Tektronix P6139B, P5050B Instructions Manual

Revision A
P6139B & P5050B 500 MHz 10X Passive Probes
Instructions
500 MHz 10X Probes
The P6139B & P5050B probes are high impedance, passive probes that are designed for use with Tektronix ground-referenced oscilloscopes, including the DPO/MSO/TDS 3000/4000/5000/7000 series oscilloscopes.
Both probes have a bandwidth of >500 MHz and 10X attenuation. The P6139B probe has a compensation range of 8 to 18 pF, while the P5050B probe has a compensation range of 15 to 22 pF.
Connecting the Probe to the Oscilloscope
Connect the probe to the oscilloscope as shown below.
Compensating the Probe
Due to variations in oscilloscope input characteristics, the low-frequency compensation of the probe may need adjustment after moving the probe from one oscilloscope channel to another.
If a 1 kHz calibrated square wave displayed at 1 ms/division shows signicant differences between the leading and trailing edges, perform the following steps to optimize low-frequency compensation:
1.
Connect the probe to the oscilloscope channel that you plan to use for your measurements.
2.
Connect the probe to the probe compensation output terminals on the oscilloscope front panel.
WARNING.
Probe Comp signal on the oscilloscope when making this adjustment.
To avoid electric shock, only use the insulated adjustment tool when making compensation adjustments.
To avoid electric shock, only connect to the
3.
Push AUTOSET or otherwise adjust your oscilloscope to display a stable waveform.
4.
Adjust the trimmer in the probe until you see a perfectly at-top square wave on the display. (See illustration.)
Standard Accessories
The accessories included with the probe are shown below.
WARNING.
To avoid electric shock when using the probe or accessories, keep ngers behind the nger guard of probe body and accessories.
To reduce risk of shock, verify that the ground lead and ground spring are fully mated before connecting the probe to the circuit under test.
Item Description
Universal IC Cap
Use this cap to prevent shorting the
probe tip between IC pins.
Press the cap on the probe tip until it snaps on, and then spin the cap to expose the probe tip toward the IC lead.
Reorder Tektronix part number 013-0366-xx
Hook tip
Press the hook tip on the probe tip and then clamp the hook on the circuit.
Rating: 300 V CAT II
Reorder Tektronix part number 013-0362-xx
Color bands
Use these bands to identify the oscilloscope channel at the probe head.
Reorder Tektronix part number 016-0633-xx (5 pairs)
Item Description
Ground lead, with alligator clip
Attach the lead to the probe head ground and then to your circuit ground.
Reorder Tektronix part number 196-3521-xx
Ground spring
This spring minimizes aberrations on high frequency signals caused by ground path inductance, giving you measurements with good signal delity.
Attach the spring to the ground band on the probe tip. Bend the spring in or out, up to ~0.75 in.
Reorder Tektronix part number 016-2028-xx
Adjustment tool
Use only this insulated tool for compensation adjustments.
Reorder Tektronix part number 003-1433-xx
Optional Accessories
The accessories shown below are available for the probes and are rated 30 V unless indicated otherwise.
Accessory Part number
MicroCKT Test Tip
Micro Hook TipRating: 300 V CAT II
BNC to Tip Adapter, Unterminated
Circuit B oard Test Point/PCB Adapter
Chassis-Mount Probe Test Jack
6” Clip-on Ground Lead
12” Alligator Ground Lead
Ground Spring, Short
206-0569-xx
013-0363-xx
013-0367-xx
016-2016-xx
131-4210-xx
196-3198-xx
196-3512-xx
016-2034-xx
*P071271000*
x
071-2710-00
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Revision A
Replacing the Probe Tip
Order Tektronix part number 206-0635-xx (P6139B) or 206-0636-xx (P5050B). The insulator sleeve, part number 204-1226-xx, is reorderable and is common to both probes.
WARNING.
