tional copyright laws and international treaty provisions.
Tektronix pro
previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
ducts are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tek.com to find contacts in your area.
Warranty
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of
shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective
product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and
replacement products used by Tektronix for warranty work may be new or reconditioned to like new performance. All replaced
parts, modules and products become the property of Tektronix.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period
and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the
defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the
product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall
be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and
care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from a ttempts by personnel
other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or
connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or
d) to service a product that has been modified or integrated with other products when the effect of such modification or integration
increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTR ON IX WITH RES PEC T TO THE PR OD UC T IN LIEU OF ANY OTHER WARRANTIES,
EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS
IS THE SOLE AND E XCLU S IVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRAN TY. TEKTRONIX
AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES
IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH
DAMAGES.
[W2 – 15AUG04]
Table of Contents
Important Safety Information..........................................................................................................iii
General Safe
Service Safety Summary ........................................................................................................v
P5200A Series H igh Voltage Differential Probes Instruction Manuali
Table of Content
s
iiP5200A Series High Voltage Differential Probes Instruction Manual
Important Safet
y Information
Important Saf
This manual contains information and warnings that must be followed by the user for safe operation and to keep the
product in a safe condition.
To safely perform service on this product, additional information is provided at the end of this section. (See page v, ServiceSafety Summary.)
ety Information
General Safety Summary
Use the prod
or any products connected to it. Carefully read all instructions. Retain these instructions for future reference.
This product shall be used in accordance with local and national codes.
For correct and safe operation of the product, it is essential that you follow generally accepted safety procedures in addition
to the safety precautions specified in this manual.
The product is designed to be used by trained personnel only.
Only qualified personnel who are aware of the hazards involved should remove the cover for repair, maintenance, or
adjustme
Before us
This pro
uct only as specified. Review the following safety precautions to avoid injury and prevent damage to this product
nt.
e, always check the product with a known s ource to be sure it is operating correctly.
duct is not intended for detection of hazardous voltages.
Use pers
While us
component manuals for warnings and cautions related to operating the system.
When incorporating this equipment into a system, the safety of that system is the responsibility of the assembler of the system.
onal protective equipment to prevent shock and arc blast injury where hazardous live conductors are exposed.
ing this product, you may need to access other parts of a larger system. Read the safety sections of the other
To Avoid Fire or Personal Injury
Use Proper Power Cord. Use only the power cord specified for this product and certified for the country of use.
Do not use the provided power cord for other products.
nd the Product.
Grou
To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input
or output terminals of the product, make sure that the product is properly grounded.
Do not disable the power cord grounding connection.
Power Disconnect. The power cord disconnects the product from the power source. See instructions for the location.
Do not position the equipment so that it is difficult to operate the power cord; it must remain accessible to the user at all
times to allow for quick disconnection if needed.
Use Proper AC Adapter. Use only the AC adapter specified for this product.
This product is indirectly grounded through the grounding conductor of the mainframe power cord.
P5200A Series H igh Voltage Differential Probes Instruction Manualiii
Important Safet
Connect and Disconnect Properly. Do not connect or disconnect probes or test leads while they are connected
to a voltage sou
y Information
rce.
Use only insul
suitable for the product.
ated voltage probes, test leads, and adapters supplied with the product, or indicated by Tektronix to be
Observe all Terminal Ratings. To avoid fire or shock hazard, observe all ratings and markings on the product. Consult
the product manual for further ratings information before making connections to the product. Do not exceed the Measurement
Category (CAT) rating and voltage or current rating of the lowest rated individual component of a product, probe, or accessory.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Do Not Opera
open. Hazardous voltage exposure is possible.
te Without Covers.
Do not operate this product with covers or panels removed, or with the case
Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present.
Do Not Operate with Suspected Failures. If you suspect that there is damage to this product, have it inspected by
qualified service personnel.
Disable the product if it is damaged. Do not use the product if it is damaged or operates incorrectly. If in doubt about safety of
the produ
Before us
use probes or test leads if they are damaged, if there is exposed metal, or if a wear indicator shows.
Examine the exterior of the product before y ou use it. Look for cracks or missing pieces.
Use only specified replacement parts.
ct, turn it off and disconnect the power cord. Clearly mark the product to prevent its further operation.
e, inspect voltage probes, test leads, and accessories for mechanical damage and replace when damaged. Do not
Do Not Operate in Wet/damp Conditions. Be aware that condensation may occur if a unit is moved from a cold to a
warm environment.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Remove the input signals before you clean the product.
Probes and Test Leads
Before connecting probes or test leads, connect the power cord from the power connector to a properly grounded power
outlet.
Keep fingers behind the protective barrier, protective finger guard, or tactile indicator on the probes.
Remove all probes, test leads and accessories that are not in use.
Use only correct Measurement Category (CAT), voltage, temperature, altitude, and amperage rated probes, test leads,
and adapters for any measurement.
ivP5200A Series High Voltage Differential Probes Instruction Manual
Important Safet
y Information
Beware of High Voltages. Understand the voltage ratings for the probe you are using and do not exceed those ratings.
The following r
ating is important to know and understand:
The maximum me
The voltage ra
information.
WARNING. To prevent electrical shock, do not exceed the maximum measurement or maximum floating voltage for the
oscilloscop
To prevent da
cause induction heating of the jaws.
e input BNC connector, probe tip.
mage to the alligator clip insulation, do not use in high A/m magnetic fields at high frequencies, which can
asurement voltage for probe tip.
ting depends on the probe and your application. Refer to the Specifi cations section of the manual for more
Connect and Disconnect Properly. Connect the probe output to the m easurement product before connecting the
probe to th
the measurement product.
e circuit under test. Disconnect the probe input from the circuit under test before disconnecting the probe from
Inspect the Probe and Accessories. Before each use, inspect probe and accessories for damage (cuts, tears, or
defects in the probe body, accessories, or cable jacket). Do not use if damaged.
Service Safety Summary
The Service safety summary section contains additional information required to safely perform service on the product. Only
qualified personnel should perform service procedures. Read this Service safety summary and the General safety summary
erforming any service procedures.
before p
To Avoid Electric Shock. Do not touch exposed connections.
Do Not S
rendering first aid and resuscitation is present.
ervice Alone.
Do not perform internal service or adjustments of this p roduct unles s another person capable of
Disconnect Power. To avoid electric shock, switch off the product power and disconnect the power cord from the mains
power before removing any covers or panels, or opening the case for servicing.
Use Care When Servicing with Power On. Dangerous voltages or currents may exist in this product. Disconnect
r, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing
powe
components.
Verify Safety After Repair. Always recheck ground continuity and mains dielectric strength after performing a repair.
P5200A Series H igh Voltage Differential Probes Instruction Manualv
Important Safet
y Information
Terms in this Manual
These terms may appear in this manual:
WARNING. Warning s tatements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Symbols and Terms on the Product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the m arking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
When this symbol is marked on the product, be sure to consult the manual to find out the nature of the
potential hazards and any actions which have to be taken to avoid them. (This symbol may also be used to
refer the user to ratings
The following symbol(s) may appear on the product:
in the manual.)
viP5200A Series High Voltage Differential Probes Instruction Manual
Compliance Info
rmation
Compliance In
This section lists the safety and environmental standards with which the instrument complies. This product is intended for
use by professionals and trained personnel only; it is not designed for use in households or by children.
Questions about the following compliance information may be directed to the following address:
Tektronix, Inc.
PO Box 500, MS 19-045
Beaverton, OR 97077, USA
www.tek.com
Safety Compliance
This section lists the safety standards with which the product complies and other safety compliance information.
EU Low Vol
Compliance was demonstrated to the following specification as listed in the Official Journal of the European Union:
Low Voltage Directive 2014/35/EU.
EN 61010-031. Particular requirements for handheld probe assemblies for electrical measurement and test equipment.
tage Directive
formation
U.S. Nationally Recognized Testing Laboratory Listing
UL 6101
0-031. Particular requirements for handheld probe assemblies for electrical measurement and test equipment.
Canadian Certification
CAN/CSA-C22.2 No. 61010-031. Particular requirements for handheld probe assemblies for electrical measurement
and test equipment.
Additional Compliances
IEC 61010-031. Particular requirements for handheld probe assemblies for electrical measurement and test equipment.
ipment Type
Equ
Test and measuring equipment.
P5200A Series H igh Voltage Differential Probes Instruction Manualvii
Compliance Info
Pollution Degree Descriptions
A measure of the contaminants that could o ccur in the environment around and within a product. Typically the internal
environment inside a product is considered to be the same as the external. Products should be used only in the environment
for which they are rated.
Pollution degree 1. No pollution or only dry, nonconductive pollution occurs. Products in this category are generally
encapsulated, hermetically sealed, or located in clean rooms.
Pollution degree 2. Normally only dry, nonconductive pollution occurs. Occasionally a temporary conductivity that is
caused by condensation must be expected. This location is a typical office/home environment. Temporary condensation
occurs only when the product is out of service.
