Tektronix P7240 Service Manual

Service Manual
P7240 4 GHz Active Probe
071-1056-00
Warning
The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
www.tektronix.com
*P071105600*
071105600
Copyright © Tektronix, Inc. All rights reserved. Licensed software products are owned by Tektronix or its suppliers and are protected by United States copyright laws and international treaty provisions.
Use, duplication, or disclosure by the Government is subject to restrictions as set forth in subparagraph (c)(1)(ii) of the Rights in T echnical Data and Computer Software clause at DFARS 252.227-7013, or subparagraphs (c)(1) and (2) of the Commercial Computer Software – Restricted Rights clause at F AR 52.227-19, as applicable.
T ektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
T ektronix, Inc., P.O. Box 500, Beaverton, OR 97077 TEKTRONIX, TEK, T ekConnect, and KlipChip are registered trademarks of Tektronix, Inc.

General Safety Summary

Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this product unless another person capable of rendering first aid and resuscitation is present.
While using this product, you may need to access other parts of the system. Read the General Safety Summary in other system manuals for warnings and cautions related to operating the system.
To Avoid Fire or Personal Injury
Connect and Disconnect Properly . Do not connect or disconnect probes or test
leads while they are connected to a voltage source. Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Do Not Operate Without Covers. Do not operate this product with covers or panels removed.
Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions. Do Not Operate in an Explosive Atmosphere. Keep Product Surfaces Clean and Dry . Provide Proper Ventilation. Refer to the manual’s installation instructions for
details on installing the product so it has proper ventilation.
P7240 4 GHz Active Probe Service Manual
i
General Safety Summary
Symbols and Terms
T erms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
T erms on the Product. These terms may appear on the product: DANGER indicates an injury hazard immediately accessible as you read the
marking. WARNING indicates an injury hazard not immediately accessible as you read the
marking. CAUTION indicates a hazard to property including the product. Symbols on the Product. The following symbols may appear on the product:
CAUTION
Refer to Manual
WARNING
High Voltage
Double
Insulated
Protective Ground
(Earth) Terminal
Not suitable for
connection to
the public telecom-
munications network
ii
P7240 4 GHz Active Probe Service Manual

Preface

Manual Structure

Related Documentation

This is the service manual for the P7240 4 GHz Active Probe. Read this preface to learn how this manual is structured and where you can find other information related to servicing this product.
This manual contains three sections – Description, Performance Verification, and Adjustment Procedures, and is intended to be used by qualified service personnel. Replaceable parts are limited to accessories and adapters, and are described in the Instruction Manual that is shipped with the probe.
Be sure to read the introductions to all procedures. These introductions provide important information needed to do the service correctly, safely, and efficiently.
The probe is shipped with the following manual:
H
P7240 4 GHz Active Probe Instruction Manual. Tektronix part number 071-0759-XX.
P7240 4 GHz Active Probe Service Manual
iii
Preface

Contacting Tektronix

Phone 1-800-833-9200*
Address Tektronix, Inc.
Department or name (if known) 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
Web site www.tektronix.com
Sales support 1-800-833-9200, select option 1*
Service support 1-800-833-9200, select option 2*
Technical support Email: support@tektronix.com
1-800-833-9200, select option 3* 6:00 a.m. – 5:00 p.m. Pacific time
* This phone number is toll free in North America. After office hours, please leave a
voice mail message. Outside North America, contact a Tektronix sales office or distributor; see the Tektronix web site for a list of offices.
iv
P7240 4 GHz Active Probe Service Manual

Description

This document describes the Performance Verification and Adjustment Proce- dures for the P7240 4 GHz Probe. There are no user-replaceable parts inside the
probe. If the probe fails the performance verification procedure, and adjusting the probe cannot compensate for the failure, contact your Tektronix service center.

