezing and vibration
Instrument secured to vibration platform during test. Total vibration time, about 55 minutes.
Guillotine-type shocks. Total of 12 shocks
Automatic resetting thermal
1-6
EN V I RO N M EN TA L C HARAC TERI STIC S (c ont)
TM 11-6625-1722-15
Characteristic
Tr a n s p o r t a t i o n
Pa c ka g e v ib ra t io n
Pa c ka g e d ro p
Ty p e 4 5 3 Ty p e R 4 5 3
MECHANICAL CHARACTERISTICS
Characteristic
Construction
Chassis
Pa n el
Cabinet Circuit boards
Overall Dimensions, Type
453 (m easured ot m axi­mum points) He ig ht
Width
Le n g t h
Overall Dimensions, Type
R4 53 ( m e a su r e d a t m a x i­mum points)
He ig ht
Pe rf orm a nc e Re q u ire m e nt
Meets National Safe Transit type of test when
packaged as shipped from Tektronix, Inc. One hour vibration slightly in excess of 1 g.
30-inc h drop o n any corne r, e dge o r fla t surface.
18-inc h drop on a ny c orner, e d ge or fla t surface.
Width
Info rmation
Le n g t h
Aluminum alloy Aluminum alloy with ano-
dized finish Blu e v in y l- c o a t e d a l um i n um Glass-epoxy laminate
Connectors
Z A XIS I NP UT
All other connec tors
Net Weight
Ty p e 4 5 3 ( i n c l u d e s f r o n t
cover without accessor-
Ty p e R 4 5 3 ( w i t h o u t a c -
hand le positioned for ca rry­ing .
St a n d a r d a c c e s so r i e s s u p p l i e d w i t h t h e Ty p e 4 5 3 a n d R4 5 3 are listed on the last pullout page of the Mechanical Parts List illu st ra t io ns.
7 inches
Operational Information
Pa c ka ge sho uld just l ea ve vib ra t io n sur­face
19 inc hes
panel;
Bin d i ng p o st
BN C
Approximately 29 pounds.
ies)
Approximately 32 pounds.
cessories)
STANDA RD A C C ESSO RIES
1-7

SECTION 2

OPERATING INSTRUCT IONS
General
O effectively use the Type 453, the operation and capa-
T bilities of the instrument must be known. This section de­scribes the operation of the front-, side- and rear-panel controls and connectors, gives first time and general operat­ing information and lists some basic applications for this instrument.
Front Cover and Handle
The front cover furnished with the Type 453 provides a dust-tight seal around the front panel. Use the cover to protect the front panel when storing or transporting the instrument. The cover also provides storage space for probes and other accessories (see Fig. 2-1).
TM 11-6625-1722-15
Fig. 2-1. Accessory storage provided in front cover.
The handle af the Type 453 can be positioned for carrying
or as a tilt-stand for the instrument. To position the handle,
press in at both pivot points (see Fig. 2-2) and turn the handle to the desired position. Several positions are pro­vided for convenient carrying or viewing. The instrument
may also be set an the rear-panel feet for operation or storage.
Operating Voltage
The Type 453 can be operated from either a 115-volt or
a 230-volt nominal line-voltage source. The Line Voltage
Fig. 2 -2 . Handle positione d to provide a sta nd for t he instrum ent
Selector assembly on the rear panel converts the instrument from one operating range to the other. In addition, this assembly changes the primary connections of the power transformer to allow selection of one of three regulating ranges. The assembly also includes the two line fuses. When the instrument is converted from 115-volt to 230-volt nominal operation, or vice versa, the assembly connects or discon­nects one of the fuses to provide the correct protection for the instrument. Use the following procedure to convert this instrument between nominal line voltages or regulating ranges.
1. Disconnect the instrument from the power source.
2. Loosen the two captive screws which hold the cover onto the voltage selector assembly; then pull to remove the cover.
3. To convert from 115-volts nominal to 230-volts nomi­nal line voltage, pull out the Voltage Selector switch bar (see Fig. 2-3]; turn it around 180° and plug it back into the remaining holes. Change the line-cord power plug to match the power-source receptacle or use a 115- to 230-volt adapter.
4. To change regulating ranges, pull out the Range Selector switch bar (see Fig. 2-3); slide it to the desired position and plug it back in. Select a range which is cen­tered about the average line voltage to which the instru­ment is to be connected (see Table 2-1).
5. Re-install the cover and tighten the two captive screws.
6. Before applying power to the instrument, check that the indicating tabs on the switch bars are protruding through the correct holes for the desired nominal line voltage and
regulating range.
2-1
TM 11-6625-1722-15
Fig. 2-3. Line Voltage Selector assembly on the rear panel (shown with cover removed).
CAUTION
The Type 453 should not be operated with the Voltage Selector or Range Selector switches in the
wrong positions for the line voltage applied. Operation of the instrument with the switches in the wrong positions may either provide incorrect aperotion or damage the instrument.

TABLE 2-1

more frequently. The air filter should be cleaned occasion-
ally to aII
OW the maximum amount of cooling air to enter
the instrument. Cleaning instructions are given in Section 4.
The Type 453 can be operated where the ambient air
temperature is between
-15°C and +55°C. Derate the maximum operating temperature 1°C for each additional 1000 feet of altitude above 5000 feet. This instrument can be stored in ambient temperatures between –55°C and
+75°C. After storage at temperatures beyond the operating limits, allow the chassis temperature to come within the operating limits before power is applied.
Rackmounting
Complete information for mounting the Type R453 in a
cabinet rack is given in Section 10 of this manual.
CONTROLS AND CONNECTORS
A brief description of the function or operation front-, side- and rear-panel controls and connectors (see Fig. 2-4). More detailed information is given
of the follows in this
section under General Operating Information.
Cathode-Ray Tube
INTENSITY
FOCUS
Controls brightness of display.
Provides adjustment for a well-defined dis­play.
SCALE ILLUM
TRACE FINDER
Controls graticule illumination.
Compresses display within graticule area independent of display position or appli­ed signals.
Operating Temperature
The Type 453 is cooled by air drawn in at the rear and
blown out through holes in the top and bottom covers. Ade­quate clearance on the top, bottom and rear must be pro­vided to allow heat to be dissipated away from the instru-
ment. The clearance provided by the feet at the bottom and
rear should be maintained. If possible, allow about one inch of clearance on the top. Do not block or restrict the air flow from the air-escape holes in the cabinet.
A thermal cutout in this instrument provides thermal pro­tection and disconnects the power to the instrument if the internal temperature exceeds a safe operating level. Opera­tion of the instrument for extended periods without the covers may cause it to overheat and the thermal cutout to open
Vertical (both
VOLTS/DIV
VARIABLE
UNCAL
POSITION
GAIN
Input Coupling (AC GND DC)
channels except as noted)
Selects vertical deflection factor (VARl­ABLE control must be in CAL position for indicated deflection factor).
Provides continuously variable deflection factor between the calibrated settings of
the VOLTS/DIV switch.
Light indicates that the VARIABLE control
is not in the CAL position.
Controls vertical position of trace.
Screwdriver adjustment to set gain of the Vertical Preamp. Line between adjust­ment and 20 mV VOLTS/DIV position in­dicates that gain should be set with VOLTS/DIV switch in this position.
Selects method of coupling input signal to Vertical Deflection System.
AC: DC component of input signal is
blocked. Low frequency limit -3 dB point) is about 1.6 hertz.
GND: Input circuit is grounded (does not
ground applied signal).
2-2
TM 11-6625-1722-15
Fig. 2-4. Front-, side- and rear-panel controls and connectors.
2-3
TM 11-6625-1722-15
DC: All components of the input signal are
passed to the Vertical Deflection System.
STEP ATTEN
BAL
INPUT
MODE
TRIGGER
INVERT (CH 2
only)
A and B Triggering (both where applicable)
EXT TRIG
INPUT
SOURCE
Screwdriver adjustment to balance Verti­cal Deflection System in the 5, 10 and 20 mV positions of the VOLTS/DIV switch.
