Tektronix P5120 200 MHz Passive High-Voltage Probe Primary User

Instructions
P5120 200 MHz Passive High-Voltage Probe
071-1463-00
www.tektronix.com
*P071146300*
071146300
Copyright © Tektronix, Inc. All rights reserved.
Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077
TEKTRONIX and TEK, TEKPROBE are registered trademarks of Tektronix, Inc.
WARRANTY
P5120 Probe
Tektronix warrants that the product listed above will be free from defects in materials and workmanship for a period of one (1) year from the date of original purchase from an authorized Tektronix distributor. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Batteries are excluded from this warranty. Parts, modules and replacement products used by Tektronix for warranty work may be new or reconditioned to like new performance. All replaced parts, modules and products become the property of Tektronix.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, shipping charges prepaid and with a copy of Customer proof of purchase. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non--Tektronix supplies; or d) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
WARRANTY (Continued)
P5120 Probe
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
Table of Contents
General Safety Summary iii............................
Features and Accessories 1.............................
Operating Basics 5...................................
Maximum Nondestructive Input Voltage 5.................
Reference Lead Length 5...............................
Probe Loading 6......................................
Adjustment Procedures 7..............................
LF Compensation 7....................................
HF Compensation 8...................................
Performance Verification 15............................
Maintenance 17.......................................
Cleaning Instructions 17.................................
Customer Support 17...................................
Preparation for Shipment 18..............................
Specifications 19......................................
Replaceable Parts 25...................................
P5120 High-Voltage Probe Instructions
i
Table of Contents
List of Figures
Figure 1: Waveform distortion from reference lead length 5....
Figure 2: Probe loading equivalent circuit 6................
Figure 3: LF compensation setup 7.......................
Figure 4: LF compensation adjustment 8...................
Figure 5: Signal source characteristics for HF compensation 8..
Figure 6: Opening the compensation box 10.................
Figure 7: HF compensation setup 11.......................
Figure 8: HF compensation adjustments 12..................
Figure 9: HF compensation regions of fast rise signal 12.......
Figure 10: Bandwidth test setup 16........................
Figure 11: Maximum voltage derating curves
(V
RMS
Figure 12: Typical input impedance 21.....................
Figure 13: Typical input phase 22.........................
Figure 14: P5120 replaceable parts 25......................
List of Tables
, DC coupled) 20..............................
Table 1: Required test equipment 15.......................
Table 2: Warranted electrical specifications 19...............
Table 3: Typical electrical characteristics 21................
Table 4: Typical mechanical characteristics 23...............
Table 5: Environmental specifications 23...................
ii
P5120 High-Voltage Probe Instructions
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified.
WARNING. Only qualified personnel should perform service procedures
To Avoid Fire or Personal Injury
Connect and Disconnect Properly. Connect the probe output to the
measurement instrument before connecting the probe to the circuit under test. Disconnect the probe input and the probe reference lead from the circuit under test before disconnecti ng the probe from the measurement instrument.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all
ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Avoid Electric Overload. To avoid injury or fire hazard, do not apply
potential to any input, including the reference inputs, that varies from ground by more than the maximum rating for that input.
Avoid Exposed Circuitry. Do not touch exposed connections and
components when power is present.
Avoid Electric Shock. To avoid injury or loss of life, do not connect or
disconnect probes or test leads while they are connected to a voltage source.
Floating. Do not float the probe reference lead to > 600 V
or > 300 V
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere. To avoid injury or fire
hazard, do not operate this product in an explosive atmosphere.
Keep Probe Surface Clean and Dry. To avoid electric shock and
erroneous readings, keep probe surface clean and dry.
P5120 High-Voltage Probe Instructions
RMS
CAT III.
RMS
CAT II
iii
General Safety Summary
Inspect Probe. Before use, inspect probe for damage and do not use if
damaged.
Product Damage Precautions
Do Not Operate With Suspected Failures. If you suspect there is damage
to this product, have it inspected by qualified service personnel.
Do Not Immerse in Liquids. Clean the probe using only a damp cloth.
Refer to cleaning instructions on page 17.
Safety Terms and Symbols
Terms in This Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. These symbols may appear on the product:
iv
P5120 High-Voltage Probe Instructions
General Safety Summary
DANGER
High Voltage
Protective Ground
(Earth) Terminal
ATTENTION
Refer to Manual
Double
Insulated
Certifications and Compliances
Compliances. Consult the product specifications for Overvoltage
Category and Safety Class.