Performance Verification

Use the performance verification procedures to verify the warranted specifica­tions of the P7240 probe. The recommended calibration interval is one year. No software is required to complete a performance verification of the probe.
The performance verification procedures check the following specifications:
H
H
H
Output zero DC attenuation accuracy Rise time

Adjustments

The P7240 probe has two electronically-adjusted controls that rarely need adjusting unless a performance verification yields a specification that is out of tolerance, or the probe fails an oscilloscope probe calibration utility.
If you must make adjustments to the probe, you need the TekConnect Adjust­ment Software for the P7240 4 GHz Probe. Refer to Equipment Setup in the Adjustment Procedures, step 7, for instructions on finding and loading the adjustment software on your instrument.
The adjustment procedures describe the following adjustments for the probe:
H
Offset zero
H
Offset gain
P7240 4 GHz Active Probe Service Manual
1
Description

Equipment Required

Refer to Table 1 for the equipment required to perform the service procedures.
T able 1: Equipment required for performance verification and adjustment procedures
Item description Performance requirement Recommended example
T ekConnect adjustment software
Sampling Oscilloscope 12.5 GHz bandwidth T ektronix TDS8000 Sampling head, with extension cable 12.5 GHz bandwidth Tektronix 80E0X with 012-1568-00 cable Oscilloscope T ekConnect Interface T ektronix TDS7404 T ekConnect calibration adapter T ekConnect Interface 067-0422-00 Calibration Step Generator 250 mV step, 30ps rise time 067-1338-0X Adapter TekConnect-to-SMA TCA-SMA Adapter Probe tip with 50Wterminator 015-0678-00 DC Power Supply 1 VDC at 1 mA T ektronix PS280 DMM (2) with leads 0.05% accuracy, 0.1mV resolution Fluke 87 or equivalent Feedthrough T ermination 50W ±0.05 BNC-to-BNC coaxial cable (2) 50W coaxial cable 012-0057-01 Coaxial cable Male-to-Male SMA 012-0649-00 Adapter SMA Male-to-Male 015-101 1-00 Adapter SMA Female-to-Female 015-1012-00 Adapter SMA Male-to-BNC Female 015-0554-00 Adapter (2) BNC Female-to-Dual Banana 103-0090-00 Adapter Y -lead adapter 196-3456-00 Adapters (2) KlipChip adapter 206-0364-00 SMA torque wrench 5/16-in, 7 in-lb. SMA adapter wrench 7/32-in
1
Required for adjustment procedures only
1
P7240 Probe Optional Applications for Windows-Based
Oscilloscopes CD
W
011-0129-00

Special Adapters Required

Some of the adapters listed in Table 1 are custom-made and available only from Tektronix. The adapters are described on the following pages.
2
P7240 4 GHz Active Probe Service Manual
Description
TekConnect-to-SMA
Adapter

Probe Tip Adapter

The TekConnect-to-SMA Adapter, Tektronix part number TCA-SMA, allows signals from an SMA cable or probe to be connected to a TekConnect input. See Figure 1. Connect and disconnect the adapter the same way as you do the P7240 Probe.
Figure 1: TekConnect-to-SMA adapter
The probe tip adapter, Tektronix part number 015-0678-00, provides a low-noise method for connecting the P7240 to signals present on SMA cables. The adapter has a test point for the probe and two SMA connectors. A 50 ohm termination is included with the adapter, and is connected to the SMA connector nearest the probe test point to minimize relections. Connect the cable from the test circuit to the other SMA connector. See Figure 2.
NOTE. When taking measurements, do not touch the probe tip adapter. Measure­ment accuracy is degraded when the probe tip adapter is handled.
SMA connector for test circuit
Probe test point
Terminator
Figure 2: Probe tip adapter
P7240 4 GHz Active Probe Service Manual
3
Description
TekConnect Interface
Calibration Adapter
The TekConnect Interface Calibration Adapter, Tektronix part number 067-0422-00, connects between the host instrument and the probe under test, and provides connectors for probe signal and offset voltage measurements. See Figure 3.
Figure 3: TekConnect Interface Calibration Adapter
When the adapter is connected to the oscilloscope, the adapter is identified as a valid calibration device. However, additional power supplies necessary to power the probe are not enabled until a TekConnect probe is connected to the adapter and identified by the oscilloscope. When a probe is detected through the adapter, the Volts/div readout on the oscilloscope displays ##.
Refer to Table 2 on page 5 for features of the calibration adapter.
4
P7240 4 GHz Active Probe Service Manual
T able 2: TekConnect Interface Calibration Adapter features
Feature Description
Description
Offset switch
GND
Latch button
Latch
VAR
Offset voltage output
Signal out
Latch button. The spring-loaded latch mechanically retains the adapter to the oscilloscope. To
release the adapter, grasp the adapter housing, depress the latch button, and pull the adapter straight out of the oscilloscope.
Offset output select switch. The offset output switch selects between ground and the offset
voltage level from the oscilloscope. Leave the switch in the ground position for the performance verification procedures. Move the switch to the variable position for the adjustment procedures.
Offset voltage. The offset voltage of the probe is accessed through the BNC connector.
Measure the offset voltage using a DVM, BNC coaxial cable, and BNC-to-dual-banana jack.
Signal out. The SMA connector on the rear of the box allows for direct monitoring of the probe
signal.

