Tektronix PWS4205, PWS4305, PWS4323, PWS4602, PWS4721 Performance Verification

xx
PWS4205, PWS4305, PWS4323, PWS4602, and PWS4721 Linear DC Power Supplies
ZZZ
Specications and Performance Verication
Technical Reference
*P077048000*
077-0480-00
xx
ZZZ
Specications and Performance Verication
Technical Reference
Warning
The servicing instructions ar only. To avoid personal injury, do not perform any servicing unless you are qualied to do s before performing service.
e for use by qualied personnel
o. Refer to all safety summaries
www.tektronix.com
077-0480-00
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specica tions and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O . B ox 50 0 Beaverto USA
For product information, sales, service, and technical support:
n, OR 97077
In North America, call 1-800-833-9200. World wide, vis it www.tektronix.com to nd contacts in your area.
Table of Contents
General Safety Summary ......................................................................................... iii
Specications ........................ .................................. ................................ ............. 1
Performance Verication ................. .................................. ................................ ....... 9
Test Rec o r d
Performance Verication Procedures ....................... ................................ ................ 19
........................ ................................ ................................ ............ 10
PWS4000 Series Specications and Performance Verication i
Table of Contents
List of Tables
Table 1: Specications...... ................................ ................................ ....................... 1
Table 2 : USB
Table 3: Terminal strip characteristics ........................................................................... 5
Table 4: Mains power characteristics ............. .................................. ............................. 6
Table 5: Mechanical characteristics ..... ................................ .................................. ....... 6
Table 6: Environment performance.......................................... ................................ ..... 7
Table 7: Keypad special operations .............................................................................. 8
Table 8 : D
Table 9: DC voltage accuracy without remote sense .......................................................... 12
Table 10: DC voltage readback accuracy ....................................................................... 13
Table 11: DC voltage line regulation............................................................................ 14
Table 12: DC voltage load regulation ....... ................................ .................................. .. 14
Table 13: DC overvoltage detection................. ................................ ............................ 14
14: DC current accuracy................................................................................... 15
Table
Table 15: DC current readback accuracy ....................................................................... 16
Table 16: DC current line regulation ............................................................................ 17
Table 17: DC current load regulation ............. .................................. ............................ 17
Table 18: Voltage noise at 7 MHz ....... ................................ .................................. ...... 18
Table 19: Voltage noise at 20 MHz.................. ................................ ............................ 18
ble 20: Current noise at 20 MHz........................ ................................ ...................... 18
Ta
Table 21: Test equipment. ................................ ................................ ........................ 20
characteristics ................. ................................ ................................ ..... 4
C voltage accuracy with remote sense.............................................................. 11
ii PWS4000 Series Specications and Performance Verication
General Safety Summary
General Safet
To Avoid Fi
re or Personal
Injury
ySummary
Review the fo this product or any products connected to it.
To avoid pot
Only qualied personnel should perform service procedures.
Use proper
certied for the country of use.
Use prope
is in the proper position for the source being used.
Ground t
of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
llowing safety precautions to avoid injury and prevent damage to
ential hazards, use this product only as specied.
power cord. Use only the power cord specied for this product and
r voltage setting. Before applying power, ensure that the line selector
he product. This product is grounded through the grounding conductor
Power disconnect. The power switch disconnects the product from the power
source. See instructions for the location. Do not block the power switch; it must remain accessible to the user at all times.
Do not operate without covers. Do not operate this product with covers or panels removed.
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by qualied service personnel.
Avoid exposed circuitry. Do not touch exposed connections and components when power is present.
Use proper fuse. Use only the fuse type and rating specied for this product.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
Provide proper ventilation. Refer to the manual's installation instructions for details
on installing the product so it has p roper ventilation.
PWS4000 Series Specications and Performance Verication iii
General Safety Summary
TermsinThisManual
Symbols and Terms on the
Product
These terms may
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
These terms may appear on the product:
DANGER in the marking.
