Tektronix P7708, P7716, P7720, P7713 User Manual

x
P7700 Series TriMode™ Probes
ZZZ
User Manual
*P071342102*
071-3421-02
xx
P7700 Series TriMode™ Probes
ZZZ
User Manual
www.tek.com
071-3421-02
Copyright © Tektronix. All rights reserved. Licens ed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specications and price ch ange privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
TriMode and TekFlex are both trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O . B ox 5 0 0 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200. Worldwide, visit www.tek.com to nd contacts in your area.
Warranty
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If any such product proves d efective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and replacement products used by Tektronix for warranty work may be new or reconditioned to like new performance. All replaced parts, modules and products become the property of Tektronix.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packag ing and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is l o cated. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned t o any other locations.
This warranty shall not apply to any defect, failure or da m age caused b y improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to s ervice a product that has been modied or integrated with other products when the effect of such modication or integration increases the time or difculty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEK TRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES. [W2 – 15AUG04]
Table of Contents
Importantsafety information................................................................................................................ iii
General safety summary. . ............................................................................................................ iii
Terms in this manual .. . ............................................................................................................... iv
Symbols and terms on the product.................................................................................................... v
Compliance information .................................................................................................................... vi
Environmental considerations ........................................................................................................ vi
Preface..................................................................................................................................... vii
Probe models ........................................................................................................................ vii
Documentation....................................................................................................................... vii
Key features ................................................................................................................................ 1
Operating considerations................................................................................................................... 2
Installation .................................................................................................................................. 4
Overview .............................................................................................................................. 4
Connect to the host instrument ....................................................................................................... 5
Connect P7700 accessories to the TekFlex connector . ............................................................................. 7
Probe comp box controls and indicators ............................................................................................. 8
Table of Contents
Functional check andcalibration.......................................................................................................... 11
Functional check..................................................................................................................... 11
TriMode probe DC calibration .......................................................................................................18
P7700 Series TriMode Probes User Manual i
Table of Contents
Basic operation ............................................................................................................................ 24
Probe handling best practices............................................................................................................. 64
S-parameter information................................................................................................................... 66
Accessories and options .................................................................................................................. 67
Maintenance ............................................................................................................................... 74
Index
Probe setup screen .................................................................................................................. 28
Improving measurement accuracy .................................................................................................. 36
Connecting to a circuit board . . . ..................................................................................................... 45
Standard accessories. . .............................................................................................................. 67
Optional accessories................................................................................................................. 70
Host instrument rmware............................................................................................................ 74
Browser tip replacement............................................................................................................. 75
Error conditions ...................................................................................................................... 76
Handling the probe................................................................................................................... 79
Cleaning the probe................................................................................................................... 80
Returning theprobe for servicing.................................................................................................... 81
ii P7700 Series TriMode Probes User Manual
Important safety information
This manual contains information and warnings that must be followed by the user for safe operation and to keep the product in a safe condition.
General safety summary
Use the product only a s specied . Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. Carefully read all instructions. Retain these instructions for future reference.
This product is not intended for detection of hazardous voltages.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the
product manual for further ratings information before making connections to the product.
Do not ap ply a potential to any terminal, including the c ommon terminal, that exceeds the maximum rating of that terminal.
Do not operate without covers. Do not operate this product with covers or panels removed, or with the case open. Hazardous
voltage exposure is possible.
Avoid exposed circuitry. Do not touch exposed connections and components when power is present.
Do not operate in wet/damp conditions. Be a wa re that condensation may occur if a unit is moved from a cold to a warm
environment.
Important safety information
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
P7700 Series TriMode Probes User Manual iii
Remove the input signals before you clean the product.
Important safety information
Probes and test leads
Remove all probes, test leads and accessories that are not in use.
Inspect the probe and accessories. Before each use, inspect probe and accessories for damage (cuts, tears, or defects in the
probe body, accessories, or cable jacket). Do not use if damaged.
Useonspecified replacement parts.
Termsinthismanual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to t his product or other property.
iv P7700 Series TriMode Pro bes User Manual
Symbols and terms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
When this symbol is marked on the product, be sure to consult the manual to nd out the nature of the potential hazards and any actions which have to be taken to avoid them. (This sym bol may also be used to refer the user to ratings in the manual.)
The following symbol(s) may appear on the product:
Important safety information
P7700 Series TriMode Probes User Manual v
Compliance information
Compliance information
This section lists the environmental standards with which the instrument complies.
Environmental considerations
This section provides information about the environmental impact of the prod uct.
Product end-of-life handling
Observe the following guidelines when recycling an instrument or compo nent:
Equipment recycling. Production of this equipment required the extraction and use of natural resources. The equ ipment may
contain substances that cou ld be harmful to the environment or human health if improperly handled at the pro duct’s end of life. To avoid release of such substances into the environment and to reduce the use of natural resources, we en courage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or re cycled appropriately.
