P7508 & P7513A TriMode Probes
Manuals and Related Documentation
The P7500 Series TriMode Probe manuals that are included with your P7508 and
P7513A probes contain all of the operating information necessary for you to
use these probes. However, the probe connection to the DC calibration board is
slightly different and is shown on the back of this sheet.
The information and specifications for the P7508 and P7513A probes in this
sheet supersede the information in the P7500 Series Technical Reference on the
Documenta
tion CD.
Probe
model Characteristic Probe only
P7508
513A
P7
P7508 &
P7513A
1
Measurements taken using an embedded probe fixtureanda250mVstep@18°Cto28°C(64°Fto82°F).
2
Warranted specification
Bandwidth
Rise time
CMRR >60 dB a
DMRR >40 dB
nnel isolation
Cha
Delay tim e 5.76 ns, ±0.1 ns
Channel isolation >15 dB to 7 GHz
Delay tim e 5.76 ns, ±0.1 ns
robe length
P
DC Calibration Board
and Socket Cable
(standard accessories)
Optional accessories
for the P7508 &
P7513A probes
>8 GHz >8 GHz >8 GHz
10–90% 55 ps
20–80%
>40dBto50MHz
>30dBto1GHz
>25 dB
>20dBto8GHz
>30dBto1GHz
>25dBto4GHz
>20 d
>40dBto50MHz
>30
>20dBto4GHz
>12dBto8GHz
>8 dB to 13 GHz
1.3 m (51.6 in) (End-to-end with no accessories attached)
oard
B
Socket Cable
Antistatic wrist strap
Bullet removal tool
Replacement bullets
1
2
2
35 ps
tDC
to4GHz
to 50 MHz
Bto8GHz
dBto1GHz
Probe with
P75TLRST solder tip
10–90% 55 ps
20–80%
>60 dB a
>40dBto50MHz
>30 dB to 1 GHz
>25 dB
>20 dB to 8 GHz
>40 dB
>30 dB to 1 GHz
>25 dB to 4 GHz
>20 d
>40dBto50MHz
>30
>15 dB to 4 GHz
>10 dB to 8 GHz
T
Tektronix part number 020-2954-XX
Order Tektronix part number 006-3415-XX
Order Tektronix part number 003-1896-XX
Order Tektronix part number 013-0359-XX
35 ps
tDC
to 4 GHz
to 50 MHz
Bto8GHz
dB to 1 GHz
ektronix part number 067-1967-XX
Probe with
P75PDPM adapter
10–90% 55 ps
20–80%
>60 dB a
>40dBto50MHz
>30dBto1GHz
>25 dB
>20dBto8GHz
Not ap
Not applicable
35 ps
tDC
to4GHz
plicable
www.tektronix.com
Copyright © Tektronix
*P001144900*
001-1449-00

P7508 & P7513A TriMode Probes Manuals and Related Documentation
DC Calibration Board
Connection
When you perfor
P7513A probes, you must use the Socket Cable included with the probes to
connect between the probe and the board. See the illustration below.
m a functional check or DC calibration on the P7508 and
Figure 1: Connecting the P7508 and P7513A probes to the DC Calibration Board with the Socket Cable
For future reference, s ave this document with your probe documentation kit.
2 P7508 & P7513A TriMode Probes Supplement