WAR NI N G read the marking.
CAUTIO
The following symbol(s) may appear on the product:
appear in this manual:
dicates an injury hazard immediately accessible as you read
indicates an injury hazard not immediately accessible as you
N indicates a hazard to property including the product.
iv PWS4000 Series Specications and Performance Verication
Specications
This section contains specications for the PWS4000 series instruments. All specications are guaranteed unless noted as "typical." Typical specications are provided for marked with the
Table 1: Specications
Parameter PWS4205 PWS4305 PWS4323 PWS4602 PWS4721
Constant voltage
n, nominal
operatio
Voltage range
This is th over which the output voltage is adjusta
Remote sense voltage
Voltage resolution, nomin
accuracy with remote sense
At 25 °C ±5 °C
accuracy without remo
At 25 °C ±5 °C
Current limit, nominal
Voltage temperature co
Constant current o
C
This is the range over which the output current is adjustable
e range
ble
range
al
Front panel:
amming:
Progr
Voltage setting
Voltage setting
te sense
efcient, typical
peration, nominal
urrent range
The unit may be set to a constant voltage over a range of currents
0to20V 0to30V 0to32V 0to60V 0to72V
The tota minimum of 0.1 V/line for remote sense.
The voltage each terminal of the remote sense function is allowed to vary from the respective output terminal is 0 V to
1mV
0.1 mV
±(0. settling+3mV)
±(0.05% of settling + 10 mV)
Dened by the constant current setting. The changeover is automatic.
±(0.02% + 3 mV) pe the25°C±5°C standard range
The unit may be set to a constant current over a range of voltages.
0Ato5A 0Ato5A 0Ato3A 0Ato2.5A 0Ato1.2A
l voltage from the – terminal to the + terminal may not exceed the rated FS voltage + 0.2 V, allowing a
1V.
1mV
0.1 mV
03% of
r °C outside
±(0. settling+3mV)
±(0.05% of settling + 10 mV)
±(0.02% + 3 mV) pe the25°C±5°C standard range
your convenience but are not guaranteed. Specications that are
symbol are checked in Performance Verication.
03% of
r °C outside
1mV
0.1 mV
03% of
±(0. settling + 3 mV)
±(0.05% of settling + 10 m V )
±(0.02% + 3 mV)
r ° C outside
pe the25°C±5°C standard range
1mV
0.5 mV
03% of
±(0. settling+6mV)
±(0.05% of settling + 20 mV)
±(0.02% + 6 mV)
r °C outside
pe the25°C±5°C standard range
1mV
0.1 mV
±(0. settling+6mV)
±(0.05% of settling + 20 mV)
±(0.02% + 5 mV)
r °C outside
pe the25°C±5°C standard range
03% of
PWS4000 Series Specications and Performance Verication 1
Specications
Table 1: Specications (cont.)
Parameter PWS4205 PWS4305 PWS4323 PWS4602 PWS4721
Current resolution, nominal
Front panel: Programming:
Current setting
accuracy
At 25 °C ±5 °
C
nominal
Current temperature coefcient, typical
nominal
Meter,
nt readback
Curre resolution, nominal
Current
readback accuracy
At 25 °C ±5 °C
Current readback temperature coefcient, typical
Voltage dback accuracy
rea
At 25 °C ±5 °C
ltage readback
Vo resolution, nominal
ltage readback
Vo temperature coefcient, typical
Maximum input VA
,typical
RMS
Line regulation
– voltage
0.1 mA
0.1 mA
±(0.05% + 2 mA) ±(0.05% +
0.1 mA
0.1 mA
2.5 mA)
0.1 mA
0.1 mA
0.1 mA
0.1 mA
±(0.05% + 2 mA) ±(0.05% +
1.5 mA)
0.1 mA
0.05 mA
±(0.05% + 1 mA)
0to20V 0to30V 0to32V 0to60V 0to72VVoltage limit,
Also see the Over Voltage Limit characteristics.
