Tektronix P2221 User manual

xx
Probe Compensation
Performance G raphs
Due to variations in oscilloscope input characteristics, the low-frequency compensation of the probe may need adjustment (in 10X attenuation only) after moving the probe from one oscilloscope channel to another.
If a 1 kHz c alibrated square wave displayed at 1 ms/division shows signicant differences between the leading and trailing edges, perform the following steps to optimize low-frequency compensation:
1. Connect the probe to the calibration signal on the oscillosco
pe front panel.
WARNING. To avoid electric shock, only connect to the
Probe Comp s
ignal on the oscilloscope when making this
adjustment.
P2221 200 MHz 1X/10X Passive Probe
Instruct
ions
2. Press AUTOS
ET or otherwise adjust your digitizing
oscilloscope to display a stable waveform.
3. Adjust the trimmer in the probe until you see a perfectly at-top square wave on the display. (See i llustration.)
WARNING. To avoid electric shock, only use the insulated
adjustment tool when making compensation adjustments.
Tab le 2 :
Characteristics Description
Temperature
Humidity
Altitude
Electrostatic Immunity The probe is not static sensitive
Environmental specications
Maximum operating
Minimum operating
Nonoperating
Operating
rating
Nonope
+50 °C (+122 °F)
0°C(+32°F)
–55°Cto+75°C (–67 °F to +167 °F)
10 to 95%
3.0 km (10,000 ft) maximum
15 km (50,000 ft) maximum
relative humidity
*P0712
33500*
071-2335-00
P2221 1X/10X Probe
The P2221 200 MHz 1X/10X Passive Probe is a c ompact passive probe with selectable 1X/10X attenuation.
The P2221 Probe has a compensation range of 10 to 25 pF, and is designed for use with Tektronix DPO/MSO2000 Series oscilloscopes.
The pr
WARNING. Do not oat the probes when used with
TDS1000/2000 Series Oscilloscopes.
NOTE. The P2221 Probe is not recommended for use with the
TPS2000 Series Oscilloscopes. Instead, use the P2220 1X/10X probe.
cting the Probe to the Oscilloscope
Conne
Connect the probe as shown in the illustration below.
WARNING. To avoid electric shock when using the probe,
keep ngers behind the guard on the probe body.
Attenuation Select Switch
This switch selects the attenuation factor of the probe.
Connecting the Probe to t he Circuit
to the back of these instructions for connection methods
Refer using the accessories included with the probe.
Specications
Table 1: Electrical and mechanical specications
Characteristic 10X position 1X position
Bandwidth
System Attenuation Accuracy
Compensation Range
System Input Resistance
System Input Capacitance
System Rise time (typical)
Maximum input voltage between tip (signal) and earth ground
Cable length
DC to 200 MHz (–3 dB)
10:1 ±2% 1:1 ±2%
10 pF-25 pF NA
10 M±3% at DC 1 M±3% at DC
17.0 pF 110 pF
2.2 ns 50 ns
10X position:
300 V CAT II 420 V
, <50% Duty factor, <1 s PW
peak
1X position:
150 V
CAT II or 150 VDCCAT II
rms
210 V
, <50% Duty factor, <1 s PW
peak
1.5 m
DC to 6 MHz (–3 dB)
-
ent Recycling. This product complies
Equipm
with the European Union’s requirements a ccording to Directive 2002/96/EC on waste electrical and electronic equipment (WEEE). For more information about recycling options, check the Support/Service section of the Tektronix Web site
tektronix.com).
(www.
3: Certications and compliances
Tab le
Characteristics Description
EC Declaration of
rmity
Confo
Measurement
gory
Cate Descriptions
Pollution Degree 2 Do not operate in environments where
Additional Safety
ndards
Sta
Compliance was demonstrated to
llowing specication as listed in
the fo the Ofcial Journal of the European Communities:
Low Voltage Directive 2006/95/EC:
EN61010-031: 2002
Category Examples of Products in
CAT III
CAT II
CAT I
conductive pollutants may be present (as dened in IEC 61010-1). Rated for indoor use only.
UL61010B-1, First Edition &
1010B-2-031, First Edition.
UL6
CAN/CSA-C22.2 No. 1010.1-92, & CAN/CSA-C22.2 No. 1010.2.031-94
IEC61010-031:2002
Pollution Degree 2
this Category
Distribution-level mains, xed installation
Local-level mains, appliances, portable equipment
Local-level mains, appliances, portable equipment
x
Standard Accessories
The following standard accessories ship with the probe and are shown in the illustration above. The accessory quantities m ay differ when multiple probes are shipped with an oscilloscope/probe system.
Item number Description Quantity
1 Hook tip 1 ea 013-0107-XX
2
3
4 Adjustment tool 1 ea 003-1433-XX
5
Reference lead
marker bands
Color
Instructions 1 ea 071-2335-XX
1 ea 196-3466-XX
f10
1seto (5 pairs)
Part number
016-1315-XX
Safety Summary
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specied.
Connect and Disconnect Properly. Connect the probe output to the measurement instrument before connecting the probe to the circuit under test. Disconnect the probe input and the probe ground from the circuit under test before disconnecting the probe from the measurement instrument.
Connect and Disconnect Properly. Do not connect or disconnect probes or test leads while they are connected to a voltage source.
Observe All Terminal Ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product. Connect the probe reference lead to earth ground only.
To avoid electric shock when using probe accessories, never exceed the lowest rating of the probe or its accessory, whichever is less, including the measurement category and voltage rating.
Do not Operate Without Covers. Do not touch exposed connections and components when power is present.
Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present.
Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualied service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Safety Terms and Symbols Terms in This Manual.
These terms may appear in this manual:
WARNING. Warning statements identify conditions or
practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or
practices that could result in damage to this product or other property.
Symbols on the Product. These symbols may appear on the product:
Contacting Tektronix
Web site: www.tektronix.com
Phone: 1–800–833–9200
Address: Tektronix, Inc.
Email:
Department or name (if known) 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
techsupport@tektronix.com
Warranty Information
For warranty information, go to www.tektronix.com/warranty
Copyright © Tektronix, Inc. All rights reserved. www.tektronix.com
Loading...