Warning: The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified
to do so. Refer to all safety summaries prior to performing service.
Supports Product Firmware V1.28 and above
Register now!
Click the following link to protect your product.
www.tek.com/register
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
•In North America, call 1-800-833-9200.
•Worldwide, visit www.tek.com to find contacts in your area.
Table of Contents
Table of Contents
List of Figures................................................................................................................................................................................5
List of Tables................................................................................................................................................................................. 6
Important safety information..........................................................................................................................................................7
General safety summary........................................................................................................................................................7
To avoid fire or personal injury........................................................................................................................................7
Probes and test leads..................................................................................................................................................... 8
Service safety summary.........................................................................................................................................................8
Terms in the manual....................................................................................................................................................... 9
Terms on the product............................................................................................................................................................. 9
Symbols on the product......................................................................................................................................................... 9
Analog inputs....................................................................................................................................................................... 10
Arbitrary function generator................................................................................................................................................. 36
Digital volt meter (DVM).......................................................................................................................................................38
Input/Output port specifications........................................................................................................................................... 40
Data storage specifications..................................................................................................................................................43
Power supply system........................................................................................................................................................... 45
Test records......................................................................................................................................................................... 51
Instrument information, self test record.........................................................................................................................51
DC Offset Accuracy test record.................................................................................................................................... 52
Analog Bandwidth test record.......................................................................................................................................56
Input impedance test record......................................................................................................................................... 60
DC Gain Accuracy test record...................................................................................................................................... 60
Random Noise, sample acquisition mode test record...................................................................................................64
Random Noise, High Res mode test record............................................................................................................... 120
Long term sample rate through AFG DC offset accuracy test records....................................................................... 226
Check DC gain accuracy................................................................................................................................................... 241
6 Series B Mixed Signal Oscilloscopes Specifications and Performance Verification3
Table of Contents
Check DC offset accuracy..................................................................................................................................................243
Check analog bandwidth....................................................................................................................................................244
Check random noise, sample acquisition mode (10, 8, and 6 GHz options)..................................................................... 245
Check random noise, High Res mode............................................................................................................................... 247
Check long term samples rate and delay time accuracy....................................................................................................248
Check digital threshold accuracy....................................................................................................................................... 249
Check AUX Out output voltage levels................................................................................................................................250
Check DVM voltage accuracy (DC)................................................................................................................................... 251
Check DVM voltage accuracy (AC)................................................................................................................................... 252
Check trigger frequency accuracy and maximum input frequency.................................................................................... 253
Arbitrary function generator............................................................................................................................................... 254
Check AFG sine and ramp frequency accuracy......................................................................................................... 254
Check AFG square and pulse frequency accuracy.....................................................................................................255
Check AFG signal amplitude accuracy.......................................................................................................................256
Check AFG DC offset accuracy..................................................................................................................................257
Figure 6: DC offset tests............................................................................................................................................................258
6 Series B Mixed Signal Oscilloscopes Specifications and Performance Verification5
List of Tables
List of Tables
Table 1: Gain expected worksheet............................................................................................................................................241
6 Series B Mixed Signal Oscilloscopes Specifications and Performance Verification6
Important safety information
Important safety information
This manual contains information and warnings that must be followed by the user for safe operation and to keep the product in a safe
condition.
To safely perform service on this product, see the Service safety summary that follows the General safety summary.
General safety summary
Use the product only as specified. Review the following safety precautions to avoid injury and prevent damage to this product or any
products connected to it. Carefully read all instructions. Retain these instructions for future reference.
This product shall be used in accordance with local and national codes.
For correct and safe operation of the product, it is essential that you follow generally accepted safety procedures in addition to the safety
precautions specified in this manual.
The product is designed to be used by trained personnel only.
Only qualified personnel who are aware of the hazards involved should remove the cover for repair, maintenance, or adjustment.
Before use, always check the product with a known source to be sure it is operating correctly.
This product is not intended for detection of hazardous voltages.
Use personal protective equipment to prevent shock and arc blast injury where hazardous live conductors are exposed.
To avoid fire or personal injury
Use proper power cordUse only the power cord specified for this product and certified for the country of use.
Ground the productThis product is grounded through the grounding conductor of the power cord. To avoid
electric shock, the grounding conductor must be connected to earth ground. Before making
connections to the input or output terminals of the product, ensure that the product is properly
grounded. Do not disable the power cord grounding connection.