(See pag
±(0.05% per °C outside the25°C±5°C standa
e3.)
+2mA)
rd range
±(0.05%
+2mA) per °C outside the25°C±5°C
rd range
standa
±(0.05%
+2mA) per °C outside the25°C±5°C
rd range
standa
±(0.05%
+1mA) per °C outside the25°C±5°C
rd range
standa
±(0.05%
0.5 mA) per °C outside the 25 °C ±5 °C st range
al measurement capability. Voltage and current are measured internally.
Intern
0.1 mA
±(0.05% of reading + 2 mA)
±(0.05% + 2 mA) per °C outside the25°C±5°C standard range
±(0.05% of reading + 2.5 mA)
±(0.05% + 2 mA) per °C outside the25°C±5°C standard range
±(0.05% of reading + 2 mA)
±(0.05% + 2 mA) per °C outside the25°C±5°C standard range
±(0.05% of reading + 1.5 mA)
±(0.05% + 1 mA) per °C outside the25°C±5°C standard range
±(0.05% of reading + 1 mA)
±(0.05% +
0.5 mA) per °C outside the 25 °C ±5 °C standard range
±(0.02% of reading + 3 mV)
V
1m
±(0.02% of reading + 3 mV)
er °C outside
p the25°C±5°C standard range
±(0.02% of reading + 2.5 mV)
±(0.02% of reading + 2 mV)
er °C outside
p the25°C±5°C standard range
±(0.02% of reading + 3 mV)
±(0.02% of reading + 3 mV)
er °C outside
p the25°C±5°C standard range
±(0.02% of reading + 6 mV)
±(0.02% of reading + 6 mV)
er °C outside
p the25°C±5°C standard range
±(0.02% of reading + 5 mV)
±(0.02% of reading + 5 mV)
er °C outside
p the25°C±5°C standard range
350 VA 500 VA 350 VA 500 VA 350 VA
Change over the full rated input voltage: ±(0.01% +1mV)
Change over the full rated input voltage: ±(0.01% +1mV)
Change over the full rated i nput voltage: ±(0.01% +1mV)
Change over the full rated input voltage: ±(0.01% +2mV)
Change over the full rated input voltage: ±(0.01% +1mV)
+
andard
2 PWS4000 Series Specications and Performance Verication
Specications
Table 1: Specications (cont.)
Parameter PWS4205 PWS4305 PWS4323 PWS4602 PWS4721
Line regulat
ion
– current
Load regul
ation
– voltage
Load regulation
– current
Voltage ripple,
20 MHz bandwidth
Voltage ripple,
7 MHz bandwidth
Current ripple,
20 MHz bandwidth
Voltage transient response settling time, load change
Voltage transient response settling time, setting change, rising
Voltage transient response settling time, setting change, falling
Isolation voltage
Over Voltage
otection setting
Pr accuracy
Over Voltage Protection “ON”
oltage, typical
v
Change over full rated i nput voltage: ±(0.05% +0.1mA)
Change from 0 to 100% of full scale current; ±(0.01% + 2 mV)
Change from 10 to 90% of full scale vol
tage;
±(0.05% +
0.1 mA)
RMS
RMS
p-p
RMS
and 20 mV
and 12 mV
RMS
and
<3mV <1.5mV
<1mV 3mV
<3mA
<400 μs to within 75 mV following a change from 0.1 A to 1 A.
<35 ms from beginning of excursion to within 75 mV of terminal value following a change from 1 V to 11 V into a 1 A load.
<35 ms from beginning of excursion to within 75 mV of terminal value following a change from 11 V to 1 V into a 1 A load.
The output can be oated up to 100 V (DC + pk AC (where the AC component cannot exceed 3 V 60 Hz max)) between the Earth Ground term inal and any of the following terminals:
Positive Output (+), front or rear panel
Negative Output (-), front or rear panel
Remote Sense (+ or -). Must maintain correct voltage relative to the related output.
All other rear panel terminals are grounded.