Vertical input connector for signal.
Selects vertical mode of operation.
CH 1: The Channel 1 signal is displayed.
CH 2: The Channel 2 signal is displayed.
ALT: Dual trace display of signal on both
channels. Display switched at end of each sweep.
CHOP: Dual trace display of signal on
both channels. Approximately one-mi­crosecond segmerlts from each channel displayed at a repetition rate of about 500 kilohertz.
ADD: Channel 1 and 2 signals are alge-
braically added and the algebraic sum is displayed on the CRT.
Selects saurce of internal trigger signal from vertical system.
NORM: Sweep circuits triggered from dis-
played channel(s). Channel 1 signal
available at CH 1 OUT connector.
CH 1 ONLY: Sweep circuits triggered only
from signal applied to the Channel 1
INPUT connector. No signal available at CH 1 OUT connector. CH 1 lights, located beside A and B SOURCE switch­es indicate when the TRIGGER switch is in the CH 1 ONLY position.
Inverts the Channel 2 signal when pulled
out.
Input connector for external trigger signal. Connector in B Triggering section of front panel also serves as external horizontal input when HORIZ DISPLAY switch is in EXT HORIZ position and B SOURCE switch is in EXT position.
Selects source of trigger signal.
INT: Internal trigger signal obtained from
Vertical Deflection System. When CH
1 light is on, trigger signal is obtained
only from the Channel 1 input signal; when the light is off, the trigger signal
is abtained from the displayed chan-
nel(s). Source of internal trigger signal
is selected by the TRIGGER switch.
LINE: Trigger signal obtained from a sam-
ple of the line voltage applied to this
instrument.
EXT: Trigger signal obtained from an ex-
ternal signal applied to the EXT TRIG INPUT connector.
nal approximately 10 times.
CH 1
COUPLING
SLOPE
LEVEL
HF STAB Decreases display jitter for high-frequency
(A Trigger­ing only) sweep rates.
A and B Sweep
DELAY-TIME
MULTIPLIER
A SWEEP
TRIG’D
UNCAL A
OR B
A AND B
TIME/DIV AND DELAY TIME
Light indicates that the internal trigger sig­nal is abtained only from the signal con­nected to the Channel 1 INPUT connector (see TRIGGER switch).
Determines method of coupling trigger signal to trigger circuit.
AC: Rejects DC and attenuates signals be-
low about 30 hertz. Accepts signals between abaut 30 hertz and 50 mega­hertz.
LF REJ: Rejects DC and attenuates signals
below about 30 kilohertz. Accepts sig­nals between about 30 kilohertz and 50 megahertz.
HF REJ: Accepts signals between about 30
hertz and 50 kilohertz; rejects DC and attenuates signals outside the above range.
DC: Accepts all trigger signals from DC to
50 megahertz or greater.
Selects portion of trigger signal which
starts the sweep.
+: Sweep can be triggered from positive-
going portion of trigger signal.
-: Sweep can be triggered from negative­going portion of trigger signal.
Selects amplitude point on trigger signal
at which sweep is triggered.
signals. Has negligible effect at lower
Provides variable sweep delay between
0.20 and 10.20 times the delay time indi-
cated by the A TIME/DIV switch.
Light indicates that A sweep is triggered and will produce a stable display with correct INTENSITY and POSITION control settings.
Light indicates that either the A or B
VARIABLE control is not in the CAL posi-
tion.
A TIME/DIV switch (clear plastic flange)
selects the sweep rate of the A sweep
circuit for A sweep only operatian and selects the basic delay time (to be multi-
plied by DELAY-TIME MULTIPLIER dial setting) for delayed sweep operation.
B TIME/DIV (DELAYED SWEEP) switch selects sweep rate of the B sweep circuit
Attenuates external trigger sig-
2-4
TM 11-6625-1722-15
A VARIABLE
B SWEEP
MODE
HORIZ
DISPLAY
MAG
A SWEEP
MODE
for delayed sweep operation only. VARl­ABLE controls must be in CAL positions for calibrated sweep rates.
Provides continuously variable A sweep rate to at least 2.5 times setting of the A TIME/DIV switch. A sweep rate is cali­brated when control is set fully clockwise to CAL.
Selects B sweep operation mode.
TRIGGERABLE AFTER DELAY TIME: B
sweep circuit will not produce a sweep until a trigger pulse is received follow­ing the delay time selected by the DELAY TIME (A TIME/DIV) switch and
the DELAY-TIME MULTIPLIER dial.
B STARTS AFTER DELAY TIME: B sweep
circuit runs immediately following delay time selected by the DELAY TIME switch and DELAY-TIME MULTIPLIER dial.
Selects horizontal mode of operation.
A: Horizontal deflection provided by A
sweep.
B sweep inoperative.
A INTEN DURING B: Sweep rate deter-
mined by A TIME/DIV switch. An inten­sified portion appears on the sweep during the B sweep time. This position
provides a check of the duration and position of the delayed sweep (B) with respect to the delaying sweep (A).
DELAYED SWEEP (B): Sweep rate deter-
mined by B TIME/DIV switch with the delay time determined by the setting of the DELAY TIME (A TIME/DIV) switch and the DELAY-TIME MULTIPLIER dial. Sweep mode determined by B SWEEP
MODE switch.
EXT HORIZ: Horizontal deflection pro-
vided by an external signal.
Increases sweep rate to ten times setting of A or B TIME/DIV switch by horizontally expanding the center division of the dis­play. Light indicates when magnifier is
on.
Determines the operating mode for A sweep.
AUTO TRIG: Sweep initiated by the ap-
plied trigger signal using the A Trig-
gering controls when the trigger signal repetition rate is above about 20 hertz. For lower repetition rates or when there is no trigger signal, the sweep free runs at the sweep rate selected by the A TIME/DIV switch to produce a bright reference trace.
NORM TRIG: Sweep initiated by the ap-
plied trigger signal using the A Trig-
gering controls. No trace is displayed when there is no trigger signal.
RESET
A SWEEP
LENGTH
POSITION
FINE
1 kHz CAL
POWER ON
Side Panel
ASTIG
B TiME/DIV­VARIABLE
PROBE LOOP
A GATE
B GATE
CH 1 OUT
SINGLE SWEEP: After a sweep is display-
ed, further sweeps cannot be presented until the RESET button is pressed. Dis­play is triggered as for NORM opera­tion using the A Triggering controls.
When the RESET button is pressed (SIN­GLE SWEEP mode), a single display will be presented (with correct triggering) when the next trigger pulse is received. RESET light (inside RESET button) remains on until a trigger is received and the sweep is completed. RESET button must be pressed before another sweep can be presented.
Adiusts length of A sweep. In the FULL
position (clockwise detent), the sweep is about 11 divisions long. As the control is rotated counterclockwise, the length of A sweep is reduced until it is less than four divisions long iust before the detent in the fully-counterclockwise position is reach­ed. In the B ENDS A position (counter­clockwise detent), the A sweep is reset at the end of the B sweep to provide the fastest possible sweep repetition rate for delayed sweep displays.
Controls horizontal position of trace.
Provides more precise horizontal position adjustment.
Calibrator output connector.
Light: Indicates that POWER switch is
on and the instrument is connected to a line voltage source.
Switch: Controls power to the instrument.
Screwdriver adjustment used in conjunc­tion with the FOCUS control to obtain a well-defined display. Does not require readjustment in normal use.
Provides continuously variable sweep rate to at least 2.5 times setting of B TIME/DIV switch. B sweep rate is calibrated when control is set fully clockwise to CAL.
Current loop providing five-milliampere square-wave current from calibrator cir­cuit.
Output connector providing a rectangular
pulse coincident with A sweep.
Output connector providing a rectangular
pulse coincident with B sweep.