Power disconnectThe power cord disconnects the product from the power source. See instructions for the
location. Do not position the equipment so that it is difficult to operate the power cord; it must
remain accessible to the user at all times to allow for quick disconnection if needed.
Connect and disconnect properlyDo not connect or disconnect probes or test leads while they are connected to a voltage
source.
Use only insulated voltage probes, test leads, and adapters supplied with the product, or
indicated by Tektronix to be suitable for the product.
Observe all terminal ratingsTo avoid fire or shock hazard, observe all rating and markings on the product. Consult the
product manual for further ratings information before making connections to the product. Do
not exceed the Measurement Category (CAT) rating and voltage or current rating of the lowest
rated individual component of a product, probe, or accessory. Use caution when using 1:1 test
leads because the probe tip voltage is directly transmitted to the product.
Do not apply a potential to any terminal, including the common terminal, that exceeds the
maximum rating of that terminal.
Do not operate without coversDo not operate this product with covers or panels removed, or with the case open. Hazardous
voltage exposure is possible.
Avoid exposed circuitryDo not touch exposed connections and components when power is present.
Do not operate with suspected
failures
If you suspect that there is damage to this product, have it inspected by qualified service
personnel.
6 Series B Mixed Signal Oscilloscopes Specifications and Performance Verification7
Important safety information
Disable the product if it is damaged. Do not use the product if it is damaged or operates
incorrectly. If in doubt about safety of the product, turn it off and disconnect the power cord.
Clearly mark the product to prevent its further operation.
Before use, inspect voltage probes, test leads, and accessories for mechanical damage and
replace when damaged. Do not use probes or test leads if they are damaged, if there is
exposed metal, or if a wear indicator shows.
Examine the exterior of the product before you use it. Look for cracks or missing pieces.
Use only specified replacement parts.
Do not operate in wet/damp
conditions
Do not operate in an explosive
atmosphere
Keep product surfaces clean and dry Remove the input signals before you clean the product.
Provide proper ventilationRefer to the installation instructions in the manual for details on installing the product so it has
Provide a safe working environmentAlways place the product in a location convenient for viewing the display and indicators.
Be aware that condensation may occur if a unit is moved from a cold to a warm environment.
proper ventilation.
Slots and openings are provided for ventilation and should never be covered or otherwise
obstructed. Do not push objects into any of the openings.
Avoid improper or prolonged use of keyboards, pointers, and button pads. Improper or
prolonged keyboard or pointer use may result in serious injury.
Be sure your work area meets applicable ergonomic standards. Consult with an ergonomics
professional to avoid stress injuries.
Use care when lifting and carrying the product. This product is provided with a handle or
handles for lifting and carrying.
Use only the Tektronix rackmount hardware specified for this product.
Probes and test leads
Before connecting probes or test leads, connect the power cord from the power connector to a properly grounded power outlet.
Keep fingers behind the protective barrier, protective finger guard, or tactile indicator on the probes. Remove all probes, test leads and
accessories that are not in use.
Use only correct Measurement Category (CAT), voltage, temperature, altitude, and amperage rated probes, test leads, and adapters for
any measurement.
Service safety summary
The Service safety summary section contains additional information required to safely perform service on the product. Only qualified
personnel should perform service procedures. Read this Service safety summary and the General safety summary before performing any
service procedures.
To avoid electric shockDo not touch exposed connections.
Do not service aloneDo not perform internal service or adjustments of this product unless another person capable of
rendering first aid and resuscitation is present.
Disconnect powerTo avoid electric shock, switch off the product power and disconnect the power cord from the
mains power before removing any covers or panels, or opening the case for servicing.
6 Series B Mixed Signal Oscilloscopes Specifications and Performance Verification8
Important safety information
Use care when servicing with poweronDangerous voltages or currents may exist in this product. Disconnect power, remove battery
(if applicable), and disconnect test leads before removing protective panels, soldering, or
replacing components.
Verify safety after repairAlways recheck ground continuity and mains dielectric strength after performing a repair.
Terms in the manual
These terms may appear in this manual:
Warning: Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION: Caution statements identify conditions or practices that could result in damage to this product or other property.
Terms on the product
These terms may appear on the product:
•DANGER indicates an injury hazard immediately accessible as you read the marking.