Overvoltage Category. Overvoltage categories are defined as follows:
CAT III: Distribution level mains, fixed installation
CAT II: Local level mains, appliances, portable equipment
CAT I: Signal level, special equipment or parts of equipment, telecommunication, electronics
P5120 High-Voltage Probe Instructions
v
General Safety Summary
vi
P5120 High-Voltage Probe Instructions
Contacting Tektronix
Phone 1-800-833-9200*
Address Tektronix, Inc.
Department or name (if known) 14200 SW Karl Braun Dri ve P.O. Box 500 Beaverton, OR 97077 USA
Web site www.tektronix.com
Contacting Tektronix
Sales support
Service support
Technical support
* This phone number is toll free in North America. After office
hours, please leave a voice mail message. Outside North America, contact a Tektronix sales office or distributor; see the Tektronix web site for a list of offices.
1-800-833-9200, select option 1*
1-800-833-9200, select option 2*
Email: techsupport@tektronix.com
1-800-833-9200, select option 3*
6:00 a.m. -- 5:00 p.m. Pacific time
P5120 High-Voltage Probe Instructions
vii
Contacting Tektronix
viii
P5120 High-Voltage Probe Instructions
Features and Accessories
The P5120 is a passive high-voltage probe designed for use with TPS2000 series instruments. The probe feat ures a floating reference and is ideal for measuring and troubleshooting power supplies and high-voltage devices. The probe provides 20X attenuation of signals up to 1000 V P5120 probe is certified to international safety standards.
The P5120 probe has features and accessories to make high-voltage measurements safer and easier. Please take a moment to familiarize yourself with these features and accessories and their uses.
WARNING. Do not substitute accessories from other products for those provided with the P5120. The accessories for the P5120 have been safety tested. The use of non-tested accessories may cause injury.
P5120 features and standard accessories
Finger Guard
Probe
Reference
6-32 Probe Tip
and derated performance up to 200 MHz. The
RMS
Probe Body -- The P5120 probe body is designed for personal
safety, ergonomic comfort, and signal fidelity.
The finger guard provides protection when the hook tips are not being used. Keep your fingers behind the finger guard whenever possible to reduce the risk of a shock from the circuit under test.
The probe tip is a 6--32 threaded post that accepts both standard and optional accessories.
Keep the probe body and accessories clean to reduce the risk of shock due to surface conduction.
P5120 High-Voltage Probe Instructions
1
Features and Accessories
P5120 features and standard accessories (Cont.)
Compensation Box -- The compensation box contains low- and
high-frequency compensation adjustments. The high-frequency adjust­ments are accessible by removing the compensation box cover. Refer to Adjustment Procedures on page 7.
Adjustment Tool -- Use the supplied adjustment tool for probe com-
pensation adjustments. Do not use metallic or conductive tools while making adjustments.
Color Markers -- The matching markers help you identify probe in-
puts at a glance. Clip one marker near the probe head and the match­ing marker near the compensation box as shown.
Matching
Markers
Small Hook Tip -- Use the small hook tip for making connections to
small conductors such as component leads.
Install the small hook tip by sliding it over the body of the probe and screwing it onto the threaded probe tip.
To use the tip, hold the probe body and pull the tip shield back. Hook the tip onto the circuit and release the shield. The hook tip will firmly hold the conductor under test.
2
P5120 High-Voltage Probe Instructions
P5120 features and standard accessories (Cont.)
Large Hook Tip -- Use the large hook tip when working with larger
components such as bolt terminals and buss bars typically found in power distribution equipment.
Install the large hook tip by sliding it over the body of the probe and screwing it onto the threaded probe tip.
To use the tip, hold the probe body and pull the tip shield back. Hook the tip onto the circuit and release the shield. The hook tip will firmly hold the conductor under test.
Reference Leads -- The P5120 probe comes with a long and a short,
floatable reference lead. Both reference leads connect to a removable crocodile clip.
Attach the reference lead to the probe body by inserting the probe head into reference lead as shown.
The length of the reference lead directly impacts the fidelity of high-fre­quency signals. The longer the reference lead, the more the signal will be distorted. For more information about the reference lead and signal quality, go to page 5.
Crocodile Clip -- The large crocodile clip attaches to the insulated
banana connector on the reference leads.
Features and Accessories
WARNING. For safe electrical connections, use the crocodile clip provided. Do not attempt to substitute the crocodile clip with any other accessory.
