Tektronix P7360A, P7340A Reference manual

Technical Reference
P7360A and P7340A 6 GHz and 4 GHz Z-Active
t Differential Probes
071-1868-01
www.tektronix.com
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX, TEK, TekConnect, and Z-Active are registered trademarks of Tektronix, Inc.
Velcro is a registered trademark of Velcro Industries B.V.
Tip-Clip is a trademark of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
H In North America, call 1-800-833-9200.
H Worldwide, visit www.tektronix.com to find contacts in your area.

Table of Contents

General Safety Summary v...................................
Introduction 1..............................................
Theory of Operation 3.......................................
Input Voltage Limits 3..............................................
Maximum Input Voltage 3.......................................
Operating Voltage Window 4.....................................
Differential-Mode Signal Range 4.................................
Common-Mode Rejection 4.........................................
Probing Techniques to Maximize Signal Fidelity 5.......................
Input Impedance and Probe Loading 6.................................
Electrical Effects of Accessories 8....................................
Reference 9.................................................
Single-Ended Measurements 9.......................................
Differential Measurements 10.........................................
Common-Mode Rejection Ratio 10.................................
Assessing CMRR Error 11........................................
Input Impedance Effects on CMRR 11..............................
Serial Bus Standards 12..............................................
Specifications 13.............................................
Warranted Characteristics 13..........................................
Typical Characteristics 14............................................
Nominal Characteristics 17...........................................
Tip-Clip Assembly Specifications 20...................................
Performance Verification 29...................................
Equipment Required 29..............................................
Special Adapters Required 31.........................................
Equipment Setup 33................................................
Output Offset Voltage 34.............................................
DC Gain Accuracy 35...............................................
Gain Check at 5X Attenuation 35..................................
Gain Check at 25X Attenuation 36.................................
Rise Time 37......................................................
Rise Time Measurement using the Probe Calibration Fixture 38..........
Rise Time Check at 25X Attenuation 39.............................
Test Record 43.....................................................
User Service 45..............................................
Probe/Adapter/Oscilloscope Compatibility 45............................
Error Condition 45..................................................
Replacement Parts 45...............................................
Preparation for Shipment 46..........................................
P7360A and P7340A Z-ActiveDifferential Probes
i
Table of Contents

List of Figures

Figure 1: Dynamic range 4....................................
Figure 2: Use the Short Flex, Small Resistor Tip-Clip assembly 5...
Figure 3: Typical probe input model 6..........................
Figure 4: Symmetric coupled line 7............................
Figure 5: Transmission line equivalent 7........................
Figure 6: Lumped element equivalent 7.........................
Figure 7: Simplified model of a differential amplifier 10............
Figure 8: Typical Common-Mode Rejection Ratio (5X attenuation) 15
Figure 9: Typical differential input impedance versus frequency 15..
Figure 10: Probe and Tip-Clip dimensions 16.....................
Figure 11: Z-Active probe dynamic range verses frequency
5X gain setting 17.........................................
Figure 12: Z-Active probe dynamic range verses frequency
25X gain setting 18.......................................
Figure 13: P7380A, P7360A, and P7340A dynamic range
5X gain setting 18.........................................
Figure 14: P7380A, P7360A, and P7340A dynamic range
25X gain setting 19........................................
Figure 15: TekConnect-to-SMA Adapter 31......................
Figure 16: Probe Calibration Fixture 31.........................
Figure 17: Adapter Fixture 32..................................
Figure 18: Preliminary test setup 33.............................
Figure 19: Setup for the output offset zero test 34..................
Figure 20: DC Gain Accuracy setup 35..........................
Figure 21: Reverse the power supply polarity on the probe inputs 36.
Figure 22: Handheld Adapter and calibration fixture 38............
Figure 23: PPM203B Articulated Arm with the Handheld Adapter 38
Figure 24: Test system rise time setup 39.........................
Figure 25: Setting the TDR parameters 40.......................
Figure 26: Test probe rise time setup 41..........................
ii
P7360A and P7340A Z-ActiveDifferential Probes

List of Tables

Table of Contents
T able 1: Offset ranges 9......................................
T able 3: Warranted electrical characteristics 13.................
T able 4: Typical electrical characteristics 14......................
Table 5: Typical mechanical characteristics 16....................
T able 6: Nominal electrical characteristics 17.....................
Table 7: Test equipment 29....................................
Table 8: Differential probe compatibility issues 45................
P7360A and P7340A Z-ActiveDifferential Probes
iii
Table of Contents
iv
P7360A and P7340A Z-ActiveDifferential Probes

General Safety Summary

Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the system. Read the General Safety Summary in other system manuals for warnings and cautions related to operating the system.
ToAvoidFireor
Personal Injury
Connect and Disconnect Properly. Connect the probe output to the measurement instrument before connecting the probe to the circuit under test. Disconnect the probe input from the circuit under test before disconnecting the probe from the measurement instrument.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Do Not Operate Without Covers. Do not operate this product with covers or panels removed.
Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present.
Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
P7360A and P7340A Z-ActiveDifferential Probes
v
General Safety Summary
Symbols and Terms
Terms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Symbol on the Product. The following symbol may appear on the product:
CAUTION
Refer to Manual
vi
P7360A and P7340A Z-ActiveDifferential Probes

Introduction

This manual discusses topics not covered or otherwise mentioned briefly in the
P7313 12.5 GHz, P7380A 8 GHz, P7360A 6 GHz, and P7340A 4 GHz Z-Active Differential Probe Family User Manual.
The following is a list of brief explanations:
H Theory of Operation — Contains probe details not mentioned in the user
manual.
H Reference — Contains information about differential measurements and how
to increase measurement accuracy.
H Specifications — Contains warranted, typical, and nominal characteristics
for the probe and probe Tip-Clip Assemblies.
H Performance Verification — Describes the procedures for verifying the
warranted specifications.
H User Service — Describes troubleshooting and probe maintenance.
P7360A and P7340A Z-ActiveDifferential Probes
1
Introduction
2
P7360A and P7340A Z-ActiveDifferential Probes

Theory of Operation

This section discusses operating considerations and probing techniques. For more detailed information about differential measurements and common-mode rejection ratio (CMRR), see the Reference sectiononpage9.
These probes are optimized for high bandwidth; they are not general purpose probes. The probe head and tips are miniaturized for electrical characteristics and access to dense circuitry, and must be handled carefully.
CAUTION. To prevent damage to the probes, use care when handling the probes. Rough or careless use can damage the probes.