•WARNING indicates an injury hazard not immediately accessible as you read the marking.
•CAUTION indicates a hazard to property including the product.
Symbols on the product
When this symbol is marked on the product, be sure to consult the manual to find out the nature of the potential hazards and
any actions which have to be taken to avoid them. (This symbol may also be used to refer the user to ratings in the manual.)
The following symbols(s) may appear on the product.
6 Series B Mixed Signal Oscilloscopes Specifications and Performance Verification9
Specifications
Specifications
This chapter contains specifications for the instrument. All specifications are guaranteed unless noted as "typical." Typical specifications
are provided for your convenience but are not guaranteed. Specifications that are marked with the ✓ symbol are checked in this manual.
All specifications apply to all models unless noted otherwise.
To meet specifications, these conditions must first be met:
•The instrument must have been calibrated in an ambient temperature between 18 °C and 28 °C (64 °F and 82 °F).
•The instrument must be operating within the environmental limits. (See Environmental specifications on page 45)
•The instrument must be powered from a source that meets the specifications. (See Power supply system on page 45)
•The instrument must have been operating continuously for at least 20 minutes within the specified operating temperature range.
•You must perform the Signal Path Compensation procedure after the warmup period. See the online help for instructions on how to
perform signal path compensation. If the ambient temperature changes more than 5 °C (9 °F), repeat the procedure.
Analog inputs
Number of analog input channelsMSO64B: 4
MSO66B: 6
MSO68B: 8
Input couplingDC, AC
Input resistance selection1 MΩ or 50 Ω
250 KΩ selectable for Performance Verification
✓DC Input Resistance, 50 Ω, DC
coupled
✓DC Input Resistance, 1MΩ DCCoupled
Input VSWR, 50 Ω DC-coupled, typical
Maximum input voltage, 50 Ω2.3 V
Maximum input voltage 1 MΩ DCcoupled
50 Ω ±3%
1 MΩ ±1%
Input Frequency
≤2.5 GHz1.41.2
>2.5 GHz and ≤6 GHz1.51.3
>6 GHz and ≤9.5 GHz1.91.8
>9.5 GHz and ≤10 GHz2.11.9
, at <100 mV/div, with peaks ≤±20 V (Pulse Width ≤ 1 μs)
RMS
5.5 V
300 V
Maximum peak input voltage at the BNC, ±425 V
, at ≥ 100 mV/div, with peaks ≤±20 V (Pulse Width ≤ 200 μs)
RMS
, DC to 10 kHz
RMS
VSWR <100 mV/div
VSWR ≥ 100 mV/div
Sensitivity range, coarse
1 MΩ500 µV/div to 10 V/div in a 1-2-5 sequence
50 Ω1 mV/div to 1 V/div in a 1-2-5 sequence
6 Series B Mixed Signal Oscilloscopes Specifications and Performance Verification10
Sensitivity range (Fine)
1 MΩAllows continuous adjustment from 500µV/div to 10V/div
50 ΩAllows continuous adjustment from 1mV/divto 1V/div
Sensitivity Resolution (Fine)≤1% of coarse sensitivity range setting
Specifications
Input capacitance 1 MΩ DC coupled,
typical
14.5 pF ±1.5 pF
Analog DC
Maximum offset ranges,Input signal cannot exceed maximum input voltage for the 50 Ω input path.
Volts/div SettingMaximum offset range, 50 Ω Input
1 mV/div - 99 mV/div±1 V
100 mV/div - 1 V/div±10 V
Volts/div SettingMaximum offset range, 1 MΩ Input
500 µV/div - 63 mV/div±1 V
DC voltage measurement accuracy,
Average acquisition mode
64 mV/div - 999 mV/div
1 V/div - 10 V/div±100 V
Input Signal cannot exceed max input voltage for the 50 Ω input path.
INL @ ≤ 2 mV/div: ±20 DL's (12-bit reference)
DNL: ±1.0 DL's (12-bit digitizing scale) when oscilloscope is in Hi-Res mode.
Analog AC
✓Analog bandwidth 50 Ω DC coupled
ModelVolts/div SettingBandwidth
MSO6XB BW-100001 mV/div – 1V/divDC - 10GHz
MSO6XB BW-80001 mV/div – 1V/divDC – 8 GHz
MSO6XB BW-60001 mV/div – 1V/divDC – 6 GHz
MSO6XB BW-40001 mV/div – 1V/divDC – 4 GHz
MSO6XB BW-25001 mV/div – 1V/divDC – 2.5 GHz
MSO6XB BW-10001 mV/div – 1V/divDC – 1 GHz
✓Analog Bandwidth, 1 MΩ
The limits are for ambient temperature of ≤30 °C and the bandwidth selection set to FULL.
Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
MSO6XB, all models:
Volts/Div SettingBandwidth
1 mV/div – 10 V/divDC – 500 MHz
500 µV/div – 995 µV/divDC – 250 MHz
1
Immediately following SPC, add 2% for every 5 °C change in ambient.
2
Immediately following SPC, add 1% for every 5 °C change in ambient.
6 Series B Mixed Signal Oscilloscopes Specifications and Performance Verification12
Specifications
Analog bandwidth TPP1000 10X
probe
The limits are for ambient temperature of ≤30 °C and the bandwidth selection set to FULL.
Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
1M Ohm20 MHz (HW), 200 MHz, 250 MHz (HW), 350 MHz, and Full (500 MHz)
Frequency response tolerance/
flatness, 50 Ω
± 0.5 dB from DC to 80% of rated bandwidth up to 8 GHz instruments
± 0.5 dB from DC To 65% of rated bandwidth for 10 GHz instruments
Combined TDP7700 and 6 Series B
MSO flatness, typical
±0.6 dB from DC to 80% of nominal BW when used with P77C292MM (SMA Probe Tip)
Not valid while using peak detect or envelope mode. Valid for probe modes A, B, and D
Phase accuracy±2.5 degrees, typical out to 9 GHz
Lower frequency limit, AC coupled,
typical
<10 Hz when AC 1 MΩ coupled. The AC coupled lower frequency limits are reduced by a factor
of 10 (<1 Hz) when 10X passive probes are used.
Upper frequency limit, 250 MHz bandwidth limited, typical
50 Ω, DC-coupled250 MHz, ± 5%
1 MΩ, DC-coupled250 MHz, ± 25%
Upper frequency limit, 20 MHz bandwidth limited, typical
50 Ω, DC-coupled20 MHz, ± 5%
1 MΩ, DC-coupled20 MHz, ± 25%
Calculated rise timeThe formula used is 0.4/BW where BW is the measured –3 dB bandwidth of the oscilloscope.
The formula accounts for the rise time contribution of the oscilloscope independent of the rise
time of the signal source.
Calculated Rise Time (10% to 90%)
3
3
Below specification is independent of oscilloscope model and is dependent on bandwidth option only.
6 Series B Mixed Signal Oscilloscopes Specifications and Performance Verification13
Effective bits (ENOB), typical
These limits apply to:
•Fastacq turned OFF
•8 channel box: ch1, ch5
•6 channel box: ch1, ch4
•4 channel box: ch1, ch3
Specifications
Model50 ΩTPP1000 Probe
1 mV-1 V5 mV-10 V
MSO6X BW-1000040 ps400 ps
MSO6X BW-800050 ps400 ps
MSO6X BW-600066.67 ps400 ps
MSO6X BW-4000100 ps400 ps
MSO6X BW-2500160 ps400 ps
MSO6X BW-1000400 ps400 ps
50 mV/div, 50 GS, Sample mode, 50 ohm,
TYP
Channel bandwidth
Frequ
ency10GHz
10
MHz
250
MHz
1 GHz 6.66.756.8577.17.37.457.77.87.958.3
2 GHz 6.556.656.756.8577.27.357.557.657.75
4 GHz 6.456.656.756.957.057.27.35
7 GHz 6.556.656.756.9
2 mV/div, 50 GS, Sample mode, 50 ohm, TYP 2 mV/div, 25 GS, HiRes mode, 50 ohm, TYP
Phase noise at 1 GHz, typical-118 dBc/Hz 10 kHz offset
Absolute amplitude accuracy, typical ±1 dB (0-8 GHz) for max 10 GHz BW.
Noise figure, typical1 mv/div, 11 dB 10 MHz to 6 GHz and 14 dB 6 GHz to 10 GHz
SNR/Dynamic Range, typical112 dB with a 1 GHz signal in a 100 MHz span with 1 kHz RBW ± 20 MHz from the carrier.
2nd harmonic distortion at -58 dBc with a 0 dBm 1 GHz signal.