Pollution degree 3. Conductive pollution, or dry, nonconductive pollution that becomes conductive due to condensation.
These are sheltered locations where neither temperature nor humidity is controlled. The area is protected from direct
sunshine, rain, or direct wind.
Pollution degree 4. Pollution that generates persistent conductivity through conductive dust, rain, or snow. Typical
outdoor locations.
Pollution Degree Rating
Pollution degree 2 (as defined in IEC 61010-1). Rated for indoor, dry location use only.
rmation
IP Rating
IP20 (as defined in IEC 60529).
Measurement and Overvoltage Category Descriptions
Measurement terminals on this product may be rated for measuring mains voltages from one or more of the following
ories (see specific ratings marked on the product and in the manual).
categ
ory II. Circuits directly connected to the building wiring at utilization points (socket outlets and similar points).
Categ
gory III. In the building wiring and distribution system.
Cate
gory IV. At the source of the electrical supply to the building.
Cate
NOTE. Only mains power supply circuits have an overvoltage category rating. Only measurement circuits have a
measurement category rating. Other circuits within the product do not have either rating.
Mains Overvoltage Category Rating
Overvoltage category II (as defined in IEC 61010-1).
viiiP5200A Series High Voltage Differential Probes Instruction Manual
Environmental C onsiderations
This section provides information about the environmental impact of the product.
Restriction of Hazardous Substances
Complies with RoHS2 Directive 2011/65/EU.
Compliance Info
rmation
Product End
Observe the following guidelines when recycling an instrument or component:
-of-Life Handling
Equipment Recycling. Production of this equipment required the extraction and use of natural resources. The
equipment may contain substances that could be harmful to the environment or human health if improperly handled at the
s end of life. To avoid release of such substances into the environment and to reduce the use of natural resources,
product’
we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused
or recycled appropriately.
This symbol indicates that this product c omplies w ith the applicable European Union requirements according
to Directives 2012/19/EU and 2006/66/EC on waste electrical and electronic equipment (WEEE) and batteries.
rmation about recycling options, check the Tektronix Web site (www.tek.com/productrecycling).
For info
P5200A Series H igh Voltage Differential Probes Instruction Manualix
Preface
Preface
This document provides operating information and specifications for the Tektronix P5200A Series high voltage differential
probes. The probes share similar functions, properties, and operating procedures, and are discussed in the firstpartofthe
manual. The specifications and performance verification procedures for the probes follow.
WARNING. Only use the accessories that are designed for your probe and that are rated at or above the voltages you are
measuring. (See Table i on page xi.) (See Table ii on page xiv.)
NameBandwidthAttenuationOscilloscope interface
P5200A50 MHz
P5202A100 MHz
P5205A100 MHz
P5210A50 MHz
50X/500XBNC
20X/200XTekProbe BNC-Level 2
50X/500XTekProbe BNC-Level 2
100X/1000XTekProbe BNC-Level 2
xP5200A Series High Voltage Differential Probes Instruction Manual
Preface
Figure i: P5200A High Voltage Differential Probe with accessories
The P5200A, P5202A & P5205A standard accessories can also be used with the P5210A probe, but only at the reduced voltage levels listed here.
2A
P520
VCATI
450
300 V CAT II
300 V CAT II
0VCATII
30
50VCATI
4
300 V CAT II
0A & P5205A
P520
0VCATII
100
600 V CAT III
1000 V C AT II
600 V CAT III
1000 V C AT II
0 V CAT III
60
000VCATII
1
600 V CAT III
P521
230
1000 V CAT III
1000 V CAT I
1000 V CAT III
1000 V CAT I
10
1
1000 V CAT III
1
0A
0VCATI
00 V CAT III
000 V CAT I
P5200A Series H igh Voltage Differential Probes Instruction Manualxi
Preface
Figure ii: P5202A High Voltage Differential Probe with accessories
xiiP5200A Series High Voltage Differential Probes Instruction Manual
Preface
Figure iii: P5205A High Voltage Differential Probe with accessories
P5200A Series H igh Voltage Differential Probes Instruction Manualxiii
Preface
Figure iv: P5210A High Voltage Differential Probe with accessories
Table ii: P5210A probe standard accessories derating table
Combined probe and accessory common-mode voltage
put voltage-to-earth ratings
and in
ssory
Acce
Extender leads
P520
450
300 V CAT II
Test probe (TATP)450 V CAT I
300 V CAT II
Small hook tip (TASH)450 V CAT I
30
arge hook tip (TALH)
L
4
300 V CAT II
1
The P5210A standard accessories can be used with these probes at the reduced voltage levels listed in this table.
1
2A
VCATI
0VCATII
50VCATI
0A & P5205A
P520
0VCATII
100
600 V CAT II
1000 V C AT II
600 V CAT II
1000 V C AT II
0VCATII
60
000VCATII
1
600 V CAT II
1
P521
0VCATI
230
0A
1000 V CAT III
1000 V CAT I
1000 V CAT II
2300 V CAT I
00 V CAT II
10
300VCATI
2
1000 V CAT II
xivP5200A Series High Voltage Differential Probes Instruction Manual
Probe Operating
Information
Probe Operati
The P5200A Series probes share many common features, including connections to the circuit, c ompensation box buttons,
and operating basics. The probes connect to the host oscilloscope through one of two probe-to-oscilloscope interfaces:
BNC – this connection is a shielded 50 Ω coaxial cable with an outer ground connection and center signal pin. The BNC
interface is used on the P5200A probe, which allows you to connect directly to most ground-referenced oscilloscopes.
TekProbe
the probe tip. This interface is used on the P5202A, P5205A, and P5210A probes, and many Tektronix oscilloscopes.
These probes can also be used with Tektronix oscilloscopes that use the TekVPI-interface, by connecting through an
optional
TPA-BNC Adapter
Level 2 – this interface adds probe communications with the oscilloscope to accomplish calibrated offset at
TPA-BNC Adapter.
ng Information
The TPA-BNC Adapter is an optional accessory that enables you to use existing TekProbe-interface probes with oscilloscopes
that feature the TekVPI probe interface. The adapter recognizes TekProbe-interface probes and supplies the necessary
power, serial communication, and offset control as used by the connected TekProbe product accessory.
WARNING. To reduce risk of shock or fire, do not exceed the ratings of the TPA-BNC adapter; it is not intended to be
connected to voltages above 30 VAC, 42 Vpk, or 60 VDC. For BNC probes, connect the probe directly to the oscilloscope.
P5200A Series H igh Voltage Differential Probes Instruction Manual1
Probe Operating
Information
Connecting to the Instrument
P5200A Probe
The P5200A probe requires an external AC adapter to power the internal circuitry. Install the P5200A probe as follows:
1. Connect the output of the probe to the BNC input of the oscilloscope or other measurement instrument. The
measurement instrument input must be ground-referenced (not floating).
2. Connect the power cord to the AC adapter.
3. Connect the output of the AC adapter to the D C input jack located on the output lead of the probe. All of the LEDs on the
probe briefly light to confirm power-on, and then indicate the settings from the previous session.
4. Adjust the vertical offset (or position) of the measurement instrument input.
5. Select the proper range setting. For example, when using the P5200A probe, to achieve higher resolution and less noise
when measuring signals below 130 V
output signal may not be accurate. Use the 500X setting instead.
If you want the oscilloscope to display the actual probe voltage instead of a scaled value, you must match the attenuation
setting of the oscilloscope to the probe attenuation setting. Use the on-screen Probe Setup menu on the oscilloscope;
the access method varies depending on oscilloscope model.
For example, on DPO/MSO4000 series oscilloscopes, press the front-panel channel number button and then press the
More button on the lower bezel until Probe Setup is highlighted. The oscilloscope attenuation setting displays in the
Probe Setup menu. Change the attenuation by turning the multipurpose knob.
WARNING. To avoid electrical shock, observe proper safety precautions when working with voltages above 60 VDC or
30 VAC
Make sure that the accessories are fully mated before connecting or disconnecting.
To avoid electrical shock or fire, make sure the test leads are in good condition. The input leads and extender leads have a
jacket wear indicator which becomes visible if the wire jacket becomes excessively worn. If the wear indicator is visible, do
not use the probe. Contact Tektronix Service for repair or replacement.
To avoid electrical shock or fire, keep the prob
The probe body and output cable are not intended to be in contact with the circuits being measured.
. T hese voltage levels pose a shock hazard. Use only the accessories specified for the probe that you are using.
RMS
, switch the attenuation to 50X. If the overrange indicator lights or flashes, the
pk
e body and output cable of the probe away from the circuits being measured.
6. Using the appropriate probe accessories, connect the inputs of the probe to the circuit points to be m easured.
2P5200A Series High Voltage Differential Probes Instruction Manual
Probe Operating
Information
Connecting P5202A, P5205A, & P5210A Probes to Oscilloscopes that feature the VPI Interface
1. Connect the TPA-BNC adapter to the oscilloscope.
2. Connect the probe to the input of the adapter.
3. Turn the probe connector clockwise to secure it. All of the LEDs on the probe briefly light to confirm power-on, and then
indicate the settings from the previous session.