Input Voltage Limits

These probes are designed to probe low-voltage circuits. Before probing a voltage, take into account the limits for maximum input voltage, the common­mode signal range, and the differential-mode signal range. For specific limits, refer to page 14.

Maximum Input Voltage

The maximum input voltage is the maximum voltage to ground that the inputs can withstand without damaging the probe input circuitry.
CAUTION. To avoid damaging the inputs of the P7360A and P7340A Differential Probes, do not apply more than ±15 V (DC + peak AC) between each input or between either probe input and ground.
P7360A and P7340A Z-ActiveDifferential Probes
3
Theory of Operation

Operating Voltage Window

Differential-Mode Signal
Range
The operating voltage window defines the maximum voltage that you can apply to each input, with respect to earth ground, without saturating the probe input circuitry. See Figure 1. A common-mode voltage that exceeds the operating voltage window may produce an erroneous output waveform even when the differe ntial- mode specification is met. For specifications, refer to page 14.
The differential-mode signal range is the maximum voltage difference between the plus and minus inputs that the probe can accept without distorting the signal. The distortion from a voltage that is too large can result in a clipped or otherwise inaccurate measurement. For specifications, refer to page 14.
1.3 V
+4.0 V
5X
25X
Figure 1: Dynamic range

Common-Mode Rejection

The common-mode rejection ratio (CMRR) is the ability of a probe to reject signals that are common to both inputs. More precisely, CMRR is the ratio of the differential gain to the common-mode gain. The higher the ratio, the greater the ability to reject common-mode signals. CMRR varies with frequency, usually decreasing at higher frequencies. For additional information about CMRR, see page 15.
5.0 V
-- 3 . 0 V
4
P7360A and P7340A Z-ActiveDifferential Probes

Probing Techniques to Maximize Signal Fidelity

Signal fidelity is an indication of how accurately a probe represents the signal being measured. The signal fidelity of the probe is best when the probe is applied directly to the circuit with the Short Flex, Small Resistor, Tip-Clip assembly shown in Figure 2. This Tip-Clip assembly achieves high signal fidelity by minimizing the distance between the probe head and the signal source. This reduces probe interconnect parasitics which tend to degrade signal fidelity. However, some probing tasks are made easier using other accessories included with the probe.
The Tip-Clip specifications starting on page 20 show pulse response illustrations that give some indication of signal fidelity with different Tip-Clip assemblies.
Signal fidelity is affected by both the probe interconnect and the speed of the signal. It is recommended that the longer the Tip-Clip assembly be used with somewhat slower speed signals for better signal fidelity.
Theory of Operation
Figure 2: Use the Short Flex, Small Resistor Tip-Clip assembly
P7360A and P7340A Z-ActiveDifferential Probes
5
Theory of Operation

Input Impedance and Probe Loading

When you connect the probe inputs to a circuit, you are introducing a new resistance, capacitance, and inductance into the circuit. Each input of the differential probe has a characteristic input impedance of 50 kΩ to ground.
60 fF
3.12 pF
3.12 pF
60 fF
Input
+
Input
--
1
See the following figures for an explanation of cpl.
2
Short Tip-Clip (blue) length
38 Ω
38 Ω
37 Ω
20 fF
20 fF
37 Ω
1
Cpl
Zodd=95
Zeven=190
L=3.7 mm
Kodd ~ Keven ~ 1
2
140 Ω
50 kΩ
310 Ω
310 Ω
50 kΩ
140 Ω
Figure 3: Typical probe input model
For signals with low source impedance and frequency, the 50 kΩ input imped­ance on each input is large enough to prevent the inputs from loading the signal sources. As the signal source impedance on an input increases, the more the probe loads the source and reduces the signal amplitude. The greater the source impedances and the higher the signal frequencies, the more you must take these factors into account. See Figure 3.
The frequency of the signal also affects signal measurement. As the frequency of the signal increases, the input impedance of the probe decreases. The lower the impedance of the probe relative to that of the source, the more the probe loads the circuit under test and reduces the signal amplitude. For a graph of input impedance versus frequency, refer to Figure 9 on page 15.
6
P7360A and P7340A Z-ActiveDifferential Probes
Cpl
ZOO= 100 ZOE= 190
L=3.7mm(air)
Figure 4: Symmetric coupled line
Z0 = ZOE= 190 Ω,
Td = 12.1 ps
Z0 = 2(ZOOZOE)/(ZOE-- ZOO) =211 ,
Td = 12.1 ps
Theory of Operation
Z0 = ZOE= 190 Ω,
Td = 12.1 ps
TD = 12.1 ps
(3.7 mm in air)
Figure 5: Transmission line equivalent
Cg = (C11+C12)/2
= 31.9 ff
Cg = --0.5C
= 14.4ff
12
Lp = Ls = L11=1.76nH
= 0.545 nH
M=L
12
11+C12
)/2
Cg = (C
= 31.9ff
Cg = --0.5C
= 14.4 ff
12
Figure 6: Lumped element equivalent
P7360A and P7340A Z-ActiveDifferential Probes
7
Theory of Operation

Electrical Effects of Accessories

The Tip-Clip accessories included with your probe help connect to different types of components. The Tip-Clip accessories are designed to provide optimum performance as a system. Each Tip-Clip accessory has distinct characteristics. While these accessories make connections easier, be aware that the Tip-Clip accessory you choose affects the signal you are measuring, depending on a variety of factors, including signal frequency, source impedance, and lead length. Refer to Specifications on page 13 for more Tip-Clip information.
8
P7360A and P7340A Z-ActiveDifferential Probes

Reference

This section contains important reference information about differential measurements and how to increase measurement accuracy.