3rd harmonic distortion at -55 dBc with a 0 dBm 1 GHz signal.
-119 dBc/Hz 100 kHz offset
-132 dBc/Hz 1 MHz offset
-140 dBc/Hz 10 MHz offset
Skew and delay
Digital skew, typical
6 Series B Mixed Signal Oscilloscopes Specifications and Performance Verification21
Digital-to-Analog skew1 ns
Digital-to-Digital skew± 320 ps from bit 0 of any TekVPI + channel to bit 0 of any TekVPI+ channel.
Digital skew within a FlexChannel < 200 ps within any TEKVPI + channel
Specifications
Delay between analog channels, full
bandwidth, typical
Deskew range and resolution-125 ns to +125 ns with a resolution of 40 ps (for Peak Detect and Envelope acquisition
Acquisition ModesSample, Peak Detect, High Res, Envelope, Average, Fast Frame
Number of digitized bits
≤ 10 ps for any two channels with input impedance set to 50 Ω, DC coupling with equal
Volts/div or above 10 mV/div
modes).
-125 ns to +125 ns with a resolution of 1 ps (for all other acquisition modes).
MSO64B
channels 1 and 3
(2 and 4 off)
MSO66B
channels 1 and 4
(2, 3, 5, and 6
off)
MSO68B
channels 1 and 5
(2, 3, 4, 6, 7, and
8 off)
Sample RateAcquisition
Mode
50 GS/sSample810 GHz
25 GS/sHi Res125 GHz
12.5 GS/sHi Res125 GHz
6.25 GS/sHi Res132.5 GHz
3.125 GS/sHi Res141 GHz
1.5625 GS/sHi Res15500 MHz
625 MS/sHi Res16200 MHz
Digitized BitsChannel
Bandwidth
Peak Detect or Envelope Mode
Minimum Detectable Pulse, typical
MSO64B all
channels
MSO66B
channels 1, 2, 4,
and 5 (3 and 6
off)
MSO68B
channels 1, 2, 5,
and 6 (3, 4, 7,
and 8 off)
MSO66B all
channels
MSO68B all
channels
25 GS/sSample810 GHz
12.5 GS/sSample1210 GHz
12.5 GS/sHi Res125 GHz
6.25 GS/sHi Res132.5 GHz
3.125 GS/sHi Res141 GHz
1.5625 GS/sHi Res15500 MHz
625 MS/sHi Res16200 MHz
12.5 GS/sSample85 GHz
6.25 GS/sSample125 GHz
3.125 GS/sHi Res131 GHz
1.5625 GS/sHi Res14500 MHz
1.25 GS/sHi Res15500 MHz
312.5 MS/sHi Res16100 MHz
6 Series B Mixed Signal Oscilloscopes Specifications and Performance Verification22
Specifications
ChannelsSample RateMinimum Pulse Width
MSO64B channels 1 and 3 (2
and 4 off)
MSO66B channels 1 and 4 (2,
3, 5, and 6 off)
MSO68B channels 1 and 5
(2,3, 4, 6, 7, and 8 off)
50 GS/s160 ps
25 GS/s160 ps
12.5 GS/s320 ps
Number of waveforms for average
acquisition mode
MSO64B channels 2 and 4
MSO66B channels 2 and 5 (3
and 6 off)
MSO68B channels 2 and 6 (3,
4, 7 and 8 off)
MSO66B channels 3 and 6
MSO68B channels 3, 4, 7,
and 8
2 to 10,240 Waveforms, default 16 waveforms
TekVPI interface
TekVPI interconnectAll analog channel inputs on the front panel conform to the TekVPI+ specification defined in the
TEKPROBE, TEKCONNECT, AND TEKVPI STANDARDS specification.
Total Probe Power
80 W maximum
MSO64B40 W maximum for channels 1 through 2 and 40 W maximum for channels 3 through 4.
MSO66B40 W maximum for channels 1 through 3 and 40 W maximum for channels 4 through 6.
MSO68B40 W maximum for channels 1 through 4 and 40 W maximum for channels 5 through 8.