Theory of Operation

Refer to Figure 4 as you read the following information. There are two adjustments in the P7240 adjustment procedure: Offset Zero and
Offset Gain. These adjustments are made using digitally controlled potentiome­ters that are controlled by a two-wire serial interface from the host oscilloscope. Because the adjustments are digitally controlled, the probe does not need to be disassembled as part of the adjustment procedure. The software application provides the control needed to make the adjustments to the probe.
P7240 4 GHz Active Probe Service Manual
5
Description
Probe tip
Probe head Compensation box Oscilloscope
Probe tip amplifier
EEPROM
Signal Out Probe ID Out
Clock In
Offset to probe
+7 V
Offset ampifier
Probe cable
+V Offset
Linear
regulator
Offset zero
–V
gain
T ekConnect
Interface
±1 V Offset
+15 V
–5 V Ground
Figure 4: P7240 simplified schematic diagram
The offset signal from the oscilloscope extends the dynamic range of the P7240 probe by compensating for some of the DC voltage that may be part of the signal applied to the probe. Compensation is accomplished by summing the offset signal with the probe input signal at the probe input attenuator, as shown in Figure 4. The host oscilloscope supplies the offset signal (with a range of ±1 volt) to the offset amplifier in the probe. Since the offset scale factor of the P7240 probe is 5X, the effective offset signal range is increased to ±5 V.
The Offset Zero adjustment is used to null out any amplifier voltage offsets in the offset signal path. With no input signal at the probe input and no offset signal from the host oscilloscope, the Offset Zero is adjusted to produce 0.0000 V at the probe output. Since the P7240 probe is designed to operate into a host oscillo­scope with a 50 load, a precision 50 feed thru terminator is used in the adjustment procedure.
The Offset Gain adjustment is used to adjust the offset amplifier gain for a precise 1X scale factor. Since the P7240 input signal scale factor has been designed with a 5X scale factor, the Offset Gain is adjusted to null out the difference between a DC input signal and a 5X-scaled offset signal from the host oscilloscope.
6
P7240 4 GHz Active Probe Service Manual

Performance Verification

Use the following procedures to verify the warranted specifications of the P7240 probe. Before beginning these procedures, refer to page 14 and photocopy the test record and use it to record the performance test results. The recommended calibration interval is one year. These procedures test the following specifica­tions:
H
Output zero
H
DC attenuation accuracy
H
Rise time

Equipment Setup

Use this procedure to set up the equipment to test the probe.
1. Connect the probe calibration adapter to the oscilloscope.
2. Connect the probe to the probe calibration adapter.
3. Turn on the oscilloscope, and enable the channel.
4. Verify that the Volts/div readout on the oscilloscope channel displays ##
(the oscilloscope recognizes the probe through the adapter).
5. Set the multimeters to read DC volts.
6. Allow 30 minutes for the equipment to warm up.
P7240 4 GHz Active Probe Service Manual
7
Peformance Verification

Output Zero

Use this procedure to verify the probe output zero.
1. Connect the test equipment as shown in Figure 5.
2. Set the offset switch on the calibration adapter to GND.
NOTE. Leave the offset switch in the ground position for all of the performance verification checks.
3. Ground the probe tip by connecting the probe tip to the probe ground socket.
(Connecting two KlipChip adapters together is recommended.)
4. Observe the multimeter display, and record the results on the test record. The displayed DC level should be 0.000 V ±10 mV.
TDS7404 Oscilloscope
terminator
BNC-SMA adapter
multimeter
BNC-to-Dual
Banana Adapter
BNC cable
Digital
50W Precision
Figure 5: Setup for output zero
Set offset switch to GND
KlipChip adapters
TekConnect Calibration adapter
Y-lead adapter
P7240 Probe
8
P7240 4 GHz Active Probe Service Manual