P5120 High-Voltage Probe Instructions
3
Features and Accessories
P5120 optional accessories
Probe Tip Extender -- Use the probe tip extender to connect the
probe to larger test sockets located in equipment or on printed circuit boards.
Connect the probe tip extender by screwing it onto the threaded probe tip.
Reference
Clip
Probe
Reference
Probe-to-BNC Adapter -- Use the probe-to-BNC adapter to connect
the probe directly to BNC connectors. The adapter is recommended for use when verifying or adjusting the probe.
Attach the adapter by pressing it onto the 6-32 probe tip and making sure that the reference clip makes contact with the probe reference.
WARNING. Do not connect the Probe-to-BNC Adapter to a floating or high-voltage source. The BNC adapter is only rated to 500 V (DC + peak AC) with the probe reference connected to earth ground.
4
P5120 High-Voltage Probe Instructions
Operating Basics
To get guaranteed performance from the P5120 probe, please read and apply the information in this section.
Maximum Nondestructive Input Voltage
The P5120 probe is designed to be used in a hi gh voltage environ­ment; however, the voltage input rating of the probe starts to decrease at about 2 to 3 MHz. Please refer to the Specifications section of these instructions for the maximum input voltage and frequency derating information.
Reference Lead Length
Always use as short a referenc e l ead as possible between the probe head and circuit reference for high frequency or fast rise time signals.
The input capacitance of the probe and the series inducta nce added by the probe tip and reference lead can result in a resonant circuit; this circuit may cause parasitic “ringing” visible within the bandwidth of your oscilloscope. Refer to Figure 1.
6-inch Reference 18-inch
Figure 1: Waveform distortion from reference lead length
P5120 High-Voltage Probe Instructions
Reference
5
Operating Basics
Parasitic ringing is not a concern when the probe is used with low frequency or slow rise time signals and is not apparent if the frequency content is beyond the bandwidth of the oscilloscope.
Probe Loading
When you touch your probe tip to a circuit element, you are introducing a new resistance, capacitance, and inductance into the circuit. Refer to Figure 2.
R
source
Probe C 11.2 pF
in
V
source
L
(reference lead)
cl
Probe R
5M
in
Figure 2: Probe loading equivalent circuit
Frequency and impedance of the source determine how much the probe loads the circuit you are measuring. As the frequency of the source starts to increase beyond 1 kHz, the input impedance of the probe begins to decrease. The lower the impedance of the probe relative to that of the source, the more the probe loads the circuit under test. Refer to Figure 12 on page 21 for a graph of frequency versus input impedance.
The probe has virtually no loading effect on sources with relatively low impedance and low frequency.
6
P5120 High-Voltage Probe Instructions
Adjustment Procedures
The low-frequency (LF) compensation of the P5120 probe must be checked every time the probe is connected to a different oscilloscope input.
The high-frequency (HF) compensation should be checked if the probe was repaired or fails to meet rated bandwidth specifications. Check the probe using the performance verification procedures on page 15 to determine if adjustment is required.
LF Compensation
Before taking any measurements with the probe, first check the LF compensation of the probe and adjust it to match the channel input it is connected to.
1. Connect the probe to the Probe Comp signal on your TPS2000 instrument and then press AUTORANGE.SeeFigure3.Usethe reference lead connection to obtain the proper waveform.
AUTOSET
Probe reference
lead to
Figure 3: LF compensation setup
P5120 High-Voltage Probe Instructions
Ch 1
Probe tip to PROBE COMP
7
Adjustment Procedures
2. Compensate the probe by adjusting the trimmer in the probe
compensation box so that the c orners of the signal are square. See Figure 4.
LF
Figure 4: LF compensation adjustment
HF Compensation
To compensate the probe for high frequencies, you will need a properly terminated signal source that has the characteristics given in Figure 5. Refer to Table 1 on page 15 for a complete list of required test equipment.
1kHz
1MHz
1 Vpp
Figure 5: Signal source characteristics for HF compensation
To access the HF adjustments of the probe, you must open the cover of the compensation box using the procedure that follows.
8
P5120 High-Voltage Probe Instructions
tr<1 ns
90%
10%
Adjustment Procedures
Opening the Compensation Box
NOTE. This procedure should be performed by qualified service personnel only.
To open the cover, obtain two small screwdrivers and use them to unlatch the tab as shown in Figure 6.
