P6460 DATA ACQUISITION
PROBE
INTRODUCTION
The P6460 is a 100 MHz data acquisition probe suitable for use with the 1240/ Acquisition Cards and the DAS Acquisition Modules. The P6460 has nine data channels and a clock line. When used with the DAS, one channels serves as a clock qualifier.
The P6460 probe can also be used with the following DAS modules:
91
S 16/32 Pattern Generation Module 92A 16/16E Data Acquisition Module 92S 16/32 Pattern Generation Module
These Instructions contain the service information for this probe. For convenience, put a copy
070-4345-00
Product Group 57/5E
1241
Logic Analyzer 1240D1 and 1240D2
91
A24 and
of
them
P6460
data acquisitio n probe.
service manual for the logic analyzer that the probe will usually be used with.
Labels
91
AE24 Data
of
the data
Two labels are included with your probe. The one with an 8 in the upper left corner and a C/Q in the upper right corner is for use with 1240/ upper left corner and a with DAS91 A24 and
Operator's Infor
This manual contains primarily service information. For operator's information, refer instrument that the probe is
1240/1241 Logic Analyzer Operator's Manual
in
the back
Copyright © 1983, 1987, Tektronix, Inc.
of
All righ
the
ts
DAS 9100 Series Operator's Manual.
reserved.
1241
Logic Analyzer. The label with a
CK
in the upper right corn
91
AE24 Data Aquisition Modules.
mat
ion
to
the operator's manual
to
be used with, either the
FIRST PRINTING JANUARY 1983
VIS
ED PRINTING DECEMB
RE
Qin
er
is for use
ER
4345-1
of
or
1987
the
the
the
P6460
ACCESSORIES
Standard Accessories:
1 Instruction Sheet
Lead Set, 10 inch Package
(each tip is a 206-0222-00)
2 Gro
with Pomona Hook Tips (344-0267-00)
2 Alligator Clips (substitute for hook
tips above)
When ordering the P6460 for the 92A 16 use the part number 010-6460-16. This insures you will receive the appropriate standard accessories for the 92A 16/16E listed below.
010-6460-16 Standard Accessories:
1 Instruction Sheet 1 Lead Set, 5 inch 2 Ground VL Sense Leads, 5 inch with
Pomona Hook Tips (344-0267 -00)
2 Alligator Clips (substitute for hook
tips above)
und·
of
12 Probe Tips
(or
VL)
Sense Leads, 5 inch
070-4345-00 012-07 4 7-00
020-1386-00 012-0989-00 344-0046-00
070-4345-00 012-0987-00
012-0989-00 344-0046-00
Optional Accessories:
Flying Lead Set, 5 inches long Flying Lead Set, 25 cm long 40-pin Univ. Probe lntfc. Kit 1240 Diagnostic Lead Set Diagnostic Lead Set, 10 inches long Package of 10 Ground (or
Leads with Pomona Hook Tips
(344-0267 -00) GPIB Connector/Adapter IC
Extractor, 16 pin
Clip,
Adapter, Test Adapter, Test Clip,
cable (requires 380-0560-05 Adapter) Adapter, Test Clip, 40 DIP, 30 cm
cable (requires 380-0560-05 Adapter) Adapter, required for use with
Adapters
16 DIP
40
VL)
DIP, 10 cm
Sense
40
DIP
012-0987-00 012-0800-00 020-1 041-00 198-5069-00 012-1 000-00
012-0989-01 103-0209-00 003-0709-00 015-0330-00
015-0339-02 015-0339-00 380-0560-05
''
GROUND CONNECTIONS AND
FREQUENCY
For clock operating frequencies up to 25 MHz, the 10-inch lead sets (012-0747-00 or 012-0800-00) with one USER'S GND lead connected are sufficient.
I
WARNING
Disconnect power before removing protective coverings, solderi
2
ng, or
replacing components.
Standard accessories for the
F
or MHz, 10-inch lead sets may be used leads are connected.
For freq
sure
USER'S
P6460,
clock operating frequencies above 25 MHz but below
uen
cies above 50 MHz and noisy environments, be
to
use the 5-inch lead sets (012-0987-00) and both
GND
lea
ds with alligator clips.