Output connector providing a sample of the signal applied to the Channel 1 lN-
PUT connector when the TRIGGER switch
is in the NORM position.
2-5
TM 11-6625-1722-15
CALIBRATOR
TRACE
ROTATION
Rear Panel
Z AXIS INPUT
Line Voltage
Selector
The following steps will demonstrate the use of the con­trols and connectors of the Type 453. It is recommended that this procedure be followed completely for familiariza­tion with this instrument.
Setup Information
1. Set the front-panel controls as follows:
CRT Controls
INTENSITY
FOCUS
SCALE ILLUM Counterclockwise
Vertical Controls (both channels if applicable)
VOLTS/DIV 20 mV
VARIABLE
POSITION Midrange
INPUT COUPLING
MODE
TRIGGER
INVERT
Triggering Controls (both A and B if applicable)
LEVEL
SLOPE
COUPLING
SOURCE
Sweep Controls
DELAY-TIME
MULTIPLIER
A and B TIME/DIV
A VARIABLE
B SWEEP MODE
Switch selects output voltoge of Calibrator.
1-volt or 0.1-volt square wave available.
Screwdriver adjustment to align trace with horizontal graticule lines.
Input connector for intensity modulation of the CRT display.
Switching assembly to select the nominal operating voltage and the line voltage range. line fuses.
Voltage Selector: Selects nominal operat-
Range Selector: Selects line voltage range
FIRST-TIME OPERATION
The assembly also includes the
ing voltage range (115V or 230V).
(low, medium, high).
Counterclockwise
Midrange
CAL
D C
CH 1
NORM
Pushed in
Clockwise (+)
+
AC
INT
0.20
.5 ms
CAL
B STARTS AFTER
DELAY TIME
HORIZ DISPLAY
MAG
POSITION
A SWEEP LENGTH
A SWEEP MODE
POWER
Side-Panel Controls
B TIME/DIV VARIABLE
CALIBRATOR
2. Connect the Type 453 to a power source that meets the voltage and frequency requirements of the instrument. If the available line voltage is outside the limits of the Line Voltage Selector assembly position (on rear panel), see Operating Voltage in this section.
3. Set the POWER switch to ON. Allow about five minutes warmup so the instrument reaches a normal operating tem­perature before proceeding.
CRT Controls
4. Advance the INTENSITY control until the trace is at the desired viewing level (near midrange).
5. Connect the 1 kHz CAL connector to the Channel 1
INPUT connector with a BNC cable.
6. Turn the A LEVEL control toward 0 until the display
becomes stable. Note that the A SWEEP TRIG’D light is on
when the display is stable.
7. Adiust the FOCUS control for a sharp, well-defined display over the entire trace length. (If focused display can-
not be obtained, see Astigmatism Adjustment in this section.)
8. Disconnect the input signal and move the trace with the Channel 1 POSITION control so it coincides with one of the horizontal graticule lines. the graticule line, see Trace Alignment Adjustment in this
section.
9. Rotate the SCALE ILLUM control throughout its range and notice that the graticule lines are illuminated as the
control is turned clockwise (most obvious with mesh or smoke­gray filter installed). Set control so graticule lines are illuminated as desired.
Vertical Controls
10. Change the CH 1 VOLTS/DIV switch from 20 mV to 5 mV. If the vertical position of the trace shifts, see Step Attenuator Balance in this section.
11. Set the CH 1 VOLTS/DIV switch to 20 mV and set the Channel 1 Input Coupling switch to AC. Connect the 1 kHz CAL connector to both the Channel 1 and 2 INPUT con­nectors with two BNC cables and a BNC T connector.
If the BNC cables and BNC T connector are not
available, make the following changes in the pro­cedure. Place the BNC jack post (supplied ac­cessory) on the 1 kHz CAL connector and connect
A
OFF
Midrange
FULL
AUTO TRIG
OFF
CAL
.1 V
If the trace is not parallel with
NOTE
2-6
TM 11-6625-1722-15
the Channel 1 and 2 INPUT connectors. Connect the probe tips to the BNC jack post. Set the CALI­BRATOR switch (on side-panel) to 1 V.
12. Turn the Channel 1 POSITION control to center the display. The display is a square wave, five divisions in amplitude with about five cycles displayed on the screen. If the display is not five divisions in amplitude, see Vertical
Gain Adjustment in this section.
13. Set the Channel 1 Input Coupling switch to GND ond
position the trace to the center horizontal line.
14. Set the Channel 1 Input Coupling switch to DC. Note that the baseline of the waveform remains at the center horizontal line (ground reference).
15. Set the Channel 1 Input Coupling switch to AC. Note that the waveform is centered about the center horizontal line.
16. Turn the Channel 1 VARIABLE control throughout its range. Note that the UNCAL light comes on when the VARl-
ABLE control is moved from the CAL position (fully clock­wise). The deflection should be reduced to about two divi­sions. Return the VARIABLE control to CAL.
17. Set the MODE switch to CH 2.
18. Turn the Channel 2 POSITION control to center the
display. The display will be similar to the previous display for Channel 1. Check Channel 2 step attenuator balance and gain as described in steps 10 through 12. The Channel 2
input Coupling switch and VARIABLE control operate as
described in steps 13 through 16.
19. Set both VOLTS/DIV switches to 50 mV.
20. Set the MODE switch to ALT and position the Channel
1 waveform to the top of the graticule area and the Chan-
nel 2 waveform to the bottom of the graticule area. Turn
the A TIME/DIV switch throughout its range. Note that the display alternates between channels at all sweep rates.
Triggering
25. Set the CALIBRATOR switch to 1 V. Rotate the A LEVEL control throughout its range. The display free runs at the extremes of rotation. Note that the A SWEEP TRIG'D light is on only when the display is triggered.
26. Set the A SWEEP MODE switch to NORM TRIG. Again rotate the A LEVEL control throughout its range. A display is presented only when correctly triggered. The A SWEEP TRIG'D light operates as in AUTO TRIG. Return the A SWEEP MODE switch to AUTO TRIG.
27. Set the A SLOPE switch to -. The trace starts on the negative part of the square wave. Return the switch to +; the trace starts with the positive part of the square wave.
28. Set the A COUPLING switch to DC. Turn the Chan­nel 1 POSITION control until the display becomes unstable (only part of square wave visible). Return the A COUPLING
switch to AC; the display is again stable. Since changing
trace position changes DC level, this shows how DC level
changes affect DC trigger coupling. Return the display to
the center of the screen.
29. Set the MODE switch to CH 2; the display should be
stable. Remove the signal connected to Channel 1; the dis-
play free runs.
Set the TRIGGER switch to NORM; the dis-
play is again stable. Note that the CH 1 lights in A and B
Triggering go out when the TRIGGER switch is changed to
NORM.
30. Connect the Calibrator signal to both the Channel 2
INPUT and A EXT TRIG INPUT connectors. Set the A
SOURCE switch to EXT. Operation of the LEVEL, SLOPE and COUPLING controls for external triggering are the same as described in steps 25 through 28.
31. Set the A SOURCE switch to EXT is the same as for EXT. Note that the A LEVEL control has less range in this position, indicating trigger signal attenua­tion. Return the A SOURCE switch to INT.
32. Operation of the B Triggering controls is similar to A Triggering.
21. Set the MODE switch to CHOP and the A TIME/DIV Note the switching between channels as
shown by the segmented trace. Set the TRIGGER switch to
CH 1 ONLY; the trace should appear more solid, since it is no longer triggered on the between-channel switching trans-
ients. Turn the A TIME/DIV switch throughout its range. A dual-trace display is presented at all sweep rates, but unlike ALT, both channels are displayed on each trace on a time­sharing basis. Return the A TIME/DIV switch to .5 ms.
22. Set the MODE switch to ADD. The display should be four divisions in amplitude. Note that either POSITION con­trol moves the display.
