Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverto
USA
For product information, sales, service, and technical support:
n, OR 97077
In North America, call 1-800-833-9200.
Worl dwid e, visi t www.tektronix.com to find contacts in your area.
Warranty
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1)
year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its
option, either will repair the defective product without charge for parts and labor, or will provide a replacement
in exchange for the defective product. Parts, modules and replacement products used by Tektronix for warranty
work may be n
the property of Tektronix.
ew or reconditioned to like new performance. All replaced parts, modules and products become
In order to o
the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible
for packaging and shipping the defective product to the service center designated by Tektronix, with shipping
charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within
the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping
charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate
maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage
result
b) to repair damage resulting from improper use or connection t o incompatible equipment; c) to repair any damage
or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modified or
integrated with other products when the effect of such modification or integration increases the time or difficulty
of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY
OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
TRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE
TEK
AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY.
TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL,
OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS
ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
[W2 – 15AUG04]
btain service under this warranty, Customer must notify Tektronix of the defect before the expiration of
ing from attempts by personnel other than Tektronix representatives to install, repair or service the product;
Table of Contents
General Safety Summary .........................................................................................iii
Service Safety Summary.............. .................................. ................................ ...........v
Compliance information ..........................................................................................vi
Table 9: Service options ............ ................................ ................................ ..............27
iiP6910 General Purpose Probe Instruction Manual
General Safety Summary
General Safet
To Avoid Fire or Personal
Injury
ySummary
Review the fo
this product or any products connected to it.
To avoid pot
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of a larger system.
Read the safety sections of the other component manuals for warnings and
cautions r
Ground the product. This product is indirectly grounded through the grounding
conductor of the mainframe power cord. To avoid electric shock, the grounding
conductor must be connected to earth ground. Before making connections to
the input or output terminals of the product, ensure that the product is properly
ground
Observe all terminal ratings. To avoid fire or shock hazard, observe all ratings
and ma
information before making connections to the product.
The i
ed.
nputs are not rated for connection to mains or Category II, III, or IV circuits.
llowing safety precautions to avoid injury and prevent damage to
ential hazards, use this product only as specified.
elated to operating the system.
rkings on the product. Consult the product manual for further ratings
Connect the probe reference lead to earth ground only.
Do not apply a potential to any terminal, including the common terminal, that
exceeds the maximum rating of that terminal.
Do not operate without covers. Do not operate this product with covers or panels
removed.
Do not operate with suspected failures. If you suspect that there is damage to this
product, have it inspected by qualified service personnel.
Avoid exposed circuitry. Do not touch exposed connections and components when
power is present.
P6910 General Purpose Probe Instruction Manualiii
General Safety Summary
TermsinThisManual
Symbols and Terms on the
Product
Do not operate i
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
These terms
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
These terms may appear on the product:
DANGER
the marking.
WARNI
read the marking.
CAUT
n wet/damp conditions.
may appear in this manual:
indicates an injury hazard immediately accessible as you read
NG indicates an injury hazard not immediately accessible as you
ION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
ivP6910 General Purpose Probe Instruction Manual
Service Safety Summary
Service Safet
y Summary
Only qualifieSafety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnect Power. To avoid electric shock, switch off the instrument power, then
disconnect the power cord from the mains power.
UseCareWhenServicingWithPowerOn. Dangerousvoltagesorcurrentsmay
exist in
disconnect test leads before removing protective p anels, soldering, or replacing
components.
To avoid electric shock, do not touch exposed connections.
d personnel should perform service procedures. Read this Service
this product. Disconnect power, remove battery (if applicable), and
P6910 General Purpose Probe Instruction Manualv
Compliance information
Compliance in
Safety compliance
Equipment type
Safety class
Pollution
degree
description
formation
This section
instrument complies.
Test and measuring equipment.
Class 1 – grounded product.
A measure of the contaminants that could occur in the environment around
and within a product. Typically the internal environment inside a product is
considered to be the same as the external. Products should be used only in the
environ
Pollution Degree 1. No pollution or only dry, nonconductive pollution occurs.
Produc
located in clean rooms.
lists the safety and environmental standards with which the
ment for which they are rated.
ts in this category are generally encapsulated, hermetically sealed, or
ollution degree
P
tion Degree 2. Normally only dry, nonconductive pollution occurs.
Pollu
Occasionally a temporary conductivity that is caused by condensation must
be expected. This location is a typical office/home environment. Temporary
condensation occurs only when the product is out of service.
Pollution Degree 3. Conductive pollution, or dry, nonconductive pollution
that becomes conductive due to condensation. These are sheltered locations
where neither temperature nor humidity is controlled. The area is protected
from direct sunshine, rain, or direct wind.
Pollution Degree 4. Pollution that generates persistent conductivity through
conductive dust, rain, o r snow. Typical outdoor locations.
Pollution Degree 2 (as defined in IEC 61010-1). Note: Rated for indoor use only.
viP6910 General Purpose Probe Instruction Manual
Environmental considerations
This section provides information about the environmental impact of the product.