0.5% of setting + 0.5 V
V
<1
Change over full rated input voltage: ±(0.05% +0.1mA)
Change from 10 to 90% of full scale vol
tage;
±(0.05% +
1.5 mA)
, guaranteed
p-p
,typical
p-p
<1mV 4mV
<4mA
and
RMS
p-p
RMS
Change over full rated input voltage: ±(0.05% +0.1mA)
Change from 10 to 90% of full scale vol
tage;
±(0.05% +
0.1 mA)
<1mV 4mV
<3mA
and
RMS
p-p
RMS
Change over full rated i nput voltage: ±(0.05% +0.05mA)
Change from 10 to 90% of full scale vol
tage;
±(0.05% +
0.5 mA)
<1mV 5mV
<3mA
and
RMS
p-p
RMS
Change over full rated input voltage: ±(0.05% +0.1mA)
Change from 10 to 90% of full scale vol ±(0.05% +
0.5 mA)
<1mV
RMS
3mV
p-p
<3mA
RMS
p-p
tage;
and
max and
PWS4000 Series Specications and Performance Verication 3
Specications
Table 1: Specications (cont.)
Parameter PWS4205 PWS4305 PWS4323 PWS4602 PWS4721
Over Voltage Protection “ON” holding current, typical
Over Voltage Protection adjustment range, typical
Over Voltage Protection response time, typical
Over Voltage Protection setting resolution
Setup memory, nominal
Software voltage limit
>250 mA
A minimum amount of current is required for operation.
1VOVP Setting FS Output Voltage
It may be possible to set very low voltages for the O VP output but behavior may be unexpected.
<10 ms from voltage sensed in excess of OVP setting to output begins to fall.
1mV
40 setup locations
The maximum settable voltage (MAX VOLT) menu entry, when set below the hardware maximum voltage limit, disallows adjusting the instrument to voltages more than the MAX VOLT limit.
Table 2: USB characteristics
Parameter All models
USB device USB 2.0 Full Speed device. 12 Mb/sec maximum. Supports USB-TMC communications with Tektronix
ensions to support TEK-USB-488 GPIB to USB adapter.
ext
mmand
Co processing time, typical
GPIB interface GPIB access using a TEK-USB-488 accessory through USB port.
mmands will be processed and committed to hardware within 400 ms.
Co
4 PWS4000 Series Specications and Performance Verication
Specications
Table 3: Termin
al strip characteristics
Parameter All models
CONTROL IN and
OUT
Port Modes available:
Trigger RI/DFI Digital I/O
Negative (-)
Digital OUT DFI or Digit
Open circui
t
5 V ±10%
voltage, typical
Short circuit
0.5 mA ±20%
current, typical
Breakdown voltage, typical
Maximum current, typical
The output may be driven to 5 V.
The Maximum current must not be exceeded.
The output will sink up to 1 mA
The Output Breakdown Voltage must not be exceeded.
Digital IN Trigger, RI or Digital IN
Open circuit
5 V ±10%
voltage, typical
Short circuit
t, typical
curren
Maximum voltage,
al
typic
VIH,typical
0.5 mA ±20%
C
5.5 VD
(V
2.4 V
)
IH
Voltage at which the sense is assured to be high.
typical
VIL,
0.8 V (V
)
IL
Voltage at which the sense is assured to be low.
terminals are referenced to chassis ground.
al OUT
PWS4000 Series Specications and Performance Verication 5
Specications
Table 4: Mains p
Parameter All models
Voltage Two ranges. Bottom-panel switch selectable.
