The servicing instructions are for use by qualified personnel only. To avoid
personal injury, do not perform any servicing unless you are qualified to do
so. Refer to all safety summaries prior to performing service.
Supports product Firmware V1.4 and above. Revision A; released January
13, 2020.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
FlexChannel, e*Scope, and TekVPI are registered trademarks of Tektronix, Inc.
5 Series MSO Specifications and Verification manual. Released 20180621
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
■
In North America, call 1-800-833-9200.
■
Worldwide, visit www.tek.com to find contacts in your area.
Table of Contents
Important safety information................................................................................................................................iii
General safety summary................................................................................................................................iii
Terms in the manual.......................................................................................................................................v
Terms on the product.....................................................................................................................................v
Symbols on the product..................................................................................................................................v
Specifications
Analog channel input and vertical specification.............................................................................................1
Serial Trigger specifications.........................................................................................................................19
Digital acquisition system.............................................................................................................................22
Digital volt meter (DVM)...............................................................................................................................22
Trigger frequency counter............................................................................................................................23
Arbitrary Function Generator system...........................................................................................................23
Display system (MSO54, MSO56, MSO58).................................................................................................25
Input/Output port specifications....................................................................................................................26
Data storage specifications..........................................................................................................................28
Power supply system...................................................................................................................................29
Test Records................................................................................................................................................32
Instrument information, self test record..................................................................................................32
Input Impedance test record...................................................................................................................32
DC Balance test record..........................................................................................................................33
DC Gain Accuracy test record................................................................................................................41
DC Offset Accuracy test record..............................................................................................................52
Analog Bandwidth test record................................................................................................................56
Random Noise, sample acquisition mode test record............................................................................89
Random Noise, High Res mode test record.........................................................................................121
Long term sample rate through AFG DC offset accuracy test records.................................................152
Check DC balance.....................................................................................................................................177
Check DC gain accuracy............................................................................................................................179
Check DC offset accuracy..........................................................................................................................181
Check analog bandwidth............................................................................................................................183
Check random noise, sample acquisition mode.........................................................................................187
Check random noise, High Res mode.......................................................................................................189
Check long term sample rate.....................................................................................................................192
Check delta time measurement accuracy..................................................................................................193
Check digital threshold accuracy...............................................................................................................195
Check AUX Out output voltage levels........................................................................................................197
Check DVM voltage accuracy (DC)...........................................................................................................198
Check DVM voltage accuracy (AC)............................................................................................................199
Check trigger frequency accuracy and maximum input frequency............................................................200
Arbitrary function generator........................................................................................................................201
Check AFG sine and ramp frequency accuracy...................................................................................201
Check AFG square and pulse frequency accuracy..............................................................................202
Check AFG signal amplitude accuracy................................................................................................203
Check AFG DC offset accuracy...........................................................................................................204
ii5 Series MSO Specifications and Performance Verification
Important safety information
This manual contains information and warnings that must be followed by the user for safe operation and to keep the product in a
safe condition.
General safety summary
Use the product only as specified. Review the following safety precautions to avoid injury and prevent damage to this product or
any products connected to it. Carefully read all instructions. Retain these instructions for future reference.
This product shall be used in accordance with local and national codes.
For correct and safe operation of the product, it is essential that you follow generally accepted safety procedures in addition to
the safety precautions specified in this manual.
The product is designed to be used by trained personnel only.
Only qualified personnel who are aware of the hazards involved should remove the cover for repair, maintenance, or adjustment.
Before use, always check the product with a known source to be sure it is operating correctly.
This product is not intended for detection of hazardous voltages.
Use personal protective equipment to prevent shock and arc blast injury where hazardous live conductors are exposed.
To avoid fire or personal injury
Use proper power cord. Use only the power cord specified for this product and certified for the country of use. Do not use the
provided power cord for other products.
Ground the product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the
grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the
product, ensure that the product is properly grounded. Do not disable the power cord grounding connection.
Power disconnect. The power cord disconnects the product from the power source. See instructions for the location. Do not
position the equipment so that it is difficult to operate the power cord; it must remain accessible to the user at all times to allow for
quick disconnection if needed.