Compliance information
Product End-of-Life
handling
Restriction of hazardous
substances
Observe the f
Equipment recycling. Production of this equipment required the extraction and
use of natural resources. The equipment may contain substances that could be
harmful to the environment or human health if improperly handled at the product’s
end of life. To avoid release of such substances into the environment and to
reduce the
an appropriate system that will ensure that most of the materials are reused or
recycled appropriately.
This pr
scope of the 2002/95/EC RoHS Directive.
ollowing guidelines when recycling an instrument or component:
use of natural resources, we encourage you to recycle this product in
This symbol indicates that this product complies with the applicable European
Union re
on waste electrical and electronic equipment (WEEE) and batteries. For
information about recycling options, check the Support/Service section of the
Tekt r on
oduct is classified as Monitoring and Control equipment, and is outside the
quirements according to Directives 2002/96/EC and 2006/66/EC
ixWebsite(www.tektronix.com).
P6910 General Purpose Probe Instruction Manualvii
Compliance information
viiiP6910 General Purpose Probe Instruction Manual
Preface
Related documentation
Preface
The followi
ng list and table provide information on the related documentation
available for your Tektronix product. For additional information, refer to the
Tektronix Web site (www.tektronix.com/manuals).
Related documentation
ItemPurpose
TLA Quick Start User Manuals
Online Help
Installation Reference SheetsHigh-level installation information
Installation Manuals
XYZs of Logic Analyzers
Declassification and Securities instructionsData security concerns specific to sanitizing
Optional Service ManualsSelf-service documentation for modules and
High-level operational overview
In-depth operation and UI help
Detailed first-time installation information
Logic analyzer basics
or removing memory devices from Tektronix
products
Collection of logic analyzer application
specific notes
TLA Product specifications and performance
verification procedures
Detailed information for controlling the logic
analyzer using .NET
Upgrade information for your logic analyzer
mainframes
P6910 General Purpose Probe Instruction Manualix
Preface
xP6910 General Purpose Probe Instruction Manual
Operating basics
Product description
This section provides a brief description o f the Tektronix P6910 General Purpose
Logic Analyzer Probe, probe accessories, probe labels, and probe accessory
connection i
nstructions.
The P6910 p
from a Tektronix logic analyzer to the SUT (system-under-test).
The follo
34 individual active channel connections
Differential and single-ended data, clock and qualification inputs
Lead set support for both single-ended and differential applications
Holder for 8-channel applications
Color-coded signal connectors
–2.5 V
Minimal loading of <1 pF and 20 kΩ to ground
Operation in normal or inverted polarity is acceptable
Any common mode voltage is acceptable as long as the maximum positive
voltage does not exceed +5 V and the maximum negative voltage does not
exceed –2.5 V
robe is a 34-channel, general purpose probe that provides connections
wing list details the capabilities and qualities of the P6910 probe:
to +5 V input operating range
P6910 probe accessory information
The P6910 probe includes accessories to connect the logic analyzer to the SUT.
The following accessories are available for the P6910 probe:
Option 1K provides accessories for probing applications for less than
100 MHz. Tektronix recommends ordering two of each of the accessories
for a 34-channel probe. Details on the accessories are provided later in this
document. (See page 9, Option 1K accessories.)
Option 2K provides a probe accessory kit (Tektronix part number,
020-2973-01 or higher) for high-performance probing applications; provides
different means to connect the logic analyzer to the SUT. Tektronix
recommends ordering two accessory kits for a 34-channel probe. Details on
the accessory kit contents are provided later in this document. (See page 10,
Option 2K accessories.)
P6910 General Purpose Probe Instruction Manual1
Operating basics
Probe label overview
Probe labels, o
P6910 General Purpose Logic Analyzer Probe Instruction Manual (Tektronix
part number, 0
downloadable from the Tektronix Web site: www.tektronix.com/manuals)
The logic analyzer probe comes with a shee t of labels to apply to the probe before
connecting the probe to the instrument and SUT.
Tektronix provides a sheet of color-coded labels to apply to the probe to help
identify the connections to the logic analyzer and to the SUT. The label color is
designed to match the color of the probe connections on the logic analyzer. A set
of custom labels is also available for custom applications.
NOTE. If you are labeling probes for either 34- or 68-channel logic analyzers, use
a combi
this probe are grouped within a dashed red line. (See Figure 1 on page 3.)
nation of two sets of probe labels for one of your probes. The labels for
ne sheet
77-0630-xx, available on the TLA Documentation CD or
2P6910 General Purpose Probe Instruction Manual
Operating basics
Figure 1: Example of a sheet of P6910 probe labels
P6910 General Purpose Probe Instruction Manual3
Operating basics
Apply the labels to the p robe
Attach the labels to logic analyzer-end of the probe and to each of the four way
stations.
Attach the color-coded labels to the probe to help you identify the logic analyzer
and probe connections when connecting the probe to the SUT.