Single-Ended Measurements

A differential probe, for example the P7360A and P7340A Differential Probes, can be used for single-ended measurements within the limits of its dynamic and offset voltage ranges. Single-ended probes such as the P7240 typically have a wider offset range than corresponding differential probes (see Table 1).
Table 1: Offset ranges
Probe examples
P7240 +/-- 5 V 4V
P7360A and P7340A Differen­tial Probes
DC Offset, 5X Dynamic
Range, 5X
PP
+4 V, --3 V 2V
PP
DC Offset, 25X Dynamic
Range, 25X
-- -- --
+4 V, --3 V 5V
-- -- --
PP
Differential probes are ideal for a class of single-ended measurements where the reference voltage is not ground:
H SSTL_1,2: V
H PECL: V
TT,VREF=VDD
REF=VCC
/2
--1.3
To measure single-ended signals in this class, connect the negative input of the differential probe to V
REF
.
A differential probe in these applications displays the true signal despite any AC or DC variation in V displays the signal plus the variation in V
from its nominal value. While a single-ended probe
REF
.
REF
Differential probes can also be used to make ground referenced single-ended measurements on either single-ended signals or differential signals like PCI Ex­press or Serial ATA. To measure ground referenced single-ended signals, connect the negative input of the differeential probe to ground.
Single-ended measurement on differential signals are used to measure common mode voltage and check for differential signal symmetry.
P7360A and P7340A Z-ActiveDifferential Probes
9
Reference

Differential Measurements

A differential probe is optimized to make measurements of high speed differen­tial signals. Differential signals are formed from two complementary signals with a common reference voltage. See Figure 7.
Devices designed for differential measurements avoid problems presented by single-ended systems. These devices include a variety of differential probes, differential amplifiers, and isolators.
A differential probe is a differential amplifier (Figure 7), which is used to make differential measurements that reject any voltage that is common to the inputs and amplifies any difference between the inputs. Voltage that is common to both inputs is often referred to as the Common-Mode Voltage (VCM) and voltage that is different as the Differential-Mode Voltage (VDM).
Common-Mode Rejection
Ratio
Common mode
V
CM
V
DM
2
+
V
DM
2
Differential
+
+
Differential
mode
mode
+
V
--
out
Figure 7: Simplified m odel of a differential amplifier
Differential amplifiers cannot reject all of the common-mode signal. The ability of a differential amplifier to reject the common-mode signal is expressed as the Common-Mode Rejection Ratio (CMRR). The CMRR is the differential-mode gain (A
) divided by the common-mode gain (ACM). It is expressed either as a
DM
ratio or in dB.
A
DM
A
CM
CMRR =
A
DM
A
CM
dB = 20 log
10
CMRR generally is highest (best) at DC and degrades with increasing frequency.
P7360A and P7340A Z-ActiveDifferential Probes
Reference

Assessing CMRR Error

Input Impedance Effects
on CMRR
Figure 8 on page 15 shows the CMRR of a Z-Active differential probe. This derating graph assumes a sinusoidal common-mode signal.
A quick way to assess the magnitude of CMRR error when the common-mode signal is not sinusoidal is to connect both leads to the same point in the circuit. The oscilloscope displays only the common-mode component that is not fully rejected by the probe. While this technique may not give you accurate measure­ments, it does allow you to determine if the magnitude of the common-mode error signal is significant. Make the probe Tip-Clip wires the same length to maximize the probe CMRR.
The lower the input impedance of the probe relative to the source impedance, the lower the CMRR for a given source impedance imbalance. Differences in the source impedance driving the two inputs lowers the CMRR. Note that single-en­ded measurements generally result in asymmetric source impedances which tend to reduce the differential mode CMRR.
P7360A and P7340A Z-ActiveDifferential Probes
11
Reference
12
P7360A and P7340A Z-ActiveDifferential Probes

Specifications

The specifications in Tables 2 through 5 apply to the P7360A and P7340A Differential Probes installed on any TekConnect instrument or Tektronix 80A03 TekConnect adapter. When the probe is used with another oscilloscope, the oscilloscope must have an input impedance of 50 Ω. The probe must have a warm-up period of at least 20 minutes and be in an environment that does not exceed the limits described in Table 2. Specifications for these probes fall into three categories: warranted, typical, and nominal characteristics.