23. Pull the INVERT switch. The display is a straight line indicating that the algebraic sum of the two signals is zero (if the Channel 1 and 2 gain is correct).
24. Set either VOLTS/DIV switch to 20 mV. The square-
wave display indicates that the algebraic sum of the two
signals is no longer zero.
Return the MODE switch to CH 1
and both VOLTS/DIV switches to .2 (if using 10X probes,
set both VOLTS/DIV switches to 20 mV). Push in the INVERT switch.
Normal and Magnified Sweep
33. Set the A TIME/DIV switch to 5 ms and the MAG switch to X10. The display should be similar to that obtain­ed with the A TIME/DIV switch set to .5 ms and the MAG switch to OFF.
34. Turn the horizontal POSITION control throughout its range; it should be possible to position the display across the complete graticule area. Now turn the FINE control. The
display moves a smaller amount and allows more precise positioning. Return the A TIME/DIV switch to .5 ms, the
MAG switch to OFF and return the start of the trace to the left graticule line.
Delayed Sweep
36. Pull the DELAYED SWEEP knob out and turn it to 50
(DELAY TIME remains at .5 ms). Set the HORIZ DISPLAY switch to A INTEN DURING B. An intensified portion, about one division in length, should be shown at the start of the trace. Rotate the DELAY-TIME MULTIPLIER dial throughout its range; the intensified portion should move along the dis­play.
2-7
TM 11-6625-1722-15
37. Set the B SWEEP MODE switch to TRIGGERABLE
AFTER DELAY TIME. Again rolate the DELAY-TIME MULTI-
PLIER dial throughout its range and note that the intensified portion appears to jump between posltive slopes of the dis­play. Set the B SLOPE switch to -; begins on the negative slope.
the intensified portion
Rotate the B LEVEL control; the intensified portion of the display disappears when the B LEVEL control is out of the trlggerable range. Return the B LEVEL control to 0.
38. Set the HORIZ DISPLAY switch to DELAYED SWEEP (B). Rotate the DELAY-TIME MULTIPLIER dial througout its range; about one-half cycle of the waveform should be
displayed on the screen (Ieading edge visible only at high
INTENSITY control setting). The display remains stable on the screen, indicating that the B sweep is triggered.
39. Set the B SWEEP MODE switch to B STARTS AFTER
DELAY TIME. Rotate the DELAY-TIME MULTIPLIER dial
throughout its range; the display moves continously across the screen as the control is rotated.
40. Rotate the DELAYTIME MIULTIPLIER dial fully counter-
clockwise and set the HORIZ DISPLAY switch to A INTEN
DURING B. Rotate the A SWEEP LENGTH control counter-
clockwise; the Iength of the display decreases. Set the con-
trol to the B ENDS A position; now the display ends after the intensified portion. Rotate the DELAY-TIME MULTIPLIER dial ond note that the sweep length increases as the display moves across the screen. Return the A SWEEP LENGTH con-
trol to FULL and the HORIZ DISPLAY switch to A.
Single Sweep
41. Set the A SWEEP MODE switch to SINGLE SWEEP. Remove the Calibrator signal from the Channel 2 INPUT connector. Press the RESET button; the RESET light should come on and remain on. Again apply the signal to the Chan­nel 2 INPUT connector; a single trace should be presented
and the RESET light should go out. Return the A SWEEP
MODE switch to AUTO TRIG.
External Horizontal
42. Connect the Calibrator signal to both the Channel 2
INPUT and EXT HORIZ (B EXT TRIG lNPUT] connectors. Set
the B SOURCE switch to EXT, B COUPLING switch to DC and the HORIZ DISPLAY switch to EXT HORIZ. lncrease the
INTENSITY control setting until the display is visible (two dots
displayed diagonally). The display should be five divisions vertically and about 3.7 divisions horizontally. Set the B
10. The display should be reduced ten times horizontally. The display can be positioned hori­zontally with the horizontal POSITION or FINE control and vertically with the Channel 2 POSITION control.
43. Connect the Calibrator signal to both the Channel 1 and 2 INPUT connectors. Set the TRIGGER switch to CH 1 ONLY and the B SOURCE switch to INT.
44. The display should be five divisions vertically and horizontally.
The display can be postioned horizontally
with the Channel 1 POSITION control and vertically with the
Channel 2 POSITION control.
45. Change the CH1 VOLTS/DIV switch to 5. The display play is reduced to two divisions horizontally. Now set the CH 2 VOLTS/DIV switch to 5. The display is reduced to two divisions vertically.
Trace Finder
46. Set the CH 1 and CH 2 VOLTS/DlV switches to 10
mv. The display i
S not visible since it exceeds the scan area
of the CRT.
47. Press the TRACE FINDER button. Note that the dis­is returned to the display area. While holding the
play TRACE FINDER button depressed, increase the vertical and horizontal deflection factors until the display is reduced to about two divisions vertically ond horizontally. Adjust the Channel 1 and 2 POSITION controls to center the display
about the center lines of the graticule. Release the TRACE
FINDER and note that the display remains within the viewing area. Disconnect the applied signal.
48. Reduce the INTENSITY control setting to normal, B SOURCE switch to INT and set the HORIZ DISPLAY switch to A.
Z-Axis Input
49. If an, External signal is available (five volts peak to peak minimum] the function of tlhe Z AXIS INPUT circuit can be demonstrated. Connect the external signal to both the
Channel 2 INPUT connector and the Z AXIS INPUT binding posts. Set the A TIME/DIV switch to display about five cycles of the waveform. The positive peaks of the waveform should be blanked and the negative peaks intensified, indi-
cating intensify modulation.
50. This completes the basic operating procedure for the
Type 453. lnstrument operation not explained here, or opera­tions which need further explanation are discussed under General Operating Information.
CONTROL SETUP CHART
Fig. 2-5 shows the front, side and rear panels of the Type
453. This chart can, be reproduced and used as a test-setup record for special measurements, applications or procedures, or it may be used as a training aid for familiarization with this instrument.
GENERAL OPERATING INFORMATION
Intensify Control
The setting of the INTENSITY control may affect the cor­rect focus of the display. Slight readjustment of the FOCUS control may be necessary when the intensity level is changed.
To protect the CRT phosphor, do not turn the INTENSITY control higher than necessary to provide a satisfactory dis-
play. The light filters reduce the observed light output from
the CRT. When using these filters, avoid odvancing the
INTENSITY control to a setting that may bum the phosphor.
When Ihe highest intensity display is desired, remove the filters and use the clear faceplate protector. Also, be care­ful that the INTENSITY control is not set too high when changing the TIME/DlV switch front a fast to a slow sweep
rate, or when changing the HORIZ DISPLAY switch from EXT HORIZ operation to the norrmal sweep mode.
Astigmatism Adjustment
If a well-defined trace cannot be obtained with the FOCUS
control, adjust the ASTIG adjustment (side panel) as fol-
lows.
2-8
TM 11-6625-1722-15
Fig. 2-5. Control setup chart for the Type 453.
2-9
TM 11-6625-1722-15
NOTE
To c h e c k f o r p r o p e r s e t t i n g o f t h e A STI G a d j u st ­ment, slowly turn the FOCUS control through the optimum setting. If the ASTIG adjustment is cor-
re c tly set, the vert ic a l a n d h orizo nt a l p o rtio ns of the trace will come into sharpest focus at the sa m e p o si t io n o f t h e FO C US c o n t ro l . This se t t ing of the ASTIG adjustment should be correct for any display. However, it may be necessary to reset the FOCUS control slightly when the INTENSITY control is changed.
1. Co nnec t a 1 V Ca libra tor sig na l to e ithe r channe l a nd se t t h e V O LTS/ D IV sw i t c h o f t h a t c h a n n e l t o p r e se n t a t w o ­division display. Set the MODE switch to display the chan-
nel selec ted.