1. Determine the channel groups that you plan to use on your logic analyzer and
identify the matching labels.
Refer to the logic analyzer to identify the number of probe connections
(dependent on the number of channels on the instrument) and the channel
color-coding that corresponds to the provided sheet of probe labels.
2. Identify the logic analyzer end of the probe. (See Figure 2 on page 5.)
3. Using a pair of flat-nosed tweezers, place the tip of the tweezers under the
upper right corner of the logic analyzer-end label.
This will be the largest label of the five labels. Grasp the corner of the label
and lift the label up and toward you.
4. Carefully align the label to the label indent located on the top of the logic
analyzer-end and apply the label.
5. Identify the four way stations. (See Figure 2 on page 5.)
Note that the logic analyzer end label contains two colors. From the logic
analyzer end follow one of the two colors down the length of the cable to
the way station.
6. Using a pair of flat-nosed tweezers, place the tip of the tweezers under the
upper right corner of the matching color-coded way station.
Grasp the corner of the label and lift the label up a nd toward you.
7. Carefully align the label to the label indent located on the top of the way
station and apply the label.
8. Repeat step 7 for the second way station of the same color (but different label
shape).
9. Repeat the above steps for the other logic analyzer-end label colors .
After applying the labels to the probes, you are ready to connect the probes to
the logic analyzer.
4P6910 General Purpose Probe Instruction Manual
Operating basics
Figure 2: Attaching the probe labels
P6910 General Purpose Probe Instruction Manual5
Operating basics
Connect the pr
obe to the instrument
The P6910 probe connects a TLA6200 Series logic analyzer or TLA7000 Series
logic analyzer with TLA7Axx or TLA7Bxx logic analyzer modules to the SUT.
Apply the labels to the probes before connecting the probes to the instrument
and to the SUT.
Connect the probes to the logic analyzer using the following steps:
1. Match the color-coded labels of the probe to the same color-coded connector
on the logic analyzer. (See Figure 3 on page 7.)
2. Identify the beveled edges of the connector inside the logic analyzer end
of the probe.
3. Align the beveled edges of the connector to its mating connector on the logic
analyzer and press into place.
4. Evenly tighten both screws on the logic analyzer end of the probe until they
are snug.
Tighten each screw to 4 in-lbs (max).
CAUTION. When attaching the probe to the logic analyzer, use care to evenly
tighten probe screws until they are snug. Under-tightening the probe screws can
lt in intermittent operation. Over-tightening can result in stripped screws.
resu
NOTE. The P6910 probe can be connected to the logic analyzer when it is
ered on. The probe head can also be connected to the SUT without turning
pow
the power off.
6P6910 General Purpose Probe Instruction Manual
Operating basics
Figure 3: Connect the probe to the logic analyzer
Connect the probe directly to the SUT
Connect the probe directly to square pin connectors on the SUT.
CAUTION. The probe can be damaged by incorrectly connecting the probe to
the SUT. Incorrect handling of the probe while connecting it to the SUT can
damage the probe or the mating connector on the SUT. Always position the probe
endicularly to the mating connector and then connect the probe.
perp
P6910 General Purpose Probe Instruction Manual7
Operating basics
1. Connect the pro
The probe is compatible with differential and single-ended signals. The (-)
lead can be con
NOTE. Connect lead sets in groups or to individual locations on the SUT. If
necessary, use the grouper from the accessories kit to hold the lead set together.
NOTE. The lead sets have true differential inputs. For single-ended use, the
negative input of the lead set can be connected to ground. However, the ground
lead from each way station is provided to connect the probe ground to the SUT
ground.
be to the square pins on the SUT.
nected to the (-) side of a differential signal or to ground.
8P6910 General Purpose Probe Instruction Manual
Operating basics
Option 1K acce
ssories
Option 1K provides probe accessories to connect the P6910 probe to the SUT for
applications for less than 100 MHz. The following table and illustration show
the probe acc
Ref
numberItemQuantity
18-channel lead set2 ea196-3470-01
21-channel lead set1 ea196-3471-01
3
4Probe grabber tips2 ea
1
Instructions for installing the probe grouper are provided later in this manual. (See page 15, Install the probe
grouper (Option 1K and 2K).)
essories.
Probe Grouper kit
Header, 2 X 8 pin
2ea
1ea
1 package of 10 tips
Tektronix
part number
020-3042-
020-2896-00
00
Figure 4: Option 1K accessories
P6910 General Purpose Probe Instruction Manual9
Operating basics
Option 2K acce
ssories
Option 2K provides probe accessories with different means to connect the P6910
probe to the SUT.
Determine which method meets your particular needs. For some cases, you may
need to either glue or solder the adapters to the SUT before connecting the probe.
Not all of th
e accessories in the kit are needed for use with this probe.
The accessory kit is identical to the one used with the P6780 MSO probe. Either
the P6780 o
r P6910 probes can connect to any of the accessories in the kit. The
accessories enable users to share accessories between either probe as well as
switching between logic analyzer and oscilloscope probing applications.