Warranted Characteristics

Warranted characteristics (Table 2) describe guaranteed performance within tolerance limits or certain type-tested requirements. Warranted characteristics that have checks in Table 2 are marked with the n symbol.
Table 2: Warranted electrical characteristics
Characteristic Description
n DC attenuation accuracy ±2%
n Output Zero ±3mV(+20to+30_C,+68to+86_F) (5X) ±15 mV on oscilloscope
±3mV(+20to+30_C,+68to+86_F) (25X) ±75 mV on oscilloscope
n Rise time <70 ps (probe only, Short Flex, Small Resistor Tip-Clip), all other Tip-Clip assemblies are
typical (P7360A) <100 ps (probe only, Short Flex, Small Resistor Tip-Clip), all other Tip-Clip assemblies are typical P7340A)
Temperature Operating: 0 to +40 _C(+32to+104_F),
Nonoperating: --55 to +75 _C (--131 to +167 _F)
Humidity Operating: 0--90% RH, tested at +0 to +40 _C(+32to+104_F)
Nonoperating: 0--90% RH, tested at --55 to +75 _C (+67 to +167 _F)
1
See warning that follows.
1
WARNING. To avoid a burn hazard at high ambient temperatures, do not touch the probe with bare hands at nonoperating temperatures above +75 _C (+167 _F). Allow sufficient time for the probe to cool before handling.
P7360A and P7340A Z-ActiveDifferential Probes
13
Specifications

Typical Characteristics

Typical characteristics (Tables 3 and 4) describe typical but not guaranteed performance.
Table 3: Typical electrical characteristics
Characteristic Description
Differential input resistance, DC coupled 104 kΩ ±2%
Common-mode input resistance, DC coupled
Differential offset range -- 3 . 0 V t o + 4 V
52 kΩ ±1kΩ
Noise
Input impedance See Figure 9 and Tip-Clip assemblies starting on page 20 for more detail information.
Bandwidth See Tip-Clip assemblies starting on page 20 for more detail information.
Small signal rise time See Tip-Clip assemblies starting on page 20 for more detail information.
Common-mode rejection ratio SeeFigure8
Maximum non destructive input voltage ±15 V
Dynamic Range
Operating Voltage Window --3.0 V to +5.0 V at (5X or 25X)
Linearity
< 31 nVHz5X
< 75 nVHz25X
(+ 2.0dB, --3 dB) for an ambient temperature range of 20 _Cto30_C
Ambient temperature range of 20 _Cto30_C
50 dB: DC — 1 MHz35 dB: >1 MHz — 1 GHz20 dB: >1 GHz — 6 GHz
(DC + peak AC)
±1 V at attenuation setting of 5X ±2.5 V at attenuation setting of 25X
±1 V at 2% at attenuation setting of 5X ±0.75 V at 1% at attenuation setting of 5X ±0.5 V at 0.5% at attenuation setting of 5X
±3 V at 2% at attenuation setting of 25X ±2.5 V at 1% at attenuation setting of 25X ±1.5 V at 0.5% at attenuation setting of 25X
between each input or between either probe inputs and ground.
DC offset drift 50 μV/_C (at the output of the probe)
0.25 mV/_C (displayed on screen with t he TekConnect interface) 5X
1.25 mV/_C (displayed on screen with t he TekConnect interface) 25X
DC voltage measurement accuracy ±〈2% of input + (2% of offset) + 15 mV + 5 mV) 5X
±〈2% of input + (2% of offset) + 75 mV + 25 mV) 25X
14
P7360A and P7340A Z-ActiveDifferential Probes
Differential Mode
Gain (5X)
dB
-- 1 4
-- 2 4
-- 3 4
-- 4 4
-- 5 4
Specifications
CMRR
Common Mode
Gain (5X)
-- 6 4
-- 7 4
-- 8 4 50 MHz
Note: Short Flex, Small Resistor Tip-Clip assembly graph
Frequency
Figure 8: Typical Common-Mode Rejection Ratio (5X attenuation)
10 GHz
Note: Short Flex, Small Resistor Tip-Clip assembly
Figure 9: Typical differential input impedance versus frequency
P7360A and P7340A Z-ActiveDifferential Probes
15
Specifications
Table 4: Typical mechanical characteristics
Characteristic Description
Dimensions, compensation box 107 mm × 41 mm × 26 mm (4.2 in × 1.6 in × 1.0 in)
Dimensions, probe head 19.43 mm × 3.30 mm × 7.6 mm (0.765 in × 0.130 in × 0.300 in)
Dimensions, cable length 1.5 m (60 in) (from the probe head to the compensation box)
Unit weight 1.406 kg (3.1 lbs) (probe, accessories and packaging)
Large Resistor, Flex
Tip-Clips, 20 mil dia wire
5.08 mm (.20 in)
3.30 mm (.130 in)
12.7 mm (.50 in)
4.77 mm (.188 in)
7.6 mm (.300 in)
25.4 mm (1.00 in)
19.43 mm (.765 in)
124.46 mm (.490 in)
Small Resistor, Flex
Tip--Clips, 8 mil dia wire
2.03 mm (.080 in)
89.78 mm (3.53in)
33.02 mm (1.3 in)
5.08 mm (.20 in)
25.4 mm (1.00 in)
26.0 mm (1.00 in)
41.0 mm (1.60 in)
107 mm
(4.2 in)
Figure 10: Probe and Tip-Clip dimensions
16
P7360A and P7340A Z-ActiveDifferential Probes

Nominal Characteristics

Nominal characteristics (Table 5) describe guaranteed traits, but the traits do not have tolerance limits.
Table 5: Nominal electrical characteristics
Characteristic Description
Input configuration Differential (two inputs, + and --)
Output coupling DC
Attenuation settings 5X and 25X
Termination Terminate output into 50 Ω
Specifications
Figure 11: Z-Active probe dynamic range verses frequency 5X gain setting
P7360A and P7340A Z-ActiveDifferential Probes
17
Specifications
Figure 12: Z-Active probe dynamic range verses frequency 25X gain setting
18
Figure 13: P7380A, P7360A, and P7340A dynam ic range 5X gain setting
P7360A and P7340A Z-ActiveDifferential Probes
Specifications
Figure 14: P7380A, P7360A, and P7340A dynam ic range 25X gain setting
P7360A and P7340A Z-ActiveDifferential Probes
19
Specifications