This symbol indicates that this product complies with the applicable European Union requirements according to Directives 2012/19/EU and 2006/66/EC on waste electrical and electronic equipment (WEEE) and batteries. For information about recycling op tions, check the Tektronix Web site (www.tek.com/productrecycling).
vi P7700 Series TriMode Probes User Manual
Preface
This manual describes the installation and operation of the P7700 Se ries TriMode Probes. Basic probe operations and concepts are presented in this manual. All documents listed below are accessible from the Tektronix Web site (www.tek.com/manuals).
Probe models
These P7700 Series TriMode Probe family includes these models:
P7708 8 GHz
P7713 13 GHz
P7716 16 GHz
P7720 20 GHz
Documentation
To read about Use these documents
Installation and Operation (overviews) Read this manual for information about how to set up and use your probe.
In-Depth Operation Use the technical reference manual and this manual.
Specications Use the technical reference manual.
Reordering accessories
Preface
Use the Accessories and Options section or refer to the insert in your accessory kit when reordering accessories.
P7700 Series TriMode Probes User Manual vii
Preface
viii P7700 Series TriMode Probes User Manual
Key features
The P7700 Series TriMode Probes allow you to take differential, single-ended, and common mode measurements with one probe connection. Key features includ e:
Low loading for low power circuits
Thin solder-in tips that can tintotight
spaces on a device under test
Flexible solder-in tips and main probe cable
Lightweight probe cable and head
TekFlex™ connector technology that makes the probe easy to connect to accessories
Optional 2.92 mm adapter for connecting to 50 RF test points
Browser accessory with adjustable tips and headlight
Full AC calibration of the probe and accessory tips with unique S-parameters
Key features
Solder-in tips operate at extended temperature range
P7700 Series TriMode Probes User Manual 1
Operating considerations
Operating considerations
Table 1: P7700 Series TriMode probes
Characteristic Description Specication
Input Voltage Dynamic range
Operating voltage window
Offset voltage range
Maximum non­destructive input voltage
Temperature
Non-operating
TekFlex solder-in tips
2.5 Vpp (single­ended)
5.0 Vpp (differential input)
±5.25 V ±10.0 V ±4.0 V
-4 V to +4 V -10 V to +10 V -4 V to +4 V
-15 V to +15 V (tip attached or detached)
Probe compensation box: 0 °C to 45 °C (32 °F to 113 °F)Operating
Probe cable, solder tips, and P77292MM adapter: -35 °C to 85 °C (-31 °F to 185 °F); Minimum Airow required 46°C to 85°C (114.8 °F to 185 °F)
Probe compensation box: -20 °C to 60 °Cg (-4 °F to 140 °F)
Probe cable, solder tips, and P77292MM adapter: -35 °C to 85 °C (-31 °F to 185 °F)
P77BRWSR P77C292MM adapter
6.0 Vpp (single­ended)
12.0 Vpp (differential input)
1.2 Vpp (single­ended)
2.0 Vpp (differential input)
-5Vto+5V
2 P7700 Series TriMode Probes User Manual
Operating considerations
Characteristic Description Specication
TekFlex solder-in
P77BRWSR P77C292MM adapter
tips
Humidity
Operating Probe Compensation Box, C able, Tips, and SMA adapter: 20% to 80%
Relative Humidity (%RH) at up to 45°C non-condensing
Non-operating
Probe Compensation Box, Cable, Tips, and P772 92MM adapter: 10% to 90% (Relative Humidity), non-condensing
Altitude Non-operating
Probe Compensation Box, Cable, Tips, and SMA adapter: 12,000 meters (39,370 feet)
Pollution Degree 2, Indoor use only
CAUTION. To avoid ESD damage to the probe, always use an antistatic wrist strap (provided with your probe), and work at
a static-approved workstation when you handle the probe.
P7700 Series TriMode Probes User Manual 3
Installation
Installation
Before you connect the probe to your instrument, read the Overview below to understand the sequence of events necessary to properly install the probe and adapters.
CAUTION. To avoid ESD damage to the probe, always use an antistatic wrist strap (provided with your probe), and work at
a static-approved workstation when you handle the probe.
Overview
1. Connect the probe to the host instrument.
If it is the rst time the probe has been connected to the oscilloscope, the oscilloscope will download the S-parameters stored in the probe. Once the oscilloscope has stored the S-paramete rs for the probe, it doesn’t matter which channel the probe is plugged into. The stored S-parameters will be available for any channel the probe is moved to.
2. Connect the probe t ip to the TekFlex connector on the probe.
When a tip is inserted into the probe for the rst time, the oscilloscope will download the S-parameter data stored in the tip.