The following table describes the contents of the accessory kit.
Tektronix part
ItemQuantity
Standard AdapterKit of 17 ea
Wire Tubing
Ferrite Bead
Wide Body Adapter
Flex Adapter
Insulated Wire
25°/55° HolderKit of 17 ea
Lead set GroundKit of 2 ea
Hand Browser
Probe Grouper
Header, 2 X 8 pin
Kitof34ea
Kitof17ea
Kitof17ea
Kitof17ea
15 ft. (4.57 m)
Kit of 2 ea
2ea
1ea
number
020-3035-01
020-3037-01
020-3034-01
020-3036-01
020-3033-01
020-3021-00
020-3032-01
020-3038-00
020-3031-00
020-3042-00
10P6910 General Purpose Probe Instruction Manual
Operating basics
Install the standard adapter
The following fi
Figure 5: Option 2K accessory kit contents
Use the general purpose standard adapters for soldering to the com
vias on the SUT.
gure shows the individual contents of the accessory kit.
ponent pins or
Install adapters on the PCB as n eeded to connect the probe to the SUT.
1. Locate the standard adapter in the accessory kit and trim the length according
to your needs.
2. Trim the wire tubing to the desired length and slide it over the adapter leads.
(SeeFigure6onpage11.)
Figure 6: Installing the standard adapter on the PCB
P6910 General Purpose Probe Instruction Manual11
Operating basics
Install the wide body
adapter
3. Attach the ferr
4. Insert the ends of the adapter onto the PCB.
5. Solder the adapter to the PCB.
6. Connect the probe lead set to the adapter.
Use the wide body adapter to firmly hold the probe lead set in place on the PCB.
The adapters can be stacked or grouped together depending on the connection
needs.
Install the wide body adapter on the PCB after installing the wire tubing and
ferrite bead:
1. Locate the wide body adapter in the accessory kit and trim the length
according to your needs.
2. Trim the wire tubing to the desired length, attach the ferrite bead (if needed
to improve the signal quality), and slide the ferrite bead and tubing over the
adapter leads.
3. Insert the ends of the adapter onto the PCB.
4. Solder the adapter to the PCB.
ite bead as needed to improve the signal quality. (See Figure 6.)
5. Identify where to place the wide body adapter on the PCB and glue it into
place. (See Figure 7 on page 12.)
Figure 7: Install the wide body adapter on the PCB
6. Connect the probe lead set to the adapter.
12P6910 General Purpose Probe Instruction Manual
Operating basics
Install the flex adapter
Use the flexible
1. Determine the location for the flex adapter on the circuit board.
The adapter connects to the SUT through the insulated wire provided with
the accessory kit.
2. Use a soldering iron to melt the insulation on the wire to connect the adapter
totheSUT.(SeeFigure8onpage13.)
Use two wires to connect differential signals from the probe to the SUT.
tip to connect to circuit board vias on the SUT.
Install the 25° /55° holder
Figure 8: Install the adapter to the SUT with the insulated wire
3. Solder one end of the wire to the circuit board as shown. (See Figure 8.)
Install the second wire in the same manner.
4. Install the flex adapter over the two wires and solder it in place.
5. Connect the probe lead set to the adapter.
Use the holders to provide strain relief for the flex adapter and standard adapter
a 25° or 55° angle.
at
1. Determine if you want to use the holder with the adapters at a 25° or 55° angle
n the circuit board.
o
Rotate the holder to work at either angle.
2. Position the holder so that the leads of the flex adapter or standard adapter
align to the connections on the circuit board and glue the holder into place.
(See Figure 9.)
P6910 General Purpose Probe Instruction Manual13
Operating basics
Install the hand browser
Figure 9: I
3. Install t
Use the hand browser to temporarily connect to signals o f interest on the SUT.
1. Insert the lead set into the hand browser. (See Figure 1 0.)
nstall the adapter holder by gluing it to the circuit board
he lead set and the adapter in the holder.
Figure 10: Insert the lead set into the hand browser
2. Position the probe tips as needed to connect to the signal of interest on the
SUT.
Theprobetipsswivel360°.
14P6910 General Purpose Probe Instruction Manual
Operating basics
Install the probe grouper
(Option 1K and
2K)
Use the probe gr
pins on the SUT.
1. Connect the le
The header te
grouper.
Figure 1
2. Place t
3. Remove the header from the lead sets.
1: Use the grouper to hold the lead sets together
ouper to group lead sets together for connections to adjacent
ad sets to the header. (See Figure 11.)
mporarily holds the lead sets in place while you install the
he grouper around the lead sets and close it in place.
4. Connect the assembly to the SUT.
Note the insertion force and clearance guidelines when connecting the probe
to the SUT. (See Figure 12 on page 16.)