Tip-Clip Assembly Specifications

Note. All specifications are typical in the following Tip-Clip assemblies, unless otherwise indicated.
Short Flex, Small Resistor, Tip-Clip Assembly
Tektronix part number: 020-2600-XX
Bandwidth: >6.0 GHz 10/90 Rise time: <70 ps 20/80 Rise time: <50 ps
*
Guaranteed
Loading: Z
Best overall signal fidelity. The small resistors are ideal for connecting to small circuit board vias and fine pitch circuitry.
290 to 6 GHz
MIN
*
20
P7360A and P7340A Z-ActiveDifferential Probes
Short Flex, Large Resistor, 1/8 W Tip-Clip Assembly
Tektronix part number: 020-2601-XX
Bandwidth: >6.0 GHz 10/90 Rise time: <70 ps 20/80 Rise time: <50 ps
Specifications
Loading: Z
High bandwidth and good signal fidelity, idea for connecting to large features.
290 to 6 GHz
MIN
P7360A and P7340A Z-ActiveDifferential Probes
21
Specifications
Medium Flex, Small Resistor,
Tip-Clip Assembly
Tektronix part number: 020-2602-XX
Bandwidth: >5.0 GHz 10/90 Rise time: <70 ps
*
20/80 Rise time: <50 ps
*
Guaranteed
Loading: Z
290 to6GHz
MIN
Good compromise between ease-of-use and maximum performance when attaching to smaller devices or circuit board vias.
22
P7360A and P7340A Z-ActiveDifferential Probes
Medium Flex, Large Resistor, 1/8 W, Tip-Clip Assembly
Tektronix part number: 020-2603-XX
Bandwidth: >5.0 GHz 10/90 Rise time: <70 ps 20/80 Rise time: <50 ps
Specifications
Loading: Z
Good compromise between ease-of-use and maximum performance when attaching to larger devices.
260 to6GHz
MIN
P7360A and P7340A Z-ActiveDifferential Probes
23
Specifications
Long Flex, Small Resistor,
Tip-Clip Assembly
Tektronix part number: 020-2604-XX
Bandwidth: >4.0 GHz 10/90 Rise time: <130 ps 20/80 Rise time: <60 ps
Loading: Z
Extended reach with good Step Re­sponse. Useful for connecting to hard to reach small vias and fine-pitch circuitry. Conveniently sized to fit between DIMM modules. Not recommended for signals faster than 4 GHz.
360 to6GHz
MIN
24
P7360A and P7340A Z-ActiveDifferential Probes
Long Flex, Large Resistor, 1/8 W, Tip-Clip Assembly
Tektronix part number: 020-2605-XX
Bandwidth: >5.0 GHz 10/90 Rise time: <90 ps 20/80 Rise time: <60 ps
Specifications
Loading: Z
Extended reach with good step response. Useful for connecting to hard-to-reach circuitry with large features. Conveniently sized to fit between DIMM modules. Not recommended for signals faster than 4GHz.
300 to6GHz
MIN
P7360A and P7340A Z-ActiveDifferential Probes
25
Specifications
Square Pin, Tip-Clip Assembly
Tektronix part number: 020-2701-XX
Bandwidth: >6.0 GHz 10/90 Rise time: <70 ps 20/80 Rise time: <50 ps
Use the Square Pin Tip-Clip assembly for probing 0.025-in diameter square pins spaced 0.1-in on center. Square pins are not an ideal transmission path for high speed electrical signals. Square pins are not recommended for signals faster than 100 ps or 3 GHz.
Square Pin Tip-Clip assembly measuring differential 100 ps risetime signal with 0.34-in long square pins.
6
3
0
-- 3
dB
-- 6
-- 9
-- 1 2
1.00E+08 1.00E+09 1.00E+10 Frequency (Hz)
Square Pin Tip-Clip assembly frequency response with 0.34-in long square pins.
26
P7360A and P7340A Z-ActiveDifferential Probes
Variable Spacing Tip-Clip Assembly
Tektronix part number: 020-2596-XX
Bandwidth: >6 GHz
: 10/90 <70 ps, 20/80 <50 ps
T
R
Loading: Z
220 to6GHz
MIN
Use the Variable Spacing Tip-Clip assembly for probing test points spaced from 0.020-in to 0.180-in apart.
Exercise care when handling the articu­lated pins.
Specifications
P7360A and P7340A Z-ActiveDifferential Probes
27
Specifications
28
P7360A and P7340A Z-ActiveDifferential Probes

Performance Verification

The following procedures verify the warranted Output offset voltage, DC attenuation accuracy, and Rise time specifications of the P7360A and P7340A Differential Probes. The recommended calibration interval is one year.
CAUTION. To avoid ESD damage to the probe, always use an antistatic
wrist strap (provided with your probe), and work at a static-approved workstation when handling the probe.