4. Adjust the vertical offset (or position) of the oscilloscope input.
NOTE. Do not attempt to adjust the offset adjustment in the probe compensation box; it is factory-preset to optimize the
range of the electronic adjustment. To make the offset adjustment, refer to the procedures. (See page 46, Offset Zero.)
5. Select the proper range setting. For example, when using the P5202A probe, to achieve higher resolution and less
noise when measuring signals below 64 V
, switch the attenuation to 20X. If the overrange indicator lights or flashes,
pk
the output signal may not be accurate. Use the 200X range setting instead.
WARNING. To avoid electrical shock, observe proper safety precautions when working with voltages above 60 VDC or
30 VAC
. These voltage levels pose a shock hazard. Use only the accessories specified with the probe that you are using.
RMS
Make sure that the accessories are fully mated before connecting or disconnecting.
To avoid electrical shock or fire, make sure the test leads are in good condition. The input leads and extender leads have a
jacket wear indicator which becomes visible if the wire jacket becomes excessively worn. If the wear indicator is visible, do
not use the probe. Contact Tektronix Service for repair or replacement.
To avoid electrical shock or fire, keep the probe body and output cable of the probe away from the circuits being measured.
The probe body and output cable are not intended to be in contact with the circuits being measured.
6. Using the appropriate probe accessories, connect the inputs of the probe to the circuit points to be measured.
P5200A Series H igh Voltage Differential Probes Instruction Manual3
Probe Operating
Disconnecting from the Instrument
Information
WARNING. To av
the instrument.
oid electrical shock, disconnect the probe inputs from the circuit before disconnecting the probe from
P5200A
1.Turn the probe connector counterclockwise.
2. Pull straight out.
P5202A, P5205A, P5210A
1. Turn the probe co
counterclockwise.
2. Pull straight out.
3. Press the latch
4. Pull straight out.
nnector
button on the adapter.
4P5200A Series High Voltage Differential Probes Instruction Manual
Probe Controls
The P5200A Series probes have several features that make probing and measurement a simpler task. Familiarize yourself
with the controls shown on the following pages. The attenuation ranges differ between probe models.
Overrange Indicator
The overrange indicator lights red if the
voltage of the input signal exceeds the
linear range of the range setting. When this
happens, the signal on the probe output
does not accurately represent the signal on
the probe input.
WARNING. The Overrange indicator
does not detect overrange condition of
common-mode voltages or voltage-to-earth
potential at the probe inputs. The O verrange
indicator only detects differentially between
the + and – inputs (not relative to ground).
Do not exceed the common-mode voltage
or input voltage-to-earth ratings of the probe
when taking measurements. (See page 27,
Overrange Detection.)
Probe Operating
Information
If you are not sure, make a single-ended
measurement of each point you are intending
to measure differentially first. Make a
single-ended measurement by tying one
input lead to ground (the "–" input) and then
connecting the other lead (the “+" input) to
the points of interest, one at a time.
Attenuation Button and Indicators
Press the button to select between the
voltage range (attenuation) settings of the
probe. The range is indicated by two LEDs
on the probe and may be displayed on
the oscilloscope screen, depending on the
oscilloscope model.
P5200A models only: To display the actual
probe voltage instead of a scaled value, you
must match the attenuation setting of the
oscilloscope to the probe attenuation setting.
Use the on-screen Probe Setup menu on the
oscilloscope to change the setting.
The Overrange LED lights if the applied
voltage exceeds the selected range. To
extinguish the LED, select a higher range.
If a higher range is not available, do not
attempt to take the measurement.
P5200A Series H igh Voltage Differential Probes Instruction Manual5
Probe Operating
Bandwidth Limit Button and
Indicators
Press the button to limit the probe bandwidth
to 5 MHz. 5 MHz is close to the switching
frequency of most switching transistors
(FETs) in switch mode power supplies
(SMPS).
The 5 MHz filter assists in the characterization
and testing of power supplies in switch mode
by removing all high frequency content, noise
and harmonics from the measurement.
Press the button again to return to the Full
position, which selects the full specified
bandwidth of the probe.
Audible Overrange On/Off Button
and Indicators
Information
Press the
enable an audible alarm that indicates when
the measured signal exceeds the selected
range. P
OFF LED and disable the audible feature.
button to light the ON LED and
ress the button again to light the
6P5200A Series High Voltage Differential Probes Instruction Manual
Functional Check
Using accessories that are shipped with your P5200A Series probe and a source that supplies AC line voltage, perform the
following procedure.
WARNING. To reduce risk of shock or fire, ensure that the accessories are fully mated before you connect to voltage
sources above 42 Vpk.
1. Connect the output of the probe to the oscilloscope input channel.
2. Connect the probe inputs to the A C voltage source.
3. Set the probe attenuation range to the highest setting and perform the check as each row of the following table indicates.
Probe Operating
Information
Figure 1: Funct
Input 1
(+ or –)
Hot
ional check setup
Input 2
ModeRange setting
(– or +)
Ground or NeutralDifferentialHigh (1000X,
500X, or 200X)
Hot
Ground or NeutralDifferentialLow (100X,
50X, or 20X)
Hot
1
If a DC offset voltage is present, zero the DC offset. (See page 46, Offset Zero.)
Hot (same
connection)
Common
Mode
High or lowNo signal
Check
Measurement instrument displays or indicates
the line voltage
Measurement instrument displays or indicates
the line voltage. Overrange indicator lights if
the input is ~20% over
1
P5200A Series H igh Voltage Differential Probes Instruction Manual7
Accessories
Accessories
The P5200A Series probes include standard accessories that make connecting to your circuit an easier task. Other
accessories are available and are described in the Optional Accessories section. (See page 15, P5200A Series ProbesOptional Accessories.)
Connecting to the Circuit
Make the connections to your circuit using the integral input leads or the accessories that best fit your application.
WARNING. To reduce risk of shock or fire, do not exceed either the voltage rating or category ratings of the probe or the
probe acces
sory, w hichever is the lesser of the two. Use only the accessories provided with the probe.
To avoid el
away from the shaded area shown in the accessory illustrations below.
To avoid electrical shock or fire, make sure the test leads are in good condition. The input leads and extender leads have a
jacket wear indicator which becomes visible if the wire jacket becomes excessively worn. If the wear indicator is visible, do
not use th
To avoid e
The probe body and output cable are not intended to be in contact with the circuits being measured.
Integra
The integral input leads extend ~9 in (0.23 m)
from the probe body and have shrouded
male ban
directly to your circuit, or use the extender
leads and the other accessories shown on
the fol
ectric shock when using the probe or accessories, keep your fingers behind the fi nger guard of the probe body and
e probe. Contact Tektronix Service for repair or replacement.
lectrical shock or fire, keep the probe body and output cable of the probe away from the circuits being measured.
l Input Leads
ana plugs. Connect the leads
lowing pages.
8P5200A Series High Voltage Differential Probes Instruction Manual
P5200A, P5202A, & P5205A Probe Standard Accesso ries
Extender Leads
These c ables e
by ~67 in (1.5 m). The banana ends connect
to all of the clip accessories that are included
with the probe
One pair of extender leads are included with
the probes.
Maximum rati
2300 V CAT I *
1000 V C AT III
*SeeSpecifi
Transient (OVT) rating for the probe that you
are using.
Reorder Te
196-3523-xx (one pair)
Hook Clips (AC280-FL)
Plug the probe test leads into the banana
plug conn
expose the hook clip and then clasp it around
the circuit test point.
Maximum
1000 V C AT III
600 V C AT IV
One pai
probes.
Reorder Tektronix part number:
AC280-
xtend the reach of the probes
s.
ngs:
cations for the Over-Voltage
ktronix part number:
ectors. Squeeze the grips to
ratings:
r of hook clips is included with the
FL (one pair)
Accessories
P5200A Series H igh Voltage Differential Probes Instruction Manual9
Accessories
Pincer Clips (AC283-FL)
The plunger probes have long probe sleeves
with retracti
connect to recessed test points that are
otherwise difficult to reach.
Maximum rati
1000 V CAT III
600 V CAT IV
One pair of p
probes.
Reorder Tektronix part number:
AC283-FL (o
ng hooks. These probes safely
ngs:
incer clips is included with the
ne pair)
Alligator
These large insulated alligator clips connect
to many circuit components.
Maximum r
1000 V CAT III
600 V CAT IV
One pair
Reorder Tektronix part number:
AC285-FL (one pair)
Clips (AC285-FL)
atings:
of clips is included with the probes.
10P5200A Series High Voltage Differential Probes Instruction Manual
P5210A Probe Standard Accessories
Accessories
WARNING. To av
CAT IV circuits. Refer to the ratings tables in the beginning of the manual. (See Table i on page xi.) (See Table ii on page xiv.)