Connect and disconnect properly. Do not connect or disconnect probes or test leads while they are connected to a voltage
source. Use only insulated voltage probes, test leads, and adapters supplied with the product, or indicated by Tektronix to be
suitable for the product.
Observe all terminal ratings. To avoid fire or shock hazard, observe all rating and markings on the product. Consult the product
manual for further ratings information before making connections to the product. Do not exceed the Measurement Category
(CAT) rating and voltage or current rating of the lowest rated individual component of a product, probe, or accessory. Use caution
when using 1:1 test leads because the probe tip voltage is directly transmitted to the product.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Do not operate without covers. Do not operate this product with covers or panels removed, or with the case open. Hazardous
voltage exposure is possible.
Avoid exposed circuitry. Do not touch exposed connections and components when power is present.
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by qualified
service personnel.
Disable the product if it is damaged. Do not use the product if it is damaged or operates incorrectly. If in doubt about safety of the
product, turn it off and disconnect the power cord. Clearly mark the product to prevent its further operation.
5 Series MSO Specifications and Performance Verification iii
Important safety information
Before use, inspect voltage probes, test leads, and accessories for mechanical damage and replace when damaged. Do not use
probes or test leads if they are damaged, if there is exposed metal, or if a wear indicator shows.
Examine the exterior of the product before you use it. Look for cracks or missing pieces.
Use only specified replacement parts.
Do not operate in wet/damp conditions. Be aware that condensation may occur if a unit is moved from a cold to a warm
environment.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry. Remove the input signals before you clean the product.
Provide proper ventilation. Refer to the installation instructions in the manual for details on installing the product so it has
proper ventilation.
Slots and openings are provided for ventilation and should never be covered or otherwise obstructed. Do not push objects into
any of the openings.
Provide a safe working environment. Always place the product in a location convenient for viewing the display and indicators.
Avoid improper or prolonged use of keyboards, pointers, and button pads. Improper or prolonged keyboard or pointer use may
result in serious injury.
Be sure your work area meets applicable ergonomic standards. Consult with an ergonomics professional to avoid stress injuries.
Use care when lifting and carrying the product. This product is provided with a handle or handles for lifting and carrying.
WARNING. The product is heavy. To reduce the risk of personal injury or damage to the device get help when lifting or carrying
the product.
Use only the Tektronix rackmount hardware specified for this product.
Probes and test leads
Before connecting probes or test leads, connect the power cord from the power connector to a properly grounded power outlet.
Keep fingers behind the protective barrier, protective finger guard, or tactile indicator on the probes.
Remove all probes, test leads and accessories that are not in use.
Use only correct Measurement Category (CAT), voltage, temperature, altitude, and amperage rated probes, test leads, and
adapters for any measurement.
iv5 Series MSO Specifications and Performance Verification
Important safety information
Terms in the manual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Terms on the product
These terms may appear on the product:
■
DANGER indicates an injury hazard immediately accessible as you read the marking.
■
WARNING indicates an injury hazard not immediately accessible as you read the marking.
■
CAUTION indicates a hazard to property including the product.
Symbols on the product
When this symbol is marked on the product, be sure to consult the manual to find out the nature of the potential
hazards and any actions which have to be taken to avoid them. (This symbol may also be used to refer the user to
ratings in the manual.)
The following symbols may appear on the product:
5 Series MSO Specifications and Performance Verification v
Important safety information
vi5 Series MSO Specifications and Performance Verification
Specifications
This chapter contains specifications for the instrument. All specifications are guaranteed unless noted as "typical." Typical
specifications are provided for your convenience but are not guaranteed. Specifications that are marked with the symbol are
checked in this manual. All specifications apply to all models unless noted otherwise.
To meet specifications, these conditions must first be met:
■
The instrument must have been calibrated in an ambient temperature between 18 °C and 28 °C (64 °F and 82 °F).
■
The instrument must be operating within the environmental limits. (See Environmental specifications on page 29.)
■
The instrument must be powered from a source that meets the specifications. (See Power supply system on page 29.)
■
The instrument must have been operating continuously for at least 20 minutes within the specified operating temperature
range.