Equipment Required

Table 6 lists the equipment required for the performance verification procedure. The types and quantities of connectors depends on the specific equipment you use.
Table 6: Test equipment
Description and quantity Performance requirement Recommended example
Sampling Oscilloscope Tektronix TDS 8200 Series
Sampling Module 20 GHz bandwidth Tektronix 80E04 or 80E03
Pulse Generator <25 ps rise time Tektronix 80E04 TDR
Sampling Head Extender Cable
TekConnect Probe Interface Module with semi-rigid cable
TekConnect-to-SMA adapter See page 31 Tektronix TCA-SMA
DMM (2), with leads
Dual Power Supply 5.0 VDC at 200 mA B+K Precision 1760A or
Coaxial cable
Test leads (1) Banana plug ends, red 012-0031-XX
Test leads (1) Banana plug ends, black 012-0039-XX
Test leads (2) Mini plunger with test clip Mueller BU-1120
Adapter
Adapter BNC(M)-to-Minigrabbers 013-0342-XX
Adapter SMA Male-to-BNC female 015-1018-XX
1m 012-1568-XX
Firmware version 2.0 Tektronix 80A03, with
174-4857-XX cable
0.1 mV and 0.01 Ω resolution
Male-to-Male BNC, 50 Ω
SMA 50 Ω termination (comes with the probe calibration fixture)
Fluke 187 or equivalent
equivalent
012-0057-XX
015-1022-XX
1
P7360A and P7340A Z-ActiveDifferential Probes
29
Performance Verification
Table 6: Test equipment (cont .)
Description and quantity Recommended example
Adapter (2) SMA Male-to-Male 015-1011-XX
Feed through termination
Probe calibration fixture See page 31 067-0419-XX
Variable Spacing, Tip-Clip Assembly
Long Flex, Small Resistor, 1/8 Watt Tip-Clip Assembly
Super glue Loctite 444 12292
SMA torque wrench 5/16-in, 7 in-lb.
1
Nine-digit part numbers (xxx-xxxx-xx) are Tektronix part numbers.
Performance requirement
BNC, 50 Ω ± 0.05 Ω
011-0129-XX
020-2596-XX
020-2604-XX
2
2
2 Standard accessories included with the probe.
H Optional Tool. A torque wrench helps to ensure reliable connections by
meeting the nominal torque values listed in these instructions.
1
30
P7360A and P7340A Z-ActiveDifferential Probes

Special Adapters Required

Performance Verification
Some of the adapters listed in Table 6 are available only from Tektronix. These adapters are described on the following pages.
TekConnect-to-SMA
Adapter

Probe Calibration Fixture

The TekConnect-to-SMA Adapter, Tektronix part number TCA-SMA, lets you connect an SMA cable to a TekConnect input. See Figure 15. Connect and disconnect the adapter the same way as you do the probe.
This adapter is an oscilloscope accessory that can be used for measurement applications, as well as these performance verification procedures.
Figure 15: TekConnect-to-SMA Adapter
Some of the procedures in this manual use a probe calibration fixture, Tektronix part number 067-0419-XX.
The calibration fixture provides a means to test the probe for common mode (CM) and differential mode (DM) measurements. SMA connectors on the front and back of the fixture allow you to apply stimulus signals.
Figure 16: Probe Calibration Fixture
P7360A and P7340A Z-ActiveDifferential Probes
31
Performance Verification
PPM203B Articulated Arm
and Probe Arm Adapter
Some of the procedures in this manual use a PPM203B Articulated Arm. This is a general purpose benchtop probe holder that provides a method for securing the probe and must be used with the Probe Arm Adapter. Use the following steps and Figure 17 to attach the Probe Arm Adapter to the PPM203B Articulated Arm.
1. Using the Hex wrench, remove the screw from the end of the articulated arm.
2. Using the Hex wrench, attach the Adapter bottom to the probe arm.
3. Loosen the Adapter top by turning the two thumb screws counter clockwise.
4. Place the probe in the loosened Adapter bottom (keyed).
5. Secure the Adapter top by tightening the two thumb screws.
Tighten
thumb
screws
Adapter top
Adapter to probe arm
Probe arm
connection
Figure 17: Adapter Fixture
Hex
wrench
Adapter
bottom
32
P7360A and P7340A Z-ActiveDifferential Probes

Equipment Setup

Performance Verification
CAUTION. To avoid ESD damage to the probe, always use an antistatic
wrist strap (provided with your probe), and work at a static-approved workstation when handling the probe.
Perform the following verification procedures in order.
Use the following procedure to set up and warm the equipment to test the probe.
1. Connect the 80A03 TekConnect probe interface to channels 3 and 4 of the TDS 8200 oscilloscope. See Figure 18.
2. Connect the 80E0X module to the 80A03 TekConnect probe interface.
3. Connect the probe to one of the 80A03 probe interface channels.
4. Turn on the oscilloscope and allow 20 minutes for the equipment to warm
up.
5. From the Utilities menu, select Utilities Compensation to compensate and save the compensation for module channels 3 and 4.
6. Photocopy the test record on page 43 to record the performance test results.
TDS/CSA 8200 Series Oscilloscope
AB
80A03
80E0X Module
CH 4 (measurement channel)
80A03 TekConnect probe interface module
P7360A/P7340A probe
Figure 18: Preliminary test setup
P7360A and P7340A Z-ActiveDifferential Probes
33
Performance Verification

Output Offset Voltage

NOTE. Before beginning these procedures, refer to page 43 and photocopy the test record (if you have not already done so) and use it to record the performance test results.
Use the following procedure to test the Output Offset voltage.
1. Connect the equipment as shown in Figure 19.
2. Plug the probe into the 80A03 module and warm up for 20 minutes, if not
done.
3. Short the two Tip-Clip leads together (see Figure 19.) We recommend using the BNC(m)-to-Minigrabber (black) with the Minigrabber not grounded.
TDS/CSA 8200 Series Oscilloscope
80A03
BNC-SMA
adapter
50 Ω Precision
termination
P7360A/P7340A
probe
BNC cable
--
+
Long Flex, Large
Resistor, 1/8 W
Tip-Clip Assembly
Figure 19: Setup for the output offset zero test
4. Set the multimeter to read DC volts.
5. Verify that the output voltage is 0 V
±3.0 mV for both the 5X and 25X
attenuation settings.
6. Record the results on the test record.
DMM
BNC-to-dual banana adapter
34
P7360A and P7340A Z-ActiveDifferential Probes