P6910 General Purpose Probe Instruction Manual15
Operating basics
WARNING. To avoid damaging the lead sets, do not exceed the clearances or
insertion for
for clearance and insertion force information.
ce for the lead sets and square pins. Refer to the following illustration
Figure 1
2: Clearance and insertion force information for the lead sets
16P6910 General Purpose Probe Instruction Manual
Reference
Clock and qual
Table 1: Logic analyzer clock and qualifier availability
Logic analyzer width
34-channel
annel
68-ch
102-channel
136-channel
ifiers
Use the clock and qualifier probe connections from the SUT to define how the
logic analyzer stores data.
Every logic analyzer has some-special purpose input channels. Inputs designated
as clocks can cause the logic analyzer to store data. Qualifier channels can be
logically
from the SUT. Routing the appropriate signals from the design to these inputs
ensures that the logic analyzer can acquire data correctly. Unused clocks can be
used as qualifier signals.
Depending on the channel width, each logic analyzer will have a different set of
clock and qualifier channels. The following table shows the availability of the
clock and qualifier channels.
Clock inputsQualifier inputs
CLK:0CLK:1CLK:2CLK:3QUAL:0QUAL:1QUAL:2QUAL:3
AND’ed and OR’ed with clocks to further define when to latch data
lock and qualifier channels are stored. The logic analyzer always stores the
All c
logic state of these channels every time it latches data.
ce clock and qualifier channels are stored in memory there is no need to
Sin
double-probe these signals for timing analysis. When switching from state to
timing analysis modes, all of the clock and qualifier signals will be visible. This
also allows you to route regular signals, those not needed for clocking, to these
channels when they are not being used for their special purpose.
It is a good practice to take advantage of these channels to increase your options
for when you will latch data. Routing several of your design's clocks and strobes
to the logic analyzer clock inputs will provide you with a greater flexibility in the
logic analyzer clocking setup menus.
As an example, look at a processor with a ma ster clock, a data strobe, and an
address strobe. Routing all three of these signals to the logic analyzer clock inputs
will enable you to latch data on the processor master clock, only when data is
strobed, or only when address is strobed. Some forethought in signal routing can
greatly expand the ways to latch and analyze data.
P6910 General Purpose Probe Instruction Manual17
Reference
A processor als
o provides a good example of signals that can be useful as
qualifiers. There are often signals that indicate data reads versus data writes
(R/W), signals that show when alternate bus masters have control of the processor
buses (DMA), and signals that show when various memory devices are being
used (ChipSel). All of these signals are good candidates for assignment to
qualifier channels.
By logically AND'ing the clock with one of these qualifiers you can program the
logic analyzer to store only data reads or data writes. Using the DMA signal as a
qualifier p
rovides a means of filtering out alternate bus master cycles. Chip selects
can limit data latching to specific memory banks, I/O ports, or peripheral devices.
Multiplexed buses
The logic analyzer can demultiplex data from multiplexed buses.
Each sig
logic analyzer channel. Refer to the following tables to determine which channel
groups to connect to feed the test data to channels on the SUT.
nal on a dual- or quad-multiplexed bus can be demultiplexed into its own
102-channel logic
analyzer
———
———
———
———
68-channel logic
analyzer
C3:7-0C3:7-0
C2:7-0C2:7-0
——
——
——
——
34-channel logic
analyzer
—
—
18P6910 General Purpose Probe Instruction Manual
Reference
Table 3: 4X demu
Base connecting channel groups
C3:7-0C2:7-0 C1:7-0
A1:7-0A0:7-0 D1:7-0
A3:7-0A2:7-0 D3:7-0
E3:7-0E2:7-0 E1:7-0
CLK:3CLK:2 QUAL:3
CLK:1CLK:0 QUA
ltiplexing source-to channel assignments
Prime channels receiving SUT test data
136-channel l
analyzer
C0:7-0
D0:7-0
D2:7-0
E0:7-0
QUAL:2
L:1
QUAL:0
When demultiplexing data there is no need to connect the destination channels to
the multiplexed bus. Data from the source channels are routed to the d estination
channels internal to the logic analyzer.
Demultiplexing affects only the main memory for the destination channels. This
means that the MagniVu memory is filled with data from whatever is connected to
the demultiplexing destination channel probe inputs. This provides an opportunity
to acquire high resolution MagniVu data on a few extra channels. Connecting the
demultiplexing destination channels to other signals will allow viewing of their
activity in the MagniVu memory but not the main memory.
ogic
102-channel l
analyzer
C2:7-0 C1:7-0
C0:7-0
A0:7-0 D1:7-0
D0:7-0
A2:7-0 D3:7-0
D2:7-0
———
———
CLK:0 Q UA
QUAL:0
ogic
L:1
68-channel lo
analyzer
A3:7-0 A2:7C2:7-0
A0:7-0 D1:7-0
D0:7-0
——
——
gic
0
34-channel lo
analyzer
A3:7-0 A2:7C2:7-0
—
gic
0
Range recognizers
When using range recognizers, probe groups and probe channels must be in
hardware order.
