Tektronix P7313SMA Quick Start Guide

P7313SMA Differential Probe Quick Start User Manual
www.tektronix.com
071-1965-00
Copyright © Tektronix. All rights r eserved. Licensed software products are owned by Tektronix or its subsidiaries or supp liers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14200 SW Karl Braun Drive P.O . B ox 5 00 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200. Worldwide, visit www.tektronix.com to nd contacts in your area.
Warranty 2
Tektronix warrants that th is product will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and replacement products used by Tektronix for warranty work may be new or reconditioned to like new performance. All replaced parts, modules and products become the property of Tektronix.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in wh ich the Tektronix service c enter is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned t o any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modied or integrated with other products when the effect of such modication or integration increases the time or difculty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR C ONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
Table of Contents
General Safety Sum m ary .. . ............................................................................................................... iii
Environmental Considerations ............................................................................................................ vii
Preface..................................................................................................................................... viii
Documentation....................................................................................................................... viii
Conventions Used in this Manual ................................................................................................... viii
Returning the Probe for Servicing.................................................................................................... ix
Key Features ............................................................................................................................... 1
Operating Considerations .................................................................................................................. 2
Installation .................................................................................................................................. 3
Connecting to the Host Instrument ................................................................................................... 3
Connecting to the Probe .. . ........................................................................................................... 4
Functional Check ........................................................................................................................... 8
Required Equipment.................................................................................................................. 9
Power-On Self Test .................................................................................................................. 10
Signal and Termination Voltage Monitor Check..................................................................................... 11
Aux Output Check ................................................................................................................... 14
DC Termination Voltage Zero Check ................................................................................................ 16
Table of Contents
Probe Calibration .......................................................................................................................... 18
Check the Signal Path Compensation Status . ...................................................................................... 18
P7313SMA Differential Probe Quick Start User Manual i
Table of Contents
Basic Operation............................................................................................................................ 21
Applications................................................................................................................................ 29
Standard Accessories . . ................................................................................................................... 43
User Service ............................................................................................................................... 49
Index
Run the Probe Calibration ........................................................................................................... 19
Probe Controls and Connections.................................................................................................... 21
Optional Accessories ................................................................................................................ 47
Probe/Adapter/Oscilloscope Compa tibility . . ........................................................................................ 49
Error Conditions...................................................................................................................... 50
ii P7313SMA Differential Probe Quick Start User Manual
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to this product or any products con nected to it.
To avoid potential hazards, use this product only as specied.
Only qualied personnel should pe rform service procedures.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals for warnings and cautions related to operating the system.
To Avoid Fire or Personal Injury
Connect and Disconnect Properly. Do not connect or disconnect probes or test leads while they are connected to a voltage
source.
Connect and Disconnect Properly. Connect the probe output to the measurement instrument before connecting the probe to the
circuit under test. Connect the p robe reference lead to the circuit under test before connecting the probe input. Disconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe from the measurement instrument.
Ground the Product. This product is ind irectly grounded through the grounding conductor of the mainframe power cord. To avoid
electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Observe All Terminal Ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the
product manual for further ratings information before making connections to the product.
General Safety Summary
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Do Not Operate Without Covers. Do not operate this product with covers or panels removed.
P7313SMA Differential Probe Quick Start User Manual iii
General Safety Summary
Do Not Operate With Suspected Failures. If you suspect that there is damage to this product, have it inspected by qualied
service personnel.
Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
iv P7313SMA Differential Probe Quick Start User Manual
General Safety Summary
Terms in this Manual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Symbols and Terms on the Product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
P7313SMA Differential Probe Quick Start User Manual v
General Safety Summary
The following symbols may appear on the product:
vi P7313SMA Differential Probe Quick Start User Manual
Environmental C onsiderations
Product End-of-Life Handling
Observe the following guidelines when recycling an instrument or component:
Equipment Recycling. Production of th is equipment required the extraction and use of natural resources. The equipment may
contain substances that could be harmful to the environment or human health if improperly handled at the product’s end of life. In order to avoid release of such substances into the environment and to reduce the use of natural resources, we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately.
The symbol shown below indicates that this product complies with the European Union’s requirements according to Directive 2002/96/EC on w aste electrical and electronic equipment (W EEE). For information about recycling options, check the Support/Service section of the Tektronix Web site (www.tektronix.com).
Restriction of Hazardous Substances
Environmental Considerations
This product has been classied as Monitoring and Control equip ment, and is outside the scope of the 2002/95/EC RoHS Directive. This product is known to contain lead, cadmium, mercury, and hexavalent chromium.
P7313SMA Differential Probe Quick Start User Manual vii
Preface
Preface
This is the quick start user manual for the P7313SMA differential probe. This manual provides operating information, applications, and functional verication procedures for the probe.
