Product Description
The P6410 logic analyzer probe connects the Tektronix
TLA series logic analyzer to your target system. The probe
consists of 16 data channels and one clock/data channel. You
can connect the probe leads separately to the target system,
or together using the groupers
groupers installed.
. The probe ships with the
Connecting the Probe to the Logic Analyzer
Connect the probe (and optionally, the retaining brackets) as
shown in the following figure. The retaining brackets and
hardware ship with the logic analyzer.
(DUT)
20 KΩ
1.4 pf (Data)
2.0 pf (Clock/Qual)
Loading and equivalent circuit
(LA module)
Labeling Your Probe
1. Determine the channel group/color that you will use for
the probe and select the appropriate label group from the
label sheet.
2. Apply the smallest of the three labels to the module end.
3. Apply the remaining two labels to the DUT ends of the
probe, on the sides without the WEEE labels (shown in
Equipment Recycling). Match the corners of each label
to the label detents in the probe heads.
2
1
2
Match color-coded labels
Connecting the Probe to the Circuit
The probe is shipped with the groupers installed so that you
can connect to 2x8 rows of 0.025--inch square pins. The
clock lead is separate. Refer to the back of these instructions
for other connection methods using the accessories that are
included with the probe.
Reference/Probe Connectors
This section provides reference information and
specifications for the probe.
Orange probe
Brown probe
Gray probe
White probe
Yellow probe
Blue probe
Violet probe
Green probe
Section names
A0
A2
C0
C2
D0
D2
E0
E2
0 7+ 0 7+ CLK+
A1
A3
C1
C3
D1
D3
E1
E3
Clock channel
CK1
CK0
Q1
CK3
CK2
Q0
Q2
Q3
6.35 mm min
(0.250 in)
17.78 mm
(0.700 in)
Pin spacing between 8- channel groups
and clock/qual leads
17.78 mm
(0.700 in)
2.54 mm
(0.100 in)
2.54 mm
(0.100 in)
2.80 mm max
(0.110 in)
23.37 mm max
(0.920 in)
8- channel probe footprint with
8- channel grouper attached
2.54 mm
(0.100 in)
5.08 mm min
(0.200 in)
4.57 mm max
(0.180 in)
11.68 mm max
(0.460 in)
4
3
4
3
Specifications
Table 1 lists the electrical and mechanical specifications for
the P6410 probe. The electrical specifications apply when
the probe is connected between a compatible logic analyzer
and a target system.
Table 1: Electrical and mechanical specifications
Characteristic
Threshold accuracy ±100 mV
Channel-to-channel-skew ≤1ns
Input resistance
(probe and lead sets)
Standard lead set 20 KΩ
Reduced bias set 20 KΩ
Input capacitance
Data 2.0 pF typical
Clock/Qual 2.0 pF typical
Minimum slew rate 0.2 ns
Maximum operating signal 6.5 V
Maximum nondestructive input
signal to probe
Minimum input pulse--width signal 2ns
Delay from probe tip to module
input connector
Probe length 1.8 m (6 ft)
Description
P--P
--3.5 V absolute input voltage min
6.5 V absolute input voltage max
±15 V
7.33 ns
Table 2 shows the environmental specifications for the
probe. The probe is designed to meet Tektronix standard
062-2847-00 class 5.
Table 2: Environmental specifications
Characteristic
Temperature
Maximum operating
Minimum operating
Nonoperating -- 5 5 °C to +75 °C
Humidity 10 to 95% relative humidity
Altitude
Operating 4.5 km (15,000 ft) maximum
Nonoperating 15 km (50,000 ft) maximum
Electrostatic Immunity The probe is not static sensitive
Description
+50
°C (+122 °F)
0 °C (+32 °F)
(--67 °F to +167 °F)
0 7- 0 7-
CLK-
Grouping footprint (16 Channels plus one clock)
Standard Accessories
The following standard accessories ship with the probe and
are shown on the back of these instructions.
Description Quantity Part number
Extension ground tip 20 ea 206-0559-XX
Probe tip 1 set of 10 131-5638-11
IC grabber 20 ea 206-0569-XX
Lead set 8ea 196-3501-XX
Ground lead set 2ea 196-3497-XX
Grouper (shipped on
probe)
Label set 1 sheet 335-1770-XX
2ea 352-1115 -XX
Warranty Information
For warranty information, go to
www.tektronix.com/warranty
Contacting Tektronix
Tektronix, Inc.
14200 SW Karl Braun Drive
PO Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical
support:
H In North America, call 1-- 800--833--9200.
H Worldwide, visit www.tektronix.com to find contacts in
your area.
Equipment Recycling. This product complies
with the European Union’s requirements
according to Directive 2002/96/EC on waste
electrical and electronic equipment (WEEE).
For more information about recycling options,
check the Support/Service section of the
Tektronix Web site (www.tektronix.com).
P6410
Logic Analyzer Probe
Instructions
www.tektronix.com
*P071219300*
071-2193-00

Using the Probe Accessories
1
Notches
2
Spread grouper to disengage retention
tangs from notches in leads
Retention tangs
Push leads forward
CLK
Remove leads
3
individually from grouper
Rotate 90° to remove
4
To reassemble, reverse the procedure
2
2
1
1
1
2
Copyright © Tektronix, Inc.
1
1
2
2