DC Attenuation Accuracy

1. Disconnect the KlipChip adapters, and connect the setup as shown in
Peformance Verification
Figure 6.
BNC-to-Dual
Banana Adapter
50W Precision
terminator
BNC-SMA
adapter
P7240
Probe
Digital
multimeter
Black (–)
TDS7404 Oscilloscope
Set offset switch to GND
BNC cable
TekConnect
calibration
adapter
Digital multimeter
Power supply
–+
KlipChip adapters
Red (+)
Figure 6: Setup for DC attenuation accuracy
Record the results of steps 2 through 5 separately. Use the results to calculate the DC accuracy of the probe in step 6.
2. Set the power supply to +1.000 V. Use the multimeter connected to the
power supply to verify that the DC voltage is as close to +1.000 V as possible. Record this measurement as Vmax.
3. Measure and record the reading of the other multimeter as M1.
4. Set the power supply to –1.000 V. Use the multimeter connected to the
power supply to verify that the DC voltage is as close to –1.000 V as possible. Record this measurement as Vmin.
5. Measure and record the reading of the other multimeter as M2.
P7240 4 GHz Active Probe Service Manual
Y-lead adapter
9
Peformance Verification
6. Determine the percent error by using the formula below:
%Error
+
M1–M2
(Vmax–Vmin) .2
–1 100%
Example:
%Error
0.2028–(–02021)
+
[1.007–(–0.998)] .2
0.4049
+
+
+)
–1 100%
0.4010
1.0097–1 100%
0.97%
–1 100%
The calculated error should be ≤2%.
7. Record the results on the test record.
NOTE. An unacceptable error value may result if a low tolerance termination is substituted for the recommended termination.

Rise Time

10
This procedure verifies that the probe meets rise time specifications. The probe rise time is calculated using rise times measured from the test system
separately, and the test system including the probe.
1. Connect the test equipment as shown in Figure 7 on page 11.
CAUTION. To prevent damage, use care when working with SMA connectors: support equipment to avoid mechanical strain on the connectors, and use a torque wrench when tightening connections.
P7240 4 GHz Active Probe Service Manual
TDS7404 Oscilloscope
Peformance Verification
Set offset switch to GND
Sampling head extender cable
80E0X
SMA
M-to-M
adapter
CSA8000/TDS8000
TekConnect Calibration adapter
TekConnect-to­SMA Adapter
Generator remote head
Calibration
step generator
Trigger input
SMA cable
Internal clock output
Figure 7: Test system rise time setup
2. Set the oscilloscope trigger to internal clock.
3. Select the channel you have connected to on the 80E0X sampling head, and
then set the oscilloscope vertical scale to 50 mV/div.
NOTE. The output of the step generator rises from a –250 mV level to ground.
4. Adjust the oscilloscope horizontal and vertical position controls to display a
signal similar to that shown in Figure 7.
5. Set the oscilloscope horizontal scale to 50 ps/div and center the waveform.
6. Use the oscilloscope measurement capability to display rise time. Increase
the stability of the pulse edge measurement by using averaging, if available. Rise time is determined from the 10% and 90% amplitude points on the waveform. Record the rise time as t
The system rise time (ts) that you measured in step 6 represents the rise time of the test system without the probe.
P7240 4 GHz Active Probe Service Manual
s.
11
Peformance Verification
After you assemble the test setup that includes the probe, as shown in Figure 8, you will measure the system and probe rise time (t calculate the probe rise time (t
) in step 14.
p
) in step 13. This is used to
s+p
7. Set the step generator control switch to standby.
8. Remove the TekConnect-SMA adapter from the test setup.
9. Connect the test setup as shown in Figure 8 on page 12.
CAUTION. To prevent damaging the SMA connectors, use a 7/32-inch wrench when connecting and disconnecting the female-to-female SMA adapter.
TDS7404 Oscilloscope
SMA
Sampling head extender cable
M-to-M
adapter
80E0X
CSA8000/TDS8000
TekConnect calibration adapter
P7240 Probe
Trigger input
Internal clock output
50 W Terminator
Probe tip adapter
SMA F-to-F adapter
Generator remote head
Calibration
step generator
SMA Cable
Figure 8: Test system and probe rise time setup
10. Set the step generator control switch to on.
11. On the TDS8000, expand the horizontal scale to locate the step edge. Set the
vertical scale to 10 mV/div, and adjust horizontal range to 100 ps/div while maintaining the edge view. For a more stable measurement display, turn averaging on.
12
P7240 4 GHz Active Probe Service Manual
Peformance Verification
12. Adjust the TDS8000 horizontal and vertical position controls to display a
signal similar to that shown in Figure 8.
NOTE. Do not touch the probe tip adapter when making calibration measure­ments. Measurement accuracy is degraded when the probe tip adapter is handled.
13. Use the TDS8000 measurement capability to display rise time. Rise time is
determined from the 10% and 90% amplitude points on the waveform. Record the rise time as t
s+p
.
14. Calculate the probe-only rise time using the following formula:
2
tp+
Ǹ
t
(s)p)
*
2
t
s
15. Check that the calculated rise time meets the probe specification. The rise
time (t
) must be v120 ps.
p
16. Record the results on the test record.
P7240 4 GHz Active Probe Service Manual
13
Peformance Verification
T est record
Probe Model: Serial Number: Certificate Number: T emperature: RH %: Date of Calibration: Technician:
Performance test Minimum Incoming Outgoing Maximum
Output zero voltage (at probe output)
± 10 mV (20_ C to 30_ C) DC attenuation accuracy (% error) N/A ________ ________ 2 % Rise time N/A ________ ________ 120 ps
– 10 mV ________ ________ + 10 mV
14
P7240 4 GHz Active Probe Service Manual