Documentation
To read about Read these documents
First Time Operation, F unctional Che ck, Operating Basics
Specications, Performance Verication The technical reference manual.
In-depth oscilloscope operation, user interface, GPIB commands The online help (from the Help menu on the
*
To access the documentation that is installed on the host instrument, click Start in the taskbar and select Programs > TekApplications.
This manual.
host instrument).
*
Conventions Used in this Manual
The following icon is used throughout this manual to indicate a step sequence.
viii P7313SMA Differential Probe Quick Start User Manual
Returning the Probe for Servicing
If your probe requires servicing, you must return it to Tektronix. If the original packaging is unt for use or not available, use the following packaging guidelines:
Preparation for Shipment
1. Use a corrugated cardboard shipping
carton having inside dimensions at least one inch greater than the probe dimensions. The box should have a carton test strength of at least 200 pounds.
2. Put the probe into an antistatic bag or wrap it to protect it from dampness.
3. Place the probe into the box and stabilize it with light packing material.
4. Seal the carton with shipping tape.
5. Refer to Contacting Tektronix at the
beginning of this manual for the shipping address.
Preface
P7313SMA Differential Probe Quick Start User Manual ix
Preface
x P7313SMA Differential Probe Quick Start User Manual
Key Features
The P7313SMA is a 13 GHz, active differential probe designed for Serial Data Analysis (SDA) compliance t esting and other applications that use differential serial busses in a 50 signaling environment. The probe connects to oscilloscopes featuring the Tektronix TekConnect oscilloscope interface. Key features include:
SMA inputs with 1 meter low skew cable
pair
DC to 13 GHz bandwidth
Selectable 800 mV and 3.6 V p-p
Differential Inpu t Voltage ranges
Selectable Termination Voltage Control
modes with +3.6 V/-2.5 V range
Auxiliary Ou tput connector for triggering
TekConnect interface
Provides Automatic Units Scaling on the
oscilloscope display
Key Features
P7313SMA Differential Probe Quick Start User Manual 1
Operating Considerations
Operating Considerations
Table 1: P7313SMA
Characteristic Description
Input Voltage
Temperature
Humidity
Altitude
Pollution Degree 2, Indoor use only
Differential:
+3.6 V/-2.5 V (DC + peak AC)
Common Mode: ±2.5 V (DC + peak AC, both ra nges; input referenced to ground)
Operating: 0 to +40 °C (+32 °F to +104 °F) Nonoperating: -55 °C to +75 °C (-67 °F to +167 °F)
Operating: +30 °C to +40 °C (+86 °F to +104 °F) 0-90% RH Nonoperating: +30 °C to +60 °C (+86 °F to +140 °F) 0-90% RH
Operating: Up to 3000 meters (10,000 feet) Nonoperating: Up to 12192 meters (40,000 feet)
2 P7313SMA Differential Probe Quick Start User Manual
Installation
Connecting to the Host Instrument
The P7313SMA probe is powered through the TekConnect interface on the host instrument.
1. To install the probe, slide it into the TekConnect receptacle on the host instrument. The probe snaps into the receptacle when fully engaged.
The host instrument performs a diagnostic check to verify basic probe functionality and to cycle the status LEDs.
Before you use your probe, perform a functional check. (See page 11, Signal and Termination Voltage Monitor Check.)
2. To release the probe, grasp the probe, press the latch button, and pull the probe straight out.
Installation
P7313SMA Differential Probe Quick Start User Manual 3
Installation
Connecting to the Probe
This overview of the probe connectors helps you perform a functional check. Find more details about the probe connectors and controls in Basic Operation. (See page 21, Basic Operation.)
SMA Input Connectors
1. The P7313SMA probe has SMA connectors
for the (+) and (-) input signals.
The SMA connectors provide a signal path through the internal 50 termination network and differential probe buffer amplier to the oscilloscope.
2. Use the matched-delay SMA cables that are supplied with the probe to connect the probe to your circuit.
3. Leave the 50 terminations on the unused inputs.
4 P7313SMA Differential Probe Quick Start User Manual
DC Termination Voltage Input Jacks
1. Use these two 0.080 inch pin jacks to
provide DC termination voltage control from an external power supply.
2. Connect the 0.080 inch pin-to-banana
cables included with your probe to these jacks.
3. Press the SELECT button t o set the Vterm
Source to EXT mode.
Installation
P7313SMA Differential Probe Quick Start User Manual 5
Installation
Termination Voltage Monitor Jacks
1. Usethesetwo0.040inchjackstomonitor
the termination voltage of the signal under test.
2. Connect the 0.040-to-0.080 inch pin-jack adapters included with your probe to the jacks.
3. Connect a digital multimeter (DMM) to the jack adapters with a pair of standard DMM test leads.
6 P7313SMA Differential Probe Quick Start User Manual