■
You must perform the Signal Path Compensation procedure after the warmup period. See the online help for instructions on
how to perform signal path compensation. If the ambient temperature changes more than 5 °C (9 °F) , repeat the
procedure.
Analog channel input and vertical specification
Number of input channels
MSO544 BNC
MSO566 BNC
MSO588 BNC
MSO58LP8 BNC
Input couplingDC, AC
Input resistance selection1 MΩ or 50 Ω
✓Input impedance 1 MΩ DC
coupled
Input capacitance 1 MΩ DC
coupled, typical
✓Input impedance 50 Ω, DC
coupled
Maximum input voltage, 1 MΩ300 V
Maximum input voltage, 50 Ohm5 V
1 MΩ ±1%
14.5 pF ±1.5 pF, 2 GHz model
13 pF ±1.5 pF, 1 GHz, 500 MHz, 350 MHz models
14 pF ±1.5 pF, MSO58LP
50 Ω ±1% (VSWR ≤1.5:1, typical)
at the BNC. Installation Category II
RMS
Derate at 20 dB/decade between 4.5 MHz and 45 MHz; derate 14 dB/decade between 45 MHz and
450 MHz. Above 450 MHz, 5.5 V
Maximum peak input voltage at the BNC: ±425 V
, with peaks ≤ ±20 V (DF ≤6.25%)
RMS
RMS
5 Series MSO Specifications and Performance Verification 1
Specifications Analog channel input and vertical specification (cont.)
DC balance✓ 0.1 div with DC-50 Ω oscilloscope input impedance (50 Ω BNC terminated)
✓ 0.2 div at 1 mV/div with DC-50 Ω oscilloscope input impedance (50 Ω BNC terminated)
0.4 div at 500 μV/div with DC-50 Ω oscilloscope input impedance (50 Ω BNC terminated)
✓ 0.2 div with DC-1 MΩ oscilloscope input impedance (50 Ω BNC terminated)
0.4 div at 500 µV/div with DC-1 MΩ scope input impedance (50 Ω BNC terminated)
NOTE. 500 µV/div is a 2X digital zoom of 1 mV/div. As such, it is guaranteed by testing the 1 mV/div setting.
Number of digitized bits8 bits at 6.25 GS/s
12 bits at 3.125 GS/s
13 bits at 1.25 GS/s
14 bits at 625 MS/s
15 bits at 312.5 MS/s
16 bits at 125 MS/s
Displayed vertically with 25 digitization levels (DL
1
) per division, (8-bits only) 10.24 divisions
dynamic range.
Sensitivity range, coarse
1 MΩ500 µV/div to 10 V/div in a 1-2-5 sequence
50 Ω500 µV/div to 1 V/div in a 1-2-5 sequence
Note: 500 μV/div is a 2X digital zoom of 1 mV/div.
Sensitivity range, fineAllows continuous adjustment from 1 mV/div to 10 V/div, 1 MΩ and from 1 mV/div to 1 V/div, 50 Ω
1 MΩ500 µV/div to 10 V/div
50 Ω500 µV/div to 1 V/div
Sensitivity resolution, fine≤1% of current setting
✓ DC gain accuracy
2 GHz model, Step Gain,
±1.2%, (±2.0% at 1 mV/div and 500 µV/div settings), de-rated at 0.100%/ °C above 30 °C
50 Ohm
2 GHz model, Step Gain,
±1.0%, (±2.0% at 1 mV/div and 500 µV/div settings), de-rated at 0.100%/ °C above 30 °C
1 Meg Ohm
1 GHz, 500 MHz, 350 MHz
±1.0%, (±2.0% at 1 mV/div and 500 µV/div settings), de-rated at 0.100%/ °C above 30 °C
models, MSO58LP, Step Gain
Variable gain±1.5%, derated at 0.100%/ °C above 30 °C.
NOTE. 500 µV/div is a 2X digital zoom of 1 mV/div. As such, it is guaranteed by testing the 1 mV/div setting.
1
DL is the abbreviation for digitization level. A DL is the smallest voltage level change that can be resolved by an 8-bit A-D Converter. This value is also known as an LSB
(least significant bit).