SERVICE INFORMATION
TEKTRONIX FIELD SERVICE
I
Service of Tektronix products is available from Tektronix Service Centers around the world. Contact your local sales or
service organization for more information.
4345
if
both USER'S GND
REV
FEB
198
-2
50
7
SPECIFICATIONS
Table 1
ELECTRICAL SPECIFICATIONS
P6460
Characteristic
MODULE-TO-PROBE SIGNALS
Max. Operating Power Requirements
PROBE-TO-MODULE SIGNALS
Data Channel Signal Differential ECL
User's Ground Sense
Input Resistance 1
Input Capacitance
Max. Non-Destructive In
put Voltage Range
Performance Requirements
þÿM©
± 1%
Supplemental Information
+ 5 V (
± 5%) at 600
lines are terminated to at 300
mA
when all output lines are
unterminated
Vee
= + 5 v
Vee=
Gnd
< 100 þÿ© to user's ground
5 pF nominal
± 40 V (DC + peak AC)
mA
when 20 output
+3
V through 60
þÿ©,
Max. Voltage Between Any Two Inputs
Operating Input
Voltage Range
Threshold Offset and Accuracy
Minimum Input Swing 0.5 V p-p centered on the threshold
Minimum
(with input 250 mV
over the threshold from + 0.5 V and - 0.5
Pu
lse Width
V)
±
60
V (DC + peak AC)
From - 40 V to input threshold's voltage
+ 10 V ( + 30 V for
± 0.25%
4 ns at threshold
of
threshold ± 50 mV
RS
-232 only)
3
P6460
ENVIRONMENTAL
Characteristic
Temperature
Operating St
orage
Humidity
Altitude
Operating Non-operating
trical Di
Elec
Table 2
SPECIFICATIONS
Description
-15°C
to
+ 55°C
-62
°C
to
+ 75
°C
95%
to
97% relative humidity
ft
scharge
4.5 km (15,000 15
km
(50,000 ft.)
5 kV maximum from 200
þÿ© series resistance
2K
STATIC PRECAUTIONS
Table
3
PHYSICAL
Characteristic
We
ight
Overall Dimensions
Pod
Cable
.)
pf
with
4. Nothing capable should be allowed on the work station surface.
5.
Keep the component leads shorted
possible.
6.
Pick up components by the body, never
SPECIFICATIONS
Description
12
oz.
4.5 in. long, 2.2 in. wide, 0.85 in. deep
78.75 in. (2 m)
of
generating
±
10%
or
holding a stat
together whenever
by
ic
charge
the leads.
Static discharge can damage in
this instrument.
Observe the following precautions to avoid
of
1. Min imize handling
2. Transport and store static-sensitive components or as­semblies
on conductive foa
static-sensitive components
3. Discharge the static volta a wrist strap while handling these components. Servic­ing static-sensitive assemblies should in a static-free
perso
in
their original conta iners, or on a metal rail, or
nnel.
static-sensitive components.
m.
Label any package that contains
work stat
any
semiconductor
stat
ic damage:
or
assemblies.
ge
from your body by wearing
be
performed only
ion by qualified service
7.
Do n
ot
slide the components over any surface.
8. Avoid handling components wo
rk-surface covering capable
charge.
9. Use a soldering iron that is connected to earth ground. 10
. Use only special anti-static suction type
in
desolder
Damage immediately apparent. tionary measures listed above when handling static-sensitive components.
g tools.
to
electrical components
NOTE
Always
in
areas that have a floor
of
generating a
or
w
may
not
be
follow the
precau
ick
-
or
stat
type
ic
4
Specification Checked
P6460
PERFORMANCE CHECK
The input resistance
of
each channel should be 1 þÿM© ± 1 %.
Required Equipment
One digital multimeter with For example, the Tektronix DM 5010.
at
least 0.2% accuracy at 1 þÿM©.
Procedure
1. With the probe disconnected, measure the resistance between the input any other channel. Call this value
to
the channel you are checking and
Ra.
2. Next, measure the resistance between the input channel you want this value
3. Finally, measure the resistance between the channels. Call this value
4.
Calculate the value channel using:
R
Rinput =
b.
to
measure and a third channel. Call
two
R
e.
of
the input resistance of the original
Ra+ Rb
-
Rc
2
to
the
other
5