To avoid risk of electric shock or fire, when using the P5210A test probe or hook tip accessories with the P5200A, P5202A
and P5205A probes, do not use on circuits above 1000 V.
Use only a ccessories that are rated for the application. Substitution of other accessories may create a shock or burn hazard.
Keep the prob
Extender Leads
These leads
by ~67 in (1.5 m), which allow you to reach
connections as far as 3 m apart. Be sure to
use both ex
leads are the same length.
However, with longer lead length, differential
noise ind
Also, because of the added inductance of the
leads, voltage measurements at frequencies
above app
as precise. For best performance, use the
20 MHz or lower-bandwidth filter on your
oscillo
The male banana-plug ends connect to the
test probes that are included with the probes.
Maximum
1000 V C AT III
600 V C AT IV
Reorde
(one pair)
uced into the input leads is greater.
scope.
r Tektronix part number: 196-3523-xx
oid risk of electric shock or fire, do not use the P5210A test probe or hook tip accessories on CAT III or
e body and accessories clean to reduce the risk of shock due to surface conduction.
extend the reach of the probes
tension leads so that the input
roximately 10 MHz may not be
ratings:
A Accessory Kit
P5210
Kit includes one pair of each of the
accessories s hown on the following pages:
Test Probes (TATP)
Small Hook Tips (TASH)
Large Hook Tips (TALH)
Reorder Tektronix part number: 020-3070-xx
P5200A Series H igh Voltage Differential Probes Instruction Manual11
Accessories
Test Probes (TATP)
Use the test probes to browse multiple test
points or to co
hook tips.
1. The test probe tip is a 6-32 threaded post
nnect the test leads to the
that accepts the large and small hook
tips provide
d with the probe.
2. The finger gua
when the hook tips are not being used.
Keep your fingers behind the finger
guard whene
risk of a shock from the circuit under test.
3. Connect the back end of the test probe
to the input test leads of the probe.
Maximum ra
2300 V CAT I*
1000 V CAT II
* See Spec
Transient (OVT) rating for the probe that you
are using.
WARNING
caution when probing circuits with raised
components. Avoid getting the metal shell
n components of different potentials.
betwee
Use TASH for probing in hard to reach areas.
WARNING. To prevent arc flash, do not use
st probe or hook tips on CAT III circuits.
the te
To probe CAT III circuits, use the AC280-FL,
AC283-FL, or AC285-FL.
rd provides protection
ver possible to reduce the
tings:
ifications for the Over-Voltage
. To prevent arc flash, use
12P5200A Series High Voltage Differential Probes Instruction Manual
Small Hook Tip (TASH)
Use the small hook tip for making
connections t
component leads.
Screw the small hook tip onto the TATP test
probe. To use
body and pull the tip shield back. Hook the
tip onto the circuit and release the shield.
o small conductors such as
the hook tip, hold the probe
Accessories
WARNING. To
when measuring voltages above 1000 V,
always keep your fingers behind the tactile
indicator.
Maximum ratings:
2300 V CAT I
1000 V C AT II
*SeeSpecifications for the Over-Voltage
Transien
are using.
Large Hook Tip (TALH)
Use the large hook tip when working with
omponents such as bolt terminals
larger c
and bus bars typically found in power
distribution equipment.
he large hook tip onto the TATP test
Screw t
probe and then clamp the hook tip onto the
circuit.
NG. To reduce the risk of shock
WARNI
when measuring voltages above 1000 V,
always keep your fingers behind the tactile
ator.
indic
reduce the risk of shock
*
t (OVT) rating for the probe that you
Maximum ratings:
VCATI*
2300
1000 V C AT II
*SeeSpecifications for the Over-Voltage
nsient (OVT) rating for the probe that you
Tra
are using.
P5200A Series H igh Voltage Differential Probes Instruction Manual13
Accessories
P5200A Probe Power Supply and Power Cord Options
The P5200A Probe requires an external DC power supply, which is included with the probe. Power cord options are available
for international locations and are listed in the table below.
Table 1: P5
ItemDescript
1
2
200A power supply and power cord options
POWER SUP
IN;9 VDC 2 A, CABLE WITH BARREL CONNECTOR OUT;SAFETY CONTROLLED
OPTION A0: CABLE ASSY PWR; 3,18 AWG, 250V/10A, 98.0 L, STR, IEC320,
RCPT X NEMA 5-15P, US, SAFETY CONTROLLED,
A2: CABLE ASSY, PWR; 3,1.0 MM SQ, 250V/10A, 2.5 METER, S TR,
Tektroni
119-7758-xx
161-006
161-0066-09
161-0066-10
161-0066-13
161-0154-00
161
161-0304-00
161-0400-00
1
x p art number
6-00
-0298-00
61-0357-00
14P5200A Series High Voltage Differential Probes Instruction Manual
P5200A Series Probes Optional Accessories
Accessories
WARNING. To re
probe accessory, whichever is the lesser of the two. Use only the accessories provided with the probe or the optional
accessories shown below.
To avoid electric shock when using the probe or accessories, keep your fingers behind the finger guard of the probe body and
away from the
To avoid elec
jacket wear indicator which becomes visible if the wire jacket becomes excessively worn. If the wear indicator is visible, do
not use the probe. Contact Tektronix Service for repair or replacement.
To avoid electrical shock or fire, keep the probe body and output cable of the probe away from the circuits being measured.
The probe bo
TPA-BNC Ad
This optional accessory enables you to use
the P5202A, P5205A, and P5210A probes
with osci
probe interface.
The TPA-BNC Adapter allows the
oscillos
power and communication to these
TekProbe-interface probes.
NOTE. Th
directly to the BNC connector on the TekVPI
interface, so they do not require this adapter.
lloscopes that feature the TekVPI
cope to recognize and provide
duce risk of shock or fire, do not exceed either the voltage rating or category ratings of the probe or the
shaded area shown in the accessory illustrations below.
trical shock or fire, make sure the test leads are in good condition. The input leads and extender leads have a
dy and output cable are not intended to be in contact with the circuits being measured.
apter
e P5200A probes connect
WARNIN
do not exceed the ratings of the TPA-BNC
adapter. Do not connect the adapter to
volta
G. To reduce risk of shock or fi re,
ges above 30 VAC, 42 Vpk, or 60 VDC.
P5200A Series H igh Voltage Differential Probes Instruction Manual15
Accessories
Handheld Probes (TP175-FL)
These probes plug onto the banana input
leads and exte
threaded to accept tip accessories that are
included with the probe.
The insulator
and retracts into CAT III and C AT IV-rated
spacings. Twist the probe body past the
detent at eac
probe into the CAT setting.
WARNING. Always verify that the probe
body is lock
measurements. Do not use in the unlocked
neutral position.
Ratings:
1000 V CAT I
1000 V CAT III
10 A
Order Tek
TP175-FL (one pair)
nder leads. The tips are
sheath at the probe tip extends
h end of the twist to lock the
ed into position before taking
I
tronix part number:
16P5200A Series High Voltage Differential Probes Instruction Manual
Pogo Pin Tip Adapters & Tips
These insulated adapters hold pogo pins and
screw on to the
handheld probes.
WARNING. To prevent electrical shock,
tighten the p
the TP175-FL probe.
Two pairs of pogo pin types are included
with the adapters; one pair have sharp, cone
points and t
for embedding in soft conductors.
WARNING. The pogo pins have very s harp
points. To
carefully when you install and remove them.
WARNING. To prevent risk of arc flash,
ensure th
inserted into the adapter. Verify that the
exposed metal portion of the tip is 19 mm
(0.75 in
threaded tips of the TP175-FL
ogo pin tip adapter completely to
he other pair have serrated edges
prevent injury, handle the pins
at the pogo pin is completely
) or less.
Accessories
WARNING. The probe input rating is derated
to 150 V CAT II, 0.1 A, when used with the
THDP an
this pogo pin adapter to measure voltages
that exceed this rating.
Maxim
150 V CAT II
0.1 A
Reor
020-3107-xx (includes 2 Tip Adapters,
2 C one-Tip Pogo Pins, &
2 Ser
d TMDP series probes. Do not use
um ratings:
der Tektronix part number:
rated-Tip Pogo Pins)
P5200A Series H igh Voltage Differential Probes Instruction Manual17
Accessories
Extended Test P
These adapters screw on to the threaded tips
of the handheld probes.
Use these adap
circuitry. The sharp tips can contact small
component leads and circuit board features.
WARNING. The
to 300 V CAT II, 1 mA, when used with the
THDP and TMDP series probes. Do not
use this ext
voltages that exceed this rating.
WARNING. The tip on this adapter is very
sharp. To p
Maximum ratings:
300 V C AT II
3A
Order Tektronix part number:
012-1724-xx (one pair)
Crocodile Clips
The crocodile cl
bolts or bus bars. The connectors are double
insulated for safety. The clips screw on to the
threaded tips o
Maximum ratings:
1000 V CAT III
10 A
One pair of clips is included with the probes.