25 Series MSO Specifications and Performance Verification
Analog channel input and vertical specification (cont.) Specifications
Offset ranges, maximum
2 GHz modelsInput signal cannot exceed maximum input voltage for the 50 Ω input path.
5 Series MSO Specifications and Performance Verification 3
Specifications Analog channel input and vertical specification (cont.)
DC voltage measurement
accuracy, sample acquisition
mode, typical
Bandwidth selections20 MHz, 250 MHz, and the full bandwidth value for your model
Bandwidth selections20 MHz, 250 MHz, and 1 GHz
✓Analog bandwidth 50 Ω DC
coupled
2 GHz models
1 GHz models
Measurement TypeDC Accuracy (In Volts)
Any Sample±(DC Gain Accuracy * |reading - (offset -
position)| + Offset Accuracy + 0.15 div + 0.6 mV)
Delta Volts between any two samples acquired
with the same scope setup and ambient
conditions
Volts/Div SettingBandwidth
10 mV/div - 1 V/divDC - 2.00 GHz
5 mV/div - 9.98 mV/divDC - 1.50 GHz
2 mV/div - 4.98 mV/divDC - 350 MHz
1 mV/div - 1.99 mV/divDC - 175 MHz
500 µV/div - 995 µV/divDC - 175 MHz
Volts/Div SettingBandwidth
1 mV/div - 1 V/divDC - 1.00 GHz
500 µV/div - 995 µV/divDC - 250 MHz
±(DC Gain Accuracy * |reading| + 0.15 div +
1.2 mV)
500 MHz models
350 MHz models
Volts/Div SettingBandwidth
1 mV/div - 1 V/divDC - 500 MHz
500 µV/div - 995 µV/divDC - 250 MHz
Volts/Div SettingBandwidth
1 mV/div - 1 V/divDC - 350 MHz
500 µV/div - 995 µV/divDC - 250 MHz
45 Series MSO Specifications and Performance Verification
Analog channel input and vertical specification (cont.) Specifications
Analog bandwidth, 1 MΩ, typical
All model bandwidths except
350 MHz
350 MHz models
Volts/Div SettingBandwidth
1 mV/div - 10 V/divDC - 500 MHz
500 µV/div - 995 µV/divDC - 250 MHz
Volts/Div SettingBandwidth
1 mV/div - 10 V/divDC - 350 MHz
500 µV/div - 995 µV/divDC - 250 MHz
Analog bandwidth with TPP0500
and TPP1000 probe, typical
Lower frequency limit, AC coupled,
typical
Upper frequency limit, 250 MHz
bandwidth limited, typical
Upper frequency limit, 20 MHz
bandwidth limited, typical
Calculated rise time, typical
The limits are for ambient temperature of ≤30 °C and the bandwidth selection set to FULL. Reduce
the upper bandwidth frequency by 1% for each °C above 30 °C.
Standard1 kpoints to 125 Mpoints in single sample increments
Standard1 kpoints to 62.5 Mpoints in single sample increments
Option 5-RL-125M125 Mpoints
Seconds/Division range
Maximum triggered acquisition
rate, typical
Max HW CapabilityNumber of Channels
6.25 GS/s1-8
Model1 K10 K100 K1 M10 M62.5 M125 M
MSO5X Standard
62.5 M
MSO5X Option 5RL-125M
Analog or digital channels: single channel [Analog or Digital 8-bit channel] on screen,
measurements and math turned off. >40 wfm/sec
FastAcq Update Rate (analog only): >500 K/second with one channel active and >100 K/second
with all eight active.
200 ps 64 s
200 ps 64 s
200 ps 640 s
200 ps 640 s
200 ps - 1000 s
200 ps - 1000 s
FastAcq Update Rate (analog and analog/digital): >400 K/second with two channels active and
>100 K/second with all eight analog channels active.
Digital channel: >40/second with one channel (8-bits) active. There is no FastAcq for digital
channels, but they do not slow down FastAcq for active analog channels.
Aperture uncertainty≤ 0.450 ps + (1 * 10
-11
* Measurement Duration)
, for measurements having duration ≤ 100 ms
RMS
5 Series MSO Specifications and Performance Verification 13
Specifications Timebase system (cont.)
