P5100A 500 MHz 100X High Voltage Probe
Instructions
Operating Information
The P5100A Probe is a high impedance probe with 100X attenuation that is designed for use with Tektronix ground-referenced oscilloscopes.
Connect the probe as shown in the illustrations below.
Compensating the Probe
Due to variations in oscilloscope input characteristics, the low-frequency compensation of the probe may need adjustment after you move the probe from one oscilloscope channel to another.
1.
Connect the probe to the oscilloscope channel that you plan to use for your measurements.
2.
Connect the probe to the probe compensation output terminals on the oscilloscope front panel.
WARNING.
Probe Comp signal on the oscilloscope when making this adjustment. To avoid electric shock, only use the insulated adjustment tool when making compensation adjustments.
3.
Push AUTOSET or otherwise adjust your oscilloscope to display a stable waveform.
4.
Adjust the trimmer in the probe until you see a perfectly at-top square wave on the display. (See illustration.)
To avoid electric shock, only connect to the
Standard Accessories
The probe includes the accessories shown below.
WARNIN G.
To avoid electric shock when using the probe or accessories, keep your ngers behind the nger guard of the probe body and away from the shaded area shown in the accessory illustrations below.
Item Description
Small hook tip (TASH)
Use this tip to access test points in tight spaces. Screw the hook tip onto the probe tip and then clamp the hook onto the circuit.
Rating: 2300 V CAT I 1000 V CAT II
Reorder Tektronix part number 013-0386-xx
Large hook tip (TALH)
Screw the hook tip onto the probe tip and then clamp the hook onto the circuit.
Rating: 2300 V CAT I 1000 V CAT II
Reorder Tektronix part number 013-0384-xx
Ground lead (TAGL)
Slide the ground lead over the probe head and snap it into place.
Connect the banana plug end directly to your circuit ground, or use the crocodile clip included with the probe.
Reorder Tektronix part numbers:
6 inch: 196-3524-xx
18 inch: 196-3525-xx
Item Description
Crocodile clip
Attach the clip to the banana end of the ground lead and then to your circuit ground.
Reorder Tektronix part number 344-0461-xx
Ground spring
Use this to limit aberrations on high frequency signals caused by ground path inductance.
You can bend the spring to reach nearby ground connections (within ~1 in).
Reorder Tektronix part number 214-5298-xx
Color bands
Use these bands to identi the oscilloscope channel at the probe head.
Reorder Tektronix part number 016-1886-xx (5 pairs)
Adjustment tool
Use only this insulated tool for compensation adjustments.
Reorder Tektronix part number 003-1433-xx
Optional Accessories
The accessories shown below are available for the probe and are rated 30 V unless indicated otherwise.
Accessory
BNC to Tip Adapter, Unterminated
Rubber Spring Tip
Part number
013-0291-xx
206-0060-xx
fy
*P071288200*
x
071-2882-00
Service Information
Performance verication and adjustment procedures are available on the Tektronix Web site. Go to www.tektronix.com/manuals and enter P5100A in the search eld.
Specications
Table 1: Electrical and mechanical specications
Bandwidth (–3 dB)
System rise time (typical)
System input capacitance
Compensation range
Propagation delay ~6.1 ns
System attenuation accuracy 100:1 ±1.75%
System input resistance @DC 40 M±1%
Cable length
Maximum input voltage
Performance Graphs
(see
500 MHz
<700 ps
2.5 pF
7–30 pF
2m
1000 V
)
CAT II, 2500 V
RMS
Performance Graphs
Characteristics Description
Additional Safety Standards
Table 2: Environmental specications
Characteristics Description
Temperature
Operating Nonoperating
pk
Humidity
Operating 5% to 95% relative humidity (%RH) up to
Nonoperating
Altitude
0°Cto+50°C(+32°Fto+122°F) –40 °C to +71 °C (–40 °F to +159.8 °F)
+30 °C, 5% to 75% RH above +30 °C up to +50 °C. Noncondensing
5% to 95% RH up to +30 °C 5% to 75% RH above +30 °C to +65 °C 5% to 45% RH above +65 °C up to +71 °C. Noncondensing
Operating: up to 3000 m (9842 ft)
Nonoperating: up to 15420 m (50,000 ft)
Safety Summary
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specied. Using the probe or accessories in a manner not specied could result in a shock or re hazard.
To Avoid Fire or Personal Injury
Ground-Referenced Oscilloscope Use.
reference lead of this probe when using with ground referenced oscilloscopes (for example, DPO, MSO, and TDS series oscilloscopes). The reference lead must be connected to earth potential (0 V).
Connect and Disconnect Properly.
to the measurement instrument before connecting the probe to the circuit under test. Disconnect the probe input and the probe reference lead from the circuit under test before
IEC61010-031; IEC 61010-031/A1:2008
Equipment Recycling.
with the European Union’s requirements according to Directive 2002/96/EC on waste electrical and electronic equipment (WEEE). For more information about recycling options, check the Support/Service section of the Tektronix Web site (www.tektronix.com).
This product complies
Do not oat the
Connect the probe output
disconnecting the probe from the measurement instrument.
Table 3: Certications and compliances
Characteristics Description
EC Declaration of Conformity
Measurement Category Product Examples
Pollution Degree 2
Compliance was demonstrated to the following specication as listed in the Ofcial Journal of the European Communities:
Low Voltage Directive 2006/95/EC: EN61010-031/A1: 2008
CAT III: Distribution-level mains, fixed installation
CAT II: Local-level mains, appliances, portable equipment
CAT I: Circuits not directly connected to mains.
Do not operate in environments where cond­uctive pollutants may be present (as dened in IEC 61010-1). Rated for indoor use only.
Avoid Electric Shock.
connect or disconnect probes or test leads while they are connected to a voltage source.
Observe All Terminal Ratings.
observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Avoid Electric Shock.
exceed the lowest rating of the probe or its accessory, whichever is less, including the measurement category and voltage rating.
Avoid Electric Overload.
not apply potential to any input, including the reference inputs, that varies from ground by more than the maximum rating for that input.
To avoid injury or loss of life, do not
To avoid re or shock hazard,
When using probe accessories, never
To avoid injury or re hazard, do
Avoid Exposed Circuitry and Do Not Operate Without Covers.
Do not touch exposed connections and components
when power is present.
Inspect The Probe and Accessories.
Before each use, inspect probe and accessories for damage (cuts, tears, defects in the probe body, accessories, cable jacket, etc.). Do not use if damaged.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Safety Terms and Symbols Terms in This Manual.
These terms may appear in this manual:
WARNING.
Warning statements identify conditions or
practices that could result in injury or loss of life.
CAUTION.
Caution statements identify conditions or practices that could result in damage to this product or other property.
Symbols on the Product.
These symbols may appear on
the product:
Contacting Tektronix
Web site: www.tektronix.com
Phone: 1-800-833-9200
Address: Tektronix, Inc.
Email:
Department or name (if known) 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
techsupport@tektronix.com
Warranty Information
For warranty information, go to www.tektronix.com/warranty.
Copyright © Tektronix, Inc. All rights reserved. www.tektronix.com
Loading...