Probe groups must be used from the most-significant probe group to the
least-significant probe group based on the following order:
Probe channels must be from the most-significant channel to the least-significant
channel based on the following order:
76543210
The above example assumes a 136-channel logic analyzer. The missing channels
in instruments with fewer than 136 channels are ignored. With merged modules,
range recognition extends across the first three modules: the master module
contains the most-significant channels.
P6910 General Purpose Probe Instruction Manual19
Reference
Probe signal names
Use signal names to simplify probe connections to the logic analyzer.
Refer to the following tables for information on signal connections. Match the
alpha character that precedes the channel identifier (for example, E3:7) to the
probe head label.
All differential data and clock/qualifiers on the probe may have the negative input
pin grounded and be used as a single-ended input.
You can find more information on 2X and 4X demultiplexing channel assignments
in the Demultiplexing Information section. (See Table 2 on page 18.) (See Table 3
on page 19.)
Table 4: Signal connections on 136- and 102-channel instruments for probe 4 and probe 3
Signal
namePodlet
Clk/QualClk/QualQ3-Q2-CK0-Q0-
Clk/Qual+Clk/QualQ3+Q2+CK0+Q0+
Data 7-
Data 7+
Data 6-6E3:6-E2:6-E1:6-E0:6-A3:6-A2:6-D3:6-D2:6-
Data 6+6E3:6+E2:6+E1:6+E0:6+A3:6+A2:6+D3:6+D2:6+
Data 5-
Data 5+
Data 4-4E3:4-E2:4-E1:4-E0:4-A3:4-A2:4-D3:4-D2:4-
Data 4+4E3:4+E2:4+E1:4+E0:4+A3:4+A2:4+D3:4+D2:4+
Data 3-3E3:3-E2:3-E1:3-E0:3-A3:3-A2:3-D3:3-D2:3-
Data 3+3E3:3+E2:3+E1:3+E0:3+A3:3+A2:3+D3:3+D2:3+
Data 2-2E3:2-E2:2-E1:2-E0:2-A3:2-A2:2-D3:2-D2:2-
Data 2+2E3:2+E2:2+E1:2+E0:2+A3:2+A2:2+D3:2+D2:2+
Data 1-1E3:1-E2:1-E1:1-E0:1-A3:1-A2:1-D3:1-D2:1-
Data 1+1E3:1+E2:1+E1:1+E0:1+A3:1+A2:1+D3:1+D2:1+
Data 0-0E3:0-E2:0-E1:0-E0:0-A3:0-A2:0-D3:0-D2:0-
Data 0+0E3:0+E2:0+E1:0+E0:0+A3:0+A2:0+D3:0+D2:0+
7
7
5
5
Probe 4 way station for 136-channel instrument
only
E3:7-E2:7-E1:7-E0:7-A3:7-A2:7-D3:7-D2:7-
E3:7+E2:7+E1:7+E0:7+A3:7+A2:7+D3:7+D2:7+
E3:5-E2:5-E1:5-E0:5-A3:5-A2:5-D3:5-D2:5-
E3:5+E2:5+E1:5+E0:5+A3:5+A2:5+D3:5+D2:5+
Probe 3 way station for 136- and 102-channel
instruments
20P6910 General Purpose Probe Instruction Manual
Reference
Table 5: Signal
Signal
name
Clk/Qual-Clk/Qual-CK1-CK2-CK3-Q1-
Clk/Qual+Clk/Qual+CK1+CK2+CK3+Q1+
Data 7-
Data 7+
Data 6-6A1:6-A0:6-D1:6-D0:6-
Data 6+6A1:6+A0:6+D1:6+D0:6+
Data 5-
Data 5+
Data 4-4A1:4-A0:4-D1:4-D0:4-
Data 4+4A1:4+A0:4+D1:4+D0:4+
Data 2+2A1:2+A0:2+D1:2+D0:2+
Data 3-3A1:3-A0:3-D1:3-D0:3-
Data 3+3A1:3+A0:3+D1:3+D0:3+
Data 2-2A1:2-A0:2-D1:2-D0:2-
Data 1-1A1:1-A0:1-D1:1-D0:1-
Data 1+1A1:1+A0:1+D1:1+D0:1+
Data 0-0A1:0-A0:0-D1:0-D0:0-
Data 0+0A1:0+A0:0+D1:0+D0:0+
connections on 102- and 136-channel instruments for probe 2 and probe 1
Podlet
7
7
5
5
Probe 2 way station for 136- and 102-channel
instruments
A1:7-A0:7-D1:7-D0:7-
A1:7+A0:7+D1:7+D0:7+
A1:5-A0:5-D1:5-D0:5-
A1:5+A0:5+D1:5+D0:5+
Probe 1 way station for 136- and 102-channel
instruments
C3:7-C2:7-C1:7-C0:7-
C3:7+C2:7+C1:7+C0:7+
C3:6-C2:6-C1:6-C0:6-
C3:6+C2:6+C1:6+C0:6+
C3:5-C2:5-C1:5-C0:5-
C3:5+C2:5+C1:5+C0:5+
C3:4-C2:4-C1:4-C0:4-
C3:4+C2:4+C1:4+C0:4+
C3:2+C2:2+C1:2+C0:2+
C3:3-C2:3-C1:3-C0:3-
C3:3+C2:3+C1:3+C0:3+