3. The p
4. Open the Calibration menu (found in the Vertical menu).
5. Perform the DC probe calibration procedure using the DC probe cal xture that ships standard with each probe. (See page 18,
6. Use the Probe Setup screen to set the probe parameters as described in the Basic Operation section.
4 P7700 Series TriMode Probes User Manual
robe performs a self test, and then one Input Mode LED remains on.
TriMode probe DC calibration.)
Connect to the host instrument
NOTE. Your TekConnect instrument may
require a rmware upgrade to support full functionality of the P7700 Series probes. Before you connect the probe, check the version requirements. (See page 74, Host instrument rmware.)
1. Slide the probe into the TekConnect receptacle. The probe clicks into place when fully engaged.
2. Optional. Turn the thumbscrew clockwise (nger-tight only) to secure the probe to the instrument.
Disconnect
3. To disconnect, turn the thumbscrew
counter-clockwise.
4. Press the latch release button and pu ll the probe away from the instrument.
Installation
P7700 Series TriMode Probes User Manual 5
Installation
Probe and probe tip power-on
After the con nection to the oscilloscope is made:
1. The probe briey turns on all LEDs during a self-test, and then the A – B Input Mode LED remains on. If a probe tip is attached, an LED on the tip also turns on when properly attached to the probe.
2. The probe transfe rs data from the probe and tip to the host instrument, and a message displays on the instrument as the transfer occurs.
The data transfer takes a few minutes, and is only don e when the host instrument discovers a new probe or new probe tip. The data transfer only occurs on instruments that are fully compatible with the probe.
3. After the data transfer is done, the probe is ready for a fun c tional check and calibration. (See page 11, Functional check.)
NOTE. If the probe’s warning light remains on, the power-on self-test likely failed.
(See page 76, Error conditions.)
6 P7700 Series TriMode Probes User Manual
Connect P7700 accessories to the TekFlex connector
All of the P7700 accessories mate with the new TekFlex (zero insertion force) connector. This connector provides an easy connection with one handed operation for attaching to the P7700 series accessories. T he following procedures show the steps for connecting probe tips to the TekFlex conne ctor; the procedures for connecting to other accessories are similar.
Connect the tip to the probe TekFlex connector as follows:
1. Pinch the TekFlex connector to open the
jaws.
2. Orient the tip with the probe head (notch to
the left) and slide the tip connector into the TekFlex connector.
A green LED will light up when the tip is inserted. The LED is the rst indication that the tip is powered and is inserted. You also need to conrm that the connector p in s on the TekFlex connector t through the alignment holes on the tip.
3. Once the tip is fully inserted, release the
pinch and the TekFlex connector closes. When closed and properly seated, the top of connector will be ush with the probe tip housing.
Installation
P7700 Series TriMode Probes User Manual 7
Installation
Probe comp box controls and indicators
Input mode buttons and LEDs
The P7700 Series solder-in tips support TriMode operation. TriMode enables you to switch the probe between four different measurement types without changing the probe’s connection:
NOTE. You can also change the Input Mode in the oscillosco pe Probe Setup screen.
Press the Input Mode button to select one of the four TriMode measurements. The modes cycle in the following sequence:
A – B (for differential signal measurement)
A – GND (for A inp ut single-ended measurement)
B – GND (for B inp ut single-ended measurement)
(A+B)/2–GND(forcommonmode measurement)
8 P7700 Series TriMode Probes User Manual
Installation
Warning LED
The Warning LED glows amber under the following conditions:
Probe power-on self test failure
Probe tip over-temperature detected
The input voltage on either the A or B input exceeds the allowable limit
The Warning LED clears when the condition causing the warning is removed. A notier message of a probe warning condition can also be seen on the oscilloscope.
CAUTION. Do not exceed the input voltage limits of the probe and probe tips. The probe or oscilloscope circuits may be damaged
if the limits are exceeded. Make sure that you understand and work within the limits of the pro be and probe tips.
P7700 Series TriMode Probes User Manual 9
Installation
TriMode probing
TheTriModefeatureallowsyoutoview two single-ended s ignals and the resultant differential waveform and common-mode voltage without moving the probe connection. Press the Input Mode button to cycle through the waveform views.
This example shows a typical signal on the A and B inputs. The resultant differential waveform and common-mode voltage are shown.
10 P7700 Series TriMode Pro bes User Manual
Functional check and calibration
After you connect the probe to the oscilloscope, you can perform a functional check using either the optional deskew xture designed for the probe or the optional P77C292MM Adapter.
CAUTION. To avoid ESD damage to the probe, always use an antistatic wrist strap (provided with your probe), and work at
a static-approved workstation when you handle the probe.