Order Tektronix part number:
344-0670-xx
robe Adapters
ters to reach into dense
probe input rating is derated
ended probe adapter to measure
revent injury, do not touch the tip.
ips connect easily to large
f the handheld probes.
(one pair)
18P5200A Series High Voltage Differential Probes Instruction Manual
Table 2: Voltage derating for P5200A Series probes optional accessories
Accessories
Combined probe
and input voltage-to-earth ratings
and accessory common-mode voltage
1,2
AccessoryP5202AP5200A & P5205AP5210A
3
Handheld pro
(TP175-FL)
Pogo pin tip adapters with either style tips
bes
450VCATI
300 V CAT II
1000 V CAT II
600 V CAT III
1000 V CAT I
1000 V CAT III
150 V CAT II150 V CAT II150 V CAT II
(020-3107-xx)
Extended test probe adapters
(012-1724-xx)
Crocodile
clips
(344-0670-xx)
1
The operating altitude of the probe is derated to 2000 m (6560 ft) when used with these accessories.
2
3
ge rating and CAT rating are derated to the voltage in this table when used with these accessories.
The volta
When using the TP175-FL test probes in CAT III circuits, the tip must be in the retracted position to prevent risk of arc flash. The exposed metal tip
is about 3.7 mm (0.15 in) in the retracted position.
300VCATI
300 V CAT II
450VCATI
300 V CAT II
300 V CAT II300 V CAT I
300 V CAT II
1000 V CAT I
600 V CAT III
I
1000 V CAT I
1000 V CAT III
P5200A Series H igh Voltage Differential Probes Instruction Manual19
Accessories
TPH1000 Probe H
The TPH1000 probe holder allows you to
make a hands-free connection when using
the optional h
The probe holder can also be used with
many other Tektronix probes.
You have two op
differential m easurements:
You can use the handheld probes with
two TPH1000 probe holders (required if
the test poin
For test poi
handheld probes with the optional
THV-Browser (shown on the following
page).
To use the probe holder, do the following:
1. Insert the probe into one of the holder
openings
the circuit under test.
2. Slide the probe forward to secure it.
older
andheld probes (TP175-FL).
tions for taking hands-free
ts are >1 inch apart).
nts <1 inch apart, use the
so that the Tektronix logo faces
CAUTION. To avoid personal injury, always
nd remove the probe by gripping the
insert a
handheld section of the probe.
3. Position the base of the probe holder on
your circuit where it can maintain stability
ontacting the test point.
while c
The weight of the probe holder keeps the
probe in place.
NG. Do not use the probe holder
WARNI
without the rubber feet; internal metal would
be exposed which presents a shock hazard.
ION. If you are probing circuitry
CAUT
with dense contacts such as IC pins,
Tektronix recommends that you use insulated
be tip accessories designed to prevent
pro
short-circuiting adjacent IC pins or circuitry.
Order Tektronix part number:
H1000
TP
20P5200A Series High Voltage Differential Probes Instruction Manual
THV-Browser
The THV-Browser allows y o u to
set and lock the spacing between
two handheld p
browse your circuit with one hand.
robe tips, and then
Handheld Browsing.
1. Place each TP175-FL
handheld probe into the
cavity and then slide the probe
forward to lock it into place.
2. Loosen the thumb screw and
adjust the spacing between
the probe tips. Graticules
near the thumb screw indicate
the spacing. The maximum
spacingis~1in(2.54mm).
3. Tighten the thumb screw.
You can now browse your circuit.
Accessories
WARNING. To avoid injury or
short circuits, do not drop the
THV-Browser on high voltage
circuitry. The browser contains
metal components.
Hands-free Probing.
If you want a stationary, hands-free
connection, attach the browser to
the TPH1000 probe holder:
4. Align the slots on the top of the
probe holder with the pins on
the bottom of the browser.
5. Rotate the browser 90° and
making sure that the Tektronix
logo on the probe holder faces
the circuit under test.
6. Position the probe tips on your
test points so that you can set
the holder on a stable surface.
Order Tektronix part number:
THV-Browser
P5200A Series H igh Voltage Differential Probes Instruction Manual21
Accessories
Tektronix 1103
Supply.
for performing service procedures on the
P5202A, P5205A
The 1103 power supply provides power to
the probe and routes the probe output signal
through a BNC c
of the supply.
Order Tektronix part number:
1103
Probe Power
Use the 1103 power supply
, and P5210A probes.
onnector on the front panel
Replacement Label (Safety Item;
Service Only).
the openings to the service-only adjustments
on the back of the probe. To maintain
the safety of the probe, the label must be
replaced after service adjustments are made
to the p robe.
If the original label becomes damaged or
lost, order a replacement label.
NOTE. This label is only replaceable on
units with serial numbers C020000 and
above
This reusable label covers
Order Tektronix part number:
335-2913-xx
22P5200A Series High Voltage Differential Probes Instruction Manual
Options
Service Options
Option C3. Calibration Service 3 years
Option C5. Calibration Service 5 years
Option D1. Calibration Data Report
Option D3. Calibration Data Report, 3 years (with Option C3)
Option D5. Calibration Data Report, 5 years (with Option C5)
Option R3. Repair Service 3 years
Option R5. Repair Service 5 years
Options
P5200A Series H igh Voltage Differential Probes Instruction Manual23
Operating Basic
s
Operating Bas
To help you use the P5200A Series High Voltage Differential Probes safely and effectively, this section provides important
information about safety limits, operating characteristics, and probing techniques.
ics
Operating Characteristics and Probing Techniques
This section explains the operating characteristics of the P5200A Series High Voltage Differential Probes and includes
techniques
Operating Limits
The probes have two operating ranges that you select with the ATTEN button. These ranges set the maximum differential
voltage that can be measured. The ranges and voltage limits differ between probe models. (See Table 3.)
that you can use to maximize the performance of the probe.
Low attenuation range
(20X/50X/100X)
130 V>140 V1300 V>1400 V
64 V>70 V640 V>700 V
130 V>140 V1300 V>1400 V
560 V>600 V5600 V>6000 V
High attenuation range
(200X/500X/1000X)
The input signals that you attempt to measure must be considered both for the differential potential between each other
and for the amplitude on each input with respect to ground (the common mode voltage specification). The maximum
common mode voltage limits vary between probes, from 450 V for the P5202A, to 2300 V for the P5210A probe. You
should consider both specifications when choosing a probe for your measurement task. Some examples that illustrate this
are shown on the following pages.
24P5200A Series High Voltage Differential Probes Instruction Manual
Operating Basic
Measurement Examples
Example 1. Consider a case where you need to measure two sinusoidal waveforms that are 180° out of phase with each
other, each with an amplitude of 1000 V
with no DC offset (centered at 0 V). (See Figure 2.)
pk
s
Figure 2: Measuring two equal-amplitude waveforms that are 180 degrees out of phase
If both w aveforms are at the same voltage potential, then the differential measurement would be 2 times the i ndividual signal
inputs (in this example, 2000 V
probes, the P5210A probe is capable of measuring this signal. (See Table 6 on page 29.) For reference, the rms
Series
). Looking at the maximum measurable differential voltage specifications for the P5200A
pk
values of the Common-Mode Voltage and Voltage-to-Earth ratings and Maximum Input Signals for each probe model are
shown in the figure above.
P5200A Series H igh Voltage Differential Probes Instruction Manual25
Operating Basic
Example 2. Next, assume that the s ame waveforms from the previous example are 120° out of phase with each other.
(See Figure 3.)
Although this is a lower potential between the inputs than in example 1, it still exceeds the differential rating (1300 V
P5200A and P5205A probes, so you must use the P5210A.
s
This phase relationship yields a maximum differential of 1.732 times the individual signal inputs, or 1732 V
)ofthe
pk
.
pk
Figure 3: Measuring two equal-amplitude waveforms that are 120 degrees out of phase
e examples with equal-amplitude signals on the inputs, the P5200A and P5205A probes can measure up to 1300 V/2
In thes
= 650 V
(460 V
pk
) on each input when the signals are 180° apart, and 1300 V/1.732 = 750 Vpk(530 V
rms
) when 120° apart.
rms
When you are taking these types of measurements, to prevent exceeding the maximum differential input voltage for the
fic probe model that you are using, refer to the respective figures for the maximum input signal voltages (V
speci
rms
).
Example 3. Your task is to measure two AC waveforms of the same phase, each with an amplitude of 300 V. However, one
waveform is centered on ground (– input), and the other is centered on an offset of 400 VDC (+ input). The common m ode
voltage is the 300 V
taken into account for both inputs. The voltage-to-earth is 300 V
is 700 V
(the 300 VAC
rms
of the P5202A probe, so it cannot be used for taking this measurement. In this case, you must use another probe; the
next closest being either the P5200A or P5205A probe.