Only qualied service personnel should replace the probe tip. Disconnect the probe before servicing. After replacing the tip, securely tighten the insulator sleeve before using the probe.
Specications
Table 1: Electrical and mechanical specications
Characteristic Specication
Bandwidth
System attenuation accuracy 10:1 ±0.5% @DC
Compensation range
System input resistance 10 M±0.5% @DC
System input capacitance
Propagation delay ~6 ns
Maximum tip input voltage
Cable length
DC to 500 MHz (–3 dB)
P6139B: 8 pF–18 pF P5050B: 15 pF–22 pF
P6139B: <8 pF P5050B: <11 pF
300 V
CAT II and DC
RMS
1.3 m
Table 2: Environmental specications
Characteristic Description
Temperature
Operating
Nonoperating
Humidity
Operating 5% to 95% relative humidity (%RH) up to
Nonoperating
Altitude
Operating
Nonoperating
–15°C(+5°F)to+65°C(+149°F)
–62 °C to +85 °C (–80 °F to +185 °F)
+30 °C, 5% to 75% RH above +30 °C up to +65 °C. Noncondensing
5% to 45% RH above +65 °C up to +85 °C. Noncondensing
3.0 km (10,000 ft) maximum
15 km (50,000 ft) maximum
Table 3: Certications and compliances
Characteristic Description
EC Declaration of Conformity
Measurement Category Product Examples
Compliance was demonstrated to the following specication as listed in the Ofcial Journal of the European Communities:
Low Voltage Directive 2006/95/EC: EN61010-031: 2008
CATIII: Distribution-level mains, xed installation
CAT II: Local-level mains, portable equipment
CAT I: Circuits not directly connected to mains
Characteristic Description
Pollution Degree 2
Additional Safety Standards
Do not operate in environments where cond– uctive pollutants may be present (as dened in IEC 61010-1). Rated for indoor use only.
UL61010-031;2010 CAN/CSA C22.2 No. 61010-031:07/A1:2010 IEC61010-031; IEC 61010-031/A1:2008
Equipment Recycling.
with the European Union’s requirements according to Directive 2002/96/EC on waste electrical and electronic equipment (WEEE). For more information about recycling options, check the Support/Service section of the Tektronix Web site (www.tektronix.com).
This product complies
Safety Summary
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specied. Using the probe or accessories in a manner not specied could result in a shock or re hazard.
To Avoid Fire or Personal Injury
Ground-Referenced oscilloscope use.
Do not oat the reference lead of this probe when using with ground referenced oscilloscopes (for example, DPO, MSO, and TDS series oscilloscopes). The reference lead must be connected to earth potential (0 V).
Connect and disconnect properly.
Connect the probe output to the measurement instrument before connecting the probe to the circuit under test. Disconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe from the measurement instrument.
Avoid electric shock.
To avoid injury or loss of life, do not connect or disconnect probes or test leads while they are connected to a voltage source.
Observe all terminal ratings.
To avoid re or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Avoid electric shock.
When using probe accessories, never exceed the lowest rating of the probe or its accessory, whichever is less, including the measurement category and voltage rating.
Avoid electric overload.
To avoid injury or re hazard, do not apply potential to any input, including the reference inputs, that varies from ground by more than the maximum rating for that i nput.
Avoid exposed circuitry and do not operate without covers.
Do not touch exposed connections and components
when power is present.
Inspect the probe and accessories.
Before each use, inspect probe and accessories for damage (cuts, tears, defects in the probe body, accessories, cable jacket, etc.). Do not use if damaged.
Do not operate in Wet/Damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
Safety Terms and Symbols Terms in This Manual.
These terms may appear in this manual:
WARNING.
Warning statements identify conditions or
practices that could result in injury or loss of life.
CAUTION.
Caution statements identify conditions or practices that could result in damage to this product or other property.
Symbols on the product.
These symbols may appear on
the product:
Contacting Tektronix
Web site: www.tektronix.com