Tektronix P8018 User manual

P8018 Handheld TDR Probe Instructions
071-1318-01
Safety symbols on the product
CAUTION
Refer to Product Manual
Environmental Considerations
This section provides information about the environmental impact of the product.
Product End-of-Life Handling
Observe the following guidelines when recycling an instrument or component:
Equipment Recycling. This product complies with the European Union’s requirements according to Directive 2002/96/EC on waste electrical and electronic equipment (WEEE). For information about recycling options, check the Support/Ser­vice section of the Tektronix Web site (www.tek­tronix.com).
Restriction of Hazardous Substances
This product has been classified as Monitor­ing and Control equipment, and is outside the scope of the 2002/95/EC RoHS Directive. This product is known to contain lead and hexavalent chromium.
Using the Probe
Connector Care
Use extra care when attaching or removing the cable from SMA connectors. Turn only the nut, not the cable. When attaching the cable, align the connectors carefully before turning the nut. Use light finger pressure to make this initial connection. Then tighten the nut lightly with a wrench. Ideally, the SMA connectors should be tightened to 56 N-cm (5 in-lb) with a torque wrench.
Probing Technique
The probe tip and grounding clip are designed to probe printed circuit boards (PCB) that incorporate ground reference points next to the signal measurement point. This is to allow the probe tip to be inserted into a PCB through-hole test point while the ground clip contacts the ground reference point.
To achieve the specified bandwidth, the ground clip must be trimmed to the shortest length possible for probing the PCB. See Figure 2 for proper probing technique and ground clip trimming.
www.tektronix.com
*P071131801*
071131801
Safety Summary
To avoid potential hazards, use this product only as specified.
Ground the product. This product is indirectly
grounded through the grounding conductor of the mainframe power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Connect and disconnect properly. Do not connect or
disconnect probes or test leads while they are connected to a voltage source.
Observe all terminal ratings. To avoid fire or shock
hazard, observe all ratings and markings on the product. Consult the host instrument documenta­tion for further ratings information before making connections to the host instrument.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Do not operate with suspected failures. If you
suspect there is damage to this product, have it inspected by qualified service personnel.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
Product Description
The P8018 is a handheld passive probe with a 50 Ω input impedance. The P8018 comes standard with a one meter SMA male-to-male cable assembly that interfaces with an 80A02 EOS/ESD Protection Module or other SMA connectable input.
Figure 1: P8018 handheld TDR probe
The P8018 probe is especially designed to be used with the Tektronix 80A02 EOS/ESD Protection Module. Together, they protect sensitive instrument-input stages, such as the sampling bridge of Tektronix electrical sampling modules, from electro-overstress (EOS) and electrostatic discharge (ESD) damage due to static charges stored in the device under test (DUT).The P8018 probe can also be used as a stand-alone, high-bandwidth probe that is usable with any SMA connectable instrument.
The P8018 provides a unique pressure-actuating probe tip that switches between DUT static-discharging protective mode and engaged-measurement mode.
Ground clip installed on
probe
PCB
Figure 2: PCB probing and grounding technique
These instructions describe the operation of the P8018 probe connected to a Tektronix 80A02 protection module which is then connected to an instrument to be protected such as a Tektronix 80E04 or SD24 electrical TDR sampling module.
Probe tip through PCB
test point via hole
Ground clip
contact trimmed as
short as possible
Copyright © Tektronix. All rights reserved.
Specifications
Table 1: Environmental and mechanical
Characteristic
Probe dimensions
Length 125 mm (4.94 in)
Height 27 mm (1.06 in)
Width 16 mm (0.64 in)
Weight (probe only) 0.054 kg (0.12 lbs)
Altitude
Operating 3,000 m (9,843 ft)
Non-operating 12,190 m (40,000 ft)
Ambient temperature
Operating --15 _C to +55 _C(+5_Fto
Non-operating -- 6 2 _C to +85 _C(--80_F
Humidity
Operating 20% to 80% relative
Non-operating 5% to 90% relative humidity
Specification
+131 _F)
to +185 _F)
humidity with a maximum wet bulb temperature of +29 _C at or below +40 _C, non-condensing. (Upper limit derates to 45% relative humidity at +40 _C.)
with a maximum wet bulb temperature of +29 _Cator below +60 _C, non-condensing. (Upper limit derates to 20% relative humidity at +60 _C.)
Table 2: Electrical
Characteristic
Probe impedance
Attenuation 1X
RF input connector Precision 26 GHz SMA
Control signal connector type
Analog bandwidth
Probe head <3 dB insertion loss, DC to
Cable DC to 20 GHz
Probe fall time (measured with 80E04 or SD24 TDR modules)
Probe delay 1 ns typical from the
Maximum voltage input
Module activation probe pressure
Specification
50 Ω +±2%
female connector
Audio mono phone mini-jack
20 GHz
<38ps
Typical reflected TDR fall time<28ps
connector end to the probe tip
This is a stable impedance section which can be calibrated and used as a reference point.
±5.0 V
200 to 260 grams
Accessories
The P8018 probe includes the following items:
S Special 1 meter SMA cable
with a parallel control line) 174-4879-xx
S 20 ground clips 016-1936-xx
WARRANTY SUMMARY
Tektronix warrants that the products that it manufactures and sells will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If a product proves defective within the respective period, Tektronix will provide repair or replacement as described in the complete warranty statement.
To arrange for service or obtain a copy of the complete warranty statement, please contact your nearest Tektronix sales and service office.
EXCEPT AS PROVIDED IN THIS SUMMARY OR THE APPLICABLE WARRANTY STATEMENT, TEKTRONIX MAKES NO WARRANTY OF ANY KIND, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL TEKTRONIX BE LIABLE FOR INDIRECT, SPECIAL OR CONSEQUENTIAL DAMAGES.