Tektronix Option 21 Field Install Instructions

Instructions
RSA6100A Series Real-Time Spectrum Analyzers
General Purpose Modulation Analysis (Option 21) Upgrade
071-1921-00
Warning
www.tektronix.com
*P071192100*
071192100
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered tradem arks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
H In North America, call 1-800-833-9200.
H Worldwide, visit www.tektronix.com to find contacts in your area.
Kit Description
Products
Kit Parts List
This kit describes the installation of the General Purpose Modulation Analysis Upgrade (Option 21) in an RSA6100A Series Real-Time Spectrum Analyzer.
RSA6100A Series AllSerialNumbers
Quantity Part number Description
1ea 071-1921-00 MANUAL,TECH:INSTALLATION,RSA6UP, OPT21
1ea NS DATA SHEET; SOFTWARE OPTION KEY AUTHORIZATION
CERTIFICATE, UPGRADE KITS
1ea NS LABEL, MANUFACTURED; OPTION KEY UPGRADE LABEL
2.100 X 2.700, SAFETY CONTROLLED
NS Not Saleable
RSA6100A Series General Purpose Modulation Analysis Installation (Option 21)
1
Kit Description
2
RSA6100A Series General Purpose Modulation Analysis Installation (Option 21)
Installation Instructions
The software that performs General Purpose Modulation Analysis is included in all spectrum analyzer application installations, but you need an option installa­tion key to activate it. To activate the General Purpose Modulation Analysis capability, you will enter the new option installation key supplied with this kit into the spectrum analyzer application.
Install Option Key
To activate the General Purpose Modulation Analysis option:
1. In the spectrum analyzer application, select Tools > Install Upgrades.The Install Upgrade wizard starts, as shown in Figure 1. Click Continue.
Figure 1: Starting the option upgrade
2. The Install Upgrade Enter Key screen appears (see Figure 2).
RSA6100A Series General Purpose Modulation Analysis Installation (Option 21)
3
Installation Instructions
Figure 2: Entering the option installation key
3. Type the option installation key into the empty text box and click Continue.
An Installation Success dialog opens, as shown in Figure 3.
Figure 3: Upgrade success notification
4. Click Close, and then exit and restart the RSA6100A application, to enable
the new feature.
5. Press the Displays button on the RSA6100A.
4
RSA6100A Series General Purpose Modulation Analysis Installation (Option 21)
Install Option Key Label
Installation Instructions
6. In the Folders panel, verify that you can now select the GP Digital Modula­tion folder.
This completes the installation of the Option Installation Key.
H Place the new option key label over the existing label on the instrument rear
panel.
This completes the installation of the General Purpose Modulation Analysis (Option 21) upgrade kit.
g End of document g
RSA6100A Series General Purpose Modulation Analysis Installation (Option 21)
5