Tektronix P6616 General-Purpose Logic Probe Instructions User manual

xx
P6616 General-Purpose Logic Probe
Instruct
1
*P0712
071-2831-01
Product Description
The P6616 general-purpose logic probe connects the Tektronix MSO/DPO5000 Series and MSO4000B Series of mixed-signal oscilloscopes to digital buses and signals on your target system. The probe contains 16 data channels split between two lead sets (GROUP 1 and GROUP 2).
All leads include a ground connection at the tip. You can connect the probe leads separately to the target system, or group the leads together using the probe tip holders.
ions
83101*
Setting Up the Probe
To set and view the digital channel parameters, do the following:
On MSO/DPO5000 Series instruments, select Vertical>Digital Setup.
On MSO/DPO4000B Series instruments, push the D15–D0 button.
The parameters listed below can be set on each digital channel:
Threshold voltage and vertical position (the default threshold setting is 1.4 V)
Signal height and position (set once for all 16 channels)
Channel label
To set and view bus characteristics, do the following:
On MSO/DPO5000 Series instruments, use the controls in the Bus Setup screen.
On MSO/DPO4000B Series instruments, push the B1 to B4 buttons.
The setup screens allow you to set and view bus characteristics such as:
Clock type
Bus type (Serial or Parallel)
Bus width
Display format (Hex, Binary, or ASCII symbols)
Parallel bus setup information is resident on some oscilloscope models. However, for other buses such as SPI and I2C, you must have the appropriate option. See your oscilloscope manual or product data sheet for nomenclature and ordering details.
Functional Check
Logic activity immediately displays on all c onnected, active channels. If you do not see an active signal:
1. Push the Trigger button.
2. Select Edge for trigger type.
3. Select the channel that you are setting up as the source.
4. On MSO/DPO4000B Series instruments, push the Autoset
button.
If you do not see an active signal, try another probe channel (or analog probe) to verify circuit activity at the test point.
Typical Application
1. Use the P6616 probe to view digital signals on a system bus.
2. Use an analog probe, such as the TPP0500 or TPP1000 active probe to view analog waveform information.
Specications
Table 1: Electrical and mechanical specications
Characteristic Description
Input channels 16 digital
Input resistance
Input capac
Input signal swing
Maximum nondest– ructive input signal
Threshold voltage ± 40 V
Minimum detectable pulse width
Maximum input toggle rate
Digital channel-to­digital channel skew
Probe length
itance
Minimum 400 mV p-p
Maximum
100 k±1.0%
3.0 pF
30 V p-p, 200 MHz (centered around the DC threshold voltage) at the probe tip
10 V p-p, 200 MHz (centered around the DC threshold voltage) at the probe tip
30 V p-p, 42 V peak, ±50 VDC
1ns
500 MHz
200 ps
1.0 m (3.28 ft)
Table 2: Environmental specications
Characteristic Description
Temperature
Operating Nonoperating
Humidity
Operating Nonoperating
Altitude
Operating Nonop
erating
0 °C to +50 °C (+32 °F to +122 °F) –55 °C to +75 °C (–67 °F to +167 °F)
5% to 95% relative humidity 10% to 95% relative humidity
4.6 km (15,092 ft) maximum 50,000 ft) maximum
15 km (
Equipment Recycling. This product complies with the European Union’s requirements according
ctive 2002/96/EC on waste electrical
to Dire and electronic equipment (WEEE). For more information about recycling options, check the Support/Service section of the Tektronix Web site (www.tektronix.com).
Safety Summary
Connect and Disconnect Properly. Connect the probe output
to the measurement instrument before connecting the probe to the circuit under test. Disconnect the probe input and the probe ground from the circuit under test before disconnecting the probe from the measurement instrument.
Observe All Terminal Ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Do not Operate Without Covers. Do not touch exposed connections and components when power is present.
Connecting the Probe to the Oscilloscope
Connect the probe as shown in the illustration below.
1. Insert the probe label-side up into the connector on the oscilloscope.
2. To remove the probe, squeeze the buttons on the side and pull out the probe.
Connecting the Probe to Your Circuit
Attach the probe to the circuit using the connectors and adapters shown on the back of these instructions. Select the best method for your needs, and then proceed to Setting up the Probe.
Accessories
The following standard accessories ship with the probe and are shown in the illustration on the following page.