C3:2-C2:2-C1:2-C0:2-
C3:1-C2:1-C1:1-C0:1-
C3:1+C2:1+C1:1+C0:1+
C3:0-C2:0-C1:0-C0:0-
C3:0+C2:0+C1:0+C0:0+
P6910 General Purpose Probe Instruction Manual21
Reference
Table 6: Signal
Signal
name
Clk/QualClk/QualCK1-CK2-CK3-
Clk/Qual+Clk/QualCK1+CK2+CK3+
Data 7-
Data 7+
Data 6-6A1:6-A0:6-D1:6-D0:6-
Data 6+6A1:6+A0:6+D1:6+D0:6+
Data 5-
Data 5+
Data 4-4A1:4-A0:4-D1:4-D0:4-
Data 4+4A1:4+A0:4+D1:4+D0:4+
Data 3-3A1:3-A0:3-D1:3-D0:3-
Data 3+3A1:3+A0:3+D1:3+D0:3+
Data 2-2A1:2-A0:2-D1:2-D0:2-
Data 2+2A1:2+A0:2+D1:2+D0:2+
Data 1-1A1:1-A0:1-D1:1-D0:1-
Data 1+1A1:1+A0:1+D1:1+D0:1+
Data 0-0A1:0-A0:0-D1:0-D0:0-
Data 0+0A1:0+A0:0+D1:0+D0:0+
connections on 68- and 34-channel instruments
Podlet
7
7
5
5
Probe 2 way station for 68-channel instrument onlyProbe 1 way station for 34-channel
A1:7-A0:7-D1:7-D0:7-
A1:7+A0:7+D1:7+D0:7+
A1:5-A0:5-D1:5-D0:5-
A1:5+A0:5+D1:5+D0:5+
instrument only
C3:7-C2:7-
C3:7+C2:7+
C3:6-C2:6-
C3:6+C2:6+
C3:5-C2:5-
C3:5+C2:5+
C3:4-C2:4-
C3:4+C2:4+
C3:3-C2:3-
C3:3+C2:3+
C3:2-C2:2-
C3:2+C2:2+
C3:1-C2:1-
C3:1+C2:1+
C3:0-C2:0-
C3:0+C2:0+
A2:7-
A2:7+
A2:6-
A2:6+
A2:5-
A2:5+
A2:4-
A2:4+
A2:3-
A2:3+
A2:2-
A2:2+
A2:1-
A2:1+
A2:0-
A2:0+
22P6910 General Purpose Probe Instruction Manual
Reference
Probe footpri
nts
Pin spacing allows for space tolerances between the podlet holder and the
clock/qualifier configurations.
Refer to the following figure for footprint information. (See Figure 13.) Negative
inputs of differential signals can be grounded to support single-ended signal inputs.
Figure 13: P6910 General-Purpose probe land footprint
P6910 General Purpose Probe Instruction Manual23
Reference
Load models
Load models are important electrical considerations when working with the
P6910 probe.
The low frequency model is typically adequate for rise and fall times of 1 ns
or greater in a typical 25 Ω source impedance environment (50 Ω runs with a
pass-through connection). For source impedance outside this range, and/or rise
and fall times less than 1 ns, use the high frequency model to determine if a
significan
t difference is obtained in the modeling result.
The following electrical model displays a single podlet load model of the probe.
(See Figu
re 14.)
Specifications
Figure 14: Single podlet load model
The following tables list the electrical and environmental specifications for the
P6910 probe.
The electrical specifications apply when the probe is connected between a
compatible logic analyzer and a SUT. Refer to the Tektronix Logic Analyzer Series
(available on the TLA Documentation CD or downloadable from the Tektronix
Web site) for a complete list of system and module specifications. The probes are
designed to meet Tektronix standard 062-2847-00 class 5.
Table 7: Mechanical and electrical specifications
CharacteristicDescription
Threshold accuracy
Input resistance
Input capacitance<1.0 pF
Minimum digital signal swing300 mV single-ended
Maximum nondestructive input signal to
probe
Delay from probe tip to input connector
±(35 mV ±1% of setting)
20 kΩ ±1%
1
±15 V
7.70 ns ±80 ps
24P6910 General Purpose Probe Instruction Manual
Table 7: Mechanical and electrical specifications (cont.)
CharacteristicDescription
Probe length
Operating range
1
P6910 single podlet input capacitance is 0.7 pF, but podlets in a g roup will have <1 pF input capacitance.
