Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously
published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
EZ-Probe is a registered trademark of Cascade Microtech, Inc.
TriMode is a trademark of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O . Bo x 5 0 0
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tektronix.com to find contacts in your area.
Warranty
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If
any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts
and labor, or will provide a replacement in exchange for the defective product. Parts, modules and replacement products used by Tektronix for
warranty work may be new or reconditioned to like new performance. All replaced parts, modules and products become the property of Tektronix.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make
suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service
center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a
location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties,
taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. T ektronix
shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix
representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to
repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modified or integrated with
other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR
IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY
PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY
INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS
ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
[W2 – 15AUG04]
Table of Contents
General safetysummary ...................................................................................................................iii
User-Replaceable Parts .............................................................................................................64
Handling the Probe ..................................................................................................................68
Cleaning the Probe ..................................................................................................................69
Returning the Probe for Servicing ...................................................................................................70
iiP7630 30 GHz TriMode Probe Quick Start User Manual
General safety summary
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it.
To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
To avoid fire or personal i njury
General safety summary
Connect and disconnect properly.
circuit under test. Connect the probe reference lead to the circuit under test before connecting the probe input. Disconnect the probe
input and the probe reference lead from the circuit under test before disconnecting the probe from the measurement instrument.
Observe all terminal ratings.
manual for further ratings information before making connections to the product.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Do not operate without covers.
Do not operate with suspected failures.
service personnel.
Avoid exposed circuitry.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
P7630 30 GHz TriMode Probe Quick Start User Manualiii
Do not touch exposed connections and components when power is present.
Connect the probe output to the measurement instrument before connecting the probe to the
To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the product
Do not operate this product with covers or panels removed.
If you suspect that there is damage to this product, have it inspected by qualified
General safety summary
Terms in this manual
These terms may appear in this manual:
WARNING.
CAUTION.
Warning statements identify conditions or practices that could result in injury or loss of life.
Caution statements identify conditions or practices that could result in damage to this product or other property.
Symbols and terms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
ivP7630 30 GHz TriMode Probe Quick Start User Manual
Environmental Considerations
This section provides information about the environmental impact of the product.
Product End-of-Life Handling
Observe the following guidelines when recycling an instrument or component:
Environmental Considerations
Equipment Recycling.
contain substances that could be harmful to the environment or human health if improperly handled at the product’s end of life. In
order to avoid release of such substances into the environment and to reduce the use of natural resources, we encourage you to
recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately.
The symbol shown below indicates that this product complies with the European Union’s requirements according to Directive
2002/96/EC on waste electrical and electronic equipment (WEEE). For information about recycling options, check the
Support/Service section of the Tektronix Web site (www.tektronix.com).
P7630 30 GHz TriMode Probe Quick Start User Manualv
Production of this equipment required the extraction and use of natural resources. The equipment may
Environmental Considerations
Restriction of Hazardous Substances
This product is classified as an industrial monitoring and control instrument accessory, and is not required to comply with the
substance restrictions of the recast RoHS Directive 2011/65/EU until July 22, 2017.
viP7630 30 GHz TriMode Probe Quick Start User Manual
Preface
This manual describes the installation and operation of the P7630 30 GHz TriMode Probe. Basic probe operations and concepts
are presented in this manual. All documents listed below are located on the Documentation CD that came with your product. You
can also access the Tektronix Web site for these documents (www.tektronix.com/manuals).
Documentation
To read aboutUse these documents
Installation and operation (overviews)Read this quick start user manual for general information about how to use
In-depth operation
SpecificationsUse the technical reference manual.
Reordering accessories
your probe.
Use the technical reference manual (located on your documentation
CD-ROM) along with the user manual.
Accessories and Options
Use the
section in this manual.