DC Gain Accuracy

Performance Verification
This test checks the DC gain accuracy of the probe at the 5X and 25X attenua­tion settings.
Gain Check at 5X
Attenuation
1. Set the probe attenuation to 5X.
2. Connect the probe with a Tip-Clip Assembly to the power supply as shown
in Figure 20. Monitor the source voltage with one of the DMMs.
DMM (V in)
TDS/CSA 8200 Series Oscilloscope
+
--
Power supply
Banana lead
Red
80A03
BNC-SMA
adapter
50 Ω Precision
termination
P7360A/P7340A
probe
Long Flex, Small
Resistor, 1/8 W
Tip-Clip Assembly
Black
--
+
BNC cable
Banana
lead
-- +
--
Test leads (2), w/mini plunger
BNC-to-dual
banana adapter
+
DMM (V out)
Figure 20: DC Gain Accuracy setup
3. Set the power supply to approximately +0.5 V. This represents 80% of the probe dynamic range in this attenuation setting. Record this source voltage
1.
as V
in
4. Record the output voltage (on the second DMM) as V
5. Disconnect the test leads from the power supplies. Leave the DMM leads
connected to the adapters.
6. Reverse the polarity of the voltage applied to the probe inputs by swapping both sets of banana leads at the power supply, as shown in Figure 21.
7. Record the actual source voltage (now a negative value), as V
P7360A and P7340A Z-ActiveDifferential Probes
out
1.
2.
in
35
Performance Verification
DMM (V in)
TDS/CSA 8200 Series Oscilloscope
P7360A/P7340A
80A03
BNC-SMA
adapter
50 Ω Precision
termination
probe
Banana
lead
Red
--
+
Long Flex, Small
Resistor, 1/8 W
Tip-Clip Assembly
BNC cable
--
+
Power supply
-- +
--
+
Test leads (2)
w/mini plunger
BNC-to-dual
banana adapter
Banana lead
Black
DMM (V out)
Figure 21: Reverse the power supply polarity on the probe inputs
8. Record the output voltage on the second DMM (now a negative value) as
2.
V
out
36
Gain Check at 25X
Attenuation
9. Calculate the gain as follows: (V
out
1--V
2) ÷ (Vin1--Vin2).
out
10. Verify that the gain is 0.2, ±2.0%.
11. Record the calculated gain for the 5X setting on the test record.
1. Set the attenuation on the probe to 25X.
2. Repeat steps 2 through 9, but in step 3, set the power supply to 1.5 V.
3. Verify that the gain is 0.04, ±2.0%.
4. Record the calculated gain on the test record.
P7360A and P7340A Z-ActiveDifferential Probes

Rise Time

Performance Verification
This procedure verifies that the probe meets the rise time specification. Two rise times are measured; the test system alone, and the test system with the probe included. The probe rise time is calculated using the two measurements.
This test uses the TDR function of the 80E04 sampling head as a fast rise time signal source. The measurements are made using an 80A03 TekConnect probe interface. Although the following procedure assigns the TDR and measurement functions to specific oscilloscope channels, any valid channel combination can be used. However, the TDR function is only available on 80E04 sampling heads.
This test checks both of the probe attenuation settings.
Review Rise Time Measurements Using the Probe Calibration Fixture on page 38 if you have not used a calibration fixture before.
P7360A and P7340A Z-ActiveDifferential Probes
37
Performance Verification
Rise Time Measurement
using the Probe
Calibration Fixture
1. Connect the fixture to the pulse generator using an SMA adapter.
2. Connect the 50
Ω termination included with the fixture to the unused SMA
connector.
3. Probe the calibration fixture using the Handheld Adapter (see Figure 23).
NOTE. It is recommended that you use the PPM203B Articulated Arm with the Probe Arm Adapter (see page 32) to stabilize the probe while verifying the differential rise time specification.
Signal
source
50 Ω
Termination
Figure 22: Handheld Adapter and calibration fixture
Figure 23: PPM203B Articulated Arm with the Handheld Adapter
PPM203B Articulated Arm
Signal source
50 Ω Termination
38
P7360A and P7340A Z-ActiveDifferential Probes
Performance Verification
Rise Time Check at 25X
Attenuation
1. Connect the standard 80A03 semi-rigid SMA connector between the 80A03 probe output and the 80E0X module input. Insert a TCA-SMA adapter into the TekConnect interface on the 80A03.
2. Connect the test equipment as shown in Figure 24. A sampling module extender cable is used with the TDR pulse generator to minimize cable loss problems.
CAUTION. To prevent mechanical strain on the connectors, use care when working with SMA connectors: Support equipment and use a torque wrench to tighten connections to 7 in-lbs.
TDS/CSA 8000 Series Oscilloscope
CH 7 and 8 (module slot)
CH 4 (measurement
channel)
80A03
80E0X sampling
module
SMA male-to-male
connector
TCA-SMA adapter
SMA male-to-male connector
Calibration fixture
80E04 TDR pulser
Sampling module extender cable
Figure 24: Test system rise time setup
NOTE. The 80A03 firmware version must be version V 2.0 or above. The firmware version label is on the rear panel of the instrument.
3. Turn on Channel 4, and set the vertical scale to 50 mV/div.
P7360A and P7340A Z-ActiveDifferential Probes
39
Performance Verification
4. Set the Channel 8 sampling head to TDR mode: Press the SETUP DIALOGS button and select the TDR tab. See Figure 25.
TDR tab
Enable outputs
Preset
Step polarity
Figure 25: Setting the TDR parameters
5. Set the Preset of Channel 8. The sampling module turns on a red light next to the SELECT channel button, indicating that TDR is activated for that channel.
TDR Preset sets Internal Clock in the Trigger menu, turns on the TDR Step in the TDR Setups menu, turns on the channel and selects the acquisition units in the TDR Setups menu, and sets the horizontal scale, position, and reference.
40
P7360A and P7340A Z-ActiveDifferential Probes
Performance Verification
6. Turn off the display for Channels 8, then only Channel 4 is shown on the screen.
7. Adjust the oscilloscope horizontal and vertical position controls to display a signal similar to that shown in Figure 24.
8. Set the oscilloscope horizontal scale to 100 ps/div and center the waveform.
9. Use the oscilloscope measurement capability to display rise time. Increase
the stability of the pulse-edge measurement by using averaging, if available. Rise time is measured from the 10% and 90% amplitude points on the waveform. Rise time can be measured using the automatic measurement capability of the TDS8200 series oscilloscopes. Record the system rise time as t
This value is used to calculate both the 5X and 25X probe rise times.
s.
The following steps instruct you to assemble the test setup that includes the probe, as shown in Figure 26. The system and probe rise time (t measure in step 17 is used to calculate the probe rise time (t
TDS/CSA 8000 Series Oscilloscope
p
) that you
s+p
)instep18.
CH 4 (measurement
channel)
80E0X sampling
module
50 Ω Termination
Variable Spacing Tip-Clip
SMA male-to-male
connector between 80E04
and calibration fixture
underneath the probe
Figure 26: Test probe r ise time setup
80A03
CH 7 and 8 (module slot)
P7360A/P7340A TekConnect and probe
Calibration fixture
80E04 TDR pulser
Sampling module extender cable
P7360A and P7340A Z-ActiveDifferential Probes
41
Performance Verification
10. Remove the calibration fixture from the TCA--SMA adapter and disconnect the TCA--SMA adapter from the 80A03 TekConnect probe interface.
11. Connect the probe to the 80A03 TekConnect probe interface.
12. Check that an SMA 50
Ω termination included with the probe calibration
fixture is connected to the open SMA output on the fixture.
13. Set the attenuation on the probe to 25X.
14. Connect the probe input to the probe calibration fixture as shown in
Figure 23 on page 38. Check that the TDR function is still active.
The test setup should now be connected as shown in Figure 26.
15. Adjust the vertical scale to 50 mV/div, averaging on.
16. Expand the horizontal scale to help locate the step edge, then adjust the
horizontal range to 100 ps/div while centering the edge view. For a more stable measurement display, turn averaging on.
17. Use the oscilloscope measurement capability to display rise time. Rise time is measured from the 10% and 90% amplitude points on the waveform. Record the rise time as t
s+p.
18. Calculate the probe rise time using the following formula:
2
tp= t
(s+p)
t
2
s
19. Record the calculated probe rise time on the test record.
Rise Time Check at 5X
Attenuation
42
20. Set the attenuation on the probe to 5X.
21. Repeat steps 16 through 19 for the 5X attenuation setting.
P7360A and P7340A Z-ActiveDifferential Probes
Performance Verification