1.8 m (6 ft)
+5 V to –2.5 V
Table 8: Environmental speci fi cations
CharacteristicDescription
Temperature
Operating0 °C to 50 °C (32 °F to 122 °F) with 5 °C/hour maximum
gradient, noncondensing
Nonoperating
Humidity
Operating5% to 95% relative humidity ≤ 30 °C (86 °F)
Nonoperating
Altitude
OperatingUp to 4600 m (15,092 ft)
Nonoperating
–55 °C to 75 °C (–67 ° to 167°F) with 5 °C/hour maximum
gradient, no media in instrument drive
5% to 75% relative humidity 30 °C to 60 °C (86 °F to 140 °F)
Non condensing
5% to 95% relative humidity ≤ 30 °C (86 °F)
5% to 75% relative humidity 30 °C to 50 °C (86 °F to 122 °F)
Non condensing
Up to 4600 m (15,092 ft)
Reference
P6910 General Purpose Probe Instruction Manual25
Reference
26P6910 General Purpose Probe Instruction Manual
Maintenance
Probe calibration information
The probe does not require calibration. If a probe failure occurs, return the entire
probe to your Tektronix representative for repair.
Probe service strategy information
The following service options are available when you order your Tektronix
product:
Table 9: Service options
OptionDescription
C3Calibration Service 3 Years
C5Calibration Service 5 Years
R3
R5
R3DW
R5DW
Includes initial certifications plus two annual calibrations
Includes initial certifications plus four annual calibrations
Repair Service 3 Years
Return product to Tektronix for servicing
Repair Service 5 Years
Return product to Tektronix for servicing
Repair Service Coverage 3 Years
(includes product warranty period). 3-year period starts at time of
instrument purchase
Repair Service Coverage 5 Years
(includes product warranty period). 5-year period starts at time of
instrument purchase
Perform the functional check
A functional check verifies basic functionality of the probe.
onnect the probe to the logic analyzer and to an a ctive signal source.
1.C
2. Open the Setup window where the probes are attached to the logic analyzer.
If your logic analyzer is a TLA7000 Series mainframe with multiple modules,
open the Setup window for the module where the probes are attached.
3. Check for signal activity in the Setup window for the attached probe.
P6910 General Purpose Probe Instruction Manual27
Maintenance
Inspect or clean the probe
Inspect and clean the probe as often as operating conditions require. Dirt acts as
an insulating blanket, preventing efficient heat dissipation. Dirt also provides an
electrical conduction path that can cause failures, especially under high-humidity
conditions.
Perform the following steps to c lean the probe:
1. Keep the probes free of dirt, dust, and contaminants to maintain a reliable
2. Remove dirt and dust with a soft brush.
3. Use only a damp cloth for more extensive cleaning.
Repackage the probe
The following information describes how to repackage the probe, to store the
probe, or to return the probes to the factory.
electrical probe connection.
Never use abrasive cleaners or organic solvents.
1. Use the original packaging, if possible.
original packaging is not available, use a corrugated cardboard shipping
If the
carton.
2. Add c
3. Enc
ushioning material to prevent the probes from moving inside the
shipping container.
lose the following information when shipping the probe to a Tektronix
Center:
ner’s address
Ow
Name and phone number of a contact person
Type of probe
Reason for return
Full description of the service required
28P6910 General Purpose Probe Instruction Manual
Replaceable parts
Replaceable parts
Parts orderin
g information
Replacement parts are available through your local Tektronix field office or
representative.
The P6910 probe contains no user-replaceable parts. However, probe accessories
can be replaced. (See page 1, P6910 probe accessory information.) Contact your
local Tekt
Changes to Tektronix products are sometimes made to accommodate improved
componen
improvements. When ordering parts, include the following information in your
order:
If you order a part that has been replaced with a different or improved part, your
local Tektronix field office or representative will contact you concerning any
change in the part number.
ronix representative for replacement information.
ts as they become available and to give you the benefit of the latest
Part number
ment type or model number
Instru
Instrument serial number
Instrument modification n umber, if applicable
P6910 General Purpose Probe Instruction Manual29
Replaceable parts
30P6910 General Purpose Probe Instruction Manual
Glossary
Functional check procedure
Functional check procedures verify the basic functionality of the probes by
confirming that the probes recognize signal activity at the probe tips.
Keep out area
Theareaof
be mounted.
Logic analyzer-end
The end of the probe which connects to the logic analyzer.
Module
The unit that plugs into a TLA7000 series mainframe which provides
instru
Podlet
A circuit contained in a flex lead and attached to a probe which provides
square-pin connections to the circuit under test for one data acquisition
channel and a reference pin.
Podlet holder
movable clip that groups eight individual podlets into a single 8-wide
Are
P6910 probe assembly. This provides ease when connecting to a row of 2 x
8 2.54 mm (0.100 in) square pins.
PCB
the printed circuit board in which only probe components m ay
ment capabilities such as logic analysis.
An acronym for Printed Circuit Board.
Probe head
The end of the probe that connects to the SUT.
SUT
System-Under-Test. Also known as the target system. The logic analyzer
connects to the SUT through the probe.
Way station
An intermediate probe part is u sed to connect the heads of the P6910 probes
to a single ribbon cable.