WARNING. Avoid damaging the tab. A damaged tab will allow the cover to easily separate and expose the user to the electrical potential of the reference connection.
To avoid damaging the tab, do not attempt to press on the tab through the hole in the cover where it latches. In the event of a damaged tab, replace the probe.
Perform the following procedure to open the compensation box.
1. Insert the flat blade of small screwdriver (A) into the location shown in Figure 6. Insert the blade until it stops against the back of the tab.
P5120 High-Voltage Probe Instructions
9
Adjustment Procedures
Figure 6: Opening the compensation box
B
Tab
Insert up to the tab
A
10
2. While holding screwdriver (A) in pla ce, insert small screwdriver
(B) between the ridge on the front of the compensation box and the edge of the cover.
3. Gently twist the blade of screwdriver (B) to separate the cover
and unlatch the tab.
4. After unlatching the tab, slide the cover back to expose the
internal adjustments.
HF Compensation Procedure
After opening the compensation box, perform the following procedure to compensate the probe for high frequencies.
1. Connect a 50 feedthrough termination to the positive-going FAST RISE output of the calibration generator. See Figure 7.
P5120 High-Voltage Probe Instructions
Calibration
generator
FAST RISE output
50 termination
Probe-to-BNC adapter
P5120 probe
Figure 7: HF compensation setup
2. Refer to page 4 and use the optional probe-to-BNC adapter to
connect the probe to the termination.
Adjustment Procedures
Ch 1
3. Set the Calibration Generator to 1 MHz.
4. Set the PULSE AMPLITUDE to 1.0 V
NOTE. If you have changed the default setting for the probe type, access the vertical menu of the oscilloscope and return the setting to 20X.
5. Set the oscilloscope to 100 mV/division.
6. Set the oscilloscope averaging to 4.
7. Set the oscilloscope to 5.0 ns/div Adjust the oscilloscope as
necessary to view the leading edge of the waveform.
8. Locate the HF1, HF2,andHF3 adjustments. See Figure 8.
P5120 High-Voltage Probe Instructions
P-P
.
11
Adjustment Procedures
Figure 8: HF compensation adjustments
9. The probe should be preset to the correct nominal position.
10. Adjust HF1 for flatness and adjust HF3 for t he be st front corner. See Figure 9. Because all adjustments interact, readjust HF1, HF2,andHF3 as necessary to achieve th e be st compromise
between the fastest rise time and minimum aberrations.
Not used
LF
HF1
HF2
HF3
12
HF1
(50 ns)
HF2
(20 ns)
HF3
(10 ns)
10%
Aberrations
Figure 9: HF compensation regions of fast rise signal
P5120 High-Voltage Probe Instructions
Adjustment Procedures
11. After compensating the HF, carefully slide the cover back until
the tab latches securely into place.
P5120 High-Voltage Probe Instructions
13
Adjustment Procedures
14
P5120 High-Voltage Probe Instructions
Performance Verification
Oscill
Calibrati
Use the procedure in this section to verify that the P5120 probe meets the warranted bandwidth specification highlighted with boldface type in the Specifications section on page 19. The steps of this procedure must be performed in the order given using a TPS2000 series instrument and other equipment that meets the minimum requirements listed in Table 1.
Table 1: Required test equipment
Description Minimum Requirements Example
Oscilloscope Use a TPS2000 series
oscilloscope only. The probe is optimized to match the input characteristics and bandwidth of the TPS2000 series
Calibration generator (for HF adjustment only)
Leveled signal generator 50 kHz to 250 MHz
Termination, 50 feedthrough
Adapter, probe tip to BNC ProbetiptomaleBNC Tektronix part number
1 ns rise time square wave Fluke 9100 Universal
50 male to female BNC
TPS2024
Calibration System with
oscope
Module (Option 250)
Tektronix part number 011-0049-02
013-0291-00
NOTE. Compensate the probe for LF before performing this procedure. The LF compensation procedure begins on page 7.
If the probe does not meet the bandwidth requirement of this procedure, perform the adjustments beginning on page 7 and retest.
on
P5120 High-Voltage Probe Instructions
15
Performance Verification
1. Connect the test setup as shown in Figure 10.
Leveled signal
generator
Test oscilloscope
50 termination
P5120 probe
Ch 1
Output
Probe tip to
BNC adapter
Figure 10: Bandwidth test setup
2. Set the leveled signal generator output frequency to 1 MHz.
NOTE. If you have changed the default setting for the probe type, access the vertical menu of the oscilloscope and return the setting to 20X.