Probe Power per Channel5 V Supply: 300 mW (60 mA max)
12 V Supply: 20 W (1.67 A max)
Low-C Passive Probe SupportSupports TPP1000 and similar probes
Digital Probe SupportSupports TLP058 VPI+ digital probes
Probe User InterfaceProbe setup menu
Probe menu button support (opens probe setup menu)
Probe warning messages and indicators
Dynamic range indication
Probe bandwidth limiting
6 Series B Mixed Signal Oscilloscopes Specifications and Performance Verification23
Specifications
Timebase system
✓Timebase factory toleranceFrequency tolerance at factory calibration is ±12 ppb (parts per billion).
At Calibration, 25 °C ambient, over any ≥1 ms interval.
Timebase temperature stability±20 PPB across the full operating range of 0 °C to 50 °C, after a sufficient soak time at the
temperature.
The instrument needs to soak at a fixed temperature for an extended period of time to insure
the time-base frequency is stable. The following is a worst case estimation for the frequency
error versus the amount of time the instrument has been soaking at a temperature.Max error (in
ppb) = ±10^[log[100/soak time(in hours)]]
For example, a 1 hour soak will have a max frequency error of ±100 ppb, but a 10 hour soak
will have a max frequency error of 10 ppb.
±300 PPB/Year, and will not exceed ±2 PPM over 10 years without calibration.
Calibration will reduce this frequency error to under ±12 PPB
Frequency tolerance change at 25 °C over a periods of 1 year and 10 years.
8 channels6.25 S/s to 12.5 GS/s25 GS/s to 2.5 TS/s
On 4-channel models, the 2 channels with 50 GS/s capability are 1 and 3 (channels 2 and 4
must be off).
On 6-channel models, the2 channels with 50 GS/s capability are 1 and 4 (channels 2, 3, 5, and
6 must be off).
On 8-channel models, the 2 channels with 50 GS/s capability are 1 and 5 (channels 2, 3, 4, 6,
7, and 8 must be off).
On 6-channel models, the 4 channels with 25 GS/s capability are 1,2, 4 and 5 (channels 3 and
6 must be off).
On 8-channel models, the 4 channels with 25 GS/s capability are 1, 2, 5, and 6 (channels 3, 4,
7, and 8 must be off).
ChannelsSample rate
2 channelsup to 25 GS/s
4 channelsup to 12.5 GS/s
8 channelsup to 6.25 GS/s
On 4-channel models, the 2 channels with 25 GS/s capability are 1 and 3 (channels 2 and 4
must be off).
On 6-channel models, the 2 channels with 25 GS/s capability are 1 and 4 (channels 2, 3, 5, and
6 must be off).
6 Series B Mixed Signal Oscilloscopes Specifications and Performance Verification24
Specifications
On 8-channel models, the 2 channels with 25 GS/s capability are 1 and 5 (channels 2, 3, 4, 6,
7, and 8 must be off).
On 6-channel models, the 4 channels with 12.5 GS/s capability are 1, 2, 4, and 5 (channels 3
and 6 must be off).
On 8-channel models, the 4 channels with 12.5 GS/s capability are 1, 2, 5 and 6 (channels 3, 4,
7, and 8 must be off).
Interpolated waveform rate (Sample
mode):
Interpolated waveform rate (High Res
mode):
Record length rangeApplies to analog and digital channels. All acquisition modes are 1 G maximum record length,
•2.5 TS/s
•1 TS/s
•500 GS/s
•250 GS/s
•100 GS/s
•50 GS/s (only with a 25 GS/s or12.5 GS/s channel on)
•25 GS/s (only with a 12.5 GS/s channel on)
•2.5 TS/s
•1 TS/s
•500 GS/s
•250 GS/s
•100 GS/s
•50 GS/s
•25 GS/s (only with a 25 GS/s or12.5 GS/s channel on)
•12.5 GS/sec (only with a 12.5 GS/s channel on)
down to 1 k minimum record length, adjustable in 1 sample increments.
Standard: 62.5 Mpoints
Option 6-RL-1: 125 Mpoints
Option 6-RL-2: 250 Mpoints
Option 6-RL-3: 500 Mpoints
Option 6-RL-4: 1 Gpoints
Horizontal scale range40 ps/div to 1000 s/div
The minimum horizontal scale is determined by the record length by dividing the record length
by 10 (because there are 10 divisions on-screen) and then dividing by the maximum sample
rate (2.5 TS/s).
40 ps/div can only be achieved with a 2.5 TS/s sample rate (maximum) and a 1000 point record
length (minimum).The table below shows minimum horizontal scales for a collection of record
lengths.