2. Set the TIME/DIV sw itc h to .2 ms
3. With the FOCUS c ontro l an d ASTIG a d just me nt set to
midra ng e, a djust the INTENSITY c ontro l so the rising portio n
of the display c an be seen.
4. Set the ASTIG ad justment so the ho rizo ntal a nd verti­cal portions of the display are equally focused, but not
nec essarily well focused.
5. Se t the FOCUS c ontrol so the vertic al p ortion of the
trac e is as thin as possible.
6. Repe a t step s 4 an d 5 fo r b est overa ll foc us. Make fina l
check at normal intensity.
Graticule
Th e g r a t i c u l e o f t h e Ty p e 4 5 3 i s i n t e r n a l l y m a r k e d o n t h e faceplate of the CRT to p rovid e acc urate, no -para lla x meas­urements. The graticule is marked with six vertic al and 10 horizontal divisions. Each division is 0.8 centimeter sq uare. In a d dition, e ac h m a jo r d iv ision is d iv id ed int a five m inor divisions at the center vertical and horizontal lines. The
vertic al gain and horizonta l timing are calibrated to the
graticule so accurate measurements can be made from the CRT. The illumination of the graticule lines can be varied with the SCALE ILLUM control.
Fi g . 2 - 6 sh o w s t h e g r a t i c u l e o f t h e Typ e 4 53 a n d d e f i n e s the various mea surement lines. The terminology defined here will be used in all discussions involving graticule measure­ments.
Fi g . 2 - 6 . D e f i n i t i o n o f m e a s u re m e n t l in e s o n Ty p e 4 5 3 g r a t i c u l e .
re mov e the f ilte r, p ress
down at the bottom of the frame
and pull the top of the filter away from the CRT faceplate
(see Fig. 2-7).
Th e t i n t e d l i g h t f i l t e r m i n i m i z e s l i g h t r e f l e c t i o n s f r o m t h e face of the CRT to imp ro ve c ontrast when viewing the dis­play under high ambient light conditions. A clear plastic faceplate protector is a lso p rovided w ith this instrument for use when neither the me sh nor the tinted filter is used. The clear faceplate protector provides the best display for wave­form p hotographs. It is a lso preferable for view ing high writing rate displays.
A filter or the facepla te protector should be used at all times ta protec t the CRT fac eplate from scratc hes. The face­plate protector and the tinted filter mount in the same holder.
Tra c e A l i g n m e n t A d j u st m e n t
If a free-runn ing t ra c e is not p a ra lle l t o the ho rizo nt a l
graticule lines, set the TRACE ROTATION adjustment as fol-
lows. Positio n the tro ce to th e c en te r ho rizonta l line . Ad just
the TRACE ROTATION adjustment (side panel) so the trace
is para llel w ith th e ho rizonta l grat ic ule line s.
Li g ht Fi lte r
Th e m e sh f i l t e r p r o v i d e d w i t h t h e Ty p e 4 5 3 p r o v i d e s s h i e l d ­ing a gain st ra d iate d EM I (e lectro-mag ne tic int erferen c e) fro m the face of the CRT. It also serves as a light filter to make the trace more visible und er ambient light conditions. To
2-10
Fi g . 2 - 7. Re m o v i n g t h e f i lt e r o r f a c e p l a t e p r o t e c t o r .
TM 11-6625-1722-15
To r e m o v e t h e l i g h t f i l t e r o r f a c e p l a t e p r o t e c t o r f r o m t h e
holder, press it out to the rear. They can be replaced by sn a p p i n g t h e m b a c k i n t o t h e h o l d e r.
Tr a c e Find e r
Th e TR A C E FI N D ER p r o v i d e s a m e a n s a f l o c a t i n g a d i s p l a y which overscans the viewing area either vertically or hori­zo nta l ly. W he n t h e TRA CE FIND ER b u t to n is p re sse d , t he display is compressed within the graticule area. T
O locate
and reposition an overscanned display, use the following
procedure.
1. Pre ss the TRA CE FINDER button.
2. While the TRACE FINDER b ut to n is held d epre ssed , inc rease the vertic al a nd ho rizo nta l deflec tio n facto rs until the vertic al deflection is reduced to a bout two divisions and the horizontal d eflection is reduced to about four d ivisions (the horizo ntal deflec tion need s to be reduced o nly when in the external horizontal mode of op eration).
3. Adjust t he v ertic al a nd horizonta l POSITION c ontrols to c enter the d isp la y ab out the vertical and horizontal c en-
ter lines.
4. Re lea se the TRA CE FINDER button; the displa y sho uld
re main within the view ing area .
Vertical Channel Se lection
Ei t h e r o f t h e i n p u t c h a n n e l s c a n b e u se d f o r si n g l e - t r a c e displays. Apply the signal to the desired INPUT connector and set the MODE switch to display the channel used. How­ever, since CH 1 ONLY triggering is provided only in Chan-
nel 1 and the invert feature only in Cha nnel 2, the correct channel must be selected to take advantage of these fea­tures. For d ual-trace displays, c onnect the signals to both
INPUT c onn ec t ors a nd se t t he M ODE switc h to o ne o f the dual-trace positions.
Vertical Ga in Adjustment
To c h e c k t h e g a i n o f e i t h e r c h a n n e l , s e t t h e V O L TS/ D I V
sw i t c h t o 2 0 m V . Se t t h e C A L IBRA TO R sw i t c h t o . 1 V a n d
connect the 1 kHz CAL connector to the INPUT of the chan-
nel used . The vertica l deflec tion should be exac tly five divisions.
If not, a d just t he f ront -p ane l G AIN a d just m ent
for exac tly five divisions of d eflec tion.
NOTE
If t he g a in o f the t wo c ha nne ls m ust b e c lo se ly matched (such as for ADD mode operation), the ADJUSTMENT p roc edure given in the Calibration se c t i o n sh o u ld b e u se d .
Th e b e s t m e a s u r e m e n t a c c u r a c y w h e n u s i n g p r o b e s i s p r o ­vided if the GAIN adjustment is made with the p robes in­st a l l e d ( se t t h e C A LIB RA TO R sw i t c h t o 1 V ) . A l so , t o p r o v id e the most accurate measurements, ca librate the vertical g ain
of the Type 453 at the temperature at which the measurement
is to be ma de.
Ste p A tt e nua t o r Ba l a nc e
To c h e c k t h e s t e p a t t e n u a t o r b a l a n c e o f e i t h e r c h a n n e l , se t t h e In p u t C o u p l i n g sw i t c h t o G N D a n d se t t h e A SW EEP
MODE swic h to AUTO TRIG to provide a free-running trace. Change the VOLTS/DIV switc h from 20 mV to 5 mV. If the trac e moves vertically, adjust the front-panel STEP ATTEN
BA L a d j u st m e n t a s f o ll o w s ( a llo w a t l e a st 1 0 m i nut e s w a rm up b efore performing this a djustment].
1. With th e Inp ut Coupling switch se t to GND a nd the
VOLTS/ DIV sw itc h set to 20 mV, m ove the tra ce to the center
horizontal line of the graticule with the vertica l POSITION control.
2. Se t the VOLTS/ DIV switc h to 5 m V a nd a djust t he STEP A TTEN B A L a d j u st m e n t t o r e t u r n t h e t r a c e t o t h e c e n t e r horizonal line.
3. Re c he ck step a ttenuator bala nc e and rep eat adjustm ent until no trac e shift oc curs a s the VOLTS/ DIV switc h is cha ng­ed from 20 mV to 5 mV.
Sig na l C o nn e c ti o ns
In gen era l, p rob es o ff er th e m o st c on ve nient me a ns o f c o n­nec ting a signal to the input of the Type 453. The Tektronix probes are shielded to prevent pickup of electrostatic inter­ferenc e. A 10X a ttenuator probe offe rs a high input imped ­ance and allows the circuit under test to perform very close to normal operating c onditions. However, a 10X p robe
also attenuates the input signal 10 times.