, but the maximum voltage-to-earth (the common mode voltage plus the signal waveform) must also be
rms
on the (– input), but on the (+ input), the voltage-to-earth
rms
plus the 400 VDC
rms
). Thus the (+ input) exceeds the maximum input voltage-to-earth rating
rms
26P5200A Series High Voltage Differential Probes Instruction Manual
Operating Basic
Overrange Detection
Differential voltage outside the operating range will overdrive the circuitry of the probe and distort the output signal. When
this differential overrange occurs, the probe detects the condition and lights the overrange indicator. With the Audible
Overrange ON, the probe will also emit an audible alarm.
WARNING. The Overrange indicator does not detect an overrange condition of common-mode voltages or voltage-to-earth
potential at the probe inputs. The Overrange indicator only detects differentially between the + and – inputs (not relative to
ground). Do not exceed the Common-Mode Voltage or Input Voltage-to-Earth ratings of the probe when taking measurements.
If you are not sure, first take a single-ended measurement of each point that you are intending to measure differentially.
Take a single-ended measurement by tying one input lead to ground (the – input) and then connecting the other lead (the +
input) to the points of interest, one at a time.
Common-Mode Rejection
The common-mode rejection ratio (CMRR) is the specified ability of a probe to reject signals that are common to both inputs.
More precisely, CMRR is the ratio of the differential gain to the common-mode gain. The higher the ratio, the greater the
f probe to reject common-mode signals.
ability o
s
Common mo
500 V
50 mV
Twistin
Twisting the input leads helps to cancel noise from high-EMI environments that is induced into the input leads.
de rejection decreases as the input frequency increases. For example, if you apply a 60 Hz line voltage of
to both input leads of the probe, the probe rejects the signal by 80 dB (typical) and the signal appears as only a
p-p
signal on the oscilloscope screen.
p-p
g the Input Leads
Probe Loading
When you touch your probe tip to a circuit element, you are introducing a new resistance, capacitance, and inductance into
e circuit. Frequency and impedance of the source determine how much the probe loads the circuit you are measuring. As
th
the frequency of the source starts to increase beyond 1 kHz, the input impedance of the probe begins to decrease.
The lower the impedance of the probe relative to that of the source, the more the probe loads the circuit under test. For a
graph of frequency versus input impedance, refer to the Specifications section. As the graph shows, the probes have virtually
loading effect on sources with relatively low impedance and low frequency.
no
P5200A Series H igh Voltage Differential Probes Instruction Manual27
Specifications
Specification
The specifications shown apply to the P5200A Series probes installed on Tektronix MSO/DSO4000 oscilloscopes. When a
probe is used with another oscilloscope, the oscilloscope must have an input impedance of 1 MΩ and a bandwidth equal
to or greater than the probe. The probe must have a warm-up period of at least 20 minutes and be in an environment
that does not exceed the limits described. (See Table 5.) The probe calibration should be run on the host instrument
before verifying the warranted probe specifications. Specifications for the P5200A Series probes fall into three categories:
warranted, typical, and nominal characteristics.
s
Warranted Specifications
Warranted c
(See Table 4.)
Table 4: Wa
SpecificationP5200AP5202AP5205AP5210A
Rise time
(small signal, 10–90%,
+20°Cto+30°C)
Gain accuracy±2%
1
haracteristics describe guaranteed performance within tolerance limits or certain type-tested requirements.
rranted electrical specifications
≤7.0 ns
Output may be slew rate limited for large amplitude signals.
1
≤3.5 ns
≥240 V/ns (200X))
(slew rate
≤3.5 ns1(slew rate
≥590 V/ns (500X))
≤7.0 ns
Table 5: Warranted environmental specifications
Specification
Temperature
Operating0 °C to 40 °C (32 °F to +104 °F)
Nonoperating
Humidity
Operating5 to 85% RH (Relative Humidity) 0 °C to +40 °C (32 °F to +104 °F)
Nonoperating
Altitude
Operating3,000 m (10,000 ft)
Nonoperating
P5200AP5202AP5205AP5210A
–30° C to +70° C (–22 °F to +158 °F)
5% to 85% RH at up to +40° C (+104 °F)
5% to 45% RH above +40° C up to +70° C (+104 to +158 °F)
up to 15,240 m (50,000 ft)
28P5200A Series High Voltage Differential Probes Instruction Manual
Typical Specifications
Typical specifications describe typical but not guaranteed performance.
Table 6: Typical electrical specifications
Specifications
Specification
Maximum measurable
differential voltage
rms
1
)
(DC + Peak AC)
(V
Maximum common mode
voltage (DC + Peak AC)
Maximum input
voltage-to-earth (V
rms
)
CAT I Maximum R ated
P5200AP5202AP5205AP5210A
3
3
50X: ±130 V
500X: ±1300 V
50X: 92 V
500X: 920 V
rms
rms
50X: ±130 V
500X: ±1300 V
1000 V CAT II
600 V CAT III
20X: ±64 V
200X: ±640 V
20X: 45 V
rms
200X: 450 V
20X: ±64 V
200X: ±640 V
300 V CAT II
450 V CAT I
rms
50X: ±130 V
500X: ±1300 V
50X: 92 V
500X: 920 V
rms
rms
50X: ±130 V
500X: ±1300 V
1000 V CAT II
600 V CAT III
NA1760 VN A2250 V
100X: ±560 V
1000X: ±5600 V
100X: 396 V
rms
1000X: 2650 V
100X: ±320 V
1000X: ±3200 V
1000 V CAT III
2300 V CAT I
Overvoltage Transient
(OVT) (V
1
2
3
4
4
)
Pk
This is the maximum measurable range between the (+) and (-) inputs of the probe. Beyond these limits, the output could be clipped. (See Figure 4.)
This rating assumes that the common mode voltage and input voltage-to-earth ratings are not exceeded.
The maximum common-mode and input voltage-to-earth ratings are the maximum amount that each input lead (+/-) can be from ground.
Applies to CAT I ratings only. The OVT peak is typically measured on top of the Peak Working Voltage.
rms
2
Figure 4: Specification table footnotes referring to the probe input limits shown on each probe label
P5200A Series H igh Voltage Differential Probes Instruction Manual29
DC to 50 MHzDC to 100 MHzDC to 100 MHzDC to 50 MHz
±10 mV
(50X & 500X)
output referred
10 MΩ ±2%5 MΩ ±2%10 MΩ ±2%40 MΩ ±2%
5MΩ ±2%2.5 MΩ ±2%5 MΩ ±2%20 MΩ ±2%
<4.0 pF per side<4.0 pF per side<4.0 pF per side<5.0 pF per side
DC: >80 dB
100 kHz:
3.2 MHz: >30 dB
50 MHz: >26 dB
<20 ns to 10% of
final
5X overdrive (1/50
gain)
9VDC
750 mA Max
00 - 240 VAC
1
9VDC/2A
>60 dB
value after
±200 mV (20X)
±2 V (200X)
input referred
DC: >80 dB
100 kHz:
3.2 MHz: >30 dB
100 MHz: >26 dB
<20 ns to 10% of
final
5X overdrive (1/20
gain)
———
———
———
>60 dB
value after
±500 mV (50X)
±5 V (500X)
input referred
DC: >80 dB
100 kHz:
3.2 MHz: >30 dB
100 MHz: >26 dB
50 μV/ °C
<20 ns to 10% of
final
5X overdrive (1/50
gain)
>60 dB
value after
±1 V (100X)
±10 V (1000X)
input referred
DC: >80 dB
100 kHz:
3.2 MHz: >30 dB
50 MHz: >26 dB
<30 ns to 10% of
value after 5X
final
overdrive (1/100
gain)
>60 dB
30P5200A Series High Voltage Differential Probes Instruction Manual
Table 8: Typical mechanical specifications
Specifications
Specification
Probe body dimensions
Input cable length
Output cable
Weight (pro
length
be only)
Nominal Specifications
Nominal specifications describe guaranteed traits, but the traits do not have tolerance limits.
Table 9: Nominal electrical specifications
Specification
Number of inputsDifferential (two inputs, + and – )
Input coupling
Output couplingDC coupling
Output terminationTerminate into 1 MΩ
Attenuation
Differential overvoltage
detection level
Input referred noise (V
1
The O
common mode voltage or input voltage-to-earth ratings of the probe are not exceeded, the test points can be measured relative to ground by probing
each separately with the + input lead while the – input lead is grounded (by taking a single-ended measurement).