Test Record

Probe Model/Serial Number: Certificate Number: Temperature: RH %: Date of Calibration: Technician:
Performance test Minimum Results Maximum
Output offset voltage 25X
± 3mV(20_Cto30_C)
-- 3 m V +3mV
5X ± 3mV(20_Cto30_C)
DC attenuation accuracy 25X 0.0392 0.0408
5X 0.196 0.204
Rise time (P7360A) 25X N/A 70 ps
5X N/A 70 ps
(P7340A) 25X N/A 100 ps
5X N/A 100 ps
-- 3 m V +3mV
P7360A and P7340A Z-ActiveDifferential Probes
43
Performance Verification
44
P7360A and P7340A Z-ActiveDifferential Probes

User Service

This section covers troubleshooting and probe maintenance.

Probe/Adapter/Oscilloscope Compatibility

Thezs differential probes are designed to work with all TekConnect-interface oscilloscopes and adapters. However, there may be some cases where probe features may not work properly.
Table 7: Differential probe compatibility issues
Symptom Likely cause
Differential probe does not work with an 80A03 TekCon­nect Probe Interface Adapter
The LED on the 80A03 Adapt­er glows red, indicating an incompatible probe.
The 80A03 Adapter requires firmware version 2.0 and above. The firmware version label is on the rear panel of the instrument. Contact Tektronix for information on updating the adapter firmware.

Error Condition

Replacement Parts

The LEDs on the probe alert you to error or status conditions affecting the probe. If the probe LEDs flash or otherwise appear to be malfunctioning, an error condition may exist. Call your Tektronix representative for service.
When the probe is functioning correctly there is a quick flash of the LEDs on the probe just after connecting to the oscilloscope.
There are no user replaceable parts within the probe. Refer to your product user manual for a list of replaceable accessories for your probe.
If your probe does not meet the specifications tested in the Performance Verification, you can send the probe to Tektronix for repair. Follow the procedure on page 46 to prevent damage to the probe during shipping.
P7360A and P7340A Z-ActiveDifferential Probes
45
User Service

Preparation for Shipment

If the original packaging is unfit for use or not available, use the following packaging guidelines:
1. Use a corrugated cardboard shipping carton having inside dimensions at least
2. Put the probe into an antistatic bag or wrap to protect it from dampness.
3. Place the probe into the box and stabilize it with light packing material.
4. Seal the carton with shipping tape.
5. Refer to Contacting Tektronix at the front of the P7313, P7380, and P7360
one inch greater than the probe dimensions. The box should have a carton test strength of at least 200 pounds.
5X/25X Differential Probes User Manual for the shipping address.
46
P7360A and P7340A Z-ActiveDifferential Probes
Loading...