Getting Started with Option SST
Y ou can use this side of the Quick Reference to start to take
measurements with the Option SST Serial ATA and Serial
Attached SCSI Analysis modules that plug into the RT--Eye
Compliance and Analysis Solution. The other side contains
a complete menu tree for Option SST software.
NOTE. For complete operating instructions
and General Safety information, refer to the
online help for the application.
The Option SST Serial ATA and Serial Attached SCSI
Analysis modules transform a digital oscilloscope into an
analysis tool to test storage devices according to the Serial
Advanced Technology Attachment or SATA (Gen I and Gen
II) and Serial Attached SCSI (SAS) standards.
Performing a Differential Output Voltage Measurement
To measure the differential output voltage of a Serial ATA
device, follow these steps:
1. Select File> Run Application> RT--Eye Serial
Compliance and Analysis from the oscilloscope menu
bar. (For supported B-- series oscilloscopes, on the
oscilloscope menu bar , click App> RT-- Eye Serial
Compliance and Analysis).
2. Select Measurements> Serial ATA. Set the Test Point to
Transmitter, the Probe Type to Differential, and the
measurement to Differential Output Voltage.
3. Connect the probes to the device under test. For
example, the test setup using a dif ferential probe for
Differential Output Voltage is shown below.
4. Select Measurement> Configure. Select the Source tab.
Set the Test Method to BIST FIS/User. Set the channel
to Ch1.
5. Select the General Config tab. Set the Usage Model as
appropriate, Device Type to Drive, and set the Number
of UI and Cable Attn as appropriate. Set the Diff Volt
Option to Option1.
6. Push the “Running Man” button to start taking
measurements. Follow the on-screen prompts to input
different test patterns to complete the measurement.
When the measurement is completed successfully, the
application automatically displays the results. You can
also view the results by selecting Results> Summary or
Results> Details from the application menu bar.
7. The application displays results for the measurement.
8. To generate a report, select Utilities> Report Generator.
9. Select the template in the Generate Report tab.
10. Select the Generate button to post the test data to the
template.
For up-to-date information on Tektronix oscilloscope solutions for Serial ATA, access the
www.tektronix.com/Measurement/scopes/
Option SST Ordering Information
Supports the TDS6804B, TDS6604/B, TDS6000C,
TDS7704B series oscilloscopes; refer to the Optional
Applications Software on Windows-Based Oscilloscopes
Installation Manual for a complete list of supported
models. The applications CD includes a PDF file of the
installation manual.
If you order the Option SST along with a TDS6804B,
TDS6604/B, TDS6000C, TDS7704B series oscilloscope,
the Option SST Serial A TA and Serial Attached SCSI
Analysis Software will already be installed and enabled.
web page.
To order an upgrade for an existing oscilloscope:
H Order TDS6UP, TDS6BUP, or TDS7BUP* Opt. SST
H Opt. RTE: RT--Eye Serial Data Compliance and
Analysis Software
H Opt. JA3 -- Jitter and Timing Analysis Software
* Supported only on TDS7704B oscilloscope
Recommended Products and Accessories
Arbitrary Waveform Generator: A WG
610/615/710/710B/715B
High Bandwidth Dif ferential Probe: P7380, P7380SMA
SMA Cables: 174--1427--00
SMA Attenuators: 015--1002--01
TDR Measurement: TDS8200/80E04
Option SST
Test Solut i o n
Reference
www.tektronix.com
*P071155901*
071-1559-01

Option SST Menu Tree
File
Recall Default
Recall
Save
Recall Recent
Preferences
Minimize
Exit
Do not override the setup in BIST FIS/User method
Prompt for correct waveform confirmation during run
Refine filter initial condition for SATA AC CM
Overcome the filter transient response by discarding the fixed time constants of data
Modules
Measurements
Serial ATA
SAS
Select
Configure
Probe Type
Test Point
Device ID
Description
Timing
Amplitude
OOB**: COMWAKE, COMINIT, COMRESET
Source
General Config
Single--ended
Differential
Transmitter
Receiver
Eye
Bit Rate
Differential Skew*
Rise and Fall Time
Unit Interval
Select All
Differential Output Voltage
AC CM Voltage*
Test Method
Live
Trigger
Ref
File
Usage Model
Device Type
Cable Attn
Timing: Eye, Bit Rate, Differential Skew*, Rise
and Fall Time, Unit Interval , Select All
Amplitude: Differential Output Voltage
AC CM Voltage*
AWG | BIST FIS/User
Ch1Ch3 | Ch1Ch4 | Ch2Ch3 | Ch2Ch4
Ch1 | Chf2 | Chf3 | Ch4
Ref1Ref2 | Ref1Ref3 | Ref1Ref4 | Ref2Ref3 | Ref2Ref4 | Ref3Ref1 | Ref3Ref2
File Browser
Gen1i | Gen1x | Gen1m | Gen2i | Gen2x | Gen2m
Host | Drive
Results
Utilities
Help
Limits
Summary
Details
Worst Case Eye
Report Generator***
Help Topics
PDF Help
About RT--EYE
Clock Config
Plot Config
Limits File File Browser
De fine Test Template
Define Report Layout
Generate Report
OOB Type
Num of UI
Diff Volt Option
Clock Recovery
Analysis Window
Transition and Non--transition Bits
All Bits
New
Edit Template File Browser
Edit
New
Edit Template File Browser
Edit
New
Edit Template File Browser
Edit
In Spec | Out of Spec | Inter Burst
Option1 | Option2
Window Size
Scan Incr
Method
250 UI | 5 UI
Const Clk: Mean | Const Clk: Median
* Added when the Probe Type is Single ended.
** Added only when the Test Point is Transmitter.
*** The Report Generator utility includes online help specific to the use of the Report Generator Editor.
Copyright © Tektronix, Inc.