1
)
rms
verrange/overvoltage indicator does not detect common mode voltage or voltage-to-earth potential at the probe inputs. To ensure that the
P5200AP5202AP5205AP5210A
185 mm x 56 mm x 2
5mm(7.3inx2.2inx1.0in)
22.9 cm (9 in)
1.5m(59in)1.5m(59in)1.5m(59in)1.5m(60in)
295 gm (9.4 o
z)
323 gm (11.4
oz)
323 gm (11.4
oz)
340 gm (12.0
P5200AP5202AP5205AP5210A
DC only
50X/500X20X/200X50X/500X100X/1000X
50X: >140 V
500X: >1400 V
50X: <40 mV
500X: <120 mV
20X: >70 V
200X: >700 V
20X: <25 mV
200X: <60 mV
50X: >140 V
500X: >1400 V
50X: <45 mV
500X: <130 mV
100X: >600 V
1000X: >6000 V
100X: <150 mV
1000X: <350 mV
oz)
P5200A Series H igh Voltage Differential Probes Instruction Manual31
Specifications
Performance Graphs
Figure 5: P5200A Series impedance plots
Figure 6: P5200A and P5205A voltage derating curves
32P5200A Series High Voltage Differential Probes Instruction Manual
Specifications
Figure 7: P
Figure 8: P5210A voltage derating curve
5202A voltage derating curve
P5200A Series H igh Voltage Differential Probes Instruction Manual33
Specifications
Figure 9: P5200A rise time (typical)
Figure 10: P5202A rise time (typical)
34P5200A Series High Voltage Differential Probes Instruction Manual
Figure 11: P5205A rise time (typical)
Specifications
Figure 12: P5210A rise time (typical)
P5200A Series H igh Voltage Differential Probes Instruction Manual35
Performance Ver
ification
Performance V
Use the following procedures to verify the warranted specifications of the P5200A Series probes. Before beginning these
procedures, photocopy the test record and use it to record the performance test results. (See Table 13 on page 41.) The
recommended calibration interval is one year.
These procedures test the following specifications:
Gain accuracy
Rise time
erification
Required Equipment
The equipm
quantities of connectors may vary depending on the specific equipment you use.
P5200A Series H igh Voltage Differential Probes Instruction Manual41
Adjustments
Adjustments
Use the following procedures to make adjustments to the P5200A Series probes. (For probes with serial numbers C019999
and below, see note and table that follow.) These procedures describe how to make adjustments to the specifications
listed below.
NOTE. Only probes with serial numbers C020000 and above have internal adjustments. (See Table 14.) Probes with serial
numbers C019999 and below, that require adjustments (other than offset zero), must be returned to Tektronix for service.
Table 14: P5200A Series probe adjustments
SpecificationAdjustment m ethod usedProbe serial number
Offset zero
Gain accuracyInternal; adjustments on PCBSerial numbers C020000 and above
DC CMRRInternal; adjustments on PCBSerial numbers C020000 and above
LF compensation
AC CMRRInternal; adjustments on PCBSerial numbers C020000 and above
External; user probe controls and comp
box adjustment
Internal; adjustments on PCBSerial numbers C020000 and above
1
All serial numbers
1
P5200A probe does not require comp box adjustment
NOTE. T
procedures to check the probe characteristics and optimize them if necessary.
he adjustments in the probes are preset at the factory for best overall performance. However, you m ay follow these
42P5200A Series High Voltage Differential Probes Instruction Manual
Adjustments
Figure 13: P5200A Series probe serial n um ber and adjustment locations
P5200A Series H igh Voltage Differential Probes Instruction Manual43
Adjustments
Equipment Required
The equipment required to perform the adjustment procedures are shown in the table below. The types and quantities of
connectors may vary depending on the specific equipment you use.
Table 15: Equipment required for adjustments
DescriptionMinimum requirem entsExample product
Oscilloscope
Generator
TekProbe power supply
Digital Multimeter (DMM)100 mV and 1 V true RMS AC ranges, < ±0.3 %
CableCoax, BNC, 50 Ω,36in
Adapter
Adapter
Adapter
Probe hook tips (2)
Adapter
Adjustment tool
Adjustment tool
Replacement rear-panel
label
1
2
3
4
1
2
34
Not required for P5200A probe
Required for the CMRR adjustment
original label is backed with a reusable adhesive. If the label does not sufficiently adhere to the probe, order a replacement.
The
Label removal is not required to access offset zero adjustments
500 MHz
Tektronix MSO/DSO4000
±100 V variable, 100 Hz square wave, calibratedFluke 9100
1
TekProbe inputsTektronix 1103
Tektronix DMM4040/4050
accuracy
Tektronix part number 012-0482-xx
BNC male-to-dual binding post
BNC female-to-SMA male
BNC male-to-dual banana male
Included with probe accessory kit
Tektronix part number 103-0035-xx
Tektronix part number 015-1018-xx
Fluke PM9081
Tektronix part number AC280–FL
TekVPI-to-BNCTektronix TPA-BNC
Insulated, slotted (straight) head
Insulated, narrow-slotted (straight) head
Reusable, adhesive-backed label that covers
Tektronix part number 003-1433-xx
Tektronix part number 003-1928-xx
Tektronix part number 335-2913-xx
adjustment access openings
WARNING. These procedures require you to remove a reusable label from the back of the probe. You must replace
the label after you complete the probe adjustments. Failure to do so may subject the user to high voltages present in
the probe during measurements.
44P5200A Series High Voltage Differential Probes Instruction Manual
Adjustment Procedures
WARNING. These procedures require the application of high voltage to the inputs of the probes. Only qualified personnel
should perfor
voltage measurements should be followed and adhered to.
WARNING. These procedures require you to remove a reusable label from the back of the probe. You must replace the
label after y
probe during measurements. If you need a replacement label, refer to the Equipment Required table for the Tektronix part
number. (See Table 15 on page 44.)
m any testing with voltage levels exceeding 30 V
ou complete the probe adjustments. Failure to do so may subject the user to high voltages present in the
. All pertinent safety rules and guidelines for elevated
rms
Adjustments
WARNING. T
adjustments described in the procedures. Failure to do so presents a potential shock hazard.
hese procedures require the application of high voltages to the probes. Use only an insulated tool to make the
Test Setup
1. Turn on the oscilloscope.
2. Connect the output of the probe to channel 1 of the oscilloscope.
3. If you are testing a P5200A probe, connect the 9 VDC output plug of the AC adapter to the DC input jack of the probe,
and then connect the AC adapter to the correct line voltage.
4. Turn on the g
P5200A Series H igh Voltage Differential Probes Instruction Manual45
enerators and let the probe and test equipment warm up for 20 minutes.
Adjustments
Offset Zero
1. Set the oscilloscope offset to 0 volts.
2. Connect the probe inputs together with the hook tips.
This procedure is identical for three of the probes. The fourth probe (P5200A) only uses a digital control for offset, so
there is no hardware adjustment.
For probes with serial numbers C199999 and below, Offset Zero is the only adjustment that can be done to the probe.
For probes with serial numbers C020000 and above, Offset Zero is the only adjustment that can be done without
removing the back label.
The adjustment for each range is independent and does not interact between the ranges.
3. This st
4. Press and hold the probe BANDWIDTH and ATTENUATION buttons until the OVERRANGE LED on the probe flashes.
5. Release the buttons. The OVERRANGE LED continues to flash, indicating that the digitally-controlled offset zero
6. Use the probe BANDWIDTH and ATTENUATION buttons to set the probe offset voltage as close to 0 V as possible,
NOTE. To display the actual P5200A probe voltage instead of a scaled value on the oscilloscope, you must match the
atte
oscilloscope; the access method depends on the oscilloscope model.
For example, on DPO/MS O4000 series oscilloscopes, press the front-panel channel number button and then press the
More button on the lower bezel until Probe Setup is highlighted. The oscilloscope attenuation setting is displayed in the
Pro
7. Press the AUDIBLE OVERRANGE button on the probe to store the adjusted offset value. The OVERRANGE LED stops
8. Select the remaining attenuation range and repeat steps 4 through 7.
ep only applies to P5202A, P5205A, and P5210A probes. Using an insulated adjustment tool, center the OFFSET
ADJUST pot in the probe comp box.
adjustment is enabled.
played on the oscilloscope. The BANDWIDTH button decreases the offset voltage and the ATTENUATION
as dis
button increases it.
nuation setting of the oscilloscope to the probe attenuation setting. Use the on-screen Probe Setup menu on the
be Setup menu. Change the attenuation by turning the multipurpose knob.
flashing to indicate that the offset value is stored and that the adjustment is disabled.
46P5200A Series High Voltage Differential Probes Instruction Manual
Accessing the Internal Adjustments
Adjustments
NOTE. Only pro
numbers C 019999 and below that require adjustments (other than offset zero) must be returned to Tektronix for service.
1. Remove the reusable back-panel label shown below to gain access to the adjustments. Store the label in a safe place to
preserve the
bes with serial numbers C020000 and above have internal adjustments. (See Figure 14.) Probes with serial
adhesive backing for reuse.
Figure 14: Internal adjustment locations (S/N C020000 and above only)
CAUTION. You must replace the reusable label after you complete the adjustment procedures. Failure to do so may subject
the user to high voltages present in the probe during measurements. If you need a replacement label, refer to the Equipment
ired table for the Tektronix replacement part number. (See Table 15 on page 44.)
Requ
P5200A Series H igh Voltage Differential Probes Instruction Manual47
Adjustments
Gain Accuracy
The equipment and equipment settings for this test differ between probes. Refer to the table for specific settings for the
probe that you are testing. (See Table 16 on page 49.)