✓Timebase accuracy±2.5 x 10-6 over any ≥1 ms time interval
DescriptionSpecification
Factory Tolerance±5.0 x10-7. At calibration, 23 °C ambient, over any ≥1 ms
interval
Temperature stability±5.0 x10-7. Tested at operating temperatures
Crystal aging±1.5 x 10-6. Frequency tolerance change at 25 °C over a period
of 1 year
✓Delta-time measurement
accuracy
The formula to calculate delta-time measurement accuracy (DTA) for a given instrument setting
and input signal is as follows (assumes insignificant signal content above Nyquist frequency):
SR 1 = Slew Rate (1st Edge) around 1st point in measurement
SR 2 = Slew Rate (2nd Edge) around 2nd point in measurement
N = input-referred guaranteed noise limit (V
TBA = timebase accuracy or Reference Frequency Error
t p = delta-time measurement duration (sec)
(assume edge shape that results from Gaussian filter response)
2
DTApp(typical) = 10 ×
DTA
RMS
The term under the square root sign is the stability and is due to TIE (Time Interval Error). The
errors due to this term occur throughout a single-shot measurement. The second term is due to
both the absolute center-frequency accuracy and the center-frequency stability of the timebase and
varies between multiple single-shot measurements over the observation interval (the amount of
time from the first single-shot measurement to the final single-shot measurement).
N
=
(
SR
1
N
(
SR
1
2
N
+
)
(
SR
2
N
+
)
(
SR
2
+(0.450 ps +(1 × 10
)
)
RMS
2
+(0.450 ps +(1 × 10
)
2
-11
× t
2
))
p
-11
× t
p
+ TBA × t
2
+ TBA × t
))
p
p
NOTE. The formulas assume negligible errors due to measurement interpolation, and apply only
when the interpolated sample rate is 25 GS/s or higher.
145 Series MSO Specifications and Performance Verification
Specifications
Trigger system
Trigger bandwidth (edge, pulse
and logic), typical
Edge-type trigger sensitivity, DC
coupled, typical
2 GHz models, Edge2 GHz
2 GHz models, Pulse
and Logic
1 GHz models,
MSO58LP
500 MHz models500 MHz
350 MHz models350 MHz
1 GHz
1 GHz
5 Series MSO
PathRangeSpecification
1 MΩ path (all
models)
50 Ω path,
1 GHz,
500 MHz,
350 MHz
models
50 Ω path,
2 GHz models
LineFixed
0.5 mV/div to
0.99 mV/div
≥ 1 mV/divThe greater of 5 mV or 0.7 div from DC to lesser of 500 MHz or
0.5 mV/div to
0.99 mV/div
1 mV/div to
9.98 mV/div
≥ 10 mV/div< 1.0 division from DC to instrument bandwidth
5 mV from DC to instrument bandwidth
instrument BW, & 6 mV or 0.8 div from > 500 MHz to instrument
bandwidth
The greater of 5.6 mV or 0.7 div from DC to the lesser of
500 MHz or instrument BW, & 7 mV or 0.8 div from > 500 MHz to
instrument bandwidth
3.0 div from DC to instrument bandwidth
1.5 divisions from DC to instrument bandwidth
MSO58LP
PathRangeSpecification
1 MΩ path (all
models)
50 Ω pathThe greater of 5.6 mV or 0.7 div from DC to the lesser of
LineFixed
AUX Trigger in200 mVPP, DC to 250 MHz
5 Series MSO Specifications and Performance Verification 15
0.5 mV/div to
0.99 mV/div
≥ 1 mV/divThe greater of 5 mV or 0.7 div from DC to lesser of 500 MHz or
4.5 div from DC to instrument bandwidth
instrument BW, & 6 mV or 0.8 div from > 500 MHz to instrument
bandwidth
500 MHz or instrument BW, & 7 mV or 0.8 div from > 500 MHz to
instrument bandwidth
Specifications Trigger system (cont.)