Th e Te k t r o n i x P60 45 Fie ld Ef fe c t Tra n sisto r p ro b e a n d a c c e ssory p o w er su p p l y o f f e r t h e sa m e h i g h - i n p u t i m p e d a n c e a s t h e 1 0 X probes. However, it is particularly useful since it provides
wide-band operatian while presenting no attenuation (1X
gain) and a low input capacitance. To obtain maximum bandwidth when using the probes, observe the grounding considerations given in the probe manual. The probe-to­connector adapters and the bayonet-ground tip provide the best frequency response. Remember that a ground strap only a few inches in length can produc e several percent of ringing wh en opera ting at th e hig he r fre q ue nc y lim it o f this
sy st e m . Se e y o u r Te kt r o n i x, I n c . c a t a l o g f o r c h a r a c t e ri st i c s
and compatibility of probes for use with this system.
In hig h-freque nc y a p p lic a tions re q uiring ma ximu m ove ra ll bandwidth, use coaxial cables terminated at both ends in their characteristic impedanc e. See the discussion on coa x­ia l cab les in th is sec tio n fo r more info rma tio n.
Hig h-lev el, lo w-fre que nc y sig na ls c a n b e c onnec ted directly to the Type 453 INPUT connectors with short unshield ed leads. Th is c ou p ling m et ho d w orks b est fo r sig na ls b elow about one kilohertz and deflection foctors above one volt/ division. When this method is used, establish a common ground between the Type 453 and the equipment under test. Attempt to position the leads away from any source of inter­ferenc e to avoid errors in the display. If interference is excessive with unshielded leads, use a coaxial cable or a probe.
Lo a d in g Ef fe c t o f t he Ty p e 4 5 3
As nearly as possible, simulate actual operating c ondi­tions in the equipment under test. Otherwise, the equipment under test may not produce a normal signa l. The 10X attenuator probe and field effect transistor probe mentioned previously offer the least circuit loading. See the probe instru c tio n man ua l fo r load ing cha ra c te ristics o f t he ind ivi­dual probes.
2-11
TM 11-6625-1722-15
When the sig nal is c oupled directly to the input of the
Ty p e 4 5 3 , t h e i n p u t i m p e d a n c e i s a b o u t o n e m e g o h m
paralleled by about 20 pF. When the signal is coupled to the input through a c oaxia l c ab le, the effeclive input ca­pacitance depends upon the type and Iength of cable used. Se e t h e f o l l o w i n g d i s c u ss i o n f o r
inf orrna tio n o n obt a ining
maximum frequency responspe with coaxial cables.
Th e s i g n a l c a b l e s u s e d t o c o n n e c t t h e s i g n a l 1 0 t h e t y p e 453 INPUT conn ec tors ha ve a Ia rg e effec t on t he ac curac y of the displayed high-frequency waveform.
To m a i t a i n t h e
high-frequenc y characteristics of the app lied signal, high­quality low-loss coaxial cable should be used. The cable sh o u l d b e t e r m i n a t e d a t t h e Ty p e 4 5 3 IN PU T c o n n e c t o r in i t s characteristic impedonce. with differing characteristic impedances,
If it is n ec e ssary t o u se c a b les
use suitab le imped-
ance-matching devices to provide the correct transition, with
minimum loss, from one impedance to the other.
Th e c h a r a c t e r i s t i c i m p e d a n c e , v e l o c i t y o f p r o p a g a t i o n a n d nature of signa l lOSSeS in a c oa xial c ab le a re det ermined by the physical and electrical characteristics of the cable. Lo sse s c a u se d b y e n e rg y d i ssip a t i o n in t h e d i e le c t ric a re
proportional to the signal frequency. Therefore, much of the high-frequency information in a fast-rise p ulse can be lost in o nly a few fee t of intercon ne c tin g c able if it is no t the c orrec t type. To be sure of the high-frequency resp onse of the system when using cables longer than about five feet, observe the transient response of the Type 453 and the
int ercon ne c tin g c ab le with a fast-rise
pulse generator (gen-
erator risetime Iess than 0.5 nanoseconds).
DC c om po nents. The pre -charging ne twork inco rpora te d in this u nit a II
OWS the inp ut-c oupling cap ac itor to charge to
the DC source voltage level when the Inp ut Coupling switc h is se t to G ND. Th e p ro c ed ure f or using this fe ature is a s
follows:
1. Be fore c onn ec ting the signa l c ontaining a DC c om po­nent to the Typ e 453 INPUT connector, set the Input Coupling sw i tc h t o G N D. Th e n c o n n e c t t h e si g n a l t o t h e IN PUT connector.
2. Wa it a bout one se c ond for the coupling c apa citor to
charge.
3. Se t the In put C ouplin g switc h to A C. Th e t rac e (dis­play) will remain on the screen and the AC component of the signal c an be measured in the normal ma nner.
Deflection Fa c tor
Th e a m o u n t o f v e r t i c a l d e f l e c t i o n p r o d u c e d b y a si g n a l is d eterm ined by the sig na l a m plitude , th e attenua tio n fac to r of the probe (if used), the setting of the VOLTS/DIV switch and the setting of the VARIABLE VOLTS/DIV control. The
calibrated deflection factors indicated by the VOLTS/DIV sw i t c h e s a p p l y o n l y w h e n t h e V A RI A BLE c o n t r o l i s se t t o the CAL position.
Th e V A R I A B L E V O L TS/ D I V c o n t r o l p r o v i d e s v a r i a b l e (unc alibrated) vertical deflection between the c alibrated se t t i n g s o f t h e V O LTS/ D IV sw i t c h . Th e V A RIA BLE c o n t r o l
extends the maximum vertical deflec tion factor of the Type 453 to a t lea st 25 volts/ divisio n (10 volts po sition).
Input Coupling
Th e C h a n n e l 1 a n d 2 l n p u t C o u p l i n g sw i t c h e s a l l o w a
choice of input caupling. The type of display desired will determine the coupling used.
Th e D C p o s i t i o n c a n b e u s e d f o r m o st a p p l i c a t i o n s . H o w ­ever, if the DC component of the signal is much larger than the AC component, the AC p osition vvill probab ly provide a better display. DC coupling should be used to display AC si g n a l s b e l o w a b o u t 1 6 h e l t z a s t h e y w i l l b e a t t e n u a t e d i n
the AC position.
In t he A C p o sitio n, t he D C c o mpone nt o f t he sig nal is blocked by a capacitor in the input circuit. The low-fre­quency response in the AC position is about 1.6 hertz (–3 dB
point). Therefore,
so m e l o w - f re q u e n c y d i st o rt i o n c a n b e expected near this frequency limit. Distortion will also appear
in sq ua re w aves w hic h hav e low-f req ue nc y c om-
ponents.
Th e G N D p o s i t i o n p r o v i d e s a g r o u n d r e f e r e n c e a t t h e
inp ut of the Typ e 453. The sig na l app lied to the inp ut con-
nec tor is internally disc onnec ted but not grounded. The inp ut c irc uit is he ld at g rou nd p ote nt ial, e limina tin g th e ne ed to externally ground the input to establish a DC ground re fe renc e.
Th e G N D p o s i t i o n c a n a l s o b e u se d t o p r e - c h a r g e t h e coupling capacitor to the average voltage level of the signal applied to the INPUT connector. This allows measurement of only the AC component of signals having both AC and
Dua l-Tra ce O pe ration
Alte rnate Mode. Th e A L T p o s i t i o n o f t h e M O D E s w i t c h
produces a disploy which alternates between Channel 1 and 2 w ith eac h swe ep of t he CRT. A ltho ug h the ALT mod e can be used at all sweep rates, the CHOP mode provides
a more satisfactory display at sweep rates below about 50
microseconds/division. At these slower sweep rates, alternate mode switching becomes visually perceptible.