Item Description Quantity Part number
Logic probe accessory kit Item 1–6 020-2662-XX
1 Extension ground tip
2 Probe tip
3
IC grabber 1 set of 20
4 Probe tip holder 2 ea 352-1115-XX
5
8” Ground lead 1 set of 2
6
3” Ground lead 1 set of 8
Instructions
1
Instructions are included with the probe, but n The instructions can be downloaded at www.tektronix.com/manuals.
1
1 set of 20
1 set of 10
1 ea 071-2831-XX
ot in the accessory kit.
020-2711-XX
131-5638-11
020-2733-XX
020-2713-XX
020-2712-XX
These optional accessories can be ordered for your probe:
Description Part number
P6960 Probe D-MAX Footprint to Square Pin Header Adapter
NEX-P6960PIN
Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present.
Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualied service personnel.
Do Not Operate in Wet/Damp Conditions. Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Safety Terms and Symbols in This Manual.
These terms may appear in this m anual:
WARNING. Warning statements identify conditions or
practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or
practices that could result in damage to this product or other property.
Symbols o n the Product. This symbol may appear on the product:
Contacting Tektronix
Web site: www.tektronix.com
Phone: 1-800-833-9200
Address: Tektronix, Inc.
Email:
Department or name (if known) 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077
A
US
techsupport@tektronix.com
Warranty Information
For warranty information, go to www.tektronix.com/warranty.
x
Connecting the Probe to the Circuit
Copyright © Tektronix, Inc. A ll rights reserved. www.tek.com
xx
P6616
汎用ロジック・プローブ
取扱説明書
2
31-01
071-28
製品の説明
P6616 型汎用ロジック・プローブは、当社 MSO/DPO5000 シリーズおよび MSO4000B シリーズ・ ミックスドシグナル・オシロスコープをターゲット・シ ステムのデジタル・バスおよび信号に接続するために使 用します。プローブは、16 のデータ・チャンネルが 2 つのリード・セットに分割されています(GROUP 1 および GROUP 2)。
すべてのリードのチップにはグランド接続があります。 プローブ・リードは、ターゲット・システムに個別に接 続したり、プローブのチップ・ホルダを使用してグルー プ化したりすることができます。
プローブのセットアップ
デジタル・チャンネルのパラメータを設定したり表示す るには、次のようにします。
MSO5000/DPO5000 シリーズの機器では、 Vertical>Digital Setup を選択します。
MSO/DPO4000B シリーズの機器では、D15–D0 ボタ ンを押します。
各デジタル・チャンネルについて、下記の
パラメータを
設定できます。
スレッショルド電圧および垂直位置(デフォルトの スレッショルド電圧は 1.4 V)
信号の高さおよび位置(全 16 チャンネルを一括設 定)
チャンネル・ラベル
バス特性を設定したり表示するには、次のようにしま す。
MSO/DPO5000 シリーズの機器では、Bus Setup スク リーンのコントロールを使用します。
MSO/DPO4000B シリーズの機器では、B1 ~ B4 のボ タンを押します。
セットアップ画面で、次のバス特性を設定したり表示し たりできます。
クロックの種類 バスの種類(シリアルまたはパラレル) バス幅 表示形式(16 進、2 進、または ASCII シンボル)
一部のオシロスコープ・モデルにはパラレル・バスの設 定情報が格納されています。しかし SPI や I2C など、 その他のバスについては適切なオプションが必要です。 名称と発注情報については、ご使用のオシロスコープの マニュアルまたは製品データ・シートを参照してくださ い。