Functional check
This procedure checks the four TriMode settings on the probe, using the FAST EDGE signal from the front panel of the oscilloscope. It veries the signal path o f both the probe A and B input signals and their combination in the four TriMode settings. The P7700 Series Probe deskew xture is used to connect the FAST EDGE signal to the probe tip input. (See Table 2.) Alternately, the P7700 Series probe with the optional coaxial input adapter can connect directly to the FAST EDGE signal. (See Table 3 on page 12.)
Table 2: Required equ ipment, functional check with probe tip
Item Description Performance Requirement Recommended Example
Oscilloscope TekConnect Interface Tektronix MSO/DPO70000C/70000DX
Probe tip
Test board
Solder tip or Browser tip P77STFLXA
Probe deskew xture P 77DESKEW
Functional check and calibration
1
2
P7700 Series TriMode Probes User Manual 11
Functional check and calibration
Table 2: Required equipment, functional check with probe tip (cont.)
Item Description Performance Requirement Recommended Example
USB cable assembly USB 2.0, A male to Micro B male, 1 m 174-6919-xx (included with
Coaxial cable SMA, 50, male-to-male
1
Nine-digit part numbers (xxx-xxxx-xx) are Tektronix part numbers
2
Optional accessory
Table 3: Required equipment, functional check with SMA adapter
Item Description Performance Requirement Recommended Example
Oscilloscope TekConnect Interface Tektronix MSO/DPO70000C/70000DX
SMA adapter Coaxial input SMA adapter P77C292MM
1
Optional accessory
1
P77DESKEW)
174-1120-xx
1
12 P7700 Series TriMode Pro bes User Manual
Functional check and calibration
Test setup with the P77DESKEW Fixture
1. Connect the probe to any channel (1-4)
of the oscilloscope (connection to CH4 is shown). Set the oscilloscope to display the connected channel.
2. Connect the USB cable assembly supplied
with the P77DESKEW xture between the USB connector on the xture board and a USB connector on the host oscilloscope. Several white LEDs under the xture Port1 and Port2 probe tip clamps will light up when the xtu re is attached to USB power.
3. Connect an SMA cable from the FAST
EDGE output connector on the oscilloscope to the A input of the Probe Deskew Fixture.
4. Connect a P7700 Series probe tip to either the Port1 or Port2 on the P77DESKEW xture. If a TekFlex solder-in tip is used
to make this connection, its tip input should be inserted into the plastic clamp of the deskew xture port. This is done by compressing the spring-loaded clamp, inserting the probe tip input into the clamp, and then releasing the clamp so that it locks the connection.
P7700 Series TriMode Probes User Manual 13
Functional check and calibration
An additional set of green LEDs will light up under the probe tip clamp when the solder-down probe tip is properly inserted into the clamp.
Finally, connect the TekFlex connector at the end of the probe main cable to the probe tip inserted into the deskew xture port. An LED on the probe tip will also light up when the tip is properly inserted into the TekFlex connector of the probe.
5. If a P77BRWSR tip is used instead of a solder-down tip to make connection to the Deskew Fixture, th e TekFlex connector of the probe should rst be connected to the P77BRWSR tip. The P77BRWSR tip inputs must t hen be pressed into place against the A and B signal traces on the P77DESKEW xture board edge, located between the Port1 and Port2 clamps. Either of the two sets of A and B signal trace connection patterns located on the P77DESKEW xture board edge may be used.
14 P7700 Series TriMode Pro bes User Manual
Functional check and calibration
Test setup with the P77C292MM adapter
1. Connect the probe to any channel (1-4)
of the oscilloscope (connection to CH4 is shown). Set the oscilloscope to display the connected channel. Connect the probe to the P77C292MM adapter.
2. Connect the A cable from the P77C292MM
adapter to the FAST E DGE output connector on the oscilloscope. Leave the B cable disconnected for now.
Test procedure
1. Set the prob e Input Mode to A-B.
2. Adjust the oscilloscope to display a stable waveform (or press the Autoset button). The FAST EDGE signal is a 1 KHz square
wave. The amplitude of the FAST EDGE signal is attenuated b y 2X, compared to a single path termination, due to the power splitter built into the P77DESKEW xture.
P7700 Series TriMode Probes User Manual 15
Functional check and calibration
3. When you see a stable square waveform, check the amplitude (use the horizontal cursors). The atten uated amplitude displayed for a DPO70000 oscilloscope FAST EDGE signal routed throu gh the Deskew Fixture should be about 100 mVpp.
4. Cycle the Input Mode button through the remaining selections and compare the displayed waveforms to the waveform measured in the setup steps.
A-B
A-GND (same amplitude and polarity as preivously measured)
B-GND (the B input is grounded; no signal is measured)
(A+B)/2- GND (half-amp litude, but the same polarity as previously measured)
16 P7700 Series TriMode Pro bes User Manual