Pro p e r in t erna l trig g e rin g in t h e A LT m o d e c a n b e o b -
tained in either the NORM or CH 1 ONLY positions of the
TRI G G E R s w i t c h . W h e n i n t h e N O R M p o si t i o n , t h e s w e e p i s
triggered from the signal on each channel. This provid es a st a b l e d i sp l a y o f t w o u n re l a t e d sig n a l s, b u t d o e s n o t in d i c a t e the time relationship between the signals. In the CH 1 ONLY position, the two signals are displayed showing true time re la tio nship . If the sig na ls a re n ot tim e relate d , th e Ch a nn el 2 wa ve fo rm will b e unsta ble in the CH 1 ONLY po sition.
Chopped Mode. Th e C H O P p o si t i o n o f t h e M O D E sw i t c h
produces a display which is electronically switched between channels. In general, the CHOP mode provides the best display at sweep rates slower than about 50 microseconds/ division, or whenever dual-trace, single-shot phenomena are to b e displayed. At faster sweep rates the chopped sw itc h­ing bec ome s ap pare nt and m ay inte rfere with th e disp la y.
Pro p e r in tern al t rig g e rin g fo r t h e CHO P mo d e is p ro vid e d with the TRIGGER switch set to CH 1 ONLY. If the NORM position is used, the sweep circuits are triggered from the between-channel switching signal and both waveforms will
2-12
TM 11-6625-1722-15
be unstable. External triggering provides the same result as CH 1 ONLY triggering.
Tw o s i g n a l s w h i c h a r e t i m e - r e l a t e d c a n b e d i s p l a y e d i n the chop ped mode showing true time relationship. If the si g n a l s a r e n o t t i m e - r e la t e d , t h e C h a n n e l 2 d i sp l a y w i ll appear unstable. Two single-shot, transient, or random
si g n a l s w h i c h o c c u r w i t h in t h e t i m e i n t e rv a l d e t e rm i n e d b y
the TIME/ DIV switch (10 times sweep rate) c an be compared using the CHOP mod e. To correctly trigger the sweep for maximum resolution, the Channel 1 signal must precede the
Channel 2 signal. Since the signals show true time relation-
sh i p , t i m e - d i f f e re n c e m e a su r e m e n t s c a n b e m a d e .
Channel 1 Output and Cascaded Operation
If a lower def lec tio n fa c to r tha n provid e d b y t he VO LTS/ DIV switc hes is d esired, Cha nne l 1 c an b e use d a s a w ide ­band preamplifier for Channel 2. Apply the input signal to the Channel 1 INPUT connector. Connect a 50-ohm BNC cable (18-inch or shorter cable for maximum cascaded fre-
quency response) between the CH 1 OUT (side panel) and
the Channel 2 INPUT c onnec tors. Set the MODE switch to CH 2 and the TRIGGER switch to NORM. With both VOLTS/ DIV switches se t to 5 mV, the d eflection factor will be less than one millivolt/division.
To p r o v i d e c a l i b r a t e d o n e m i l l i v o l t / d i v i s i o n d e f l e c t i o n factor, c onnect the .1 volt Calibrator signa l to the Channel 1 INPUT c o nn ec t or. Set t he C H 1 VOLTS/ D IV sw itc h t o . 1
and the CH 2 VOLTS/DIV switch to 5 mV. Adjust the Chan­nel 2 VARIABLE VOLTS/ DIV control to produc e a disp la y exactly five divisions in amplitude. The casc aded deflection factor is determined by dividing the C H 1 VOLTS/ DIV switc h se t t i n g b y 5 ( C H 2 V O LTS/ D IV sw i t c h a n d V A RIA BLE c o n t r o l re main a s set a bov e). For exa mple, with th e CH 1 VO LTS/ DIV sw itch se t to 5 mV the calibrated de flec tion fac tor will be 1 millivolt/division; CH 1 VOLTS/DIV switch set to 10 mV, 2 millivolts/ divisio n, e tc .
Th e f o l l o w i n g o p e r a t i n g c o n si d e r a t i o n s a n d b a s i c a p p l i -
cations may suggest other uses for this feature.
1. If A C c oupling is d esired , set the C ha nn el 1 Inp ut Coupling switch to AC and leave the Channel 2 Input Coupling switc h set to DC. When both Input Coupling sw i t c h e s a r e se t t o D C , D C si g n a l c o u p l in g i s p r o v id e d .
2. Kee p b oth v ertic al POSITION c ontro ls set nea r m id-
ra ng e.
one of the POSITION controls being turned away from
midrange, correct operation can be obtained by keeping
the Cha nnel 2 POSITION control near midrange and using
the Channel 1 POSITION control to position the trac e near
the desired locatian. Then, use the Channel 2 POSITION control far exact positioning. This method will keep both
Input Pre amps o perating in t he ir line a r ra nge.
least 25 millivo lts/ division of CRT d isp lay in all CH 1 VOLTS/ DIV switc h positions.
DIV contro l have no effect on the signal availab le at the CH 1 OUT connector.
ance matching stage with or without voltage gain. The
If th e in put sig n a l ha s a DC le ve l w hic h n ec essit ate s
3. The outp ut vo lta ge a t the C H 1 OUT c on nectar is a t
4. The MODE switch a nd Cha nne l 1 VA RIABLE VOLTS/
5. The C ha nn el 1 Input Pream p c a n b e used as an impe d -
inp ut re sista nc e is on e me g ohm a nd the o utp ut re sista nc e is about 50 ohms.
6. The dynamic ra ng e of the Ch anne l 1 Inp ut Pre amp is equal to about 20 times the CH 1 VOLTS/DIV setting. The CH 1 OUT signal is nominally at 0 volt DC for a 0 volt DC inp ut le ve l [Ch annel 1 POSITIO N c ontro l c ente red ). Th e Chanel 1 POSITION control can be used to center the out­put signal within the dynamic range of the amplifier.
7. If dua l-trace ope ration is used , the signal a pplie d to
the Channel 1 INPUT connector is disp la yed when Channel
1 is turned on. When Channel 2 is turned on , the amp lified
si g n a l i s d i sp l a y e d . Th u s, C h a n n e l 1 t r a c e c a n b e u se d t o monitor the input signal while the amplified signal is dis-
played by Channel 2.
8. In spec ial a pplic a tions w he re the fla t fre quency re -
sp o n se o f t h e Ty p e 4 5 3 i s n o t d e si re d , a f il t e r i n se r t e d between the CH 1 OUT and Channel 2 INPUT connector allows the oscilloscope to essentially take on the frequency
re sp onse of the filte r. C om bine d w ith met ho d 7, the inp ut c an be monitored by Channel 1 and the filtered signal displayed
by Channel 2.
9. By using Channe l 1 a s a 5X lo w-le ve l v oltag e p re­amplifier (5 mV position), the Channel 1 signal available at the CH 1 OUT connector can be used for any application
where a low-impedance preamplifier signal is needed. Re m e m b e r t h a t i f a 5 0- o h m lo a d i m p e d a n c e i s u se d , t h e si g n a l a m p l i t u d e w i ll b e a b o u t o n e - h a l f.
Algebraic Addition
General. Th e A D D p o si t i o n o f t h e M O D E s w i t c h c a n b e
used to d isplay the sum or d ifferenc e of two signals, for common-mode rejection to remove an undesired signal or for DC offset (ap plying a DC voltage to one channel to offse t the DC component of a signal on the other c hannel).
Th e c o m m o n - m o d e r e j e c t i o n r a t i o o f t h e Ty p e 4 5 3 i s greater than 20:1 at 20 megahertz for signal amplitudes up to eight times the VOLTS/ DIV switch setting. Rejec tion ratios of 100:1 can typically be achieved between DC and 5 mega­hertz b y ca reful a djustm ent of the gain of either c hannel while observing the displayed common-mode signal.