Trigger jitter, typical≤ 5 ps
≤ 7 ps
≤ 40 ps
≤ 200 ps
≤ 220 ps
AUX In trigger skew between
instruments, typical (MSO58LP
only)
Edge-type trigger sensitivity, not
DC coupled, typical
±100 ps jitter on each instrument with 150 ps skew; ≤350 ps total between instruments.
Skew improves for sinusoidal input voltages ≥500 mV
Trigger CouplingTypical Sensitivity
NOISE REJ2.5 times the DC Coupled limits
HF REJ1.0 times the DC Coupled limits from DC to 50 kHz. Attenuates signals
LF REJ1.5 times the DC Coupled limits for frequencies above 50 kHz.
Lowest frequency for successful
45 Hz
operation of Set Level to 50%
function, typical
for sample mode and edge-type trigger
RMS
for edge-type trigger and FastAcq mode
RMS
for non edge-type trigger modes
RMS
for AUX trigger in, Sample acquisition mode, edge trigger (MSO58LP only)
RMS
for AUX trigger in, FastAcq acquisition mode, edge trigger (MSO58LP only)
RMS
above 50 kHz.
Attenuates signals below 50 kHz.
Logic-type, logic qualified trigger,
or events-delay sensitivities, DC
coupled, typical
Pulse-type runt trigger
sensitivities, typical
Pulse-type trigger width and glitch
sensitivities, typical
Logic-type triggering, minimum
logic or rearm time, typical
2.0 divisions, at vertical settings ≥5 mV/div.
2.0 division at vertical settings ≥5 mV/div.
2.0 divisions at vertical settings ≥5 mV/div.
Triggering typePulse widthRearm timeTime skew needed for
Logic160 ps + t
Time qualified logic320 ps + t
t
is rise time of the instrument.
rise
rise
rise
160 ps + t
320 ps + t
rise
rise
100% and no
triggering
>360 ps / <150 ps
>360 ps / <150 ps
1
1
For Logic, time between channels refers to the length of time a logic state derived from more than one channel must exist to be recognized. For Events, the time is the
minimum time between a main and delayed event that will be recognized if more than one channel is used.
165 Series MSO Specifications and Performance Verification
Trigger system (cont.) Specifications
Minimum clock pulse widths for
setup/hold time violation trigger,
typical
Setup/hold violation trigger, setup
and hold time ranges, typical
Pulse type trigger, minimum pulse,
rearm time, transition time
Minimum pulsewidth, clock active
320 ps + t
t
rise
rise
is rise time of the instrument.
2
Minimum pulsewidth, clock inactive
320 ps +t
rise
3
FeatureMinMax
Setup Time0 ns20 s
Hold Time0 ns20 s
Setup + Hold Time320 ps22 s
Input coupling on clock and data channels must be the same.
For Setup Time, positive numbers mean a data transition before the clock.
For Hold Time, positive numbers mean a data transition after the clock edge.
Setup + Hold Time is the algebraic sum of the Setup Time and the Hold Time programmed by the
user.
Pulse classMinimum pulse widthMinimum rearm time
Runt160 ps + t
Time-Qualified Runt160 ps + t
Width160 ps + t
Slew Rate (minimum transition
160 ps + t
rise
rise
rise
rise
160 ps + t
160 ps + t
160 ps + t
160 ps + t
rise
rise
rise
rise
time)
For trigger class width, pulse width refers to the width of the pulse being measured. Rearm time
refers to the time between pulses.
For trigger class runt, pulse width refers to the width of the pulse being measured. Rearm time
refers to the time between pulses.
For trigger class slew rate, pulse width refers to the delta time being measured. Rearm time refers
to the time it takes the signal to cross the two trigger thresholds again.
t
is rise time of the instrument.
rise
Active pulsewidth is the width of the clock pulse from its active edge (as defined in the Clock Edge
menu item) to its inactive edge
Inactive pulsewidth is the width of the pulse from its inactive edge to its active edge.
Transition time trigger, delta time
160 ps to 20 s.
range
Time range for glitch, pulse width,
160 ps to 20 s.
timeout, time-qualified runt, or
time-qualified window triggering
2
Active pulsewidth is the width of the clock pulse from its active edge (as defined in the Clock Edge menu item) to its inactive edge.