3. Set the oscilloscope to 10 mV/division. (The effective setting with the 20X attenuation of the probe is 200 mV/division.)
4. Adjust the leveled signal generator for a display of exactly 6 divisions peak-to-peak.
5. Increase the leveled signal generator frequency until the display decreases to 4.2 divisions peak-to-peak. (This represents the .707 point or half-power amplitude.)
6. Note that the frequency is greater than 200 MHz for the TPS2024 and 100 MHz for the TPS2012 and 2014. If not, repeat the HF compensation to achieve a faster risetime.
16
P5120 High-Voltage Probe Instructions
Maintenance
This section contains information necessary for the maintenance and repair of the P5120 probe.
NOTE. The P5120 probe has no replaceable components. Accessory parts may be ordered as indicated in the replaceable parts list.
Cleaning Instructions
Clean the probe using a soft cloth dampened with a mild detergent and water solution. Do not use any other chemicals or abrasives.
Customer Support
Who to call When to call Where to call
Operational Support Call if you need
help using your probe or to order accessories
Service Support Call if your probe
needs repair
1-800-TEK-WIDE (1-800-835-9433)
If outside the United States or Canada, contact your nearest Tektronix Field Office
Call your nearest Tektronix Service Center
P5120 High-Voltage Probe Instructions
17
Maintenance
Preparation for Shipment
If you need to ship your probe, pack it care fully to prevent damage while in transit. If the original packaging is unfit for use or not available, use the following packaging guidelines:
1. Use a corrugated cardboard shipping carton having inside dimensions at least one inch greater than the probe dimensions. The box should have a carton test strength of at least 200 pounds.
2. Put the probe into a plastic bag or wrap to protect it from dampness in transit.
3. Place the probe into the box and stabilize it with light packing material.
4. Seal the carton with shipping tape.
18
P5120 High-Voltage Probe Instructions
Specifications
The warranted specifications of Table 2 that appear in boldface type are specifications checked in the Performance Verification section. Specifications that are not warranted are typical (Tables 3 and 4) and are provided as general information for your convenience.
The specifications apply to a P5120 probe installed on a Tektronix TPS2000 series oscilloscope. The instrument must have a warm-up period of at least 20 minutes and be in an environment that does not exceed the limits described in Table 5.
Table 2: Warranted electrical specifications
Attenuation (DC coupled system)
Bandwidth
(--3 dB, DC coupled system)
Maximum Rated Input Voltage IEC 1010-1 Installation Category
Between probe tip and probe reference or between probe tip and earth ground (Refer to Figure 11 a.)
Between probe reference and earth ground (Refer to Figure 11 b.)
Safety UL 61010B--2--031; 2003
IEC Installation Category Category II
IEC Pollution Degree Degree 2
20X, ±2%
DC to 200 MHz (TPS2024) DC to 100 MHz (TPS2012, TPS2014)
II III
1000 600
600 300
CSA 22.2 No. 1010.2--031, 1994 EN 61010-031 2002
P5120 High-Voltage Probe Instructions
19
Specifications
1,000 V
Voltage (V )Voltage (V )
1,000 V
Category I & II
600 V
Category III
300 V
100 V
10 V
1V
10 Hz
a. Probe tip to reference or earth ground
600 V
Category I & II
300 V
Category III
100 V
100 Hz
1KHz
P5120 Voltage derating
10 KHz
P5120 Voltage derating
