xx
Probe Compensation
Performance G raphs
Due to variations in oscilloscope input characteristics, the
low-frequency compensation of the probe may need adjustment
(in 10X attenuation only) after moving the probe from one
oscilloscope channel to another.
If a 1 kHz c alibrated square wave displayed at 1 ms/division
shows significant differences between the leading and trailing
edges, perform the following steps to optimize low-frequency
compensation:
1. Connect the probe to the calibration signal on the
oscillosco
pe front panel.
WARNING. To avoid electric shock, only connect to the
Probe Comp s
ignal on the oscilloscope when making this
adjustment.
P2221
200 MHz 1X/10X Passive Probe
Instruct
ions
2. Press AUTOS
ET or otherwise adjust your digitizing
oscilloscope to display a stable waveform.
3. Adjust the trimmer in the probe until you see a perfectly
flat-top square wave on the display. (See i llustration.)
WARNING. To avoid electric shock, only use the insulated
adjustment tool when making compensation adjustments.
Tab le 2 :
Characteristics Description
Temperature
Humidity
Altitude
Electrostatic Immunity The probe is not static sensitive
Environmental specifications
Maximum operating
Minimum operating
Nonoperating
Operating
rating
Nonope
+50 °C (+122 °F)
0°C(+32°F)
–55°Cto+75°C
(–67 °F to +167 °F)
10 to 95%
3.0 km (10,000 ft) maximum
15 km (50,000 ft) maximum
relative humidity
*P0712
33500*
071-2335-00
P2221 1X/10X Probe
The P2221 200 MHz 1X/10X Passive Probe is a c ompact
passive probe with selectable 1X/10X attenuation.
The P2221 Probe has a compensation range of 10 to 25 pF,
and is designed for use with Tektronix DPO/MSO2000 Series
oscilloscopes.
obe has no user- or Tektronix-serviceable parts.
The pr
WARNING. Do not float the probes when used with
TDS1000/2000 Series Oscilloscopes.
NOTE. The P2221 Probe is not recommended for use with the
TPS2000 Series Oscilloscopes. Instead, use the P2220 1X/10X
probe.
cting the Probe to the Oscilloscope
Conne
Connect the probe as shown in the illustration below.
WARNING. To avoid electric shock when using the probe,
keep fingers behind the guard on the probe body.
Attenuation Select Switch
This switch selects the attenuation factor of the probe.
Connecting the Probe to t he Circuit
to the back of these instructions for connection methods
Refer
using the accessories included with the probe.
Specifications
Table 1: Electrical and mechanical specifications
Characteristic 10X position 1X position
Bandwidth
System Attenuation
Accuracy
Compensation
Range
System Input
Resistance
System Input
Capacitance
System Rise time
(typical)
Maximum input
voltage between tip
(signal) and earth
ground
Cable length
DC to 200 MHz
(–3 dB)
10:1 ±2% 1:1 ±2%
10 pF-25 pF NA
10 MΩ ±3% at DC 1 MΩ ±3% at DC
17.0 pF 110 pF
≤2.2 ns ≤50 ns
10X position:
300 V CAT II
420 V
, <50% Duty factor, <1 s PW
peak
1X position:
150 V
CAT II or 150 VDCCAT II
rms
210 V
, <50% Duty factor, <1 s PW
peak
1.5 m
DC to 6 MHz
(–3 dB)
-
ent Recycling. This product complies
Equipm
with the European Union’s requirements a ccording
to Directive 2002/96/EC on waste electrical
and electronic equipment (WEEE). For more
information about recycling options, check the
Support/Service section of the Tektronix Web site
tektronix.com).
(www.
3: Certifications and compliances
Tab le
Characteristics Description
EC Declaration of
rmity
Confo
Measurement
gory
Cate
Descriptions
Pollution Degree 2 Do not operate in environments where
Additional Safety
ndards
Sta
Compliance was demonstrated to
llowing specification as listed in
the fo
the Official Journal of the European
Communities:
Low Voltage Directive 2006/95/EC:
EN61010-031: 2002
Category Examples of Products in
CAT III
CAT II
CAT I
conductive pollutants may be present (as
defined in IEC 61010-1). Rated for indoor
use only.
UL61010B-1, First Edition &
1010B-2-031, First Edition.
UL6
CAN/CSA-C22.2 No. 1010.1-92, &
CAN/CSA-C22.2 No. 1010.2.031-94
IEC61010-031:2002
Pollution Degree 2
this Category
Distribution-level mains,
fixed installation
Local-level mains,
appliances, portable
equipment
Local-level mains,
appliances, portable
equipment
x

Standard Accessories
The following standard accessories ship with the probe
and are shown in the illustration above. The accessory
quantities m ay differ when multiple probes are shipped with
an oscilloscope/probe system.
Item
number Description Quantity
1 Hook tip 1 ea 013-0107-XX
2
3
4 Adjustment tool 1 ea 003-1433-XX
5
Reference lead
marker bands
Color
Instructions 1 ea 071-2335-XX
1 ea 196-3466-XX
f10
1seto
(5 pairs)
Part
number
016-1315-XX
Safety Summary
Review the following safety precautions to avoid injury and
prevent damage to this product or any products connected to it.
To avoid potential hazards, use this product only as specified.
Connect and Disconnect Properly. Connect the probe output
to the measurement instrument before connecting the probe
to the circuit under test. Disconnect the probe input and the
probe ground from the circuit under test before disconnecting
the probe from the measurement instrument.
Connect and Disconnect Properly. Do not connect or
disconnect probes or test leads while they are connected to a
voltage source.
Observe All Terminal Ratings. To avoid fire or shock hazard,
observe all ratings and markings on the product. Consult the
product manual for further ratings information before making
connections to the product. Connect the probe reference lead to
earth ground only.
To avoid electric shock when using probe accessories, never
exceed the lowest rating of the probe or its accessory, whichever
is less, including the measurement category and voltage rating.
Do not Operate Without Covers. Do not touch exposed
connections and components when power is present.
Avoid Exposed Circuitry. Do not touch exposed connections
and components when power is present.
Do Not Operate With Suspected Failures. If you suspect there is
damage to this product, have it inspected by qualified service
personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Safety Terms and Symbols Terms in This Manual.
These terms may appear in this manual:
WARNING. Warning statements identify conditions or
practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or
practices that could result in damage to this product or
other property.
Symbols on the Product. These symbols may appear on the
product:
Contacting Tektronix
Web site: www.tektronix.com
Phone: 1–800–833–9200
Address: Tektronix, Inc.
Email:
Department or name (if known)
14200 SW Karl Braun
Drive P.O. Box 500
Beaverton, OR 97077
USA
techsupport@tektronix.com
Warranty Information
For warranty information, go to www.tektronix.com/warranty
Copyright © Tektronix, Inc. All rights reserved. www.tektronix.com