WARNING. Dangerous voltages will be present on the calibration generator output terminals and connection cables. Always
verify that the generator is in the standby mode before you make any connections to the generator.
1. Verify that the generator output is off.
2. If you are adjusting a P5200A probe, connect the probe output directly to the DMM through a BNC-to-male banana jack
adapter. Otherwise, connect the probe to the D MM through the Tektronix 1103 power supply. Set the DMM to AC volts.
3. Connect the probe inputs to the front outputs of the generator, using adapters if necessary.
4. Set the probe attenuation to the lower (most sensitive) range for the probe that you are adjusting.
5. Set the generator square wave output frequency and voltage to the values shown in the table for the probe that you
are adjusting. (See Table 16 on page 49.)
6. Enable the generator output.
7. Adjust the low-range DC gain pot in the probe to ≤2% of the expected output.
WARNING. Use only an insulated tool to make the adjustment. Failure to do so presents a potential shock hazard.
8. Disable the generator output.
48P5200A Series High Voltage Differential Probes Instruction Manual
Adjustments
9. Set the probe attenuation to the next range and set the generator output voltage to the value shown in the table.
10. E nable the generator output and adjust the high-range DC gain pot in the probe to ≤2% of the expected output.
11. Disable the generator output.
Table 16: Adjust gain accuracy equipment settings
ProbeGenerator square wave outputProbe output voltage
WARNING. Dangerous voltages will be present on the calibration generator output terminals and connection cables. Always
verify that the generator is in the standby mode before you make any connections to the generator.
1. Verify that the generator output is off.
2. Connect both of the probe inputs to the red (+) banana connector on the front output terminals of the generator. Use a
BNC-banana adapter if necessary.
3. Set the sine wave output of the generator to the voltage and frequency listed in the table. (See Table 17 on page 50.)
4. Set the oscilloscope horizontal to 10 ms/div.
5. Set the probe attenuation to the lower (most sensitive) range of the probe.
P5200A Series H igh Voltage Differential Probes Instruction Manual49
Adjustments
6. Enable the generator output. Set the oscilloscope vertical to display the signal. For a stable display, connect the
generator Sens
e output to another channel and trigger off of that channel.
7. Using the narr
on the oscilloscope. Use averaging or hi-res filters to make viewing the 40 Hz signal easier.
WARNING. Use only an insulated tool to make the adjustment. Failure to do so presents a potential shock hazard.
8. Disable the generator output.
Table 17: DC CMRR
ModelRangeVoltage (rms)Voltage (p-p)Frequency
P5200A50X353.53 V1000 V40 Hz
P5202A20X200 V566 V40 Hz
P5205A50X353.53 V1000 V40 Hz
P5210A100X353.53 V1000 V40 Hz
ow-bladed tool, adjust the DC CMRR pot in the probe to minimize the amplitude of the waveform displayed
test equipment settings
ProbeGenerator sine wave output
50P5200A Series High Voltage Differential Probes Instruction Manual
LF Compensation
Adjustments
NOTE. The P520
probe models have two long-term +LF adjustments and two long-term –LF adjustments each.
1. Verify that the generator output is off.
2. Connect the probe inputs to the signal output connector on the back of the generator, using adapters if necessary.
Connect the red probe lead to the signal, and the black lead to ground.
2A probe only has one long-term +LF adjustment and one long-term –LF adjustment. The other three
3. Set the probe attenuation to the lower range for the probe that you are adjusting.
4. Set the oscilloscope horizontal to 4 μs/div, Acq mode = Average 16.
5. Set the generator fast-rise output (rise time waveform) frequency to 10 kHz.
6. Set the generator fast-rise output voltage to 50 Vp-p.
7. Enable the generator output. Set the oscilloscope vertical to display the signal.
8. Make the adjustments in the following order: long-term +LF, long-term +LF*, short-term +LF. (The long-term +LF*
adjustment is not used in the P5202A probe.) Repeat this sequence as necessary to optimize the square-wave response.
WARNING. Use only an insulated tool to make the adjustment. Failure to do so presents a potential shock hazard.
P5200A Series H igh Voltage Differential Probes Instruction Manual51
Adjustments
9. Disable the generator output.
10. Reverse the probe input leads to the generator.
11. Invert the signal and trigger slope on the oscilloscope to display the rising edge of the signal.
12. Enable the generator output and make the adjustments in the following order: long-term –LF, long-term –LF*, short-term
13. Disable the generator output.
Table 18: LF compensation test equipment settings
ModelRangeVoltage (p-p)Frequency
P5200A50X50 V10 kHz
P5202A20X50 V10 kHz
P5205A50X50 V10 kHz
P5210A100X50 V10 kHz
–LF. (The long-term –LF* adjustment is not used in the P5202A probe.) Repeat this sequence as necessary to optimize
the square-wave response.
ProbeGenerator fast-rise output
AC CMRR
WARNING. Dangerous voltages will be present on the calibration generator output terminals and connection cables. Always
verify that the generator is in the standby mode before you make any connections to the generator.
1. Verify that the generator output is off.
2. Connect both of the probe inputs to the red (+) banana connector on the front output of the generator. Use a
BNC-banana adapter if necessary.
3. Set the sine wave output of the generator to 297 Vp-p (105 Vrms) @100 kHz.
4. Set the oscilloscope horizontal to 10 μs/div.
5. Set the probe bandwidth to full and the attenuation to the lower range of the probe.
52P5200A Series High Voltage Differential Probes Instruction Manual
6. Enable the generator output. Adjust the oscilloscope vertical to display the signal.
Adjustments
7. Make only slight
the following
the P5202A probe.)
WARNING. Use only an insulated tool to make the adjustment. Failure to do so presents a potential shock hazard.
8. Disable the generator output.
Table 19: AC CMRR
ModelRangeVoltage (p-p)Frequency
P5200A50X297 V100 kHz
P5202A20X297 V100 kHz
P5205A50X297 V100 kHz
P5210A100X297 V100 kHz
adjustments to only the +LF pots to optimize the CMRR (minimize the signal). Adjust the pots in
order: short-term +LF, long-term +LF, long-term +LF*. (The long-term +LF* adjustment is not used in
test equipment settings
ProbeGenerator sine wave output
This completes the adjustment procedures.
P5200A Series H igh Voltage Differential Probes Instruction Manual53
Troubleshootin
g
Troubleshooting
Host Instrument Firmware
Some instruments may require a firmware upgrade to support full functionality of the latest probes that are offered by
Tektronix. Instruments with lower versions of firmware may not display a ll probe controls and indicators on screen, and in
some cases m ay require you to power-cycle the instrument to restore normal instrument operation. If you are having problems
with your probe and suspect that you need to upgrade your instrument firmware, go to www.tektronix.com/probe-support
to download the latest firmware.
To check the firmware version on W indows-bas ed instruments, from the menu bar, click H elp/About TekScope. On
Linux-based instruments, press the Utilities button on the front panel.
Error Conditions
LEDs Do Not Remain Lit
If none of the LEDs remain lit after you connect the probe, a probe/oscilloscope interface fault may exist. Perform the
following steps until you clear the fault or isolate the problem:
If you are using a P5200A probe, check that the DC adapter is plugged in and functional.
Disconnect and reconnect the probe to restart the power-on diagnostic sequence.
Connect the probe to a different channel on the oscilloscope.
Disconnect the probe from the oscilloscope, power-cycle the oscilloscope, and then reconnect the probe.
Connect the probe to a different oscilloscope.
If the symptoms remain (they follow the probe), then the probe is defective and must be returned to Tektronix for repair.
Signal Display
probe is connected to an active signal source and you do not see the signal displayed on the oscilloscope:
If the
k that the probe accessories that you are using are fully mated.
Chec
k the probe connection on your circuit.
Chec
orm a functional check on the probe.
Perf
54P5200A Series High Voltage Differential Probes Instruction Manual
Cleaning
Protect the probe from adverse weather conditions. The probe is not waterproof.
CAUTION. To prevent damage to the probe, do not expose it to sprays, liquids, or s olvents. Avoid getting moisture inside
the probe when cleaning the exterior.
Clean the exterior surfaces of the probe with a dry, lint-free cloth or a soft-bristle brush. If dirt remains, use a soft cloth or
swab dampened with a 75% isopropyl alcohol solution. Use only enough solution to dampen the cloth or swab. Do not use
abrasive compounds on any part of the probe.
Service
There are no user-serviceable parts in the P5200A Series probes. If your probe requires service, contact your Tektronix
service representative or repair center for instructions on returning your probe for repair.
Troubleshootin
g
P5200A Series H igh Voltage Differential Probes Instruction Manual55
Troubleshootin
g
56P5200A Series High Voltage Differential Probes Instruction Manual
Index
Index
A
Accessories
P5200A Series optional, 15
P5200A standard, 9
P5202A stand
P5205A standard, 9
P5210A standard, 11