機能チェック
接続されたすべてのアクティブなチャンネルについて、 ロジック動作が直ちに表示されます。アクティブな信号 が表示されない場合は、次の操作を行ってください。
1. Trigger ボタンを押します。
2. トリガのタイプとして Edge を選択します。
3. ソースとしてセットアップするチャンネルを選択し ます。
4. MSO/DPO4000B シリーズの機器では、Autoset ボ タンを押します。
アクティブな信号が表示されない場合は、他のプロー ブ・チャンネル(またはアナログ・プローブ)を使用し て、テスト・ポイントの回路動作を確認してください。
主な用途
1. P6616 型プローブは、システム・バスのデジタル信 号の観測に使用します。
2. アナログ波形情報を観測するには、TPP0500 型ま たは TPP1000 型能動プローブなどのアナログ・プ ローブを使用します。
1
取扱説明書はプローブに付属しています。アクセサリ・キットにはあ りません。 取扱説 明書は www.tektronix.com/manuals からダ ウンロードするこ とができます。
下記のオプショナル・アクセサリをご購入いただけま す。
説明 部品番号
P6960 型プローブ D-MAX フット プリント用スクエア・ピン・ヘッ ダ・アダプタ
NEX-P6960PIN
仕様
表 3: 電気仕様と機械
特性 説明
入力チャンネル
入力抵抗
入力キャパシタン ス
入力信号スイング
最小値
最大値
最大非破壊 入力信号
スレッショルド電 圧
検出可能な最小パ ルス幅
最大入力トグル・ レート
デジタル・チャン ネル間スキュー
プローブ長
仕様
デジタル 16 チャンネル
100 kΩ ±1.0
3.0 pF
400 mVp-p
30 V p-p、≤200 MHz (プローブ・ チップで DC スレ を中心に)
10 V p-p、≥200 MHz (プローブ・ チップで DC スレッショルド電圧 を中心に)
30 V p-p、42 V ピーク、±50 VDC
±40 V
1ns
500 MHz
200 ps
1.0 m(3.28 フィート)
%
ッショルド電圧
表 4: 環境仕様
特性 説明
温度
動作時 非動作時
湿度
動作時 非動作時
高度
動作時 非動作時
0~+50℃ –55 ~ +75 ℃
相対湿度 5 ~ 95% 相対湿度 10 ~ 95%
最高 4.6 Km(15,092 フィート) 最高 15 Km(50,000 フィート)
プローブとオシロスコープの接続
以下の図に示すようにプローブを接続します。
1. ラベル面を上にして、プローブをオシロスコープの コネクタに挿入します。
2. プローブを取り外すには、両サイドのボタンを押し てプローブを引き抜きます。
プローブと測定回路の接続
プローブを回路に接続するには、本書の末尾に記載され ているコネクタとアダプタを使用します。最適な方法を 選んで、次の「プローブのセットアップ」に記載された 指示に従ってください。
x
アクセサリ
本プローブには下記の表に示すスタンダード・アクセサ リが付属しています。後の図を参照してください。
目説 量 部
ロジック・プローブ・
アクセサリ・キット
1
延長グランド・チップ
2
プローブ・チップ
3
IC グラバ
4
プローブ・チップ・ホ ルダ
5
8 インチ・グランド・ リード
6
3 インチ・グランド・ リード
取扱説明書
1
項目 1–6
20 本入り セット
10 本入り セット
0 本入り
2 セット
2個
2 本入り セット
8 本入り セット
1個
番号
020-2662-XX
020-2711-XX
131-5638-11
020-2733-XX
352-1115-XX
020-2713-XX
20-2712-XX
0
071-2831-XX
機器のリサイクル本製品は WEEE Directive
/96/EC (廃棄電気・電子機器に関する
2002 指令)に基づく EU の諸要件に準拠してい ます。リサイクル方法の詳細については、 当社 Web サイト(www.tektronix.com)の 「Support/Service」を参照してください。
安全にご使用いただくために
接続と切断の手順を守ってください: 測定対象の回路にプ
ローブを接続する前に、プローブ出力を計測機器に接続 してください。計測機器からプローブを外す前に、測定 対象の回路からプローブの入力とグランドを外してくだ さい。
すべての端子の定格に従ってください: 火災や感電の危険 を避けるために、本製品のすべての定格とマーキングに 従ってください。本製品に電源を接続する前に、定格の 詳細について、製品マニュアルを参照してください。
カバーを外した状態では使用しないでください: 電源が投入 されているときに、露出した接続部分や部品に触れない でください。
回路の露出を避けてください: 電源が投入されているとき に、露出した接続部分や部品に触れないでください。
故障の疑いがあるときは使用しないでください: 本製品に 故障の疑いがある場合、資格のあるサー ビス担当者に 検査してもらってください。
湿気の多いところでは使用しないでください: 爆発しやすい 環境では動作させないでください。
製品の表面を清潔で乾燥した状態に保ってください:
安全に関する用語と記号
このマニュアルでは次の用語を使用します。
警告: 人体や生命に危害をおよぼすおそれのある状
態や行為を示します。
注意: 本製品やその他の接続機器に損害を与えるお
それのある状態や行為を示します。
Loading...
+ 5 hidden pages