100 KHz
Frequency (Hz)
1MHz
10 MHz
100 MHz
200 MHz
10 V
1V
10 Hz
b. Probe reference to earth ground
100 Hz
1KHz
10 KHz
Frequency (Hz)
Figure 11: Maximum voltage derating curves (V
20
P5120 High-Voltage Probe Instructions
100 KHz
, DC coupled)
RMS
1MHz
10 MHz
100 MHz
200 MHz
Table 3: Typical electrical characteristics
Specifications
Rise Time (System) Checked using optional BNC adapter
< 3.6 ns (TPS2012, 2014) < 2.2 ns (TPS2024) Typical
Input Resistance (System)
5M
SeeFigure12
Input Phase (System) (See Figure 13)
Compensation Range 15 pF to 25 pF
Input Capacitance (System)
11.2 pF Probe connected to instrument and properly compensated. Checked at 1 MHz
10 M
5M
1M
100 k
10 k
OHMS
1k
P5120 Impedance
100
10
1
10 Hz
100 Hz
1KHz
10 KHz
Frequency (Hz)
Figure 12: Typical input impedance
P5120 High-Voltage Probe Instructions
100 KHz
1MHz
10 MHz
100 MHz
200 MHz
21
Specifications
0
-- 1 0
-- 2 0
-- 3 0
-- 4 0
-- 5 0
Degrees
-- 6 0
-- 7 0
-- 8 0
-- 9 0
--100
10 Hz
100 Hz
Figure 13: Typical input phase
P5120 Phase
1KHz
10 KHz
Frequency (Hz)
100 KHz
1MHz
10 MHz
100 MHz
200 MHz
22
P5120 High-Voltage Probe Instructions
Table 4: Typical mechanical characteristics
Specifications
Net Weight (including accessories)
Cable Length 3m(+--3cm)(9.84ft)
0.26 kg (0.57 lb)
Table 5: Environmental specifications
Temperature Operating: 0 °Cto+50°C
(+32 °F to +122 °F)
Nonoperating: --40 °Cto+71°C
(--40 °F to +159 °F)
Humidity Operating:
High: 50 °C (122 °F)/60% RH
Low: 30 °C(86°F)/60% RH
Nonoperating: High: 55 °Cto71°C (131 °F to 160 °F), 60% RH
Low: 0 °Cto30°C(32°Fto86°F), 90% RH
Transportation Tektronix Standard 062-0937-00 Revision C
Altitude Operating: 4,600 m (15,092 ft)
Nonoperating: 12,192 m (40,000 ft)
P5120 High-Voltage Probe Instructions
23
Specifications
24
P5120 High-Voltage Probe Instructions
Replaceable Parts
1
6
Figure 14: P5120 replaceable parts
Standard accessories
7
8
Optional accessories
2
3
45
P5120 High-Voltage Probe Instructions
25
Replaceable Parts
3
34404610011
CLIPELECCROCODILE82MML4MM
TK248
3
XK081
W
L
4
600
3--143
3--0
2
196342400
111SCREWDRIVER:ADJUSTMENTTOOL
LEADSETASSY:COMMONP51201)180L1)60L80009
003143302
L
L
L
Mfr. part no.
XK--081
003143302--
016188600
013029001
Mfr.
code
80009
TK2483
80009
TK2469
1) 6.0
DRIVER: ADJUSTMENT TOO
EAD SET ASSY: COMMON, P5120, 1) 18.0
CLIP,ELEC: CROCODILE, 82MM L, 4MM
SCRE
MARKER BAND SET
Name & description
Qty
Serial no.
Effective Dscont
part no.
Tektr o n ix
no.
index
Fig. &
STANDARD ACCESSORIES
-- -- -- -- -- -- -- -- -- -- -- 1 PROBE, HI VOLTAGE: FLOATING, 10X, 200 MHZ, P5120
14--1
HOOKTIP, SET: HOOKTIP ASSY, LARGE AND SMALL
1
1
344--0461--00
003--1433--02
016--1886--00
013--0290--01
-- 2
--6 196--3424--00 1
-- 3
-- 4
-- 5
071--1463--XX 1 MANUAL, TECH: INSTRUCTIONS, P5120 80009 0711463XX
OPTIONAL ACCESSORIES
--7 206--0060--00 1 TIP, PROBE 80009 206006000
--8 013--0291--00 1 ADAPTER, PROBE: PROBE TO BNC 80009 013029100
26 P5120 High--Voltage Probe Instructions
Replaceable Parts
D--4300 ESSEN 11 (W GERMANY)
MESSTECHNISCHE WERKE
BEAVERTON OR 97077--0001
CATHOSTRASSA 7
PO BOX 500
CROSS INDEX - MFR. CODE NUMBER TO MANUFACTURER
Mfr. code Manufacturer Address City, state, zip code
TRIQUEST CORP
0JR05
TK2483 HEI NRICH C KOSMEIER
GMBH--ESSEN
TK2469 UN ITREK CORPORATION 3000 LEWIS & CLARK HWY VANCOUVER WA 98661
80009 TEKTRONIX INC 14150 SW KARL BRAUN DR
27P5120 High-Voltage Probe Instructions
Replaceable Parts
28
P5120 High-Voltage Probe Instructions
Loading...