Deflection Fac tor. Th e o v e r a l l d e f l e c t i o n i n t h e A D D position of the MODE switch when both VOLTS/DIV switches
are set to the same position is the same as the deflection factor indicate d by eithe r VOLTS/ DIV sw itch. The a mpli­tude of an added mod e display can be determined directly from the resulta nt CRT deflection multiplied by the deflec ­tion factor indicated by either VOLTS/DIV switch. How­ever, if the CH 1 and CH 2 VOLTS/ DIV switches are set to different deflection factors, resultant voltage is difficult to determine from the CRT display. In this case, the voltage amplitude of the resultant display can be determined accu­ra tely only if th e amp litud e o f the sig na l a p p lied to eit he r channel is known.
Pre c a u tion s.
be observed when using the ADD mode.
1. Do no t exceed the inp ut voltag e
453.
Th e f o l l o w i n g g e n e r a l
precautions should
ra ting o f the Type
2-13
TM 11-6625-1722-15
2. Do n ot a pply sig na ls th at e xcee d on e quiva lent o f about 20 times the VOLTS/DIV switch setting. For example, with a VOLTS/DIV switch setting of .5, the voltage applied to that channel should not exceed a bout 10 volts. Larger voltages may disto rt the disp la y.
3. Use ve rtic al POSITION c ontro l se ttin gs w hic h m ost nearly position the signa l of ea ch cha nnel to mid-sc reen when viewed in either the CH 1 or CH 2 positions of the MODE switch. This insures the greatest dynamic range for ADD mode opera tion.
4. For similiar re sp onse from ea c h c ha nnel, set both Input
Coupling switches to the same position.
Trig g e r So u r c e
INT. Fo r m o st a p p l ic a t i o n s, t h e sw e e p c a n b e t ri g g e r e d
int erna lly. In th e INT p o sit ion o f t he Trig g ering SO URC E sw i t c h , t h e t r ig g e r si g n a l i s o b t a i n e d f r o m t h e V e r t i c a l
Defle ction System. The TRIGGER sw itch pro vide s furthe r
se l e c t i o n o f t h e i n t e r n a l t ri g g e r si g n a l ; o b t a i n e d f r o m t h e
Channel 1 signal in the CH 1 ONLY position, or from the displayed signal when in the NORM position. For single­trac e displays of either channel, the NORM position pro­vides the most co nvenient operation. However, for dual­trac e displays special c onsiderations must be made to provide the c orrec t display. Set Dual-Trace Op eration in this sec tion
for dual-tra ce trig gering information.
LI N E. Th e L I N E p o s i t i o n o f t h e SO U RC E s w i t c h c o n n e c t s
a sample of the power-line frequency to the Trigger Gen-
erator circuit. Line triggering is useful when the input signal
is time -re late d t o t he lin e f req ue nc y. It is a lso u se fu l f or
providing a stable display of a line-frequency component
in a c om plex wa ve fo rm .
EX T. An externa l signal conected to the EXT TRIG INPUT connector can be used to trigger the sweep in the EXT position of the Triggering SOURCE switch. The external sig­nal must b e time-related to the disp la yed signal for a stable display. An external trigger signal can be used to provide a triggered display when the internal signal is too low in amplitude for correct triggering, or contains signal com­p onents on wh ich it is not de sired to trigge r. It is also useful when signal tracing in amplifiers, phase-shift networks, wave­sh a p i n g c i rc u i t s, e t c . Th e si g n a l f r o m a sin g l e p o i n t i n t h e circuit under test can be connected to the EXT TRIG INPUT
connector through a signal probe or cable. The sweep is
then trigg ered by the sa me signal at a ll times a nd allows amplitude, time relationship or waveshape changes of si g n a l s a t v a ri o u s p o i n t s i n t h e c i rc u i t t o b e e x a m i n e d w i t h ­out resetting the trigger controls.
frequency components of the trigger signa l which can trig ger the sweep.
AC. Th e A C p o si t i o n b l o c k s t h e D C c o m p o n e n t o f t h e trigger signal. Signals with low-frequency components below about 30 hertz are attenuated. In general, AC coupling can be used for most applications. However, if the trigger signal contains unwanted components or if the sweep is to be triggered at a low repetition rate or a DC level, one of the re maining COUPLING sw itc h p osition s w ill p rovid e a bet ter display.
Th e t r i g g e r i n g p o i n t i n t h e A C p o s i t i o n d e p e n d s o n t h e average voltage level of the trigger signal. If the trigger si g n a l s o c c u r in a ra n d o m f a sh i o n , t h e a v e r a g e v o l t a g e l e v e l will vary, causing the triggering point to vary also. This shift of the triggering point may be enough so it is impossible to
maintain a stable display. In such cases, use DC coupling.
LF REJ . In th e LF REJ p o sitio n, DC is re je c te d a nd sig na ls
below about 30 kilohertz are attenuated. Therefore, the
sw e e p w i ll b e t r ig g e re d o n l y b y t h e h i g h e r - f re q u e n c y c o m -
ponents of the signal. This position is particularly useful
for providing sta ble triggering if the trigger signal contains
line -frequenc y com po ne nt s. Also , in the ALT positio n of the MODE switch, the LF REJ position provides the best display
at high sweep rates when comparing two unrelated signals
(TRIGGER switc h set to NORM).
HF REJ. Th e H F R EJ p o s i t i o n p a s se s a l l l o w - f r e q u e n c y sig na ls b et we en a bout 30 hertz a nd 50 kilo hertz. DC is re jecte d a nd sig na ls o utsid e the g ive n ra ng e are a tt enuate d . When trigg ering from co mplex w avefo rms, this p ositio n is useful for providing sta ble d isp lay of low-frequenc y compo­nents.
DC. DC c oupling ca n b e used to pro vide sta ble trigg ering with low-frequency signals which would be attenuated in the AC position, or with low-repetition rate signals. The LEVEL c o n t ro l c a n b e a d j ust e d t o p r o vi d e t r ig g e r in g a t t h e desired DC level on the waveform. When using internal triggering, the setting of the Channel 1 and 2 POSITION controls affects the DC trigger level.
DC trigg er coup ling should not b e used in the ALT dua l­trac e mod e if the TRIGGER switch is set to NORM. If used , the sweep will trigger on the DC level of one tra ce a nd then either lock out completely or free run on the other trace. Correct DC triggering for this mode c an be obtained with the TRIGGER switch set to CH 1 ONLY.
Trig g e r Slo p e
sa m e a s d e sc r ib e d f o r EX T e x c e p t t h a t t h e e x t e rn a l t r ig g e r in g si g n a l i s a t t e n u a t e d 1 0 t i m e s. A t t e n u a t i o n o f h i g h - a m p l it u d e external triggering signals is desirable to broaden the range of the Triggering LEVEL control. sw i t c h i s se t t o L F REJ, a t t e n u a t i o n is a b o u t 2 0 :1 .
Tr ig g e r C o u p l i n g
Fo u r m e t h o d s o f c o u p l in g t h e circuits can be selected with sw i t c h e s. Ea c h p o si t i o n p e rm i t s
2-14
When the COUPLING
trigger signal to the trigger
the Trig gering COUPLING
se l e c t i o n o r re j e c t io n o f t h e
Th e t r i g g e r i n g SL O P E s w i t c h d e t e r m i n e s w h e t h e r t h e t r i g g e r circuit responds on the positive-going or negative-going portion of the trigger signal. When the SLOPE switch is in the + (positive-going) p osition, the display sta rts with the positive-going portion of the waveform; in the - (negative­going) position, the display starts with the negative-going
portion of the waveform (see Fig. 2-8]. When several cycles of a signal appear in the display, the setting of the SLOPE sw i t c h is o f t e n u n i m p o r t a n t . Ho w e v e r , if o n l y a c e rt a i n
portion of a cycle is to be displayed, correct setting of the SLO P E s w i t c h i s i m p o r t a n t t o p r o v i d e a d i s p l a y w h i c h st a r t s on the desired slope of the input signal.
Loading...
+ 229 hidden pages