Passive High-voltage Probes
P5100 • P5120 • P5122 • P6015A Data Sheet
P5122
For TPS2000 Series Only
DC to 200 MHz
P5120
Features & Benefits
P5100
DC to 250 MHz
1000 V CAT III (2500 V DC + Peak AC)
100X with Readout Coding
UL3111-1, EN61010-1, IEC61010-2 -031, CSA1010.1, CSA1010.2.031
7 to 30 pF Compensation Range
P5120
For TPS2000 Series
DC to 200 MHz
1000 V
Floatableupto600V
Preferred Probe for Ripple Measurements
20X
15 to 25 pF Compensation Range
UL61010B-2-031, EN61010-2-031, IEC61 010-2-031, CSA61010.2.031
RMS
CAT II
CAT II or 300 V
RMS
RMS
CAT III
1000 V
Floatable up to 600 V
100X
IEC61010-031
CATII when DC-coupled*
RMS
RMS
CATIII
1
P6015A
HighVoltage–20kVDC/40kVPeak(100msPulseWidth)
High Bandwidth – 75 MHz
Silicone Dielectric
Optional 1000X Readout Coding
Wide Compensation Range (7-49 pF)
Heavy-duty Versatile Ground Lead and Alligator Clip
Applications
Power Supply Design
Design Motor Drive
Electronic Ballast
Power Semiconductors
Switch Mode Control
UPS Systems
Power Converters
*1The P5122 probe should not be used for AC-coupled measurements on signals with greater than 300 V DC
offset. The P5120 is the recommended probe for measuring ripple on high-voltage DC supplies.
Data Sheet
P5120 and P 5122 IsolatedChannel™
Applications
In many applications, it is important to be able to isolate the measurement
from earth ground and also to isolate t he common voltage between
channels. The P5120 or P5122, coupled with the TPS2000 Series digital
storage oscil
earth ground and full isolation between the channels. The P5120 is the
recommended probe for measuring ripple on high-voltage DC supplies. T he
P5122 probe should not be used for AC-coupled measurements on signals
with greater than 300 V DC offset.
loscopes, deliver both the isolation for the measurem ent from
P6015A
P5100 High-voltage Probe
The P5100 is a low input capacitance high-voltage probe (2.5 kV) designed
for higher-frequency applications. The probe can be compensated to match
plug-ins and oscilloscopes with nominal input capacitances of 7-30 pF. A
variety of screw-on accessories provide easy connection to the device
under test.
P6015A High-v
For heavy-duty high-performance measurements of voltages over 2.5 kV,
the P6015A is
to 20 kV
bandwidth enables you to capture fast, high-voltage signals.
The P6015A uses an environmentally safe silicone compound for a
dielectric and never needs refilling. Other features include: A 7-49 pF
compensation range, small compensation box that fits on adjacent amplifier
inputs, and a readout option for use with most Tektronix digital scopes. With
the readout option, displayed voltage amplitude values will be the actual
signal value
Note: Using the readout version with other than Tektronix digital scopes
may result i
the industry standard. You can measure DC voltages up
and pulses up to 40 kV (peak, 100 ms duration). The 75 MHz
RMS
rather than understated by a factor of 1,000 .
n an erroneous readout display.
oltage Probe
2 www.tektronix.com