3
Inactive pulsewidth is the width of the pulse from its inactive edge to its active edge.
5 Series MSO Specifications and Performance Verification 17
Specifications Trigger system (cont.)
Time accuracy for pulse, glitch,
timeout, or width triggering
Time RangeAccuracy
1 ns to 500 ns±(160 ps +Time Base Error * Setting).
520 ns to 1 s±(160 ps +Time Base Error * Setting).
B trigger after events, minimum
pulse width and maximum event
frequency, typical
B trigger, minimum time between
arm and trigger, typical
Minimum pulse width: 160 ps + t
Maximum event frequency: Instrument bandwidth.
t
is rise time of the instrument.
rise
320 ps
For trigger after time, this is the time between the end of the time period and the B trigger event.
For trigger after events, this is the time between the last A trigger event and the first B trigger event.
B trigger after time, time range160 ps to 20 seconds
B trigger after events, event range1 to 65,471
Trigger level ranges
SourceRange
Any Channel±5 divs from center of screen
LineFixed at about 50% of line voltage
rise
SourceRange
Any Channel±5 divs from center of screen
Aux In Trigger±5 V
LineFixed at about 50% of line voltage
This specification applies to logic and pulse thresholds.
Trigger level accuracy, DC
For signals having rise and fall times ≥10 ns:
coupled, typical
SourceRange
Any Input Channel±0.20 div
LineN/A
Trigger holdoff range0 ns to 10 seconds
185 Series MSO Specifications and Performance Verification
Specifications
Serial Trigger specifications
Maximum serial trigger bits128 bits
Optional serial bus interface
triggering
I2CAddress Triggering: 7 & 10 bits of user-specified addresses supported
Data Trigger: 1 - 5 Bytes of user-specified data
Trigger on: Start, Repeated Start, Stop, Missing Ack, Data, Address, or Address & Data
Maximum Data Rate: 10 Mb/s
SPIData Trigger: 1 - 16 Bytes of user-specified data
Trigger on: SS Active, Data
Maximum Data Rate: 20 Mb/s
CANData Trigger: 1 - 8 Bytes of user-specified data, including qualifiers of equal to (=), not equal to
(≠), less than (<), greater than (>), less than or equal to (<=), greater than or equal to (>=)
Trigger on: Start of Frame, Type of Frame, Identifier, Data, Identifier & Data, End of Frame,
Missing Ack, or Bit Stuffing Errors
Frame Type: Data, Remote, Error, Overload
Identifier: Standard (11 bit) and Extended (29 bit) identifiers
Maximum Data Rate: 1 Mb/s
LINIdentifier Trigger: 6 bits of user-specified data, equal to (=)
Data Trigger: 1 - 8 Bytes of user-specified data, including qualifiers of equal to (=), not equal to
(≠), less than (<), greater than (>), less than or equal to (<=), greater than or equal to (>=), inside
range, outside range
Error Trigger: Sync, Identifier Parity, Checksum
Trigger on: Sync, Identifier, Data, Identifier & Data, Wakeup Frame, Sleep Frame, or Error
Maximum Data Rate: 100 kb/s
5 Series MSO Specifications and Performance Verification 19
Specifications Serial Trigger specifications (cont.)