Adjustment Procedures

These instructions are for personnel who are familiar with servicing the product. If you need further details for disassembling or reassembling the product, refer to the appropriate product manual. Contact your nearest Tektronix, Inc., Service Center or Tektronix Factory Service for installation assistance.
CAUTION. To prevent static discharge damage, service the product only in a static-free environment. Observe standard handling precautions for static-sensi­tive devices when servicing the probe. Always wear a grounded wrist strap, grounded foot strap, and static resistant apparel when servicing the probe.

Equipment Setup

Use this procedure to set up the equipment to adjust the probe.
1. Turn on the oscilloscope and enable the channel you intend to use.
2. Connect the probe calibration adapter to the oscilloscope.
3. Connect the probe to the probe calibration adapter.
4. Verify that the Volts/div readout on the oscilloscope channel displays ##
(the oscilloscope recognizes the probe through the adapter).
5. Set the multimeters to read DC volts.
6. Allow 30 minutes for the equipment to warm up.
7. While the equipment is warming, verify that the application has been loaded
onto the oscilloscope:
a. Minimize the scope application window by going to FileMinimize. b. Go to: StartProgramsTektronixTekConnect Adjust c. If the application is not loaded on the oscilloscope, get the application:
Retreive the application CD that came with the oscilloscope, insert the CD in the drive, select StartRun, and then enter D:\setup.exe
Or, download the application from the Tektronix website at: www.tektronix.com/Software and Drivers/Accessories
8. When the application starts, the display appears as shown in Figure 9.
P7240 4 GHz Active Probe Service Manual
15
Adjustment Procedures

Application Setup

Figure 9: TekConnect probe adjustment application
1. Make the following selections in the application dialog box: a. Select P7240 in the Probe Type to Adjust list. b. Select the oscilloscope channel that the adapter is connected to in the
Channel to Use list.
c. Select Adjustment in the Task to Perform list.
2. Click Continue.
3. The P7240 Probe Adjustments Dialog Box appears. See Figure 11 on
page 18.
NOTE. An error message appears if one of the following conditions exists:
H
A probe is not connected to the adapter.
H
A probe model other than a P7240 is connected to the adapter.
16
H
The P7240 connected to the adapter is defective (unable to properly communicate with the adapter).
P7240 4 GHz Active Probe Service Manual