Frequency
Fuse rating nominal
ower characteristics
50/60 Hz
,
110 VAC setting: 5.0 A TH 250 V
220 VAC setting: 2.5 A TH 250 V
110 V: 110 VAC t
220 V: 220 VAC to 240 VAC (±10% )
o 120 VAC (±10%)
Table 5: Mechanical characteristics
Parameter PWS4205 PWS4305 PWS4323 P WS4602 PWS4721
Weight, nominal
Includes boots, handles, and shorting adapters
Dimensions, nominal
Cooling method, nominal
Required cooling clearance
Construction materials, nominal
16.0 lbs (7.3 kg)
With boots and handle:
Height: 105 mm (4.15 in)
Width: 242 mm (9.52 in)
Depth: 384 mm (15.12 in)
Without boots and handle:
Height: 91 mm (3.57 in)
Width: 217 mm (8.55 in)
Depth: 362 mm (14.24 in)
Fan
Rear of instrument must be separated by at least 5 cm (2 in) from any airow restrictions to assure adequate cooling.
Bottom of instrument must have 1 cm (0.4 in) from any airow restrictions. The protective boot maintains the proper bottom c learance.
For rackmount conguration, follow the clearance requirements provided with the rackmount installation instructions.
Chassis – Aluminum Front Panel Plastics – PC/ABS blend
16.0 lbs (7.3 kg)
16.0 lbs (7.3 kg)
15.3 lbs (7.0 kg)
16.0 lbs (7.3 kg)
6 PWS4000 Series Specications and Performance Verication
Specications
Table 6: Enviro
Parameter All models
Temperature
Humidity
Altitude
nment performance
Operating: +0
Nonoperating: -20 °C to 70 ° C
Operating:
5% to 95% relative humidity (% RH) at up to 40 °C, noncondensing
Nonoperati
5% to 95% relative humidity (RH) at up to +40 ° C,
5% to 60% RH above +40 °C up to +70 C, noncondensing
Operating: 100% capability up to 2,000 meters
Nonoperat
°C to +40 °C
ng:
ing: Up to 4,000 meters
PWS4000 Series Specications and Performance Verication 7
Specications
Table 7: Keypad
Operation Key press
Display the instrument rmware version. Shift + 8
Display the internal temperature of the instrument. Shift + . + 1
Perform a factory r eset. All setups and lists are erased. Shift + . + 2
Activate the factory service port.
The factory service port provides no user functions.
Unlock the keypad.
special operations
Shift + . + 3
If the password used to lock the keypad is forgotten, enter the four nu keypad.
Cycling the instrument power also unlocks the keypad.
merical digits of the instrument name to unlock the
8 PWS4000 Series Specications and Performance Verication
Performance Verication
This section contains performance verication procedures for the specications marked with the verication
These procedures cover the PWS4205, PWS4305, PWS4323, PWS4602, and PWS4721 mod are testing.
procedures. (See Table 21 on page 20.)
symbol. Additional test equipment is required to complete the
els. Ignore checks that do not apply to the specic model you
Print the t test results for your power supply.
NOTE. Completion of the performance verication procedure does not update the
stored time and date of the latest successful adjustment. The date and time are updated only when the factory adjustment procedures are successfully completed.
The performance verication procedures verify the performance of your instrument. They do not adjust your instrument. If your instrument fails any of the performance verication tests, you should contact Tektronix service.
WAR N I
this product and any associated instrumentation. Although some instruments and accessories are used with nonhazardous voltages, there are situations where hazardous conditions may be present. This product is intended for use by qualied personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Before using the product, carefully
ad and follow all installation, operation, and maintenance information. Refer
re to this manual for complete product specications. Before performing any maintenance, disconnect the line cord and all test cables. Operators of this instrument must be protected from electric shock at all times. The responsible body must make sure that operators are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit can operate at or above 1000 volts, no conductive part of the circuit may be exposed.
est record on the following page s, and use it to record the performance
NG. Observe all safety precautions listed in this manual before using
WAR N ING. Do not loosen any screw on this product. There are no user
serviceable components inside.
CAUTION. Use properly rated load wires. All load wires must be heavy enough not
to overheat when carrying the maximum short-circuit output current of the power supply. If there is more than one load, then any pair of load wires must be capable of safely carrying the full-rated short-circuit output current of the power supply.
PWS4000 Series Specications and Performance Verication 9