Preface
P7630 30 GHz TriMode Probe Quick Start User Manualvii
Preface
viiiP7630 30 GHz TriMode Probe Quick Start User Manual
Key Features
The P7630 TriMode Probe allows you to take differential, single-ended, and common mode measurements with one probe
connection. Key features include:
Revolutionary TriMode operation
TekConnect interface
>30 GHz Bandwidth (typical)
<16 ps Rise time 10-90%
Key Features
<1 mV
adapters
Automatic tip adapter type identification
DSP-corrected response provided for
improved measurement fidelity
P7630 30 GHz TriMode Probe Quick Start User Manual1
Input voltage range
(DC + peak AC, input referenced
to ground)
Maximum nondestructive voltage±5 V±8 V
Temperature
Humidity
Altitude
Pollution degree2, Indoor use only
Operating
Nonoperating
Operating
Nonoperating
Operating
Nonoperating
1.2 V
2.0 V
(single-ended)
p-p
(differential)
p-p
6.0 V
10.0 V
(single-ended)
p-p
(differential)
p-p
+4.0 V, –4.0 V+5.0 V, –5.0 V
0 to +40 °C (+32 °F to +104 °F)
-20 °C to +60 °C (-4 °F to +140 °F)
Up to +40 °C (+104 °F) 20%-80% RH
+30 °C to +46 °C (+86 °F to +115 °F) 0-90% RH
3000 meters (9842 feet)
12000 meters (39,370 feet)
CAUTION.
To avoid ESD damage to the probe, always use an antistatic wrist strap (provided with your probe), and work at
a static-approved workstation when you handle the probe.
2P7630 30 GHz TriMode Probe Quick Start User Manual
Installation
Installation
Before you connect the probe to your instrument, read the
properly install the probe and adapters.
CAUTION.
a static-approved workstation when you handle the probe.
Overview
1.
2.
3.
4.
5.
6.
7.
Overview
To avoid ESD damage to the probe, always use an antistatic wrist strap (provided with your probe), and work at
Connect a P76xxx adapter to the probe. This step should be done before you connect the probe to the oscilloscope.
Connect the probe to the oscilloscope.
The probe performs a self test, and then one Input Mode LED remains lit.
The oscilloscope discovers the probe (if this is the first time they are connected), and downloads S-parameter data from the
probe. The oscilloscope then computes the filters. A message displays until the process is complete.
The Probe Setup screen displays (if this is a first-time connection).
Perform a functional check on the probe.
Use the Probe Setup screen to set the probe parameters as described in the
below to understand the sequence of events necessary to
Basic Operation
section.
P7630 30 GHz TriMode Probe Quick Start User Manual3
Installation
Connect TriMode Adapters to the Probe Body
TriMode adapters are required to complete the connection between the probe and your circuit, and are briefly described below. The
adapters connect the P7630 probe to your circuit through 2.92 mm or SMP cables, or through P7500 Series solder tips, such as the
P75PST Performance Solder Tip. The adapter inputs are polarized, with the A input marked in red and the B input marked in black.
The adapters should be connected to the probe before the probe is connected to the instrument.
1. P76CA-292
Use this adapter with high quality,
skew-matched cables that have male
2.92 mm connectors at the probe end. The
other end of the cables can be customized with
connectors that mate to your circuit.
2. P76CA-292C
This adapter includes a pair of skew-matched,
6 inch cables with male 2.92 mm connectors.
3. P76CA-SMP
This adapter includes a pair of skew-matched,
6 inch cables with female SMP connectors.
4. P76TA
Use this adapter with P7500 Series probe
solder tips. (See page 39,
Circuit Board
.)
Connecting to a
4P7630 30 GHz TriMode Probe Quick Start User Manual
Connect the Adapters
All of the adapters mate to the P7630 probe
head through a keyed, multi-pin connector
that transfers the adapter information to the
oscilloscope for automatic identification.
Connect them as follows:
Orient the adapter to the probe body with
1.
the A and B inputs on top, as shown.
Insert the adapter into the probe head.
2.
Attach the adapter to the probe head with
3.
the hex tool included with the probe. Use
the long end of the tool as shown, and only
use enough force to ensure a secure fit.
CAUTION.
be careful not to damage the input cables on
adapters with integral cables.
As you tighten the adapter screw,
Installation
CAUTION.
screw. You can damage the connectors if you
apply too much torque to the screw.
P7630 30 GHz TriMode Probe Quick Start User Manual5
Do not overtighten the adapter
Installation
Connect to the Host Instrument
Your TekConnect instrument may
NOTE.
require a firmware upgrade to support full
functionality of the P7630 probe. Before
you connect the probe, check the version
requirements. (See page 61, Host Instrument
Firmware.)
Slide the probe into the TekConnect
1.
receptacle. The probe clicks into place
when fully engaged.
Turn the thumbscrew clockwise (finger-tight
2.
only) to secure the probe to the instrument.
Disconnect
To disconnect, turn the thumbscrew
3.
counter-clockwise.