Record lengthMinimum Horizontal Scale
1 kS40 ps/div
10 kS400 ps/div
100 kS4 ns/div
Table continued…
6 Series B Mixed Signal Oscilloscopes Specifications and Performance Verification25
Specifications
Record lengthMinimum Horizontal Scale
1 MS40 ns/div
10 MS400 ns/div
62.5 MS2.5 us/div
125 MS with optional memory length5 us/div
250 MS with optional memory length10 us/div
500 MS with optional memory length20 us/div
1 GS with optional memory length40 us/div
1000 s/div is the maximum horizontal scale which limits the acquisition length to 10000 s (2
hours, 46 minutes and 40 seconds).
Below a record length of 62.5 kS, the horizontal scale is further limited. The maximum
horizontal scale can be calculated by dividing the record length by 10 (because there are 10
division on-screen) and then dividing by the minimum sample rate (6.25 S/s).The table below
shows maximum horizontal scales for a collection of record length.
Record lengthMaximum Horizontal Scale
1 kS16 s/div
5 kS80 s/div
10 kS160 s/div
25 kS400 s/div
50 kS800 s/div
62.5 kS1000 s/div
Sample jitter (Aperture uncertainty),
typical
Time durationTypical jitter
<1 μs80 fs
<1 ms130 fs
Delta-time measurement accuracy, typical
Delta-time measurement accuracy,
typical
Formula for DTA:SR1 = Slew Rate (1st Edge) around 1st point in measurement
SR2 = Slew Rate (2nd Edge) around 2nd point in measurement
N = RSS of input-referred noise (volts rms) and Dynamic noise estimate (volts rms).
Dynamic noise estimate*
Note: Dynamic noise is noise that appears with a signal applied (such as distortion or
interleave errors).
tj = aperture uncertainty (sec rms—80fs for short durations)
TBA = timebase accuracy or reference frequency error (which is 20ppb)
6 Series B Mixed Signal Oscilloscopes Specifications and Performance Verification26
tp = delta-time measurement duration (sec)
Specifications
Delta Time Measurement Accuracy
(DTA), reference example
Trigger holdoff range0 ns minimum to 10 seconds maximum
Trigger level ranges, typical
Time Range for Glitch, Pulse Width,
Timeout, Time-qualified Runt or Timequalified Window, Transition Time
Trigger
Setup/Hold Violation Trigger, Setup
and Hold Time Ranges
SourceRange
Analog Inputs±5 divs from center of screen
LineN/A
AUX Input±5 V
40 ps to 20 s
FeatureMinMax
Setup Time0 ns20 s
Hold Time0 ns20 s
Setup + Hold Time80 ps22 s
Input coupling on clock and data channels must be the same.
For Setup Time, positive numbers mean a data transition before the clock.
For Hold Time, positive numbers mean a data transition after the clock edge.
Setup + Hold Time is the algebraic sum of the Setup Time and the Hold Time programmed by
the user.
Oscilloscopes Trigger position is equal to the Hold Time value.
Trigger jitter, typicalAnalog Inputs
•Trigger jitter ≤ 1.5 ps RMS for Sample mode, Edge-type trigger, FastAcq, and Pulse width
modes
•Trigger jitter ≤ 40 ps RMS for non-Edge-type trigger modes
6 Series B Mixed Signal Oscilloscopes Specifications and Performance Verification29
Specifications
Trigger Bandwidth – Edge, Pulse,
Logic, typical
Trigger level accuracy, DC-coupled,
typical
ModelInstrument
bandwidth
MSO6XB10 GHz10 GHz4 GHz
MSO6XB8 GHz8 GHz4 GHz
MSO6XB6 GHz6 GHz4 GHz
MSO6XB4 GHz, 2.5 GHz, 1
GHz
For signals having rise and fall times ≥10 ns, the limits are as follows:
SourceRange
Any Input Channel±0.20 div
LineN/A
This limit is checked by SPC at very low frequency (nearly DC).
This limit does not include frequency dependent effects, edge type trigger sensitivity not DC
coupled, or trigger position error.
Set the trigger level to the desired value. Using an adjustable DC source, inject a voltage into
the instrument. Adjust the voltage downward (if checking negative slope) or upward (if checking
positive slope) until the scope triggers. The difference between the trigger level setting and the
voltage that actually caused the trigger is the trigger level accuracy.