Offset Zero

Adjustment Procedures
If an error message appears, click OK, correct the condition that caused the error message, and then click Continue.
4. Proceed to the Offset Zero procedure.
NOTE. To avoid corrupted test results, do not disconnect the probe under test or the probe calibration adapter until you have completed the service procedures. If the probe or adapter are disconnected before completing the procedures, you must reconnect the probe and adapter, and exit and restart the application.
Use this procedure to adjust the probe offset zero.
1. Connect the test equipment as shown in Figure 10.
TDS7404 Oscilloscope
terminator
BNC-SMA adapter
Set offset switch to GND
Y-lead
adapter TekConnect Calibration adapter
P7240 Probe
KlipChip adapters connected together
BNC-to-Dual
Banana Adapter
50WBNC cable
Digital
multimeter
50W Precision
Figure 10: Setup for offset zero
2. Set the offset switch on the calibration adapter to GND.
3. Click Offset Zero in the application window. See Figure 11 on page 18.
P7240 4 GHz Active Probe Service Manual
17
Adjustment Procedures
Adjust slider to set offset voltage on DMM to 0.0000 V
Index value covers a ±20 mV range
Click Store Setting after voltage is set
Channel Offset is disabled in this step
Figure 11: Offset Zero window
4. Set the offset zero voltage on the DMM to 0.0000 V by adjusting the Offset
Zero Control. The index value in the Offset Zero Control varies from 0 to 3969, and corresponds to approximately ±20 mV of offset voltage range.
Use one of the following methods to make the adjustment:
H
Use your finger or the mouse to drag the on-screen slider up or down.
H
Click or touch the up/down arrow keys above and below the slider. Each click adjusts the offset voltage by about 10 mV.
H
To increment by ten clicks or about 100 mV, click in the bar above or below the slider.
18
P7240 4 GHz Active Probe Service Manual

Offset Gain

Adjustment Procedures
5. When the offset zero voltage displayed on the voltmeter is 0.0000 V
±0.0001 V, click Store Setting to save the offset zero voltage level.
NOTE. In the Summary Info box, the After field updates with the control value from step 5. This is a temporary value until Accept is selected later, after all adjustments have been made.
If you select Accept now, the application requests confirmation: If you click Yes, the application exits; if you click No, you are able to make further adjustments. You must complete all adjustments before selecting Accept.
6. Proceed to the Offset Gain procedure.
Use this procedure to adjust the probe offset gain.
1. Set the DC power supply to –2.5 volts. Use a DMM to measure the voltage.
Record this voltage as M1.
Digital
multimeter
#1
BNC-to-Dual
Banana Adapter
BNC cable
50W Precision
terminator
BNC-SMA adapter
Figure 12: Setup for offset gain
2. Connect the test equipment as shown in Figure 12.
TDS7404 Oscilloscope
Set offset switch to VAR
TekConnect Calibration adapter
P7240 Probe
Digital
multimeter
BNC cable
#2
BNC-to-Dual
Banana Adapter
DC Power Supply
KlipChip adapters
3. Set the offset switch on the calibration adapter to VAR.
+
Y-lead adapter
4. Click OFFSET GAIN in the application window. See Figure 13 on page 20.
P7240 4 GHz Active Probe Service Manual
19
Adjustment Procedures
Adjust slider to set offset gain on DMM #1 to value calcu­lated in step 6.
Click Store Setting after gain and offset are set.
Adjust slider to set DMM #2 to 0.5 V.
Figure 13: Offset Gain window
NOTE. Channel offset is multiplied by the 5X attenuator in the probe. The offset voltage (+ 0.5 V) measured at the BNC connector on the Calibration Adapter nulls out the –2.5 V applied to the probe input.
5. Set the Channel Offset control to display 0.5 V on DMM #2.
Use one of the following methods:
H
Use your finger or the mouse to drag the on-screen slider up or down.
H
Click or touch the up/down arrow keys above and below the slider. Each click adjusts the offset voltage by about 1 mV.
H
To increment by ten clicks or 10 mV, click in the bar above or below the slider.
20
Record this voltage as M2.
P7240 4 GHz Active Probe Service Manual
Adjustment Procedures
6. Determine the offset gain target value by using the formula below:
Offset gain target value
+
M1
5
)
M2
7. The Offset Gain voltage is displayed on DMM #1. Adjust the Offset Gain
Control to the offset gain target value, ±1 mV, using one of the following methods:
H
Use your finger or the mouse to drag the on-screen slider up or down.
H
Click or touch the up/down arrow keys above and below the slider. Each click adjusts the offset voltage by about 1 mV.
H
To increment by ten clicks or 10 mV, click in the bar above or below the slider.
8. When the offset gain displayed on DMM #1 is within ±1 mV of the offset
gain target value, click Store Setting to save the offset gain level.
In the Summary Info box, the AFTER field updates with the control value from step 8. This is a temporary value until Accept is selected.
9. Click Accept to save all the values that you set in these procedures.
10. Click Yes to update the contents.

Verify Operation

This completes the adjustments to the probe.
Do the Performance Verification procedures to verify that the probe meets specifications.
g
End of document
g
P7240 4 GHz Active Probe Service Manual
21
Adjustment Procedures
22
P7240 4 GHz Active Probe Service Manual
Loading...