Press the latch release button and pull the
4.
probe away from the instrument.
6P7630 30 GHz TriMode Probe Quick Start User Manual
Probe Power-On
After the initial connection to the oscilloscope
is made:
The probe briefly lights all LEDs during a
1.
self-test, and then the A – B Input Mode
LED remains lit.
The probe transfers data to the host
2.
instrument, and a message displays on the
instrument as the transfer occurs.
The data transfer takes a few minutes,
and is only done when the host instrument
discovers a new probe. The data transfer
only occurs on instruments that are fully
compatible with the probe.
After the data transfer is done, the probe is
3.
ready for a functional check and calibration.
(See page 11,
Functional Check
.)
Installation
P7630 30 GHz TriMode Probe Quick Start User Manual7
Installation
Control Box Controls and Indicators
Input Mode Button and LEDs
Press the Input Mode button to select one of
the four TriMode measurements. The modes
cycle in the following sequence:
A – B (for differential signal measurement)
A – GND (for A input single-ended
measurement)
B – GND (for B input single-ended
measurement)
(A + B)/2 – GND (for common mode
measurement)
You can also change the Input Mode
NOTE.
in the oscilloscope Probe Setup screen. (See
Figure 4 on page 25.)
8P7630 30 GHz TriMode Probe Quick Start User Manual
Overload LED
The Overload LED glows amber under the
following conditions when using the P76CA-xxx
adapters:
• the input voltage on either the A or B input
exceeds ±4.5 V
• the termination voltage driver current on
the A or B input exceeds 50 mA
(See page 23,
The Overload LED clears when the input signal
is removed.
The Overload LED may flash briefly
NOTE.
when you adjust the termination voltage levels;
this is normal.
Termination Voltage
.)
Installation
CAUTION.
the limits are exceeded. Make sure that you understand and work within the limits of the probe and adapters.
P7630 30 GHz TriMode Probe Quick Start User Manual9
Do not exceed the input voltage limits of the probe and adapters. The probe or oscilloscope circuits may be damaged if
Installation
TriMode Probing
The TriMode feature allows you to view
two single-ended signals and the resultant
differential waveform and common-mode
voltage without moving the probe connection.
Press the Input Mode button to cycle through
the waveform views.
This example shows a typical HDMI signal (one
half-lane) on the A and B inputs. The resultant
differential waveform and common-mode
voltage are shown.
10P7630 30 GHz TriMode Probe Quick Start User Manual
Functional C heck and Calibration
After you connect the probe to the oscilloscope, perform a functional check using the procedure below.
Functional Check
This procedure checks the four TriMode settings on the probe, using the FAST EDGE connection on the front panel of the
oscilloscope. The A-B (differential mode) is set up and verified first, and then the remaining input modes are checked and
compared to the differential mode measurement.
Table 2: Required Equipment
Item descriptionPerformance requirementRecommended example
OscilloscopeTekConnect InterfaceTektronix DPO72504D or DPO73304D
Probe adapter
Coaxial cableSMA, 50 Ω, male-to-male
Adapter
Probe calibration
fixture
P76CA-xxx or P76TA probe
adapter
SMP-to-SMAFairview Microwave part number SM8810
Interconnect between P76TA
adapter and oscilloscope
Functional Check and Calibration
P76CA-292C (connects directly to oscilloscope FAST EDGE connector)
P76CA-292 (requires a coaxial cable)
P76CA-SMP (requires an SMP-to-SMA adapter)
P76TA (requires a coaxial cable and the calibration fixture listed below)
Tektronix part number 174-1120-xx
Tektronix part number 067-1821-xx
P7630 30 GHz TriMode Probe Quick Start User Manual11
Functional Check and Calibration
Test Setup
Connect the probe to any channel (1–4) of
1.
the oscilloscope.
Set the oscilloscope to display the channel.
2.
Connect one of the probe inputs to the
3.
FAST EDGE output of the oscilloscope,
using the P76CA-292C adapter.
You can also make the connection with one
of the other P76xxx adapters, but you must
use additional equipment. Refer to the
Required Equipment
on page 11.)
Leave the other probe input open.
4.
Set the termination and offset voltages to
5.
0 volts in the Probe Setup screen. (From
the oscilloscope menu, s elect Vertical and
then select Probe Cal.)
table. (See Table 2
12P7630 30 GHz TriMode Probe Quick Start User Manual