Frame ID Trigger: 11 bits of user-specified data, including qualifiers of equal to (=), not equal to
(≠), less than (<), greater than (>), less than or equal to (<=), greater than or equal to (>=)
Cycle Count Trigger: 6 bits of user-specified data , including qualifiers of equal to (=), not equal to
(≠), less than (<), greater than (>), less than or equal to (<=), greater than or equal to (>=)
Header Fields Trigger: 40 bits of user-specified data comprising Indicator Bits, Identifier, Payload
Length, Header CRC, and Cycle Count, equal to (=)
Data Trigger: 1 - 16 Bytes of user-specified data, with 0 to 253, or "don't care" bytes of data offset,
including qualifiers of equal to (=), not equal to (≠), less than (<), greater than (>), less than or
equal to (<=), greater than or equal to (>=), Inside Range, Outside Range
End Of Frame: User-chosen types Static, Dynamic (DTS), and All
Trigger on: Start of Frame, Frame ID, Indicator Bits, Cycle Count, Header Fields, Data, Identifier &
Data, End of Frame, or Error
Maximum Data Rate: 40 Mb/s
Audio (I2S)Data Trigger: 32 bits of user-specified data in a left word, right word, or either, including qualifiers
of equal to (=), not equal to (≠), less than (<), greater than (>), less than or equal to (<=), greater
than or equal to (>=), inside range, outside range
Trigger on: Word Select, Data
Maximum Data Rate: 12.5 Mb/s
Left Justified (LJ)
Data Trigger: 32 bits of user-specified data in a left word, right word, or either, including qualifiers
of equal to (=), not equal to (≠), less than (<), greater than (>), less than or equal to (<=), greater
than or equal to (>=), inside range, outside range
Trigger on: Word Select, Data
Maximum Data Rate: 12.5 Mb/s
Right Justified (RJ)
Data Trigger: 32 bits of user-specified data in a left word, right word, or either, including qualifiers
of equal to (=), not equal to (≠), less than (<), greater than (>), less than or equal to (<=), greater
than or equal to (>=), inside range, outside range
Trigger on: Word Select, Data
Maximum Data Rate: 12.5 Mb/s
TDM
Data Trigger: 32 bits of user-specified data in a channel 1-64, including qualifiers of equal to (=),
not equal to (≠), less than (<), greater than (>), less than or equal to (<=), greater than or equal to
(>=), inside range, outside range
Trigger on: Frame Sync, Data
Maximum Data Rate: 25 Mb/s
205 Series MSO Specifications and Performance Verification
Serial Trigger specifications (cont.) Specifications
RS232Bit Rate: 50 bps - 10 Mbps
Data Bits: 7, 8, or 9
Parity: None, Odd, or Even
Trigger on: Start, End of Packet, Data, Parity Error
MIL-STD-1553Bit Rate: 1 Mb/s
Trigger on:
Sync
Word Type (Command, Status, Data)
Command Word (set RT Address (=, ≠, <, >, ≤, ≥, inside range, outside range), T/R, Sub-address/
Mode, Data Word Count/Mode Code, and Parity individually)
Status Word (set RT Address ((=, ≠, <, >, ≤, ≥, inside range, outside range), Message Error,
Instrumentation, Service Request Bit, Broadcast Command Received, Busy, Subsystem Flag,
Dynamic Bus Control Acceptance (DBCA), Terminal Flag, and Parity individually)
Idle Time (minimum time selectable from 4 μs to 100 μs; maximum time selectable from 12 μs to
100 μs; trigger on < minimum, > maximum, inside range, outside range)
For MIL-STD-1553, Trigger selection of Command Word will trigger on Command and ambiguous
Command/Status words. Trigger selection of Status Word will trigger on Status and ambiguous
Command/Status words
Trigger On: Start of Frame, MAC Address, MAC Length/Type, IP Header, TCP Header, Client
Data, End of Packet, Idle, FCS (CRC) Error, MAC Q-Tag control Information
5 Series MSO Specifications and Performance Verification 21
Specifications
Digital acquisition system
Digital channel maximum sample
rate
Transition detect (digital peak
detect)
Digital-To-Analog trigger skew1 ns
Digital to digital skew320 ps from bit 0 of any TekVPI channel to bit 0 of any TekVPI channel.
Digital skew within a FlexChannel160 ps within any TekVPI channel
6.25 GS/s
Displayed data at sample rates less than 6.25 GS/s (decimated data), that contains multiple
transitions between sample points will be displayed with a bright white colored edge.
De-rated at 0.100%/°C of |reading - offset - position| above 30 °C
RMS
+DC, AC
RMS
Signal ± 5 divisions from screen center
AC:± 2% (40 Hz to 1 kHz) with no harmonic content outside 40 Hz to 1 kHz range
AC, typical: ± 2% (20 Hz to 10 kHz)
For AC measurements, the input channel vertical settings must allow the VPP input signal to cover
between 4 and 10 divisions and must be fully visible on the screen